Installation Manual
TLA 700 Series
Logic Analyzer
070-9774-02
Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
Download from Www.Somanuals.com. All Manuals Search And Download.
TEKTRONIX SERVICE OPTIONS
Tektronix offers the following service options. These options are modular, flexible, and easy to order with your instru-
ment. Designed to ease installation and start up, to support tracking of calibration to requirements of ISO9000, and
to provide for extended repair coverage, these options help fix your long-term maintenance costs and eliminate un-
planned expenditures. And these options can be converted from service at Tektronix service depots to service on-site
(see Option S3), which helps keep downtime to a minimum.
Product installation service1
Three years of calibration services
Test data
Option IN
Option C3
Option D3
Option R3
Option S1
Option S3
Provides initial installation of the product and
familiarizes new users with some of its opera-
tion features
Provides initial Certification on delivery, plus
two more annual calibrations from your ser-
vice center
Provides test data on delivery and when annu-
al calibration services are provided (three to-
tal, requires Option C3)
Three years repair coverage
One year uplift to on–site service1,2
Three year uplift to on-site service1,2
Provides three years of repair coverage for the
instrument, including displays and accessories
sold with the instrument
Converts the standard one year, “return to de-
pot” warranty to provide service on site for
one year
Converts any C3, D3, and R3 options pur-
chased to on-site services for three years
Tektronix Service Options are available at the time you order your instrument. Contact your local Tektronix Sales
Office for more information.
1
Availability of installation and on-site services depends on the type of product and may vary by geography.
2
Uplift options are ordered with the mainframe products and cover individual modules.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
v
vii
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Offerings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ix
x
xi
xii
Basic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Check the Shipping List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Check the Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Install the Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cover Empty Slots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connect the Standard and Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connect the Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power On the Logic Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Powering Off the Logic Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Perform the Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Create an Emergency Startup Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Back Up User Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
1
2
5
6
7
9
10
12
12
13
13
Installing Microprocessor Support Software . . . . . . . . . . . . . . . . . . . .
15
The Next Step . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Online Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TLA 700 Series QuickStart Training Board and Manual . . . . . . . . . . . . . . . . . . .
Other Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17
17
17
17
In Case of Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hardware Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TLA 700 Startup Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing Podlets on the P6417 Logic Analyzer Probe . . . . . . . . . . . . . . . . . . .
19
19
20
20
23
25
Appendix A: Merging Logic Analyzer Modules . . . . . . . . . . . . . . . . . .
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Merge Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installing the Merged Modules in the Logic Analyzer . . . . . . . . . . . . . . . . . . . .
Removing Merged Modules from the Logic Analyzer . . . . . . . . . . . . . . . . . . . .
Storing the Merge Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
33
33
37
38
38
Appendix B: Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Upgrading Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installing TPI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Updating Module Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Controller BIOS Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reinstalling Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Customizing TLA 700 Series Setups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
41
41
43
50
52
56
61
Appendix C: Power Cord and Line Fuse Requirements for the
Benchtop Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
65
i
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
List of Figures
Figure 1: Location of ground connections . . . . . . . . . . . . . . . . . . . . . . .
Figure 2: Installing modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3: Installing the empty slot panel covers . . . . . . . . . . . . . . . . . .
Figure 4: Locations of external connectors . . . . . . . . . . . . . . . . . . . . . .
Figure 5: Connecting the logic analyzer probes . . . . . . . . . . . . . . . . . .
Figure 6: Line fuse and power cord connector locations . . . . . . . . . . .
Figure 7: On/Standby switch locations . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 8: TLA 700 startup sequence . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9: Moving probe podlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10: Disassembling the probe cable . . . . . . . . . . . . . . . . . . . . . . .
4
5
6
7
9
10
11
24
25
27
Figure 11: Removing the faulty channel from the probe cable
header housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28
29
30
31
32
34
35
36
37
39
50
66
Figure 12: Removing a podlet cable from the rubber comb . . . . . . . .
Figure 13: Replacing a podlet in the podlet holder . . . . . . . . . . . . . . .
Figure 14: Installing the color-coded ring . . . . . . . . . . . . . . . . . . . . . . .
Figure 15: Installing the cable in the cable header housing . . . . . . . .
Figure 16: Removing the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 17: Feeding the Merge cable through the cover . . . . . . . . . . . .
Figure 18: Seating the cover on the chassis . . . . . . . . . . . . . . . . . . . . . .
Figure 19: Lining up the two modules . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 20: Dressing the Merge cable before installing the cover . . . .
Figure 21: Flash programming pins . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 22: Power cord identification chart . . . . . . . . . . . . . . . . . . . . . .
ii
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
List of Tables
Table 1: Mainframe operating requirements . . . . . . . . . . . . . . . . . . . .
3
8
Table 2: Additional accessory connection information . . . . . . . . . . . .
Table 3: Line fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 5: Software not covered by the TLA 700 Series software setup
Table 6: Controller BIOS setups for the TLA 700 application . . . . .
Table 7: Power for instrument modules . . . . . . . . . . . . . . . . . . . . . . . .
10
21
41
53
65
iii
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
iv
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the instrument.
Read the General Safety Summary in other system manuals for warnings and
cautions related to operating the system.
To Avoid Fire or
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is indirectly grounded through the grounding
conductor of the mainframe power cord. To avoid electric shock, the grounding
conductor must be connected to earth ground. Before making connections to the
input or output terminals of the product, ensure that the product is properly
grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marking on the product. Consult the product manual for further ratings
information before making connections to the product.
The common terminal is at ground potential. Do not connect the common
terminal to elevated voltages.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
v
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
Protective Ground
(Earth) Terminal
CAUTION
Refer to Manual
Double
Insulated
vi
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Caution When Servicing the CRT. To avoid electric shock or injury, use
extreme caution when handling the CRT. Only qualified personnel familiar with
CRT servicing procedures and precautions should remove or install the CRT.
CRTs retain hazardous voltages for long periods of time after power is turned off.
Before attempting any servicing, discharge the CRT by shorting the anode to
chassis ground. When discharging the CRT, connect the discharge path to ground
and then the anode. Rough handling may cause the CRT to implode. Do not nick
or scratch the glass or subject it to undue pressure when removing or installing it.
When handling the CRT, wear safety goggles and heavy gloves for protection.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
X-Radiation. To avoid x-radiation exposure, do not modify or otherwise alter the
high-voltage circuitry or the CRT enclosure. X-ray emissions generated within
this product have been sufficiently shielded.
vii
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Service Safety
viii
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Preface
The TLA 700 Series Installation Manual is intended to help you install your
TLA 700 Series Logic Analyzer. Although your TLA 700 Series Logic Analyzer
may arrive with all modules installed, you should still follow the steps in this
manual to complete the installation process.
The manual consists of the following sections:
H
H
Basic Installation describes the hardware installation process and provides
information to start the logic analyzer the very first time.
Installing Microprocessor Support Software describes the steps for installing
microprocessor support software that does not come installed on your logic
analyzer.
H
H
The Next Step provides information on other TLA 700 Series documentation
that you may need to operate your TLA 700 Series Logic Analyzer.
In Case of Problems provides information on handling problems or
difficulties that may occur during the installation process. It also provides
procedures for replacing probe podlets and a list of phone numbers for
Tektronix product support.
H
H
Appendix A: Merging Logic Analyzer Modules provides the procedures for
combining two logic analyzer modules to create a single, wider module.
Appendix B: Upgrades provides information on upgrading software,
installing TPI, (TLA 700 Programmatic Interface), upgrading firmware for
the individual modules, and for using the BIOS setup for the controller in
your logic analyzer.
H
Appendix C: Power Cord and Line Fuse Requirements for the Benchtop
Mainframe provides information on selecting the proper power cord and line
fuse for the benchtop mainframe.
ix
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Preface
Related Documentation
In addition to this installation manual, the following documentation is available
for your TLA 700 Series logic analyzer:
H
The TLA 700 Series Logic Analyzer User Manual provides basic operating
information for the TLA 700 Series Logic Analyzer.
H
H
H
The online help provides information about the TLA 700 Series user
interface. To view the online help, select Help Topics from the Help menu.
A series of microprocessor support instruction manuals provide operating
and service instructions for the individual microprocessor support packages.
The TLA 7QS Training Manual provides training exercises to help you learn
key features of the TLA 700 Series Logic Analyzer. The training manual is
designed to be used with the TLA 700 Series QuickStart training board.
H
H
The TLA 7QS Technical Reference Manual provides technical and service
information for the TLA 700 Series QuickStart training board.
The TLA 700 Series Performance Verification and Adjustment Technical
Reference Manual provides performance verification and adjustment
procedures for the major components of the TLA 700 Series Logic Analyzer.
The manual includes the performance verification and adjustment software.
H
H
H
H
H
The TLA 711 Color Benchtop Chassis Service Manual provides module-lev-
el service information for the benchtop chassis.
The TLA 711 Color Benchtop Controller Service Manual provides module-
level service information for the benchtop mainframe controller module.
The TLA 704 Color Portable Mainframe Service Manual provides module-
level service information for the portable mainframe.
The TLA 7Lx/TLA 7Mx Logic Analyzer Service Manual provides module-
level service information for the logic analyzer module.
The TLA 7Dx/TLA 7Ex Digitizing Oscilloscope Service Manual provides
module-level service information for the DSO module.
x
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Preface
Service Offerings
Tektronix provides service to cover repair under warranty as well as other
services that are designed to meet your specific service needs.
Whether providing warranty repair service or any of the other services listed
below, Tektronix service technicians are well equipped to service the TLA 700
Series Logic Analyzers. Tektronix technicians train on Tektronix products; they
have access to the latest information on improvements to the TLA 700 Series
Logic Analyzers as well as the latest new product upgrades. Services are
provided at Tektronix Services Centers and on-site at your facility, depending on
your location.
Warranty Repair Service
Tektronix warrants this product for one year from date of purchase. (The
warranty appears behind the title page in this manual.) Tektronix technicians
provide warranty service at most Tektronix service locations worldwide. The
Tektronix product catalog lists all service locations worldwide or you can visit us
surement/Service. See our latest service offerings and contact us by email.
Calibration and Repair
Service
In addition to warranty repair, Tektronix Service offers calibration and other
services which provide cost-effective solutions to your service needs and quality-
standards compliance requirements. Our instruments are supported worldwide by
the leading-edge design, manufacturing, and service resources of Tektronix to
provide the best possible service.
The following services can be tailored to fit your requirements for calibration
and/or repair of the TLA 700 Series Logic Analyzers.
Service Options. Tektronix Service Options can be selected at the time you
purchase your instrument. You select these options to provide the services that
best meet your service needs. These service options are listed on the Tektronix
Service Options page following the title page of this manual.
Service Agreements. If service options are not added to the instrument purchase,
then service agreements are available on an annual basis to provide calibration
services or post-warranty repair coverage for the TLA 700 Series Logic
Analyzers. Service agreements may be customized to meet special turn-around
time and/or on-site requirements.
Service on Demand. Tektronix also offers calibration and repair services on a
“per-incident” basis that is available with standard prices for many products.
xi
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Preface
Self Service. Tektronix supports repair to the replaceable-part level by providing
for circuit board exchange.
Use this service to reduce down-time for repair by exchanging circuit boards for
remanufactured ones. Tektronix ships updated and tested exchange boards. Each
board comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales
engineer for more information on any of the Calibration and Repair Services just
described.
Contacting Tektronix
Product
Support
For application-oriented questions about a Tektronix measure-
ment product, call toll free in North America:
1-800-TEK-WIDE (1-800-835-9433 ext. 2400)
6:00 a.m. – 5:00 p.m. Pacific time
Or contact us by e-mail:
For product support outside of North America, contact your
local Tektronix distributor or sales office.
Service
Support
Contact your local Tektronix distributor or sales office. Or visit
our web site for a listing of worldwide service locations.
In North America:
For other
information 1-800-TEK-WIDE (1-800-835-9433)
An operator will direct your call.
To write us Tektronix, Inc.
P.O. Box 1000
Wilsonville, OR 97070-1000
xii
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
This chapter describes all of the steps needed to install the TLA 700 Series Logic
Analyzer. It is written from the perspective that you purchased most of the items
uninstalled and you intend to install all of the pieces. If you purchased a logic
analyzer with modules already installed, you should still review all of the
information in this chapter and perform those steps that apply to your specific
situation.
This chapter deals mainly with hardware installation. The basic operating
software is already installed on the internal hard disk. However, if you ordered
additional software, such as microprocessor support software, you must install
that software yourself. Refer to Installing Microprocessor Support Software on
page 15.
Check the Shipping List
Verify that you have received all of the parts of your logic analyzer. Use the
shipping lists that came with your logic analyzer to compare against the actual
contents of your order. You should also verify the following:
H
H
Verify that you have the correct power cords for your geographical area.
Verify that you have backup copies of the installed software. Store the
backup software in a safe location where you can easily retrieve the software
for maintenance purposes.
NOTE. Keep the software packaging available because you will need it to enter
the Windows 95 software registration number when you first power on the logic
analyzer. (See Step 6 under Power on the Logic Analyzer on page 11.)
H
H
Verify that you have the correct module types and probes.
Verify that you have all the standard and optional accessories that you
ordered.
Remember to fill out and send in the customer registration card. The registration
card is packaged with this manual.
1
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Check the Environmental Requirements
Read this section before attempting any installation procedures. This section
describes site considerations, power requirements, and ground connections for
your logic analyzer.
Site Considerations:
Portable Mainframe
The portable mainframe is designed to operate on a bench or on a cart in the
normal position (on the bottom feet). For proper cooling, at least two inches
(5.1 cm) of clearance is recommended on the rear and sides of the mainframe.
You can also operate the mainframe while it rests on the rear feet. If you operate
the mainframe while it is resting on the rear feet, make sure that you properly
route any cables coming out of the rear of the mainframe to avoid damaging
them.
CAUTION. Keep the bottom of the mainframe clear of obstructions to ensure
proper cooling.
Site Considerations:
Benchtop Mainframe
The benchtop mainframe is designed to operate on a bench or in a rackmount
environment. For proper cooling, at least two inches (5.1 cm) of clearance is
recommended on the rear and sides of the mainframe.
WARNING. Because of the size and weight of the benchtop mainframe, use care
when lifting or moving the mainframe to avoid personal injury while performing
the installation procedures. Use two people to lift or move the mainframe.
2
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Operating Requirements
Table 1 lists the operating requirements for the logic analyzers. Refer to the
TLA 700 Series Logic Analyzer User Manual for a complete specification list.
Table 1: Mainframe operating requirements
Requirement
TLA 704 Color Portable mainframe
90–250 V , 45–66 Hz, continuous range
TLA 711 Color Benchtop mainframe
90–250 V , 45–66 Hz, continuous range
Source voltage
RMS
RMS
100–132 V , 360–440 Hz, continuous range 100–132 V , 360–440 Hz, continuous range
RMS
RMS
Steady state input current
6 A , maximum
15 A , maximum at 90 VAC
RMS
RMS
6.3 A , maximum at 207 VAC
RMS
Maximum power consumption
600 W
1350 W line power
Temperature
(No media in floppy disk drive)
+5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
+5° C to +50° C (+41° to +122° F), 15° C/hr
(55° F/hr) maximum gradient, noncondensing
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
Humidity
(No media in floppy disk drive)
20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
20% to 80% relative humidity, noncondensing.
Maximum wet bulb temperature: +29° C
(84° F) (derates relative humidity to ~22% @
50° C (+122° F))
Altitude
To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
To 10,000 ft (3,040 m)
(derated 1° C (34° F) per 1000 ft (304.8 m)
above 5000 ft (1524 m))
3
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Chassis Ground
Connections
Figure 1 shows chassis ground connections. Use the chassis ground connections
to connect the grounds of one or more instruments to the logic analyzer to ensure
a common ground connection between instruments.
WARNING. Do not remove the safety ground screw from the benchtop mainframe.
The safety ground screw must always be in place to ensure the proper bonding of
the power supply to the mainframe.
Chassis ground
Safety ground
(Benchtop
mainframe only)
Figure 1: Location of ground connections
4
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Install the Modules
CAUTION. Do not install or remove any modules while the mainframe is powered
on; doing so can damage the modules or the logic analyzer.
If your logic analyzer does not have the modules preinstalled, install the modules
as shown in Figure 2. You can install the modules in any slot. Use a Phillips
screwdriver with a Number 1 tip to tighten the retaining screws after seating the
modules in place.
If you want to merge two logic analyzer modules to create a wider module, refer
to Appendix A: Merging Logic Analyzer Modules on page 33.
Portable mainframe
Benchtop mainframe
Retaining screws
Injector/ejector handles
Figure 2: Installing modules
5
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Cover Empty Slots
If you have any unused (empty) slots in your mainframe, you must install blank
slot panel covers to meet EMC specifications. Install a blank slot panel cover for
each empty slot as shown in Figure 3. Make sure that the EMI shielding is in
contact with the adjacent panel or module cover. Installing the covers provides
proper EMI shielding and reduces audible noise.
CAUTION. Use only Tektronix TLA 700 Series slot panel covers on the benchtop
mainframe. Do not use non-Tektronix covers, otherwise the mainframe may not
meet cooling and EMC requirements.
EMI shielding strip
Double slot panel cover
EMI shielding strip
Single slot panel cover
Double slot panel cover
Figure 3: Installing the empty slot panel covers
6
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Connect the Standard and Other Accessories
The accessory connections are the same as those you would make on a personal
computer. The connection points are shown in Figure 4. Use the icons that
appear in the circled areas as a guide. See Table 2 on page 8 for additional
connection information.
Icon
Description
Monitor
RS-232
Printer
Mouse
Keyboard
Floppy
disk drive
PC card
(LAN or other)
Figure 4: Locations of external connectors
7
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Table 2: Additional accessory connection information
Item
Description
Monitor
If you use a non-standard monitor, you may need to change the
the Windows 95 display settings to achieve the proper
resolution for your monitor. Refer to Set the Display Driver on
page 58 or to Adjust the Display Desktop Settings for an
External Monitor on page 62. You may also want to refer to the
documentation for your display monitor for installation
information.
Printer
Connect the printer to the EPP (enhanced parallel port)
connector directly. If your printer has a DB–25 connector, use
the adapter cable that came with your logic analyzer to connect
to the EPP connector.
Rackmount
Other
Refer to the TLA 711 Benchtop Mainframe Rackmount
Installation Instructions for information on installing the
rackmount kit.
Refer to the TLA 700 Series Application online release notes
for additional accessory installation information not covered in
this manual. You should also refer to the user or installation
documentation that came with your accessory.
8
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Connect the Probes
Connect the logic analyzer probes and the optional retaining brackets as shown
in Figure 5. Then connect the oscilloscope probes.
Match color-coded labels
Attach optional
probe retainer
brackets
Figure 5: Connecting the logic analyzer probes
9
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Power On the Logic Analyzer
Follow these steps to power on the logic analyzer for the first time.
1. Check that the line fuse is correct for your application. See Table 3 and
Figure 6.
Table 3: Line fuses
Line voltage
Rating
Tektronix part number
Portable mainframe
90 V to 132 V operation
207 V to 250 V operation
Benchtop mainframe
90 V to 132 V operation
103 V to 250 V operation
207 V to 250 V operation
8 A, fast blow, 250 V
6.3 A, fast blow, 250 V
159-0046-00
159-0381-00
20 A, slow blow, 250 V
15 A, fast blow, 250 V
6.3 A, fast blow, 250 V
159-0379-00
159-0256-00
159-0381-00
AC power and fuse
Figure 6: Line fuse and power cord connector locations
CAUTION. Connect the keyboard, mouse, and other accessories before applying
power to the mainframe. Connecting the accessories after powering on the
mainframe can damage the accessories. Refer to Figure 4 on page 7 for
connection information.
10
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
2. Connect the power cord.
CAUTION. Although the benchtop mainframe can use the power cord with the
15 A plug, mainframes operating at low line (90 VAC) with four or more
instrument modules may require the power cord with the 20 A plug. If you have
four or more modules in your mainframe, you must determine the correct fuse
and power cord combination to avoid overloading the power distribution system.
Refer to Power Cord and Line Fuse Requirements for the Benchtop Mainframe
on page 65 for further information.
3. If you have an external monitor, connect the power cord and power on the
monitor.
4. Press the On/Standby switch to power on the logic analyzer (see Figure 7 for
the switch location).
Figure 7: On/Standby switch locations
5. Wait for the boot routine and low-level self test to complete.
6. Follow the instructions on the screen and enter the Windows 95 Product
Identification number (located above the bar code) on the Windows 95
software package that came with your logic analyzer. Enter other information
as required.
The internal setup software will automatically configure your logic analyzer
and install all required devices depending on the installed accessories.
7. Restart your logic analyzer when the setup program is complete.
The next time you power on the mainframe, the logic analyzer application starts
automatically.
11
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Powering Off the Logic Analyzer
The portable mainframe has a built-in soft power-off function that safely powers
off the mainframe when you press the On/Standby switch. If you have the
benchtop mainframe, you must power off the mainframe using the Windows 95
shutdown process before depressing the On/Standby switch.
CAUTION. When powering off the benchtop mainframe, use the Windows 95 shut
down procedure. Powering off the benchtop mainframe prematurely can corrupt
the software on the hard disk.
Perform the Incoming Inspection
Incoming inspection consists of verifying the basic operation of the logic
analyzer. The Power-on diagnostics check the basic functionality. These
diagnostics run every time you power on the logic analyzer.
You can also verify more detailed functionality by running the self calibration
and extended diagnostics.
NOTE. Allow a 30-minute warm-up before running the self calibration.
Disconnect any attached probes from the modules. Then select the System menu,
and point to Calibration and Diagnostics. Run the self calibration followed by
the extended diagnostics by selecting the proper tab. Results of the tests display
on the individual property page.
NOTE. The time required to run the self calibration on the logic analyzer
modules depends on the number of acquisition channels. Modules with a large
number of channels may take several minutes to run the self calibration.
If you want to perform detailed performance verification procedures, refer to the
TLA 700 Series Performance Verification and Adjustment Technical Reference
Manual (available as an optional accessory).
Checking the Logic
Analyzer Probes
(Optional)
Connect the logic analyzer probes to a signal source, start an acquisition, and
verify that the acquired data is displayed in either the listing or waveform
windows.
12
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
Checking the DSO Probes
(Optional)
Connect the oscilloscope probes to the Probe Compensation connector on the
front panel of the DSO module. You can then run the Calibrate Probe function in
each vertical setup page for the module.
Checking the Mainframe
(Optional)
To check the mainframe diagnostics not covered by the TLA 700 Series
Application software. Run the QAPlus diagnostics or the TLA 700 Mainframe
Diagnostics located under the Windows 95 Start menu under the Tektronix
TLA 700 programs. Exit the TLA 700 Series Application before running the
external diagnostics.
Create an Emergency Startup Disk
Now that you have completed the basic installation process, you should create an
emergency startup disk that you can use to restart your logic analyzer in case of a
major hardware or software failure. You should create this disk and then store it
in a safe place.
The emergency startup disk contains basic files to restart your logic analyzer. It
also contains files to check and format the hard disk.
Follow these steps to create the emergency startup disk:
1. Exit the TLA 700 Series application.
2. Click on the Windows 95 Start button, point to Settings, and click Control
Panel.
3. In the Control Panel window, double-click Add/Remove Programs.
4. Select the Startup Disk property page.
5. Insert a floppy disk into the disk drive and follow the on-screen instructions
to create the startup disk.
Back Up User Files
You should always back up your user files on a regular basis. Use the Win-
dows 95 Back Up tool to back up files stored on the hard disk. The Back Up tool
is located in the System Tools folder in the Accessories folder. Start the tool and
determine which files and folders that you want to back up. Use the Windows 95
online help for information on using the Back Up tool.
13
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Basic Installation
14
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Installing Microprocessor Support Software
Each microprocessor support package comes with setup software and required
support files. Install the microprocessor software using the setup program on the
floppy disk (the following procedure assumes the instrument is already
powered on).
NOTE. Before you install any software, verify that microprocessor support
software is compatible with the logic analyzer software.
1. Insert the floppy disk in the disk drive.
2. Click the Windows 95 Start button, point to Settings, and click Control
Panel.
3. In the Control Panel window, double-click Add/Remove Programs.
4. Follow the instructions on the screen for installing the software from the
floppy disk.
Similarly, you can remove or uninstall software by following the above
instructions and selecting Uninstall.
15
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Installing Microprocessor Support Software
16
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
The Next Step
Now that you have set up your logic analyzer, you are ready to learn how to use
the logic analyzer for various applications. This chapter provides information on
other documentation that you may need to operate your logic analyzer.
Online Help
The online help is the primary source of information for the logic analyzer. The
online help provides operational details on the screen without having you refer to
printed documentation. Most of the online help information overlaps basic
operating information found in the user manual.
The online help also contains basic operating information for microprocessor
support products.
An online example is provided to guide you through the basic steps in using your
logic analyzer. To access the online help example, start the TLA 700 Series
Application, click on Help, Help Topics, and Welcome.
TLA 700 Series QuickStart Training Board and Manual
The TLA 7QS Training Manual is intended to be used with the TLA 700 Series
QuickStart Training Board to provide examples of how to use the logic analyzer
to solve typical problems. The TLA 700 Series QuickStart Training Board
provides a means of connecting the logic analyzer to signal sources for exercises
described in the training manual. The training board can be used with the logic
analyzer module and the DSO module.
A TLA 7QS Technical Reference Manual is also available to provide technical
and service information on the TLA 700 Series QuickStart Training Board.
Other Documentation
Refer to Related Documentation in the Preface for a list of printed documenta-
tion for your TLA 700 Series Logic Analyzer.
17
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
The Next Step
18
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
This chapter provides information that addresses problems that you may
encounter while installing your logic analyzer. This chapter does not identify
module-specific problems relating to performance verification or adjustments.
For information on running performance verification procedures or adjustment
procedures, refer to the instructions in the TLA 700 Series Performance
Verification and Adjustment Technical Reference Manual, available as an
optional accessory to your logic analyzer.
Diagnostics
The following diagnostic tools are available with your logic analyzer:
H
Power on diagnostics. These diagnostics run when you first power on the
logic analyzer or when you first start the TLA 700 Series application. If any
diagnostic failures occur during power on, the Calibration and Diagnostics
property page appears.
NOTE. Before running the extended diagnostics on the logic analyzer module,
disconnect any attached probes.
H
Extended diagnostics. These diagnostics test the logic analyzer more
thoroughly than the power on diagnostics. The TLA 700 Series extended
diagnostics test the DSO module as well as the logic analyzer module. You
can use the extended diagnostics to isolate problems down to an individual
module. The extended diagnostics are part of the TLA 700 Series Applica-
tion.
H
H
TLA 700 Mainframe diagnostics. The TLA Mainframe diagnostics are a
separate Windows 95 application located in the TLA 700 program group.
These diagnostics check operation of the mainframe beyond the basic PC
circuitry. These diagnostics also check the front panel controls of the
portable mainframe.
QAPlus diagnostics. The QAPlus diagnostics are a separate Windows 95
application located in the Windows 95 Start Programs menu. The diagnostics
check the basic operation of the controller including the following items:
system board, video, hard and floppy disks, memory, communications port,
printer port, mouse and keyboard.
Use the diagnostics to isolate problems to software or hardware within the logic
analyzer. For additional help on the diagnostics, contact your local Tektronix
representative.
19
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Software Problems
Your TLA 700 Series Logic Analyzer comes with most software already
installed. For any suspected software problems, try to isolate the problem to the
TLA 700 Series application software or to other installed software.
Before running any of the diagnostics, you should check the online release notes
to verify the logic analyzer software matches the module firmware.
Run the QAPlus diagnostics software to identify hardware or software problems.
Follow the QAPlus online help instructions for running the diagnostics software.
The diagnostics are located in the Start menu under Programs/ QAPlus_Win-
Win.
Many software problems can be due to corrupted or missing software files. In
most cases the easiest way to solve software problems is to reinstall the software
and follow the on-screen instructions. Refer to Upgrading Software on page 41
for instructions on reinstalling or upgrading software.
Refer to Table 4 on page 21 for a list of software and hardware troubleshooting
information and recommended action.
If you suspect problems with the TLA 700 Series software, contact your local
Tektronix representative or refer to Contacting Tektronix on page xii.
Hardware Problems
Hardware problems can have several causes. The first step to identifying a
problem is to review the installation instructions in this manual to verify that you
have properly installed the logic analyzer.
If you are certain that you have installed the logic analyzer correctly, run the
TLA 700 Series extended diagnostics (located under the System menu) to
identify any problems with the individual modules.
If your logic analyzer powers up so that you have access to the desktop, run the
QAPlus diagnostics software to identify possible controller hardware problems.
Follow the QAPlus online help instructions for running the diagnostics software.
The diagnostics are located in the Start menu under Programs/ QAPlus_Win-
Win.
You can also run the external TLA 700 Mainframe diagnostics to identify
problems not covered by other diagnostics. The TLA 700 Mainframe diagnostics
are located under the Start menu under the Tektronix TLA 700 programs.
Table 4 lists some of the symptoms of hardware problems and possible solutions.
20
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
CAUTION. Do not attempt to remove or reinstall modules in the logic analyzer
while the logic analyzer is receiving power. Power off the logic analyzer and
disconnect the power cord before removing and installing modules. Failure to do
so can damage the logic analyzer or the individual modules.
Table 4: Troubleshooting
Symptom
Possible causes and recommended action
Mainframe does not power on
Verify that all power cords are connected to the mainframe and to the power source.
Check that the mainframe receives power when you press the On/Standby switch; check that fans
start and that front-panel indicators light.
Check that power is available at the power source.
Check for failed fuses (see Table 3 on page 10 for mainframe fuse information).
Mainframe failure; contact your local Tektronix service center.
Check the monitor power cord connection.
Monitor does not power on
Monitor display is blank
Check for failed fuse.
Monitor failure; contact the vendor of your monitor for corrective action.
Check that the monitor is connected to the mainframe; replace the cable if necessary.
If portable mainframe display is blank, try connecting external monitor; if both displays are blank,
contact your local Tektronix service center.
External monitor controls turned down; adjust monitor controls for brightness and contrast.
Check the controller BIOS setups for the monitor.
Faulty monitor; contact the vendor of your monitor for corrective action.
Mainframe powers on but does Power off mainframe and check that all modules are fully inserted.
not complete the power-on
If mainframe is a benchtop mainframe, check the status of the SYSTEM FAIL and TEST LEDs on
the benchtop controller. If either LED stays on, contact your local Tektronix service center.
sequence
Check the status of the READY and ACCESSED LEDs on the front panel of the application
modules. The READY LED turns on when the module passes the power on diagnostics and when
the module is ready to communicate with the controller. The ACCESSED LED turns on any time the
controller accesses the module.
Check for disk in floppy disk drive; make sure mainframe boots from the hard disk drive.
Check for faulty module. Remove modules one at a time and power on the mainframe. If mainframe
completes the power-on sequence, replace faulty module.
Possible software failure or corrupted hard disk; see Software Problems at the beginning of this
chapter.
Power-on diagnostics fail
Isolate problem to faulty mainframe or to faulty module. Multiple diagnostic failures across modules
indicate a faulty mainframe. Diagnostic failures confined to an single module most likely indicate a
faulty module. Contact your local Tektronix service center.
21
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Table 4: Troubleshooting (Cont.)
Symptom
Possible causes and recommended action
Mainframe does not recognize
accessories such as monitor,
printer, or keyboard
Check that accessories are properly connected or installed. Try connecting other standard PC
accessories or contact your local Tektronix service center.
LA Module merge not recog-
nized
Modules not merged; check merge connection.
Faulty module; replace with known-good module or contact your local Tektronix service center.
Windows 95 comes up but the Mainframe not set up to start TLA 700 application at power on. Start application from the desktop, by
TLA 700 application does not
double-clicking on the TLA 700 icon; see the TLA 700 application online help on Powering On and
Off for setting the startup sequence.
Faulty or corrupt software; reinstall the TLA 700 application software.
Exit the Safe mode and restart the mainframe.
Windows 95 comes up in Safe
mode
Incompatible hardware and hardware driver software. Either install hardware driver or remove the
incompatible hardware.
TLA 700 application starts but
modules do not display in
System window
Power off mainframe and check that all modules are fully inserted.
Flash programming jumper installed. Power off mainframe and remove module. Remove jumper (see
Figure 21 on page 50), and reinstall module.
Module address switches not set correctly. Power off mainframe and remove module. Set address
switches to FF and reinstall module (refer to Figure 21 on page 50 for address switch locations).
Module failure; replace with known-good module or contact your local Tektronix service center.
Mainframe failure; contact your local Tektronix service center.
Portable Mainframe will not
power off with On/Standby
switch
The Mainframe Utilities may be set up to disable hard power off. Check the setting of the Mainframe
Utilities (the Mainframe Utilities are located in the Windows 95 Control Panel).
The Windows 95 operating system problem. Try powering off the mainframe using the Windows 95
shutdown procedure. If the mainframe still does not power off, disconnect power cord and reconnect
after 10 seconds to reboot the mainframe.
22
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
TLA 700 Startup Sequence
The following information is intended to provide troubleshooting hints in case
the logic analyzer fails to complete the startup sequence. You may want to refer
to Figure 8 on page 24 while reading the following paragraphs.
At power on, the logic analyzer starts the mainframe and module kernel tests. If
the mainframe passes the kernel tests, it begins to boot the Windows 95
operating system. If the mainframe fails the kernel tests, it displays the error
codes, beeps, and terminates the startup sequence.
The Windows 95 operating system starts the resource manager. The resource
manager performs the following tasks:
H
H
H
H
H
H
Starts the mainframe power-on self tests
Verifies the power-on self test status
Inhibits any failed modules
Records the power-on self test failures
Determines the logic analyzer configuration
Executes the system controller power-on diagnostics
After completing all of the above tasks, the logic analyzer starts the TLA 700
application. The TLA 700 application performs the following tasks:
H
H
H
Performs the power-on diagnostics on all installed modules
Performs the power-on diagnostics on the TLA 700 system
Records the Pass/Fail status in the Calibration and Diagnostics property
sheet
If no failures occur, the TLA 700 application is ready to use for regular tasks.
23
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Power on
Module 1
kernel test
Pass/Fail
Module 2
kernel test
Pass/Fail
Module 3
kernel test
Pass/Fail
Module n
kernel test
Pass/Fail
Mainframe kernel test
Fail
Terminate and
display error
code
Pass/Fail ?
Pass
Verify kernel self test status
Inhibit failed modules
Log failures and build configuration
MS Windows 95
boot
Start TLA 700 application
Start resource
manager
Module power-on diagnostics
System controller
tests
TLA 700 system power-on diagnostics
Records failures in Calibration and
Diagnostic property sheet
TLA 700 ready
Figure 8: TLA 700 startup sequence
24
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Replacing Podlets on the P6417 Logic Analyzer Probe
Although your logic analyzer was fully tested at the factory prior to shipment,
there may be reasons to move or replace podlets on your logic analyzer probes.
Moving Probe Podlets
Depending on your application, you can move or reposition probe podlets as
shown in Figure 9. When moving podlets in the podlet holder keep the ground
side of the podlets together.
Remove podlet
cables
Pull open the podlet holder
Figure 9: Moving probe podlets
25
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Replacing Probe Podlets
The podlet replacement procedure consists of identifying the faulty podlet and
replacing it with a new one. A podlet replacement kit is available as an optional
accessory.
NOTE. Refer to the TLA 700 Online Help for part numbers of the TLA 700
accessories.
Equipment and Required Material. You will need the following equipment and
material:
H
H
H
H
Replacement podlet
Masking tape
Small diagonal cutters
Screwdriver with #1 Phillips tip
Replacement Procedure. Perform the following steps to replace a faulty probe
podlet:
1. Identify and mark the faulty podlet.
2. Disconnect the probe from the logic analyzer and set the probe cable on a
clean working surface.
NOTE. If you need to replace more than one podlet, replace the podlets one at a
time to avoid mixing them up.
26
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
3. Using the screwdriver with the #1 Phillips tip, remove the four screws on the
probe housing on the module end of probe cable (see Figure 10).
Top housing
Cable header
housing
Rubber comb
Bottom housing
Podlet holder
Figure 10: Disassembling the probe cable
4. Remove the top and bottom probe housings and set them aside.
27
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
5. Remove the remaining top and bottom probe housings and set them aside.
Do not mix the probe connector-end housings with the podlet connector
housings; the labels are different at each end of the probe.
6. Identify the faulty probe channel at the module end of the cable.
7. Refer to Figure 11 and use a small pointed tool, such as a straightened paper
clip or a mechanical pencil, to release the faulty cable from the cable header
housing. Discard the small plastic retaining snap.
Paper clip
Cable header housing
Retaining snap
Podlet cable
Figure 11: Removing the faulty channel from the probe cable header housing
8. Identify the faulty probe channel at the podlet end of the probe.
CAUTION. Be sure that you have identified the correct podlet cable before you cut
the cable.
9. Cut the faulty podlet just above the color-coded ring.
10. Remove the color-coded ring and set it aside; you will need to install it on
the replacement cable later in this procedure.
28
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
11. Refer to Figure 12 and bend the rubber comb to remove the faulty podlet
cable from the comb.
Remove faulty podlet cable from rubber comb
Cut above color-coded ring
Figure 12: Removing a podlet cable from the rubber comb
12. Attach the module end of the replacement podlet cable to the cut end of the
faulty podlet cable in the following manner:
a. Lay the two cables in parallel with a four-inch (10 cm) to six-inch
(15 cm) overlap.
b. Wrap the overlapping portion of the two podlet cables tightly with
masking tape to temporarily splice them together.
13. Pull the faulty cable from the module end through the mesh sleeve so that
the replacement cable is drawn through the mesh sleeve. Stop pulling when
the replacement cable is located at the proper position in the cable wire
bundle.
29
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
14. Remove the faulty podlet cable from the podlet holder and replace it with the
new cable as shown in Figure 13.
Align the ground connectors
Pull open
Figure 13: Replacing a podlet in the podlet holder
15. Take the color-coded ring that you removed in step 10 on page 28 and slide
it onto the installation tool. (The installation tool is included in the podlet
replacement kit.)
CAUTION. Do not damage the cable with the installation tool when installing the
color-coded ring.
16. Refer to Figure 14 and place the installation tool over the replacement podlet
cable as shown. Slide the color-coded ring down the installation tool onto the
replacement cable. Position the ring so that it is in the same relative position
as the other color-coded rings (approximately 0.5-inch from the podlet).
30
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
Installation tool
Install color-coded ring on
the replacement cable and
slide the ring down the tool.
Replacement
cable
Figure 14: Installing the color-coded ring
17. Push the new podlet cable into the rubber comb.
18. Unwrap and remove the masking tape from the splice of the two cables and
discard the faulty cable.
19. Remove the protective 1x2 housing from the new podlet cable. Use the same
procedure as in Step 7 on page 28 to remove the plastic retaining snap.
Discard the retaining snap and the protective housing.
31
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
In Case of Problems
20. Insert the new podlet cable partially into the cable header housing. Place a
new retaining snap (included with the kit) on the podlet cable as shown in
Figure 15 and fully insert the cable and retaining snap into the cable header
housing. Tug lightly on the podlet cable to ensure the retaining snap is in
place.
Cable header housing
Retaining snap
Podlet cable
Figure 15: Installing the cable in the cable header housing
21. Reinstall the top and bottom probe housings on the module end of the probe
cable in the following manner (refer to Figure 10 on page 27 if necessary):
a. Line up the two tabs on the probe header with the two embedded slots in
the bottom (deep) probe housing.
NOTE. It may be necessary to push or pull the cables in the sleeve so that the
probe header lines up properly with the slots in the bottom probe housing.
b. Place the two metallic guides on the probe sleeve in the slots in the
bottom probe housing.
c. Line up the three tabs of the probe header with the three embedded slots
in the top probe housing.
d. Install and tighten the four screws in the probe housing.
e. Smooth the sleeve by hand to ensure the sleeve fully covers the cables
when replacing the probe housing over the comb.
22. Repeat step 21 to install the probe housing over the rubber comb on the
podlet end of the probe cable.
32
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
This appendix describes how to merge two logic analyzer modules to form a
wider module.
There are no specific rules to merging modules. However, the following
guidelines may be helpful.
H
H
H
H
H
You can merge modules with 102 channels or greater.
The maximum merged module width is two modules.
The two modules should have the same firmware version.
The two modules should have the same state speed.
When merging modules of unequal channel widths, place the module with
the higher number of channels in the lower numbered slot.
H
Merged modules of unequal memory depths will assume the depth of the
shallower module.
Equipment Required
Merge Procedure
You will need the following equipment to complete these procedures:
H
H
Torque screwdriver with a size T-10 Torx tip
Two logic analyzer modules
CAUTION. Static discharge can damage any semiconductor component in the
logic analyzer module. Discharge the static voltage from your body by wearing a
grounded antistatic wrist strap while performing the Merge Procedure.
Perform the following steps to merge two logic analyzer modules:
1. Power off the logic analyzer and remove the modules.
2. Determine which module will be designated as the slave module (this
module will be placed in the higher numbered slot).
3. Lay the slave module on the right side (as viewed from the front panel).
33
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
4. Use the screwdriver with the T-10 Torx tip and remove the eight screws from
the cover (Six screws on the cover, and two screws on the rear panel, see
Figure 16).
5. Remove the four flat-head screws near the front of the module (102-channel
modules only have two flat-head screws).
6. Remove the two screws holding the Merge cable bracket to the cover.
7. Remove the top part of the cable bracket and set it aside.
8. Remove the cover and locate the Merge cable.
Merge cable bracket
Remove screws (2)
Remove screws (4)
Remove screws (8)
Module cover
Chassis
Figure 16: Removing the cover
34
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
9. Replace the cover while feeding the Merge cable through the hole in the
cover (see Figure 17). Do not twist the cable while feeding it through the
hole, otherwise the cable will not connect properly to the adjacent module.
Remove screws (2)
Merge cable bracket
Merge cable
Figure 17: Feeding the Merge cable through the cover
10. Turn the Merge cable bracket over so that the guide pins point up and place
the bracket over the Merge cable connector.
11. Install the two screws that hold the Merge cable bracket in place.
CAUTION. To prevent damage to the module during the installation process,
reinstall the cover exactly as described in steps 12 through 16. If the cover is not
properly seated, the module can be damaged when you install it in the logic
analyzer or it will not meet EMC requirements.
35
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
12. Push forward on the cover so the tab on the front edge of the cover inserts
into the rear of the front subpanel. Make sure that the cover is fully seated
(no gaps) against the front and rear chassis flanges (see Figure 18).
Make sure tab inserts into
slot on rear of front panel
Leave no vertical gap
Figure 18: Seating the cover on the chassis
13. While holding the cover in place, install the four screws nearest the front of
the module (two on the top and two on the bottom), to secure the cover to
the chassis.
14. Install the flat-head screws near the front of the module.
15. Slide the rear panel on the chassis and install the two rear panel screws.
16. Install the top and bottom rear screws.
36
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
17. Check and tighten all screws to 4 in lbs.
18. Place the Master module adjacent to the Slave module so that the two guide
pins from the Slave module line up with the guide pin holes in the Master
module (see Figure 19).
Guide pins
Master module
Slave module
Figure 19: Lining up the two modules
19. Gently push the two modules together so that the Merge cable of the Slave
module mates with the Merge cable connector on the Master module.
Installing the Merged Modules in the Logic Analyzer
After merging the two modules, perform the following steps to install the
module pair in the logic analyzer:
1. Hold the module pair such that the two modules do not become separated
and line up the modules with the slot guides in the logic analyzer.
2. Push both of the modules in the logic analyzer until they rest against the rear
panel connector. Use the injector/ejector handles to fully seat the modules
one at a time. Tighten the module retainer screws.
37
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
3. Verify that the modules are fully seated before powering on the logic
analyzer.
4. After powering on the logic analyzer, run the merge calibration to calibrate
the module pair (select the system menu, point to Calibration and Diagnos-
tics, point to the Self Calibration property page, and select the merged pair in
the list box).
Removing Merged Modules from the Logic Analyzer
Perform the following steps to remove merged modules from the logic analyzer:
1. Identify the two logic analyzer modules that you want to remove.
2. Loosen the retainer screws for both modules (top and bottom).
NOTE. Use the ejector handles to just loosen the modules from the rear panel
connector before sliding both modules out of the logic analyzer.
3. Use the ejector handles to disconnect the first module from the rear panel
connector.
4. Use the ejector handles to disconnect the second module from the rear panel
connector.
5. Slide both modules out of the logic analyzer at the same time.
6. Gently pry the two modules apart.
Storing the Merge Cable
Perform the following steps to store the Merge cable inside the logic analyzer
module:
1. Use the screwdriver with the T-10 Torx tip to remove the two screws that
hold the Merge cable bracket in place.
2. Remove the Merge cable bracket.
3. Use the screwdriver with the T-10 Torx tip to remove the screws from the
side cover and rear cover (see Figure 16 on page 34).
4. Push the Merge cable through the side cover and remove the cover.
5. Dress the Merge cable as shown in Figure 20.
38
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
Merge cable
Figure 20: Dressing the Merge cable before installing the cover
CAUTION. To prevent damage to the module during the installation process,
reinstall the cover exactly as described in steps 6 through 8. If the cover is not
properly seated, the module can be damaged when you install it in a mainframe
or it will not meet EMC requirements.
6. Push forward on the cover so the tab on the front edge of the cover inserts
into the rear of the front subpanel. Make sure that the cover is fully seated
(no gaps) against the front and rear chassis flanges (see Figure 18 on
page 36).
7. While holding the cover in place, install the four screws nearest the front of
the module (two on the top and two on the bottom), to secure the cover to
the chassis.
8. Install the flat-head screws near the front of the module.
9. Slide the rear panel on the chassis and install the two rear panel screws.
10. Install the top and bottom rear screws.
11. Install the Merge cable bracket so that the guide pins point into the module.
12. Install and tighten the two screws on the Merge cable bracket.
13. Verify that you have installed and tightened all screws to 4 in lbs on the
module.
39
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A: Merging Logic Analyzer Modules
40
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
This appendix describes procedures for upgrading your logic analyzer software
or firmware. The following sections are included:
H
H
H
Restoring or upgrading software in your logic analyzer
Installing TPI, the TLA 700 Programmatic Interface (page 43)
Restoring or upgrading firmware in your logic analyzer module or DSO
module (page 50)
H
H
H
Restoring or changing the Controller BIOS setup (page 52)
Reinstalling Software (page 56)
Customizing the TLA 700 Series setups (page 61)
Upgrading Software
This section provides information for restoring or upgrading software in your
logic analyzer. In addition to the TLA 700 software, there are other software
programs that are installed separately; Table 5 lists some of the software and
installation information.
Table 5: Software not covered by the TLA 700 Series software setup
Software
Installation information
1
Microsoft Windows 95 OS
Refer to the Windows 95 documentation or contact your local
Microsoft representative.
QAPlus Diagnostic Software
Refer to the QAPlus online documentation or to Reinstall the
QAPlus Diagnostic Software on page 59.
Logic Analyzer Module & DSO Refer to the TLA 700 Series Performance Verification and
Module Performance Verifica- Adjustment Technical Reference Manual for instructions on
tion and Adjustment
installing and using the PV/Adjust software.
Microprocessor Support Soft- Refer to Installing Microprocessor Support Software beginning
ware
on page 15 in this manual.
PC Card Software
Refer to the instructions that come with your PC Card.
Refer to the instructions that come with your software.
Other software
1
For information on installing Microsoft Windows 95 software not available on the
soft.com/windows/software/cdextras.htm
41
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Upgrading or Reinstalling
TLA 700 Software
You can update or reinstall the TLA 700 software by executing the setup.exe
program on the Disk #1 of installation disks or by performing the following
steps.
NOTE. The following instructions assume that you are reinstalling the TLA 700
application software only. If you are installing the software on a new hard disk
drive, refer to the instructions under Reinstalling Software on page 56.
1. If the TLA 700 application is running, exit the application before continuing.
2. Insert the TLA 700 application floppy disk 1 in the disk drive.
3. Click Start, point to Settings, and click Control Panel.
4. In the Control Panel window, double-click Add/Remove Programs.
5. Click on the Install button and follow the instructions on the screen for
installing the software from the floppy disk.
If the setup program finds an earlier version of the TLA 700 Application, the
program will ask you to confirm your selection and then will remove the
software. At the end of the removal process, the program will ask you to
remove the floppy disk and then restart the logic analyzer to complete the
file removal process.
6. After the logic analyzer restarts, reinsert the floppy disk and repeat steps 1
through 5 to install the application software.
When executed, the setup program will replace old versions of the software after
you have confirmed your choice. The program also offers to delete shared files
which appear to be no longer needed by any application.
The program will not remove saved TLA files or microprocessor support
packages.
Uninstalling TLA 700
Software
You can also remove or uninstall the TLA 700 software using the Uninstall
program. Use the Add/Remove Programs tool under the Control Panel to
uninstall software.
NOTE. The TLA 700 Software Setup program automatically invokes the Uninstall
program before reinstalling the TLA 700 software.
42
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Installing TPI
Use these procedures to install the TLA 700 Programmatic Interface (TPI). In
these procedures, client refers to the user written application that controls the
logic analyzer through the TPI. Server refers to the TLA 700 application
software. For additional information on installation, refer to the online help.
You have two methods of using TPI with your client application:
H
You can run the client application locally on the logic analyzer. No special
setup is required. To verify that a client application can connect to the
TLA 700 server, see Running a Client Application on the Logic Analyzer.
H
You can run your client application remotely across the network. In this case,
both the logic analyzer and the remote host require special setup procedures.
See Setting Up TPI for Remote Access on the Logic Analyzer to begin this
setup procedure.
In the following procedures, <install directory> refers to the directory where the
TPI client is installed on your client machine. This directory is C:\Program
Files\TLA 700 by default.
If you need to use the type library (tla700.tlb), this file will be located in
C:\Program Files\TLA 700\System\TPI on the logic analyzer and/or in the
<install directory>\System\TPI on your client machine.
Running a Client
Application on the Logic
Analyzer
To verify that a client application on the logic analyzer can connect to the
TLA 700 application, do the following steps:
1. Start the TLA 700 application on the logic analyzer.
2. Navigate to and run the following program: C:\Program Files\
TLA 700\Samples\TPI Samples\Vc++\test client\testclient.exe
3. The Tektronix Test TPI Client dialog appears. Click the Connect to
TLA 700! button.
4. Check that the text in the dialog changes to Connected! This indicates the
test client application connected to the logic analyzer. Click Exit.
The client application will run locally on the logic analyzer. For remote
operation, see Setting Up TPI for Remote Access on the Logic Analyzer.
Setting Up TPI for Remote
Access on the Logic
Analyzer
If you want to run a client application remotely across the network, you must set
up the logic analyzer using the following procedure:
1. Install and configure TCP/IP. (networking)
43
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
NOTE. If you have difficulty configuring the network setup, contact your system
administrator.
2. Download the following from Microsoft’s web site. Microsoft’s web site is:
H
H
Distributed COM for Microsoft Windows 95 (DCOM95)
DCOM for Microsoft Windows 95 configuration utility (only required for
user level access)
Install the downloaded files and restart the TLA 700 after each installation. For
more information on DCOM, refer to the Microsoft website.
3. You can choose to have share-level or user-level access to the logic analyzer
as provided by Microsoft Windows 95.
Share-level access allows a password to be assigned to each shared resource. For
Microsoft Windows 95-only networks, share-level access is the only option.
User-level access allows a group of users to have access to each shared resource.
NOTE. For TPI to work with share-level access, authentication is turned off and
any COM client can call into any COM server running on the logic analyzer.
Setting the Access Level on the Logic Analyzer. To set up the access level on the
logic analyzer, do the following steps:
1. Click Start, point to Settings, and select Control Panel.
2. From the Control Panel, select Network, then click the Access Control tab.
3. To select share-level access, choose Share-level access control, and then
click OK.
To select user-level access, choose User-level access control and enter the
name of the domain that will be used to validate user access. Click OK.
4. Click Start again, and then Shut Down. In the Shut Down Windows dialog
box, select Restart the Computer? and then click the Yes button.
5. If you are setting up share level access, navigate to and double-click
C:\Program Files\TLA 700\System\TPI\Share Level Access Server.reg
If you are setting up user level access, navigate to and double-click
C:\Program Files\TLA 700\System\TPI\User Level Access Server.reg
6. A dialog appears, indicating successful registration. Click OK.
44
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
7. Restart the logic analyzer, using the procedure in step 4.
8. Start the TLA 700 application on the logic analyzer if it is not already
running.
This completes the setup of TPI on the logic analyzer for operating with a remote
client machine.
Next, you need to set up the client machine so that it can connect to the logic
analyzer. Refer to the appropriate setup:
H
H
H
Setting Up a Client Machine that is Running Microsoft Windows 95
on page 45.
Setting Up a Client Machine that is Running Microsoft Windows NT
on page 48.
Setting Up a Client Machine Running on Any Other Platform
on page 49.
Setting Up a Client
Machine that is Running
Microsoft Windows 95
After you set up the logic analyzer, you must set up the client machine using the
following procedure:
1. Install and configure TCP/IP. (networking)
2. Load the Tektronix TPI Client Install diskettes supplied with the TLA 700
instrument.
3. Double-click on Setup.exe.
4. Download the following from Microsoft’s web site. Microsoft’s web site is:
H
H
Distributed COM for Microsoft Windows 95 (DCOM95)
DCOM for Microsoft Windows 95 configuration utility (only required for
user level access)
Install the downloaded files and restart the TLA 700 after each installation.For
more information on DCOM, refer to the Microsoft website.
You must set up the client machine to match the access level you chose for the
logic analyzer (share-level or user-level access). Refer to the appropriate access
type to continue with the setup procedure:
H
H
Setting Up Share-level Access for Microsoft Windows 95 on page 46.
Setting Up User-level Access for Microsoft Windows 95 on page 47.
NOTE. You can switch between user- and share-level access later by redoing the
Setting the Access Level procedure on page 44, from step 3 onward.
45
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Setting Up Share-level Access for Microsoft Windows 95.To set up share-level
access for a client machine running Microsoft Windows 95, do the following
procedure on the client machine:
1. Click Start, point to Settings, and select Control Panel.
2. From the Control Panel, select Network, then click the Access Control tab.
3. Choose Share-level access control, and then click OK. If prompted, insert the
Windows 95 CD, or provide a filepath to the stored Windows 95 files.
4. The System Settings Change dialog appears. Click the Yes button.
5. Click Start, then Shut Down. In the Shut Down Windows dialog box, select
Restart the Computer? and then click the Yes button.
6. Navigate to and double-click <install directory>\System\TPI\Share Level
Access Client.reg.
7. A dialog appears, indicating successful registration. Click OK.
8. Restart the client machine, using the procedure in step 5.
9. In Microsoft Windows 95, click Start and then Run. Either locate the regedit
file using Browse, or enter regedit, then click OK to run regedit.
10. Click on the following registry key: HKEY_CLASSES_ROOT\AppID\
{C67DAA22-4972-11d1-9CAC-00805F0D8271}
11. Using Edit>New>StringValue, add a named value, RemoteServerName.
12. Click the new value, RemoteServerName, and select Edit>Modify.
The Edit String dialog appears.
13. Enter the computer name of the logic analyzer. This is the name used to
identify the logic analyzer on the network. Click OK.
To verify that the setup is complete, do the following:
1. Navigate to and run <install directory>\Samples\TPI Samples\Vc++\test
client\testclient.exe on the client machine.
2. The Tektronix Test TPI Client dialog appears. Click the Connect to
TLA 700! button to see if the client can connect to the logic analyzer.
3. Attempting Connection to TLA 700 appears in the dialog text box. (The first
time you connect, it may take a few minutes.) The text changes to Con-
nected! when the connection is made.
This completes the setup of the logic analyzer and the Windows 95 client
machine for running a client application across the network.
46
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Setting Up User-level Access for Microsoft Windows 95. To set up user-level access
for a client machine running Microsoft Windows 95, do the following procedure:
1. Click Start, point to Settings, and select Control Panel.
2. From the Control Panel, select Network, then select the Access Control tab.
3. Choose User-level access control and enter the name of the domain that will
be used to validate user access. Select OK.
4. On the client machine, click Start and then Shut Down. In the Shut Down
Windows dialog, select Restart the Computer? and then click the Yes button.
5. Navigate to and double-click <install directory>\System\TPI\User Level
Access Client.reg.
6. Restart the client machine, using the procedure in step 4.
7. In Microsoft Windows 95, click Start, and then select Run. Either locate the
dcomcnfg file using Browse, or enter dcomcnfg in the edit field. Click OK.
8. Double-click “Tektronix TLA 700 Application” in the Applications page.
9. In the Location page, uncheck the “Run application on this computer” box
and check the “Run application on the following computer” box. Enter the
name of the logic analyzer in the edit field.
To verify that the setup is complete:
1. Run <install directory>\Samples\TPI Samples\Vc++\test client\test-
client.exe on the client machine.
2. The Tektronix Test TPI Client dialog appears. Click the Connect to
TLA 700! button to see if the client can connect to the logic analyzer.
3. Attempting Connection to TLA 700 appears in the dialog text box.
(The first time you connect, it may take a few minutes.) The text changes to
Connected! when the connection is made.
This completes the setup of the logic analyzer and the client machine for running
a client application across the network.
NOTE. You can switch between share-level and user-level access later by
uninstalling the Tektronix TPI Client and DCOM95 by using the Windows 95
control panel. Redo the procedure from step 2 of Setting Up a Client Machine
that is Running Microsoft Windows 95 on page 45.
47
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Setting Up a Client
Machine that is Running
Microsoft Windows NT
After you set up the logic analyzer, you must set up the client machine using the
following procedure:
1. Install and configure TCP/IP. (networking)
NOTE. If you have difficulty configuring the network setup, contact your system
administrator.
2. Load the Tektronix TPI Client Install diskettes supplied with the logic
analyzer.
3. Double-click on Setup.exe.
4. Click Start and then Shut Down. In the Shut Down Windows dialog, select
Restart the Computer? and then click the Yes button.
5. Depending on the type of access control you chose for the logic analyzer,
double-click <install directory>\System\TPI\Share Level Access
Client.reg or User Level Access Client.reg.
6. Restart the client machine, using the procedure in step 4.
7. In Microsoft Windows NT, click Start, and then select Run. Either locate the
dcomcnfg file using Browse, or enter dcomcnfg. Then click OK.
8. Double-click Tektronix TLA 700 Application in the Applications page.
9. In the Location page, check the “Run application on the following computer”
box. Enter the name of the logic analyzer in the edit field.
To verify that the setup is complete, do the following:
1. Run <install directory>\Samples\TPI Samples\Vc++\test client\test-
client.exe on the client machine.
2. The Tektronix Test TPI Client dialog appears. Click the Connect to
TLA 700! button to see if the client can connect to the logic analyzer.
3. Attempting Connection to TLA 700 appears in the dialog text box.
(The first time you connect, it may take a few minutes.) The text changes to
Connected! when the connection is made.
This completes the setup of the logic analyzer and the Windows NT client
machine for running your client application across the network.
48
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Setting Up a Client
Machine Running on Any
Other Platform
If the client application requires use of the type library, you can generate it on
your platform using tla700.odl in C:\Program Files\TLA 700\System\TPI\src
on the logic analyzer.
Make sure that you perform the following steps:
1. Ensure that DCOM is working on your platform.
2. Merge C:\Program Files\TLA 700\System\TPI\Client.reg into your
registry.
3. Depending on the type of access control you chose for the logic analyzer,
merge C:\Program Files\TLA 700\System\TPI\Share Level Access
Client.reg or User Level Access Client.reg into your registry.
4. Add a string value named RemoteServerName to the key
HKEY_CLASSES_ROOT\AppID\
{C67DAA22-4972-11d1-9CAC-00805F0D8271}
5. Enter the computer name of the logic analyzer as the value of this string.
This is the name used to identify the logic analyzer on the network.
This completes the setup of the logic analyzer for operating with a remote client
machine using a platform other than Microsoft Windows NT or Microsoft
Windows 95.
49
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Updating Module Firmware
Perform the following steps if you need to update or reinstall the flash ROM-
based firmware on your application modules. You can update multiple modules
and module types during a single firmware update session.
1. Power off the logic analyzer.
2. Disconnect the power cord.
3. Disconnect any probes on the modules that you want to upgrade.
4. Remove the module from the logic analyzer.
5. Refer to Figure 21 and locate the Flash programming pins on the rear of the
module.
Flash programming pins
Figure 21: Flash programming pins
6. Install a jumper on the Flash programming pins (use one of the spare
jumpers that came with your logic analyzer software).
50
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
7. Reinstall the modules in the logic analyzer.
8. Reconnect the power cord and power on the logic analyzer.
NOTE. When the TLA 700 application starts, any modules where you have
installed the Flash programming jumper will not display in the system menu.
9. Exit the TLA 700 application.
10. Click Start, point to Programs, point to Tektronix TLA 700, and click
TLA 700 Firmware Loader.
11. Select the modules that you want to update from the list of modules
displayed in the Supported list box near the top of the menu. If you are
updating more than one module at a time, note the slot location of the
module in the selection box.
12. Select Load Firmware from the Execute menu.
13. Click the proper .LOD file for the module in the C:\Program Files\ TLA 700
Firmware directory. If you are updating more than one module make sure
that you select the proper firmware file for the module in the listed slot
number.
14. Click Ok. You will be asked to confirm your action; click Yes when
prompted.
NOTE. The program will not allow you to load firmware to an incompatible
module. For example, the program will not load DSO firmware to a logic
analyzer module.
15. If you are updating another module, select the proper .LOD file for the
module.
16. Click Ok. You will be asked to confirm your action; click Yes when
prompted.
The program will load the firmware for each module one at a time. The
process takes approximately five minutes per module.
17. When the load operation is complete, select each of the modules in the
Supported list and then select Checksum from the Execute menu.
18. If each module passes the checksum test, exit the program.
19. Power off the logic analyzer and disconnect the power cord.
51
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
20. Remove the module from the logic analyzer and remove the jumper from the
Flash programming pins on the rear of the module. Keep the jumper for
future updates.
21. Attach a new label with the new firmware version to the module.
22. Reinstall the module in the logic analyzer and reconnect the probes.
23. Reconnect the power cord and power on the logic analyzer.
24. Verify that the power-on diagnostics pass.
25. Run the self calibration on each module after a 30-minute warm-up.
Controller BIOS Setup
The logic analyzer has a built-in setup program where you can set some basic
input and output characteristics. In most cases, there is no need to access the
setup program. If you need to access the setup program, power on the logic
analyzer and press function key F2 on the keyboard (ALT, CTRL, F2 keys
simultaneously on the portable mainframe front panel) before the logic analyzer
boots the Windows 95 operating system.
Table 6 on page 53 lists the default settings for the BIOS setup. You may need to
refer to this information after updating the software or making other changes.
CAUTION. Do not change any of the settings if you are unfamiliar with the setup
program. Selecting an incorrect setup can cause the logic analyzer to malfunc-
tion.
If you need to change any of the settings to match those listed in Table 6, follow
the on-screen instructions.
NOTE. If you need to upgrade the Controller BIOS version, refer to the Mainte-
nance chapter of the TLA 711 Color Benchtop Controller Service Manual or the
TLA 704 Color Portable Mainframe Service Manual for the upgrade instruc-
tions.
52
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Table 6: Controller BIOS setups for the TLA 700 application
Parameter
Main menu
Factory setting
Submenu parameter
Factory setting
System Time
Set to current date
Set to current time
1.44 MB, 3 1/2
System Date
Diskette A
Diskette B
Not Installed
IDE Adapter 0 Master
(C:[hard disk size])
Autotype Fixed Disk
Type
Select Enter
Hard disk size
Number of cylinders
Number of heads
Number of sectors/track
None
Cylinders
Heads
Sectors/track
Write Precomp
Multi-sector transfers
LBA Mode Control
32-Bit I/O
Set by Autotype
Set by Autotype
Disabled
Transfer Mode
Set by Autotype
IDE Adapter 0 Slave
IDE Adapter 1 Master
IDE Adapter 1 Slave
Video System
(None)
(None)
(None)
EGA/VGA
Memory Cache
External Cache
Enabled
Enabled
Enabled
Cache System BIOS area:
Cache Video BIOS area:
Cache Memory Region
CC00 – CFFF:
Disabled
Disabled
Disabled
Disabled
Disabled
Enabled
Enabled
D000 – D3FF:
D400 – D7FF:
D800 – DBFF:
DC00 – DFFF:
Memory Shadow
System Shadow
Video Shadow:
Regions with Legacy Expan-
sion ROMS
CC00 – CFFF:
ROM
53
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)
Parameter
Factory setting
Submenu parameter
D000 – D3FF:
Factory setting
ROM
D400 – D7FF:
ROM
D800 – DBFF:
DC00 – DFFF:
Boot Delay
ROM
ROM
Boot Sequence:
A: then C:
0
Boot Sequence:
Setup Prompt
A: then C:
Enabled
Enabled
Enabled
Enabled
Off
POST Errors
Floppy Check
Summary Screen
NUMLock
NUMLOCK:
OFF
Key Click
Disabled
30/sec
1/4 sec
Keyboard auto-repeat rate
Keyboard auto-repeat delay
System Memory
Extended Memory
Advanced menu
640 KB
Installed memory – 640 K
Integrated Peripherals
COM A
3F8, IRQ4
2F8, IRQ3
378, IRQ7
ECP
COM B
LPT Port
LPT Mode
Diskette Controller
Local Bus IDE Adapter
DRAM Speed
DMA Aliasing
8 Bit I/O Recovery
16 Bit I/O Recovery
IRQ12 Used By
ECC/Parity
Enabled
both
1
Advanced Chipset Control
60 ns
Enabled
4.5
4.5
PS/2 Mouse
Disabled
Plug & Play O/S
No
Reset Configuration Data
Large Disk Access Mode:
Enable SimulScan Mode:
No
DOS
Disabled
54
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)
Parameter
Flash ATA PC Card Booting:
Factory setting
Submenu parameter
Factory setting
Disabled
Power menu
2
APM:
Enabled
3
Power Savings
Customize
Standby Timeout:
Disabled
Disabled
MAX
Suspend Timeout:
Standby CPU Speed:
Fixed Disk Timeout:
Disabled
ON
CRT:
Standby Timer Reset Events
Keyboard
Enabled
Mouse
PS/2 (IRQ12)
Standby Break Events
IRQ0
IRQ1
Disabled
Auto
IRQ3
Auto
IRQ4
Auto
IRQ5
Disabled
Disabled
Disabled
Disabled
Disabled
Disabled
Disabled
Auto
IRQ6
IRQ7
IRQ8
IRQ9
IRQ10
IRQ11
IRQ12
IRQ13
Disabled
Disabled
Disabled
IRQ14
IRQ15
Standby Wakeup Events
Keyboard
Enabled
55
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)
Parameter
Mouse
Factory setting
Submenu parameter
Factory setting
PS/2 (IRQ12)
1
2
3
DRAM speed was set to 70 ns for TLA 700 Series controllers with BIOS version 1.00.11 or earlier.
APM is set to Disabled for TLA 700 Series controllers with BIOS version 1.00.11 or earlier.
Power Savings is set to Customize when the APM is enabled. This allows Windows to manage power use for the
instrument as indicated on the Power property page of the Power icon in the Control Panel. Power Savings was set to
Disabled with BIOS version 1.00.11 or earlier.
Reinstalling Software
There may be occasions when you have to reinstall software or restore software
after replacing the hard disk. The procedures in this section explain how to
reinstall software on the hard disk. If you replace the hard disk, you must also
configure the controller BIOS. To configure the controller BIOS, follow steps
under Configure the Controller BIOS on page 56.
On rare occasions you may need to reformat the hard disk. If so, follow the steps
under Reformatting the Hard Disk on page 57. Otherwise, follow the steps under
Reinstalling the Windows 95 Software on page 58.
NOTE. You should perform all of the procedures in this section in sequence.
Configure the Controller
BIOS
This procedure is necessary after replacing the hard disk or when the CMOS
settings are corrupted or lost. You will need the the emergency startup disk that
you created when you installed your logic analyzer (see Create an Emergency
Startup Disk on page 13). To configure the Controller BIOS, complete the
following steps:
1. Power off the logic analyzer and insert the emergency startup disk in the
floppy disk drive.
2. Power on the logic analyzer.
3. When prompted, press function key F2 on the keyboard before the logic
analyzer boots the Windows 95 operating system.
4. In the Main Setup menu, use the Down Arrow key to scroll to the line
labeled IDE Adapter 0 Master.
5. Press the Enter key to display the hard disk detail submenu.
56
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
6. Press the Enter key again to cause the logic analyzer to automatically
determine and set the hard disk drive parameters.
7. Press the ESC key twice and then the Enter key twice to exit and save the
BIOS setup.
Reformatting the Hard
Disk
You should not reformat your hard disk drive unless it is absolutely necessary. If
you find that you need to format the hard disk, (if possible) make sure that you
have backup copies of your user files so that you can reinstall them later.
You must perform the following procedure after you replace the hard disk, or
when the partitioning or the format of the hard disk has been damaged. You will
use the Setup program on the Windows 95 Setup Boot floppy disk to partition
and format the hard disk.
Complete the following steps to format the hard disk:
1. Insert the Windows 95 Setup Boot floppy disk in the floppy disk drive.
2. Power on the logic analyzer so that it boots from the floppy disk.
3. Press function key F3 to exit the Setup program; press F3 again to confirm
your selection.
You must temporarily exit the Setup program to properly execute the FDISK
command with the /X switch. If you do not execute the FDISK /X command
separately, the primary DOS partition will not be compatible with the
TLA 700 Series BIOS.
4. At the prompt, enter the command FDISK /X.
5. Enter N when you are asked if your want to select Large Disk Support.
6. Select option 1 to create a new partition on the hard disk.
7. Select option 1 again to specify that the new partition is to be the primary
DOS partition.
8. Enter Y to use the maximum available size for the DOS partition.
9. Press the CTRL, ALT, and Delete keys simultaneously to reboot the logic
analyzer and restart the Setup program.
When the Setup program restarts, it will caution you that there is an
operating system already installed. Disregard this message. The message
appears because you partitioned the hard disk in the previous steps.
10. Continue with the Setup program and follow the instructions on the display.
The Setup program will format the hard disk and exit the program when
complete.
57
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Reinstalling the
Windows 95 Software
The following procedure assumes that the hard disk is already properly for-
matted. To reinstall the Windows 95 software, complete the following steps:
1. Start the logic analyzer from the Windows 95 Setup Boot floppy disk.
2. When the install program prompts you for an installation, select Typical.
3. Enter the Windows 95 Product Identification number located above the bar
code on the Windows 95 software package.
4. Enter your name when prompted.
5. Enter your company name when prompted (optional).
6. When prompted, select the appropriate hardware categories (for example,
Network).
7. Install the Windows 95 floppy disks one at a time until you have installed all
of the floppy disks.
8. Enter the current date, time, and time zone when prompted.
9. Restart the logic analyzer when prompted. The logic analyzer should start up
in Windows 95. You are now ready to install the TLA 700 Series application
software.
Set the Display Driver
After installing the Windows 95 software, you must set the display driver for
your monitor. Otherwise, using the default display driver may not provide the
optimum display settings. To set the display driver, complete the following steps:
1. Click Start, point to Settings, click on Control Panel, and open the Display
icon.
2. Select the Settings property page.
3. Click on the Advanced Properties button.
4. Select the Adapter tab page, and then click on the Change button.
5. Click on the Have Disk button.
6. Insert the floppy disk labeled Cirrus Logic Video Display Driver
GDS7543 – Rel 2.40 for Windows 95 and click OK.
7. Select the driver named Cirrus Logic 7543 2.40 PCI(Cirrus) and then click
OK.
NOTE. If a Version Conflict dialog box appears and asks you if you want to keep
the file, answer Yes.
58
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
8. After the installation is complete, close all of the dialogs and property sheets.
However, keep the Control Panel open.
9. When the prompt appears asking you if you want to reboot now, enter No.
Reinstall the QAPlus
Diagnostic Software
To reinstall the QAPlus diagnostic software, complete the following steps:
1. Insert the QAPlus floppy disk into the disk drive.
2. In the Control Panel window, double-click Add/Remove Programs.
3. Click on the Install button to begin installing the software.
4. When the program prompts you to confirm the creation of the C:\win–win
directory, click Continue.
NOTE. If you use a set of QAPlus disks which have been previously installed, you
will not be able to enter a user name. The program will warn you that the disks
have been previously installed and then ask you to acknowledge that fact before
proceeding.
5. When the program asks you if you want to create a set of icons for the
program manager, click Create.
After completing the installation, the program opens up the Notepad
application and displays the release notes for the QAPlus software.
6. Close the Notepad application and press the CTRL, ALT, and Delete keys
simultaneously to display a list of current tasks.
7. Select the QAPlus task, and click End task to kill the task.
8. After a few seconds, a dialog box displays advising you that the application
is not responding. Click End Task again to kill the application.
9. When the desktop reappears, close the Program Manager folder.
10. Click Start, point to Programs, and click on MS-DOS Prompt to open an
MS-DOS window.
11. In the MS-DOS window, change directories to the C:\win-win directory.
12. Insert the QAPlus/WIN disk labeled QAplus\win-win Manual
Version 7.01.
13. Copy the file A:\Qawwusro.exe to the C:\win-win directory.
14. Exit and close the MS-DOS window.
59
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Reinstall the TLA 700
Series Application
Software
To reinstall the TLA 700 Series application software, complete the following
steps:
1. Insert the TLA 700 application floppy disk 1 in the disk drive.
2. Click Start, point to Settings, and click on Control Panel.
3. In the Control Panel window, double-click Add/Remove Programs.
4. Click on the Install button and follow the instructions on the screen for
installing the software from the floppy disk.
5. When the installation is complete, the program will prompt you to reboot the
system; click on OK to reboot.
Reinstalling other
Application Software
To reinstall any other application software such as microprocessor support
software, complete the following steps:
1. Insert the first floppy disk into the disk drive.
2. Click Start, point to Settings, and click Control Panel.
3. In the Control Panel window, double-click Add/Remove Programs.
4. Click on the Install button and follow the instructions on the screen for
installing the software from the floppy disk.
Restoring User Files
After you have reinstalled all of the application software files, you should restore
any backed-up user files. Use the Windows 95 Back Up tool to restore any saved
files from floppy disks. Use the online help for instructions on restoring the files.
60
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Customizing TLA 700 Series Setups
After restoring all of the user and application files, you must perform the
following steps to complete the software installation process. The steps in this
section will return your logic analyzer to the same settings it had when it was
shipped from the factory. The following steps are included in this section:
H
H
H
H
H
Creating a shortcut for the QAPlus online manual
Adjusting the display desktop settings for an external monitor
Adjusting the refresh rate
Setting up the screen saver
Setting up the Virtual Memory
NOTE. You must exit the TLA 700 Application before completing the procedures
in this section. The TLA 700 Application starts automatically each time you
power on the logic analyzer.
Create a Shortcut for the
QAPlus Online Manual
Complete the following steps to create a shortcut from the Start menu for the
QAPlus online manual:
1. Use the Windows Explorer (right-click on Start and select Explore) to open
the directory C:\Windows\Start Menu\Programs\QAPlus_WIN-WIN.
2. Select the QAPlus_WIN-WIN directory. From the File menu, select New
and then Create Shortcut.
3. Enter the name of the file C:\WIN-WIN\Qawwusro.exe in the field of the
Create Shortcut dialog.
4. Click Next and set the name of the shortcut to QAPlus_WIN-WIN Manual.
5. Click Finish to install the shortcut in the Start menu.
6. Close the Windows Explorer.
61
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Adjust the Display
Desktop Settings for an
External Monitor
The following steps are only necessary if your logic analyzer uses an external
monitor. The following steps are a generalized procedure for a standard Plug &
Play monitor. If you know the type of monitor that you will use, select the
monitor type in steps 6 and 7 and adjust the remaining parameters for your
specific monitor.
1. Click Start, point to Settings, click on Control Panel, and open the Display
icon.
2. Select the Settings property page.
3. Click on the Advanced Properties button.
4. Select the Monitor tab page and then click on the Change button.
5. Click on the radio button to Show All Devices.
6. Select Standard Monitor Types under the Manufacturers column.
7. Select Plug & Play Monitor under the Model Type column.
8. Click OK to close the selection window.
9. Click Close to close the Advanced Properties dialog.
10. Set the following parameters in the Settings property page:
H
H
Color Palette
Desktop Area
256 Colors
640 x 480
11. Leave the Display property page open and continue with the next procedure
to adjust the refresh rate.
Adjust the Refresh Rate
A higher monitor refresh rate provides a brighter, clearer display. It also reduces
eye fatigue from screen flicker. The following recommendations set the refresh
rate to the highest possible settings and may not be optimal for all display types.
You may want to try different settings to determine which work best for your
specific monitor.
1. Select the Monitor Refresh property page.
2. Click the Set Highest button to use the highest available refresh rate for each
desktop selection.
3. Click OK to close the Monitor Refresh property page.
You will see a warning dialog informing you that the new settings are about
to be tested and that the refresh settings will revert to the old settings within
10 seconds if you do not confirm them.
62
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
4. Click OK. After the new settings are in effect, click Yes to keep the new
settings.
5. Leave the Display property page open and continue with the next procedure
to select and configure the screen saver.
Set Up the Screen Saver
The screen saver includes the option of reducing the intensity of the TLA 700
Series LCD display backlight to help prolong the life of the display. Select this
recommended screen saver unless you will always use an external monitor.
1. Select the Screen Saver property page.
2. Select Tektronix LCD Saver from the list of screen savers.
3. Set the Wait period for ten minutes.
4. Click the Settings button to open the screen saver dialog.
5. Set the speed to Fast and the Intensity to Reduced (60%).
6. Click OK to close the dialog.
7. Click OK to close the Display window.
Set Up the Mainframe
Utilities
The Mainframe Utilities provide additional settings for your logic analyzer (the
utilities can only be used with the portable mainframe).
1. Click Start, point to Settings, click on Control Panel, and open the Main-
frame Utilities icon.
2. Select the Power Off Mode tab page.
3. Click on the radio button named Enable Soft Power Off.
4. Select the LCD Mode tab page.
5. Click on the radio button named LCD Full On.
6. Click OK to close the utility.
63
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B: Upgrades
Setup the Virtual Memory
The following virtual memory settings ensure optimal performance of the
TLA 700 Series application software.
1. Open the System icon in the Control Panel.
2. Select the Performance property page and select the Virtual Memory button.
3. In the Virtual Memory dialog, click the radio button Let me specify my own
virtual memory settings.
4. Enter a minimum value of 100 MB and a maximum value of 500 MB.
5. Click OK to close the dialog.
6. Select Yes to when prompted to confirm your settings.
7. Click OK to close the System window.
8. When the prompt appears asking you if you want to reboot now, enter No.
Enable 32-Bit Support for
the PCMCIA Interface
Perform the following steps to enable the 32-bit support for the PCMCIA
Interface:
NOTE. You should complete this procedure even if you do not intend to use the
PCMCIA interface.
1. Open the System icon in the Control Panel.
2. Select the Performance property page and then click on the text PC card
(PCMCIA) devices are not using 32-bit support.
3. Click on the Details button.
4. In the Help window, click on the button to enable 32-bit PC card support.
5. When asked if you are using a PC card to install Windows, select No and
click Next.
6. When asked to review system files to select real-mode PC card drivers for
Windows to delete, select No and click Next.
7. Close the Help window.
8. Click Finish.
9. When asked if it is ok to shut down the computer to complete the installation
of 32-bit support, select Yes.
64
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C: Power Cord and Line Fuse Requirements for
the Benchtop Mainframe
The benchtop mainframe comes with two power cords and three fuses (one fuse
is already installed). You must determine the correct fuse and power cord for
your configuration. This is important to avoid overloading the power distribution
system and ensures that you comply with the National Electrical Code.
The power consumption depends on the number and type of instrument modules
installed in the benchtop mainframe. Table 7 lists the power consumed for each
module.
To determine the total power consumption perform the following steps:
1. Use Table 7 to determine the power consumption for each module in the
benchtop mainframe.
2. Add the power for each module to determine the total power consumption.
3. Determine at which line voltage you will be operating the mainframe.
4. Refer to Figure 22 on page 66 to determine the proper power cord and line
fuse for your mainframe.
Table 7: Power for instrument modules
Module type
Benchtop Controller
TLA 7L1
Power (watts)
40
55
73
94
109
57
76
99
115
56
81
56
81
TLA 7L2
TLA 7L3
TLA 7L4
TLA 7M1
TLA 7M2
TLA 7M3
TLA 7M4
TLA 7D1
TLA 7D2
TLA 7E1
TLA 7E2
65
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C: Power Cord and Line Fuse Requirements for the Benchtop Mainframe
For power usage in the nonshaded region of Figure 22, use either the power cord
with the 15 A plug (two parallel prongs and ground) or the power cord with the
20 A plug (two perpendicular prongs and ground). For high power usage
combined with low input line voltages (shaded region), use only the power cord
with the 20 A plug. Select the proper fuse based on the ranges shown in
Figure 22.
Power Cord with 20A plug (NEMA 5–20P) ONLY!
800
700
600
500
400
300
200
100
0
Power Cord with 15A plug (NEMA 5–15P)
Power Cord with 20A plug (NEMA 5–20P)
90
110
115
20AT Fuse
15AF Fuse
130
150
170
190
210
230
250
Input Line Voltage (Volts AC)
6.3AF Fuse
Figure 22: Power cord identification chart
For example, assume that your configuration consists of four TLA 7M4 Logic
Analyzer Modules and one TLA 7E2 DSO Module. Also assume that you will
be operating the mainframe at 90 VAC.
Add 40 watts (for the controller) + 81 watts (for the DSO Module) + 4 X
115 watts (for the logic analyzer modules) to come up with a total value of
581 watts. Because the power consumption (at 90 VAC) is approaching the
shaded area of the graph in Figure 22, you should consider using the 20A power
cord.
66
TLA 700 Series Installation Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
|