Teledyne Marine Radio 465M User Manual

INSTRUCTION MANUAL  
MODEL 465M  
OZONE MONITOR  
TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TELEDYNE API)  
9480 Carroll Park Drive  
San Diego, CA 92121-5201  
TOLL-FREE: 800-324-5190  
TEL: 858-657-9800  
FAX: 858-657-9816  
Copyright 2007-2012  
06160D DCN 6409  
23 MARCH 2012  
Teledyne Advanced Pollution Instrumentation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY MESSAGES  
Important safety messages are provided throughout this manual for the purpose of avoiding  
personal injury or instrument damage. Please read these messages carefully. Each safety message  
is associated with a safety alert symbol and placed throughout this manual and inside the  
instrument. The symbols with messages are defined as follows:  
WARNING: Electrical Shock Hazard  
HAZARD: Strong oxidizer  
GENERAL WARNING/CAUTION: Read the accompanying message  
for specific information.  
CAUTION: Hot Surface Warning  
Do Not Touch: Touching some parts of the instrument without  
protection or proper tools could result in damage to the part(s) and/or the  
instrument.  
Technician Symbol: All operations marked with this symbol are to be  
performed by qualified maintenance personnel only.  
Electrical Ground: This symbol inside the instrument marks the central  
safety grounding point for the instrument.  
CAUTION  
GENERAL SAFETY HAZARD  
This instrument should only be used for the purpose and in the  
manner described in this manual. If you use it in a manner other  
than that for which it was intended, unpredictable behavior could  
ensue with possible hazardous consequences.  
NEVER use any gas analyzer to sample combustible gas(es).  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page i  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CONSIGNES DE SÉCURITÉ  
Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but  
d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement  
ces consignes. Chaque consigne de sécurité est représentée par un pictogramme d’alerte de  
sécurité; ces pictogrammes se retrouvent dans ce manuel et à l’intérieur des instruments. Les  
symboles correspondent aux consignes suivantes :  
AVERTISSEMENT : Risque de choc électrique  
DANGER : Oxydant puissant  
AVERTISSEMENT GÉNÉRAL / MISE EN GARDE : Lire la consigne  
complémentaire pour des renseignements spécifiques  
MISE EN GARDE : Surface chaude  
Ne pas toucher : Toucher à certaines parties de l’instrument sans  
protection ou sans les outils appropriés pourrait entraîner des dommages  
aux pièces ou à l’instrument.  
Pictogramme « technicien » : Toutes les opérations portant ce symbole  
doivent être effectuées uniquement par du personnel de maintenance  
qualifié.  
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le  
point central de la mise à la terre sécuritaire de l’instrument.  
MISE EN GARDE  
Cet instrument doit être utilisé aux fins décrites et de la manière décrite  
dans ce manuel. Si vous utilisez cet instrument d’une autre manière que  
celle pour laquelle il a été prévu, l’instrument pourrait se comporter de  
façon imprévisible et entraîner des conséquences dangereuses.  
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz  
combustibles!  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY  
WARRANTY POLICY (02024F)  
Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of  
Teledyne Instruments, Inc., provides that:  
Prior to shipment, TAPI equipment is thoroughly inspected and tested.  
Should equipment failure occur, TAPI assures its customers that prompt  
service and support will be available.  
COVERAGE  
After the warranty period and throughout the equipment lifetime, TAPI  
stands ready to provide on-site or in-plant service at reasonable rates similar  
to those of other manufacturers in the industry. All maintenance and the first  
level of field troubleshooting are to be performed by the customer.  
NON-TAPI MANUFACTURED EQUIPMENT  
Equipment provided but not manufactured by TAPI is warranted and will be  
repaired to the extent and according to the current terms and conditions of  
the respective equipment manufacturer’s warranty.  
Product Return  
All units or components returned to Teledyne API should be properly  
packed for handling and returned freight prepaid to the nearest designated  
Service Center. After the repair, the equipment will be returned, freight  
prepaid.  
The complete Terms and Conditions of Sale can be reviewed at  
CAUTION – Avoid Warranty Invalidation  
Failure to comply with proper anti-Electro-Static Discharge (ESD)  
handling and packing instructions and Return Merchandise  
Authorization (RMA) procedures when returning parts for repair or  
calibration may void your warranty. For anti-ESD handling and  
packing instructions please refer to “Packing Components for  
Return to Teledyne API’s Customer Service” in the Primer on  
Electro-Static Discharge section of this manual, and for RMA  
procedures please refer to our Website at http://www.teledyne-  
api.com under Customer Support > Return Authorization.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
1. INTRODUCTION..........................................................................................................9  
1.1. PREFACE..............................................................................................................................................9  
2. SPECIFICATIONS AND AGENCY APPROVALS ................................................11  
2.1. SPECIFICATIONS.................................................................................................................................11  
2.2. APPROVALS .......................................................................................................................................12  
2.2.1. Safety .......................................................................................................................................12  
2.2.2. EMC.........................................................................................................................................12  
2.2.3. Other Type Certifications ........................................................................................................12  
3. PRINCIPLE OF OPERATION..................................................................................13  
4. GETTING STARTED.................................................................................................15  
4.1. MECHANICAL INSTALLATION (FOR NEMA 4X ENCLOSURE).............................................................15  
4.1.1. Ventilation Clearance..............................................................................................................15  
4.2. MECHANICAL INSTALLATION (19” RACK VERSION)..........................................................................16  
4.2.1. Ventilation Clearance..............................................................................................................16  
4.3. AC POWER CONNECTION (FOR MONITORS IN NEMA 4X ENCLOSURES)...........................................17  
4.4. PNEUMATIC CONNECTIONS................................................................................................................19  
4.4.1. Ozone Inlet Connection ...........................................................................................................20  
4.4.2. Exhaust Connection.................................................................................................................21  
4.5. ELECTRICAL I/O CONNECTIONS.........................................................................................................23  
4.5.1. Location of I/O Connectors in the NEMA Configuration........................................................23  
4.5.2. Analog Output..........................................................................................................................23  
4.5.3. Status Outputs..........................................................................................................................24  
4.5.4. Hi-Current Relay Outputs........................................................................................................25  
4.5.5. Digital Serial Interfaces ..........................................................................................................25  
5. FREQUENTLY ASKED QUESTIONS.....................................................................27  
6. OPTIONAL HARDWARE AND SOFTWARE........................................................29  
6.1. RACK MOUNT SLIDES........................................................................................................................29  
6.2. LOW FLOW ALARM ...........................................................................................................................29  
7. OPERATION ...............................................................................................................31  
7.1. FRONT PANEL OVERVIEW..................................................................................................................31  
7.1.1. Ozone Concentration Display..................................................................................................31  
7.1.2. Menu Navigation Keypad ........................................................................................................31  
7.1.3. Concentration Alarm Status/Reset...........................................................................................32  
7.1.4. Status LED’s............................................................................................................................32  
7.2. FRONT PANEL MENUS .......................................................................................................................33  
7.2.1. Menu Navigation Basics..........................................................................................................33  
7.2.2. Editing Numeric Values...........................................................................................................33  
7.2.3. DIAG Menu..............................................................................................................................36  
7.2.4. VIEW Menu..............................................................................................................................39  
7.2.5. SETUP-VARS Menu.................................................................................................................40  
7.2.6. SETUP-ALARM Menu.............................................................................................................41  
7.2.7. CONFIG Menu ........................................................................................................................41  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page v  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. DIGITAL COMMUNICATIONS..............................................................................43  
8.1. DIGITAL COMMUNICATIONS OVERVIEW ............................................................................................43  
8.2. CHANGING COM PORT HARDWARE CONFIGURATION .......................................................................43  
8.3. RS-232 CONFIGURATION...................................................................................................................45  
8.4. RS-485 CONFIGURATION...................................................................................................................46  
8.5. ETHERNET CONFIGURATION ..............................................................................................................47  
8.5.1. Hardware Setup for Configuring the Ethernet Module ...........................................................47  
8.5.2. Verifying Network Hardware Connection ...............................................................................47  
8.5.3. Determining a Dynamically Assigned IP Address (DHCP Mode)...........................................48  
8.5.4. Manually Assigning a Static IP Address..................................................................................48  
8.6. MODBUS REGISTER MAPS...............................................................................................................50  
8.6.1. Notes on MODBUS registers...................................................................................................50  
8.6.2. Real-Time Concentration and Status Registers (Read Only)...................................................50  
8.6.3. Instrument Setup and Configuration Registers (Read / Write) ................................................50  
9. MAINTENANCE.........................................................................................................51  
9.1. MAINTENANCE SCHEDULE.................................................................................................................51  
9.2. INSTRUMENT LAYOUT .......................................................................................................................52  
9.3. REPLACING INTERNAL PARTICULATE FILTER ....................................................................................54  
9.4. UV LAMP ADJUSTMENT ....................................................................................................................56  
9.5. UV LAMP REPLACEMENT ..................................................................................................................56  
9.6. SAMPLE PUMP REPLACEMENT ...........................................................................................................57  
9.7. AUTO-ZERO VALVE REPLACEMENT...................................................................................................57  
9.8. CONFIGURING THE ANALOG OUTPUT.................................................................................................58  
9.9. CLEANING EXTERIOR SURFACES OF THE M465M..............................................................................59  
9.10. BOOT LOADER OPERATION..............................................................................................................60  
9.10.1. Boot Loader – Hardware Setup.............................................................................................60  
9.10.2. Boot Loader – Software Setup...............................................................................................60  
10. TROUBLESHOOTING ............................................................................................63  
10.1. REFERENCE DRAWINGS ...................................................................................................................63  
10.1.1. Pneumatic Diagram ..............................................................................................................63  
10.1.2. Interconnect Diagram ...........................................................................................................64  
10.2. TROUBLESHOOTING USING FRONT PANEL STATUS LED’S OR STATUS OUTPUTS ............................66  
10.2.1. Sensor OK..............................................................................................................................67  
10.2.2. Invalid Reading .....................................................................................................................67  
10.2.3. Check Lamp...........................................................................................................................67  
10.2.4. Pneumatic Error....................................................................................................................67  
11. CALIBRATION.........................................................................................................69  
11.1. FACTORY CALIBRATION ..................................................................................................................69  
12. TECHNICAL ASSISTANCE ...................................................................................71  
13. M465M SPARE PARTS............................................................................................73  
13.1. SPARE PARTS LIST FOR RACK MOUNT VERSION..............................................................................73  
13.2. SPARE PARTS LIST FOR NEMA VERSION.........................................................................................74  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FIGURES  
Figure 4-1: M465M Mounting Hole Locations and Dimensions.................................................................16  
Figure 4-2. NEMA AC Power Connection to Monitor................................................................................18  
Figure 4-3: Pneumatic Connections, Rack Mount Configuration................................................................19  
Figure 4-4: Pneumatic Connections, NEMA Configuration ........................................................................20  
Figure 4-5: Rear Panel Electrical I/O Connections......................................................................................21  
Figure 4-6: NEMA Electrical I/O Connections............................................................................................22  
Figure 4-7: NEMA I/O Label.......................................................................................................................23  
Figure 4-8: Status Output Schematic............................................................................................................24  
Figure 7-1: Front Panel Display...................................................................................................................31  
Figure 7-2: Front Panel Menu Diagram .......................................................................................................34  
Figure 7-3: Front Panel Menu Diagram .......................................................................................................35  
Figure 7-4: Span Cal Menu..........................................................................................................................39  
Figure 8-1: Com Configuration Jumper Settings (SW2)..............................................................................44  
Figure 9-1: Instrument Layout, Rack Mount Configuration ........................................................................52  
Figure 9-2: Instrument Layout, NEMA Configuration ................................................................................53  
Figure 9-3: Internal Particulate Filter Replacement .....................................................................................55  
Figure 9-4: Mainboard – Analog Output Configuration ..............................................................................59  
Figure 10-1: Pneumatic Diagram (Rack Mount Version) ............................................................................63  
Figure 10-2: Pneumatic Diagram (NEMA Version with Sample Conditioner) ...........................................64  
Figure 10-3: Interconnect Diagram..............................................................................................................65  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLES  
Table 4-1: M465M Ventilation Clearances, NEMA Version........................................................................15  
Table 4-2: M465M Ventilation Clearances, Rack Mount Version................................................................16  
Table 4-3: Status Output Definitions............................................................................................................24  
Table 4-4: Relay Functions...........................................................................................................................25  
Table 7-1: Alarm LED’s...............................................................................................................................32  
Table 7-2: Status LED’s ...............................................................................................................................32  
Table 7-3: Analog Step Test Values.............................................................................................................36  
Table 7-4: Signal I/O List.............................................................................................................................37  
Table 7-5: VIEW Menu Parameters .............................................................................................................39  
Table 7-6: VARS List...................................................................................................................................40  
Table 7-7: Alarm Configuration Settings .....................................................................................................41  
Table 8-1: RS-232 Connector Details ..........................................................................................................45  
Table 8-2: RS-232 Port Setup.......................................................................................................................45  
Table 8-3: RS-485 Connector Details ..........................................................................................................46  
Table 8-4: RS-485 Port Setup.......................................................................................................................46  
Table 8-5: Ethernet LED Definitions ...........................................................................................................47  
Table 9-1: Maintenance Schedule ................................................................................................................51  
Table 9-2: Required Software ......................................................................................................................60  
Table 10-1: Status LED/Output Definitions.................................................................................................66  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. INTRODUCTION  
1.1. Preface  
Teledyne API is pleased that you have purchased the M465M. We offer a full one year  
any support required so that you may utilize our equipment to the fullest extent.  
We hope you will not experience any problems with the Teledyne API Model 465M  
(M465M) but if you do, the built-in tests and diagnostics should allow you to quickly and  
easily find the problem. In addition, our full time customer service department is always  
available to answer your questions.  
Teledyne API, Customer Service,  
9480 Carroll Park Drive  
San Diego, California 92121-5201USA  
Toll-free Phone: 800-324-5190  
Phone: 858-657-9800  
Fax: 858-657-9816  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. SPECIFICATIONS AND AGENCY  
APPROVALS  
2.1. Specifications  
Measuring principle  
Ranges  
UV absorption (Beer Lambert Law)  
User selectable to any full scale range from:  
0-1000 PPM to 0-10,000 PPM  
0-5 g/Nm3 to 0-50 g/Nm3  
0-1 wt% to 0-5 wt%  
Measurement Units  
PPM, wt%, g/Nm3  
Accuracy  
1% of Full Scale  
Precision/Repeatability  
Display Resolution  
Response Time (95%)  
0.5% of Full Scale  
1 PPM, 0.001 wt%, 0.01 g/Nm3  
<45 seconds to 95%  
Pressure, Temperature (NTP = 273.15K, 760  
mmHg)  
Compensation  
Gas Inlet Pressure Range  
Sample Flow Rate  
Temperature Range  
11.0 – 16.0 psia  
0.8 LPM  
o
5-45 C  
Rack Mount (3U Panel) - 5.22” (133 mm) x 19.0"  
(483 mm) x 15.3" (388 mm)  
Dimensions (H x W x D)  
Weight  
With NEMA 4X enclosure – 16.01” (407 mm) x  
15.77” (401 mm) x 8.02” (204 mm)  
Rack Mount – 13.6 lbs (6.17 kg)  
With NEMA 4X enclosure– 16.5 lbs (7.5 kg)  
100-240 VAC, 47-63 Hz, 74 W Max  
2000 meters  
0 – 5 V or 4-20 mA isolated output  
System OK, Invalid Reading, UV Lamp Fault,  
Pneumatic Error  
Power  
Maximum Operating Altitude  
Analog Output  
Status Outputs  
RS-232 (optional RS-485)  
57.6 Kbaud, DB-9 connector  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
2.2. Approvals  
This section presents Safety and Electromagnetic Compatibility (EMC) compliance  
approvals for the Model 465M monitor.  
2.2.1. Safety  
IEC 61010-1:2001, Safety requirements for electrical equipment  
for measurement, control, and laboratory use.  
CE: 2006/95/EC, Low-Voltage Directive  
North American:  
cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04  
NEMKO-CCL (US): UL No. 61010-1 (2nd Edition)  
2.2.2. EMC  
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial  
Immunity  
EN 55011 (CISPR 11), Group 1, Class A Emissions  
FCC 47 CFR Part 15B, Class A Emissions  
CE: 2004/108/EC, Electromagnetic Compatibility Directive  
2.2.3. Other Type Certifications  
For additional certifications, please contact Customer Service:  
Toll-free 800-324-5190  
Phone:  
Phone: 858-657-9800  
Fax: 858-657-9816  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
3. PRINCIPLE OF OPERATION  
The detection of ozone molecules is based on absorption of 254 nm UV light due to an  
internal electronic resonance of the O molecule. The M465M uses a mercury lamp  
3
constructed so that a large majority of the light emitted is at the 254nm wavelength. Light  
from the lamp shines through an absorption cell through which the sample gas is passed.  
The ratio of the intensity of light passing through the absorption cell and the intensity of  
light measured by a reference detector forms a ratio I/Io. This ratio forms the basis for  
the calculation of the ozone concentration.  
The Beer-Lambert equation, shown below, calculates the concentration of ozone from the  
ratio of light intensities.  
106  
14.695psi  
o  
CO    
n  
3
L 273   
Where:  
I = Intensity of light passed through the sample  
Io = Intensity of light through sample free of ozone  
= absorption coefficient  
L = path length  
CO = concentration of ozone in parts per million  
3
= sample temperature in Kelvin  
= pressure in pounds per square inch (absolute)  
As can be seen the concentration of ozone depends on more than the intensity ratio.  
Temperature and pressure influence the density of the sample. The density changes the  
number of ozone molecules in the absorption path which impacts the amount of light  
removed from the light beam. These effects are addressed by directly measuring  
temperature and pressure and including their actual values in the calculation. The  
absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254  
-1  
-1  
nm light. Most current measurements place this value at 308 cm atm at STP. The  
value of this number reflects the fact that ozone is a very efficient absorber of UV  
radiation which is why stratospheric ozone protects the life forms lower in the  
atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
length determines how many molecules are present in the column of gas in the absorption  
tube.  
The intensity of light is converted into a voltage by the detector/preamp module. The  
voltage is converted into a number by a high resolution A/D (analog-to-digital) converter.  
The digitized signal, along with the other variables, is used by the CPU to compute the  
concentration using the above formula.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. GETTING STARTED  
The M465M is shipped with the following standard equipment:  
1. Power cord (Rack mount version only)  
2. Instruction manual.  
Upon receiving the M465M please do the following:  
1. Verify that there is no apparent shipping damage. (If damage has occurred  
please advise shipper first, then Teledyne API.)  
4.1. Mechanical Installation (for NEMA 4X  
Enclosure)  
Mount the enclosure securely to a vertical surface.  
All four mounting holes should be used to secure the monitor.  
Use stainless steel, 5/16” diameter bolts.  
4.1.1. Ventilation Clearance  
When installing the M465M be sure to leave sufficient ventilation clearance.  
Table 4-1: M465M Ventilation Clearances, NEMA Version  
Area  
Minimum required clearance  
Back of the monitor  
Sides of the monitor  
Above and below the monitor  
None  
1 inch  
1 inch  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Figure 4-1: M465M Mounting Hole Locations and Dimensions  
4.2. Mechanical Installation (19” Rack Version)  
The Rack Mount version of the M465M was designed to be operated as a bench-top  
monitor or to be mounted in a standard 19” RETMA instrumentation rack. For rack  
installations, the four mounting feet should be removed from the bottom of the monitor.  
4.2.1. Ventilation Clearance  
When installing the M465M be sure to leave sufficient ventilation clearance.  
Table 4-2: M465M Ventilation Clearances, Rack Mount Version  
Area  
Minimum required clearance  
Back of the monitor  
Sides of the monitor  
Above and below the monitor  
3 inches  
1 inch  
1/2 inch  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
4.3. AC Power Connection (for Monitors in  
NEMA 4X Enclosures)  
CAUTION  
Disconnect power to the AC mains before making or removing  
any electrical connections to the M465M.  
CAUTION  
A proper earth ground connection must be made to the  
receptacle labeled “Earth Ground” on the 3 pin AC connector.  
Failure to do so may result in a shock hazard and malfunction  
of the monitor  
1. Install a ½” conduit fitting for routing the electrical wiring into the monitor through  
the hole provided in the bottom face of the NEMA enclosure; an appropriate sealed  
conduit connector should be used.  
2. Attach the leads of the power line to the AC power connector (see Figure 4-6).  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3. For the NEMA configuration, the electrical connection must be hard-wired to  
the 3-pin connector on the mainboard labeled J23. AC power connection to the  
monitor should be made with 12-14 AWG stranded copper wire, connected to the  
monitor as follows:  
a. Earth Ground (green): Connect the earth ground wire to the screw terminal lug on  
the chassis as shown in the Figure below. Do not connect  
the earth ground wire directly to the PCB connector on the  
mainboard PCB.  
b. Line (Black):  
Connect the Line wire to the connector on the mainboard  
PCB labeled “Line”.  
c. Neutral (White):  
Connect the Neutral wire to the connector on the  
mainboard PCB labeled “Neutral”.  
Figure 4-2. NEMA AC Power Connection to Monitor  
4. Connect the power cord to an appropriate power outlet (see the serial number  
tag for correct voltage and frequency).  
CAUTION  
Verify that the instrument is set up for proper line voltage and  
frequency. Observe local electrical codes when connecting  
power to the monitor.  
5. Turn on the M465 H by switching the switch on the left side of the rear panel, or  
applying power to the monitor (NEMA configuration.) The front panel display  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
should light with a sequence of messages, including the instrument serial number  
and software revision. (When the instrument first powers up, the display will  
show ‘xxxx’, indicating that the instrument is in start-up mode, waiting for the  
UV detector readings to stabilize).  
After the start-up period, the instrument will begin displaying the ozone concentration  
and the optional stream selector will begin operating, if installed.  
4.4. Pneumatic Connections  
Figure 4-3: Pneumatic Connections, Rack Mount Configuration  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Figure 4-4: Pneumatic Connections, NEMA Configuration  
4.4.1. Ozone Inlet Connection  
Sample gas connections are made to the ¼” stainless steel compression fitting on the rear  
1/4" O.D. FEP (Teflon) or Stainless Steel tubing is recommended to connect the sample  
source to the monitor. Any fittings used in the sample lines should be constructed of  
Stainless Steel or Teflon.  
The M465M has an internal sample pump and is designed to sample gas that is close to  
ambient pressure. Ensure that the sample gas pressure is in the range of 11 – 16 psia  
(pounds per square inch absolute.)  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
4.4.2. Exhaust Connection  
Connect a ¼” OD vent line to the “Exhaust” fitting on the rear panel. This line must be  
vented outdoors or to an appropriate discharge system capable of handling ozone. Local  
regulations regarding the discharge of ozone to the atmosphere must be observed. Note  
that the monitor may be equipped with an internal ozone scrubber for removal of ozone  
from the exhaust stream. For safety reasons, the exhaust port must still be vented  
appropriately as described above.  
CAUTION  
Exhaust gas from the M465M may contain dangerous  
levels of ozone! Do not vent this gas into enclosed or  
occupied areas.  
Figure 4-5: Rear Panel Electrical I/O Connections  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Figure 4-6: NEMA Electrical I/O Connections  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4.5. Electrical I/O Connections  
4.5.1. Location of I/O Connectors in the NEMA  
Configuration  
In the NEMA configuration of the M465M, the I/O connectors are located internally in  
usually made via a conduit connection to the NEMA enclosure. A standard conduit  
penetration may be added to one of the walls of the enclosure for this purpose, or the  
existing penetration may be used.  
The I/O label inside the NEMA cover provides information on the functions of the  
Figure 4-7: NEMA I/O Label  
4.5.2. Analog Output  
labeled “ANA.” The analog output can be configured for 0-5V voltage or 4-20 mA  
current output. See Section 7.2.3.1 to automatically step-test the analog output.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
The analog output is automatically scaled to the range of the instrument, i.e. if the  
monitor range is 1000 PPB, then 5V would correspond to 1000 PPB and 0V would  
correspond to 0 PPB. See Section 7.2.5 for information on changing the range of the  
monitor.  
4.5.3. Status Outputs  
There are four instrument status outputs located at the “General I/O” connector. These  
status outputs mirror the state of the four diagnostic LED’s on the front panel. If any of  
these status outputs are in a different state than shown in Table 4-3, then some adjustment  
or troubleshooting should be performed. See Section 10.2 for more detailed information.  
Electrically these outputs are SPST dry contact relay closures with a common contact,  
labeled ‘C’ on the rear panel. They are capable of switching up to 50 VDC, 250mA  
maximum current. There are no polarity requirements, so the outputs can be used to sink  
or source current from the ‘C’ pin, within the maximum voltage and current specified.  
Schematically, the outputs look as shown in Figure 4-8 below:  
Figure 4-8: Status Output Schematic  
The outputs are defined as follows:  
Table 4-3: Status Output Definitions  
Status  
Output #  
Name  
Normal Operating  
State  
1
2
3
4
Sensor OK  
On  
Off  
Off  
Off  
Invalid Reading  
Check Lamp  
Pneumatic Error  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
4.5.4. Hi-Current Relay Outputs  
Three form C relay outputs are provided on the rear panel on the nine pin connector (See  
The relays are dry contact type form C (SPDT) relays with Normally Open (N.O.),  
Normally Closed (N.C.) and Common (COM) contacts. The relays are capable of  
driving loads of up to 250VAC, 5A. For maximum contact life, the relays should only be  
used to drive resistive loads. Contact life may be dramatically shortened if inductive  
loads are driven without any provision for minimizing high voltage “inductive kick” that  
can occur. See Section 7.2.6 to configure the concentration for the alarm trigger.  
Table 4-4: Relay Functions  
Relay #  
Function  
Normal  
Operating State  
1
System OK Diagnostic On  
(Identical to Sensor OK  
status output as  
2
3
Hi Alarm  
Off  
Off  
Hi-Hi Alarm  
4.5.5. Digital Serial Interfaces  
There is one external communications port available on the M465M that can be  
configured as RS232, RS485 (Half Duplex,) or Ethernet (10Mbit.) This section provides  
only a brief overview of the digital serial connections, for more detailed information on  
configuring and using the digital serial interfaces, see Section 8.  
4.5.5.1. RS232 Connection  
With the instrument configured for RS232, connection can be made to the DB-9  
connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O”  
connector. See Section 8.2 for additional details.  
When using RS232, a 9-pin serial cable (TAPI P/N WR77) can be used to connect the  
instrument to a standard PC (personal computer) serial port.  
4.5.5.2. RS485 Connection  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
With the instrument configured for RS485, connection can be made to the DB-9  
connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O”  
connector. See Section 8.2 for additional details.  
4.5.5.3. Ethernet Connection  
With the instrument configured for Ethernet, connection should be made to the RJ-45  
port labeled “Ethernet.” Connection can be made using a standard Category 5 or better  
network patch cable. T-API can provide an appropriate cable, please see the spare parts  
list in Appendix A. See Section 8.2 for additional details.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. FREQUENTLY ASKED QUESTIONS  
1. Q: How long does the UV Lamp last?  
A: The typical lifetime is about 2-3 years.  
The lamp output will steadily decrease until lamp adjustment can no longer be  
performed. The "Check Lamp" warning LED will tell the operator when this  
adjustment should be attempted. When this adjustment can no longer be  
performed, the lamp should be replaced. See Chapter 9 for more information.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. OPTIONAL HARDWARE AND  
SOFTWARE  
6.1. Rack Mount Slides  
Rack mount slides can be attached to the sides of the Rack Mount version of the M465M,  
allowing it to slide completely out of a RETMA rack, facilitating maintenance activities.  
6.2. Low Flow Alarm  
The M465M can be fitted with an internal electronic flow-switch that enables the  
instrument to detect gas flow and provide a Pneumatic Error if the flow is outside it’s  
normal limits.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. OPERATION  
7.1. Front Panel Overview  
The M465M front panel provides a number of functions in addition to displaying the  
current ozone concentration being measured Figure 7-1 below shows the display and  
keypad portion of the front panel.  
Figure 7-1: Front Panel Display  
7.1.1. Ozone Concentration Display  
The 4-character LED display in the center of the front panel normally shows the current  
ozone concentration being measured by the M465M. While displaying concentration,  
one of the unit-of-measure indicators (i.e., WT%, PPM) should be backlit, indicating the  
current measurement units. If none of these indicators are backlit, then the display is in  
Menu mode or Startup mode and is not displaying the concentration. If the display is in  
Menu mode, the display can be returned to Concentration mode by repeatedly pressing  
the CFG button until one of the unit-of-measure indicators is lit. Note that even when the  
display is in Menu mode, normal measurement operation will continue in the  
background.  
NOTE  
Even when the display is in Menu mode, normal measurement  
and stream selector operation will continue in the background.  
7.1.2. Menu Navigation Keypad  
The menu navigation keypad (CFG, ENT and UP/DOWN buttons) is used to navigate  
the various menu items available on the M465M. The menus allow the user to view  
operational parameters of the instrument as well as view and modify the configuration of  
the instrument. See Section 7.2 for more detailed information.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
7.1.3. Concentration Alarm Status/Reset  
The M465M front panel displays the concentration alarm status. An LED labeled  
‘ALARM STATUS’ shows the alarm status as follows:  
Table 7-1: Alarm LED’s  
Alarm Status LED Color  
Alarm State  
Green  
Amber  
Red  
No alarm active  
‘HI’ alarm active  
‘HI-HI’ alarm active  
If the alarms are configured for latching operation, then the RESET button can be used  
to attempt to reset all alarms.  
NOTE  
Alarms can only be reset if the concentration measured has  
subsequently dropped below the alarm limit.  
If the alarms are configured for non-latching operation, they will be automatically cleared  
when the concentration falls below the alarm threshold. The RESET button has no effect  
when the alarms are configured for non-latching operation.  
7.1.4. Status LED’s  
There are four Status LED’s on the front panel to the right of the concentration display.  
These LED’s provide a quick view of the instrument’s Status and provides indication of  
any problems that may exist. The normal state for these LED’s after the initial warm-up  
period is as follows:  
Table 7-2: Status LED’s  
Status LED  
Normal State  
ON  
Sensor OK  
Invalid Reading  
Check Lamp  
Pneumatic Error  
OFF  
OFF  
OFF  
If any of these LED’s are in a different state than shown above, then some adjustment or  
troubleshooting should be performed. See Section 10.2 for more detailed information.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
7.2. Front Panel Menus  
The M465M has an interactive menu structure that can be operated from the front panel.  
From within this menu structure, the operator can view real-time parameters such as  
temperature and pressure, view configuration information, and edit setup parameters such  
as alarm limits.  
See Figure 7-2 and Figure 7-3 shows the menu structure as well as the buttons that are  
used to navigate it.  
Starting with the display in Concentration mode, press CFG to enter Menu mode. At this  
point the unit-of-measure LED should turn off and the first level of the Menu tree will be  
active. Pressing the UP/DOWN arrows will now select from the items available at this  
menu level. Press ENT to enter one of these sub-menus or press CFG to back out of this  
menu level back to Concentration mode.  
7.2.1. Menu Navigation Basics  
1. Press CFG from Concentration mode to enter Menu mode.  
2. Once in Menu mode, CFG will move backwards one level in the menu structure  
and ENT will move forward.  
3. Each menu level contains an EXIT selection. Selecting EXIT and pressing ENT  
will exit the Menu mode and return the display to Concentration mode.  
7.2.2. Editing Numeric Values  
At certain places in the menu structure the operator can edit numeric values. The menu  
buttons behave somewhat differently when in this mode.  
Press CFG to select a digit to edit. The selected digit will blink. Next press the  
UP/DOWN arrows to increment/decrement the selected digit. Press CFG again to select  
another digit. Pressing CFG at the last digit will wrap back to the first digit.  
Once all the digits have been edited, press ENT to accept the new value.  
To abort the editing process without changing the original value, press and hold the CFG  
button until the display returns to the previous menu level.  
NOTE  
Limit checking is enforced, while editing values. The display  
will flash to indicate the user has selected an invalid or out of  
range value. The ENT button will not operate until a valid value  
is entered.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 7-2: Front Panel Menu Diagram  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 7-3: Front Panel Menu Diagram  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
7.2.3. DIAG Menu  
testing and configuring external equipment that may be connected to the rear panel I/O of  
the instrument.  
7.2.3.1. Analog Step-Test  
When placed in this mode, the instrument will automatically step the analog output  
through 5 points from 0 to 100 % and display the current value on the front panel. The  
operator can suspend the automatic cycling and manually control the stepping by  
pressing either one of the UP / DOWN arrow buttons. To resume automatic cycling, exit  
the function by pressing CFG and re-enter by pressing ENT.  
The 5 points will correspond to the following analog output values depending on whether  
the output is configured for 0-5V or 4-20mA:  
Table 7-3: Analog Step Test Values  
Point  
0%  
0-5V Output  
0.00 V ± 0.02  
1.25 V ± 0.02  
2.50 V ± 0.02  
3.75 V ± 0.02  
5.00 V ± 0.02  
4-20 mA Output  
4 mA ± 0.02  
25%  
50%  
75%  
100%  
8 mA ± 0.02  
12 mA ± 0.02  
16 mA ± 0.02  
20 mA ± 0.02  
Press CFG to exit from the Analog Step-Test function.  
7.2.3.2. Signal I/O  
The Signal I/O menu allows the operator to manually control the various relay outputs  
and control inputs available on the rear panel of the M465M. This function is useful for  
testing or debugging external control systems (i.e. dataloggers or PLC’s) to which the  
instrument may be connected.  
See Section 4.5 for additional details on making connections to these I/O signals on the  
rear panel.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Table 7-4: Signal I/O List  
Signal Name  
AUX 1 IN  
Rear Panel Label  
CTRL IN - 1  
Function  
UNDEFINED  
UNDEFINED  
SENSOR OK  
AUX 2 IN  
CTRL IN - 1  
RELAY 1  
RELAY 1  
RELAY 2  
RELAY 2  
HI ALARM  
RELAY 3  
RELAY 3  
HI-HI ALARM  
SENSOR OK  
STAT OUT 1  
STAT OUT 2  
STAT OUT 3  
STAT OUT 4  
STAT AUX 1  
STAT AUX 2  
STATUS OUT – 1  
STATUS OUT – 2  
STATUS OUT – 3  
STATUS OUT – 4  
AUX OUT – 1  
AUX OUT – 2  
INVALID READING  
CHECK LAMP  
PNEUMATIC ERROR  
UNDEFINED  
UNDEFINED  
7.2.3.3. Memory Reset  
This function is currently disabled.  
7.2.3.4. Zero Calibration  
The zero calibration allows the instrument to calibrate it’s internal ozone offset factor.  
This should only be done with a source of zero air connected to the “Sample In” port of  
the M465M. Allow the instrument to stabilize on the zero air source before attempting to  
zero. This normally takes 10-15 minutes.  
If the zero calibration is enabled, the internal ozone offset factor will automatically be  
reported as zero every 6 minutes. If the zero calibration is disabled, then manual  
calibration is necessary based on the customer’s applications. For normal applications,  
the zero calibration is not necessary.  
When entering the Zero Calibration menu, the prompt “ENT TO CAL” appears on the  
display. Simply confirm the calibration by pressing the ENT button to perform the  
calibration (to abort the calibration, press and hold the CFG button to return to the  
previous menu level.) After pressing ENT, the instrument will automatically exit the  
menu mode and return to Concentration mode. The concentration reading should quickly  
go to zero.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NOTE  
While measuring zero air, a certain amount of noise or  
“dithering” of the concentration about the zero point will occur  
and is normal. This noise is typically 1-3 PPM in magnitude.  
7.2.3.5. Span Calibration  
The Span Calibration also allows the instrument to calibrate its internal ozone offset  
factor. The Span Calibration however, is done with some controlled concentration gas  
connected to the “Sample In” port of the M465M. It is recommended that a span gas is  
concentrated to around 80% of the operating range. Set the SPAN VALUE to the  
measured concentration of your calibration gas. This normally takes 10-15 minutes.  
NOTE  
A Span Calibration should only be performed with a stable  
source of Ozone and a reference monitor measuring the same gas  
supply.  
If you are unsure regarding the suitability of a particular source  
of calibration gas, contact Customer Service at Teledyne API for  
assistance (see Section 12.)  
See Figure 7-4 shows the Span Cal menu. After the instrument has stabilized on the  
source of span gas, navigate to the Span Cal menu (See Figure 7-4) and Press ENT. Next  
enter the Span Target concentration (the actual concentration of ozone being supplied to  
the monitor) and press ENT. The display will next show a confirmation menu, ENT TO  
SPAN. Press ENT to perform the Span Calibration, or CFG to abort back to the start. If  
the calibration is successful, the display will return to the concentration menu and the  
monitor reading should change adjust to read very close to the target value. If the  
calibration cannot be performed, an OUT OF RANGE error will be displayed. Press  
ENT to confirm and the display will return to the start of the Span Cal menu.  
If the OUT OF RANGE error occurs, it means the Span Cal cannot be performed  
because it would result in an out of range slope value for the monitor. This means that  
either the sensor in the monitor is malfunctioning, causing improper readings, or the  
actual ozone concentration being supplied to the monitor is different than the target value  
being entered.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 7-4: Span Cal Menu  
7.2.4. VIEW Menu  
The VIEW menu allows the operator to view various measurement parameters in real-  
time. This can be useful for diagnosing various instrument or system problems.  
NOTE  
These values are updated in real-time as they are repeatedly  
scrolled on the display.  
Table 7-5: VIEW Menu Parameters  
Parameter  
Meaning  
Units  
Normal Range*  
MEAS  
UV detector reading, measure cycle  
mV  
250 – 1230  
UV detector reading, reference  
cycle  
REF  
mV  
psia  
ºC  
250 – 1230  
9.0 – 18.0  
50.0 – 52.0  
PRESSURE Sample pressure  
LAMP  
UV lamp temperature  
TEMP  
SAMPLE  
Sample temperature  
TEMP  
ºC  
5.0 –50.0  
ON  
FLOW  
Sample Gas Flow Rate**  
ON/OFF  
*After initial 20 minute warm-up period.  
** Optional Hardware.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
7.2.5. SETUP-VARS Menu  
The VARS menu allows viewing and editing of various global setup variables that effect  
how the M465M operates. These variables are stored in the instrument’s non-volatile  
Table 7-6: VARS List  
Default  
Value  
VAR  
Meaning  
Range  
Controls the latching behavior of  
the concentration alarms. See  
details.  
Number of readings above alarm  
limit required to trigger alarms  
Boxcar filter length for  
concentration filter. Higher  
values will smooth out the  
concentration readings but will  
also increase the response time.  
Enables/disables operation of  
optional flow switch  
LATCHING,  
NON-  
LATCHING  
ALARM MODE  
ALARM DELAY  
LATCHING  
5
1 - 20  
1 - 30  
BC FILT SIZE  
30  
FLOW SWITCH  
ENABLE  
OFF  
1
ON, OFF  
1 - 247  
MODBUS address for COM  
ports.  
Concentration unit-of-measure  
setting.  
COMM ADDRESS  
UNIT  
PPM, WT%,  
g/Nm3  
PPM  
Adjustment for display scroll  
delay. Higher values result in  
slower scrolling speed.  
SCROLL DELAY  
20  
6 - 100  
Enables/Disables auto-zero  
calibration  
Time interval for performing the  
auto-zero calibration (minutes)  
Molecular weight of carrier gas  
(ie. feed gas for ozone  
AZERO ENABLE  
ON  
6
ON/OFF  
3 - 1440  
AZERO INTERVAL  
generator.) This value is only  
used for the calculation of wt%  
concentration units. It is not  
used for PPM or g/Nm3  
calculations.  
GAS CARRIER  
WEIGHT  
32  
20.0 – 40.0  
5 – 50 g/Nm3  
1000 – 9999 PPM  
0.5 – 5.0 wt%  
Full-scale concentration range.  
Used to scale the analog output.  
ANALOG RANGE  
5000  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
7.2.6. SETUP-ALARM Menu  
The ALARM menu allows the operator to view and edit properties of the concentration  
alarms. The table below summarizes the alarm configuration settings.  
Table 7-7: Alarm Configuration Settings  
Property  
Meaning  
Default  
Value  
Range  
HI ALARM  
HI alarm trigger value  
10 PPM  
10 - 9999  
PPM  
HIHI ALARM HI-HI alarm trigger value  
20 PPM  
10 - 9999  
PPM  
ON/OFF HI  
ALARM  
HI alarm enable, enables HI alarm ON  
operation  
ON/OFF  
ON/OFF HI HI HI-HI alarm enable, enables HI-HI ON  
ALARM alarm operation  
ON/OFF  
7.2.7. CONFIG Menu  
The CONFIG menu displays important information about the instrument. The Serial  
Number, COMM address settings, and the latest revision of software installed are  
information available in the CONFIG menu. Settings can be changed under other menus  
(see Sections 7.2.3, 7.2.5, and 7.2.6); however the CONFIG menu just contains  
instrument-specific information.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. DIGITAL COMMUNICATIONS  
8.1. Digital Communications Overview  
The M465M comes equipped with a digital communications (Com) port that can be  
connected to a computer or digital data acquisition system. This COM port uses the  
standardized MODBUSprotocol and can be configured for RS232, RS485, or Ethernet  
(10 Mbit,) allowing for connections to a wide variety of devices and networks. See  
Section 8.2 for more information on configuring the COM port.  
The COM port can be used for data acquisition, alarm triggering, and instrument  
configuration. All the functions that are available at the front panel of the instrument can  
also be performed over the COM port.  
NOTE  
The APICOM software provided by Teledyne API does not currently  
support communications with the M465M. Contact T-API for more  
information on compatible MODBUS software that can be used with  
the M465M.  
8.2. Changing COM port Hardware  
Configuration  
CAUTION  
RISK OF ELECTRICAL SHOCK. THE OPERATIONS  
OUTLINED IN THIS SECTION ARE TO BE PERFORMED BY  
QUALIFIED MAINTENANCE PERSONNEL ONLY!  
The COM port can be configured for RS232, RS485, or Ethernet. To change the COM  
port hardware configuration, perform the following steps:  
1. Disconnect power from the M465M.  
2. Remove the six screws and the top cover.  
3. Locate the DIP switches labeled SW3 on the rear portion of the mainboard.  
4. Turn ON the switches specified for the desired COM configuration (See Figure  
5. Locate SW2 (DTE-DCE switch) on mainboard.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
6. For RS-232 operation, set SW2 to either DTE or DCE (See Section 8.3) For RS-  
485 operation, set SW2 to DCE.  
7. Re-install the instrument cover.  
8. Re-connect power to the instrument.  
9. The software portion of the COM port configuration will now automatically be  
completed after the next instrument boot-up.  
SW3  
1
2
3
4
5
6
7
8
9 10  
RS232 Configuration  
ETHERNET  
Close Indicated Positions Only  
RS485  
RS232  
SW3  
3
5
6
7
8
9 10  
1
2
4
RS485 Configuration  
ETHERNET  
Close Indicated Positions Only  
RS485  
RS232  
SW3  
1
2
3
4
5
6
7
8
9 10  
Ethernet Configuration  
ETHERNET  
Close Indicated Positions Only  
RS485  
RS232  
Figure 8-1: Com Configuration Jumper Settings (SW2)  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8.3. RS-232 Configuration  
The RS-232 configuration is usually used when making a one-to-one connection between  
the instrument and a single computer or PLC. The communications protocol used for the  
RS-232 configuration is MODBUS RTU. For details on the MODBUS RTU  
specification, please see http://www.modbus.org/.  
With the instrument configured for RS-232, connection can be made to the DB-9  
connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O”  
connector. Only one of these connections should be used, the other must be left un-  
Table 8-1: RS-232 Connector Details  
Function  
DB-9  
DB-9  
General I/O  
Connector Pin Connector Pin Connector  
(SW2 – DTE) (SW2 – DCE)  
Transmit (Tx)  
Receive (Rx)  
Ground  
3
2
5
2
3
5
‘T’  
‘R’  
‘G’  
NOTE  
A DTE-DCE switch is provided so that the instrument can be  
connected to another device using either a straight-through or  
crossover (Null-Modem) cable. See Section 8.2 for details on  
changing this configuration.  
The serial port setup for RS-232 configuration is shown in Table 8-2 below:  
Table 8-2: RS-232 Port Setup  
Property  
Baud Rate  
Data Bits  
Parity  
Value  
57600  
8
None  
1
Stop Bits  
Flow Control  
None  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
8.4. RS-485 Configuration  
RS-485 is commonly used for factory automation busses with a computer or PLC acting  
as a master. The communications protocol used for the RS-485 configuration is  
MODBUS RTU. For details on the MODBUS RTU specification, please see  
With the instrument configured for RS-485, connection can be made to the DB-9  
connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O”  
Table 8-3: RS-485 Connector Details  
Function  
DB-9  
General I/O  
Connector Pin Connector  
(SW2 – DCE)  
RS485 - A  
RS485 - B  
Ground  
2
3
5
‘T’  
‘R’  
‘G’  
NOTE  
For RS-485 configuration, the DTE-DCE switch should always  
be set to DCE. See Section 8.2 for details on changing this  
configuration.  
The serial port setup for RS-485 configuration is shown in Table 8-4 below:  
Table 8-4: RS-485 Port Setup  
Property  
Baud Rate  
Data Bits  
Parity  
Value  
57600  
8
None  
1
Stop Bits  
Flow Control  
None  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
8.5. Ethernet Configuration  
Ethernet is commonly used for factory automation busses with a computer or PLC acting  
as a master. The communications protocol used for the Ethernet configuration is  
MODBUS TCP/IP. For details on the MODBUS TCP/IP specification, please see  
A Windows software application, DeviceInstaller, is available for configuring the  
Ethernet module used in the M465M. This application is available for download here:  
8.5.1. Hardware Setup for Configuring the Ethernet  
Module  
To make the first-time configuration of the Ethernet module as easy as possible, the  
module comes from the factory with DHCP enabled. The M465M can simply be plugged  
into a network with a DHCP server and an IP Address should be automatically assigned  
to, and stored in, the device.  
A simple network can easily be constructed for this purpose by using a small broadband  
router commonly used for home networks. The M465M and a PC can then be connected  
to the router and the router’s DHCP server will assign IP Addresses to both the PC and  
the M465M, enabling them to communicate.  
8.5.2. Verifying Network Hardware Connection  
With the M465M connected to a network and powered up, check the status of the two  
LEDs on the Ethernet connector on the rear panel. The left-hand LED should be lit,  
either amber or green indicating link status and the right-hand led should blink  
periodically, indicating activity on the network. See Table 8-5 below for more details on  
the status LED’s.  
If neither of the LED’s are lit, then the Ethernet module may not be powered up or the  
instrument may not be connected to a valid network. Check the COM port configuration  
per Section 8.2 and make sure that it is set to “Ethernet.”  
Table 8-5: Ethernet LED Definitions  
Link LED (Left Side)  
Activity LED (Right Side)  
Color  
Off  
Meaning  
Color  
Off  
Meaning  
No Link  
No Activity  
Half-Duplex  
Full-Duplex  
Amber  
Green  
10 Mbps Connection  
Amber  
100 Mbps Connection Green  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
8.5.3. Determining a Dynamically Assigned IP  
Address (DHCP Mode)  
The DeviceInstaller software application (See Section 8.5) can be used to search a  
network for instruments and determine the IP Address assigned by a DHCP server.  
1. Click Start->Programs->Lantronix DeviceInstaller->DeviceInstaller. If  
your PC has more than one network adapter, a message displays. Select an  
adapter and click OK.  
2. Click on the Search icon. After a moment a list of instruments on the network  
will be displayed. The M465M should be listed as an XPORT-IAP type device.  
3. Double-click on the device in the right-hand section of the DeviceInstaller  
window; a list of configuration parameters will be shown, including the IP  
Address. If multiple devices are shown, the correct one can be determined by  
matching the Hardware Address shown with the address printed on the label of  
the Ethernet module in the M465M.  
NOTE  
The instrument cover must be removed to inspect this address in  
the instrument.  
8.5.4. Manually Assigning a Static IP Address  
For applications where the IP Address must remain constant, a static IP Address can be  
manually assigned to the instrument and the DHCP function must be disabled.  
1. Click Start->Programs->Lantronix DeviceInstaller->DeviceInstaller. If  
your PC has more than one network adapter, a message displays. Select an  
adapter and click OK.  
NOTE  
If the unit already has an IP address (e.g., DHCP has assigned an  
IP address), click the Search icon and select the unit from the list  
of Lantronix device servers on the local network.  
2. Click the Assign IP icon .  
3. If prompted, enter the hardware address (on the product label) and click Next.  
4. Select Assign a specific IP address and click Next.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
5. Enter the IP address. The Subnet mask displays automatically based on the  
IP address; if desired, you may change it. On a local network, you can leave the  
Default gateway blank (all zeros). Click Next.  
6. Click the Assign button and wait several seconds until a confirmation message  
displays. Click Finish.  
7. Select the device from the main window list and select Ping from the Tools  
menu. The Ping Device dialog box shows the IP address of the selected unit.  
8. From the Tools menu, click the Ping button. The results display in the Status  
window. Click the Clear Status button to clear the window so you can ping the  
device again.  
NOTE  
If you do not receive “Reply” messages, make sure the unit is  
properly attached to the network and that the IP address assigned  
is valid for the particular network segment you are working with.  
If you are not sure, check with your systems administrator.  
9. Click the Close button to close the dialog box and return to the main window.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8.6. MODBUS Register Maps  
8.6.1. Notes on MODBUS registers  
Concentration values in the MODBUS registers are always expressed in units of PPB  
(parts-per-billion,) regardless of the concentration units setting of the monitor.  
8.6.2. Real-Time Concentration and Status Registers  
(Read Only)  
Decimal  
Star  
#
Address Description  
Register Type t Bit Bits  
1
Stream Concentration (g/Nm3)  
Float Inverse  
Byte  
0
0
8
0
0
0
0
0
32  
8
3
Hi Alarm Status  
3
Hi Hi Alarm Status  
Byte  
8
19  
21  
23  
25  
27  
Sensor Reference Value  
Sensor Measure Value (mV)  
Sensor Pressure Value (mV)  
Sensor Gas Temperature (K)  
Sensor UV Lamp Temperature (oC)  
Float Inverse  
Float Inverse  
Float Inverse  
Float Inverse  
Float Inverse  
32  
32  
32  
32  
32  
8.6.3. Instrument Setup and Configuration Registers  
(Read / Write)  
Decimal  
Address Description  
Register  
Type  
Star  
t Bit Bits  
#
1030  
1032  
1034  
1034  
1035  
1038  
Hi Alarm Value (PPM)  
Float Inverse  
Float Inverse  
Byte  
0
0
0
8
0
0
32  
32  
8
Hi Hi Alarm Value (PPM)  
Hi Alarm Enable (PPM)  
Hi Hi Alarm Enable (PPM)  
Alarm Latch Mode  
Byte  
8
Byte  
8
Analog Range (PPM)  
Float Inverse  
32  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
9. MAINTENANCE  
CAUTION  
RISK OF ELECTRICAL SHOCK. THE OPERATIONS  
OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY  
QUALIFIED MAINTENANCE PERSONNEL ONLY!  
9.1. Maintenance Schedule  
Table 9-1 below outlines the suggested maintenance procedures and intervals for  
ensuring the M465M continues to operate accurately and reliably.  
Table 9-1: Maintenance Schedule  
Maintenance Item  
Recommended Interval  
Section  
Replace particulate filter 6 months*1  
element  
Adjust UV lamp  
As Indicated by ‘Check Lamp’  
LED or status output  
Replace lamp  
As required; when adjustment can 9.5  
no longer be performed.  
Replace sample pump  
2 years  
Replace Auto-Zero Valve 2 years  
Rebuild cells, replace o-  
rings  
3 years  
Return  
to  
factory  
*1 When external sample line pre-filters are used. If pre-filters are not  
used, internal sample filter should be replaced every month.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
9.2. Instrument Layout  
will be referenced in the procedures that follow.  
Note the caution areas where high voltage (line voltage) may be present when  
power is connected to the instrument.  
Figure 9-1: Instrument Layout, Rack Mount Configuration  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Figure 9-2: Instrument Layout, NEMA Configuration  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
9.3. Replacing Internal Particulate Filter  
1. Disconnect power from the M465M.  
2. Loosen the six screws from the top cover (Rack Mount Configuration, see Figure  
3. Remove the six screws and top cover from the instrument.  
4. Locate the particulate housing filter on the rear panel (Rack Mount Configuration,  
see Figure 9-1) or the bottom panel (NEMA Configuration, see Figure 9-2.)  
5. Loosen the four screws on the sample filter body (see Figure 9-3.)  
6. Remove the fours screws, the four washers and the sample filter body from the  
sample filter base.  
7. Remove the o-ring and inspect for any cracks or deformities. If there are no  
cracks or deformities, place the o-ring back in the sample filter body.  
NOTE  
If the o-ring is cracked or deformed, discard the o-ring and  
replace with a new one.  
8. Remove the two sample filter retainers and glass fiber filter element from the  
sample filter body.  
9. Discard the glass fiber filter element.  
10. Place a new glass fiber filter element in between the two sample filter retainers  
and place in the sample filter body.  
11. Place the sample filter body on the sample body base and secure with the four  
screws and fours washers.  
12. Reinstall the top cover on the instrument and secure with the six screws or close  
the front panel.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 9-3: Internal Particulate Filter Replacement  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
9.4. UV Lamp Adjustment  
1. Instrument should be running and warmed up for at least 20 minutes.  
2. With instrument running, remove the six screws and the top cover (Rack Mount  
Configuration) or open front panel (NEMA Configuration.)  
3. Locate the Reference UV Detector adjustment pot, PR3, on the Mainboard PCA  
4. Navigate the front panel menu to VIEW menu and scroll to REF display and press  
ENT. At this point there should be a scrolling display similar to “REF = XXXX  
MV.” See Section 7.2 for details on menu navigation.  
5. While observing the REF value on the display, slowly turn the pot to adjust the  
value. The target adjustment range is as high as possible within the range of 800  
– 1150 mV.  
6. Locate the Measure UV Detector adjustment pot, PR4, on the mainboard PCA  
7. Navigate the front panel menu to VIEW menu and scroll to MEAS display and  
press ENT. At this point there should be a scrolling display similar to “MEAS =  
XXXX MV.” See Section 7.2 for details on menu navigation.  
8. While observing the MEAS value on the display, slowly turn the pot to adjust the  
value. The target adjustment range is as high as possible within the range of 800  
– 1150 mV.  
9. If the required adjustment cannot be achieved by adjusting the UV Detector pot  
alone, then additional adjustment can be made by loosening the two UV lamp  
setscrews on the UV lamp housing (see Figure 9-1 or Figure 9-2) and rotating the  
lamp. Rotate the lamp very slowly while observing the REF or MEAS value on  
the display. Make sure the lamp does not pull out and remains seated in the  
housing while it is being rotated. Re-tighten the two setscrews when a desired  
point has been reached.  
10. If necessary, additional “fine tuning” can now be done with the UV Detector  
adjustment pots per steps 4-8.  
11. Re-Install instrument cover and observe REF and MEAS values on display for a  
couple minutes to verify it does not drift out of the adjustment range.  
9.5. UV Lamp Replacement  
1. Disconnect power from the M465M.  
2. Remove the six screws and the top cover (Rack Mount Configuration) or open  
front panel (NEMA Configuration.)  
3. Loosen the two UV lamp setscrews on the UV lamp housing (see Figure 9-1 or  
4. Unplug the lamp power cord from the connector labeled J18 on the mainboard  
PCA.  
5. Carefully slide the lamp out of housing.  
6. Install the new lamp, seating it in the lamp housing until it stops.  
7. Re-tighten the two UV lamp setscrews.  
8. Plug the lamp power cord into J18 on the mainboard PCA.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
9. Reconnect power to the instrument and turn on power switch. Let instrument  
warm up for at least 20 minutes.  
10. Perform UV lamp adjustment procedure per Section 9.4.  
9.6. Sample Pump Replacement  
1. Disconnect power from the M465M.  
2. Remove the six screws and the top cover (Rack Mount Configuration) or open  
front panel (NEMA Configuration.)  
3. Locate sample pump assembly (See Figure 9-1 or Figure 9-2.)  
4. Cut off the clamps holding the Tygon tubing to the pump nipples and remove  
tubing. Note which tubing is connected to inlet and outlet.  
5. Unplug the pump power connector from the mainboard PCA at connector J12.  
6. Remove the four screws attaching the sample pump assembly to the chassis.  
7. Turn over pump assembly and remove the two screws holding the sheet-metal  
base to the pump.  
8. Install the base on the new pump.  
9. Install new pump assembly in chassis.  
10. Cut off the ends of the used Tygon tubing and reconnect tubing; note that outlet  
fitting of pump should be connected to tubing routed to the ‘Exhaust’ fitting on  
rear panel.  
11. Re-install new tubing clamps or cable-ties to secure tubing connections.  
9.7. Auto-Zero Valve Replacement  
1. Disconnect power from the M465M.  
2. Remove the six screws and the top cover (Rack Mount Configuration) or open  
front panel (NEMA Configuration.)  
3. Locate the auto-zero valve (See Figure 9-1 or Figure 9-2.)  
4. Unplug the two-pin valve connector from the mainboard PCA.  
5. Remove the silver retainer clip from the top of the sensor valve. A pair of pliers  
may be used to slide off the retainer clip.  
6. The valve coil can now be removed by sliding upwards.  
7. Remove the two (2) mounting screws using a #2 Phillips screwdriver. Note that  
there are four screw heads visible on the top of the valve body, only two (2) of  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
these should be removed. Use the new valve as a reference to determine which  
two screws need to be removed.  
8. Note the orientation of the valve body on the manifold so that the new valve can  
be installed in the same orientation.  
9. Remove valve body from sensor manifold.  
10. Clean any residue or dirt off the surface of the manifold using a lint-free cloth and  
distilled or DI water.  
11. Install the new valve by reversing steps 1-7.  
9.8. Configuring the Analog Output  
The analog output can be configured for either 0-5V DC or 4-20mA operation. To  
change or verify the configuration of the analog output:  
1. Disconnect power from the M465M.  
2. Remove the six screws and the top cover (Rack Mount Configuration) or open  
front panel (NEMA Configuration.)  
3. Locate the Mainboard PCA (see Figure 9-1 or Figure 9-2.)  
4. Set the desired operation as shown in Figure 9-4.  
5. Re-Install the top cover or re-secure the front panel (NEMA Configuration.)  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SW1  
R67  
CPU ALIVE  
D14  
R1  
VCC  
TP1  
3.3V  
DGND  
TP3  
J1  
J2  
TP2  
R64  
+
C1  
C2  
X1  
U3  
RS-485/  
R2 R3 R4  
C3  
C4  
Analog Out  
Expansion  
PIC18F8720  
J3  
C5  
VR1  
U4  
R6  
R5  
XPORT XE  
C6  
DAC  
TP4  
U1  
R10  
U2  
R8 R7 R9  
R11  
R66  
R13 R14 C9  
C8  
C7  
LT1181  
JP3  
JP4  
FLOW  
R15  
SW2  
D1  
C10  
C11 C12  
Q1  
C15  
SW3  
C14  
74HC373  
C13  
74HC373  
R17  
R18  
U5  
C16  
C17  
R16  
SWITCH DIP 10  
U8  
TP5  
C18  
JP1  
ISO124  
U6  
R27  
U7  
ISO_GND  
J4  
C20  
PCF8575  
U10  
U9  
ETHERNET  
RS485  
RS232  
Close Indicated Positions Only  
C19  
TP6  
R19  
R20  
C23  
XTR110  
C22  
C30  
D2  
D3  
TP7  
74HC373  
C21  
C24  
7407  
ISO_V  
SRAM128KX8-SOP  
C27  
R24  
R22  
R23  
R21  
+
C25  
U14  
SPAN  
C28  
U11  
U12  
J5  
5.0VREF  
C26  
.
.
.
J6  
JP2  
JP3  
TP8  
R26  
U13  
C31  
R25  
C41  
C29  
+
0-5V OUTPUT  
Q1  
PR1  
ZERO  
D4  
U18  
U15 C32  
PRESS  
C34  
U16  
M41T0  
R27  
R29  
C33  
J24  
C35  
U19  
U17  
SW4  
R28  
K1  
K2  
K3  
C36  
R30  
PR2  
JP4  
Y1  
R31  
I2  
JP3  
Q1  
C38  
DCP010515  
Analog In  
Expansion  
C37  
R32  
C40  
U20  
C39  
U21  
TP9  
C45  
D5  
C50  
J7  
D6  
HCF4052B  
D7  
VR3  
C42  
VR5 R33  
U22  
R36  
VR2  
C44  
C43  
R37  
R38  
C49  
D8  
J11  
R34  
U23  
LM7805  
R35  
C46  
I1  
TP10  
R45  
TP11  
R27  
N.C.  
D9  
R39  
R40  
R41  
C48  
C47  
U24  
COM  
N.O.  
N.C.  
4.096VREF  
PR3  
TP12  
R44  
R46  
C53  
J8  
R42  
C51  
R43  
JP4  
TP14  
D10  
VR4  
REF_DET  
R47  
R50  
SDA  
TP13  
COM  
N.O.  
N.C.  
SCL  
TP15  
U25  
R49  
C52  
R48  
REF  
PR4  
C55  
C57  
U26  
D11  
+15V  
TP19  
4-20mA OUTPUT  
C54  
R51  
R52  
TP18  
C56  
C58  
J9  
-9V  
TP16  
TC1044  
COM  
N.O.  
TP17  
TP20  
J13  
+12VL  
U27  
J10  
C59  
R53  
MEAS  
R54  
C61  
D12  
J14  
J15  
J12  
+12V  
HEATER  
C62  
R65  
C63  
C60  
D15  
MEAS_DET  
TP21  
VR6  
TP22  
R55  
D13  
J22  
C64  
Q2  
J17  
J20  
PUMP MODULE AUX  
+12V  
PS1  
VALVE  
+15V  
K4  
+9V  
+15V +15V  
LAMP INVERTER  
J16  
16  
C65  
C66  
J18  
15  
14  
13  
12  
11  
R56  
U28  
K5  
K7  
R57  
J19  
R58  
JP5  
R59  
K6  
C69  
C67  
C68  
VR7  
R60  
10  
9
Q3  
C70  
PS2  
R61  
R62  
J21  
K8  
K9  
8
U29  
R63  
7
6
5
4
3
2
AC  
G1  
K10  
G2  
J23  
EARTH  
1
LINE  
NEUT  
C71  
C72  
DANGER!  
HIGH VOLTAGE  
PCA 05072  
REV:  
TELEDYNE API  
(C)2004  
Figure 9-4: Mainboard – Analog Output Configuration  
9.9. Cleaning Exterior Surfaces of the M465M  
If necessary, the front panel mask and keyboard of the M465M can be cleaned with a  
damp cloth. Do not attempt to clean any of the other surfaces of the instrument. Do not  
submerge any part of the instrument in water or cleaning solution.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
9.10. Boot Loader Operation  
The main function of a boot loader is to load or update the firmware to the  
microcontroller on the Printed Circuit Board (PCB) of an instrument. There are two  
types of boot loader setup: hardware and software, described in the following sections.  
NOTE  
When an external MODBUS host device is attempting to  
communicate with the instrument over the serial or Ethernet port  
while the instrument is starting up, the instrument may halt. In  
order to prevent the problem from occurring, ensure that there are  
no devices attempting communication with the instrument during  
start-up. Once the sign-on message is displayed on the front  
panel, it is safe to begin communications.  
9.10.1. Boot Loader – Hardware Setup  
Verify that the instrument is configured for an RS-232 connection. See Section 8.2 for  
details on changing the COM port configuration. The instrument can be connected to the  
PC COM port using a 9-pin serial cable (TAPI P/N WR77).  
9.10.2. Boot Loader – Software Setup  
Table 9-2 lists the required firmware loader software and firmware options.  
Table 9-2: Required Software  
SOFTWARE TYPE  
FILE TYPE/VERSION  
FirmwareLoader.EXE  
PC Software  
05498-00_XX.HEX (for ppb, ppm, ugm)  
05498-01_XX.HEX (for ppb, ppm, pphm)  
Instrument Firmware  
The firmware loader is available on Teledyne API’s website at http://www.teledyne-  
api.com/software/.  
1. Install the firmware loader software onto a PC.  
2. Verify the cable connections. Do not power up the instrument.  
3. Run the firmware loader application on the PC.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
4. Select the COM port that the instrument is connected to.  
5. Select the HEX file for the appropriate firmware.  
6. Refer to Section 9.10.2.1 to upload the firmware or to Section 9.10.2.2 to verify  
the firmware.  
9.10.2.1. Upload Firmware  
1. Press “Write”. The screen will display, “Wait for boot loader”.  
2. When the screen displays, “Power cycle instrument now”, power up the  
instrument.  
NOTE  
Search for boot loader will time out in 20 seconds. If the  
instrument is not powered up in 20 seconds, the screen will  
display, “No boot loader detected.” Go back to step 1 to rewrite  
the firmware.  
The firmware loader software should display a message “Boot loader found,  
uploading…”. If this message is not displayed, power down the instrument and  
receck the COM port configuration and cable (see Section 9.10.1) and retry.  
3. When the upload is complete, a message, “Write-flash completed successfully”  
should be displayed. If it is an unsuccessful upload, a message, “Write-flash  
error” is displayed. If a Write-flash error occurs, ensure that the correct PC COM  
port is selected for the instrument (see Section 9.10.2, step 4).  
NOTE  
It is not recommended to stop the writing process half way  
through. This will cause the firmware to not be fully uploaded.  
9.10.2.2. Verify Firmware  
The verify function compares the instrument’s current firmware to the selected HEX file  
(see Section 9.10.2, step 5) and confirms whether or not the files are a match.  
1. Press “Verify”. The screen will display, “Wait for boot loader”.  
2. When the screen displays, “Power cycle instrument now”, power up the  
instrument.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NOTE  
Search for boot loader will time out in 20 seconds. If the  
instrument is not powered up in 20 seconds, the screen will  
display, “No boot loader detected.” Go back to step 1 to verify  
the firmware.  
The firmware loader software should display a message “Boot loader found,  
verifying…”. If this message is not displayed, power down the instrument and  
receck the COM port configuration and cable (see Section 9.10.1) and retry.  
3. When the upload is complete, a message, “Verify-flash completed successfully”  
should be displayed. If it is an unsuccessful upload, a message “Verify-flash  
error” is displayed. If a Verify-flash error occurs, ensure that the correct PC  
COM port is selected for the instrument (see Section 9.10.2, step 4).  
NOTE  
A verification process can be stopped at any point in time. This  
will not affect the firmware that is already uploaded on the  
instrument. The instrument will automatically go to the firmware  
in 10 seconds.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10. TROUBLESHOOTING  
CAUTION  
RISK OF ELECTRICAL SHOCK. THE OPERATIONS  
OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY  
QUALIFIED MAINTENANCE PERSONNEL ONLY!  
10.1. Reference Drawings  
The drawings contained in this section are for general reference and may be useful when  
performing certain troubleshooting activities.  
10.1.1. Pneumatic Diagram  
Figure 10-1 and Figure 10-2 below is a pneumatic diagram that can be referenced when  
performing troubleshooting on the monitor.  
O3 Gas In  
Filter  
Flow-Meter  
Azero Valve  
O3 Destruct  
Flow Switch  
(Optional)  
O3  
Measurement  
Cell  
Exhaust Out  
Flow Control  
Orifice  
Pump  
Figure 10-1: Pneumatic Diagram (Rack Mount Version)  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
O3 Gas In  
Dryer  
Water  
Drop-Out  
Filter  
Flow-Meter  
Azero Valve  
O3 Destruct  
Flow Switch  
(Optional)  
O3  
Measurement  
Cell  
Filter  
Exhaust Out  
Flow Control  
Orifice  
Pump  
Figure 10-2: Pneumatic Diagram (NEMA Version with Sample Conditioner)  
10.1.2. Interconnect Diagram  
modules in the M465M.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
RACK-MOUNT  
CN0000073 IEC TYPE  
VERSION ONLY  
POWER ENTRY MODULE  
SW0000051 POWER  
SWITCH/CIRCUIT  
BREAKER  
05072-0100 MAINBOARD PCA  
J23 AC IN  
AC/DC Power  
Supply  
J19  
FAN (REAR  
PANEL)  
J15  
FLOW SWITCH  
(OPTIONAL)  
J24 “ANALOG IN  
EXPANSION”  
03484-XXXX SENSOR ASSY  
SW0000062 PRESSURE  
SENSOR  
J7  
J17  
J20  
04511-0100 HEATER/  
TEMP SENSOR PCA  
02571-0100 MEAS  
DETECTOR PCA  
J16  
02571-0100 REF  
DETECTOR PCA  
J8  
03742 UV LAMP ASSY  
J18  
05313-0100 DISPLAY PCA  
J4  
+15V  
J14 “MODULE +15”  
J4  
J2  
I2C  
J2  
N/C  
04702 PUMP ASSY  
(M465M ONLY)  
12V  
15V  
J12 “PUMP +12V”  
05125 AUTO-ZERO  
VALVE ASSY  
J13 “VALVE”  
Figure 10-3: Interconnect Diagram  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
10.2. Troubleshooting Using Front Panel Status  
LED’s or Status Outputs  
The M465M has 4 front-panel status LED’s that reflect the current operating status of the  
monitor, and indicate fault conditions. There are also four relay Status Outputs on the  
rear panel (or internally in the NEMA configuration) that also reflect the state of these  
status LED’s. See Section 4.5.3 for more information on the Status Outputs.  
Table 10-1: Status LED/Output Definitions  
Status  
Output # Label  
Status LED  
Normal  
Operating  
State  
Triggers  
Critical  
Warning?  
1
2
Sensor OK  
On  
Yes  
No  
Reference < 125.0 mV  
Reference > 1230.0 mV – 0.5 Wt%  
O3 Concentration < F.S. Range  
O3 Concentration > F.S. Range  
Reference < 250.0 mV  
Pressure < 9.0 psia  
Pressure > 18.0 psia  
Invalid Reading Off  
3
4
Check Lamp  
Pneumatic  
Error  
Off  
Off  
No  
No  
Flow < 500 cc/min*  
*If Flow Switch installed  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
10.2.1. Sensor OK  
The Sensor OK LED indicates the status of the O3 sensor module in the monitor. The  
normal state of this LED (or Status Output) is On. If this LED remains off after the  
normal warm-up period, then a failure has occurred and the monitor should be  
removed from service and repaired.  
The most common cause of this warning is a failure of the UV Lamp. A UV Lamp  
Adjustment (See Section 9.4) should be attempted. If the UV Lamp cannot be  
adjusted to meet the specified values, then a UV Lamp Replacement should be  
performed (See Section 9.5.)  
10.2.2. Invalid Reading  
The Invalid Reading LED indicates that the instrument is reading a value that cannot  
be represented properly on the analog output. Since the analog output is limited to 0-  
5V (or 4-20mA,) it cannot properly represent negative values, or values in excess of  
the full-scale range.  
If the monitor is consistently reading negative values, then a zero calibration should  
be performed, see Section 7.2.3.4.  
If the monitor is consistently reading values in excess of the full-scale range, then the  
range value.  
10.2.3. Check Lamp  
The Check Lamp LED indicates that the UV Lamp intensity has dropped below  
250mV, a level where UV Lamp Adjustment (See Section 9.4) should be made at the  
next convenient opportunity. Note that this is a non-critical warning, unless it is also  
accompanied by Sensor OK LED turning off, then the instrument does not need to be  
removed from service at this time. If the UV Lamp intensity drops below 125mV,  
then the Sensor OK LED will also turn off, indicating that the monitor must be  
immediately serviced or taken off-line.  
10.2.4. Pneumatic Error  
The Pneumatic Error LED indicates that one of the pneumatic parameters, flow or  
pressure, has gone outside of normal ranges. Note that this is a non-critical warning.  
Unless it is also accompanied by Sensor OK LED turning off, the instrument does not  
need to be removed from service at this time.  
The first step in troubleshooting a Pneumatic Error is determining which parameter  
has caused the warning. At the monitor front panel, navigate to the VIEW menu (See  
Section 7.2.4.) Examine the Flow and Pressure values and compare them to the limits  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
10.2.4.1. Pressure Too High  
The monitor inlet is being pressurized. The monitor inlet should be allowed to  
sample gas at ambient pressure. Disconnect all sample lines from the monitor and  
read the pressure again. If the value then drops to within the acceptable range,  
then one of the sample lines is at elevated pressure.  
If the inlet is being pressurized, then the design of the sample system should be  
reviewed and corrected.  
10.2.4.2. Pressure Too Low  
Something may be restricting the flow through the monitor, causing the sample  
pump to pull a vacuum on the sensor module. First, unplug the sample pump  
power plug on the Mainboard PCA and observe the pressure reading. If the  
pressure reading is still too low, then likely there is a problem with the pressure  
sensor on the Sensor Module and the Sensor Module should be replaced. Next  
reconnect the pump and disconnect all sample lines from the monitor and see if  
the pressure increases to normal levels. If it does, then the problem is somewhere  
in the sampling system external to the monitor. If the pressure is still below the  
limit with the sample lines removed, then there is a restriction in the pneumatics  
of the monitor upstream of the sensor assembly. The most likely cause is a  
plugged particulate filter. Disconnect the outlet fitting on the sample filter and  
observe the pressure. If it returns to normal levels then the restriction is in the  
particulate filter or inlet manifold assembly (multi-stream configurations only.)  
Replace particulate filter per Section 9.3.  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11. CALIBRATION  
CAUTION  
RISK OF ELECTRICAL SHOCK. THE OPERATIONS  
OUTLINED IN THE FOLLOWING SECTIONS OF THIS  
CHAPTER ARE TO BE PERFORMED BY QUALIFIED  
MAINTENANCE PERSONNEL ONLY!  
11.1. Factory Calibration  
The M465M is calibrated to internal T-API calibration standards to shipment. A  
calibration certificate for your instrument can be purchased from Teledyne API if  
required. Teledyne API also recommends that the M465M be re-calibrated once a year.  
Teledyne API can provide NIST traceable calibration services at our factory or on-site.  
Please contact our Customer Service department for details on these services (see Section  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12. TECHNICAL ASSISTANCE  
If this manual does not solve your problems, technical assistance may be obtained from:  
Teledyne API, Customer Service,  
9480 Carroll Park Drive  
San Diego, California 92121-5201USA  
Toll-free Phone: 800-324-5190  
Phone: 858-657-9800  
Fax: 858-657-9816  
Before you contact Teledyne Instruments’ Customer service, fill out the problem report  
form in Appendix C, which is also available online for electronic submission at  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
THIS PAGE IS INTENTIONALLY LEFT BLANK  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
13. M465M SPARE PARTS  
13.1. Spare Parts List for Rack Mount Version  
Part Number  
025710100  
028510000  
036750000  
037420000  
045110100  
046740000  
050720400  
051250000  
053130400  
053200000  
DR0000006  
HW0000120  
OP0000022  
OR0000079  
OR0000080  
OR0000081  
OR0000082  
OR0000099  
OR0000105  
PS0000042  
SW0000029  
SW0000051  
WR0000008  
Description  
PCA, UV DETECTOR PREAMP  
KIT, EXPENDABLES, 25MM GLASS FIBER FILTER  
AKIT, ELEMENTS, 47MM, 30 MICRON (10 PCS)  
ASSY, UV LAMP (BIR)  
PCA, HEATER/SENSOR  
ASSY, PUMP, 12VDC  
PCA, MAINBOARD, M465H/M, CE  
ASSY, VALVE, HI-LEVEL O3, SIMRIZ  
PCA, FRONT PANEL, SNGL CHL, M465, CE  
PCA, FRONT PANEL DAUGHTER BOARD  
DRYER, 24", 1/4" SS FITTINGS  
SHOCKMOUNT, GROMMET ISOLATOR  
WINDOW, SAPPHIRE, .039" THK X .75" DIA  
ORING, 2-014, CV-75 COMPOUND  
ORING, 2-018, CV-75 COMPOUND  
ORING, 2-022, CV-75 COMPOUND  
ORING, 2-122, CV-75 COMPOUND  
ORING, 2-129, CV-75 COMPOUND  
0-RING, 2-105V  
PS, 60W SWITCHING, 2X4, 15V  
PRESSURE TRANSDUCER, SS, 0-50PSIA  
SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB  
POWER CORD, 10A  
06182E M465M SPL, NEMA  
(DCN5191)  
11/26/08  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
13.2. Spare Parts List for NEMA Version  
Part Number  
Description  
025710100  
PCA, UV DETECTOR PREAMP  
028510000  
036280000  
036750000  
037420000  
045110100  
046740000  
048990300  
050720400  
051250000  
053130400  
053200000  
061600000  
061470000  
061550000  
061560000  
DR0000006  
FL0000017  
HW0000120  
OP0000022  
OR0000079  
OR0000080  
OR0000081  
OR0000082  
OR0000099  
PS0000042  
KIT, EXPENDABLES, 25MM GLASS FIBER FILTE  
ASSY, WATER DROP OUT FILTER, PFA  
AKIT, ELEMENTS, 47MM, 30 MICRON (10 PCS)  
ASSY, UV LAMP (BIR)  
PCA, HEATER/SENSOR  
ASSY, PUMP, 12VDC  
ASSY, HI-CON O3 SCRUBBER, CH71  
PCA, MAINBOARD, M465H/M, CE  
ASSY, VALVE, HI-LEVEL O3, SIMRIZ  
PCA, FRONT PANEL, SNGL CHL, M465, CE  
PCA, FRONT PANEL DAUGHTER BOARD  
MANUAL, OPERATORS, M465M  
GASKET, ADAPTOR PLATE, M465M, NEMA  
GASKET, FLOWMETER  
GASKET, KEYBOARD, FRONT PNL, M465 NEMA  
DRYER, 24", 1/4" SS FITTINGS  
FILTER ELEMENT, 25MM GLASS FIBER, 1.5UM  
SHOCKMOUNT, GROMMET ISOLATOR  
WINDOW, SAPPHIRE, .039" THK X .75" DIA  
ORING, 2-014, CV-75 COMPOUND  
ORING, 2-018, CV-75 COMPOUND  
ORING, 2-022, CV-75 COMPOUND  
ORING, 2-122, CV-75 COMPOUND  
ORING, 2-129, CV-75 COMPOUND  
PS, 60W SWITCHING, 2X4, 15V  
06183E M465M SPL, Rack  
(DCN5191)  
11/26/08  
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 74  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

Tascam Music Mixer US 1641 User Manual
Tatung Car Video System TLM 1506 User Manual
Texas Instruments Stereo System DEM PCM1717 User Manual
TRENDnet Network Hardware TPE113GI User Manual
Tri M Systems Automobile Parts V104 User Manual
UltraViolet Devices Water System MP13A User Manual
Uniden Scanner BCT 10 User Manual
Vector TV Mount PWA 14 User Manual
Viking Oven F20362D User Manual
Wagan Power Supply 400 watt User Manual