Teledyne Automobile Parts HFM 60 User Manual

TELEDYNE HASTINGS  
INSTRUCTION  
INSTRUCTION MANUAL  
INSTRUMENTS  
MODEL HFM-60  
FLOWMETER  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
1.0 G EN ERAL IN FO RM AT IO N ............................................................. 5  
1.1 Features ...............................................................................................................................5  
1.2 Specifications .......................................................................................................................6  
1.3 Optional 4-20 mA Current Output ......................................................................................7  
1.4 Digital Display (900 tiltable) .................................................................................................7  
1.5 WallTransformer Power Supply ...........................................................................................7  
1.6 Other Accessories .................................................................................................................7  
2.0 IN ST AL L AT IO N .................................................................................. 9  
2.1 Receiving Inspection ............................................................................................................9  
2.2 Power Requirements .............................................................................................................9  
2.3 Output Signal .......................................................................................................................9  
2.4 Mechanical Connections.......................................................................................................9  
2.5 Electrical Connections ........................................................................................................ 10  
2.6 Operation ........................................................................................................................... 11  
3.0 T H EORY O F O PERAT IO N .............................................................1 3  
3.1 Overall Functional Description ........................................................................................... 13  
3.2 Sensor ................................................................................................................................ 13  
3.3 Base ................................................................................................................................... 14  
3.4 Shunt ................................................................................................................................. 14  
3.5 Electronic Circuitry ........................................................................................................... 16  
4.0 M AIN T EN AN C E ................................................................................1 7  
4.1 Authorized Maintenance..................................................................................................... 17  
4.2 Troubleshooting ................................................................................................................. 17  
4.3 Adjustments ....................................................................................................................... 18  
4.4 End Cap Removal .............................................................................................................. 19  
4.5 Printed Circuit Board Replacement .................................................................................... 20  
4.6 Sensor Replacement ........................................................................................................... 20  
5.0 WARRAN T Y AN D REPAIR .............................................................2 1  
5.1 Warranty Policy.................................................................................................................. 21  
5.2 Non-Warranty Repair Policy .............................................................................................. 21  
6.0 D IAG RAM S AN D D RAWIN G S ......................................................2 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 1  
General Information  
T he Hastings Mass Flowmeter, HFM-60 is designed to accurately measure mass flow up to 70  
slpm with an accuracy of better than 3%. Hastings mass flow instruments do not require any  
periodic maintenance under normal operating conditions with clean gases. No damage will occur  
from the use of moderate overpressures (~150 psi) or overflows. Calibrations for other gases, such  
as oxygen, helium and argon, are available upon special order. Instruments are normally calibrated  
with the appropriate standard calibration gas (nitrogen) then a correction factor is used to adjust the  
output for the intended gas.  
1.1  
Features  
T he HFM-60 is intrinsically linear. Should recalibration (a calibration standard is  
required) in the field be desired, the customer needs to simply set the zero and  
span points.  
T he HFM-60 incorporates a removable/replaceable main PC board/sensor module  
which virtually eliminates long down time due to clogging. Clogging is the most  
common cause of failure in the industry. T he HFM-60 sensor is less likely to be  
clogged due to the large internal diameter (0.020”).  
Each flowmeter has a shunt which can be quickly and easily exchanged in the field  
to different ranges. Calibration, however, is required.  
T he filters located upstream of the shunt and sensor eliminate most of the larger  
particulates and impurities which tend to clog the sensor tube.  
T he HFM-60 is available in ranges from 0-10 sccm to 0-60 slpm (air equivalent  
flow). Special ranges are available upon request.  
T he HFM-60 is constructed of 6061-T 6 aluminum alloy. Buna-N is standard for the  
O-rings and seals. Other elastomers are available upon request, e.g., Kalrez,Viton.  
T he temperature coefficient of the HFM-60 is less than 0.1%/°C from 0-50°C.  
T he 4-20 mA option gives the user the advantages of a current loop output to  
minimize environmental noise pickup.  
T he optional 0-90° tiltable digital display allows convenient real time indication of the  
flow rate.  
T he low pressure drop across this instrument is ideal for leak detection applications.  
page 5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.2  
Specifications  
Accuracy .........................................................................................................................<±3% F.S.  
Repeatability ................................................................................................. ±0.2% of F.S.(typical)  
±0.4% of F.S. (max.)  
Maximum pressure ............................................................................................................. 150 psi.  
Pressure coefficient ................................................................................................... 0.01%/psi (N2)  
Operating temperature ............................................................................ 0-50°C in non-condensing  
environment  
Temperature coefficient (zero) ..................................................................... ±1000 ppm/°C, typical  
(±2000 ppm/°C, max)  
Temperature coefficient (span) ....................................................................... ±500 ppm/°C, typical  
(±1000 ppm/°C, max)  
Leak integrity ........................................................................................................... 1x10-9 std. cc/s.  
Standard Flow ranges...................................................................... 10, 20, 50, 100, 200, 500 sccm.  
1, 2, 5, 10, 20, 40, 70 slpm.  
Standard Output .............................................................................................................. 0-5VDC.  
Optional Output .............................................................................................................. 4 -20 mA.  
Power ...................................................................... +15 to +25VDC, 60 mA (0 - 5VDC output).  
+15 to +25VDC, 80 mA (4-20 mA output).  
Wetted Materials ...................................................................... 6061 Al, 304L stainless steel, nylon,  
Buna N, Monel 400, Pb40Sn solder,  
(Cu30Zn brass in 40 to 60 slpm units,  
and epoxy in 10 to 200 sccm units)  
page 6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.3  
1.4  
O ptional 4-20 m A Current O utput  
An option to the standard 0-5VDC output is the 4-20 mA current output that is proportional to  
flow. T he 4 - 20 mA signal is produced from the 0 - 5VDC output of the flowmeter. T he current  
loop output is useful for remote applications where pickup noise could substantially affect the  
stability of the voltage output. T he load impedance should be between 0 and 400 .  
D igital D isplay (0-90° Tiltable)  
1.4.1 Local Display  
For real time indication of the flow rate, a very useful option available is a 3.5 digit LCD digital  
display module mounted directly on top of the flowmeter case. T he display can be rotated 90°  
from the horizontal position to the fully vertical for better visibility. T he LCD digits are large and  
clear and can be easily discerned at distances of 20 feet.  
1.4.2 Rem ote Display  
T he tiltable display module may be removed from the flowmeter case and mounted at any distance  
away from the flowmeter body to permit convenient observation. Standard lengths of 7', 15' and  
25' cable (also special orders of any desired length) are available.  
1.5  
1.6  
Wall Transfor m er Power Supply  
T he input voltage required for operation of the flowmeter is +15 to +25VDC, at a maximum of  
60 mA (80 mA for the 4-20 mA output option). A 115VAC wall power supply rectified to +18.5  
VDC is available as a power supply option. T he wall power supply comes with a 6' length of 2  
conductor cable which can be connected directly to the HFM-60 connector.  
O ther Accessor ies  
1.6.1 Totalizer (ATR-1J)  
T he Hastings FlowTotalizer integrates the 0-5VDC signal generated by the flowmeter to give a  
total flow reading. Count rates from 0 to 999 counts per minute are selectable by internal setting.  
1.6.2 Hastings Model 400/200/40/10 Power Supply  
Hastings power supplies (other than the aforementioned wall transformer) are available in either  
one, two or four channel versions. T hey convert 115 or 230VAC to the +15VDC required to  
operate the flowmeter. Interface terminals for the +15VDC input and the 0-5VDC linear output  
signal are located on the rear of the panel. Also, a cable can be supplied with the power supply that  
provides the +15VDC on pin 11 of a “D” connector and the 0 - 5VDC output measurement on  
pin 6. Pins 5 and 12 are common and pin 7 is chassis ground. T hroughout this manual, when  
reference is made to a power supply, it is assumed the customer is using a Hastings wall power  
supply or a Model 200/400/40/10 supply.  
page 7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 2  
Installation  
T his section contains the necessary steps to assist in getting a new flowmeter in operation as quickly  
and easily as possible. Please read the following thoroughly before attempting to install the instru-  
ment.  
WAR N IN G  
T he customer should determine if their process gas is compatible with  
the wetted materials of the flowmeter as specified in section 1.2. Some  
corrosivegases may damage the flowmeter materials and elastomer  
O-ring. T his could result in incorrect flow measurement, or leakage over  
time.  
2.1  
Receiving and Inspection  
Carefully unpack the Hastings HFM-60 and any accessories that have also been ordered. Inspect  
for any obvious signs of damage to the shipment. Immediately advise the carrier who delivered the  
shipment if any damage is suspected. Check each component shipped with the packing list. Insure  
that all parts are present (i.e., flowmeter, wall transformer power supply, cables, etc.). Optional  
equipment or accessories will be listed separately on the packing list. T here may also be one or  
more OPT-options on the packing list. T hese normally refer to special ranges or special gas  
calibrations. T hey may also refer to special helium leak tests, high pressure tests or special modifi-  
cations such as high temperature O-ring materials. In most cases, these are not separate parts, but  
special options or modifications built into the flowmeter.  
2.2  
2.3  
2.4  
Power Requirem ents  
T he HFM-60 requires +15 to+25VDC, 60mA (0 - 5VDC output); 80mA (4 -20mA output). T he  
supply voltage should be regulated such that there is no more than 50 mV ripple. Surge suppres-  
sors are recommended to prevent power spikes reaching the instrument. T hese power require-  
ments are satisfied by the Hastings wall transformer or by the Hastings wall power supply described  
in Section 1.6.1 or 1.6.2.  
O utput Signal  
T he standard output of the flowmeter is a 0-5VDC signal proportional to the flow rate. T he input  
power is applied at pins 1 (+15 to+25VDC) and 2 (ground). T he output is obtained on pins 3 (0 to  
+5VDC) and 4 (common) of the 4 pin connector. It is recommended that the load resistance be  
no less that 2 k. When the 4-20 mA option (see Section 1.3) is selected the output is also on pins  
3 and 4. T he load impedance must be no greater than 400 .  
Mechanical Connections  
T he flowmeter may be mounted in any position as long as the direction of gas flow through the  
instrument follows the arrow marked on the bottom of the flowmeter case label. T he preferred  
orientation is with the inlet and outlet fittings in a horizontal plane. Changing the mounting  
orientation requires re-zeroing of the instrument at zero flow.  
T he smallest of the internal passageways in the HFM-60 is the diameter of the sensor tube, which  
is 0.020 and the annular clearance for the 500 sccm shunt which is 0.009", so the instrument  
requires adequate filtering of the gas supply to prevent blockage or clogging of the tube. T he  
screen diffusers assist in filtering to some extent where the filter mesh is as small as 0.010".  
page 9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
T here is a threaded hole in the bottom of the transducer that can be used to secure it to a mount-  
ing bracket, if desired. Other holes for special mounting can be added to the end cap as desired.  
T he standard inlet and outlet fittings for the HFM-60 are 1/4" NPT to 1/4" Swagelok (optional  
VCR and VCO fittings are available from the factory). T he O-rings for the end cap and the sensor  
are Buna N (optional Kalrez, Neoprene orViton O-rings are available from the factory). It is  
suggested that all connections be checked for leaks after installation. T his can be done by pressuriz-  
ing the instrument (do not exceed 150 psig unless the flowmeter is specifically rated for higher  
pressures) and applying a diluted soap solution to the flow connections.  
2.5  
Electrical Connections  
If a wall transformer power supply from Hastings Instruments is used, installation consists of  
connecting the 4 conductor cable to the 4 pin connector located on the side of the flowmeter.T he  
power is supplied to the instrument on pin 1, +15 to +25VDC, and pin 2, ground. T he output is  
measured on pin 3, 0 - 5VDC or 4-20 mA and pin 4, ground. If a Hastings Model 400/200/40/10  
power supply is used, installation consists of connecting a 4 conductor cable to the terminal strip on  
the rear of the supply to the 4 pin connector located on the side of the flowmeter (similar to that  
described above). Also, a cable with a male “Dconnector can be supplied that mates with the  
power supply. T he “D” connector cable provides +15VDC on pin 11 and the 0-5VDC output on  
pin 6. Pins 5 and 12 are common and pin 7 is chassis ground.  
2.6  
O peration  
T he standard instrument output is a 0 - 5VDC out and the signal is proportional to the flow i.e., 0  
volts = zero flow and 5 volts = 100% of rated flow. T he low noise 4 - 20 mA option is also  
proportional to flow (4 mA = zero flow and 20 mA = 100% of rated flow).  
2.6.1 OperatingTem perature  
For proper operation, the combination of ambient temperature and gas temperature must be such  
that the flowmeter temperature remains between 0 and 50°C. Most accurate measurement of flow  
will be obtained if the flowmeter is zeroed at operating temperature as temperature shifts result in  
some zero offset.  
2.6.2 Zero Check  
Connect the wall transformer power supply to the instrument or if using a Hastings power supply,  
turn the power supply on. Allow approximately 10% flow for 1/2 hour warm-up. Stop all flow  
through the instrument and wait 2 minutes. Caution: Do not assume that all metering valves  
completely shut off the flow. Even a slight leakage will cause an indication on the meter and an  
apparent zero shift. For the standard 0-5VDC output, adjust the zero potentiometer located on the  
lower outlet side of the flowmeter until the meter indicates zero. For the optional 4-20 mA output,  
adjust the zero potentiometer so that the meter indicates slightly more than 4 mA, i.e. 4.03 to 4.05  
mA. T his slight positive adjustment ensures that the 4-20 mA current loop transmitter is not in its  
cut-off region. T he error induced by this adjustment is approximatly 0.3% of full scale. T his zero  
should be checked periodically during normal operation. Zero adjustment is required if there is a  
change in ambient temperature, vertical orientation of the flowmeter, or a change in input voltage.  
2.6.3 Blending of Gases  
If more than one flowmeter is used to mix gases and each is calibrated for a given gas, a desired  
mixture of the gases can be achieved. As an example, suppose a total flow rate of 50 slpm of gases  
A+B+C is required; 15% of the total flow should be gas B, 20% of the total flow should be gas C  
and the balance is gas A. T hen gas A should have a total flow of 0.65 (50) = 32.5 splm, gas B  
should have a total flow of 0.15(50) = 7.5 splm and gas C should have a total flow of 0.20(50) =  
10 splm. Further, let us say that flowmeter A has a 0-60 slpm range, flowmeter B has a 10 slpm  
range and flowmeter C has a 40 slpm range. Since all 3 flowmeters have 0-5VDC out then  
flowmeter A should be set to read 5/60 = A/32.5 or A = 2.71volts, flowmeter B should be set to  
read5/10 = B/7.5 or B = 3.75 volts and flowmeter C should be set to read 5/40 = C/10 or C =  
1.25 volt. Ideally, however, flow controllers are a far better way to accomplish the blending, simply  
because there are no problems with pressure fluctuations.  
page 10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.6.4 Range Changes:  
T he range of the flowmeter can be changed in the field if recalibration facilities are available. T he  
flowmeter may require a different shunt which can be purchased from the factory. A listing of  
available shunts (order codes) and the corresponding air equivalent flow ranges are as follows:  
Order Code  
01  
Shunt Range  
0-10 sccm  
Order Code  
07  
Shunt Range  
0-1 slpm  
02  
03  
04  
05  
0-20 sccm  
0-50 sccm  
0-100 sccm  
0-200 sccm  
08  
09  
10  
11  
0-2 slpm  
0-5 slpm  
0-10 slpm  
0-20 slpm  
06  
0-500 sccm  
12  
13  
0-40 slpm  
0-70 slpm  
Gases other than air or nitrogen have different properties, and the range of each shunt will be  
different than those listed in the order codes above.  
T he flow rate of interest determines the size of the shunt required. As previously indicated, 13  
separate shunts are required for the range of flow spanning 10 sccm to 70 slpm full scale. For the  
200 sccm flow range, four Monel 400 tubes of 0.027" i.d. are located within the aluminum shunt  
plug to accommodate the full 200 sccm. Lower flow rates require proportionally fewer tubes and  
the unused holes are sealed with epoxy (or selected eutectic solder if epoxy is undesired). A single  
304L stainless steel plug with an annular spacing of 0.009" with the base accommodates the 500  
sccm flow range. Increasing the flow requires more annular passageways which is accomplished by  
adding concentric cylinder shells with increasing gap dimensions. Eventually, a maximum annular  
gap dimension for laminar flow is obtained (~0.030"). For a maximum flow rate of 60 slpm, three  
annular regions using two brass concentric cylinders is required. To change ranges, a new shunt  
must be installed and a calibration conducted to do further measurements.  
2.6.5 Spanning the Digital Display  
An exchange of shunt requires resetting of the digital display. Ideally, the flow should be estab-  
lished to the full scale level consistent with the selected shunt and the display pot adjusted for that  
full scale flow rate. It is also possible to disconnect the phone jack connector to the main body and  
input 5.00VDC and set the display consistent with the particular full scale flow for the selected  
shunt. Verify the full scale voltage/current output and adjust the display pot for that full scale flow  
rate. It is also possible to disconnect the phone jack between the display and flowmeter, and supply  
external power (+5VDC) and a full scale voltage (0-5VDC) to the display. Refer to the schemat-  
ics of PCB-846 in section 6 of this manual. T he display may then be adjusted consistent with the  
particular full scale flow for the selected shunt. Note that there may be need to change the display  
board solder jumpers in order to exhibit the new full scale flow rate.  
Largest D isplay Value  
Jumper Setting  
All jumpers open  
JP 1 closed  
1999.  
199.9  
19.99  
JP 2 closed  
1.999  
JP 3 closed  
page 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 3  
Theory of Operation  
T his section contains an overall functional description of the HFM-60. Detailed schematics and  
parts lists can be found at the end of the manual in Section 5.0. In this section and other sections  
throughout this manual, it is assumed that the customer is using the wall transformer power supply  
supplied by Hastings or a Hastings Power Supply.  
3.1  
O verall Functional D escription  
T he HFM-60 consists of a sensor, a base, shunt and electronic circuitry. T he sensor measures the  
gas flow rate from 0 to 10 sccm. T he shunt divides the flow such that the flow through the sensor  
is a precise percentage of the flow through the shunt. T he flow through both the sensor and shunt  
is laminar. T he circuit board amplifies the sensor output from the two Chromel P thermocouples  
and and provides an analog output of either 0-5VDC or 4-20 mA. T he output can also be  
indicated on the digital display.  
3.2  
Sensor  
T he Hastings HFM-60 operates on a unique thermal electric principle whereby a metallic capillary  
tube is heated uniformly by a resistance winding attached at the center of the capillary (see figure  
3.1). T hermocouplesT C-1 andT C-2 are positioned on each side of the heater coil and are welded  
at equal distances from the mid-point of the capillary. At zero flow, theT C outputs are equal.  
When flow occurs through the tubing, heat is transferred from the heated region to the gas and  
from the gas back to the cool wall downstream. T hermocoupleT C-1 measures the temperature  
upstream before the gas is heated and thermocoupleT C-2 measures the gas temperature down-  
stream after it is heated. During flow, the temperature distribution along the sensor tube becomes  
skewed (see figure 3.2). Conventionally, the difference in temperature of the two thermocouples  
provides a voltage difference that is proportional to flow.  
Heater (VDC Source)  
TC-2  
TC-1  
O-ring  
PC Board  
Nylon  
Spacer  
flow  
figure 3.1  
For a constant power input, the differential voltage is a function of the mass flow rate and the heat  
capacity of the gas. Since the heat capacity of many gases is relatively constant over wide ranges  
of temperature and pressure, the flowmeter may be calibrated directly in mass units for those gases.  
Changes in gas composition usually require application of a multiplication factor to the air calibra-  
tion to account for the difference in heat capacity. T he flowmeter is capable of measuring a wide  
range of gases. T he Monel 400 sensor tube has an internal diameter of 0.0205" and is heated by  
600W of MWS-800 wire. T his diameter allows approximately 10 sccm full scale flow with p of  
0.5 inches of H2O. T he sensor is encapsulated by a plastic cover and filled with fiberglass insulation  
to minimize convective heat losses.  
page 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ZERO  
FLOW  
SMALL  
FLOW  
TC-1  
TC-2  
L/2  
0
L/2  
Length of tube  
figure 3.2  
3.3  
3.4  
Base  
T he 6061 aluminum alloy base has a 1" square cross-section and is 3.375" long (with end cap).  
T he internal flow channel is 0.812" diameter with in and out 304L stainless steel fittings 1/4" NPT  
to 1/4" Swagelok fittings.  
Shunt  
T he flow rate of interest determines the size of the shunt required. As previously indicated, 13  
separate shunts are required for the range of flow spanning 10 sccm to 70 slpm full scale. T he  
geometry of the shunt for 0- 200 sccm is shown in figure 3.3(a). Four Monel 400 tubes of 0.027"  
i.d. are located within the aluminum plug to accommodate the full 200 sccm. Lower flow rates  
require proportionally fewer tubes and the unused holes are sealed with epoxy (or selected eutectic  
solder if epoxy is undesired). At the lowest flow rate (10 sccm), the shunt is completely blocked and  
all of the flow is routed through the sensor. Figure 3.3(b) shows the geometry for higher flow rates.  
A single 304L stainless steel plug with an annular spacing of 0.009" accommodates the 500 sccm flow  
range. Increasing the flow requires more annular passageways which is accomplished by adding  
concentric cylinder shells with increasing gap dimensions. Eventually, a maximum annular gap  
dimension for laminar flow is obtained (~0.030"). For a maximum flow rate of 70 slpm, three  
annular regions using two brass concentric cylinders is required. T his patent pending shunt technol-  
ogy also includes inboard sensor ports which ensure laminar flow without the turbulence associated  
with end effects. T his unique flow geometry provides an exceedingly linear shunt.  
A schematic diagram of the integral flowmeter (without the digital display module) employing a 70 slpm  
shunt is shown in figure 3.4. Note the inboard location of the sensors inlet and outlet with respect to  
the inlet and outlet of the shunt. Also note the screen disk and washer which are used to minimize the  
radial velocity gradient entering the shunt. T his results in a more uniform flow velocity profile and  
therefore, more uniform flow in the annular passages of the shunt.  
page 14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
epoxy  
flow  
Aluminum  
10 sccm - 200 sccm  
(typical configuration)  
figure 3.3 (a)  
annular flow  
passages  
flow  
500 sccm - 70 slm  
(typical configuration)  
figure 3.3 (b)  
page 15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.5  
Electronic Circuitr y  
T he HFM-60 employs a thermal flow sensor (capillary tube described in section 3.2) to measure  
the flow which is proportional to the total flow through the instrument. T he sensor develops a  
differential voltage output signal proportional to flow which is approximately 1 mV full scale  
magnitude. A differential amplifier and an inverting amplifier boost the signal strength by a factor  
of ~5000 at full scale flow. T he amplified output can be measured on pins 3 and 4 of the external  
connector. If a Hastings power supply is employed, the 5 volt output is also sent to the terminals  
on the back and to the decoding circuitry in the display which converts it to a 3 and 1/2 digit  
output. An optional 4-20 mA analog output on pins 3 and 4 is also available in lieu of an output  
voltage. T he addition of a 4-20 mA current loop transmitter on the secondary pc board (mounted  
parallel to the main pc board) is required to provide this current loop. A jumper change is made on  
the main PC board to establish the selected output mode. T he digital display option includes the  
tiltable LCD display module which is removable for remote display. T his separate module contains  
its own pc board and mounts directly on the top of the main cover. T his local or remote digital  
unit provides a continual digital display in addition to the 0-5VDC or 4-20 mA output. A standard  
4 conductor phone jack and cable is used to connect the display module to the main pc board. T he  
circuit diagrams for the aforementioned functions are presented in Section 6.  
T his section contains service and calibration information. Some portions of the instrument are  
delicate. Use extreme care when servicing the instrument. T he potentiometer positions and the  
electrical components referred to in the troubleshooting section can be found in Section 4.3 on the  
electrical component layout drawing.  
cover  
PC  
board  
flow  
shunt plug  
(60 slm shown)  
end cap  
fine  
screen  
base  
shunt  
tubes  
O-ring  
coarse  
screen  
spacer  
figure 3.4  
page 16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 4  
Maintenance  
4.1  
4.2  
Authorized Maintenance  
With proper care in installation and use, the flowmeter will require little or no maintenance. If  
maintenance does become necessary, most of the instrument can be cleaned or repaired in the field.  
Some procedures may require recalibration. Do not attempt these procedures unless facilities are  
available. Entry into the sensor or tampering with the printed circuit board will void warranty. Do  
not perform repairs on these assemblies while the unit is still under warranty.  
Troubleshooting  
Sym ptom : Output reads strong indication of flow with no flow present. Zero pot has no effect.  
C ause:  
Action:  
Power shorted out.  
Unplug power supply for a few seconds, then plug it back in. If this is ineffective,  
disconnect the power supply from the unit and check the power supply voltage  
(+15 to +25 VDC on pin 1).  
Symptom: Output of flowmeter is proportional to flow, but extremely small and not correctable  
by span pot.  
C ause:  
Action:  
Sensor is not being heated.  
Shut off gas supply and disconnect the power to the flowmeter. Remove cover and pc  
board from unit. Check the resistance between w1 and w2 on PCB-844 (refer to  
figures in section 6) of the flowmeter. T he resistance should be approximately 560 W.  
T he resistance between w3 and w4 should be approximately 2-3W. If not, the  
sensor unit needs to be replaced.  
Symptom: Sensor has proper resistance readings, but little or no output with flow.  
C ause:  
Action:  
Plugged sensor.  
Shut off gas supply and disconnect the power to the flowmeter. Remove cover and pc  
board from unit. If sensor has evidence of clogging, clean or replace as applicable.  
Symptom: Flowmeter reads other than 0.00VDC with no flow or there is a small flow when the  
flowmeter reads 0.00VDC.  
C ause:  
Action:  
Zero pot is out of adjustment.  
Shut off all flow. For the standard 0-5VDC output, adjust the zero potentiometer  
located on the lower outlet side of the flowmeter until the meter indicates zero. For  
the optional 4-20 mA output, adjust the zero potentiometer so that the meter  
indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. T his slight positive  
adjustment ensures that the 4-20 mA transmitter is not in its cut-off region. T he  
error induced by this adjustment is approximately 0.3% of full scale.  
Symptom: Flowmeter is out of calibration and non-linear.  
C ause:  
Action:  
Leaks in the gas inlet or outlet fittings.  
Check all fittings for leaks by placing soap solution on all fittings between gas supply and  
final destination of gas. Check flowmeter for leaks. Replace“O”-rings if required or  
recalibrate as necessary.  
page 17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.3  
Adjustm ents  
Calibration P rocedure (See figure 4.1)  
1. Connect power to pins 1 and 2 as specified in Section 2.5. Allow the instrument to warm up  
for 30 minutes with 10% flow.  
2. Completely shut off the flow and wait for 2 minutes. For the standard 0-5VDC output,  
adjust the zero potentiometer located on the lower outlet side of the flowmeter until the  
meter indicates zero. For the optional 4-20 mA output, adjust the zero potentiometer so that  
the meter indicates slightly more than 4 mA, i.e. 4.03 to 4.05 mA. T his slight positive  
adjustment ensures that the 4-20 mA transmitter is not in its cut-off region. T he error  
induced by this adjustment is approximately 0.3% of full scale.  
3. Turn on gas supply to inlet of instrument and insure the flow rate is an accurate maximum  
flow for the shunt employed. Ensure that the full range can still be obtained at minimum  
inlet pressure.  
4. Adjust Span pot until the flow reference reads full scale (5.00VDC or 20 mA). Perform this  
step only if a calibrated reference flowmeter is available.  
5. Adjust the span pot of the display so that it reads correct full scale flow.  
6. Record flowmeter and flow reference outputs for flow rates of 20%, 40%, 60%, 80% and  
100% and make sure data are within ± 3% of full scale.  
flow  
span  
zero  
figure 4.1  
page 18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.4  
End Cap Rem oval  
T he end cap on the inlet side must be removed to gain access to the filter and shunt assembly.  
First remove power and shut off the supply of gas to the instrument. Disconnect the Swagelok  
fittings on the inlet and outlet sides of the transducer and remove it from the system plumbing.  
Remove the two Allen head screws holding the end cap to the instrument (see figure 4.2). Care-  
fully remove the end cap, retaining ring, filter, spacer, filter washer and ball bearing (if present) and  
shunt, noting their order and proper orientation. T he shunt can be severely damaged if dropped.  
Examine the filter and shunt. If either is damaged, dirty or blocked, clean and replace as applicable.  
Reassemble in the reverse order of disassembly. Secure the endcap with 4 in lb of torque on each  
18-8 stainless steel socket head cap screw. Use of a fastener other than the one mentioned here  
may result in leakage at the seal. Recalibration of the HFM-60 is necessary.  
60 slm  
40 slm  
10 sccm to 200 sccm  
figure 4.2  
page 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.5  
P rinted Circuit Board Replacem ent  
In the event that any of the PC boards fail, they are easily removed from the instrument and  
replaced with a spare. T his ease in disassembly and replacement substantially reduces instrument  
downtime.  
1. Replacement of the 4-20 mA option PC board: Unplug the power cable from the transducer.  
If a digital display exists, unplug the phone jack connection and remove the flowmeter cover.  
Remove the 4 brass spacers and lift off the 4-20 mA board. Be careful not to damage the  
main board and 4-20 mA board connector.  
2. Replacement of the main PC board: Continue from 1. Remove the 4 brass spacers from the  
main board and gently lift the board from the flowmeter base. When installing the replacement  
board, insure that the brass spacers are uniformly and securely tightened (finger tight) since the  
compression seals the “O”-rings on the base, PC board and the sensor tube which, in turn,  
maintains the integrity of the flow in the sensor.  
4.6  
Sensor Replacem ent  
T he sensor is an intimate part of the main PC board and can not be separately removed from the  
main board without great difficulty. T herefore, sensor replacement requires the replacement of the  
entire main board. Follow instructions for removing the main board as shown in Section 4.5 and  
consult Section 5 to acquire a new main board/sensor unit.  
page 20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 5  
Warranty and Repair  
5.1 War ranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment to be free  
from defects in material and workmanship. T his warranty does not apply to defects or failures resulting from  
unauthorized modification, misuse or mishandling of the product. T his warranty does not apply to batteries  
or other expendable parts, nor to damage caused by leaking batteries or any similar occurrence. T his  
warranty does not apply to any instrument which has had a tamper seal removed or broken.  
T his warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to  
fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential  
damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if Hastings  
Instruments determines, in good faith, that it is defective in materials or workmanship during the warranty  
period. Defective instruments should be returned to Hastings Instruments, shipm ent prepaid, together  
with a written statement of the problem and a Return Material Authorization (RMA) number. Please consult  
the factory for your RMA number before returning any product for repair. Collect freight will not be  
accepted.  
5.2 Non-War ranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form and  
a written description of the problem with the instrument. If the repair cost is higher, you will be contacted  
for authorization before we proceed with any repairs. If you then choose not to have the product repaired, a  
minimum will be charged to cover the processing and inspection. Please consult the factory for your RMA  
number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
(757) 723-6531  
1-800-950-2468  
FAX  
(757) 723-3925  
EMAIL  
hastings_instruments@teledyne.com  
INTERNET ADDRESS http:/ / www.hastings-inst.com  
Repair Forms may be obtained from the “Information Request” section of the  
Hastings Instruments web site.  
page 21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 6  
Diagrams and Drawings  
page 23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
page 30  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Telex Stereo System ANT FP User Manual
Texas Instruments Stereo Amplifier TPA3100D2 User Manual
Tokina Camera Accessories TVR3314DC User Manual
Toshiba Projection Television TLP T720 User Manual
Traulsen Refrigerator TR35772 User Manual
Tripp Lite Switch U222 004 R User Manual
Ultra Products Network Card ULT31344 User Manual
Uniden Cordless Telephone WDSS 5305 User Manual
Video Products Computer Hardware VGACMP TV User Manual
Wayne Water Pump 321602 001 User Manual