Tektronix TV Converter Box 070 9774 02 User Manual

Installation Manual  
TLA 700 Series  
Logic Analyzer  
070-9774-02  
Warning  
The servicing instructions are for use by qualified  
personnel only. To avoid personal injury, do not  
perform any servicing unless you are qualified to  
do so. Refer to all safety summaries prior to  
performing service.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TEKTRONIX SERVICE OPTIONS  
Tektronix offers the following service options. These options are modular, flexible, and easy to order with your instru-  
ment. Designed to ease installation and start up, to support tracking of calibration to requirements of ISO9000, and  
to provide for extended repair coverage, these options help fix your long-term maintenance costs and eliminate un-  
planned expenditures. And these options can be converted from service at Tektronix service depots to service on-site  
(see Option S3), which helps keep downtime to a minimum.  
Product installation service1  
Three years of calibration services  
Test data  
Option IN  
Option C3  
Option D3  
Option R3  
Option S1  
Option S3  
Provides initial installation of the product and  
familiarizes new users with some of its opera-  
tion features  
Provides initial Certification on delivery, plus  
two more annual calibrations from your ser-  
vice center  
Provides test data on delivery and when annu-  
al calibration services are provided (three to-  
tal, requires Option C3)  
Three years repair coverage  
One year uplift to on–site service1,2  
Three year uplift to on-site service1,2  
Provides three years of repair coverage for the  
instrument, including displays and accessories  
sold with the instrument  
Converts the standard one year, “return to de-  
pot” warranty to provide service on site for  
one year  
Converts any C3, D3, and R3 options pur-  
chased to on-site services for three years  
Tektronix Service Options are available at the time you order your instrument. Contact your local Tektronix Sales  
Office for more information.  
1
Availability of installation and on-site services depends on the type of product and may vary by geography.  
2
Uplift options are ordered with the mainframe products and cover individual modules.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
v
vii  
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Service Offerings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
ix  
x
xi  
xii  
Basic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Check the Shipping List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Check the Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Install the Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Cover Empty Slots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Connect the Standard and Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Connect the Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Power On the Logic Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Powering Off the Logic Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Perform the Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Create an Emergency Startup Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Back Up User Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
1
1
2
5
6
7
9
10  
12  
12  
13  
13  
Installing Microprocessor Support Software . . . . . . . . . . . . . . . . . . . .  
15  
The Next Step . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Online Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
TLA 700 Series QuickStart Training Board and Manual . . . . . . . . . . . . . . . . . . .  
Other Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
17  
17  
17  
17  
In Case of Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Software Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Hardware Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
TLA 700 Startup Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Replacing Podlets on the P6417 Logic Analyzer Probe . . . . . . . . . . . . . . . . . . .  
19  
19  
20  
20  
23  
25  
Appendix A: Merging Logic Analyzer Modules . . . . . . . . . . . . . . . . . .  
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Merge Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Installing the Merged Modules in the Logic Analyzer . . . . . . . . . . . . . . . . . . . .  
Removing Merged Modules from the Logic Analyzer . . . . . . . . . . . . . . . . . . . .  
Storing the Merge Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
33  
33  
33  
37  
38  
38  
Appendix B: Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Upgrading Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Installing TPI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Updating Module Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Controller BIOS Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Reinstalling Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Customizing TLA 700 Series Setups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
41  
41  
43  
50  
52  
56  
61  
Appendix C: Power Cord and Line Fuse Requirements for the  
Benchtop Mainframe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
65  
i
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
List of Figures  
Figure 1: Location of ground connections . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2: Installing modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 3: Installing the empty slot panel covers . . . . . . . . . . . . . . . . . .  
Figure 4: Locations of external connectors . . . . . . . . . . . . . . . . . . . . . .  
Figure 5: Connecting the logic analyzer probes . . . . . . . . . . . . . . . . . .  
Figure 6: Line fuse and power cord connector locations . . . . . . . . . . .  
Figure 7: On/Standby switch locations . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 8: TLA 700 startup sequence . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 9: Moving probe podlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 10: Disassembling the probe cable . . . . . . . . . . . . . . . . . . . . . . .  
4
5
6
7
9
10  
11  
24  
25  
27  
Figure 11: Removing the faulty channel from the probe cable  
header housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
28  
29  
30  
31  
32  
34  
35  
36  
37  
39  
50  
66  
Figure 12: Removing a podlet cable from the rubber comb . . . . . . . .  
Figure 13: Replacing a podlet in the podlet holder . . . . . . . . . . . . . . .  
Figure 14: Installing the color-coded ring . . . . . . . . . . . . . . . . . . . . . . .  
Figure 15: Installing the cable in the cable header housing . . . . . . . .  
Figure 16: Removing the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 17: Feeding the Merge cable through the cover . . . . . . . . . . . .  
Figure 18: Seating the cover on the chassis . . . . . . . . . . . . . . . . . . . . . .  
Figure 19: Lining up the two modules . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 20: Dressing the Merge cable before installing the cover . . . .  
Figure 21: Flash programming pins . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 22: Power cord identification chart . . . . . . . . . . . . . . . . . . . . . .  
ii  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
List of Tables  
Table 1: Mainframe operating requirements . . . . . . . . . . . . . . . . . . . .  
3
8
Table 2: Additional accessory connection information . . . . . . . . . . . .  
Table 3: Line fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Table 4: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Table 5: Software not covered by the TLA 700 Series software setup  
Table 6: Controller BIOS setups for the TLA 700 application . . . . .  
Table 7: Power for instrument modules . . . . . . . . . . . . . . . . . . . . . . . .  
10  
21  
41  
53  
65  
iii  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
iv  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Safety Summary  
Review the following safety precautions to avoid injury and prevent damage to  
this product or any products connected to it. To avoid potential hazards, use this  
product only as specified.  
Only qualified personnel should perform service procedures.  
While using this product, you may need to access other parts of the instrument.  
Read the General Safety Summary in other system manuals for warnings and  
cautions related to operating the system.  
To Avoid Fire or  
Personal Injury  
Use Proper Power Cord. Use only the power cord specified for this product and  
certified for the country of use.  
Connect and Disconnect Properly. Do not connect or disconnect probes or test  
leads while they are connected to a voltage source.  
Ground the Product. This product is indirectly grounded through the grounding  
conductor of the mainframe power cord. To avoid electric shock, the grounding  
conductor must be connected to earth ground. Before making connections to the  
input or output terminals of the product, ensure that the product is properly  
grounded.  
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings  
and marking on the product. Consult the product manual for further ratings  
information before making connections to the product.  
The common terminal is at ground potential. Do not connect the common  
terminal to elevated voltages.  
Do Not Operate Without Covers. Do not operate this product with covers or panels  
removed.  
Do Not Operate With Suspected Failures. If you suspect there is damage to this  
product, have it inspected by qualified service personnel.  
Do Not Operate in Wet/Damp Conditions.  
Do Not Operate in an Explosive Atmosphere.  
Provide Proper Ventilation. Refer to the manual’s installation instructions for  
details on installing the product so it has proper ventilation.  
v
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Safety Summary  
Symbols and Terms  
Terms in this Manual. These terms may appear in this manual:  
WARNING. Warning statements identify conditions or practices that could result  
in injury or loss of life.  
CAUTION. Caution statements identify conditions or practices that could result in  
damage to this product or other property.  
Terms on the Product. These terms may appear on the product:  
DANGER indicates an injury hazard immediately accessible as you read the  
marking.  
WARNING indicates an injury hazard not immediately accessible as you read the  
marking.  
CAUTION indicates a hazard to property including the product.  
Symbols on the Product. The following symbols may appear on the product:  
WARNING  
High Voltage  
Protective Ground  
(Earth) Terminal  
CAUTION  
Refer to Manual  
Double  
Insulated  
vi  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Safety Summary  
Only qualified personnel should perform service procedures. Read this Service  
Safety Summary and the General Safety Summary before performing any service  
procedures.  
Do Not Service Alone. Do not perform internal service or adjustments of this  
product unless another person capable of rendering first aid and resuscitation is  
present.  
Disconnect Power. To avoid electric shock, disconnect the main power by means  
of the power cord or, if provided, the power switch.  
Use Caution When Servicing the CRT. To avoid electric shock or injury, use  
extreme caution when handling the CRT. Only qualified personnel familiar with  
CRT servicing procedures and precautions should remove or install the CRT.  
CRTs retain hazardous voltages for long periods of time after power is turned off.  
Before attempting any servicing, discharge the CRT by shorting the anode to  
chassis ground. When discharging the CRT, connect the discharge path to ground  
and then the anode. Rough handling may cause the CRT to implode. Do not nick  
or scratch the glass or subject it to undue pressure when removing or installing it.  
When handling the CRT, wear safety goggles and heavy gloves for protection.  
Use Care When Servicing With Power On. Dangerous voltages or currents may  
exist in this product. Disconnect power, remove battery (if applicable), and  
disconnect test leads before removing protective panels, soldering, or replacing  
components.  
To avoid electric shock, do not touch exposed connections.  
X-Radiation. To avoid x-radiation exposure, do not modify or otherwise alter the  
high-voltage circuitry or the CRT enclosure. X-ray emissions generated within  
this product have been sufficiently shielded.  
vii  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Safety  
viii  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
The TLA 700 Series Installation Manual is intended to help you install your  
TLA 700 Series Logic Analyzer. Although your TLA 700 Series Logic Analyzer  
may arrive with all modules installed, you should still follow the steps in this  
manual to complete the installation process.  
The manual consists of the following sections:  
H
H
Basic Installation describes the hardware installation process and provides  
information to start the logic analyzer the very first time.  
Installing Microprocessor Support Software describes the steps for installing  
microprocessor support software that does not come installed on your logic  
analyzer.  
H
H
The Next Step provides information on other TLA 700 Series documentation  
that you may need to operate your TLA 700 Series Logic Analyzer.  
In Case of Problems provides information on handling problems or  
difficulties that may occur during the installation process. It also provides  
procedures for replacing probe podlets and a list of phone numbers for  
Tektronix product support.  
H
H
Appendix A: Merging Logic Analyzer Modules provides the procedures for  
combining two logic analyzer modules to create a single, wider module.  
Appendix B: Upgrades provides information on upgrading software,  
installing TPI, (TLA 700 Programmatic Interface), upgrading firmware for  
the individual modules, and for using the BIOS setup for the controller in  
your logic analyzer.  
H
Appendix C: Power Cord and Line Fuse Requirements for the Benchtop  
Mainframe provides information on selecting the proper power cord and line  
fuse for the benchtop mainframe.  
ix  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
Related Documentation  
In addition to this installation manual, the following documentation is available  
for your TLA 700 Series logic analyzer:  
H
The TLA 700 Series Logic Analyzer User Manual provides basic operating  
information for the TLA 700 Series Logic Analyzer.  
H
H
H
The online help provides information about the TLA 700 Series user  
interface. To view the online help, select Help Topics from the Help menu.  
A series of microprocessor support instruction manuals provide operating  
and service instructions for the individual microprocessor support packages.  
The TLA 7QS Training Manual provides training exercises to help you learn  
key features of the TLA 700 Series Logic Analyzer. The training manual is  
designed to be used with the TLA 700 Series QuickStart training board.  
H
H
The TLA 7QS Technical Reference Manual provides technical and service  
information for the TLA 700 Series QuickStart training board.  
The TLA 700 Series Performance Verification and Adjustment Technical  
Reference Manual provides performance verification and adjustment  
procedures for the major components of the TLA 700 Series Logic Analyzer.  
The manual includes the performance verification and adjustment software.  
H
H
H
H
H
The TLA 711 Color Benchtop Chassis Service Manual provides module-lev-  
el service information for the benchtop chassis.  
The TLA 711 Color Benchtop Controller Service Manual provides module-  
level service information for the benchtop mainframe controller module.  
The TLA 704 Color Portable Mainframe Service Manual provides module-  
level service information for the portable mainframe.  
The TLA 7Lx/TLA 7Mx Logic Analyzer Service Manual provides module-  
level service information for the logic analyzer module.  
The TLA 7Dx/TLA 7Ex Digitizing Oscilloscope Service Manual provides  
module-level service information for the DSO module.  
x
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
Service Offerings  
Tektronix provides service to cover repair under warranty as well as other  
services that are designed to meet your specific service needs.  
Whether providing warranty repair service or any of the other services listed  
below, Tektronix service technicians are well equipped to service the TLA 700  
Series Logic Analyzers. Tektronix technicians train on Tektronix products; they  
have access to the latest information on improvements to the TLA 700 Series  
Logic Analyzers as well as the latest new product upgrades. Services are  
provided at Tektronix Services Centers and on-site at your facility, depending on  
your location.  
Warranty Repair Service  
Tektronix warrants this product for one year from date of purchase. (The  
warranty appears behind the title page in this manual.) Tektronix technicians  
provide warranty service at most Tektronix service locations worldwide. The  
Tektronix product catalog lists all service locations worldwide or you can visit us  
on our Customer Services World Center web site at http://www.tek.com/Mea-  
surement/Service. See our latest service offerings and contact us by email.  
Calibration and Repair  
Service  
In addition to warranty repair, Tektronix Service offers calibration and other  
services which provide cost-effective solutions to your service needs and quality-  
standards compliance requirements. Our instruments are supported worldwide by  
the leading-edge design, manufacturing, and service resources of Tektronix to  
provide the best possible service.  
The following services can be tailored to fit your requirements for calibration  
and/or repair of the TLA 700 Series Logic Analyzers.  
Service Options. Tektronix Service Options can be selected at the time you  
purchase your instrument. You select these options to provide the services that  
best meet your service needs. These service options are listed on the Tektronix  
Service Options page following the title page of this manual.  
Service Agreements. If service options are not added to the instrument purchase,  
then service agreements are available on an annual basis to provide calibration  
services or post-warranty repair coverage for the TLA 700 Series Logic  
Analyzers. Service agreements may be customized to meet special turn-around  
time and/or on-site requirements.  
Service on Demand. Tektronix also offers calibration and repair services on a  
“per-incident” basis that is available with standard prices for many products.  
xi  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
Self Service. Tektronix supports repair to the replaceable-part level by providing  
for circuit board exchange.  
Use this service to reduce down-time for repair by exchanging circuit boards for  
remanufactured ones. Tektronix ships updated and tested exchange boards. Each  
board comes with a 90-day service warranty.  
For More Information. Contact your local Tektronix service center or sales  
engineer for more information on any of the Calibration and Repair Services just  
described.  
Contacting Tektronix  
Product  
Support  
For application-oriented questions about a Tektronix measure-  
ment product, call toll free in North America:  
1-800-TEK-WIDE (1-800-835-9433 ext. 2400)  
6:00 a.m. – 5:00 p.m. Pacific time  
Or contact us by e-mail:  
For product support outside of North America, contact your  
local Tektronix distributor or sales office.  
Service  
Support  
Contact your local Tektronix distributor or sales office. Or visit  
our web site for a listing of worldwide service locations.  
In North America:  
For other  
information 1-800-TEK-WIDE (1-800-835-9433)  
An operator will direct your call.  
To write us Tektronix, Inc.  
P.O. Box 1000  
Wilsonville, OR 97070-1000  
xii  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
This chapter describes all of the steps needed to install the TLA 700 Series Logic  
Analyzer. It is written from the perspective that you purchased most of the items  
uninstalled and you intend to install all of the pieces. If you purchased a logic  
analyzer with modules already installed, you should still review all of the  
information in this chapter and perform those steps that apply to your specific  
situation.  
This chapter deals mainly with hardware installation. The basic operating  
software is already installed on the internal hard disk. However, if you ordered  
additional software, such as microprocessor support software, you must install  
that software yourself. Refer to Installing Microprocessor Support Software on  
page 15.  
Check the Shipping List  
Verify that you have received all of the parts of your logic analyzer. Use the  
shipping lists that came with your logic analyzer to compare against the actual  
contents of your order. You should also verify the following:  
H
H
Verify that you have the correct power cords for your geographical area.  
Verify that you have backup copies of the installed software. Store the  
backup software in a safe location where you can easily retrieve the software  
for maintenance purposes.  
NOTE. Keep the software packaging available because you will need it to enter  
the Windows 95 software registration number when you first power on the logic  
analyzer. (See Step 6 under Power on the Logic Analyzer on page 11.)  
H
H
Verify that you have the correct module types and probes.  
Verify that you have all the standard and optional accessories that you  
ordered.  
Remember to fill out and send in the customer registration card. The registration  
card is packaged with this manual.  
1
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Check the Environmental Requirements  
Read this section before attempting any installation procedures. This section  
describes site considerations, power requirements, and ground connections for  
your logic analyzer.  
Site Considerations:  
Portable Mainframe  
The portable mainframe is designed to operate on a bench or on a cart in the  
normal position (on the bottom feet). For proper cooling, at least two inches  
(5.1 cm) of clearance is recommended on the rear and sides of the mainframe.  
You can also operate the mainframe while it rests on the rear feet. If you operate  
the mainframe while it is resting on the rear feet, make sure that you properly  
route any cables coming out of the rear of the mainframe to avoid damaging  
them.  
CAUTION. Keep the bottom of the mainframe clear of obstructions to ensure  
proper cooling.  
Site Considerations:  
Benchtop Mainframe  
The benchtop mainframe is designed to operate on a bench or in a rackmount  
environment. For proper cooling, at least two inches (5.1 cm) of clearance is  
recommended on the rear and sides of the mainframe.  
WARNING. Because of the size and weight of the benchtop mainframe, use care  
when lifting or moving the mainframe to avoid personal injury while performing  
the installation procedures. Use two people to lift or move the mainframe.  
2
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Operating Requirements  
Table 1 lists the operating requirements for the logic analyzers. Refer to the  
TLA 700 Series Logic Analyzer User Manual for a complete specification list.  
Table 1: Mainframe operating requirements  
Requirement  
TLA 704 Color Portable mainframe  
90–250 V , 45–66 Hz, continuous range  
TLA 711 Color Benchtop mainframe  
90–250 V , 45–66 Hz, continuous range  
Source voltage  
RMS  
RMS  
100–132 V , 360–440 Hz, continuous range 100–132 V , 360–440 Hz, continuous range  
RMS  
RMS  
Steady state input current  
6 A , maximum  
15 A , maximum at 90 VAC  
RMS  
RMS  
6.3 A , maximum at 207 VAC  
RMS  
Maximum power consumption  
600 W  
1350 W line power  
Temperature  
(No media in floppy disk drive)  
+5° C to +50° C (+41° to +122° F), 15° C/hr  
(55° F/hr) maximum gradient, noncondensing  
(derated 1° C (34° F) per 1000 ft (304.8 m)  
above 5000 ft (1524 m))  
+5° C to +50° C (+41° to +122° F), 15° C/hr  
(55° F/hr) maximum gradient, noncondensing  
(derated 1° C (34° F) per 1000 ft (304.8 m)  
above 5000 ft (1524 m))  
Humidity  
(No media in floppy disk drive)  
20% to 80% relative humidity, noncondensing.  
Maximum wet bulb temperature: +29° C  
(84° F) (derates relative humidity to ~22% @  
50° C (+122° F))  
20% to 80% relative humidity, noncondensing.  
Maximum wet bulb temperature: +29° C  
(84° F) (derates relative humidity to ~22% @  
50° C (+122° F))  
Altitude  
To 10,000 ft (3,040 m)  
(derated 1° C (34° F) per 1000 ft (304.8 m)  
above 5000 ft (1524 m))  
To 10,000 ft (3,040 m)  
(derated 1° C (34° F) per 1000 ft (304.8 m)  
above 5000 ft (1524 m))  
3
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Chassis Ground  
Connections  
Figure 1 shows chassis ground connections. Use the chassis ground connections  
to connect the grounds of one or more instruments to the logic analyzer to ensure  
a common ground connection between instruments.  
WARNING. Do not remove the safety ground screw from the benchtop mainframe.  
The safety ground screw must always be in place to ensure the proper bonding of  
the power supply to the mainframe.  
Chassis ground  
Safety ground  
(Benchtop  
mainframe only)  
Figure 1: Location of ground connections  
4
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Install the Modules  
CAUTION. Do not install or remove any modules while the mainframe is powered  
on; doing so can damage the modules or the logic analyzer.  
If your logic analyzer does not have the modules preinstalled, install the modules  
as shown in Figure 2. You can install the modules in any slot. Use a Phillips  
screwdriver with a Number 1 tip to tighten the retaining screws after seating the  
modules in place.  
If you want to merge two logic analyzer modules to create a wider module, refer  
to Appendix A: Merging Logic Analyzer Modules on page 33.  
Portable mainframe  
Benchtop mainframe  
Retaining screws  
Injector/ejector handles  
Figure 2: Installing modules  
5
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Cover Empty Slots  
If you have any unused (empty) slots in your mainframe, you must install blank  
slot panel covers to meet EMC specifications. Install a blank slot panel cover for  
each empty slot as shown in Figure 3. Make sure that the EMI shielding is in  
contact with the adjacent panel or module cover. Installing the covers provides  
proper EMI shielding and reduces audible noise.  
CAUTION. Use only Tektronix TLA 700 Series slot panel covers on the benchtop  
mainframe. Do not use non-Tektronix covers, otherwise the mainframe may not  
meet cooling and EMC requirements.  
EMI shielding strip  
Double slot panel cover  
EMI shielding strip  
Single slot panel cover  
Double slot panel cover  
Figure 3: Installing the empty slot panel covers  
6
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Connect the Standard and Other Accessories  
The accessory connections are the same as those you would make on a personal  
computer. The connection points are shown in Figure 4. Use the icons that  
appear in the circled areas as a guide. See Table 2 on page 8 for additional  
connection information.  
Icon  
Description  
Monitor  
RS-232  
Printer  
Mouse  
Keyboard  
Floppy  
disk drive  
PC card  
(LAN or other)  
Figure 4: Locations of external connectors  
7
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Table 2: Additional accessory connection information  
Item  
Description  
Monitor  
If you use a non-standard monitor, you may need to change the  
the Windows 95 display settings to achieve the proper  
resolution for your monitor. Refer to Set the Display Driver on  
page 58 or to Adjust the Display Desktop Settings for an  
External Monitor on page 62. You may also want to refer to the  
documentation for your display monitor for installation  
information.  
Printer  
Connect the printer to the EPP (enhanced parallel port)  
connector directly. If your printer has a DB–25 connector, use  
the adapter cable that came with your logic analyzer to connect  
to the EPP connector.  
Rackmount  
Other  
Refer to the TLA 711 Benchtop Mainframe Rackmount  
Installation Instructions for information on installing the  
rackmount kit.  
Refer to the TLA 700 Series Application online release notes  
for additional accessory installation information not covered in  
this manual. You should also refer to the user or installation  
documentation that came with your accessory.  
8
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Connect the Probes  
Connect the logic analyzer probes and the optional retaining brackets as shown  
in Figure 5. Then connect the oscilloscope probes.  
Match color-coded labels  
Attach optional  
probe retainer  
brackets  
Figure 5: Connecting the logic analyzer probes  
9
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Power On the Logic Analyzer  
Follow these steps to power on the logic analyzer for the first time.  
1. Check that the line fuse is correct for your application. See Table 3 and  
Figure 6.  
Table 3: Line fuses  
Line voltage  
Rating  
Tektronix part number  
Portable mainframe  
90 V to 132 V operation  
207 V to 250 V operation  
Benchtop mainframe  
90 V to 132 V operation  
103 V to 250 V operation  
207 V to 250 V operation  
8 A, fast blow, 250 V  
6.3 A, fast blow, 250 V  
159-0046-00  
159-0381-00  
20 A, slow blow, 250 V  
15 A, fast blow, 250 V  
6.3 A, fast blow, 250 V  
159-0379-00  
159-0256-00  
159-0381-00  
AC power and fuse  
Figure 6: Line fuse and power cord connector locations  
CAUTION. Connect the keyboard, mouse, and other accessories before applying  
power to the mainframe. Connecting the accessories after powering on the  
mainframe can damage the accessories. Refer to Figure 4 on page 7 for  
connection information.  
10  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
2. Connect the power cord.  
CAUTION. Although the benchtop mainframe can use the power cord with the  
15 A plug, mainframes operating at low line (90 VAC) with four or more  
instrument modules may require the power cord with the 20 A plug. If you have  
four or more modules in your mainframe, you must determine the correct fuse  
and power cord combination to avoid overloading the power distribution system.  
Refer to Power Cord and Line Fuse Requirements for the Benchtop Mainframe  
on page 65 for further information.  
3. If you have an external monitor, connect the power cord and power on the  
monitor.  
4. Press the On/Standby switch to power on the logic analyzer (see Figure 7 for  
the switch location).  
Figure 7: On/Standby switch locations  
5. Wait for the boot routine and low-level self test to complete.  
6. Follow the instructions on the screen and enter the Windows 95 Product  
Identification number (located above the bar code) on the Windows 95  
software package that came with your logic analyzer. Enter other information  
as required.  
The internal setup software will automatically configure your logic analyzer  
and install all required devices depending on the installed accessories.  
7. Restart your logic analyzer when the setup program is complete.  
The next time you power on the mainframe, the logic analyzer application starts  
automatically.  
11  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Powering Off the Logic Analyzer  
The portable mainframe has a built-in soft power-off function that safely powers  
off the mainframe when you press the On/Standby switch. If you have the  
benchtop mainframe, you must power off the mainframe using the Windows 95  
shutdown process before depressing the On/Standby switch.  
CAUTION. When powering off the benchtop mainframe, use the Windows 95 shut  
down procedure. Powering off the benchtop mainframe prematurely can corrupt  
the software on the hard disk.  
Perform the Incoming Inspection  
Incoming inspection consists of verifying the basic operation of the logic  
analyzer. The Power-on diagnostics check the basic functionality. These  
diagnostics run every time you power on the logic analyzer.  
You can also verify more detailed functionality by running the self calibration  
and extended diagnostics.  
NOTE. Allow a 30-minute warm-up before running the self calibration.  
Disconnect any attached probes from the modules. Then select the System menu,  
and point to Calibration and Diagnostics. Run the self calibration followed by  
the extended diagnostics by selecting the proper tab. Results of the tests display  
on the individual property page.  
NOTE. The time required to run the self calibration on the logic analyzer  
modules depends on the number of acquisition channels. Modules with a large  
number of channels may take several minutes to run the self calibration.  
If you want to perform detailed performance verification procedures, refer to the  
TLA 700 Series Performance Verification and Adjustment Technical Reference  
Manual (available as an optional accessory).  
Checking the Logic  
Analyzer Probes  
(Optional)  
Connect the logic analyzer probes to a signal source, start an acquisition, and  
verify that the acquired data is displayed in either the listing or waveform  
windows.  
12  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
Checking the DSO Probes  
(Optional)  
Connect the oscilloscope probes to the Probe Compensation connector on the  
front panel of the DSO module. You can then run the Calibrate Probe function in  
each vertical setup page for the module.  
Checking the Mainframe  
(Optional)  
To check the mainframe diagnostics not covered by the TLA 700 Series  
Application software. Run the QAPlus diagnostics or the TLA 700 Mainframe  
Diagnostics located under the Windows 95 Start menu under the Tektronix  
TLA 700 programs. Exit the TLA 700 Series Application before running the  
external diagnostics.  
Create an Emergency Startup Disk  
Now that you have completed the basic installation process, you should create an  
emergency startup disk that you can use to restart your logic analyzer in case of a  
major hardware or software failure. You should create this disk and then store it  
in a safe place.  
The emergency startup disk contains basic files to restart your logic analyzer. It  
also contains files to check and format the hard disk.  
Follow these steps to create the emergency startup disk:  
1. Exit the TLA 700 Series application.  
2. Click on the Windows 95 Start button, point to Settings, and click Control  
Panel.  
3. In the Control Panel window, double-click Add/Remove Programs.  
4. Select the Startup Disk property page.  
5. Insert a floppy disk into the disk drive and follow the on-screen instructions  
to create the startup disk.  
Back Up User Files  
You should always back up your user files on a regular basis. Use the Win-  
dows 95 Back Up tool to back up files stored on the hard disk. The Back Up tool  
is located in the System Tools folder in the Accessories folder. Start the tool and  
determine which files and folders that you want to back up. Use the Windows 95  
online help for information on using the Back Up tool.  
13  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Installation  
14  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installing Microprocessor Support Software  
Each microprocessor support package comes with setup software and required  
support files. Install the microprocessor software using the setup program on the  
floppy disk (the following procedure assumes the instrument is already  
powered on).  
NOTE. Before you install any software, verify that microprocessor support  
software is compatible with the logic analyzer software.  
1. Insert the floppy disk in the disk drive.  
2. Click the Windows 95 Start button, point to Settings, and click Control  
Panel.  
3. In the Control Panel window, double-click Add/Remove Programs.  
4. Follow the instructions on the screen for installing the software from the  
floppy disk.  
Similarly, you can remove or uninstall software by following the above  
instructions and selecting Uninstall.  
15  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installing Microprocessor Support Software  
16  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Next Step  
Now that you have set up your logic analyzer, you are ready to learn how to use  
the logic analyzer for various applications. This chapter provides information on  
other documentation that you may need to operate your logic analyzer.  
Online Help  
The online help is the primary source of information for the logic analyzer. The  
online help provides operational details on the screen without having you refer to  
printed documentation. Most of the online help information overlaps basic  
operating information found in the user manual.  
The online help also contains basic operating information for microprocessor  
support products.  
An online example is provided to guide you through the basic steps in using your  
logic analyzer. To access the online help example, start the TLA 700 Series  
Application, click on Help, Help Topics, and Welcome.  
TLA 700 Series QuickStart Training Board and Manual  
The TLA 7QS Training Manual is intended to be used with the TLA 700 Series  
QuickStart Training Board to provide examples of how to use the logic analyzer  
to solve typical problems. The TLA 700 Series QuickStart Training Board  
provides a means of connecting the logic analyzer to signal sources for exercises  
described in the training manual. The training board can be used with the logic  
analyzer module and the DSO module.  
A TLA 7QS Technical Reference Manual is also available to provide technical  
and service information on the TLA 700 Series QuickStart Training Board.  
Other Documentation  
Refer to Related Documentation in the Preface for a list of printed documenta-  
tion for your TLA 700 Series Logic Analyzer.  
17  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Next Step  
18  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
This chapter provides information that addresses problems that you may  
encounter while installing your logic analyzer. This chapter does not identify  
module-specific problems relating to performance verification or adjustments.  
For information on running performance verification procedures or adjustment  
procedures, refer to the instructions in the TLA 700 Series Performance  
Verification and Adjustment Technical Reference Manual, available as an  
optional accessory to your logic analyzer.  
Diagnostics  
The following diagnostic tools are available with your logic analyzer:  
H
Power on diagnostics. These diagnostics run when you first power on the  
logic analyzer or when you first start the TLA 700 Series application. If any  
diagnostic failures occur during power on, the Calibration and Diagnostics  
property page appears.  
NOTE. Before running the extended diagnostics on the logic analyzer module,  
disconnect any attached probes.  
H
Extended diagnostics. These diagnostics test the logic analyzer more  
thoroughly than the power on diagnostics. The TLA 700 Series extended  
diagnostics test the DSO module as well as the logic analyzer module. You  
can use the extended diagnostics to isolate problems down to an individual  
module. The extended diagnostics are part of the TLA 700 Series Applica-  
tion.  
H
H
TLA 700 Mainframe diagnostics. The TLA Mainframe diagnostics are a  
separate Windows 95 application located in the TLA 700 program group.  
These diagnostics check operation of the mainframe beyond the basic PC  
circuitry. These diagnostics also check the front panel controls of the  
portable mainframe.  
QAPlus diagnostics. The QAPlus diagnostics are a separate Windows 95  
application located in the Windows 95 Start Programs menu. The diagnostics  
check the basic operation of the controller including the following items:  
system board, video, hard and floppy disks, memory, communications port,  
printer port, mouse and keyboard.  
Use the diagnostics to isolate problems to software or hardware within the logic  
analyzer. For additional help on the diagnostics, contact your local Tektronix  
representative.  
19  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Software Problems  
Your TLA 700 Series Logic Analyzer comes with most software already  
installed. For any suspected software problems, try to isolate the problem to the  
TLA 700 Series application software or to other installed software.  
Before running any of the diagnostics, you should check the online release notes  
to verify the logic analyzer software matches the module firmware.  
Run the QAPlus diagnostics software to identify hardware or software problems.  
Follow the QAPlus online help instructions for running the diagnostics software.  
The diagnostics are located in the Start menu under Programs/ QAPlus_Win-  
Win.  
Many software problems can be due to corrupted or missing software files. In  
most cases the easiest way to solve software problems is to reinstall the software  
and follow the on-screen instructions. Refer to Upgrading Software on page 41  
for instructions on reinstalling or upgrading software.  
Refer to Table 4 on page 21 for a list of software and hardware troubleshooting  
information and recommended action.  
If you suspect problems with the TLA 700 Series software, contact your local  
Tektronix representative or refer to Contacting Tektronix on page xii.  
Hardware Problems  
Hardware problems can have several causes. The first step to identifying a  
problem is to review the installation instructions in this manual to verify that you  
have properly installed the logic analyzer.  
If you are certain that you have installed the logic analyzer correctly, run the  
TLA 700 Series extended diagnostics (located under the System menu) to  
identify any problems with the individual modules.  
If your logic analyzer powers up so that you have access to the desktop, run the  
QAPlus diagnostics software to identify possible controller hardware problems.  
Follow the QAPlus online help instructions for running the diagnostics software.  
The diagnostics are located in the Start menu under Programs/ QAPlus_Win-  
Win.  
You can also run the external TLA 700 Mainframe diagnostics to identify  
problems not covered by other diagnostics. The TLA 700 Mainframe diagnostics  
are located under the Start menu under the Tektronix TLA 700 programs.  
Table 4 lists some of the symptoms of hardware problems and possible solutions.  
20  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
CAUTION. Do not attempt to remove or reinstall modules in the logic analyzer  
while the logic analyzer is receiving power. Power off the logic analyzer and  
disconnect the power cord before removing and installing modules. Failure to do  
so can damage the logic analyzer or the individual modules.  
Table 4: Troubleshooting  
Symptom  
Possible causes and recommended action  
Mainframe does not power on  
Verify that all power cords are connected to the mainframe and to the power source.  
Check that the mainframe receives power when you press the On/Standby switch; check that fans  
start and that front-panel indicators light.  
Check that power is available at the power source.  
Check for failed fuses (see Table 3 on page 10 for mainframe fuse information).  
Mainframe failure; contact your local Tektronix service center.  
Check the monitor power cord connection.  
Monitor does not power on  
Monitor display is blank  
Check for failed fuse.  
Monitor failure; contact the vendor of your monitor for corrective action.  
Check that the monitor is connected to the mainframe; replace the cable if necessary.  
If portable mainframe display is blank, try connecting external monitor; if both displays are blank,  
contact your local Tektronix service center.  
External monitor controls turned down; adjust monitor controls for brightness and contrast.  
Check the controller BIOS setups for the monitor.  
Faulty monitor; contact the vendor of your monitor for corrective action.  
Mainframe powers on but does Power off mainframe and check that all modules are fully inserted.  
not complete the power-on  
If mainframe is a benchtop mainframe, check the status of the SYSTEM FAIL and TEST LEDs on  
the benchtop controller. If either LED stays on, contact your local Tektronix service center.  
sequence  
Check the status of the READY and ACCESSED LEDs on the front panel of the application  
modules. The READY LED turns on when the module passes the power on diagnostics and when  
the module is ready to communicate with the controller. The ACCESSED LED turns on any time the  
controller accesses the module.  
Check for disk in floppy disk drive; make sure mainframe boots from the hard disk drive.  
Check for faulty module. Remove modules one at a time and power on the mainframe. If mainframe  
completes the power-on sequence, replace faulty module.  
Possible software failure or corrupted hard disk; see Software Problems at the beginning of this  
chapter.  
Power-on diagnostics fail  
Isolate problem to faulty mainframe or to faulty module. Multiple diagnostic failures across modules  
indicate a faulty mainframe. Diagnostic failures confined to an single module most likely indicate a  
faulty module. Contact your local Tektronix service center.  
21  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Table 4: Troubleshooting (Cont.)  
Symptom  
Possible causes and recommended action  
Mainframe does not recognize  
accessories such as monitor,  
printer, or keyboard  
Check that accessories are properly connected or installed. Try connecting other standard PC  
accessories or contact your local Tektronix service center.  
LA Module merge not recog-  
nized  
Modules not merged; check merge connection.  
Faulty module; replace with known-good module or contact your local Tektronix service center.  
Windows 95 comes up but the Mainframe not set up to start TLA 700 application at power on. Start application from the desktop, by  
TLA 700 application does not  
double-clicking on the TLA 700 icon; see the TLA 700 application online help on Powering On and  
Off for setting the startup sequence.  
Faulty or corrupt software; reinstall the TLA 700 application software.  
Exit the Safe mode and restart the mainframe.  
Windows 95 comes up in Safe  
mode  
Incompatible hardware and hardware driver software. Either install hardware driver or remove the  
incompatible hardware.  
TLA 700 application starts but  
modules do not display in  
System window  
Power off mainframe and check that all modules are fully inserted.  
Flash programming jumper installed. Power off mainframe and remove module. Remove jumper (see  
Figure 21 on page 50), and reinstall module.  
Module address switches not set correctly. Power off mainframe and remove module. Set address  
switches to FF and reinstall module (refer to Figure 21 on page 50 for address switch locations).  
Module failure; replace with known-good module or contact your local Tektronix service center.  
Mainframe failure; contact your local Tektronix service center.  
Portable Mainframe will not  
power off with On/Standby  
switch  
The Mainframe Utilities may be set up to disable hard power off. Check the setting of the Mainframe  
Utilities (the Mainframe Utilities are located in the Windows 95 Control Panel).  
The Windows 95 operating system problem. Try powering off the mainframe using the Windows 95  
shutdown procedure. If the mainframe still does not power off, disconnect power cord and reconnect  
after 10 seconds to reboot the mainframe.  
22  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
TLA 700 Startup Sequence  
The following information is intended to provide troubleshooting hints in case  
the logic analyzer fails to complete the startup sequence. You may want to refer  
to Figure 8 on page 24 while reading the following paragraphs.  
At power on, the logic analyzer starts the mainframe and module kernel tests. If  
the mainframe passes the kernel tests, it begins to boot the Windows 95  
operating system. If the mainframe fails the kernel tests, it displays the error  
codes, beeps, and terminates the startup sequence.  
The Windows 95 operating system starts the resource manager. The resource  
manager performs the following tasks:  
H
H
H
H
H
H
Starts the mainframe power-on self tests  
Verifies the power-on self test status  
Inhibits any failed modules  
Records the power-on self test failures  
Determines the logic analyzer configuration  
Executes the system controller power-on diagnostics  
After completing all of the above tasks, the logic analyzer starts the TLA 700  
application. The TLA 700 application performs the following tasks:  
H
H
H
Performs the power-on diagnostics on all installed modules  
Performs the power-on diagnostics on the TLA 700 system  
Records the Pass/Fail status in the Calibration and Diagnostics property  
sheet  
If no failures occur, the TLA 700 application is ready to use for regular tasks.  
23  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Power on  
Module 1  
kernel test  
Pass/Fail  
Module 2  
kernel test  
Pass/Fail  
Module 3  
kernel test  
Pass/Fail  
Module n  
kernel test  
Pass/Fail  
Mainframe kernel test  
Fail  
Terminate and  
display error  
code  
Pass/Fail ?  
Pass  
Verify kernel self test status  
Inhibit failed modules  
Log failures and build configuration  
MS Windows 95  
boot  
Start TLA 700 application  
Start resource  
manager  
Module power-on diagnostics  
System controller  
tests  
TLA 700 system power-on diagnostics  
Records failures in Calibration and  
Diagnostic property sheet  
TLA 700 ready  
Figure 8: TLA 700 startup sequence  
24  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Replacing Podlets on the P6417 Logic Analyzer Probe  
Although your logic analyzer was fully tested at the factory prior to shipment,  
there may be reasons to move or replace podlets on your logic analyzer probes.  
Moving Probe Podlets  
Depending on your application, you can move or reposition probe podlets as  
shown in Figure 9. When moving podlets in the podlet holder keep the ground  
side of the podlets together.  
Remove podlet  
cables  
Pull open the podlet holder  
Figure 9: Moving probe podlets  
25  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Replacing Probe Podlets  
The podlet replacement procedure consists of identifying the faulty podlet and  
replacing it with a new one. A podlet replacement kit is available as an optional  
accessory.  
NOTE. Refer to the TLA 700 Online Help for part numbers of the TLA 700  
accessories.  
Equipment and Required Material. You will need the following equipment and  
material:  
H
H
H
H
Replacement podlet  
Masking tape  
Small diagonal cutters  
Screwdriver with #1 Phillips tip  
Replacement Procedure. Perform the following steps to replace a faulty probe  
podlet:  
1. Identify and mark the faulty podlet.  
2. Disconnect the probe from the logic analyzer and set the probe cable on a  
clean working surface.  
NOTE. If you need to replace more than one podlet, replace the podlets one at a  
time to avoid mixing them up.  
26  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
3. Using the screwdriver with the #1 Phillips tip, remove the four screws on the  
probe housing on the module end of probe cable (see Figure 10).  
Top housing  
Cable header  
housing  
Rubber comb  
Bottom housing  
Podlet holder  
Figure 10: Disassembling the probe cable  
4. Remove the top and bottom probe housings and set them aside.  
27  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
5. Remove the remaining top and bottom probe housings and set them aside.  
Do not mix the probe connector-end housings with the podlet connector  
housings; the labels are different at each end of the probe.  
6. Identify the faulty probe channel at the module end of the cable.  
7. Refer to Figure 11 and use a small pointed tool, such as a straightened paper  
clip or a mechanical pencil, to release the faulty cable from the cable header  
housing. Discard the small plastic retaining snap.  
Paper clip  
Cable header housing  
Retaining snap  
Podlet cable  
Figure 11: Removing the faulty channel from the probe cable header housing  
8. Identify the faulty probe channel at the podlet end of the probe.  
CAUTION. Be sure that you have identified the correct podlet cable before you cut  
the cable.  
9. Cut the faulty podlet just above the color-coded ring.  
10. Remove the color-coded ring and set it aside; you will need to install it on  
the replacement cable later in this procedure.  
28  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
11. Refer to Figure 12 and bend the rubber comb to remove the faulty podlet  
cable from the comb.  
Remove faulty podlet cable from rubber comb  
Cut above color-coded ring  
Figure 12: Removing a podlet cable from the rubber comb  
12. Attach the module end of the replacement podlet cable to the cut end of the  
faulty podlet cable in the following manner:  
a. Lay the two cables in parallel with a four-inch (10 cm) to six-inch  
(15 cm) overlap.  
b. Wrap the overlapping portion of the two podlet cables tightly with  
masking tape to temporarily splice them together.  
13. Pull the faulty cable from the module end through the mesh sleeve so that  
the replacement cable is drawn through the mesh sleeve. Stop pulling when  
the replacement cable is located at the proper position in the cable wire  
bundle.  
29  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
14. Remove the faulty podlet cable from the podlet holder and replace it with the  
new cable as shown in Figure 13.  
Align the ground connectors  
Pull open  
Figure 13: Replacing a podlet in the podlet holder  
15. Take the color-coded ring that you removed in step 10 on page 28 and slide  
it onto the installation tool. (The installation tool is included in the podlet  
replacement kit.)  
CAUTION. Do not damage the cable with the installation tool when installing the  
color-coded ring.  
16. Refer to Figure 14 and place the installation tool over the replacement podlet  
cable as shown. Slide the color-coded ring down the installation tool onto the  
replacement cable. Position the ring so that it is in the same relative position  
as the other color-coded rings (approximately 0.5-inch from the podlet).  
30  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
Installation tool  
Install color-coded ring on  
the replacement cable and  
slide the ring down the tool.  
Replacement  
cable  
Figure 14: Installing the color-coded ring  
17. Push the new podlet cable into the rubber comb.  
18. Unwrap and remove the masking tape from the splice of the two cables and  
discard the faulty cable.  
19. Remove the protective 1x2 housing from the new podlet cable. Use the same  
procedure as in Step 7 on page 28 to remove the plastic retaining snap.  
Discard the retaining snap and the protective housing.  
31  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In Case of Problems  
20. Insert the new podlet cable partially into the cable header housing. Place a  
new retaining snap (included with the kit) on the podlet cable as shown in  
Figure 15 and fully insert the cable and retaining snap into the cable header  
housing. Tug lightly on the podlet cable to ensure the retaining snap is in  
place.  
Cable header housing  
Retaining snap  
Podlet cable  
Figure 15: Installing the cable in the cable header housing  
21. Reinstall the top and bottom probe housings on the module end of the probe  
cable in the following manner (refer to Figure 10 on page 27 if necessary):  
a. Line up the two tabs on the probe header with the two embedded slots in  
the bottom (deep) probe housing.  
NOTE. It may be necessary to push or pull the cables in the sleeve so that the  
probe header lines up properly with the slots in the bottom probe housing.  
b. Place the two metallic guides on the probe sleeve in the slots in the  
bottom probe housing.  
c. Line up the three tabs of the probe header with the three embedded slots  
in the top probe housing.  
d. Install and tighten the four screws in the probe housing.  
e. Smooth the sleeve by hand to ensure the sleeve fully covers the cables  
when replacing the probe housing over the comb.  
22. Repeat step 21 to install the probe housing over the rubber comb on the  
podlet end of the probe cable.  
32  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
This appendix describes how to merge two logic analyzer modules to form a  
wider module.  
There are no specific rules to merging modules. However, the following  
guidelines may be helpful.  
H
H
H
H
H
You can merge modules with 102 channels or greater.  
The maximum merged module width is two modules.  
The two modules should have the same firmware version.  
The two modules should have the same state speed.  
When merging modules of unequal channel widths, place the module with  
the higher number of channels in the lower numbered slot.  
H
Merged modules of unequal memory depths will assume the depth of the  
shallower module.  
Equipment Required  
Merge Procedure  
You will need the following equipment to complete these procedures:  
H
H
Torque screwdriver with a size T-10 Torx tip  
Two logic analyzer modules  
CAUTION. Static discharge can damage any semiconductor component in the  
logic analyzer module. Discharge the static voltage from your body by wearing a  
grounded antistatic wrist strap while performing the Merge Procedure.  
Perform the following steps to merge two logic analyzer modules:  
1. Power off the logic analyzer and remove the modules.  
2. Determine which module will be designated as the slave module (this  
module will be placed in the higher numbered slot).  
3. Lay the slave module on the right side (as viewed from the front panel).  
33  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
4. Use the screwdriver with the T-10 Torx tip and remove the eight screws from  
the cover (Six screws on the cover, and two screws on the rear panel, see  
Figure 16).  
5. Remove the four flat-head screws near the front of the module (102-channel  
modules only have two flat-head screws).  
6. Remove the two screws holding the Merge cable bracket to the cover.  
7. Remove the top part of the cable bracket and set it aside.  
8. Remove the cover and locate the Merge cable.  
Merge cable bracket  
Remove screws (2)  
Remove screws (4)  
Remove screws (8)  
Module cover  
Chassis  
Figure 16: Removing the cover  
34  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
9. Replace the cover while feeding the Merge cable through the hole in the  
cover (see Figure 17). Do not twist the cable while feeding it through the  
hole, otherwise the cable will not connect properly to the adjacent module.  
Remove screws (2)  
Merge cable bracket  
Merge cable  
Figure 17: Feeding the Merge cable through the cover  
10. Turn the Merge cable bracket over so that the guide pins point up and place  
the bracket over the Merge cable connector.  
11. Install the two screws that hold the Merge cable bracket in place.  
CAUTION. To prevent damage to the module during the installation process,  
reinstall the cover exactly as described in steps 12 through 16. If the cover is not  
properly seated, the module can be damaged when you install it in the logic  
analyzer or it will not meet EMC requirements.  
35  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
12. Push forward on the cover so the tab on the front edge of the cover inserts  
into the rear of the front subpanel. Make sure that the cover is fully seated  
(no gaps) against the front and rear chassis flanges (see Figure 18).  
Make sure tab inserts into  
slot on rear of front panel  
Leave no vertical gap  
Figure 18: Seating the cover on the chassis  
13. While holding the cover in place, install the four screws nearest the front of  
the module (two on the top and two on the bottom), to secure the cover to  
the chassis.  
14. Install the flat-head screws near the front of the module.  
15. Slide the rear panel on the chassis and install the two rear panel screws.  
16. Install the top and bottom rear screws.  
36  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
17. Check and tighten all screws to 4 in lbs.  
18. Place the Master module adjacent to the Slave module so that the two guide  
pins from the Slave module line up with the guide pin holes in the Master  
module (see Figure 19).  
Guide pins  
Master module  
Slave module  
Figure 19: Lining up the two modules  
19. Gently push the two modules together so that the Merge cable of the Slave  
module mates with the Merge cable connector on the Master module.  
Installing the Merged Modules in the Logic Analyzer  
After merging the two modules, perform the following steps to install the  
module pair in the logic analyzer:  
1. Hold the module pair such that the two modules do not become separated  
and line up the modules with the slot guides in the logic analyzer.  
2. Push both of the modules in the logic analyzer until they rest against the rear  
panel connector. Use the injector/ejector handles to fully seat the modules  
one at a time. Tighten the module retainer screws.  
37  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
3. Verify that the modules are fully seated before powering on the logic  
analyzer.  
4. After powering on the logic analyzer, run the merge calibration to calibrate  
the module pair (select the system menu, point to Calibration and Diagnos-  
tics, point to the Self Calibration property page, and select the merged pair in  
the list box).  
Removing Merged Modules from the Logic Analyzer  
Perform the following steps to remove merged modules from the logic analyzer:  
1. Identify the two logic analyzer modules that you want to remove.  
2. Loosen the retainer screws for both modules (top and bottom).  
NOTE. Use the ejector handles to just loosen the modules from the rear panel  
connector before sliding both modules out of the logic analyzer.  
3. Use the ejector handles to disconnect the first module from the rear panel  
connector.  
4. Use the ejector handles to disconnect the second module from the rear panel  
connector.  
5. Slide both modules out of the logic analyzer at the same time.  
6. Gently pry the two modules apart.  
Storing the Merge Cable  
Perform the following steps to store the Merge cable inside the logic analyzer  
module:  
1. Use the screwdriver with the T-10 Torx tip to remove the two screws that  
hold the Merge cable bracket in place.  
2. Remove the Merge cable bracket.  
3. Use the screwdriver with the T-10 Torx tip to remove the screws from the  
side cover and rear cover (see Figure 16 on page 34).  
4. Push the Merge cable through the side cover and remove the cover.  
5. Dress the Merge cable as shown in Figure 20.  
38  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
Merge cable  
Figure 20: Dressing the Merge cable before installing the cover  
CAUTION. To prevent damage to the module during the installation process,  
reinstall the cover exactly as described in steps 6 through 8. If the cover is not  
properly seated, the module can be damaged when you install it in a mainframe  
or it will not meet EMC requirements.  
6. Push forward on the cover so the tab on the front edge of the cover inserts  
into the rear of the front subpanel. Make sure that the cover is fully seated  
(no gaps) against the front and rear chassis flanges (see Figure 18 on  
page 36).  
7. While holding the cover in place, install the four screws nearest the front of  
the module (two on the top and two on the bottom), to secure the cover to  
the chassis.  
8. Install the flat-head screws near the front of the module.  
9. Slide the rear panel on the chassis and install the two rear panel screws.  
10. Install the top and bottom rear screws.  
11. Install the Merge cable bracket so that the guide pins point into the module.  
12. Install and tighten the two screws on the Merge cable bracket.  
13. Verify that you have installed and tightened all screws to 4 in lbs on the  
module.  
39  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A: Merging Logic Analyzer Modules  
40  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
This appendix describes procedures for upgrading your logic analyzer software  
or firmware. The following sections are included:  
H
H
H
Restoring or upgrading software in your logic analyzer  
Installing TPI, the TLA 700 Programmatic Interface (page 43)  
Restoring or upgrading firmware in your logic analyzer module or DSO  
module (page 50)  
H
H
H
Restoring or changing the Controller BIOS setup (page 52)  
Reinstalling Software (page 56)  
Customizing the TLA 700 Series setups (page 61)  
Upgrading Software  
This section provides information for restoring or upgrading software in your  
logic analyzer. In addition to the TLA 700 software, there are other software  
programs that are installed separately; Table 5 lists some of the software and  
installation information.  
Table 5: Software not covered by the TLA 700 Series software setup  
Software  
Installation information  
1
Microsoft Windows 95 OS  
Refer to the Windows 95 documentation or contact your local  
Microsoft representative.  
QAPlus Diagnostic Software  
Refer to the QAPlus online documentation or to Reinstall the  
QAPlus Diagnostic Software on page 59.  
Logic Analyzer Module & DSO Refer to the TLA 700 Series Performance Verification and  
Module Performance Verifica- Adjustment Technical Reference Manual for instructions on  
tion and Adjustment  
installing and using the PV/Adjust software.  
Microprocessor Support Soft- Refer to Installing Microprocessor Support Software beginning  
ware  
on page 15 in this manual.  
PC Card Software  
Refer to the instructions that come with your PC Card.  
Refer to the instructions that come with your software.  
Other software  
1
For information on installing Microsoft Windows 95 software not available on the  
soft.com/windows/software/cdextras.htm  
41  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Upgrading or Reinstalling  
TLA 700 Software  
You can update or reinstall the TLA 700 software by executing the setup.exe  
program on the Disk #1 of installation disks or by performing the following  
steps.  
NOTE. The following instructions assume that you are reinstalling the TLA 700  
application software only. If you are installing the software on a new hard disk  
drive, refer to the instructions under Reinstalling Software on page 56.  
1. If the TLA 700 application is running, exit the application before continuing.  
2. Insert the TLA 700 application floppy disk 1 in the disk drive.  
3. Click Start, point to Settings, and click Control Panel.  
4. In the Control Panel window, double-click Add/Remove Programs.  
5. Click on the Install button and follow the instructions on the screen for  
installing the software from the floppy disk.  
If the setup program finds an earlier version of the TLA 700 Application, the  
program will ask you to confirm your selection and then will remove the  
software. At the end of the removal process, the program will ask you to  
remove the floppy disk and then restart the logic analyzer to complete the  
file removal process.  
6. After the logic analyzer restarts, reinsert the floppy disk and repeat steps 1  
through 5 to install the application software.  
When executed, the setup program will replace old versions of the software after  
you have confirmed your choice. The program also offers to delete shared files  
which appear to be no longer needed by any application.  
The program will not remove saved TLA files or microprocessor support  
packages.  
Uninstalling TLA 700  
Software  
You can also remove or uninstall the TLA 700 software using the Uninstall  
program. Use the Add/Remove Programs tool under the Control Panel to  
uninstall software.  
NOTE. The TLA 700 Software Setup program automatically invokes the Uninstall  
program before reinstalling the TLA 700 software.  
42  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Installing TPI  
Use these procedures to install the TLA 700 Programmatic Interface (TPI). In  
these procedures, client refers to the user written application that controls the  
logic analyzer through the TPI. Server refers to the TLA 700 application  
software. For additional information on installation, refer to the online help.  
You have two methods of using TPI with your client application:  
H
You can run the client application locally on the logic analyzer. No special  
setup is required. To verify that a client application can connect to the  
TLA 700 server, see Running a Client Application on the Logic Analyzer.  
H
You can run your client application remotely across the network. In this case,  
both the logic analyzer and the remote host require special setup procedures.  
See Setting Up TPI for Remote Access on the Logic Analyzer to begin this  
setup procedure.  
In the following procedures, <install directory> refers to the directory where the  
TPI client is installed on your client machine. This directory is C:\Program  
Files\TLA 700 by default.  
If you need to use the type library (tla700.tlb), this file will be located in  
C:\Program Files\TLA 700\System\TPI on the logic analyzer and/or in the  
<install directory>\System\TPI on your client machine.  
Running a Client  
Application on the Logic  
Analyzer  
To verify that a client application on the logic analyzer can connect to the  
TLA 700 application, do the following steps:  
1. Start the TLA 700 application on the logic analyzer.  
2. Navigate to and run the following program: C:\Program Files\  
TLA 700\Samples\TPI Samples\Vc++\test client\testclient.exe  
3. The Tektronix Test TPI Client dialog appears. Click the Connect to  
TLA 700! button.  
4. Check that the text in the dialog changes to Connected! This indicates the  
test client application connected to the logic analyzer. Click Exit.  
The client application will run locally on the logic analyzer. For remote  
operation, see Setting Up TPI for Remote Access on the Logic Analyzer.  
Setting Up TPI for Remote  
Access on the Logic  
Analyzer  
If you want to run a client application remotely across the network, you must set  
up the logic analyzer using the following procedure:  
1. Install and configure TCP/IP. (networking)  
43  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
NOTE. If you have difficulty configuring the network setup, contact your system  
administrator.  
2. Download the following from Microsoft’s web site. Microsoft’s web site is:  
H
H
Distributed COM for Microsoft Windows 95 (DCOM95)  
DCOM for Microsoft Windows 95 configuration utility (only required for  
user level access)  
Install the downloaded files and restart the TLA 700 after each installation. For  
more information on DCOM, refer to the Microsoft website.  
3. You can choose to have share-level or user-level access to the logic analyzer  
as provided by Microsoft Windows 95.  
Share-level access allows a password to be assigned to each shared resource. For  
Microsoft Windows 95-only networks, share-level access is the only option.  
User-level access allows a group of users to have access to each shared resource.  
NOTE. For TPI to work with share-level access, authentication is turned off and  
any COM client can call into any COM server running on the logic analyzer.  
Setting the Access Level on the Logic Analyzer. To set up the access level on the  
logic analyzer, do the following steps:  
1. Click Start, point to Settings, and select Control Panel.  
2. From the Control Panel, select Network, then click the Access Control tab.  
3. To select share-level access, choose Share-level access control, and then  
click OK.  
To select user-level access, choose User-level access control and enter the  
name of the domain that will be used to validate user access. Click OK.  
4. Click Start again, and then Shut Down. In the Shut Down Windows dialog  
box, select Restart the Computer? and then click the Yes button.  
5. If you are setting up share level access, navigate to and double-click  
C:\Program Files\TLA 700\System\TPI\Share Level Access Server.reg  
If you are setting up user level access, navigate to and double-click  
C:\Program Files\TLA 700\System\TPI\User Level Access Server.reg  
6. A dialog appears, indicating successful registration. Click OK.  
44  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
7. Restart the logic analyzer, using the procedure in step 4.  
8. Start the TLA 700 application on the logic analyzer if it is not already  
running.  
This completes the setup of TPI on the logic analyzer for operating with a remote  
client machine.  
Next, you need to set up the client machine so that it can connect to the logic  
analyzer. Refer to the appropriate setup:  
H
H
H
Setting Up a Client Machine that is Running Microsoft Windows 95  
on page 45.  
Setting Up a Client Machine that is Running Microsoft Windows NT  
on page 48.  
Setting Up a Client Machine Running on Any Other Platform  
on page 49.  
Setting Up a Client  
Machine that is Running  
Microsoft Windows 95  
After you set up the logic analyzer, you must set up the client machine using the  
following procedure:  
1. Install and configure TCP/IP. (networking)  
2. Load the Tektronix TPI Client Install diskettes supplied with the TLA 700  
instrument.  
3. Double-click on Setup.exe.  
4. Download the following from Microsoft’s web site. Microsoft’s web site is:  
H
H
Distributed COM for Microsoft Windows 95 (DCOM95)  
DCOM for Microsoft Windows 95 configuration utility (only required for  
user level access)  
Install the downloaded files and restart the TLA 700 after each installation.For  
more information on DCOM, refer to the Microsoft website.  
You must set up the client machine to match the access level you chose for the  
logic analyzer (share-level or user-level access). Refer to the appropriate access  
type to continue with the setup procedure:  
H
H
Setting Up Share-level Access for Microsoft Windows 95 on page 46.  
Setting Up User-level Access for Microsoft Windows 95 on page 47.  
NOTE. You can switch between user- and share-level access later by redoing the  
Setting the Access Level procedure on page 44, from step 3 onward.  
45  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Setting Up Share-level Access for Microsoft Windows 95.To set up share-level  
access for a client machine running Microsoft Windows 95, do the following  
procedure on the client machine:  
1. Click Start, point to Settings, and select Control Panel.  
2. From the Control Panel, select Network, then click the Access Control tab.  
3. Choose Share-level access control, and then click OK. If prompted, insert the  
Windows 95 CD, or provide a filepath to the stored Windows 95 files.  
4. The System Settings Change dialog appears. Click the Yes button.  
5. Click Start, then Shut Down. In the Shut Down Windows dialog box, select  
Restart the Computer? and then click the Yes button.  
6. Navigate to and double-click <install directory>\System\TPI\Share Level  
Access Client.reg.  
7. A dialog appears, indicating successful registration. Click OK.  
8. Restart the client machine, using the procedure in step 5.  
9. In Microsoft Windows 95, click Start and then Run. Either locate the regedit  
file using Browse, or enter regedit, then click OK to run regedit.  
10. Click on the following registry key: HKEY_CLASSES_ROOT\AppID\  
{C67DAA22-4972-11d1-9CAC-00805F0D8271}  
11. Using Edit>New>StringValue, add a named value, RemoteServerName.  
12. Click the new value, RemoteServerName, and select Edit>Modify.  
The Edit String dialog appears.  
13. Enter the computer name of the logic analyzer. This is the name used to  
identify the logic analyzer on the network. Click OK.  
To verify that the setup is complete, do the following:  
1. Navigate to and run <install directory>\Samples\TPI Samples\Vc++\test  
client\testclient.exe on the client machine.  
2. The Tektronix Test TPI Client dialog appears. Click the Connect to  
TLA 700! button to see if the client can connect to the logic analyzer.  
3. Attempting Connection to TLA 700 appears in the dialog text box. (The first  
time you connect, it may take a few minutes.) The text changes to Con-  
nected! when the connection is made.  
This completes the setup of the logic analyzer and the Windows 95 client  
machine for running a client application across the network.  
46  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Setting Up User-level Access for Microsoft Windows 95. To set up user-level access  
for a client machine running Microsoft Windows 95, do the following procedure:  
1. Click Start, point to Settings, and select Control Panel.  
2. From the Control Panel, select Network, then select the Access Control tab.  
3. Choose User-level access control and enter the name of the domain that will  
be used to validate user access. Select OK.  
4. On the client machine, click Start and then Shut Down. In the Shut Down  
Windows dialog, select Restart the Computer? and then click the Yes button.  
5. Navigate to and double-click <install directory>\System\TPI\User Level  
Access Client.reg.  
6. Restart the client machine, using the procedure in step 4.  
7. In Microsoft Windows 95, click Start, and then select Run. Either locate the  
dcomcnfg file using Browse, or enter dcomcnfg in the edit field. Click OK.  
8. Double-click “Tektronix TLA 700 Application” in the Applications page.  
9. In the Location page, uncheck the “Run application on this computer” box  
and check the “Run application on the following computer” box. Enter the  
name of the logic analyzer in the edit field.  
To verify that the setup is complete:  
1. Run <install directory>\Samples\TPI Samples\Vc++\test client\test-  
client.exe on the client machine.  
2. The Tektronix Test TPI Client dialog appears. Click the Connect to  
TLA 700! button to see if the client can connect to the logic analyzer.  
3. Attempting Connection to TLA 700 appears in the dialog text box.  
(The first time you connect, it may take a few minutes.) The text changes to  
Connected! when the connection is made.  
This completes the setup of the logic analyzer and the client machine for running  
a client application across the network.  
NOTE. You can switch between share-level and user-level access later by  
uninstalling the Tektronix TPI Client and DCOM95 by using the Windows 95  
control panel. Redo the procedure from step 2 of Setting Up a Client Machine  
that is Running Microsoft Windows 95 on page 45.  
47  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Setting Up a Client  
Machine that is Running  
Microsoft Windows NT  
After you set up the logic analyzer, you must set up the client machine using the  
following procedure:  
1. Install and configure TCP/IP. (networking)  
NOTE. If you have difficulty configuring the network setup, contact your system  
administrator.  
2. Load the Tektronix TPI Client Install diskettes supplied with the logic  
analyzer.  
3. Double-click on Setup.exe.  
4. Click Start and then Shut Down. In the Shut Down Windows dialog, select  
Restart the Computer? and then click the Yes button.  
5. Depending on the type of access control you chose for the logic analyzer,  
double-click <install directory>\System\TPI\Share Level Access  
Client.reg or User Level Access Client.reg.  
6. Restart the client machine, using the procedure in step 4.  
7. In Microsoft Windows NT, click Start, and then select Run. Either locate the  
dcomcnfg file using Browse, or enter dcomcnfg. Then click OK.  
8. Double-click Tektronix TLA 700 Application in the Applications page.  
9. In the Location page, check the “Run application on the following computer”  
box. Enter the name of the logic analyzer in the edit field.  
To verify that the setup is complete, do the following:  
1. Run <install directory>\Samples\TPI Samples\Vc++\test client\test-  
client.exe on the client machine.  
2. The Tektronix Test TPI Client dialog appears. Click the Connect to  
TLA 700! button to see if the client can connect to the logic analyzer.  
3. Attempting Connection to TLA 700 appears in the dialog text box.  
(The first time you connect, it may take a few minutes.) The text changes to  
Connected! when the connection is made.  
This completes the setup of the logic analyzer and the Windows NT client  
machine for running your client application across the network.  
48  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Setting Up a Client  
Machine Running on Any  
Other Platform  
If the client application requires use of the type library, you can generate it on  
your platform using tla700.odl in C:\Program Files\TLA 700\System\TPI\src  
on the logic analyzer.  
Make sure that you perform the following steps:  
1. Ensure that DCOM is working on your platform.  
2. Merge C:\Program Files\TLA 700\System\TPI\Client.reg into your  
registry.  
3. Depending on the type of access control you chose for the logic analyzer,  
merge C:\Program Files\TLA 700\System\TPI\Share Level Access  
Client.reg or User Level Access Client.reg into your registry.  
4. Add a string value named RemoteServerName to the key  
HKEY_CLASSES_ROOT\AppID\  
{C67DAA22-4972-11d1-9CAC-00805F0D8271}  
5. Enter the computer name of the logic analyzer as the value of this string.  
This is the name used to identify the logic analyzer on the network.  
This completes the setup of the logic analyzer for operating with a remote client  
machine using a platform other than Microsoft Windows NT or Microsoft  
Windows 95.  
49  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Updating Module Firmware  
Perform the following steps if you need to update or reinstall the flash ROM-  
based firmware on your application modules. You can update multiple modules  
and module types during a single firmware update session.  
1. Power off the logic analyzer.  
2. Disconnect the power cord.  
3. Disconnect any probes on the modules that you want to upgrade.  
4. Remove the module from the logic analyzer.  
5. Refer to Figure 21 and locate the Flash programming pins on the rear of the  
module.  
Flash programming pins  
Figure 21: Flash programming pins  
6. Install a jumper on the Flash programming pins (use one of the spare  
jumpers that came with your logic analyzer software).  
50  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
7. Reinstall the modules in the logic analyzer.  
8. Reconnect the power cord and power on the logic analyzer.  
NOTE. When the TLA 700 application starts, any modules where you have  
installed the Flash programming jumper will not display in the system menu.  
9. Exit the TLA 700 application.  
10. Click Start, point to Programs, point to Tektronix TLA 700, and click  
TLA 700 Firmware Loader.  
11. Select the modules that you want to update from the list of modules  
displayed in the Supported list box near the top of the menu. If you are  
updating more than one module at a time, note the slot location of the  
module in the selection box.  
12. Select Load Firmware from the Execute menu.  
13. Click the proper .LOD file for the module in the C:\Program Files\ TLA 700  
Firmware directory. If you are updating more than one module make sure  
that you select the proper firmware file for the module in the listed slot  
number.  
14. Click Ok. You will be asked to confirm your action; click Yes when  
prompted.  
NOTE. The program will not allow you to load firmware to an incompatible  
module. For example, the program will not load DSO firmware to a logic  
analyzer module.  
15. If you are updating another module, select the proper .LOD file for the  
module.  
16. Click Ok. You will be asked to confirm your action; click Yes when  
prompted.  
The program will load the firmware for each module one at a time. The  
process takes approximately five minutes per module.  
17. When the load operation is complete, select each of the modules in the  
Supported list and then select Checksum from the Execute menu.  
18. If each module passes the checksum test, exit the program.  
19. Power off the logic analyzer and disconnect the power cord.  
51  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
20. Remove the module from the logic analyzer and remove the jumper from the  
Flash programming pins on the rear of the module. Keep the jumper for  
future updates.  
21. Attach a new label with the new firmware version to the module.  
22. Reinstall the module in the logic analyzer and reconnect the probes.  
23. Reconnect the power cord and power on the logic analyzer.  
24. Verify that the power-on diagnostics pass.  
25. Run the self calibration on each module after a 30-minute warm-up.  
Controller BIOS Setup  
The logic analyzer has a built-in setup program where you can set some basic  
input and output characteristics. In most cases, there is no need to access the  
setup program. If you need to access the setup program, power on the logic  
analyzer and press function key F2 on the keyboard (ALT, CTRL, F2 keys  
simultaneously on the portable mainframe front panel) before the logic analyzer  
boots the Windows 95 operating system.  
Table 6 on page 53 lists the default settings for the BIOS setup. You may need to  
refer to this information after updating the software or making other changes.  
CAUTION. Do not change any of the settings if you are unfamiliar with the setup  
program. Selecting an incorrect setup can cause the logic analyzer to malfunc-  
tion.  
If you need to change any of the settings to match those listed in Table 6, follow  
the on-screen instructions.  
NOTE. If you need to upgrade the Controller BIOS version, refer to the Mainte-  
nance chapter of the TLA 711 Color Benchtop Controller Service Manual or the  
TLA 704 Color Portable Mainframe Service Manual for the upgrade instruc-  
tions.  
52  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Table 6: Controller BIOS setups for the TLA 700 application  
Parameter  
Main menu  
Factory setting  
Submenu parameter  
Factory setting  
System Time  
Set to current date  
Set to current time  
1.44 MB, 3 1/2  
System Date  
Diskette A  
Diskette B  
Not Installed  
IDE Adapter 0 Master  
(C:[hard disk size])  
Autotype Fixed Disk  
Type  
Select Enter  
Hard disk size  
Number of cylinders  
Number of heads  
Number of sectors/track  
None  
Cylinders  
Heads  
Sectors/track  
Write Precomp  
Multi-sector transfers  
LBA Mode Control  
32-Bit I/O  
Set by Autotype  
Set by Autotype  
Disabled  
Transfer Mode  
Set by Autotype  
IDE Adapter 0 Slave  
IDE Adapter 1 Master  
IDE Adapter 1 Slave  
Video System  
(None)  
(None)  
(None)  
EGA/VGA  
Memory Cache  
External Cache  
Enabled  
Enabled  
Enabled  
Cache System BIOS area:  
Cache Video BIOS area:  
Cache Memory Region  
CC00 – CFFF:  
Disabled  
Disabled  
Disabled  
Disabled  
Disabled  
Enabled  
Enabled  
D000 – D3FF:  
D400 – D7FF:  
D800 – DBFF:  
DC00 – DFFF:  
Memory Shadow  
System Shadow  
Video Shadow:  
Regions with Legacy Expan-  
sion ROMS  
CC00 – CFFF:  
ROM  
53  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)  
Parameter  
Factory setting  
Submenu parameter  
D000 – D3FF:  
Factory setting  
ROM  
D400 – D7FF:  
ROM  
D800 – DBFF:  
DC00 – DFFF:  
Boot Delay  
ROM  
ROM  
Boot Sequence:  
A: then C:  
0
Boot Sequence:  
Setup Prompt  
A: then C:  
Enabled  
Enabled  
Enabled  
Enabled  
Off  
POST Errors  
Floppy Check  
Summary Screen  
NUMLock  
NUMLOCK:  
OFF  
Key Click  
Disabled  
30/sec  
1/4 sec  
Keyboard auto-repeat rate  
Keyboard auto-repeat delay  
System Memory  
Extended Memory  
Advanced menu  
640 KB  
Installed memory – 640 K  
Integrated Peripherals  
COM A  
3F8, IRQ4  
2F8, IRQ3  
378, IRQ7  
ECP  
COM B  
LPT Port  
LPT Mode  
Diskette Controller  
Local Bus IDE Adapter  
DRAM Speed  
DMA Aliasing  
8 Bit I/O Recovery  
16 Bit I/O Recovery  
IRQ12 Used By  
ECC/Parity  
Enabled  
both  
1
Advanced Chipset Control  
60 ns  
Enabled  
4.5  
4.5  
PS/2 Mouse  
Disabled  
Plug & Play O/S  
No  
Reset Configuration Data  
Large Disk Access Mode:  
Enable SimulScan Mode:  
No  
DOS  
Disabled  
54  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)  
Parameter  
Flash ATA PC Card Booting:  
Factory setting  
Submenu parameter  
Factory setting  
Disabled  
Power menu  
2
APM:  
Enabled  
3
Power Savings  
Customize  
Standby Timeout:  
Disabled  
Disabled  
MAX  
Suspend Timeout:  
Standby CPU Speed:  
Fixed Disk Timeout:  
Disabled  
ON  
CRT:  
Standby Timer Reset Events  
Keyboard  
Enabled  
Mouse  
PS/2 (IRQ12)  
Standby Break Events  
IRQ0  
IRQ1  
Disabled  
Auto  
IRQ3  
Auto  
IRQ4  
Auto  
IRQ5  
Disabled  
Disabled  
Disabled  
Disabled  
Disabled  
Disabled  
Disabled  
Auto  
IRQ6  
IRQ7  
IRQ8  
IRQ9  
IRQ10  
IRQ11  
IRQ12  
IRQ13  
Disabled  
Disabled  
Disabled  
IRQ14  
IRQ15  
Standby Wakeup Events  
Keyboard  
Enabled  
55  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Table 6: Controller BIOS setups for the TLA 700 application (Cont.)  
Parameter  
Mouse  
Factory setting  
Submenu parameter  
Factory setting  
PS/2 (IRQ12)  
1
2
3
DRAM speed was set to 70 ns for TLA 700 Series controllers with BIOS version 1.00.11 or earlier.  
APM is set to Disabled for TLA 700 Series controllers with BIOS version 1.00.11 or earlier.  
Power Savings is set to Customize when the APM is enabled. This allows Windows to manage power use for the  
instrument as indicated on the Power property page of the Power icon in the Control Panel. Power Savings was set to  
Disabled with BIOS version 1.00.11 or earlier.  
Reinstalling Software  
There may be occasions when you have to reinstall software or restore software  
after replacing the hard disk. The procedures in this section explain how to  
reinstall software on the hard disk. If you replace the hard disk, you must also  
configure the controller BIOS. To configure the controller BIOS, follow steps  
under Configure the Controller BIOS on page 56.  
On rare occasions you may need to reformat the hard disk. If so, follow the steps  
under Reformatting the Hard Disk on page 57. Otherwise, follow the steps under  
Reinstalling the Windows 95 Software on page 58.  
NOTE. You should perform all of the procedures in this section in sequence.  
Configure the Controller  
BIOS  
This procedure is necessary after replacing the hard disk or when the CMOS  
settings are corrupted or lost. You will need the the emergency startup disk that  
you created when you installed your logic analyzer (see Create an Emergency  
Startup Disk on page 13). To configure the Controller BIOS, complete the  
following steps:  
1. Power off the logic analyzer and insert the emergency startup disk in the  
floppy disk drive.  
2. Power on the logic analyzer.  
3. When prompted, press function key F2 on the keyboard before the logic  
analyzer boots the Windows 95 operating system.  
4. In the Main Setup menu, use the Down Arrow key to scroll to the line  
labeled IDE Adapter 0 Master.  
5. Press the Enter key to display the hard disk detail submenu.  
56  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
6. Press the Enter key again to cause the logic analyzer to automatically  
determine and set the hard disk drive parameters.  
7. Press the ESC key twice and then the Enter key twice to exit and save the  
BIOS setup.  
Reformatting the Hard  
Disk  
You should not reformat your hard disk drive unless it is absolutely necessary. If  
you find that you need to format the hard disk, (if possible) make sure that you  
have backup copies of your user files so that you can reinstall them later.  
You must perform the following procedure after you replace the hard disk, or  
when the partitioning or the format of the hard disk has been damaged. You will  
use the Setup program on the Windows 95 Setup Boot floppy disk to partition  
and format the hard disk.  
Complete the following steps to format the hard disk:  
1. Insert the Windows 95 Setup Boot floppy disk in the floppy disk drive.  
2. Power on the logic analyzer so that it boots from the floppy disk.  
3. Press function key F3 to exit the Setup program; press F3 again to confirm  
your selection.  
You must temporarily exit the Setup program to properly execute the FDISK  
command with the /X switch. If you do not execute the FDISK /X command  
separately, the primary DOS partition will not be compatible with the  
TLA 700 Series BIOS.  
4. At the prompt, enter the command FDISK /X.  
5. Enter N when you are asked if your want to select Large Disk Support.  
6. Select option 1 to create a new partition on the hard disk.  
7. Select option 1 again to specify that the new partition is to be the primary  
DOS partition.  
8. Enter Y to use the maximum available size for the DOS partition.  
9. Press the CTRL, ALT, and Delete keys simultaneously to reboot the logic  
analyzer and restart the Setup program.  
When the Setup program restarts, it will caution you that there is an  
operating system already installed. Disregard this message. The message  
appears because you partitioned the hard disk in the previous steps.  
10. Continue with the Setup program and follow the instructions on the display.  
The Setup program will format the hard disk and exit the program when  
complete.  
57  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Reinstalling the  
Windows 95 Software  
The following procedure assumes that the hard disk is already properly for-  
matted. To reinstall the Windows 95 software, complete the following steps:  
1. Start the logic analyzer from the Windows 95 Setup Boot floppy disk.  
2. When the install program prompts you for an installation, select Typical.  
3. Enter the Windows 95 Product Identification number located above the bar  
code on the Windows 95 software package.  
4. Enter your name when prompted.  
5. Enter your company name when prompted (optional).  
6. When prompted, select the appropriate hardware categories (for example,  
Network).  
7. Install the Windows 95 floppy disks one at a time until you have installed all  
of the floppy disks.  
8. Enter the current date, time, and time zone when prompted.  
9. Restart the logic analyzer when prompted. The logic analyzer should start up  
in Windows 95. You are now ready to install the TLA 700 Series application  
software.  
Set the Display Driver  
After installing the Windows 95 software, you must set the display driver for  
your monitor. Otherwise, using the default display driver may not provide the  
optimum display settings. To set the display driver, complete the following steps:  
1. Click Start, point to Settings, click on Control Panel, and open the Display  
icon.  
2. Select the Settings property page.  
3. Click on the Advanced Properties button.  
4. Select the Adapter tab page, and then click on the Change button.  
5. Click on the Have Disk button.  
6. Insert the floppy disk labeled Cirrus Logic Video Display Driver  
GDS7543 – Rel 2.40 for Windows 95 and click OK.  
7. Select the driver named Cirrus Logic 7543 2.40 PCI(Cirrus) and then click  
OK.  
NOTE. If a Version Conflict dialog box appears and asks you if you want to keep  
the file, answer Yes.  
58  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
8. After the installation is complete, close all of the dialogs and property sheets.  
However, keep the Control Panel open.  
9. When the prompt appears asking you if you want to reboot now, enter No.  
Reinstall the QAPlus  
Diagnostic Software  
To reinstall the QAPlus diagnostic software, complete the following steps:  
1. Insert the QAPlus floppy disk into the disk drive.  
2. In the Control Panel window, double-click Add/Remove Programs.  
3. Click on the Install button to begin installing the software.  
4. When the program prompts you to confirm the creation of the C:\win–win  
directory, click Continue.  
NOTE. If you use a set of QAPlus disks which have been previously installed, you  
will not be able to enter a user name. The program will warn you that the disks  
have been previously installed and then ask you to acknowledge that fact before  
proceeding.  
5. When the program asks you if you want to create a set of icons for the  
program manager, click Create.  
After completing the installation, the program opens up the Notepad  
application and displays the release notes for the QAPlus software.  
6. Close the Notepad application and press the CTRL, ALT, and Delete keys  
simultaneously to display a list of current tasks.  
7. Select the QAPlus task, and click End task to kill the task.  
8. After a few seconds, a dialog box displays advising you that the application  
is not responding. Click End Task again to kill the application.  
9. When the desktop reappears, close the Program Manager folder.  
10. Click Start, point to Programs, and click on MS-DOS Prompt to open an  
MS-DOS window.  
11. In the MS-DOS window, change directories to the C:\win-win directory.  
12. Insert the QAPlus/WIN disk labeled QAplus\win-win Manual  
Version 7.01.  
13. Copy the file A:\Qawwusro.exe to the C:\win-win directory.  
14. Exit and close the MS-DOS window.  
59  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Reinstall the TLA 700  
Series Application  
Software  
To reinstall the TLA 700 Series application software, complete the following  
steps:  
1. Insert the TLA 700 application floppy disk 1 in the disk drive.  
2. Click Start, point to Settings, and click on Control Panel.  
3. In the Control Panel window, double-click Add/Remove Programs.  
4. Click on the Install button and follow the instructions on the screen for  
installing the software from the floppy disk.  
5. When the installation is complete, the program will prompt you to reboot the  
system; click on OK to reboot.  
Reinstalling other  
Application Software  
To reinstall any other application software such as microprocessor support  
software, complete the following steps:  
1. Insert the first floppy disk into the disk drive.  
2. Click Start, point to Settings, and click Control Panel.  
3. In the Control Panel window, double-click Add/Remove Programs.  
4. Click on the Install button and follow the instructions on the screen for  
installing the software from the floppy disk.  
Restoring User Files  
After you have reinstalled all of the application software files, you should restore  
any backed-up user files. Use the Windows 95 Back Up tool to restore any saved  
files from floppy disks. Use the online help for instructions on restoring the files.  
60  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Customizing TLA 700 Series Setups  
After restoring all of the user and application files, you must perform the  
following steps to complete the software installation process. The steps in this  
section will return your logic analyzer to the same settings it had when it was  
shipped from the factory. The following steps are included in this section:  
H
H
H
H
H
Creating a shortcut for the QAPlus online manual  
Adjusting the display desktop settings for an external monitor  
Adjusting the refresh rate  
Setting up the screen saver  
Setting up the Virtual Memory  
NOTE. You must exit the TLA 700 Application before completing the procedures  
in this section. The TLA 700 Application starts automatically each time you  
power on the logic analyzer.  
Create a Shortcut for the  
QAPlus Online Manual  
Complete the following steps to create a shortcut from the Start menu for the  
QAPlus online manual:  
1. Use the Windows Explorer (right-click on Start and select Explore) to open  
the directory C:\Windows\Start Menu\Programs\QAPlus_WIN-WIN.  
2. Select the QAPlus_WIN-WIN directory. From the File menu, select New  
and then Create Shortcut.  
3. Enter the name of the file C:\WIN-WIN\Qawwusro.exe in the field of the  
Create Shortcut dialog.  
4. Click Next and set the name of the shortcut to QAPlus_WIN-WIN Manual.  
5. Click Finish to install the shortcut in the Start menu.  
6. Close the Windows Explorer.  
61  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Adjust the Display  
Desktop Settings for an  
External Monitor  
The following steps are only necessary if your logic analyzer uses an external  
monitor. The following steps are a generalized procedure for a standard Plug &  
Play monitor. If you know the type of monitor that you will use, select the  
monitor type in steps 6 and 7 and adjust the remaining parameters for your  
specific monitor.  
1. Click Start, point to Settings, click on Control Panel, and open the Display  
icon.  
2. Select the Settings property page.  
3. Click on the Advanced Properties button.  
4. Select the Monitor tab page and then click on the Change button.  
5. Click on the radio button to Show All Devices.  
6. Select Standard Monitor Types under the Manufacturers column.  
7. Select Plug & Play Monitor under the Model Type column.  
8. Click OK to close the selection window.  
9. Click Close to close the Advanced Properties dialog.  
10. Set the following parameters in the Settings property page:  
H
H
Color Palette  
Desktop Area  
256 Colors  
640 x 480  
11. Leave the Display property page open and continue with the next procedure  
to adjust the refresh rate.  
Adjust the Refresh Rate  
A higher monitor refresh rate provides a brighter, clearer display. It also reduces  
eye fatigue from screen flicker. The following recommendations set the refresh  
rate to the highest possible settings and may not be optimal for all display types.  
You may want to try different settings to determine which work best for your  
specific monitor.  
1. Select the Monitor Refresh property page.  
2. Click the Set Highest button to use the highest available refresh rate for each  
desktop selection.  
3. Click OK to close the Monitor Refresh property page.  
You will see a warning dialog informing you that the new settings are about  
to be tested and that the refresh settings will revert to the old settings within  
10 seconds if you do not confirm them.  
62  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
4. Click OK. After the new settings are in effect, click Yes to keep the new  
settings.  
5. Leave the Display property page open and continue with the next procedure  
to select and configure the screen saver.  
Set Up the Screen Saver  
The screen saver includes the option of reducing the intensity of the TLA 700  
Series LCD display backlight to help prolong the life of the display. Select this  
recommended screen saver unless you will always use an external monitor.  
1. Select the Screen Saver property page.  
2. Select Tektronix LCD Saver from the list of screen savers.  
3. Set the Wait period for ten minutes.  
4. Click the Settings button to open the screen saver dialog.  
5. Set the speed to Fast and the Intensity to Reduced (60%).  
6. Click OK to close the dialog.  
7. Click OK to close the Display window.  
Set Up the Mainframe  
Utilities  
The Mainframe Utilities provide additional settings for your logic analyzer (the  
utilities can only be used with the portable mainframe).  
1. Click Start, point to Settings, click on Control Panel, and open the Main-  
frame Utilities icon.  
2. Select the Power Off Mode tab page.  
3. Click on the radio button named Enable Soft Power Off.  
4. Select the LCD Mode tab page.  
5. Click on the radio button named LCD Full On.  
6. Click OK to close the utility.  
63  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Upgrades  
Setup the Virtual Memory  
The following virtual memory settings ensure optimal performance of the  
TLA 700 Series application software.  
1. Open the System icon in the Control Panel.  
2. Select the Performance property page and select the Virtual Memory button.  
3. In the Virtual Memory dialog, click the radio button Let me specify my own  
virtual memory settings.  
4. Enter a minimum value of 100 MB and a maximum value of 500 MB.  
5. Click OK to close the dialog.  
6. Select Yes to when prompted to confirm your settings.  
7. Click OK to close the System window.  
8. When the prompt appears asking you if you want to reboot now, enter No.  
Enable 32-Bit Support for  
the PCMCIA Interface  
Perform the following steps to enable the 32-bit support for the PCMCIA  
Interface:  
NOTE. You should complete this procedure even if you do not intend to use the  
PCMCIA interface.  
1. Open the System icon in the Control Panel.  
2. Select the Performance property page and then click on the text PC card  
(PCMCIA) devices are not using 32-bit support.  
3. Click on the Details button.  
4. In the Help window, click on the button to enable 32-bit PC card support.  
5. When asked if you are using a PC card to install Windows, select No and  
click Next.  
6. When asked to review system files to select real-mode PC card drivers for  
Windows to delete, select No and click Next.  
7. Close the Help window.  
8. Click Finish.  
9. When asked if it is ok to shut down the computer to complete the installation  
of 32-bit support, select Yes.  
64  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix C: Power Cord and Line Fuse Requirements for  
the Benchtop Mainframe  
The benchtop mainframe comes with two power cords and three fuses (one fuse  
is already installed). You must determine the correct fuse and power cord for  
your configuration. This is important to avoid overloading the power distribution  
system and ensures that you comply with the National Electrical Code.  
The power consumption depends on the number and type of instrument modules  
installed in the benchtop mainframe. Table 7 lists the power consumed for each  
module.  
To determine the total power consumption perform the following steps:  
1. Use Table 7 to determine the power consumption for each module in the  
benchtop mainframe.  
2. Add the power for each module to determine the total power consumption.  
3. Determine at which line voltage you will be operating the mainframe.  
4. Refer to Figure 22 on page 66 to determine the proper power cord and line  
fuse for your mainframe.  
Table 7: Power for instrument modules  
Module type  
Benchtop Controller  
TLA 7L1  
Power (watts)  
40  
55  
73  
94  
109  
57  
76  
99  
115  
56  
81  
56  
81  
TLA 7L2  
TLA 7L3  
TLA 7L4  
TLA 7M1  
TLA 7M2  
TLA 7M3  
TLA 7M4  
TLA 7D1  
TLA 7D2  
TLA 7E1  
TLA 7E2  
65  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix C: Power Cord and Line Fuse Requirements for the Benchtop Mainframe  
For power usage in the nonshaded region of Figure 22, use either the power cord  
with the 15 A plug (two parallel prongs and ground) or the power cord with the  
20 A plug (two perpendicular prongs and ground). For high power usage  
combined with low input line voltages (shaded region), use only the power cord  
with the 20 A plug. Select the proper fuse based on the ranges shown in  
Figure 22.  
Power Cord with 20A plug (NEMA 5–20P) ONLY!  
800  
700  
600  
500  
400  
300  
200  
100  
0
Power Cord with 15A plug (NEMA 5–15P)  
Power Cord with 20A plug (NEMA 5–20P)  
90  
110  
115  
20AT Fuse  
15AF Fuse  
130  
150  
170  
190  
210  
230  
250  
Input Line Voltage (Volts AC)  
6.3AF Fuse  
Figure 22: Power cord identification chart  
For example, assume that your configuration consists of four TLA 7M4 Logic  
Analyzer Modules and one TLA 7E2 DSO Module. Also assume that you will  
be operating the mainframe at 90 VAC.  
Add 40 watts (for the controller) + 81 watts (for the DSO Module) + 4 X  
115 watts (for the logic analyzer modules) to come up with a total value of  
581 watts. Because the power consumption (at 90 VAC) is approaching the  
shaded area of the graph in Figure 22, you should consider using the 20A power  
cord.  
66  
TLA 700 Series Installation Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tatung DVD Recorder TDR 22XX User Manual
Therma Stor Products Group Dehumidifier 200 MAX User Manual
Tiffen Carrying Case F 10 User Manual
Toshiba Laptop U840W User Manual
TP Link Network Router Rev 100 1910010810 User Manual
Triarch Indoor Furnishings 33190 1 User Manual
Uniden Cordless Telephone XS 1410 User Manual
WatchGuard Technologies Automobile Electronics DV 1 User Manual
Weber Gas Grill 54265 User Manual
Weber Gas Grill 55009 User Manual