Teledyne Oxygen Equipment 3000ZA 3X User Manual

OPERATING INSTRUCTIONS FOR  
MODEL 3000ZA-3X  
Trace Oxygen Analyzer  
P/N M77579  
11/29/11  
DANGER  
Toxic gases and or flammable liquids may be present in this monitoring system.  
Personal protective equipment may be required when servicing this instrument.  
Hazardous voltages exist on certain components internally which may persist  
for a time even after the power is turned off and disconnected.  
Only authorized personnel should conduct maintenance and/or servicing.  
Before conducting any maintenance or servicing, consult with authorized  
supervisor/manager.  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Specific Model Information  
The instrument for which this manual was supplied may  
incorporate one or more options not supplied in the standard instrument.  
Commonly available options are listed below, with check boxes. Any  
that are incorporated in the instrument for which this manual is supplied  
are indicated by a check mark in the box.  
Instrument Serial Number: _______________________  
Options Included in the Instrument with the Above Serial Number:  
3000ZA-C:  
In addition to all the standard features,  
this model also has separate ports for  
zero and span gases, and built-in  
control valves. The internal valves are  
entirely under the control of the  
3000ZA electronics to automatically  
switch between gases in  
synchronization with the analyxer’s  
operation.  
3000ZA-Rack Mnt: The 19" Relay Rack Mount units are  
available with either one or two 3000  
series analyzers installed in a standard  
19" panel and ready to mount in a  
standard rack.  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3000ZA-3X  
Important Notice  
Model 3000ZA complies with all of the requirements of the  
Commonwealth of Europe (CE) for Radio Frequency Interference,  
Electromagnetic Interference (RFI/EMI), and Low Voltage Directive  
(LVD).  
The following International Symbols are used throughout the  
Instruction Manual. These symbols are visual indicators of important  
and immediate warnings and when you must exercise CAUTION while  
operating the instrument. See also the Safety Information on the next  
page.  
STAND-BY: Instrument is on Stand-by, but circuit is active  
GROUND: Protective Earth  
CAUTION: The operator needs to refer to the manual for  
further information. Failure to do so may compromise the  
safe operation of the equipment.  
CAUTION: Risk of Electrical Shock  
Teledyne Analytical Instruments  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Safety Messages  
Your safety and the safety of others is very important. We have  
provided many important safety messages in this manual. Please read  
these messages carefully.  
A safety message alerts you to potential hazards that could hurt you  
or others. Each safety message is associated with a safety alert symbol.  
These symbols are found in the manual and inside the instrument. The  
definition of these symbols is described below:  
GENERAL WARNING/CAUTION: Refer to the  
instructions for details on the specific danger. These cautions  
warn of specific procedures which if not followed could  
cause bodily Injury and/or damage the instrument.  
CAUTION: HOT SURFACE WARNING: This warning is  
specific to heated components within the instrument. Failure  
to heed the warning could result in serious burns to skin and  
underlying tissue.  
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous  
voltages appear within this instrument. This warning is  
specific to an electrical hazard existing at or nearby the  
component or procedure under discussion. Failure to heed  
this warning could result in injury and/or death from  
electrocution.  
Technician Symbol: All operations marked with this  
symbol are to be performed by qualified maintenance  
personnel only.  
NOTE: Additional information and comments regarding a  
specific component or procedure are highlighted in the form  
of a note.  
No  
Symbol  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
3000ZA-3X  
CAUTION:  
THE ANALYZER SHOULD ONLY BE USED FOR THE  
PURPOSE AND IN THE MANNER DESCRIBED IN  
THIS MANUAL.  
IF YOU USE THE ANALYZER IN A MANNER OTHER  
THAN THAT FOR WHICH IT WAS INTENDED,  
UNPREDICTABLE BEHAVIOR COULD RESULT  
POSSIBLY ACCOMPANIED WITH HAZARDOUS  
CONSEQUENCES.  
This manual provides information designed to guide you through  
the installation, calibration and operation of your new analyzer. Please  
read this manual and keep it available.  
Occasionally, some instruments are customized for a particular  
application or features and/or options added per customer requests.  
Please check the front of this manual for any additional information in  
the form of an Addendum which discusses specific information,  
procedures, cautions and warnings that may be peculiar to your  
instrument.  
Manuals do get lost. Additional manuals can be obtained from  
Teledyne Analytical Instruments at the address given in the Appendix.  
Some of our manuals are available in electronic form via the internet.  
Teledyne Analytical Instruments  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
This is a general purpose instrument designed for use in a non-hazardous  
area. It is the customer's responsibility to ensure safety especially when  
combustible gases are being analyzed since the potential of gas leaks  
always exist.  
The customer should ensure that the principles of operation of this  
equipment is well understood by the user. Misuse of this product in any  
manner, tampering with its components, or unauthorized substitution of  
any component may adversely affect the safety of this instrument.  
Since the use of this instrument is beyond the control of Teledyne  
Analytical Instruments, no responsibility by Teledyne Analytical  
Instruments, its affiliates, and agents for damage or injury from misuse  
or neglect of this equipment is implied or assumed.  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3000ZA-3X  
Table of Contents  
Safety Messages.............................................................v  
Introduction.....................................................................1  
1.1 Overview  
1
1
1
2
3
5
1.2 Typical Applications  
1.3 Main Features of the Analyzer  
1.4 Model Designations  
1.5 Front Panel (Operator Interface)  
1.6 Rear Panel (Equipment Interface)  
Operational Theory.........................................................7  
2.1 Introduction  
2.2 ZrO2 Sensor  
7
7
7
8
9
9
2.2.1 Principles of Operation  
2.3 Sample System  
2.4 Zirconium Oxide Sensor Application Notes  
2.5 Electronics and Signal Processing  
Installation.....................................................................13  
3.1 Unpacking the Analyzer  
3.2 Mounting the Analyzer  
3.3 Rear Panel Connections  
3.3.1 Gas Connections  
13  
13  
15  
15  
17  
17  
18  
24  
26  
3.3.2 Electrical Connections  
3.3.2.1 Primary Input Power  
3.3.2.2 50-Pin Equipment Interface Connector  
3.3.2.3 RS-232 Port  
3.4 Installing the Zirconium Oxide Sensors  
Teledyne Analytical Instruments  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
3.5 Testing the System  
26  
Operation.......................................................................27  
4.1 Introduction  
27  
28  
28  
28  
29  
29  
29  
30  
30  
30  
30  
32  
33  
33  
34  
34  
34  
36  
36  
36  
37  
37  
37  
38  
38  
39  
39  
39  
40  
40  
4.2 Using the Data Entry and Function Buttons  
4.2.1 Style Conventions  
4.2.2 Keys Description  
4.2.2.1 Arrow Keys  
4.2.2.2 ENTER  
4.2.2.3 ESCAPE  
4.2.2.4 Special Function Keys  
4.3 Menu/Screen Structure  
4.3.1 Analyze Screen  
4.3.2 System Menu  
4.3.3 Password routine  
4.3.4 Logout Routine  
4.3.5 Model Routine  
4.3.6 Self-Test Routine  
4.3.7 Filter Routine  
4.3.8 Analog-Out Adjust Routine  
4.3.9 Auto-Calibration Timer  
4.3.10 Change Stream  
4.4 Span Menu  
4.4.1 Span Sensor  
4.4.2 Span Value  
4.4.3 Span Timer  
4.4.4 Span Finish  
4.4.5 Span Begin  
4.5 Zero Menu  
4.5.1 Zero Finish  
4.5.2 Zero Value  
4.5.3 Zero Begin  
4.5.4 Zero Timer  
Teledyne Analytical Instruments  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3000ZA-3X  
4.6 Alarms Menu  
4.6.1 Alarm Active  
41  
41  
41  
42  
42  
42  
43  
43  
44  
44  
4.6.2 Alarm Polarity  
4.6.3 Alarm Failsafe  
4.6.4 Alarm Latching  
4.6.5 Alarm Setpoint  
4.7 Range Menu  
4.7.1 Range Select  
4.7.2 Manual Range Select  
4.7.3 Range Setpoints  
Maintenance..................................................................45  
5.1 Routine Maintenance  
5.2 Sensor Replacement  
5.2.1 Removing the Sensor  
5.3 Fuse Replacement  
5.4 System Self Diagnostic Test  
5.5 Major Internal Components  
5.6 Cleaning  
45  
45  
45  
47  
48  
49  
50  
50  
5.7 Troubleshooting  
Appendix .......................................................................53  
A-1 Specifications  
53  
55  
56  
56  
57  
A-2 Recommended 2-Year Spare Parts List  
A-3 Drawing List  
A-4 19-inch Relay Rack Panel Mount  
A.5 Application Notes  
Teledyne Analytical Instruments  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
List of Figures  
Figure 1-1: Model 3000ZA Front Panel  
Figure 1-2: Model 3000ZA Rear Panel  
Figure 2-1: Model 3000ZA Sample System  
Figure 2-2: Flow Diagram with Options  
Figure 2-3: Component Location  
3
5
8
9
11  
12  
14  
14  
15  
18  
23  
24  
31  
47  
47  
48  
49  
56  
Figure 2-4: Electronic Block Diagram  
Figure 3-1: Front Panel of the Model 3000TA  
Figure 3-2: Required Front Door Clearance  
Figure 3-3: Rear Panel of the Model 3000ZA  
Figure 3-4: Equipment Interface Connector Pin Arrangement  
Figure 3-5: Remote Probe Connections  
Figure 3-6: FET Series Resistance  
Figure 4-1: Hierarchy of Available Functions  
Figure 5-1: Sensor Block with Middle Sensor Installed  
Figure 5-2: Removing Fuse Block from Housing  
Figure 5-3: Installing Fuses  
Figure 5-4: Rear Panel Removal  
Figure A-1: Single and Dual 19" Rack Mounts  
Teledyne Analytical Instruments  
xi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3000ZA-3X  
List of Tables  
Table 3-1: Analog Output Connections  
Table 3-2: Alarm Relay Contact Pins  
Table 3-3: Remote Calibration Connections  
Table 3-4: Range ID Relay Connections  
Table 3-5: Commands via RS-232 Input  
Table 3-6: Required RS-232 Options  
Table 5-1: Self-Test Result Display  
Table 5-2: Troubleshooting  
19  
20  
21  
22  
25  
25  
48  
50  
Teledyne Analytical Instruments  
xii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Introduction  
Introduction  
1.1 Overview  
The Teledyne Analytical Instruments Model 3000ZA Trace  
Oxygen Analyzer is a versatile microprocessor-based instrument for  
detecting oxygen at the parts-per-million (ppm) level in a variety of  
gases. This manual covers the Model 3000ZA General Purpose flush-  
panel and/or rack-mount units with CE mark. These units are for indoor  
use in a non-hazardous environment.  
1.2 Typical Applications  
A few typical applications of the Model 3000ZA are:  
Monitoring inert gas blanketing  
Air separation and liquefaction  
Chemical reaction monitoring  
Semiconductor manufacturing  
Petrochemical process control  
Quality assurance  
Gas analysis certification.  
1.3 Main Features of the Analyzer  
The Model 3000ZA Trace Oxygen Analyzer is sophisticated yet  
simple to use. The main features of the analyzer include:  
A 2-line alphanumeric display screen, driven by  
microprocessor electronics that continuously prompts  
and informs the operator.  
High resolution, accurate readings of oxygen content for  
low ppm levels. Large, bright, meter readout.  
Teledyne Analytical Instruments  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
3000ZA-3X  
Three Zirconium oxide sensors to provide higher range  
of analysis flexibility. The analyzer can cover ranges as  
low as 0-10 ppm or as high as 0-95%.  
No reference gas required  
Versatile analysis over a wide range of applications.  
Microprocessor based electronics: 8-bit CMOS  
microprocessor with 32 kB RAM and 128 kB ROM.  
Three user definable output ranges (from 0-1 ppm  
through 0-95%) allow best match to users process and  
equipment.  
Auto ranging allows analyzer to automatically select the  
proper preset range for a given measurement. Manual  
override allows the user to lock onto a specific range of  
interest.  
Two adjustable concentration alarms and a system  
failure alarm.  
Extensive self-diagnostic testing, at startup and on  
demand, with continuous power-supply monitoring.  
CE compliance  
RS-232 serial digital port for use with a computer or  
other digital communication device.  
Four analog outputs: two for measurement (0–1 VDC  
and isolated 4–20 mA DC) and two for range  
identification.  
Convenient and versatile, steel, flush-panel or rack-  
mountable case with slide-out electronics drawer.  
1.4 Model Designations  
3000ZA:  
Standard model for sample under pressure  
3000ZA-C: In addition to all the standard features, this  
model also has separate ports for zero and  
span gases, and built-in control valves. The  
internal valves are entirely under the control  
of the 3000ZA electronics to automatically  
Teledyne Analytical Instruments  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Introduction  
switch between gases in synchronization  
with the analyzer’s operation.  
3000ZA-M:  
3000ZA-V:  
This model has current output signals (4-20-  
mA) for ppm range and range ID, in  
addition to voltage outputs.  
Gas flow through the cell block in this  
model is driven by vacuum downstream  
from the cell block instead of upstream  
pressure. The internal restrictor is located  
downstream from the cell block to support  
this configuration. All other standard  
features are present in this model.  
1.5 Front Panel (Operator Interface)  
The standard 3000ZA is housed in a rugged metal case with all  
controls and displays accessible from the front panel. See Figure 1-1.  
The front panel has thirteen buttons for operating the analyzer, a digital  
meter, an alphanumeric display, and a window for viewing the sample  
flowmeter.  
Figure 1-1: Model 3000ZA Front Panel  
Teledyne Analytical Instruments  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
3000ZA-3X  
Function Keys:  
Six touch-sensitive membrane switches are used to change the specific  
function performed by the analyzer:  
Analyze Perform analysis for oxygen content of a sample  
gas.  
System Perform system-related tasks (described in detail  
in chapter 4, Operation.).  
Span  
Zero  
Span calibrate the analyzer.  
Zero calibrate the analyzer.  
Alarms Set the alarm setpoints and attributes.  
Range  
Set up the 3 user definable ranges for the  
instrument.  
Data Entry Keys:  
Six touch-sensitive membrane switches are used to input data to the  
instrument via the alphanumeric VFD display:  
Left & Right Arrows Select between functions currently  
displayed on the VFD screen.  
Up & Down Arrows Increment or decrement values of  
functions currently displayed.  
Enter Advances VFD display to the next screen in a series  
or returns to the Analyze screen if none remain.  
Escape Backs VFD display to the previous screen in a series  
or returns to the Analyze screen if none remain.  
Digital Meter Display:  
The meter display is a Light Emitting Diode (LED) device that produces  
large, bright, 7-segment numbers that are legible in any lighting. It produces  
a continuous readout from 0-250 ppm. It is accurate across all analysis  
ranges without the discontinuity inherent in analog range switching.  
Alphanumeric Interface Screen:  
The VFD screen is an easy-to-use interface from operator to analyzer. It  
displays values, options, and messages that give the operator immediate  
feedback.  
Flowmeter:  
Monitors the flow of gas past the sensor. Readout is 0.2 to 2.4 standard  
liters per minute (SLPM).  
Teledyne Analytical Instruments  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Introduction  
Standby Button  
:
The Standby turns off the display and outputs but circuitry is still  
operating.  
CAUTION:  
THE POWER CABLE MUST BE UNPLUGGED TO  
FULLY DISCONNECT POWER FROM THE  
INSTRUMENT. WHEN CHASSIS IS EXPOSED OR  
WHEN ACCESS DOOR IS OPEN AND POWER CABLE  
IS CONNECTED, USE EXTRA CARE TO AVOID  
CONTACT WITH LIVE ELECTRICAL CIRCUITS.  
Access Door:  
For access to the ZrO sensors, the front panel swings open when the  
2
latch in the upper right corner of the panel is pressed all the way in with  
a narrow gauge tool. Accessing the main circuit board requires  
unfastening rear panel screws and sliding the unit out of the case.  
1.6 Rear Panel (Equipment Interface)  
The rear panel, shown in Figure 1-2, contains the gas and electrical  
connectors for external inlets and outlets. Some of those depicted are  
optional and may not appear on your instrument. The connectors are  
described briefly here and in detail in Chapter 3 Installation.  
Figure 1-2: Model 3000ZA Rear Panel  
Teledyne Analytical Instruments  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
3000ZA-3X  
Power Connection Universal AC power source.  
Gas Inlet and Outlet One inlet (must be externally valved)  
and one exhaust out. Three inlets  
when option “C” is ordered.  
9-Pin RS-232 Port  
Serial digital concentration signal  
output and control input.  
50-Pin Equipment Interface Port  
Analog Outputs: 0–1 VDC oxygen concentration plus  
0-1 VDC range ID  
Alarm Connections: 2 concentration alarms and 1  
system alarm.  
Remote Valve: Used in the 3000ZA for controlling  
external solenoid valves only.  
Remote Span/Zero: Digital inputs allow external  
control of analyzer calibration.  
Calibration Contact: To notify external equipment that  
instrument is being calibrated and readings are not  
monitoring sample.  
Range ID Contacts: Four separate, dedicated, range  
relay contacts. Low, Medium, High, Cal.  
Network I/O: Serial digital communications for local  
network access. For future expansion. Not currently  
implemented.  
Optional  
Calibration Gas Ports: Separate fittings for zero, span, and  
sample gas input, internal valves  
for automatic gas switching.  
Current Signal Output: Additional isolated 4–20 mA  
DC plus 4–20 mA DC range ID.  
Note: If you require highly accurate Auto-Cal timing, use external  
Auto-Cal control where possible. The internal clock in the  
Model 3000ZA is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
Teledyne Analytical Instruments  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operational Theory  
Operational Theory  
2.1 Introduction  
The analyzer is composed of three subsystems:  
1. Three Zirconium Oxide (ZrO2) Sensors. Each one has a  
specific range of analysis: Low range sensor of 0-200  
ppm, medium range sensor of 0-2 %, and high range  
sensor of 0-95%.  
2. Sample System  
3. Electronic Signal Processing, Display and Control  
The sample system is designed to accept the sample gas and  
transport it through the analyzer without contaminating or altering the  
sample prior to analysis. The ZrO2 sensor is a device that translates the  
amount of oxygen present in the sample into an electrical current. The  
electronic signal processing, display and control subsystem simplifies  
operation of the analyzer and accurately processes the sampled data. The  
microprocessor controls all signal processing, input/output and display  
functions for the analyzer.  
2.2 ZrO2 Sensor  
2.2.1 Principles of Operation  
Teledyne’s zirconium oxide sensor is a miniature solid state sensor.  
It uses a stabilized zirconia disc as an electrolyte with a sensing  
electrode (the cathode) and a counter electrode (the anode) on each side  
of the disc. The zirconia is heated to a temperature of about 500°C via a  
built-in heater. By applying a suitable reference voltage (approximately  
0.75 volt) across the cathode and anode, the heated zirconia acts as an  
electrolyte which is capable of moving oxygen ions within its crystalline  
structure. This allows for the reduction of oxygen molecules reaching  
the cathode and the generation of a current equivalent to the parts per  
million oxygen in the sample gas.  
Teledyne Analytical Instruments  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
3000ZA-3X  
One of the three sensors has been modified (0-200 ppm range) to  
sense low levels of oxygen—less than 1 ppm and generate a current  
proportional to the concentration. The sensor is designed for long term  
stability and an operating life of 5 years.  
2.3 Sample System  
The sample system delivers gases to the ZrO2 sensors, located on  
the cell block, from the analyzer rear panel inlet. Depending on the  
mode of operation either sample or calibration gas is delivered.  
The Model 3000ZA sample system is designed and fabricated to  
ensure that the oxygen concentration of the gas is not altered as it travels  
through the sample system.  
The sample system for the standard instrument incorporates  
compression tube fittings for sample inlet and 1/4” outlet tube  
connections at the rear panel. The sample or calibration gas that flows  
through the system is monitored by a flowmeter downstream from the  
cell. Figure 2-1 shows the piping layout for the standard model.  
Figure 2-2 shows the flow diagram for sampling systems used for  
the three configurations offered as options for this instrument. In the  
standard instrument, calibration gases can be connected directly to the  
Sample In port by teeing to the port with appropriate valves.  
Figure 2-1: Model 3000ZA-3X Sample System  
Teledyne Analytical Instruments  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operational Theory  
Figure 2-2: Flow Diagram with Options  
2.4 Zirconium Oxide Sensor Application Notes  
The following sample gases require conditioning prior to entering  
the sensors:  
Flammable gases such as methane, alcohol and carbon  
monoxide may cause a measurement error. They must be  
filtered from the sample gas before entering the cell.  
Gases containing halogen atoms (F, Cl, Br, etc.) such as  
Freon, must be filtered before entering the cell. The sensor  
will be damaged by the decomposition of Freon.  
Gases containing SOx, H2S, silicone vapor and adhesives  
must not enter the sensor. These components adversely affect  
the performance of the sensor.  
Dust and oil should be filtered from the sample gas.  
Water vapor in contact with the sensor will damage the  
sensor. Use a moisture trap or filter to remove any water  
vapor in the sample.  
2.5 Electronics and Signal Processing  
The Model 3000ZA Trace Oxygen Analyzer uses an 8031  
microcontroller with 32 kB of RAM and 128 kB of ROM to control all  
signal processing, input/output, and display functions for the analyzer.  
Teledyne Analytical Instruments  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
3000ZA-3X  
System power is supplied from a universal power supply module designed  
to be compatible with any international power source. Figure 2-3 shows  
the location of the power supply and the main electronic PC boards.  
The signal processing electronics including the microprocessor,  
analog to digital, and digital to analog converters are located on the  
motherboard at the bottom of the case. The preamplifier board is  
mounted on top of the motherboard as shown in the figure. These boards  
are accessible after removing the back panel. Figure 2-4 is a block  
diagram of the analyzer electronics.  
In the presence of oxygen the zirconia sensor generates a current. A  
current to voltage amplifier converts this current to a voltage, which is  
amplified in the second stage amplifier. The output from the second  
stage amplifier is sent to an 18 bit analog to digital converter controlled  
by the microprocessor. In the block diagram only one sensor is shown,  
but in reality there are three separate amplifier circuits for each sensor.  
The output of each sensor is then multiplexed to the analog to digital  
converter.  
The high operating temperature of the ZrO2 sensor is required for  
proper operation of the sensor. The amplifier PCB contains a circuit that  
drives the heater embedded in the sensor. The voltage across the heater  
rises slowly to avoid damaging the heater/sensor assembly. After a few  
minutes, the sensor will be at the proper operating temperature.  
The digital concentration signal along with input from the control  
panel is processed by the microprocessor, and appropriate control  
signals are directed to the display, alarms and communications port. The  
same digital information is also sent to a 12 bit digital to analog  
converter that produces the 4-20 mA dc and the 0-1 VDC analog  
concentration signal outputs, and the analog range ID outputs.  
Signals from the power supply are also monitored, and through the  
microprocessor, the system failure alarm is activated if a malfunction is  
detected.  
Teledyne Analytical Instruments  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operational Theory  
Figure 2-3: Component Location  
Teledyne Analytical Instruments  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operational Theory  
3000ZA-3X  
Figure 2-4: Electronic Block Diagram  
Teledyne Analytical Instruments  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
Installation  
Installation of the Model 3000ZA Analyzer includes:  
1. Unpacking  
2. Mounting  
3. Gas connections  
4. Electrical connections  
5. Installing the sensor  
6. Testing the system.  
3.1 Unpacking the Analyzer  
The analyzer is shipped with all the materials you need to install  
and prepare the system for operation. Carefully unpack the analyzer and  
inspect it for damage. Immediately report any damage to the shipping  
agent.  
3.2 Mounting the Analyzer  
The Model 3000ZA is for indoor use in a general purpose area. It is  
NOT for hazardous environments of any type.  
The standard model is designed for flush panel mounting. Figure 3-1  
is an illustration of the 3000TA standard front panel (the 3000ZA is  
similar) and mounting bezel. There are four mounting holes—one in each  
corner of the rigid frame. Drawings at the back of this manual illustrate  
the dimensions and mounting hole spacing for this unit.  
On special order, a 19" rack-mounting panel can be provided. For  
rack mounting, one or two 3000 series analyzers can be flush-panel  
mounted on the rack panel. See Appendix for dimensions of the  
mounting panel.  
Teledyne Analytical Instruments  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
Figure 3-1: Front Panel of the Model 3000TA (Model 3000ZA similar)  
All operator controls, except the flow control needle valve, are  
mounted on the control panel, which is hinged on the left edge and  
doubles as the door that provides access to the sensor and cell block  
inside the instrument. The door latch is spring loaded and will swing  
open when the button in the center of the latch (upper right corner) is  
pressed all the way in with a narrow gauge tool (less than 0.18 inch  
wide). Allow clearance for the door to open in a 90-degree arc of radius  
7.125 inches. See Figure 3-2.  
Figure 3-2: Required Front Door Clearance  
Teledyne Analytical Instruments  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
As a precaution, open the cell block to check that all sensors are  
firmly mounted or have not come loose during shipping.  
3.3 Rear Panel Connections  
Figure 3-3 shows the Model 3000ZA rear panel. There are ports for  
gas, power, and equipment interface. The Zero In and Span In ports are  
not included on the standard model but are available as options.  
Figure 3-3: Rear Panel of the Model 3000ZA with Span and Zero Ports  
3.3.1 Gas Connections  
Before using this instrument, it should be determined if the unit will  
be used for pressurized service or vacuum service and low pressure  
applications. The kit included with the instrument contains two  
restrictors and a union for 1/4” diameter tubing. Notice that there are  
two 1-3/4” long, 1/4” diameter restrictors. There is an open end and a  
Teledyne Analytical Instruments  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
closed end with a small circular orifice. The restrictor without the blue  
sticker is for low pressure applications and vacuum service. For high  
pressure applications (5-50 psig) use the restrictor that has a blue sticker  
on the body.  
Pressurized service: Connect the restrictor with the blue dot  
to the Sample In port. The small circular orifice should face  
AWAY from the back of the unit (against the direction of gas  
flow).  
Note: Remove the blue sticker from the restrictor before using.  
Low pressure applications: (Less than 5 psig). Connect the  
restrictor without the blue sticker in the same manner.  
Vacuum Service: For applications involving pressure less  
than 5-10 in Hg) connect the restrictor without the blue  
sticker to the Exhaust Out port. The small circular orifice  
should face TOWARD the back of the unit (against the  
direction of gas flow).  
CAUTION: OPERATING THE UNIT WITHOUT RESTRICTORS CAN  
CAUSE DAMAGE TO THE SENSOR.  
The unit is manufactured with 1/4” tube fittings and includes 6 mm  
adapters for metric system installations. For a safe connection:  
1. Insert the tube into the fitting and finger tighten the nut until  
the tubing cannot be rotated freely by hand in the fitting. This  
may require an additional 1/8 turn beyond finger tight.  
2. Hold the fitting body steady with a backup wrench. Using  
another wrench, rotate the nut another 1-1/4 turns.  
SAMPLE IN:  
In the standard model, gas connections are made at the SAMPLE IN and  
EXHAUST OUT connections. Calibration gases must be tee'd into the  
Sample inlet with appropriate valves.  
The inlet gas pressure should be reasonably regulated. Pressures  
between 2 and 50 psig are acceptable as long as the pressure, once  
established, will keep the front panel flowmeter reading in an acceptable  
range (0.5 to 2.0 SLPM). For non-pressurized sample or very low  
Teledyne Analytical Instruments  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
pressure, (less than 2 psig) vacuum service plumbing is recommended.  
Exact figures will depend on your process.  
If greater sample flow is required for improved response time,  
install a bypass in the sampling system upstream of the analyzer input.  
EXHAUST OUT:  
Exhaust connections must be consistent with the hazard level of the  
constituent gases. Check local, state, and federal laws, and ensure that  
the exhaust stream vents to an appropriately controlled area, if required.  
ZERO IN and SPAN IN (Optional):  
These are additional ports for inputting span and zero gas. There are  
electrically operated valves inside for automatic switching between  
calibration and sample gases. These valves are completely under control  
of the 3000ZA electronics. They can be externally controlled only  
indirectly through the remote cal inputs as described below. Note  
however that each sensor is calibrated separately. Thus three span gas  
mixtures are needed. It is recommended to leave connected the one for  
the range that is used most, and calibrate the other two sensors manually.  
Pressure, flow, and safety considerations are the same as prescribed for  
the SAMPLE IN inlet above.  
3.3.2 Electrical Connections  
For safe connections, no uninsulated wiring should be able to come  
in contact with fingers, tools or clothing during normal operation.  
CAUTION:  
USE SHIELDED CABLES. ALSO, USE PLUGS THAT  
PROVIDE EXCELLENT EMI/RFI PROTECTION. THE  
PLUG CASE MUST BE CONNECTED TO THE CABLE  
SHIELD, AND IT MUST BE TIGHTLY FASTENED TO  
THE ANALYZER WITH ITS FASTENING SCREWS.  
ULTIMATELY, IT IS THE INSTALLER WHO ENSURES  
THAT THE CONNECTIONS PROVIDE ADEQUATE  
EMI/RFI SIELDING.  
3.3.2.1 PRIMARY INPUT POWER  
The power cord receptacle and fuse block are located in the same  
assembly. Insert the power cord into the power cord receptacle.  
Teledyne Analytical Instruments  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
CAUTION:  
3000ZA-3X  
POWER IS APPLIED TO THE INSTRUMENT'S  
CIRCUITRY AS LONG AS THE INSTRUMENT IS  
CONNECTED TO THE POWER SOURCE. THE RED  
SWITCH ON THE FRONT PANEL IS FOR SWITCHING  
POWER ON OR OFF TO THE DISPLAYS AND  
OUTPUTS ONLY.  
The universal power supply requires an 85–250 VAC, 47-63 Hz  
power source.  
Fuse Installation: The fuse block, at the right of the power cord  
receptacle, accepts US or European size fuses. A jumper replaces the  
fuse in whichever fuse receptacle is not used. Fuses are not installed at  
the factory. Be sure to install the proper fuse as part of installation. (See  
Fuse Replacement in Chapter 5, Maintenance.)  
3.3.2.2 50-PIN EQUIPMENT INTERFACE CONNECTOR  
Figure 3-4 shows the pin layout of the Equipment Interface  
Connector. The arrangement is shown as seen when the viewer faces the  
rear panel of the analyzer. The pin numbers for each input/output function  
are given where each function is described in the paragraphs below.  
Figure 3-4: Equipment Interface Connector Pin Arrangement  
Analog Outputs: There are four DC output signal pins—two pins per  
output. For polarity, see Table 3-1. The outputs are:  
0–1 VDC % of Range:Voltage rises linearly with  
increasing oxygen, from 0 V at 0  
ppm to 1 V at full scale ppm. (Full  
scale = 100% of programmable  
range.)  
0–1 VDC Range ID: 0.20 V = Low Range, 0.5 V =  
Medium Range, 0.80 V = High  
Range.  
Teledyne Analytical Instruments  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
4–20 mA DC % Range: Current increases linearly with  
increasing oxygen, from 4 mA at 0  
ppm to 20 mA at full scale ppm.  
(Full scale = 100% of  
programmable range)  
4–20 mA DC Range ID: 7.2 mA = Low Range  
12 mA = Medium Range  
16.8 mA = High Range  
Table 3-1: Analog Output Connections  
Pin  
3
Function  
+ Range ID, 4-20 mA, floating  
– Range ID, 4-20 mA, floating  
+ ppm Range, 4-20 mA, floating  
– ppm Range, 4-20 mA, floating  
+ Range ID, 0-1 VDC  
4
5
6
8
23  
24  
7
– Range ID, 0-1 V dc, negative ground  
+ ppm Range, 0-1 VDC  
– ppm Range, 0-1 VDC, negative ground  
Alarm Relays: The nine alarm-circuit connector pins connect to the  
internal alarm relay contacts. Each set of three pins provides one set of  
Form C relay contacts. Each relay has both normally open and normally  
closed contact connections. The contact connections are shown in Table  
3-2. They are capable of switching up to 3 amperes at 250 VAC into a  
resistive load. The connectors are:  
Threshold Alarm 1:  
Can be configured as high (actuates when concentration  
is above threshold), or low (actuates when concentration  
is below threshold).  
Can be configured as failsafe or non-failsafe.  
Can be configured as latching or non-latching.  
Can be configured out (defeated).  
Teledyne Analytical Instruments  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
Threshold Alarm 2:  
Can be configured as high (actuates when concentration  
is above threshold), or low (actuates when concentration  
is below threshold).  
Can be configured as failsafe or non-failsafe.  
Can be configured as latching or non-latching.  
Can be configured out (defeated).  
System Alarm:  
Actuates when DC power supplied to circuits is  
unacceptable in one or more parameters. Permanently  
configured as failsafe and latching. Cannot be defeated.  
Actuates if self test fails.  
(Reset by pressing  
Then press again and any other button Up and Down  
simultaneously to resume.  
button to remove power to relay.  
Further detail can be found in Chapter 4, Section 4.6.  
Table 3-2: Alarm Relay Contact Pins  
Pin  
45  
28  
46  
42  
44  
43  
36  
20  
37  
Contact  
Threshold Alarm 1, normally closed contact  
Threshold Alarm 1, moving contact  
Threshold Alarm 1, normally open contact  
Threshold Alarm 2, normally closed contact  
Threshold Alarm 2, moving contact  
Threshold Alarm 2, normally open contact  
System Alarm, normally closed contact  
System Alarm, moving contact  
System Alarm, normally open contact  
Digital Remote Cal Inputs: Accept 0 V (off) or 24 VDC (on) inputs for  
remote control of calibration. (See Remote Calibration Protocol below.)  
See Table 3-3 for pin connections.  
Teledyne Analytical Instruments  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
Zero: Floating input. 5 to 24 V input across the + and – pins puts the  
analyzer into the Zero mode. Either side may be grounded at  
the source of the signal. 0 to 1 volt across the terminals allows  
Zero mode to terminate when done. A synchronous signal must  
open and close the external zero valve appropriately. See  
Remote Probe Connector. (The –C option internal valves  
operate automatically.)  
Span: Floating input. 5 to 24 V input across the + and – pins puts the  
analyzer into the Span mode. Either side may be grounded at  
the source of the signal. 0 to 1 volt across the terminals allows  
Span mode to terminate when done. A synchronous signal  
must open and close external span valve appropriately. See  
Figure 3-5 Remote Probe Connector. (The –C option internal  
valves operate automatically.)  
Cal Contact: This relay contact is closed while analyzer is spanning  
and/or zeroing. (See Remote Calibration Protocol below.)  
Table 3-3: Remote Calibration Connections  
Pin  
9
Function  
+ Remote Zero  
– Remote Zero  
+ Remote Span  
– Remote Span  
Cal Contact  
11  
10  
12  
40  
41  
Cal Contact  
Remote Calibration Protocol: To properly time the Digital Remote Cal  
Inputs to the Model 3000ZA Analyzer, the customer's controller must  
monitor the Cal Relay Contact.  
When the contact is OPEN, the analyzer is analyzing, the Remote  
Cal Inputs are being polled, and a zero or span command can be sent.  
When the contact is CLOSED, the analyzer is already calibrating. It  
will ignore your request to calibrate, and it will not remember that  
request.  
Teledyne Analytical Instruments  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
Once a zero or span command is sent, and acknowledged (contact  
closes), release it. If the command is continued until after the zero or  
span is complete, the calibration will repeat and the Cal Relay Contact  
(CRC) will close again.  
For example:  
1. Test the CRC. When the CRC is open, send a zero command  
until the CRC closes (The CRC will quickly close.)  
2. When the CRC closes, remove the zero command.  
3. When CRC opens again, send a span command until the  
CRC closes. (The CRC will quickly close.)  
4. When the CRC closes, remove the span command.  
When CRC opens again, zero and span are done, and the sample is  
being analyzed.  
Note: The Remote Valve connections (described below) provides  
signals to ensure that the zero and span gas valves will be  
controlled synchronously. If you have the –C Internal valve  
option—which includes additional zero and span gas  
inputs—the 3000ZA automatically regulates the zero, span  
and sample gas flow.  
Range ID Relays: There are four dedicated Range ID relay contacts. The  
first three ranges are assigned to relays in ascending order—Low range is  
assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and  
High range is assigned to Range 3 ID. Table 3-4 lists the pin connections.  
Table 3-4: Range ID Relay Connections  
Pin  
21  
38  
22  
39  
19  
18  
34  
35  
Function  
Range 1 ID Contact  
Range 1 ID Contact  
Range 2 ID Contact  
Range 2 ID Contact  
Range 3 ID Contact  
Range 3 ID Contact  
Spare Contact  
Spare Contact  
Teledyne Analytical Instruments  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
Network I/O: Pins 13 (+) and 29 (–) are for serial digital input/output  
for local network protocol. At this printing, this port is not yet  
functional. It is to be used for future options to the instrument.  
Remote Valve Connections: The 3000ZA is a single-chassis  
instrument, which has no Remote Valve Unit. Instead, the Remote Valve  
connections are used as a method for directly controlling external  
sample/zero/span gas valves. See Figure 3-5.  
Figure 3-5: Remote Probe Connections  
The voltage from these outputs is nominally 0 V for the OFF and  
15 VDC for the ON conditions. The maximum combined current that  
can be pulled from these output lines is 100 mA. (If two lines are ON at  
the same time, each must be limited to 50 mA, etc.) If more current  
and/or a different voltage is required, use a relay, power amplifier, or  
other matching circuitry to provide the actual driving current.  
In addition, each individual line has a series FET with a nominal  
ON resistance of 5 ohms (9 ohms worst case). This can limit the  
obtainable voltage, depending on the load impedance applied. See  
Figure 3-6.  
Teledyne Analytical Instruments  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
Figure 3-6: FET Series Resistance  
3.3.2.3 RS-232 PORT  
The digital signal output is a standard, full duplex RS-232 serial  
communications port used to connect the analyzer to a computer,  
terminal, or other digital device. It requires a standard 9-pin D  
connector.  
The output data is status information, in digital form, updated every  
two seconds. Status is reported in the following order:  
The concentration in ppm or percent  
The range in use (HI, MED, LO)  
The span of the range (0-100 ppm, etc)  
Which alarms—if any—are disabled (AL–x DISABLED)  
Which alarms—if any—are tripped (AL–x ON).  
Each status output is followed by a carriage return and line feed.  
Three input functions using RS-232 have been implemented to  
date. They are described in Table 3-5.  
Teledyne Analytical Instruments  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Installation  
Table 3-5: Commands via RS-232 Input  
Command  
as<enter>  
az<enter>  
st<enter>  
Description  
Immediately starts an autospan.  
Immediately starts an autozero.  
Toggling input. Stops/Starts any  
status message output from the  
RS-232, until st<enter> is sent  
again.  
The RS-232 protocol allows some flexibility in its implementation.  
Table 3-6 lists certain RS-232 values that are required by the 3000ZA  
implementation.  
Table 3-6: Required RS-232 Options  
Parameter  
Baud  
Setting  
2400  
8 bits  
none  
1
Byte  
Parity  
Stop Bits  
Message Interval  
2 seconds. When CRC opens  
again, zero and span are done,  
and the sample is being analyzed.  
Range ID Relays: There are three dedicated Range ID relay contacts.  
Three ranges are assigned to relays in ascending order—Low range is  
assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and  
High range is assigned to Range 3 ID.  
Network I/O: A serial digital input/output for local network protocol.  
At this printing, this port is not yet functional. It is to be used in future  
options to the instrument.  
RS-232 Port: The digital signal output is a standard RS-232 serial  
communications port used to connect the analyzer to a computer,  
terminal, or other digital device. It requires a standard 9-pin D  
connector.  
Teledyne Analytical Instruments  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
3000ZA-3X  
The data is status information, in digital form, updated every two  
seconds. Status is reported in the following order:  
The concentration in ppm or percent  
The range in use (HI, MED, LO)  
The span of the range (0-100 ppm, etc)  
Which alarms—if any—are disabled (AL–x DISABLED)  
Which alarms—if any—are tripped (AL–x ON).  
3.4 Installing the Zirconium Oxide Sensors  
If the ZrO2 sensors are not installed in the cell block when the  
instrument is shipped, they must be installed before the analyzer is  
placed in service. If they are not installed, the sensors will be packaged  
with their Teledyne Analytical Instruments part number marked on the  
bag.  
When the ZrO2 sensor needs to be installed or replaced, follow the  
procedures in Section 5.2.  
If sensors were installed at the factory, remove the cover, by  
loosening the eight screws that secure it and check that each sensor is  
firm in its respective socket.  
3.5 Testing the System  
Before applying power to the instrument:  
Check the integrity and accuracy of the gas connections.  
Make sure there are no leaks.  
Check the integrity and accuracy of the electrical  
connections. Make sure there are no exposed conductors.  
Check that inlet sample pressure is within the accepted range  
(see Section 3.3.1).  
Power up the system, and test it by repeating the Self-  
Diagnostic Test as described in Chapter 4, Section 4.3.5.  
Teledyne Analytical Instruments  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
Operation  
4.1 Introduction  
Once the analyzer has been installed, it can be configured for your  
application. To do this you will:  
Set system parameters.  
Establish a security password, if desired, requiring the  
operator to log in.  
Establish and start an automatic calibration cycle, if desired.  
Calibrate the instrument.  
Define the three user selectable analysis ranges, then choose  
autoranging or select a fixed range of analysis, as required.  
Set alarm setpoints, and modes of alarm operation (latching,  
failsafe, etc).  
Before you configure your 3000ZA, these default values are in  
effect:  
Ranges: R1 = 100 ppm, R2 = 1.00 %, R3 = 100 %  
(although sensor maxes out around 95% O2).  
Auto Ranging: ON  
Alarm Relays: Active, 80.00 and 90.00 ppm, HI, Not failsafe,  
Not latching.  
Zero: Auto, every 7 days at 12 hours, disabled.  
Span: Auto, every 7 days at 12 hours, disabled.  
Span values: High sensor = 85%, Medium sensor = 2 %, and  
Low sensor= 80 ppm. Span Timer 7 min.  
Password: TAI  
If you choose not to use password protection, the default password  
is automatically displayed on the password screen when you start up and  
you have access to all functions of the analyzer.  
Teledyne Analytical Instruments  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
4.2 Using the Data Entry and Function Buttons  
4.2.1 Style Conventions  
The following typeface conventions are used when referring to  
screen names, key presses and screen readout:  
Screens:  
Arial 12 pt. type in capital letters.  
Example: ANALYZE or MAIN  
screen or menu.  
Key presses:  
<key> The particular keystroke to  
enter is placed between < and >.  
Example: <ENTER> or  
<ESCAPE> or <UP> or  
<DOWN>.  
Only when the keystroke is to be  
entered will it be placed between  
the brackets. If discussing a  
particular key it will be typed  
using all caps. Example: this is the  
ENTER key.  
Screen Modes:  
Times New Roman 12 pt. italic.  
Example: Analysis Mode or Setup  
Mode.  
Screen Readout:  
Arial Narrow, 12 pt bold.  
Example: AUTOCAL or Zero in 12  
days.  
4.2.2 Keys Description  
Note: All menus time out after 45 seconds elapse with no button  
presses, the analyzer returns to the ANALYSIS SCREEN,  
if parameters have been altered but not saved (with the  
ENTER key), the alterations are lost.  
Navigation and Data Entry is possible using the UP and DOWN  
KEYS, the ENTER key, the ESCAPE key and the 7 special function  
keys (ANALYZE, SYSTEM, SPAN, ZERO, ALARMS, RANGE, and  
Teledyne Analytical Instruments  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
STANDBY. It is important to read and understand the MENU  
STRUCTURE section to fully understand navigation and data entry.  
4.2.2.1 ARROW KEYS  
Pressing the ARROW KEYS select menus and modifies values.  
1. Item Selection Menus are identified by the flashing arrow  
on the left. In Item Selection Menus the UP/DOWN arrow  
keys move to the next lower or upper item.  
2. Item Modification Menus are identified by a flashing  
parameter/descriptor. In Item Modification Menus, the  
UP/DOWN arrow keys increment or decrement numeric  
values or toggle ON/OFF, YES/NO, or ENABLE/DISABLE  
type values.  
4.2.2.2 ENTER  
The ENTER key is used in several context-sensitive ways.  
1. In an Item Selection Menu the ENTER key selects the line  
indicated by the arrow, and enters the Item Modification  
Menu.  
2. In an Item Modification Menu the ENTER key accepts the  
displayed value as the new parameter and returns to the Item  
Selection Menu.  
4.2.2.3 ESCAPE  
The ESCAPE key is used in several context-sensitive ways.  
1. In an Item Selection Menu the ESCAPE key restores the  
next higher Item Selection Menu in the hierarchy of menus  
(with the Analyze Mode as the final, highest screen  
available).  
2. In an Item Modification Menu the ESCAPE key returns to  
the Item Selection Menu and restores the original value of  
the parameter (discarding the displayed/modified value).  
3. In sequential operations such as the Password Entry  
Routine the ESCAPE key restores the original value of the  
parameter and backs up to the previous letter or operation in  
the sequence.  
Teledyne Analytical Instruments  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
4.2.2.4 SPECIAL FUNCTION KEYS  
Pressing any of the 7 Function keys (ANALYZE / SYSTEM /  
SPAN / ZERO / ALARMS /RANGE / STANDBY) Cause the display  
to show the top level menu for the respective function. If another menu  
operation was being performed, any modified parameters (that had not  
yet been accepted by using the ENTER key) will be discarded and  
replaced by the original values.  
4.3 Menu/Screen Structure  
The 3000 screen setup consists of several classes of screens and  
items. No item occupies more than 1 line. It is important to read and  
understand the NAVIGATION AND DATA ENTRY section to fully  
understand the menu structure.  
4.3.1 Analyze Screen  
The ANALYSIS SCREEN displays the following:  
Line 1:  
-
-
-
-
The O2 concentration in ppm or '%' (percent)  
If in Span calibration the message "SPAN".  
If in Zero calibration the message "ZERO".  
Otherwise, the message "ANLZ".  
Line 2 (display mode 1):  
-
-
-
-
-
-
If Alarm 1 is alarming, the message "AL-1".  
If Alarm 2 is alarming, the message "AL-2".  
If in Range 1 the message "R1: 0-nnnn.nn ppm".  
If in Range 2 the message "R2: 0-nnnn.nn ppm".  
If in Range 3 the message "R3: 0-nnnn.nn ppm".  
or, if range is in percent, "Rx: 0-nnn.nn %".  
4.3.2 System Menu  
The SYSTEM screen can be accessed by pressing the SYSTEM  
key and it has the following functions:  
Teledyne Analytical Instruments  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
-----------------------  
PASSWORD  
LOGOUT  
MODEL  
SELF TEST  
FILTER  
ANALOG-OUT ADJUST  
AUTO-CALIBRATION  
CHANGE STREAM  
-----------------------  
Figure 4-1: Hierarchy of Available Functions  
Teledyne Analytical Instruments  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
4.3.3 Password routine  
-----------------------  
Enter Password  
'T' 'A' 'I'  
-----------------------  
If the password is 'TAI' then the letters 'T' 'A' 'I' will be supplied to  
the user when this screen is entered. This is an 'easy guess' password. If  
the password is any other string, then the letters 'A' 'A' 'A' will be  
supplied regardless of the actual password. The user must then adjust  
the letters to the correct password.  
Pressing ENTER advances to the next letter. When at the 3rd letter,  
Pressing ENTER causes the letters to be compared to the actual  
password. If the letters don't match, the user is told 'Password Incorrect'.  
If the letters match the password, then all menu items now become  
available and the user is offered the chance to change the password by  
means of the following screens.  
-----------------------  
-> Change Password  
-----------------------  
Pressing ESCAPE causes a return to the Analyze Mode.  
Pressing ENTER proceeds as follows.  
-----------------------  
Enter New Password  
'T' 'A' 'I'  
-----------------------  
Old password is supplied as starting letters. Enter password.  
-----------------------  
Reenter New Password  
Teledyne Analytical Instruments  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
'A' 'A' 'A'  
-----------------------  
The user is prompted to re-enter the password and the letters are  
restored to 'A' 'A' 'A'. This prevents the user from accidentally setting  
the password to an unintended (and, perhaps unguessable) password.  
-----------------------  
Password Mismatch  
-----------------------  
The entered and re-entered password did not match. No change  
was made to the original password.  
-----------------------  
Password Changed!  
-----------------------  
The password is replaced by the entered password.  
4.3.4 Logout Routine  
-----------------------  
Restrict Access?  
ENTER-YES ESCAPE-NO  
-----------------------  
Pressing ENTER causes access to most menus to now be restricted.  
4.3.5 Model Routine  
-----------------------  
3000 ZO-x3 Analyzer  
Teledyne Analytical Instruments  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
V1.0  
-----------------------  
Entering the MODEL screen causes the above screen to be  
displayed for 3 seconds. There are no modifiable parameters.  
4.3.6 Self-Test Routine  
-----------------------  
Begin Self-Test?  
ENTER-YES ESCAPE-NO  
-----------------------  
Pressing ENTER causes the analyzer to run the same diagnostic  
that runs on start-up. See section 5.4 for results display.  
4.3.7 Filter Routine  
-----------------------  
Digital Filter  
->Index: 5  
-----------------------  
Pressing ENTER causes the number after 'Index:' to start  
flashing. Then pressing UP or DOWN selects a number ranging from 1  
to 10. Lower filter values cause the analyzer to converge on new inputs  
more quickly, but also allow more noise in the sample.  
4.3.8 Analog-Out Adjust Routine  
-----------------------  
4-20mA Offset: 0  
Teledyne Analytical Instruments  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
4-20mA Gain: 0  
4-20mA Limit: FALSE  
Track in Cal: Yes  
-----------------------  
4-20mA Offset: Pressing ENTER causes the number after 'Offset:'  
to start flashing. Then pressing UP or DOWN selects a number ranging  
from –341 to 500. The analog output may be sampled using a DMM at  
this time. When the parameter is zero, the analog voltage output should  
be very close to 0.0 (4.00 madc if the 4-20 madc is being adjusted).  
Adjusting the parameter up or down causes the analog output to go up or  
down respectively. When in Analyze mode, this offset value is added to  
the analog output as determined by percentage of range. Note that due to  
difference in electronic offsets between both analog outputs, the 0-1 vdc  
and the 4-20 madc, only one may be adjusted. In other words adjusting  
the 0-1 vdc output to have 0.00 vdc output may lead the 4-20 madc  
output to have an offset.  
4-20mA Gain: Pressing ENTER causes the number after 'Offset:'  
to start flashing. Then pressing UP or DOWN selects a number ranging  
from –500 to 500. The analog output may be sampled using a DMM at  
this time. When the parameter is zero, the analog output should be very  
close to 1.00 vdc or 20.0mA plus the offset caused by the Offset  
parameter (above). Adjusting the parameter up or down causes the  
analog output to go up or down respectively. When in Analyze mode,  
this gain value is factored with the analog output as determined by  
percentage of range .  
The Offset and Gain parameters are unitless numbers. Their effect  
is determined empirically using a DMM. They affect the analog output  
according to the following formula:  
Output = ([% of Range] * GAIN) + OFFSET  
4-20mA Limit: Pressing ENTER causes TRUE/FALSE to start  
flashing. Then pressing UP or DOWN toggles the value. When TRUE,  
the analog output cannot put less than 4mA out of the analog output  
(during ANALYZE MODE). When FALSE, the output may be lower than  
4mA by approximately 10%.  
Teledyne Analytical Instruments  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
Track in Cal: Analog output tracking during calibration user  
preference setting. The setting is remembered after power is turned off.  
Select “Yes” to have analog output track during calibration (either span  
or zero). Select “No” to hold analog output to last reading before  
entering calibration mode. If analog output is not tracking during  
calibration, it will hold the output for a minute after the end of  
calibration to allow calibration gas to be purged out by sample gas.  
4.3.9 Auto-Calibration Timer  
This function sets the timer for auto calibration. It is useful when  
control valves are available and the remote valve control outputs are used.  
Note that if used, only one sensor may be calibrated. Whatever sensor is  
selected from the span menu. So make sure to pre-select the sensor on  
section 4.4.1  
4.3.10 Change Stream  
Use this function with cal valve option to switch on any valve:  
Analyze, Zero or Span. This will allow the user to check the analyzer  
response to cal gas without entering calibration mode.  
4.4 Span Menu  
The SYSTEM screen is as follows:  
-----------------------  
Span Sensor: Low  
Span 200 ppm  
Span Timer: 7 min.  
Span AUTO Finish  
Span Begin  
-----------------------  
Teledyne Analytical Instruments  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
4.4.1 Span Sensor  
-----------------------  
->Span Sensor: High  
-----------------------  
Pressing ENTER causes the selection after 'Sensor:' to start flashing  
(Low / Mid / High). Then pressing UP or DOWN selects a different  
sensor. Note that the span value directly below will change according to  
which sensor is currently selected.  
The high sensor covers ranges from 2 to 95% O2. Span gas that is  
near the top of this scale should be used to calibrate this sensor.  
The medium sensor covers ranges from 200 ppm to 2% O2. Span  
gas that is near the top of this scale should be used to calibrate this  
sensor.  
The low sensor covers ranges from 0 to 100 ppm O2. Span gas that  
is near the top of this scale should be used to calibrate this sensor.  
4.4.2 Span Value  
-----------------------  
->Span 85%  
-----------------------  
Pressing ENTER causes the number after 'Span' to start flashing.  
The number has a range of 1 ppm to 100% and should be set equal to the  
concentration of the span gas used. This number corresponds to the  
sensor selected above. Pressing UP or DOWN adjusts the number.  
4.4.3 Span Timer  
-----------------------  
->Span Timer: 7 min.  
-----------------------  
This function sets the hold time in span mode before the analyzer  
actually performs the span. Adjust to give enough time to purge with  
span gas.  
Teledyne Analytical Instruments  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
4.4.4 Span Finish  
-----------------------  
->Span AUTO Finish  
-----------------------  
Pressing ENTER causes the AUTO or MANUAL after 'Span' to start  
flashing. Pressing UP or DOWN toggles the value. The value  
determines the method by which the span settling period is complete and  
a span calculation is automatically performed. If the value is MANUAL,  
then when the span begins, the user is prompted to press ENTER at the  
time he/she determines the sensor has settled on the span sample. If the  
value is AUTO, the analyzer will automatically countdown the time when  
the sensor has settled on the span sample, and will proceed to calculate  
the new span slope, and return to ANALYZE MODE. The timer is now  
fixed at 7 minutes. Note that the user may press ENTER to complete the  
span even though the value is AUTO.  
4.4.5 Span Begin  
-----------------------  
->Span Begin  
-----------------------  
-----------------------  
Begin Span?  
ENTER-YES ESCAPE-NO  
-----------------------  
Pressing ENTER causes the screen to prompt to begin the span  
routine (as displayed above). Pressing ENTER again causes the  
analyzer to commence the span routine, and begin analyzing in SPAN  
MODE.  
Teledyne Analytical Instruments  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
4.5 Zero Menu  
The ZERO screen is as follows:  
-----------------------  
Zero 0.00 ppm  
Zero Timer: 7 min.  
Zero Auto Finish  
Zero Begin  
-----------------------  
4.5.1 Zero Finish  
-----------------------  
->Zero AUTO Finish  
-----------------------  
Pressing ENTER causes the AUTO or MANUAL after 'Zero' to start  
flashing. Pressing UP or DOWN toggles the value. The value  
determines the method by which the zero settling period is complete and  
a zero calculation is automatically performed. If the value is MANUAL,  
then when the zero begins, the user is prompted to press ENTER at the  
time he/she determines the sensor has settled on the zero sample. If the  
value is AUTO, the analyzer automatically countdowns the time when  
the sensor has settled on the zero sample, and will proceed to calculate  
the new zero offset, and return to ANALYZE MODE. The timer is now  
fixed to 7 minutes. Note that the user may press ENTER to complete the  
span even though the value is AUTO.  
4.5.2 Zero Value  
-----------------------  
->Zero 0.00 ppm  
-----------------------  
Pressing ENTER causes the number after 'Zero' to start flashing.  
The number has a range of 0 ppm to 100 ppm and should be set equal to  
Teledyne Analytical Instruments  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
any offset of the zero gas used. Pressing UP or DOWN adjusts the  
number.  
4.5.3 Zero Begin  
-----------------------  
->Zero Begin  
-----------------------  
-----------------------  
Begin Zero?  
ENTER-YES ESCAPE-NO  
-----------------------  
Pressing ENTER causes the screen to prompt to begin the zero  
routine (as displayed above). Pressing ENTER again causes the  
analyzer to commence the zero routine, and begin analyzing in ZERO  
MODE.  
4.5.4 Zero Timer  
-----------------------  
->Zero Timer: 7 min.  
-----------------------  
This function sets the hold time in zero mode before the analyzer  
actually performs the zero adjustment. Give enough time to purge the  
analyzer with zero gas.  
Teledyne Analytical Instruments  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
4.6 Alarms Menu  
The ALARMS screen is as follows:  
-----------------------  
AL-1 ACTIVE  
AL-1 HIGH  
AL-1 NON-FAILSAFE  
AL-1 NON-LATCHING  
AL-1 20.00 ppm  
AL-2 ACTIVE  
AL-2 HIGH  
AL-2 NON-FAILSAFE  
AL-2 NON-LATCHING  
AL-2 50.00 ppm  
-----------------------  
Note that alarm functionality explanation below applies to both  
AL-1 and AL-2 identically.  
4.6.1 Alarm Active  
-----------------------  
->AL-1 ACTIVE  
-----------------------  
Pressing ENTER causes the word ACTIVE or DEFEATED to start  
flashing. Then pressing UP or DOWN toggles between the 2 words.  
Defeating an alarm causes the alarm to return to a non-alarming  
condition and prevents further tripping of the alarm.  
4.6.2 Alarm Polarity  
-----------------------  
->AL-1 HIGH  
-----------------------  
Teledyne Analytical Instruments  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
Pressing ENTER causes the word HIGH or LOW to start flashing.  
Then pressing UP or DOWN toggles between the 2 words. When HIGH,  
the alarm will trip if O2 concentration is above the setpoint as discussed  
below.  
4.6.3 Alarm Failsafe  
-----------------------  
->AL-1 FAILSAFE  
-----------------------  
Pressing ENTER causes the word FAILSAFE or NON-FAILSAFE to  
start flashing. Then pressing UP or DOWN toggles between the 2  
words. When FAILSAFE, the alarm relay will be energized in a non-  
tripped state, and de-energized when tripped.  
4.6.4 Alarm Latching  
-----------------------  
->AL-1 LATCHING  
-----------------------  
Pressing ENTER causes the word LATCHING or NON-LATCHING to  
start flashing. Then pressing UP or DOWN toggles between the 2  
words. When LATCHING, an alarm will remain tripped until defeated;  
even if the O2 concentration returns to a level that does not trip the  
alarm. When NON-LATCHING, a tripped alarm will become un-tripped if  
the O2 concentration returns to a non-tripping level.  
4.6.5 Alarm Setpoint  
-----------------------  
->AL-1 80.00 ppm  
-----------------------  
Pressing ENTER causes the setpoint concentration to start flashing.  
Then pressing UP or DOWN adjusts the setpoint up or down  
respectively. Setpoints may range from 0.00 ppm to 100.00 %.  
Teledyne Analytical Instruments  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Operation  
4.7 Range Menu  
The RANGE screen is as follows:  
-----------------------  
Range Select: Auto  
Man Rng: R1  
R1: 0-1000.00 ppm  
R2: 0- 1.00%  
R3: 0- 6.00%  
-----------------------  
4.7.1 Range Select  
-----------------------  
Range Select: Auto  
-----------------------  
Pressing ENTER causes the word Auto or Dgtl or Man to start  
flashing. Then pressing UP or DOWN cycles through the 3 words.  
When Auto is selected, the analyzer will select the range  
automatically. The analyzer chooses the least range that will  
accommodate the current O2 concentration. The analyzer will switch to  
the next lower range (if available) when the concentration falls below  
90% of the lower range setpoint and will switch to the next higher range  
(if available) when the concentration exceeds the range's setpoint.  
When Man(ual), the analyzer will remain in the range specified in  
the 'Man Rng' field.  
When Dgtl (Digital), the analyzer will select the range according to  
the signals on the remote digital cal contacts on the rear panel (and  
therefore disabling their original intent of being used aa remote  
calibration initiation inputs).  
Teledyne Analytical Instruments  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
3000ZA-3X  
4.7.2 Manual Range Select  
-----------------------  
Man Rng: R1  
-----------------------  
Pressing ENTER causes the range R1 or R2 or R3 to start flashing.  
Then pressing UP or DOWN cycles through the 3 ranges. The selected  
range is the effective range if the Range Select (above) is set to  
Man(ual).  
4.7.3 Range Setpoints  
-----------------------  
R1: 0-1000.00 ppm  
-----------------------  
Pressing ENTER causes the value to the right of Rn: to start  
flashing. Then pressing UP or DOWN adjusts the value. These are the  
range setpoints for R1, R2, and R3. They may be set to a value in the  
range of 1.00 ppm to 100% with the stipulation that the each range must  
have a setpoint equal to or greater than the setpoint of the range below it  
plus 1ppm.  
Teledyne Analytical Instruments  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
Maintenance  
5.1 Routine Maintenance  
Aside from normal cleaning and checking for leaks at the gas  
connections, routine maintenance is limited to replacing the sensors and  
fuses, and recalibration. For recalibration, see Section 4.4 Calibration.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
5.2 Sensor Replacement  
The ZrO2 sensors are sealed transducers with no electrolyte to  
change or electrodes to clean. When the sensor reaches the end of its  
useful life, it is replaced. The spent sensor should be discarded  
according to local regulations.  
5.2.1 Removing the Sensor  
The ZrO2 sensors is located inside the nylon cell block behind the  
front panel.  
To remove an existing cell (refer to Figure 5-1):  
CAUTION:  
POWER IS APPLIED TO THE INSTRUMENT  
CIRCUITRY WHENEVER THE INSTRUMENT IS  
CONNECTED TO THE POWER SOURCE. THE RED  
SWITCH ON THE FRONT PANEL IS FOR SWITCHING  
POWER ON OR OFF TO THE DISPLAYS AND  
OUTPUTS ONLY.  
1. Remove power to the instrument by unplugging the power  
cord at the power source.  
Teledyne Analytical Instruments  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
2. Open the front panel door by pressing the release button on  
the top right corner of the door all the way in with a narrow  
gauge tool.  
3. Remove the eight screws that hold the cover of the sensor  
block. As you remove the cover make sure the O-rings stay  
in place. If they drop, place them back carefully in the groove  
found on the cell block, not on the cover. See Figure 5.1.  
4. You can remove the sensor that you want to replace with  
your fingers or using small pliers. Just pull the sensor straight  
out.  
CAUTION:  
SENSORS MIGHT BE HOT! AFTER REMOVING  
POWER. LET THE SENSORS COOL FOR A FEW  
MINUTES.  
5. Install the new sensors paying attention to relative position in  
the sensor block. Note orientation of sensor tab aligning with  
tab location on the silkscreen of the mounting PCB. Sensor is  
held by friction of the pins sockets. The locations of the  
Sensors are:  
On the left side P/N S1439 (0-200 ppm range )  
On the center P/N S1492 (0 – 2 % range )  
On the right side P/N S1442 (0 – 95 % range )  
CAUTION:  
MAKE SURE OF THE CORRECT POSITION OF EACH  
SENSOR. DO NOT PLACE A SENSOR IN THE  
WRONG SOCKET OR THE ANALYZER WILL FAIL!  
Teledyne Analytical Instruments  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
Figure 5-1: Sensor Block with Middle Sensor Installed  
5.3 Fuse Replacement  
1. Place small screwdriver in notch, and pry cover off, as shown  
in Figure 5-2.  
Figure 5-2: Removing Fuse Block from Housing  
Teledyne Analytical Instruments  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
2. To change between American and European fuses, remove  
the single retaining screw then flip the Fuse Block over 180  
degrees. Replace the screw.  
3. Replace fuse as shown in Figure 5-3.  
4. Reassemble housing as shown in Figure 5-2.  
American Fuses  
Figure 5-3: Installing Fuses  
European Fuses  
5.4 System Self Diagnostic Test  
1. Press the SYSTEM button to enter the system mode.  
2. Use the ◄► arrow keys to move to More, and press  
ENTER.  
3. Use the ◄► arrow keys to move to Self-Test, and press  
ENTER.  
The following Items are tested:  
Table 5-1: Self-Test Result Display  
“5V - Test : GOOD”  
“15V - Test : GOOD”  
“DAC A - Test : GOOD”  
“DAC B - Test : GOOD”  
( +5VDC power supply )  
(+/-15 VDC power supply )  
(% of range analog output )  
( Range ID analog output )  
“Pre L - Test : GOOD” ( Tests offset of low range amplifier )  
“Pre M - Test : GOOD” ( Tests offset of medium range amplifier)  
“Pre H - Test : GOOD” ( Tests offset of high range amplifier )  
Teledyne Analytical Instruments  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
If any item above fails this test, it will be listed on the VFD  
display followed by the “BAD” message instead of “GOOD”  
5.5 Major Internal Components  
The zirconium sensors block is accessed by unlatching and  
swinging open the front panel, as described earlier. Other internal  
components are accessed by removing the rear panel and sliding out the  
entire chassis. See Figure 5-4, below. The gas piping is illustrated in  
Figures 2-1 and Figure 2-2 and the major electronic components  
locations are shown in Figure 2-3, in Chapter 2.  
CAUTION:  
SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
Figure 5-4: Rear Panel Removal  
(Remove only those screws marked with an X)  
The 3000ZA contains the following major components:  
Analysis Section  
Zirconium oxide sensors  
Cell block  
Teledyne Analytical Instruments  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
Sample system  
Power Supply  
Microprocessor  
Displays  
5 digit LED meter  
2 line, 20 character, alphanumeric, VFD display  
RS-232 Communications Port  
See the drawings in the Drawings section in back of this manual for  
details.  
5.6 Cleaning  
If instrument is unmounted at time of cleaning, disconnect the  
instrument from the power source. Close and latch the front-panel access  
door. Clean outside surfaces with a soft cloth dampened slightly with plain  
clean water. Do not use any harsh solvents such as paint thinner or benzene.  
For panel-mounted instruments, clean the front panel as prescribed  
in the above paragraph. DO NOT wipe front panel while the instrument  
is controlling your process.  
5.7 Troubleshooting  
Table 5-2: Troubleshooting  
Problem  
Possible Cause  
Solution  
Erratic readings of the Analyzer may have  
oxygen concentration been calibrated in an  
Recalibrate analyzer,  
making sure the  
proper gas is fed, the  
proper sensor is  
as reported by the  
analyzer  
inaccurate fashion.  
selected, and the  
proper value of span  
gas is input.  
Teledyne Analytical Instruments  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
Atmospheric oxygen  
Increase flow rate  
diffusing through vent and/or length of vent  
and altering the  
oxygen level the  
sensor sees  
tubing to dilute or  
minimize oxygen  
diffusion from vent to  
sensor.  
Inaccurate zero  
Turn analyzer off,  
then back on again.  
Press the UP and  
DOWN key  
simultaneously while  
the instrument is  
doing the warm up  
count down.  
operation (i.e. user has  
zeroed the instrument  
using a gas concen-  
tration much higher  
than normal for a zero  
gas).  
Press the ENTER to  
confirm returning the  
analyzer to factor  
defaults.  
Re-span each of the  
sensors without doing  
the zero.  
If the analyzer seems  
to be working Ok, up  
to this point, retry to  
zero.  
Teledyne Analytical Instruments  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
Teledyne Analytical Instruments  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
Appendix  
A-1 Specifications  
Packaging: General Purpose  
• Flush panel mount (Standard)  
• Relay rack mount. Contains either one  
or two instruments in one 19" relay rack  
mountable plate (Optional).  
Sensor: Three zirconium oxide sensor.  
Cell Block: Nylon  
Sample System: All wetted parts of 316 stainless steel.  
90 % Response Time: 65 seconds at 25 °C (77 °F) on 10 ppm  
range, 100 seconds on 1 ppm range  
Ranges: Three programmable ranges from 0-1 ppm  
to 0-100 % oxygen (note that the  
maximum concentration of analysis is  
95% O2)  
Autoranging with range ID output.  
Alarms: One system-failure alarm contact to detect  
power failure.  
Two adjustable concentration threshold  
alarm contacts with fully programmable  
setpoints.  
Displays: 2-line by 20-character, VFD screen, and  
one 5 digit LED display.  
Digital Interface: Full duplex RS-232 communications port.  
Power: Universal power supply 85-250 VAC, at  
47-63 Hz.  
Teledyne Analytical Instruments  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
Operating Temperature: 5-35 °C  
Accuracy: ±2% of full scale at constant temperature  
(at calibrated range).  
±5% of full scale over operating  
temperature range, except 0-1 ppm  
analysis range, once thermal equilibrium  
is reached.  
±0.2 ppm on 0-1 ppm analysis range, once  
thermal equilibrium is reached.  
Analog outputs: 0-1 VDC percent-of-range,  
0-1 VDC range ID.  
4-20 mA DC (isolated) percent-of-range,  
4-20 mA DC (isolated) range ID.  
Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep  
(6.96 in high, 8.7 in wide, 12.2 in deep).  
Teledyne Analytical Instruments  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
A-2 Recommended 2-Year Spare Parts List  
Qty.  
Part Number Description  
1
1
1
1
1
2
C65507A  
C62371-A  
B77585A  
C76925  
F10  
Back Panel Board  
Front Panel Board  
ZrO2 Amplifier Board  
Main Computer Board  
Fuse, 2A, 250V 3AG Slow Blow  
F1296  
Fuse, 2A, 250V 5x20mm  
Slow Blow  
1
1
1
1
1
A68314  
S1439  
Back Panel Connector Kit  
Zirconium oxide sensor for 0- 200 ppm range  
Zirconium oxide sensor for 0- 2 % range  
Zirconium oxide sensor for 0- 95 % range  
Restrictor kit  
S1492  
S1442  
A68729  
A minimum charge is applicable to spare parts orders.  
Note: Orders for replacement parts should include the part  
number (if available) and the model and serial number of  
the instrument for which the parts are intended.  
Orders should be sent to:  
TELEDYNE Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500, Fax (626) 961-2538  
or your local representative.  
Teledyne Analytical Instruments  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
A-3 Drawing List  
D-77579 Final Assembly  
D-75738 Outline Drawing  
A-4 19-inch Relay Rack Panel Mount  
Figure A-1: Single and Dual 19" Rack Mounts  
(dimensions in mm)  
Teledyne Analytical Instruments  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
A.5 Application Notes  
3000 SERIES ANALYZERS  
APPLICATION NOTES ON RESTRICTORS,  
PRESSURES, AND FLOW RECOMMENDATIONS  
The 3000 series analyzers require reasonably regulated sample  
pressures. While the 3000 analyzers are not sensitive to variations of  
incoming pressure (provided they are properly vented to atmospheric  
pressure) the pressure must be maintained to provide a useable flow rate  
through the analyzer. Any line attached to sample vent should be ¼” or  
larger in diameter.  
FLOW RATE RECOMMENDATIONS:  
A usable flow rate for a 3000 series analyzer is one which can be  
measured on the flowmeter. This is basically 0.2 - 2.4 SLPM . The  
optimum flow rate is 1 SLPM (mid scale). Note that response time is  
dependent on flow rate; a low flow rate will result in slow response to  
O2 changes in the sample stream. The span flow rate should be the  
approximately same as the sample flow rate.  
CELL PRESSURE CONCERNS:  
The sensors used in 3000 series analyzers are optimized to function  
at atmospheric pressure. At pressures other than atmospheric the  
diffusion rate of O2 will be different than the optimum value. Higher  
pressures will produce faster O2 diffusion rates resulting in higher O2  
reading and shorter cell life. To use a 3000 series analyzer at a cell  
pressure other than atmospheric, the analyzer must be calibrated with a  
known calibration gas at the new cell pressure to adjust for the different  
diffusion rate. Cell pressures below 2/3 atmospheric are not  
recommended because they tend to cause excessive internal expansion  
which may result in seal failure.  
For operation at cell pressures other than atmospheric care must be  
taken not to change the sample pressure rapidly or cell damage may  
occur. For cell pressures above atmospheric, caution must be exercised  
to avoid over pressuring the cell holder. (percent analyzers will require  
some type of cell retainer to prevent the cell from being pushed out by  
the pressure). For operation at pressures below atmospheric pressure a  
suffix C (clamped) cell is required.  
Teledyne Analytical Instruments  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
RESTRICTION DEVICES:  
For proper operation, all 3000 series analyzers require a flow  
restriction device. This device is typically a restrictor or a valve. The  
restriction device serves two functions in the sample path.  
1. It is used to limit the flow rate of the sample through the  
analyzer. A restrictor is chosen to operate over a range of  
pressures and provide a useable flow rate over that range.  
2. The restriction device provides a pressure drop. This device  
is selected to provide the only significant pressure drop in the  
sample path.  
RESTRICTOR KIT  
The current revision of the 3000 series analyzers are supplied with  
a kit containing two restrictors and a union which are user installed.  
These parts are supplied to give the end user more flexibility when  
installing the analyzer. The restrictor kit is suitable for high and low  
positive pressure applications as well as vacuum service applications  
(see manual for installation instructions).  
The standard restrictor (BLUE DOT) is recommended for pressures  
between 5 and 50 PSIG. For positive low pressure application (5 psig or  
less) the unmarked restrictor is better suited. For unpressurized sample  
applications the marked restrictor should be used and configured for  
vacuum service.  
Note: For extremely low positive pressure applications (less then  
2 psig) the vacuum service configuration should provide  
higher performance (higher flow rates).  
For vacuum service, the end user must supply a vacuum pump and  
a by-pass valve for the pump. A vacuum level of 5–10 inches of  
mercury should provide the optimum flow rate. CAUTION: flow  
restrictors have very small orifices and may be plugged by small  
particles (.005” dia or larger) A sample filter must be included in the  
sample line prior to the restrictor! (A 60 micron filter is recommended).  
3000TA EXAMPLES:  
Example 1: With an incoming pressure of 10 psig the standard restrictor  
(blue dot) will provide a flow rate of 0.76 SLPM. Upstream of the  
restrictor the sample line pressure will be 10 psig, while down stream  
Teledyne Analytical Instruments  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trace Oxygen Analyzer  
Maintenance  
(including the cell) the pressure will be at atmospheric pressure  
(analyzer vented to atmospheric pressure). Note, all other pressure  
drops in the sample path are insignificant at these flow rates. This  
insures that the cell operates at atmospheric pressure. At very high flow  
rates (off scale of flow-meter), pressure drops other than the restriction  
device could become significant, and result in pressurizing the cell.  
Example 2: A 3000TA is configured for vacuum service as follows. The  
unmarked restrictor is placed in the sample vent port. The downstream  
end of the restrictor is then connected to a vacuum pump and by-pass  
valve. The by-pass valve is adjusted to provide a flow rate of 1 SLPM.  
The sample pressure between the pump and the restrictor will be  
approximately -7 inches of mercury, while the pressure in the balance  
of the sample system including the cell will be approximately at  
atmospheric pressure (provided the sample flow into the analyzer is not  
blocked).  
BY-PASS:  
To improve the system response, a by-pass can be added to increase  
the sample flow rate to the analyzer by a factor of ten. A by-pass typically  
provides a sample flow path around the analyzer of 2 - 18 SCFH.  
CALIBRATION GAS:  
3000 series analyzer requirements for units with Auto-Cal options  
are as follows:  
The customer must supply control valves (or restrictors) for any SPAN  
or ZERO gas source which is attached to the Auto-Cal ports. The valve  
should be adjusted to the same flow rate as the sample gas. When restrictors  
are used, the gas pressure must be adjusted to achieve the proper flow rate.  
OPERATION WITHOUT A RESTRICTOR DEVICE:  
Operation without a restrictor device is not recommend as  
mentioned above. A 3000TA without any flow restrictor device was  
tested on 11-19-97. This results in a flow rate of 2.4 SLPM @ 1 PSIG.  
This is a cv of 0.023 for the standard sample system.  
REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on  
flow rates at various pressures.  
Teledyne Analytical Instruments  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
3000ZA-3X  
TAI PART NUMBERS  
Restrictor kit:  
Union (SS)  
LP Restrictor  
Std. Restrictor  
Nut  
A68729  
U11  
R2323 (low pressure / vac. service)  
R2324 Blue dot  
N73  
F73  
Ferrule  
Ferrule  
F74  
Both ferrules are required  
CONVERSIONS:  
1 PSI  
=
2.04 inches of mercury (in. Hg.)  
0.476 SLPM  
1 SCFH =  
Teledyne Analytical Instruments  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tascam Computer Hardware IF DM User Manual
Taylor Scale 7553 User Manual
Teac CD Player X 03 User Manual
Teac TV Receiver 7250 User Manual
Teledyne Food Saver DNV 33D User Manual
Thermador Oven SC272 User Manual
Timex Watch W 44 User Manual
Toshiba Cash Register EO1 11095 User Manual
Toshiba Computer Hardware TLCS 900 User Manual
Triarch Indoor Furnishings 25765 User Manual