Sierra Monitor Corporation Carbon Monoxide Alarm 5100 03 IT User Manual |
MODEL 5100-XX-IT
IT Series
TOXIC GAS SENSOR MODULE
Version 2.0
APPLICABILITY & EFFECTIVITY
Effective for all Model 5100-XX-IT and 5100-XX-IT-SS Modules manufactured after March 1, 2008
including.
5100-03-IT
5100-04-IT
5100-05-IT
5100-06-IT
Instruction Manual Part Number T12020
Rev A
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE OF CONTENTS
1.
PRODUCT DESCRIPTION .............................................................................................................................1
1.1
1.2
1.3
1.4
GENERAL........................................................................................................................................................1
PRODUCT CONFIGURATION........................................................................................................................1
THEORY OF OPERATION..............................................................................................................................1
MODES OF OPERATION................................................................................................................................1
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
1.4.6
SENTRY INTERFACE..............................................................................................................................1
MODBUS OPERATION............................................................................................................................2
ANALOG OPERATION ............................................................................................................................2
REMOTE SENSOR AND DISPLAY (APPENDIX D)................................................................................2
REMOTE ALARM RESET (DIGITAL INPUT) (Figure 3-7) ......................................................................2
OPTIONAL INTEGRAL RELAYS.............................................................................................................2
MECHANICAL ............................................................................................................................................................2
1.4.7
1.4.8
1.4.9
ENCLOSURE ...........................................................................................................................................2
TRANSMITTER ELECTRONICS .............................................................................................................4
SENSOR ASSEMBLY..............................................................................................................................4
1.5
1.6
INTERCONNECT WIRING..............................................................................................................................4
POWER REQUIREMENTS .............................................................................................................................4
2.
CAUTIONS & WARNINGS .............................................................................................................................5
2.1
2.2
2.3
INTRODUCTION .............................................................................................................................................5
IT MODULES - GENERAL...............................................................................................................................5
WIRING............................................................................................................................................................5
3.
INSTALLATION ..............................................................................................................................................7
3.1
3.2
SENSOR MODULE LOCATIONS....................................................................................................................7
WIRING............................................................................................................................................................7
3.2.1
3.2.2
3.2.3
3.2.4
ANALOG 4-20 mA OPERATION..............................................................................................................7
MODBUS OPERATION USING RS-485 CONNECTION ........................................................................7
SENTRY OPERATION USING SMC SENTRY CONNECTION..............................................................7
GENERAL ................................................................................................................................................7
3.3
3.4
3.5
ENCLOSURE INSTALLATION........................................................................................................................8
TRANSMITTER AND SENSOR INSTALLATION............................................................................................8
MODULE ADDRESS SWITCH......................................................................................................................10
4.
OPERATION .................................................................................................................................................17
4.1
4.2
4.3
4.4
4.5
INTRODUCTION – HUMAN-MACHINE INTERFACE SYSTEM ...................................................................17
HUMAN-MACHINE INTERFACE ..................................................................................................................17
MAIN MENU...................................................................................................................................................18
CONFIGURE SET-POINTS...........................................................................................................................19
MAINTENANCE FUNCTIONS.......................................................................................................................22
5.
CALIBRATION..............................................................................................................................................24
5.1
5.2
CALIBRATION FREQUENCY .......................................................................................................................24
CALIBRATION PREPARATION....................................................................................................................24
5.2.1
CALIBRATION GAS DELIVERY METHODS.........................................................................................24
5.3
CALIBRATION PROCEDURE.......................................................................................................................24
5.3.1
SENSOR EXPOSURE TO GAS.............................................................................................................25
6.
SERVICE .......................................................................................................................................................26
6.1
6.2
6.3
6.4
SENSOR MODULE CONFIGURATION ........................................................................................................26
ENCLOSURE REPLACEMENT ....................................................................................................................26
TRANSMITTER REPLACEMENT .................................................................................................................27
SENSOR REPLACEMENT............................................................................................................................27
Contents
Download from Www.Somanuals.com. All Manuals Search And Download.
6.5
6.5.2
6.5.3
6.5.4
INSTALLATION INSPECTION ......................................................................................................................27
INSPECTION AND TROUBLESHOOTING GUIDE...............................................................................28
IF MODULE DOES NOT RESPOND TO GAS ......................................................................................28
IF THE MODULE DISPLAYS “STARTING” FOR MORE THAN 1 HOUR .............................................28
7.
APPENDICES ...............................................................................................................................................29
APPENDIX A - SPECIFICATIONS ...........................................................................................................................29
APPENDIX B: - MODEL NUMBERS & PARTS LIST ................................................................................................31
APPENDIX C: LIMITED WARRANTY ......................................................................................................................32
APPENDIX D: REMOTE SENSOR/DISPLAY DRAWING 5394-52..........................................................................33
APPENDIX E: MODBUS MEMORY MAP.................................................................................................................34
Contents
Download from Www.Somanuals.com. All Manuals Search And Download.
1. PRODUCT DESCRIPTION
1.1 GENERAL
The Model 5100-XX-IT Toxics Gas Sensor Module is a member of the Information Technology ”IT” family of
gas sensor transmitter modules. Members of the IT Toxics family include:
•
•
•
•
5100-03-IT Oxygen Deficiency Gas Sensor Module
5100-04-IT Carbon Monoxide Gas Sensor Module
5100-05-IT Hydrogen Sulfide Gas Sensor Module
5100-06-IT Chlorine Gas Sensor Module
IT modules offer a broad array of features, including:
•
•
•
•
•
•
•
•
•
Integral alphanumeric LED display
180 day calibration frequency
4-20 mA output
Modbus® RTU interface
SMC Sentry interface
316 Stainless steel enclosure option
Auxiliary integral relay option
Remote display option
Low maintenance and operation costs
IT modules are designed, and approved for installation and operation in hazardous locations.
1.2 PRODUCT CONFIGURATION
Various module configuration options are available. Where applicable, these options are factory configured
prior to shipment. Options which can be re-configured by the installer or field technician are fully described in
this manual.
1.3 THEORY OF OPERATION
Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode and electrolyte. The
components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical
reaction and generate a current. The cells are diffusion limited so the rate the gas enters the cell is solely
dependent on the gas concentration. The current generated is proportional to the rate of consumption of the
subject gas in the cell.
1.4 MODES OF OPERATION
1.4.1 SENTRY INTERFACE
IT gas sensor modules can be installed on Sierra Monitor Sentry Model 5000 controllers. The gas sensor
module communicates as a toxic gas sensor module (Type 4 – communication) and are automatically
detected by the Sentry controller. When it is installed in a Sentry system the IT module must have a unique
address which can be established by setting an address between 1 and 8 on the Module Address Switch
® Registered trademark of Schneider Electric
Page: 1
Download from Www.Somanuals.com. All Manuals Search And Download.
accessible from the cover plate as illustrated in Figure 3-1. Figure 3-6 in this manual provides the wiring
terminations for connections to the Sentry controller.
When the module is operated in conjunction with a Sentry controller, the alarms outputs in the module can
both be set to “Sentry”, allowing the Sentry controller to manage alarm acknowledge and reset rather than
the sensor module. The alarm relay outputs are triggered by the alarm values established in the module,
and can be independent of the settings in the Sentry controller.
The Sentry configuration allows daisy chain installation using the Sentry multiplex capability, thus reducing
costs by avoiding the requirement for wiring junction boxes. The connector card has two sets of
connections, allowing for a continuous run to the next module.
1.4.2 MODBUS OPERATION
An RS-485 Modbus RTU serial interface allows direct connection to standard PLCs and DCSs. The
Module Address Switch (section 3.5) allows the user to select up to 16 different Modbus addresses. Also,
an additional 238 Modbus addresses (for a total of 254 different Modbus Addresses) are available via
menu selection. Figure 3-6 in this manual provides the wiring terminations for Modbus connections. The
5100-XX-IT provides the additional terminal connectors to enable the user to connect In/Out terminations of
a RS-485 connection.
1.4.3 ANALOG OPERATION
The 4-20 mA interface allows direct connection to standard analog controllers or PLCs. The 5100-XX-IT 4-
20 mA connection can be wired as a Type 3 (3-wire) or Type 4 (4-wire) circuit as described by ANSI/ISA-
50.00.01-1975 Standard (see figure 3-5).
1.4.4 REMOTE SENSOR AND DISPLAY (APPENDIX D)
The Remote Sensor and Display option (5394-52) can be used to remotely mount the sensor up to 100’
from the transmitter.
1.4.5 REMOTE ALARM RESET (DIGITAL INPUT) (Figure 3-7)
An input is available for connection of remote alarm reset/acknowledge. Figure 3-4 provides the wiring
termination for connecting the remote alarm reset. This only resets local alarms, not Sentry alarms. This
input can be wired as supervised or non-supervised.
Note that when the Remote Alarm Reset is operated in a supervised mode that both the switch and resistor
must be present as outlined in Figure 3-7.
1.4.6 OPTIONAL INTEGRAL RELAYS
The optional relays are integral to the gas sensor module and are rated as SPDT, 250VAC, 8 Amps for the
High Alarm and Low Alarm relays and SPDT, 250 VAC, 2 Amp for the Trouble relay.
If the gas sensor module is provided with the optional relays, it will include Terminal P4 on the interface
board (Figure 3-2). Relay output connections are on P4.
MECHANICAL
The sensor module of comprised of the following three primary components:
1.4.7 ENCLOSURE
Standard on the 5100-XX-IT is an explosion-proof, rain-tight cast aluminum electrical housing (Figure 1-1)
with three ¾” FNPT conduit hubs. The 5100-XX-IT-SS (Figure 1-2) has a 316 Stainless Steel enclosure.
Both enclosure covers have a viewing window. The design of the enclosure allows 3-way mounting choices
as shown in figure 1-3.
Page: 2
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 1-1
Model 5100-05-IT Toxic Gas Sensor Module – Aluminum Enclosures, Dimensions
Figure 1-2
Model 5100-05-IT Toxic Gas Sensor Module – 316 Stainless Steel Enclosures, Dimensions
.
Page: 3
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 1-3
Model 5100-XX-IT Toxic Gas Sensor Module – Mounting Options
1.4.8 TRANSMITTER ELECTRONICS
Electronic Assembly consist of one top transmitter board (connected to the cover plate) and on lower
interface board. Connectors for wiring for power, signal interface and alarm relays are located on the
interface board assembly
1.4.9 SENSOR ASSEMBLY
The sensor assembly includes an explosion proof housing containing the gas sensor and a wiring harness
for connection to the transmitter. The sensor assembly threads into one hub of the enclosure. The
exposed end of the sensor assembly is threaded to allow connection of a rain-shield or calibration gas
delivery fitting.
1.5 INTERCONNECT WIRING
Not supplied with the sensor module, but necessary to the installation and operation is the multi conductor
wiring which connects the module to its power source and controller. Before this wiring is installed it is
important to read and understand the control system installation instructions to determine wiring requirements
and alternatives.
1.6 POWER REQUIREMENTS
IT modules operate on DC power between 10 VDC and 30 VDC. Regulated DC power must be supplied from
a separate source, or from an approved Sentry or IT controller.
Page: 4
Download from Www.Somanuals.com. All Manuals Search And Download.
2. CAUTIONS & WARNINGS
2.1 INTRODUCTION
Although IT Transmitter Modules are designed and constructed for installation and operation in industrial
applications including "hostile" environments, caution should be taken to insure that the installation is made in
compliance with this instruction manual and that certain procedures and conditions are avoided. This chapter
discusses the necessary cautions. Read the entire chapter prior to installation of the equipment.
2.2 IT MODULES - GENERAL
Avoid installing sensor modules where they will be unnecessarily exposed to wind, dust, water (esp. direct
hose down), shock, or vibration. Observe temperature range limitations.
Sensors may be adversely affected by prolonged exposure to certain materials. Loss of sensitivity, or
corrosion, may be gradual if such materials are present in low concentrations. These materials include:
Halides (compounds containing chlorine, fluorine, bromine, iodine), acid vapors, caustic liquids or mists.
Care has been taken by the manufacturer to ship your modules in protective packaging to avoid contamination
prior to installation. It is recommended that the modules remain protected during installation and that the
covering be removed immediately prior to system start-up.
During normal use the sensor is protected from dirt and oil contamination by a sintered metal cover. If this
cover becomes clogged, the response of the sensor will be reduced. Protect the sensor from contamination
by careful placement, or by use of rain and dust shields.
Sensor modules must not be painted. Paint may contain compounds which will contaminate the sensor. Paint
will cause clogging of the sintered metal cover and will cause difficulties during attachment of the calibration
head or other maintenance activity. It is recommended that the module be tagged "DO NOT PAINT".
2.3 WIRING
The manufacturer recommends that extra caution be taken where the installation is near any sources of
electromagnetic or radio frequency interference. Precautions include:
•
Avoid running sensor module cable close to high power cables, radio transmission lines, or cables
subject to pulses of high current.
•
•
•
Avoid running cables near large electric motors or generators.
When the sensor module is to be operated in analog (4-20mA output) mode shielded cable is required.
When the risk of interference is present use shielded cables. In conduit installations the shield should
be connected to the conduit. In cable applications the shield should be connected to the cable
connector.
•
All splices must be via either a termination hardware system or soldered. Improperly spliced cable can
result in corrosion, resistance changes and system errors.
NOTE
Installation and wiring must be in accordance with the National Electrical Code.
AC Voltage conductors are not to be run in the same conduit as DC voltage
conductors.
Page: 5
Download from Www.Somanuals.com. All Manuals Search And Download.
THIS PAGE INTENTIONALLY LEFT BLANK
Page: 6
Download from Www.Somanuals.com. All Manuals Search And Download.
3. INSTALLATION
NOTE
All IT modules are factory are pre-configured and calibrated.
All modules are tagged to indicate the configuration including the sensor module number.
Identify all components during unpacking and install using the factory configuration.
3.1 SENSOR MODULE LOCATIONS
Select locations for each sensor modules based on the following:
•
•
•
Modules should be placed close to the potential source of gas.
Modules should be placed in areas accessible for calibration.
Sensors should be pointed down and the conduit should include an inverse trap to reduce moisture
(condensation) from accumulating in the electronics enclosure.
•
Remote calibration fitting (5360-00) should be used to facilitate calibration gas delivery. Run
polyurethane tubing (1/4” O.D. x 1/8” I.D.) from fitting to an accessible location.
3.2 WIRING
3.2.1 ANALOG 4-20 mA OPERATION
The 4-20 mA output for the 5100-XX-IT can be either 3-wire or 4-wire operation. If using 3-wire operation,
use a minimum of 18 AWG, shielded, 3-conductor cable up to 2000’. For 4-wire operation, use a minimum
of 2 each, 18 AWG, twisted, shielded, pair up to 2000’.
3.2.2 MODBUS OPERATION USING RS-485 CONNECTION
Use a minimum of 18 AWG, 2 conductor for DC power connection. No shield required. In addition use a
minimum of 24 AWG, low capacitance, shielded data cable for RS-485 half duplex communication. The
installation may be planned in a manner which provides up to 32 sensor modules on a single home run.
Refer to Figure 3-3 units ship with BIAS jumpers connected. For all installations, except very long cable
runs, these jumpers connected. The TERM jumpers should be installed in the last (furthest) module of
every loop.
3.2.3 SENTRY OPERATION USING SMC SENTRY CONNECTION
Wire shall be rated as 600 volt tray cable, such as Belden (or equivalent) 27331 or 27331 AS (shielded). If
high RFI or EMI levels exist wiring should be protected by conduit or shield. The installation may be
planned in a manner which provides up to 4 sensor modules on a single home run to a Sentry controller.
NOTE:
Be sure to follow all local electric code and safety requirements when installing the 5100-XX-IT Gas Sensor Module
3.2.4 GENERAL
Install conduit as required by local code or construction specifications. When sensor modules are to be
multiplexed for Sentry or RS-485 communication two alternatives may be planned:
•
Install splice boxes above each sensor module. Use multi-position positive contact terminals to
connect daisy chain wiring and provide a pigtail to connect to the sensor module transmitter board.
Page: 7
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
The drain wire of shielded cable must NOT be used as one of the conductors.
Installation and wiring must be in accordance with the National Electrical Code. Temperature
rating of cable wire must be at least 75oC. If cable runs through higher temperature
environments, it must be specified for that environment.
3.3 ENCLOSURE INSTALLATION
To protect the transmitter and sensor assembly they should be removed from the enclosure and preserved
until final installation and wiring termination.
Prior to installation and wiring.
1. Remove the transmitter from the module housing by:
•
•
•
•
Unscrew the two captive panel screws in the face plate.
Lift the transmitter out of the housing.
Unplug the sensor cable from transmitter connector P2.
Remove the sensor assembly from the enclosure hub.
2. Install the module housing onto the end of the supply conduit and/or bolt into position as required.
NOTES
When housing earth grounding is required for the installation a grounding lug is located in the
base of the enclosure. Install the earth ground wire under the green lug.
When installing the enclosure be sure to mount the FM Approved plate near the sensor
enclosure. This plate states “Caution: this area must be known to be free of flammable
concentrations prior to opening the enclosure.”
3.4 TRANSMITTER AND SENSOR INSTALLATION
When all pre-wire is complete:
1. Install sensor assembly in the open hub on the module enclosure. The sensor assembly thread
must be fully seated into the hub and tightened to maintain explosion proof assembly.
2. Connect the wires which return to the controller to interface board connectors P1, P2, P3 and P4
according to Figure 3-2.
3. Connect the sensor assembly cable to top transmitter board connector P2.
4. Align the headers between the top transmitter board and the lower interface board and push
together.
5. Turn rotary switch to correct sensor address.
NOTE
The interface board should be wired to remote controllers first before installing the front
transmitter board.
Page: 8
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Carefully return the transmitter to the enclosure installing it over the two stand-off’s. Tighten the
retaining screws into the stand-offs.
P1
1
2
3
4
5
6
7
PCB Label
Switch
Function
IN +
IN -
IN +
IN -
Digital Input SW +
Digital Input SW -
4-20 mA Input +
4-20 mA Input -
Ground
4-20
GND
4-20
OUT +
OUT -
4-20 mA Output +
4-20 mA Output -
P2
1
2
PCB Label
Function
RS 485 (+) (A)
RS 485 (-) (B)
+
-
RS 485
3
4
5
S
+
-
RS 485 shield (Isolated GND)
RS 485 (+) (A)
RS 485 (-) (B)
RS 485
6
S
RS 485 shield (Isolated GND)
P3B
1
2
3
PCB Label
Function
Power
Signal
P
S
G
Ground
P3A
4
5
P
S
G
Power
Signal
Ground
6
P4 Connections are only available when the optional Relays are included
P4
1
2
3
4
5
6
7
8
PCB Label
Function
N/C
Low Alarm Relay NC
Low Alarm Relay COM
Low Alarm Relay NO
High Alarm Relay NC
High Alarm Relay COM
High Alarm Relay NO
Trouble Alarm Relay NC *
Trouble Alarm Relay COM*
Trouble Alarm Relay NO*
LO ALM
COM
N/O
N/C
COM
N/O
N/C
HI ALM
TRBL
ALM
COM
N/O
9
* Trouble relay is fail safe so it is energized for normal operation,
functions are labeled for normal operation.
Table 3-1
Sensor Module External Interfaces (See Figure 3-2)
7. Establish the module address according to section 3.5.
NOTE
The starting delay period normally takes approximately 3 minutes but under some
circumstances can take longer.
Page: 9
Download from Www.Somanuals.com. All Manuals Search And Download.
3.5 MODULE ADDRESS SWITCH
For digital interface applications the module address switch (or Modbus node) must be set per Table 3-2:
POSITION
ADDRESS
POSITION
ADDRESS
1
2
3
4
5
6
7
8
Sensor 1
Sensor 2
Sensor 3
Sensor 4
Sensor 5
Sensor 6
Sensor 7
Sensor 8
9
A
B
C
D
E
F
0
Sensor 09
Sensor 10
Sensor 11
Sensor 12
Sensor 13
Sensor 14
Sensor 15
Software
Menu
Table 3-2
Sensor Module Address Switch Positions
NOTE
For Sentry applications only sensor addresses 1-8 are allowed. If using Modbus output sensor
addresses 1-15 are available. Position 0 allows the Modbus Address to be set by software
menu, in the range 16-254.
Figure 3-1
Transmitter Face Plate
Page: 10
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 3-2
Interface Board Connectors
Figure 3-3
RS-485: 120 ohm termination not selected, BIAS shunted for 100K ohm bias
Page: 11
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 3-4
4-20mA Circuit Types
Page: 12
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 3-5
4-20mA Circuit Type Connections for 5100-XX-IT
Page: 13
Download from Www.Somanuals.com. All Manuals Search And Download.
For end of line,
refer to Figure 3-3
Figure 3-6
Wiring Connections for Modbus and Sentry Interface
Page: 14
Download from Www.Somanuals.com. All Manuals Search And Download.
P1
+
+
+
4.7K OHM (CUSTOMER SUPPLIED)
+
-
SWITCH IN
SWITCH IN
5100-05-IT REMOTE ALARM RESET (SUPERVISED)
Figure 3-7
Wiring Connections for Remote Alarm Reset
Page: 15
Download from Www.Somanuals.com. All Manuals Search And Download.
THIS PAGE INTENTIONALLY LEFT BLANK
Page: 16
Download from Www.Somanuals.com. All Manuals Search And Download.
4. OPERATION
4.1 INTRODUCTION – HUMAN-MACHINE INTERFACE SYSTEM
The Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A magnetic tool
(5358-50) is supplied for this purpose. The menu system is used to configure alarm set-points, calibrate the
sensor module, and for maintenance procedures and alarms acknowledge.
4.2 HUMAN-MACHINE INTERFACE
The module menu system is operated by means of directing the magnet stick toward each of four independent
hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are
located above and below the faceplate display and are labeled M , E , ▲ and ▼as shown in Figure 4-1. The
key functions are as follows:
•
•
•
•
Key M : MODE
Key E : ENTER
Key ▲: UP (+)
Key ▼: DOWN (-)
Key M
Key /\
Key E
Key \/
Figure 4-1:
IT Module – Menu Switch Locations
Page: 17
Download from Www.Somanuals.com. All Manuals Search And Download.
4.3 MAIN MENU
Table 4-1 describes the primary man-machine interface operation.
Key Function Display
Description
Reference
M
E
Mode
Enter
Up
Switch [M]
S T
M
E
Switch [E]
S T
M
E
Switch [▲] Previous Menu
Switch [▼] Next Menu
First screen at power up: Model No.
S T
M
E
Down
S T
5100-XX
VXX-XX-- Second screen at power up: Version No.
STARTING Third screen at start up: Starting Delay
XXXPPM
Normal condition - default display
M
E
Mode ALMRSET: Mode Function - Alarm Reset
S T
Banner: "Press [E] to reset alarm"
M
E
RESET
Alarm Reset
S T
XXXPPM
Default Display
M
E
Mode ALMRSET: Mode Function - Alarm Reset
S T
M
E
Mode
Mode
Mode
Mode
Enter
CALIB:--
SETUP:--
MAINT:--
EXIT-?--
XXXPPM
XXXPPM
Mode Function - Calibrate
Mode Function - Set Point Adjustments
Mode Function - Maintenance
Exit Menu
Table 5-1
Table 4-3
Table 4-4
S T
M
S T
M
S T
M
S T
M
S T
E
E
E
E
Apply Selected Mode (Exit)
Default Display (Once a minute the sensor displays module address)
Table 4-1 Master Menu
Table 4-2 describes the operational display values of the human-machine interface system.
DISPLAY
STARTING
XXX PPM
LXXXPPM
HXXXPPM
DESCRIPTION
Delay from loss of power at start-up
Concentration
Low Alarm
High Alarm
Measures gas, concentration exceeds
100% to Full Scale
CXXXPPM
Calibration Mode
Acknowledged Function
Table 4-2 Operation Display Values
Page: 18
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
4.4 CONFIGURE SET-POINTS
The sensor module set-points menu is used to initially set-up the alarm set points, relay actions, gas type and
range, 4-20 mA action and RS-485/Sentry address and baud rates.
•
Alarm Set-points: Once the Set-up menu is selected, press [E] to activate the Alarm Set-point
screen. Use the [▲] or [▼] keys to select Low Alarm or High Alarm menu. Key [▲] will adjust the
setpoint upwards and Key [▼] will adjust the value downwards. Once it reaches the desired
setpoint, Key [E] will accept it and ACK will appear.
Set-points can be configured using this menu up to the following values:
5100-03-IT
5100-04-IT
5100-05-IT
5100-06-IT
O2
19.5%
C0
500 PPM
H2S 20 PPM
Cl2 5 PPM
•
•
Alarm Relay Set-up: Once the Set-up menu is selected, press key [▼] once and press [E] to
activate the Relay Set-up menu. Use the [▲] or [▼] keys to select High Alarm or Low Alarm relay
menu and press [E]. Use the [▲] or [▼] keys to select the correct alarm relay action for the
application, Latch, Sentry or Non-Latch. Selecting “Sentry” enables the Sentry controller to make all
alarm action decisions. * indicates the current selection.
Range: Once the Set-up menu is selected, press key [▼] twice and press [E] to activate the Range
Set-up menu. Use the [▲] or [▼] keys to select Range menu and press [E]. When “Range” is
selected menu provides any choices of ranges available for the gas type selected. Use the [▲] or
[▼] keys to select the desired range. If the “User” range is selected, use the [▲] or [▼] keys to
adjust the high end of the range desired.
•
4-20mA: Once the Set-up menu is selected, press key [▼] three times and press [E] to activate the
4-20mA Set-up menu. Use the [▲] or [▼] keys to select Calib or CalibOut menu and press [E].
The “Calib” section of the menu allows the user to calibrate the 4 mA and 20 mA outputs. To
calibrate the 4 mA and 20 mA outputs it is necessary to have an amp meter connected to the 5100-
XX-IT and upon selecting the 4 mA output calibration then the [▲] or [▼] keys can be used to adjust
the 4 mA reading on the amp meter until it reads 4 mA. Similar steps can then be performed for the
20 mA output. The CalibOut section allows the user to select the 4-20 mA output action desired
during calibration. * indicates the current selected value. Available selections include:
Track – the 4-20mA value tracks the calibration gas exposed to the gas sensor module
Zero – the 4-20mA value is held at 0mA during calibration
C1.50mA – the 4-20mA value is held at 1.50mA during calibration
C4.00mA – the 4-20mA value is held at 4.0mA during calibration.
•
RS-485 - Once the Set-up menu is selected, press key [▼] four times and press [E] to activate the
RS-485/Sentry Set-up menu. Use the [▲] or [▼] keys to select Address or Baud rate menu and
press [E]. Note that the 5100-XX-IT has a rotary switch on the faceplate and it is used to select
addresses 1-15. When connected to Sentry the user can select 1-8 and using Modbus RS-485 the
user can select addresses 1-15. For Modbus addresses above 15, set the rotary switch to 0 and
then use the “Address” menu to select any address between 16 and 254. The Baud rate menu
allows the user to select a baud rate of 38400, 19200, 9600, 4800 or 2400. * indicates current
selection.
Page: 19
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
Key Function Display
Description
Reference
--0%LEL-
Default Display
M
E
Mode ALMRSET: Mode Function - Alarm Reset
S T
M
E
Mode
Mode
Enter
Down
Down
Down
Down
CALIB:--
Mode Function - Calibrate
S T
M
S T
M
S T
M
S T
M
S T
M
S T
M
S T
E
SETUP:--- Mode Function - Set Point Adjustments
E
Alarms
Relays
Gas
S.P. Function - Alarm Adjust
* A Below
* B Below
* Page 21
* Page 21
* Page 21
E
S.P. Function - Relays Adjust
E
S.P. Function - Range Adjust
E
4-20mA
RS-485
S.P. Function - 4-20 mA Adjust
S.P. Function - RS-485/Sentry Output Adjust
E
High Alarm Set Point Adjustment Example
M
E
Enter
Enter
H.Alarm
S.P. Function - High Alarm Adjust
*A
S T
M
S T
E
HASP:60- High Alarm Set Point: current = 60
Use S or T keys to adjust to new set point
M
E
Down (x5) HASP:55- High Alarm Set Point: new = 55
S T
M
E
Enter
ACK
Momentary Acknowledge of new Set Point
S.P. Function - High Alarm Adjust
S T
H.Alarm
Relays Set Point Adjustment Example
S.P. Function - High Alarm Relay Adjust
S.P. Function - Low Alarm Relay Adjust
M
E
Enter
Down
H. Relay
L.Relay
*B
S T
M
S T
E
M
E
Use S or T keys to adjust to new relay action
(Latch, Sentry, NonLatc) * indicates current
NOTE: Sentry indicates that Sentry controls relay
action and not the IT Sensor Module
Enter
Latch
S T
M
S T
M
S T
E
Down
Down
Sentry
E
*Sentry
High Alarm Relay set to Sentry
Table 4-3A Set-Up Configuration
Page: 20
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
Gas Range Adjustment Example
M
E
Enter
Enter
Range
S.P. Function - Range Adjust
*C
S T
M
S T
E
Select [E] to select or S or T to select another and
press [E]
*100 PPM
M
E
Down
Down
10 PPM
USER
Press [E] if selecting 0-10 PPM range
S T
M
S T
E
Press [E] if selecting User adjustable range
M
S T
E
Select [E] to select or S or T to select another and
press [E]
Enter
100 PPM
4-20 mA Adjustment Example
M
E
Enter
Enter
Enter
Enter
Mode
Down
Enter
Calib
S.P. Function - Calib Adjust
*D
S T
M
S T
M
S T
M
S T
M
S T
M
S T
M
S T
E
Out: 4mA Use S or T keys to select 4 mA or 20mA
E
4mA
ACK
Selects 4 mA
E
Momentary Acknowledge of new Set Point
S.P. Function - Calib Adjust
E
Calib
E
CalibOut
Track
S.P. Function - Output during Calibration Adjust
E
Use S or T keys to select Track, Zero, C1.50mA, C4.00mA
Track = Output during calibration tracks the calibration gas,
Zero = Output is Zero during calibration,
C1.50mA = Output during calibration is 1.50 mA, C4.00mA =
Output during calibration is 4.0 mA
M
E
Enter
*Track
* = Current selection
S T
RS-485 Adjustment Example
M
E
Enter
Enter
Address
S.P. Function - RS-485 Address Adjust
*E
S T
M
E
Addr:016
Use S or T keys to enter new address
S T
M
E
Enter
Enter
Down
ACK
Address
Baud
New address selected
S T
M
S T
M
S T
E
S.P. Function - RS-485 Address Adjust
S.P. Function - RS-485 Baud Rate Adjust
E
M
E
Enter
*38400
Press [E] to select or [S] or [T] to select another
S T
Table 4-3B Set-Up Configuration
Page: 21
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
4.5 MAINTENANCE FUNCTIONS
The maintenance menu enables the operator to view sensor and software versions. Sierra Monitor technical
support has access to other values as needed.. The maintenance menu operation is described in Table 4-4.
Key Function Display
Description
Reference
XXXPPM
Default Display
M
E
Mode ALMRSET: Mode Function - Alarm Reset
S T
M
E
Mode
Mode
Mode
Enter
Enter
CALIB:--
SETUP:-
MAINT:--
Mode Function - Calibrate
S T
M
S T
M
S T
M
S T
M
S T
E
Mode Function - Set Point Adjustments
Mode Function - Maintenance
E
E
Ver1.00aA Module Software Version Number
CCC001 Module Custom Configuration Control number
E
Table 4-4
Maintenance Menu
Page: 22
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
THIS PAGE INTENTIONALLY LEFT BLANK
Page: 23
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
5. CALIBRATION
5.1 CALIBRATION FREQUENCY
The 5100-XX-IT has been calibrated in the factory prior to shipment. It is recommended that the user check
calibration before placing in service. The toxic sensor module must be calibrated every six months. Periodic
functional tests are advisable for critical applications and hostile environments.
The sensor module microprocessor software includes high level self checking algorithms which provide
continuous sensor diagnostic and self adjustment. Users may elect to increase calibration periods based on
low drift experience during the first two calibration periods.
5.2 CALIBRATION PREPARATION
Calibration of the Toxic Gas Sensor Module is accomplished by simple menu based steps and application of
span gas.
NOTE
If an error is made during any stage of the calibration process, hold the magnet stick at the
Switch [M] for 10 seconds. A scrolling display will indicate “Calibration aborted” and the
sensor module will exit the calibration activity and return to normal operating mode. The
calibration procedure can then be restarted
Calibration must be performed only when the area is known to be clear of the applicable toxic gas. If
necessary, use a portable instrument to confirm that there is no background toxic gas.
For compliance with Factory Mutual (FM) Apparatus, the Sierra Monitor Model 1250-01, 1256-01, 1260-XX are
the only FM Approved calibration gas delivery device. Use Model 5358-00 or 5358-01 Calibration Adapter or
Model 5360-00 Gas Delivery Fitting.
5.2.1 CALIBRATION GAS DELIVERY METHODS
Calibration gas is can be delivered to the sensors via the following delivery devices:
Model 5358-00: Calibration Adapter - used with portable calibrators. See 5-1.
Model 5360-00: Calibration Gas Delivery fitting - permanently installed fitting which
allows tubing to be run to a convenient delivery location
5.3 CALIBRATION PROCEDURE
Table 5-1 shows the step by step process of the calibration procedure.
The procedures requires that the menu “keys” be activated using the magnet stick.
Each key press steps through the process of setting the zero value for clean air and
then setting the span value.
Sample Gas In
At each of these steps, apply calibration gas of the value corresponding to the
setting accepted on the sensor module display.
Figure 5-1
Model 5358-01 Calibration
Adapter
Page: 24
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
5.3.1 SENSOR EXPOSURE TO GAS
Calibration gas must be delivered to the sensor using the flow rate and duration listed in below:
Model
Gas
Flow
Period
5100-03-IT Oxygen
300 cc/min
300 cc/min
300 cc/min
300 cc/min
Until Stable
Until Stable
Until Stable
Until Stable
5100-04-IT Carbon Monoxide
5100-05-IT Hydrogen Sulfide
5100-06-IT Chlorine
Key Function Display
Description
Reference
XXXPPM
Default Display
M
E
Mode ALMRSET: Mode Function - Alarm Reset
S T
M
E
Mode
CALIB:--
Mode Function - Calibrate
S T
M
E
Banner: Apply zero gas, enter <E> when done
Enter CAL-0PPM--
S T
Operation: Confirm area clear of gas, or apply zero air to
sensor.
M
E
Enter
Enter
ACK
Zero gas setting acknowledged
S T
Banner: Select span, enter <E> to calibrate sensor
Banner: Apply span gas, then enter <E> to calibrate gas sensor
Operation: Apply calbration gas.
10PPM-SP
C 4 PPM
C 4 PPM
CXXXPPM
CAL-OK
Sub A
Sub B
M
S T
E
Operation: As gas is applied the reading will increase - wait 3
minutes
M
S T
E
Enter
Calibration Passed - now remove gas
Operation: Five minute time out before sensor is returned to
service.
WAIT-300
Sub-Routines
Banner: Select span, enter <E> when done
Operation: Change Span Gas Value to 25PPM
Operation: User adjustable value
10PPM-SP
25PPM-SP
Adj-SPAN
25PPM-SP
Sub A
Sub B
Sub C
M
E
Up
Up
S T
M
S T
M
S T
M
S T
E
E
Operation: Ready for user adjustment
Enter
Enter
E
Banner: Apply 25PPM gas, then enter <E> to calibrate gas
sensor
C 25PPM
C 25PPM
CAL-FAIL
WAIT-300
(Any)
Operation: Apply calbration gas.
Operation: No calibration gas applied, or sensor did not respond
correctly.
Operation: Five minute time out before sensor is returned to
service.
Operation: Hold magnet over Mode Switch for ten senconds to
abort calibration
M
S T
E
Enter
Mode
M
S T
E
Banner: Calibration Aborted
XXXPPM
Default Display
Table 5-1
Calibration
Page: 25
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
6. SERVICE
6.1 SENSOR MODULE CONFIGURATION
The gas sensor module is comprised of the following sub-assemblies (Figure 6-1):
5100-XX-IT
SPL21810
SPL21825
XXXXXXX
5200-XX-IT
Gas Sensor Module
Aluminum Enclosure (SPL32178 316SS Enclosure)
Transmitter Assembly
Sensor Assembly (See Appendix B)
Sensor (See Appendix B)
6.2 ENCLOSURE REPLACEMENT
Figure 6-1
Module Components
Page: 26
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
The enclosure should be replaced if the cover threads or conduit threads have been damaged, or if the
enclosure has been damaged sufficiently that it no longer meets the required NEMA classification.
To replace the enclosure follow the transmitter and sensor assembly removal instructions, remove the
damaged enclosure from its conduit, install a new enclosure and replace the transmitter and sensor assembly.
6.3 TRANSMITTER REPLACEMENT
The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be
adjusted for calibration. This may occur due to age, corrosion or failed components.
To replace the transmitter assembly:
a. Remove the cover of the main enclosure
b. Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90o to relieve the
wiring service loop
c. Unplug the sensor connector from the transmitter
d. Plug connector into new transmitter (be sure to match numbers between connector and socket).
e. Restore power and allow a minimum of 3 hours for stabilization before re-calibration
6.4 SENSOR REPLACEMENT
The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of
the housing. The gas sensor needs replacement when:
The “CAL-FAIL” message appears after calibration
The sensor output signal is noisy, causing erroneous gas level readings
To replace the sensor:
a. Confirm that system power has been removed
b. Remove the gas sensor module enclosure cover,
c. Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90o to
relieve the wiring service loop.
d. Unplug the sensor connector from the transmitter
e. Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its
harness
f. Unscrew sensor housing cover from the sensor (be sure to loosen the set screw first)
g. Carefully pull the old sensor straight up from the socket
h. Press the new sensor into the socket.
i. Reverse the preceding steps to install the sensor assembly.
j. Allow the new sensor to stabilize for a minimum of 3 hours and then calibrate using the procedure in
Section 6.
6.5 INSTALLATION INSPECTION
Prior to system start-up or trouble shooting the entire system should be visually inspected. The following are
guidelines for that inspection:
6.5.1.1 CONTROLLER INSTALLATION
•
•
•
Controller installed in conformance to instruction manual recommendations.
AC power is correctly grounded.
Hot AC and relay connections have safety covers installed.
Page: 27
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
6.5.1.2 SENSOR MODULE INSTALLATION
•
•
•
Module installation in conformance with this manual.
Modules accessible for calibration.
Wiring terminations clean and correct.
6.5.1.3 MOISTURE TRAPS AND RAINSHIELDS
•
Conduit seals and drains installed to avoid moisture build up in electronics enclosure. Water
accumulation in sensor module enclosures is a major cause of damage and system failures - take
precautions to seal electrical conduits and provide moisture traps and drains to avoid water damage
•
Rain-shields installed where applicable.
6.5.1.4 STANDARD VOLTAGES
•
DC Voltage to be applied to the sensor module must be between 10 VDC and 30 VDC.
6.5.2 INSPECTION AND TROUBLESHOOTING GUIDE
The inspection and troubleshooting guide can be used to step through the system start-up and to determine
the corrective action if a fault occurs.
6.5.3 IF MODULE DOES NOT RESPOND TO GAS
1. Repeat calibration procedure.
2. Remove the gas and wait for the timer to completely count down.
3. Apply 50 PPM and verify that the sensor sees 50 PPM gas after calibration.
4. If the sensor still does not see gas, power cycle the unit and repeat calibration.
6.5.4 IF THE MODULE DISPLAYS “STARTING” FOR MORE THAN 1 HOUR
1. Make sure the sensor is placed in an ambient room temperature environment.
2. Power cycle the sensor.
3. Ensure that the sensor is not exposed to the gas of interest during warm-up.
6.5.4.1 IF THE MODULE DOES NOT DISPLAY THE CORRECT PPM
1. Power cycle the unit
2. Recalibrate the sensor.
6.5.4.2 IF THE DISPLAY SHOWS ‘F’
1. Power down the unit
2. Open the enclosure and unplug the sensor assembly from the transmitter board.
3. Plug the sensor back into the transmitter board carefully and ensure a secure fit.
4. Power up the unit.
6.5.4.3 IF THE DISPLAY SHOWS ‘C’
1. Complete calibration and exit to operating mode.
6.5.4.4 IF THE DISPLAY SHOWS ‘S’
1. Check connections with Sentry Connections.
Page: 28
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
7. APPENDICES
APPENDIX A - SPECIFICATIONS
Sensor type:
Electrochemical
5100-04-IT CO
0-1000 PPM
0-100, 0-500, variable
+/- 1 PPM
+/- 1 PPM
+/- 1 PPM
0.5 PPM
<25 sec.
5100-03-IT O2
5100-05-IT H2S
0-100 PPM
0-10, variable
+/- 0.5 PPM
+/- 1 PPM
+/- 1 PPM
0.1 PPM
5100-06-IT Cl2
0-10 PPM
0-5, variable
+/- 0.5 PPM
+/- 0.5 PPM
+/- 0.5 PPM
0.1 PPM
Max. Range
User Settings Range
Zero Drift
Repeatability
Linearity
Resolution
Response Time (1)
Accuracy
0-25 %Vol
+/- 0.1%
+/- 0.2%
+/- 0.2%
+/- 0.1%
<10 sec
+/- 0.2%
2 years
<30 sec.
<60 sec
+/- 1 PPM
2 years
+/- 0.5 PPM
2 years
+/- 0.5 PPM
2 years
Sensor Life
(1) Response time to 90% full signal value for applied concentration
(2) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest
Output:
Display:
Fixed and Scrolling LED
Relays (Optional):
Analog Output (Optional)
Signal Output:
8 Amp, High Alarm, Low Alarm, 2 Amp Trouble all SPDT
Analog 4-20 mA (Trouble 0 mA, Calibration 1.5 mA)
Sentry bus
Serial RS-485 Modbus RTU
Input:
Remote Alarm Reset:
Normally open digital input (supervised or non-supervised)
Electrical Data:
Power consumption:
Connection type:
Input voltage:
2 watts
3 wire
24 VDC nominal: 10-30VDC
Operating Range:
5100-03-IT O2
5100-04-IT CO
-4 to 122 oF
-20 to 50 oC
5 – 99%
5100-05-IT H2S
-40 to 122 oF
-40 to 50 oC
5 – 99%
5100-06-IT Cl2
-4 to 122 oF
-20 to 50 oC
5 – 99%
Ambient Temp (oF)
Ambient Temp (oC)
Relative Humidity
5 to 122 oF
-15 to 50 oC
5 – 99%
Enclosure:
Material - Standard:
Material - Stainless Steel Option
Dimensions:
Polyester powder-coated, sand-cast, copper-free aluminum
316 Stainless Steel (FM approval pending)
7.5 x 5.0 x 4.8 inches Al, 7.5 x 5.0 x 4.4 inches SS
(21.6 x 12.7 x 12.2 cm)
(H x W x D)
Weight:
2.7 lb. (1.3 Kg) (standard Aluminum enclosure)
5.4 lb. (2.6 Kg) 316 Stainless Steel
NEMA 4X
Housing:
Electrical Classification
Performance:
Explosion proof, Class 1, Div. I, Groups B, C, D
Class I, Zone 1, groups IIB+H2 IP66
Approvals:
FM Approved
Performance and Safety Approval Pending
Page: 29
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
Modbus:
Baud:
38400 (Adjustable 2400 to 38400 Baud)
Parity:
None
Stop bit:
Data bits:
1
8
Flow Control:
Memory map:
None
Appendix E
Warranty:
Limited Warranty 2 years
Specifications subject to change without notice
Ordering Information:
5100
|- 5100 Series |
|- 03 = O2
-
XX
-
IT
|
|
-
YY
|
|
-
ZZ
|
|
|
|- 04 = CO |- IT Series |
|- 05 = H2S
|- 06 = Cl2
|- Al (Aluminum Enclosure) |- 01= Sentry, RS-485, 4-20mA, & Digital Input
|- SS (Stainless Steel) |- 02= Sentry, RS-485, 4-20mA, Digital Input & Relays
Page: 30
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
APPENDIX B: - MODEL NUMBERS & PARTS LIST
Sensor Module
5100-03-IT
5100-03-IT-SS
5100-04-IT
Gas Sensor Module, Oxygen
Gas Sensor Module, Oxygen, Stainless Steel Enclosure
Gas Sensor Module, Carbon Monoxide
5100-04-IT-SS
5100-05-IT
Gas Sensor Module, Carbon Monoxide, Stainless Steel Enclosure
Gas Sensor Module, Hydrogen Sulfide
5100-05-IT-SS
5100-06-IT
Gas Sensor Module, Hydrogen Sulfide, Stainless Steel Enclosure
Gas Sensor Module, Chlorine
5100-06-IT-SS
Gas Sensor Module, Chlorine, Stainless Steel Enclosure
Options
5394-52
5311-00
Remote Sensor/Display Option
Rainshield
Calibration Items
1250-01
Gas Sensor Calibration Kit, Type A (O2, CO, H2S)
Gas Sensor Calibration Kit, Type C (Cl2)
Gas Cylinder, Air, (Type A) , 105 litres
Gas Cylinder, CO 100 PPM, (Type A) 57 litres
Gas Cylinder, H2S in N2, 25 PPM, (Type A) 57 litres
Gas Cylinder, Cl2, 5 PPM, (Type C) 104 litres
Gas Cylinder, CO, 1000 PPM, (Type A), 105 liters
Calibrator Head Standard
Regulator Type A Calibrator
Regulator Type C Calibrator
Calibration Adapter - Direct, Standard
Calibration/Configuration Magnetic Tool
1250-03
1260-00
1260-04
1260-05
1260-06
1260-13
5360-00
1256-01
1265-03
5358-01
5358-50
Spare Parts
5200-03-IT
5200-04-IT
5200-05-IT
5200-06-IT
SPL21834
SPL21832
SPL21830
SPL21833
SPM27082
SPM27083
SPM27079
SPM27084
SPL21824
SPL21810
SPL21825
SPL21825
SPL27826
Sensor, for 5100-03-IT
Sensor, for 5100-04-IT
Sensor, for 5100-05-IT
Sensor, for 5100-06-IT
Sensor Assembly, Aluminum, for 5100-03-IT
Sensor Assembly, Aluminum, for 5100-04-IT
Sensor Assembly, Aluminum, for 5100-05-IT
Sensor Assembly, Aluminum, for 5100-06-IT
Sensor Assembly, 316SS, for 5100-03-IT
Sensor Assembly, 316SS, for 5100-04-IT
Sensor Assembly, 316SS, for 5100-05-IT
Sensor Assembly, 316SS, for 5100-06-IT
Transmitter
Enclosure, Transmitter, Aluminum
Enclosure Transmitter, 316SS
Interface Board with Relay
Interface Board without Relay
Page: 31
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
APPENDIX C: LIMITED WARRANTY
SIERRA MONITOR CORPORATION warrants its products to be free from defects in workmanship or material
under normal use and service for two years after date of shipment. SMC will repair or replace without charge
any equipment found to be defective during the warranty period. Final determination of the nature and
responsibility for defective or damaged equipment will be made by SMC personnel.
All warranties hereunder are contingent upon proper use in the application for which the product was intended
and do not cover products which have been modified or repaired without SMC approval or which have been
subjected to accident, improper maintenance, installation or application, or on which original identification
marks have been removed or altered. This Limited Warranty also will not apply to interconnecting cables or
wires, consumables (ie. calibration gases, batteries, sensors), nor to any damage resulting from battery
leakage.
In all cases SMC’s responsibility and liability under this warranty shall be limited to the cost of the equipment.
The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty
provision and compliance with such instruction shall be a condition of this warranty.
Except for the express warranty stated above, SMC disclaims all warranties with regard to the products sold
hereunder including all implied warranties of merchantability and fitness and the express warranties stated
herein are in lieu of all obligations or liabilities on the part of SMC for damages including, but not limited to,
consequential damages arising out of/or in connection with the use or performance of the product.
Page: 32
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
APPENDIX D: REMOTE SENSOR/DISPLAY DRAWING 5394-52
Page: 33
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 5100-XX-IT Toxic Gas Sensor Module
APPENDIX E: MODBUS MEMORY MAP
Read Register Table
Register
40001
40002
40003
40004
40005
40006
40007
40008
40009
40010
40011
40012
40013
40014
40015
40016
40017
40018
40019
40020
40021
40022
40023
40024
40025
Description
Concentration
Temperature
High Alarm Relay
Low Alarm Relay
Low Alarm Value
High Alarm Value
Reserved
Reserved
Reserved
Reserved
Trouble Bits
Trouble
High Alarm Immediate Like 40003. but never latched
Low Alarm Immediate
Random ID
Restart Count
Run time – high
Run time – low
Max. gas value
Model
Range
Units
Temperature Units
Gas Scale
Temperature Scale
Comments
Gas Concentration multiplied by 100
Temperature in degrees Celsius scalled by a factor of 100
Boolean indicating the High Alarm Status, 0 No Alarm, 1 High Alarm
Boolean indicating the Low Alarm Status, 0 No Alarm, 1 Low Alarm
Low alarm set point, used for activating Alarms
High alarm set point, used for activating High Alarms
0 for no trouble
1 for any trouble, 0 for no trouble
Like 40004, but never latched
Randomly generated permanent ID of sensor
In minutes
In minutes
*scale ((see 40024)
51005, 51003, 51004, 51006, etc
*scale (see 40024)
3 is PPM
0 is C, 1 is F
Value like 1 or 100
Value like 1 or 100
Page: 34
Download from Www.Somanuals.com. All Manuals Search And Download.
|