Philips Work Light 200 User Manual

R E F 1 0 5 4 7 2 8  
1 0 7 2 9 1 9  
J H 6 / 7 / 1 0  
CLINICAL MANUAL  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FOR CLINICIAN’S USE ONLY  
Accessing Prescription Setting Screens  
WARNING  
The information on this page is ONLY for health care professionals. Remove this page from the manual  
before giving the manual to the patient!  
Full Menu Access Mode  
The ventilator has two levels of menu access, Full and Limited. Full  
Menu Access allows you to alter all available settings. Limited Menu  
access permits the user to alter only those prescription parameters  
Note: The Full Menu Access key  
sequence can be performed  
either from the Power Off screen  
that affect patient comfort, such as Rise Time, Flex, and Ramp Start  
Pressure, if those parameters are available as part of the prescription.  
The ventilator defaults to Full Menu Access mode.  
or from the Monitor screen.  
Note: When the airflow is off  
and AC power is removed from  
the device for more than 5  
minutes, the device will enter  
a low power mode to save  
battery life. When the device is  
in low power mode, the Setup  
key sequence will be ignored.  
Press the Start/Stop button, or  
connect AC power, or insert an  
SD card to exit the low power  
mode.  
After accessing the Main Menu, if the device is in Limited Menu  
Access mode, you can use the following Setup key sequence to enter  
Full Menu Access mode and be able to change prescription settings:  
Press the Down button and the Alarm Indicator/Audio  
Pause button simultaneously for several seconds. This will  
temporarily place the device in Full Menu Access mode.  
When you perform this key sequence from the Monitor screen,  
the Main Menu screen appears and an audible indicator sounds  
indicating you are now in Full Menu Access mode.  
When you perform this key sequence when the airflow is off, the  
Setup screen appears and an audible indicator sounds.  
Note: Philips Respironics  
recommends that after you are  
finished changing prescription  
settings, and you give the  
device to the patient, you set  
the device back to Limited  
Menu Access mode so  
You can go into the Options menu and permanently change the  
Menu Access setting to Full Menu Access. Otherwise, the device  
will return to the Menu Access mode stored in the setting once you  
exit the menu screens or if one minute passes without pressing any  
device buttons. If you are in the Setup mode, and an SD card is in the  
device, “Write Event Log to SD Cardwill appear in the menu.  
patients cannot change their  
prescription settings.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
R E F 1 0 5 4 7 2 8  
1 0 7 2 9 1 9  
J H 6 / 7 / 1 0  
Download from Www.Somanuals.com. All Manuals Search And Download.  
© 2010 Koninklijke Philips Electronics N.V. All rights reserved.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
Table of Contents  
FOR CLINICIAN’S USE ONLY Accessing Prescription Setting Screens................................................... i  
Chapter 1. Introduction....................................................................................................................................... 1  
Package Contents.................................................................................................................1  
Intended Use..........................................................................................................................2  
Warnings and Cautions ......................................................................................................3  
Warnings.........................................................................................................................3  
Cautions ..........................................................................................................................8  
Notes ..............................................................................................................................10  
Contraindications...............................................................................................................11  
System Overview................................................................................................................11  
Symbols..................................................................................................................................12  
Front Panel ...................................................................................................................12  
Rear and Side Panels.................................................................................................12  
How to Contact Philips Respironics.............................................................................13  
Chapter 2. System Description........................................................................................................................15  
Front Panel Features..........................................................................................................15  
Buttons ..........................................................................................................................15  
Visual Indicators .........................................................................................................16  
Display Screen.............................................................................................................16  
Side and Rear Panel Features .........................................................................................17  
Table of Contents  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ii  
Chapter 3. Modes, Features, and Alarms......................................................................................................19  
Therapy Modes....................................................................................................................19  
Breath Types.................................................................................................................20  
Therapy Mode Table .................................................................................................21  
Pressure Control Ventilation Therapy Modes...................................................22  
Volume Control Ventilation Therapy Modes....................................................27  
Therapy Mode Features...................................................................................................30  
Flex Comfort Feature................................................................................................30  
Ramp ..............................................................................................................................31  
Rise Time .......................................................................................................................32  
AVAPS Feature.............................................................................................................32  
Flow Pattern Types ....................................................................................................33  
Sigh Feature.................................................................................................................35  
Dual Prescription Feature .......................................................................................35  
Triggering ....................................................................................................................36  
BTPS Compensation..................................................................................................41  
Ventilator Alarms ................................................................................................................41  
Loss of Power Alarm..................................................................................................41  
Ventilator Inoperative Alarm .................................................................................41  
Ventilator Service Required Alarm.......................................................................41  
Check Circuit Alarm ..................................................................................................42  
Low Circuit Leak Alarm ...........................................................................................42  
High Expiratory Pressure Alarm ...........................................................................42  
Low Expiratory Pressure Alarm ............................................................................42  
High Internal Oxygen Alarm .................................................................................42  
Circuit Disconnect Alarm ........................................................................................43  
Apnea Alarm ...............................................................................................................43  
High Vte Alarm............................................................................................................43  
Low Vte Alarm ............................................................................................................43  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
iii  
High Vti Alarm.............................................................................................................44  
Low Vti Alarm .............................................................................................................44  
High Respiratory Rate Alarm .................................................................................44  
Low Respiratory Rate Alarm ..................................................................................44  
High Inspiratory Pressure Alarm ..........................................................................44  
Low Inspiratory Pressure Alarm ...........................................................................45  
High Minute Ventilation Alarm ............................................................................45  
Low Minute Ventilation Alarm ..............................................................................45  
Low Battery Alarm.....................................................................................................46  
High Temperature Alarm .......................................................................................46  
Replace Detachable Battery Alarm......................................................................46  
Ventilator Service Recommended Alarm..........................................................46  
AC Power Disconnected Alarm.............................................................................47  
Keypad Stuck Alarm .................................................................................................47  
Battery Discharging Stopped due to Temperature Info Message............47  
Battery Not Charging due to Temperature Info Message ..........................47  
Battery Not Charging Info Message....................................................................47  
Check External Battery Info Message .................................................................48  
Battery Depleted Info Message ...........................................................................48  
External Battery Disconnected Info Message..................................................48  
Detachable Battery Disconnected Info Message...........................................48  
Start On Battery Info Message ..............................................................................48  
Card Error Info Message .........................................................................................48  
Chapter 4. Ventilator Setup...............................................................................................................................49  
Position the Device............................................................................................................50  
Install the Air Filter .............................................................................................................50  
Supply Power to the Device............................................................................................50  
Using AC Power ..........................................................................................................51  
Using DC Power..........................................................................................................52  
Device Power Source Indicators...........................................................................54  
Table of Contents  
Download from Www.Somanuals.com. All Manuals Search And Download.  
iv  
Battery Disposal..........................................................................................................56  
First Time Use ..............................................................................................................56  
Connect the Breathing Circuit to the Ventilator......................................................57  
Connect a Water Trap .......................................................................................................60  
Connect Supplemental Oxygen (Optional) ..............................................................61  
Connect the Remote Alarm (Optional).......................................................................62  
Chapter 5. Viewing and Changing Settings................................................................................................63  
Keypad Lock Feature.........................................................................................................63  
Accessing the Startup and Monitor Screens ............................................................64  
Monitor Screen Indicators ......................................................................................66  
On-Screen Button Panel..........................................................................................71  
Navigating the Menu Screens........................................................................................71  
Changing and Viewing Settings in Full Menu Access Mode...............................72  
Changing the Device Settings and Alarms.......................................................73  
Device Settings Common to All Therapy Modes...................................73  
Additional Settings Specific to Therapy Modes .....................................78  
Continuous Positive Airway Pressure (CPAP) Mode..............................78  
Spontaneous (S) Mode....................................................................................80  
Spontaneous/Timed (S/T) Mode.................................................................82  
Timed (T) Mode .................................................................................................82  
Pressure Control (PC) Mode...........................................................................83  
Pressure Control Synchronized Intermittent Mandatory  
Ventilation (PC-SIMV) Mode..........................................................................83  
Control Ventilation (CV) Mode .....................................................................84  
Assist Control (AC) Mode................................................................................85  
Synchronized Intermittent Mandatory Ventilation  
(SIMV) Mode........................................................................................................85  
Viewing and Changing Options Menu Items ..................................................86  
Viewing the Alarm Log ............................................................................................89  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
v
Viewing the Event Log .............................................................................................90  
Viewing Device Information ..................................................................................90  
Updating Prescriptions Using the SD Card ...............................................................91  
Changing and Viewing Settings in Limited Menu Access Mode.......................94  
Activating Your Primary or Secondary Prescription ......................................95  
Viewing and Changing My Settings Menu Items...........................................96  
Connecting the Ventilator to the Patient...................................................................98  
Chapter 6. Ventilator Alarms ............................................................................................................................99  
Audible and Visual Alarm Indicators.........................................................................100  
Audio Pause and Alarm Reset Features...................................................................105  
What to Do When An Alarm Occurs..........................................................................106  
Alarm Summary Table....................................................................................................107  
Chapter 7. Cleaning and Maintenance.......................................................................................................121  
Cleaning the Ventilator..................................................................................................121  
Cleaning and Replacing the Air Inlet Filter.............................................................122  
Replacing the Air Inlet Path Foam.............................................................................123  
Cleaning the Patient Circuit.........................................................................................124  
Cleaning Instructions (Reusable Circuits) ......................................................124  
Preventive Maintenance ...............................................................................................126  
Chapter 8. Troubleshooting............................................................................................................................129  
Chapter 9. Accessories......................................................................................................................................133  
Adding a Humidifier .......................................................................................................133  
Adding Supplemental Oxygen to the Device .......................................................133  
Using a Remote Alarm Unit..........................................................................................134  
Using a Nurse Call System............................................................................................135  
Using a Secure Digital (SD) Card ................................................................................135  
Using the Philips Respironics DirectView Software ............................................136  
Using the Optional In-Use Bag ...................................................................................136  
Traveling with the System............................................................................................137  
Airline Travel......................................................................................................................137  
Table of Contents  
Download from Www.Somanuals.com. All Manuals Search And Download.  
vi  
Chapter 10. System Checkout Procedures................................................................................................139  
Tools Required..................................................................................................................139  
Visual Inspection..............................................................................................................139  
Initial Setup........................................................................................................................140  
Settings and Alarms Tests.............................................................................................140  
Battery Function Verification.......................................................................................148  
Alarm and Event Log Clean-Up ..................................................................................150  
Results .................................................................................................................................150  
Chapter 11. Technical Specifications...........................................................................................................151  
Chapter 12. Glossary .........................................................................................................................................155  
Chapter 13. EMC Information ........................................................................................................................161  
Index ........................................................................................................................................................................165  
Limited Warranty.................................................................................................................................................169  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
1. Introduction  
This chapter provides an overview of the Trilogy200 device.  
Package Contents  
The Trilogy200 system may include the following components.  
Some components are optional accessories that may not be  
packaged with the device.  
Package Contents  
Reusable Gray Foam Filters  
In-Use Bag  
Trilogy200  
Clinical Manual  
Disposable Active Circuit, No Water Trap  
Secure Digital  
(SD) Card  
(pre-installed)  
Disposable Passive Circuit, No Water Trap  
Detachable  
Battery  
Flexible Tubing  
Flexible Trach Adapter  
(Not Shown)  
Universal Porting Block  
Exhalation Valve  
(pre-installed)  
Line (pre-assembled  
O2 Inlet  
Quick Connects  
to AED)  
AC Power Cord  
Active Exhalation  
Device (AED)  
Proximal  
Pressure Line  
(pre-assembled  
to AED)  
Flow Sensor  
Whisper Swivel II  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
Intended Use  
The Philips Respironics Trilogy200 system provides continuous  
or intermittent ventilatory support for the care of individuals who  
require mechanical ventilation. Trilogy200 is intended for pediatric  
through adult patients weighing at least 5 kg (11 lbs.).  
The device is intended to be used in home, institution/hospital, and  
portable applications such as wheelchairs and gurneys, and may be  
used for both invasive and non-invasive ventilation. It is not intended  
to be used as a transport ventilator.  
The system is recommended to be used only with various  
combinations of Philips Respironics-approved patient circuit  
accessories, such as patient interface devices, humidifiers, water  
traps, and circuit tubing.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Warnings and Cautions  
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.  
Warnings  
A warning indicates the possibility of injury to the user or operator.  
Patient Monitoring Prior to placing a patient on the ventilator, a clinical assessment should  
be performed to determine:  
The device alarm settings  
Needed alternative ventilation equipment  
If an alternative monitor (i.e., an alarming Pulse Oximeter or  
Respiratory Monitor) should be used  
Alternative  
Ventilation  
For ventilator dependent patients, always have alternate ventilation  
equipment, such as a back-up ventilator, manual resuscitator, or similar  
device, available.  
Ventilator dependent patients should be continuously monitored by  
qualified personnel. These personnel should be prepared to provide  
alternate therapy in the event of ventilator failure or inoperative  
equipment.  
Patient Disconnect For ventilator dependent patients, do not rely on any single alarm  
Protection  
to detect a circuit disconnect condition. The Low Tidal Volume, Low  
Minute Ventilation, Low Respiratory Rate, and Apnea alarms should  
be used in conjunction with the Circuit Disconnect and Low Peak  
Inspiratory Pressure alarms.  
Test the operation of the circuit disconnect function daily and  
whenever a change is made to the patient circuit. An increase in circuit  
resistance can prevent proper operation of some alarms.  
Speaking valves, Heat Moisture Exchangers (HMEs), and filters create  
additional circuit resistance and may affect the performance of alarms  
chosen for circuit disconnect protection.  
Do not set the Low Peak Inspiratory Pressure alarm too low, or the  
system may not detect large circuit leaks or a patient disconnect.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Personnel  
Qualifications  
Trilogy200 is a restricted medical device designed for use by  
Respiratory Therapists or other trained and qualified caregivers under  
the supervision of a physician.  
The prescription and other device settings should only be changed on  
the order of the supervising physician.  
The operator of the ventilator is responsible to read and understand  
this manual before use.  
Modes of  
Ventilation  
The device can provide therapies typically associated with both  
ventilator dependent and non-dependent patients. The mode of  
ventilation, circuit type, and alarm strategies should be chosen after a  
clinical evaluation of each patient’s needs.  
C-Flex, Bi-Flex, and AVAPS are intended for use by adult patients.  
SD Card  
Prescription  
Changes  
When you change the device prescription, alarms, and other settings  
using the SD card, Trilogy200 requires that the caregiver review and  
verify the changes prior to the changes being used by the device. The  
caregiver or health care professional is responsible to ensure that the  
prescription settings are correct and compatible with the patient after  
using this feature. Installing the wrong prescription for a particular  
patient may result in improper therapy, lack of appropriate safety  
monitoring, and risk of death or injury to the patient.  
Electrical  
Interference  
This device is intended for use in the electromagnetic environment  
specified in Chapter 13 of this manual. The user of this device should  
make sure it is used in a compatible environment.  
Portable and mobile RF communications equipment should be  
used no closer to any part of the device, including cables, than the  
recommended separation distance calculated using the information  
provided in Chapter 13 of this manual.  
Battery Back-up  
Power  
The internal battery is NOT intended to serve as a primary power  
source. It should only be used when other sources are not available or  
briefly when necessary; for example, when changing power sources.  
The ventilator has a two-stage low battery alarm. The medium priority  
alarm indicates that approximately 20 minutes of operation remain,  
and the high priority alarm indicates that less than 10 minutes of  
operation remain. Actual run time may be more or less than this and  
varies with battery age, environmental conditions, and therapy.  
Immediately seek an alternate power source when the “Low Battery”  
alarm appears. Complete power failure and loss of therapy is imminent.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Operating  
and Storage  
Temperatures  
Do not use this device if the ambient temperature is warmer than  
40˚ C (104˚ F). If the device is used at room temperatures warmer than  
40˚ C, the temperature of the airflow may exceed 43˚ C. This could  
cause system alarms, thermal irritation, or injury to the patient’s airway.  
Bacteria Filter  
Philips Respironics recommends that a main line outlet bacteria filter  
(Part Number 342077) be used whenever the device is used for invasive  
therapy or if the ventilator may be used on multiple patients.  
Patient Circuits  
(General)  
The ventilator should only be used with patient interfaces (e.g., masks,  
circuits and exhalation ports) recommended by Philips Respironics.  
Proper operation of the device, including alarms, with other circuits has  
not been verified by Philips Respironics and is the responsibility of the  
health care professional or respiratory therapist.  
When adding any components to the breathing system, the flow  
resistance and dead space of the added components such as  
humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and  
filters should be carefully considered in relation to the potential for  
adverse effects on the patient’s ventilatory management and device  
alarms.  
Passive Circuits  
Active Circuits  
An exhalation port is required when using a passive circuit.  
For the passive circuit, at low expiratory pressures, the flow through  
the exhalation port may be inadequate to clear all exhaled gas from  
the tubing – some rebreathing may occur. Rebreathing of exhaled air  
for longer than several minutes can in some circumstances lead to  
suffocation.  
Only use the active exhalation devices designed for Trilogy200. Philips  
Respironics has not verified proper operation of other active exhalation  
devices, and their use may result in improper or unsafe device  
operation.  
With active exhalation circuits, the exhalation device must be operating  
properly for the ventilator to deliver therapy. The exhalation device  
should be inspected on a daily basis and replaced whenever necessary.  
System Checkout  
Do not use the ventilator on a patient until a system checkout has been  
performed. See Chapter 10 of this manual.  
To make sure the device is operating properly at start-up, always verify  
that the audible tone sounds and the alarm LEDs light red and then  
yellow momentarily. Contact Philips Respironics or an authorized  
service center for service if these indications do not occur at start-up.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
Remote Alarms  
When using a remote alarm, make sure you fully test the remote alarm  
connector and cable by verifying that:  
Annunciated alarms on the ventilator are also  
annunciated on the remote alarm.  
Disconnecting the remote alarm cable from the  
ventilator or from the remote alarm results in an alarm  
notification at the remote alarm.  
The remote alarm should be tested daily.  
Oxygen  
When administering fixed-flow supplemental oxygen, the oxygen  
concentration may not be constant. The inspired oxygen concentration  
will vary, depending on the pressures, patient flows and circuit leak.  
Substantial leaks may reduce the inspired oxygen concentration to less  
than the expected value. Appropriate patient monitoring should be  
used, as medically indicated, such as an alarming pulse oximeter.  
This device DOES NOT alarm for loss of the low flow oxygen supply.  
This device is intended to be connected to a low flow (0-15 l/min)  
oxygen source such as an oxygen concentrator or other oxygen source  
equipped with a pressure regulator (set to 50 PSI or less) and a flow  
regulator/meter.  
Do not connect the device to an unregulated or high pressure oxygen  
source.  
The device may result in incorrect flow and tidal volume measurements  
and improper operation of related alarms if you add low flow oxygen  
directly into the patient circuit or mask instead of using the oxygen  
inlet on the back of the ventilator.  
Oxygen supports combustion. Oxygen should not be used while  
smoking or in the presence of an open flame.  
If oxygen is used with the device, the oxygen flow must be turned  
off when the device is not in use. Explanation of the Warning:  
When the device is not in operation and the oxygen flow is left on,  
oxygen delivered into the tubing may accumulate within the device’s  
enclosure.  
Fire or Explosion  
The ventilator should not be operated in the presence of flammable  
gasses. This could cause a fire or explosion.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
Alarms  
Respond immediately to any alarm. It may indicate a potentially  
life-threatening condition. Refer to the Alarms and Troubleshooting  
chapters for more information.  
Visually monitor the patient and ventilator at all times during an Alarm  
Silence period. Allowing alarms to continue without intervention may  
result in harm to the patient.  
If the high priority “Low Internal Batterymessage appears,  
immediately connect the ventilator to an alternate power source. If no  
alternate power source is available, immediately place the patient on  
an alternate source of ventilation.  
If the “Ventilator Inoperablealarm occurs, immediately place the  
patient on an alternate source of ventilation.  
You should not rely on any single alarm to detect a circuit disconnect  
condition. The Low Tidal Volume, Low Minute Ventilation, Low  
Respiratory Rate, and Apnea alarms should be used in conjunction with  
the Circuit Disconnect alarm.  
Make sure the alarm volume is set loud enough to be heard by the  
caregiver. Consider the use of a remote alarm.  
Trilogy200 offers the following circuit type selections:  
Passive  
Active Flow  
Active PAP (Proximal Airway Pressure )  
The Passive circuit type provides an ESTIMATE of Vte.  
Only the Active Flow circuit type directly measures exhaled tidal  
volume (Vte).  
The Active PAP circuit type DOES NOT measure Vte and only provides  
for an indication of the delivered tidal volume (Vti).  
Improperly  
Functioning  
Ventilator  
If you notice any unexplained changes in the performance of the  
device, if it is making unusual sounds, if the device or detachable  
battery are dropped, if water is spilled into the enclosure, or if the  
enclosure is cracked or broken, discontinue use and contact Philips  
Respironics or an authorized service center for service.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
Maintenance  
Follow the service recommendations provided in Chapter 7 of this  
manual.  
Periodically inspect electrical cords, cables, and the detachable battery  
pack for damage or signs of wear. Discontinue use and replace if  
damaged.  
Repairs and adjustments must be performed by Philips Respironics-  
authorized service personnel only. Unauthorized service could cause  
death or injury, invalidate the warranty, or result in costly device  
damage.  
Cleaning  
To avoid electrical shock, always unplug the power cord from the wall  
outlet before cleaning the ventilator.  
(Refer to Chapter 7  
for detailed cleaning  
instructions.)  
Do not immerse the device in any fluids or spray the device with water  
or cleaners. Clean the device with a cloth dampened with an approved  
cleaner.  
If the device has been exposed to rain or dampness, dry the device  
including the area around the power cord connection with the power  
cord disconnected from the device before applying AC power.  
Cautions  
A caution indicates the possibility of damage to the device.  
Storage  
The internal and detachable batteries will self-discharge in storage. If it  
is desired to keep the batteries fully charged (for example, as a back-  
up ventilator), plug the device into AC power for about eight hours  
every 16 days. Alternatively, the ventilator may be left continuously  
connected to AC power without battery degradation.  
Allowing the batteries to fully discharge will not harm the batteries or  
lose device settings, but may require a longer battery charge time prior  
to use.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9
Operating  
and Storage  
Temperatures  
The device may only be operated at temperatures between 5˚ C and  
40˚ C (41˚ F and 104˚ F).  
Do not operate the device in direct sunlight or near a heating  
appliance because these conditions can increase the temperature of  
the airflow delivered to the patient.  
Prolonged operation or storage at elevated temperatures may reduce  
the service life of the battery and other internal components of the  
ventilator.  
The ventilator has an internal and detachable Lithium-Ion Battery. Do  
not expose the device or detachable battery to temperatures above  
40˚ C (104˚ F) during use, or above 60˚ C (140˚ F) during storage. This  
will reduce battery life and may increase the risk of fire or damage the  
battery.  
Condensation  
Condensation may affect operation or accuracy of the device. If the  
device has been exposed to either very hot or very cold temperatures  
during storage, allow it to adjust to ambient temperature before  
starting therapy.  
Air Filter  
The reusable foam inlet filter is required to protect the ventilator from  
dirt and dust. Wash periodically and replace when damaged for proper  
operation.  
Cooling Air Vents  
Battery Life  
Do not block the cooling air vents located on the base and the rear  
of the device. This may cause the device to overheat in high ambient  
temperatures or at high therapy settings.  
The internal and detachable batteries wear out based on the amount  
of use (hours or full charge-discharge cycles). The battery capacity and  
life are also reduced by operation at higher temperatures.  
Detachable Battery Only use the Philips Respironics Trilogy Detachable Battery with the  
ventilator.  
Cleaning  
Do not steam autoclave the ventilator. Doing so will destroy the  
ventilator.  
Do not immerse the device in liquid or allow any liquid to enter the  
enclosure or inlet filter.  
Do not spray water or any other solutions directly onto the ventilator.  
Do not use harsh detergents, abrasive cleaners, or brushes to clean the  
ventilator system. Use only cleaning agents and methods listed in this  
manual.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
Patient Circuit  
Exhalation valves, patient circuits, and water traps are shipped clean,  
not sterile. Cleaning and disinfection of these parts should follow  
individual institution processes and conform to guidelines provided by  
Philips Respironics with each accessory.  
External DC Power  
Do not use the same external battery to operate both the ventilator  
and any other equipment such as power chairs.  
An external battery should only be connected to the ventilator using  
the Philips Respironics Trilogy External Battery Cable. This cable is  
fused, pre-wired, and properly terminated to ensure safe connection  
to a standard deep-cycle, lead acid battery. Use of any other adapter or  
cable may cause improper operation of the ventilator.  
The ventilator should only be connected to an automotive electrical  
system using the Philips Respironics Trilogy Automotive Adapter  
(when available). This adapter is fused, filtered, and designed for safe  
connection to a standard automotive electrical system. Use of any  
other adapter or cable may cause improper operation of the ventilator.  
Do not operate the ventilator from a car electrical system when  
starting the vehicle or jump-starting the vehicle. Electrical transients  
during starting may cause improper operation of the ventilator.  
Electrostatic  
Discharge (ESD)  
Do not use antistatic or conductive hoses or conductive patient tubing  
with the device.  
Notes  
This product does not contain natural latex rubber or dry  
natural rubber in patient or operator accessible areas or  
in the air path or breathing circuit.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11  
Contraindications  
If the patient has any of the following conditions, consult their health  
care professional before using the device in a non-invasive mode:  
Inability to maintain a patent airway or adequately clear  
secretions  
At risk for aspiration of gastric contents  
Diagnosed with acute sinusitis or otitis media  
Epistaxis, causing pulmonary aspiration of blood  
Hypotension  
System Overview  
This ventilator provides both pressure control and volume modes of  
therapy. The device can provide non-invasive or invasive ventilation.  
It can be used to provide total therapy to patients as they progress  
from non-invasive to invasive ventilation.  
When prescribed, the device provides numerous special features to  
help make patient therapy more comfortable. For example, the ramp  
function allows you to lower the pressure when trying to fall asleep.  
The air pressure will gradually increase until the prescription pressure  
is reached. Additionally, the Flex comfort feature provides increased  
pressure relief during the expiratory phase of breathing.  
The ventilator can be operated using several different power  
sources, including an internal Lithium-Ion battery. This battery is  
automatically used when the detachable Lithium-Ion battery pack,  
external Lead Acid battery, or AC power are not available.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12  
Symbols  
The following symbols appear on the device.  
Front Panel  
Symbol  
Description  
Therapy Start/Stop  
Alarm Indicator/Audio Pause  
AC Power Indicator  
~
Rear and Side Panels  
Symbol  
Description  
AC Power Connector  
~
Secure Digital (SD) Card Slot  
Serial Port Connector  
Remote Alarm Connector  
Ethernet Connector  
DC Power Connector  
Oxygen Inlet  
O2  
Consult accompanying instructions for use.  
Type BF Applied Part  
Class II (Double Insulated)  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
13  
Drip Proof Equipment  
For Airline Use. Complies with RTCA-D0160F section 21, category M.  
How to Contact Philips Respironics  
To have your device serviced, contact Philips Respironics Customer  
Service department at 1-724-387-4000 or 1-800-345-6443.  
Chapter 1 Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
14  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
2. System Description  
This chapter describes the front and rear panel device controls and  
features.  
2
1
Front Panel Features  
The front panel contains the control buttons, visual indicators, and  
display screen.  
3
4
5
Buttons  
Front Panel Controls and  
Display Screen  
The following buttons are included on the front panel of the device.  
Note: When you start therapy,  
the display backlight and the  
backlights on the buttons turn  
on, the red and yellow alarm  
LEDs turn on momentarily, and  
an audible indicator sounds  
to indicate that therapy has  
started. The Startup screen  
appears on the display.  
1.  
Start/Stop Button  
This button turns the airflow on or off, starting or stopping  
therapy.  
2.  
Alarm Indicator and Audio Pause Button  
This button serves two purposes: it temporarily silences  
the audible portion of an alarm, and it also acts as an alarm  
indicator. When silencing an alarm, if the cause of the alarm is  
not corrected, the alarm sounds again after one minute. Each  
time the button is pressed, the alarm silence period resets to one  
minute. See Chapter 6 for more information.  
WARNING  
To make sure the device is operating  
properly at start-up, always verify  
that the audible tone sounds and  
the alarm LEDs light red and then  
yellow momentarily. Contact Philips  
Respironics or an authorized service  
center for service if these indications  
do not occur at start-up.  
3.  
Up/Down Button  
This button allows you to navigate the display menu and edit  
device settings.  
Chapter 2 System Description  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16  
4.  
Left and Right Buttons  
These buttons allow you to select display options or perform  
certain actions specified on-screen.  
Visual Indicators  
Several power and alarm indicators appear on the front panel.  
5. AC Power LED  
In the lower right corner of the front panel, a green LED (~)  
indicates that AC power is applied to the device. This light  
remains on as long as adequate AC power is available.  
6. Keypad Backlight LEDs  
The Start/Stop, Up/Down, and Left/Right buttons all have a white  
LED that lights up if the keypad backlight is turned on in the  
device Options menu. See Chapter 5 for more information.  
7. Red Alarm LED  
On the Alarm Indicator/Audio Pause button, a red light flashes to  
indicate a high priority alarm.  
8. Yellow Alarm LED  
Note: See Chapter 6 for more  
information about high,  
medium, and low priority  
alarms.  
On the Alarm Indicator/Audio Pause button, a yellow light  
flashes to indicate a medium priority alarm. A solid yellow light  
indicates a low priority alarm.  
Display Screen  
The display screen allows you to view settings, system status  
information, real-time patient data, alarms, and logs. You can also  
modify certain settings on the display screen.  
See Chapter 5 for more information on viewing and modifying  
device settings.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
17  
Side and Rear Panel Features  
2
The ventilator’s side and rear panels contain the following connectors  
and features, shown at right.  
3
1. AC Power Inlet  
You can plug the AC power cord into this connector, located on  
the right side of the ventilator.  
2. Breathing Circuit Connection  
1
The breathing circuit connector is located on the right side of  
the device. You can connect your circuit tubing system here. See  
Chapter 4 for details.  
Right Side Panel  
3. Exhalation Porting Block  
The porting block used here depends on the type of exhalation  
device you are using. The Universal Exhalation Porting Block is  
shown here. See Chapter 4 for more information.  
4. Secure Digital (SD) Card Slot  
On the left side of the device is a slot for the optional SD Card.  
You can have the patient record usage and therapy information  
from the device on the SD card.  
4
Left Side Panel  
Chapter 2 System Description  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18  
5. Serial Connector  
You can use this connector to connect the device to a computer  
running PC Direct or Sleepware software or to other Philips  
Respironics devices such as Alice 5 and AOM. Use the Trilogy  
RS232 Serial Cable to connect the ventilator to the external  
device or computer.  
11  
10  
12  
8
9
5
6
7
6. Remote Alarm/Nurse Call Connector  
If you are using an optional remote alarm or nurse call system  
with the ventilator, you can connect the Philips Respironics  
remote alarm adapter cable or nurse call adapter cable to this  
connector.  
Rear Panel  
7. Ethernet Connector (when available)  
You can connect a PC or router to this connector to upload  
therapy information to a secure web site so you can review  
therapy information remotely or remotely troubleshoot and  
service the device.  
8. External Battery Connector (DC Power Inlet)  
You can connect an external, stand-alone lead acid battery here,  
using the Trilogy External Battery cable.  
9. Oxygen (O2) Inlet Connector  
If using low flow supplemental oxygen, connect the oxygen  
source to this connector using one of the O2 Inlet Quick Connects  
provided with the device.  
10. Air Inlet and Filter  
Insert the filter supplied with the device into the air inlet.  
11. Detachable Battery Pack Slot  
If you are using the Philips Respironics Lithium-Ion detachable  
battery pack to power the device, attach it here.  
12. Cord Retainer  
Secure the power cord using the cord retainer to prevent  
someone from accidentally disconnecting the power cord. See  
Chapter 4 for more information.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
3. Modes, Features, and Alarms  
Therapy Modes  
The device provides Pressure Control Ventilation (PCV) and Volume  
Control Ventilation (VCV) for non-invasive and invasive patients.  
Pressure Control ventilation delivers a prescribed pressure to  
the patient according to set breath rate and set inspiration time  
parameters. This means that each breath is controlled so that a  
prescribed amount of pressure is delivered to the patient. The device  
offers six different Pressure Control modes of operation:  
CPAP – Continuous Positive Airway Pressure  
S – Spontaneous Ventilation  
S/T – Spontaneous/Timed Ventilation  
T – Timed Ventilation  
PC – Pressure Control Ventilation  
PC-SIMV – Pressure Controlled Synchronized Intermittent  
Mandatory Ventilation  
Volume Control ventilation delivers a prescribed inspired tidal  
volume to the patient according to set breath rate and set inspiratory  
time parameters. This means that each breath is controlled so that a  
prescribed tidal volume is delivered to the patient. The device offers  
three different Volume Control modes of operation:  
AC – Assist Control Ventilation  
CV – Control Ventilation  
SIMV – Synchronized Intermittent Mandatory Ventilation  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
20  
Breath Types  
There are four breath types that apply to the Volume Control and  
Pressure Control ventilation therapy modes:  
Spontaneous  
Mandatory  
Assisted  
Sigh  
Spontaneous Breath  
A Spontaneous breath is triggered by the patient. Breaths are  
initiated by the patient’s inhalation effort, and air delivery is  
controlled based on the current pressure or volume setting. Breaths  
are terminated by either the ventilator settings or by the patient’s  
exhalation effort, depending on the mode selected.  
Mandatory Breath  
A Mandatory breath (or machine breath) is completely controlled by  
the ventilator. The ventilator controls both the beginning (triggering)  
and end (cycling) of the inspiratory phase.  
Assisted Breath  
An Assisted breath is controlled by both the patient and the ventilator.  
Breaths are initiated by the patient’s effort and air delivery is controlled  
by the current pressure or volume settings. Volume Assisted breaths  
will deliver the prescribed Tidal Volume within the prescribed  
Inspiratory Time. Pressure Assisted breaths will deliver the prescribed  
Inspiratory Pressure for the prescribed Inspiratory Time. Breaths are  
terminated when the Inspiratory Time setting has been reached.  
Sigh  
A Sigh breath is a breath where 150% of the prescribed volume  
is delivered. The device will deliver this breath once every 100  
Mandatory or Assist breaths when the Sigh setting is enabled. Sigh  
breaths are only available in volume modes of ventilation.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
21  
Therapy Mode Table  
The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table  
are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity,  
Leak Compensation, and Flow Cycle settings will display.  
Note: Pressure Support, referred to in the table below and later in this manual, is defined as IPAP - EPAP or Pressure -  
PEEP (PC-SIMV).  
Therapy Modes  
CPAP  
S
S/T  
T
PC  
PC-SIMV  
CV  
AC  
SIMV  
Dual Prescription  
Circuit Type  
CPAP  
IPAP  
AVAPS (On, O)*  
IPAP Max Pressure  
IPAP Min Pressure  
EPAP  
Pressure  
Pressure Support (PS)  
PEEP  
Tidal Volume  
Breath Rate  
Inspiratory Time  
Trigger Type*  
Flow Trigger Sensitivity  
Leak Compensation  
Flow Cycle Sensitivity  
Ramp Length  
Ramp Start Pressure  
Flex *  
Rise Time  
Flow Pattern  
Sigh  
Circuit Disconnect  
Apnea  
Apnea Rate  
High Vte*  
Low Vte*  
High Vti*  
Low Vti*  
High Minute Ventilation  
Low Minute Ventilation  
High Respiratory Rate  
Low Respiratory Rate  
High Inspiratory Pressure  
Low Inspiratory Pressure  
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only  
available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit  
type. Leak compensation is only available with the Active Flow circuit type.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
22  
Pressure Control Ventilation Therapy Modes  
Pressure Control ventilation modes deliver a prescribed pressure to  
the patient.  
Continuous Positive Airway Pressure (CPAP) Mode  
In the Continuous Positive Airway Pressure (CPAP) mode, the device  
delivers a continuous pressure to the patient at all times. All breaths  
in this mode are Spontaneous breaths.  
Spontaneous (S) Mode  
In Spontaneous (S) mode, the device delivers bi-level pressure  
support. This mode provides only spontaneous breaths. In this  
mode, an Inspiratory Positive Airway Pressure (IPAP) is delivered  
during inhalation and a lower Expiratory Positive Airway Pressure  
(EPAP) is delivered during exhalation. The following illustration  
describes these concepts.  
= Pressure Supported  
Spontaneous Breath  
IPAP = 26 cm H2O  
1
EPAP = 6 cm H2O  
PS = IPAP-EPAP = 20 cm H2O  
PRESSURE  
5 seconds  
4 seconds  
8 seconds  
26  
20  
10  
6
1
1
1
1
Enter S mode  
S Mode  
TIME  
5
10  
15  
20  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
23  
Spontaneous/Timed (S/T) Mode  
In Spontaneous/Timed (S/T) mode, the device delivers bi-level  
pressure support. This mode provides Spontaneous and Mandatory  
breaths. A Mandatory breath is delivered if the patient does not  
spontaneously breathe within the prescribed Breath Rate (BPM)  
setting. This ensures that the patient receives a minimum number  
of breaths per minute. In this mode, an IPAP is delivered during  
inhalation and a lower EPAP is delivered during exhalation. The  
duration of a Spontaneous breath is determined by the patient effort.  
The duration of a Mandatory breath is determined by the inspiratory  
time setting. The following illustration describes these concepts.  
= Pressure Supported  
Spontaneous Breath  
= Mandatory Breath  
(note longer inspiratory  
time)  
IPAP = 26 cm H2O  
1
2
EPAP = 6 cm H2O  
Rate = 10 BPM  
PS = IPAP-EPAP = 20 cm H2O  
PRESSURE  
6 seconds  
6 seconds  
6 seconds  
26  
20  
S/T Mode  
10  
6
1
1
2
1
Enter S/T mode  
TIME  
5
10  
15  
20  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
24  
Timed (T) Mode  
In Timed (T) mode, the device delivers bi-level pressure support.  
This mode delivers Mandatory breaths only. A Mandatory breath  
is delivered according to the prescribed BPM and inspiratory time  
settings. This also means that the ventilator will not respond to  
patient effort. In this mode, an IPAP is delivered during inhalation and  
a lower EPAP is delivered during exhalation. The following illustration  
describes these concepts.  
IPAP = 26 cm H2O  
= Mandatory Breath  
2
EPAP = 6 cm H2O  
Rate = 10 BPM  
PS = IPAP-EPAP = 20 cm H2O  
PRESSURE  
6 seconds  
6 seconds  
6 seconds  
6 seconds  
26  
20  
10  
6
2
2
2
2
Enter T mode  
T Mode  
TIME  
5
10  
15  
20  
25  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
25  
Pressure Control (PC) Mode  
In Pressure Control (PC) mode, the device delivers bi-level pressure  
support. This mode delivers Assist and Mandatory breaths. This  
mode is identical to S/T mode, except that all breaths have a fixed  
inspiratory time.  
IPAP = 26 cm H2O  
= Mandatory Breath  
= Assist Breath  
2
3
EPAP = 6 cm H2O  
Rate = 10 BPM  
PS = IPAP-EPAP = 20 cm H2O  
PRESSURE  
6 seconds  
6 seconds  
6 seconds  
26  
20  
PC Mode  
10  
6
3
3
2
3
Enter PC mode  
TIME  
5
10  
15  
20  
Pressure Control - Synchronized Intermittent Mandatory  
Ventilation (PC-SIMV) Mode  
The Pressure Control – Synchronized Intermittent Mandatory  
Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and  
Mandatory breaths. This mode uses a time window to decide  
what type of breaths should be delivered. This time window is the  
maximum time between breaths according to the set Breath Rate.  
When you enter PC-SIMV mode, the time window is started. If  
the patient does not trigger a breath, the ventilator will provide  
a Mandatory breath when the time expires and then begin the  
process again. Mandatory breaths deliver the Pressure setting during  
inhalation and a lower Peak End Expiratory Pressure (PEEP)  
during exhalation.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26  
During the time window, if patient effort is detected, either a  
Spontaneous or an Assist breath is delivered for the first effort. If the  
last breath delivered was Mandatory, then a Spontaneous breath is  
delivered for the first patient effort of the window. If the last breath  
delivered was an Assist or Spontaneous breath, then an Assist breath  
is delivered for the first patient effort of the window. Assist breaths  
deliver the Pressure setting during inhalation and the PEEP setting  
during exhalation.  
During the time window, if patient effort continues to be detected  
(after the first patient triggered breath), Spontaneous breaths are  
delivered for the remainder of the window. The figure below provides  
example breath patterns in PC-SIMV mode.  
Pressure = 26 cm H2O  
Pressure Support = 10 cm H2O  
PEEP = 6 cm H2O  
= Spontaneous  
Breath  
= Mandatory Window  
M
S
1
= Spontaneous Window  
= Mandatory Breath  
= Assist Breath  
2
3
Rate = 5 BPM  
PRESSURE  
12 seconds  
12 seconds  
12 seconds  
12 seconds  
M
S
M
S
M
M
26  
20  
Sample Breath Patterns in  
PC-SIMV Mode  
10  
6
2
1
1
3
1
2
Enter PC-SIMV mode  
TIME  
5
10  
15  
20  
25  
30  
35  
40  
45  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
27  
Volume Control Ventilation Therapy Modes  
Volume Control ventilation modes deliver a prescribed inspired tidal  
volume according to a set Breath Rate and a set Inspiratory Time.  
Control Ventilation (CV) Mode  
In Control Ventilation (CV) mode, the device delivers volume control  
therapy. This mode provides only Mandatory breaths. A Mandatory  
breath is delivered according to the prescribed BPM setting for the  
prescribed Inspiratory Time. This also means that the ventilator  
will not respond to patient effort. In this mode, the Tidal Volume is  
delivered during inhalation and PEEP is delivered during exhalation.  
The following illustration describes these concepts.  
Tidal Volume = 500 mL  
Rate = 10 BPM  
= Mandatory Breath  
2
Inspiratory Time = 1.5 seconds  
Flow Pattern = Square  
FLOW  
6 seconds  
6 seconds  
6 seconds  
6 seconds  
50  
Enter CV mode  
Control Mode  
2
2
2
10  
15  
0
TIME  
5
20  
-50  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
28  
Assist Control (AC) Mode  
In Assist Control (AC) mode, the device delivers volume control  
therapy. This mode provides Assist and Mandatory breaths. An Assist  
breath is started when there is patient effort, but it is ended when  
the Inspiratory Time setting has been met. A Mandatory breath  
is delivered if the patient does not spontaneously breathe within  
the prescribed BPM setting. This ensures that the patient receives  
a minimum number of breaths per minute. In this mode, the Tidal  
Volume is delivered during inhalation and PEEP is delivered during  
exhalation. The following illustration describes these concepts.  
= Mandatory Breath  
= Assist Breath  
Tidal Volume = 500 mL  
Rate = 10 BPM  
2
3
Inspiratory Time = 2.0 seconds  
Flow Pattern = Square  
FLOW  
6 seconds  
5 seconds  
6 seconds  
50  
Enter A/C mode  
2
3
2
10  
15  
0
TIME  
5
20  
Assist Control Mode  
-50  
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode  
In Synchronized Intermittent Mandatory Ventilation (SIMV) mode, the  
device delivers both volume control and pressure control therapy.  
This mode provides Spontaneous, Assist, and Mandatory breaths.  
This mode uses a time window to decide what type of breaths should  
be delivered. This time window is the maximum time between  
breaths according to the set Breath Rate.  
When you enter SIMV mode, the time window is started. If the  
patient does not provide any effort, the ventilator will provide a  
Mandatory Volume breath when the time expires and then begin the  
process again. Mandatory breaths deliver the Tidal Volume setting  
during inhalation and provide a lower Peak End Expiratory Pressure  
(PEEP) during exhalation.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
29  
During the time window, if patient effort is detected, either a  
Spontaneous or an Assist breath is delivered for the first effort. If the  
last breath delivered was Mandatory, then a Spontaneous breath is  
delivered for the first patient effort of the window. If the last breath  
delivered was an Assist or Spontaneous breath, then an Assist breath  
is delivered for the first patient effort of the window. Assist breaths  
deliver the prescribed Tidal Volume during inhalation and the PEEP  
setting during exhalation.  
During the time window, if patient effort continues to be detected  
(after the first patient triggered breath), Spontaneous breaths are  
delivered for the remainder of the window. Spontaneous breaths  
deliver the prescribed Pressure Support setting above PEEP during  
inhalation and PEEP during exhalation. The figure below provides  
example breath patterns in SIMV mode.  
Tidal Volume = 500 mL  
Breath Rate = 5 BPM  
= Pressure Supported  
Breath  
= Mandatory Window  
M
S
1
= Spontaneous Window  
Inspiratory Time = 3.0 seconds  
Pressure Support = 10 cm H2O  
PEEP= 6 cm H2O  
= Mandatory Breath  
= Assist Breath  
2
3
PRESSURE  
12 seconds  
12 seconds  
12 seconds  
12 seconds  
M
S
M
S
M
M
26  
20  
10  
6
2
1
1
3
1
2
SIMV Mode  
Enter SIMV mode  
TIME  
5
10  
15  
20  
25  
30  
35  
40  
45  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
30  
Therapy Mode Features  
The device has several additional features that enhance patient  
comfort.  
Flex Comfort Feature  
Note: Flex is not available if  
AVAPS is enabled or if an active  
circuit is used.  
The device consists of a special comfort feature called Flex. The  
device provides the Flex feature in CPAP mode and S mode. This  
feature is only available when Auto-Trak is enabled.  
Note: The patient has access to  
this setting, if Flex is enabled.  
C-Flex  
When in CPAP mode, if C-Flex is enabled, it enhances patient comfort  
by providing pressure relief during the expiratory phase of breathing.  
In the following diagram, the dashed lines represent normal CPAP  
therapy in comparison to the bold line representing C-Flex. C-Flex  
levels of 1, 2, or 3 progressively reflect increased pressure relief.  
Note: C-Flex is not available if  
CPAP is set to 4 cm H2O.  
Note: C-Flex, Bi-Flex and AVAPS  
are intended for use by adult  
patients.  
PRESSURE  
Inhalation  
Exhalation  
CPAP  
1
2
C-Flex in Comparison to  
Traditional CPAP Therapy  
3
TIME  
C-Flex pressure relief is determined by the C-Flex setting and the  
amount of patient flow. C-Flex returns to the set pressure by the end  
of exhalation, when the airway is most vulnerable to closure.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
31  
Bi-Flex  
Note: Bi-Flex is only available  
up to 25 cm H2O in S mode.  
In S mode, the Bi-Flex attribute adjusts therapy by inserting a small  
amount of pressure relief during the latter stages of inspiration and  
during the beginning part of exhalation. In the following diagram,  
the bold lines represent Bi-Flex in comparison to the dashed line  
representing normal BiPAP therapy. Bi-Flex levels of 1, 2, or 3  
progressively reflect increased pressure relief that will take place at  
the end of inspiration and at the beginning of expiration.  
PRESSURE  
BiPAP  
IPAP  
Bi-Flex in Comparison to  
Traditional Bi-Level Therapy  
EPAP  
1
Bi-Flex  
2
3
TIME  
Ramp  
Note: IPAP will not ramp below  
4 cm H2O.  
The device is equipped with a linear ramp function. In CPAP, S, S/T, T,  
and PC modes, the Ramp feature will reduce the pressure and then  
gradually increase (ramp) the pressure to the prescription pressure  
setting so patients can fall asleep more comfortably. The figure below  
illustrates how the ramp function works.  
The Ramp Function  
TIME  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
32  
Rise Time  
In S, S/T, PC, T, PC-SIMV, and SIMV modes, rise time is the amount  
of time it takes the device to change from the expiratory pressure  
setting to the inspiratory pressure setting. Rise time levels of 1, 2,  
3, 4, 5, or 6 progressively reflect slowed response of the pressure  
increase that will take place at the beginning of inspiration. Adjust  
the rise time to find the most comfortable setting for the patient.  
PRESSURE  
Rise Time  
IPAP  
EPAP  
Rise Time  
TIME  
AVAPS Feature  
Note: AVAPS is only available if  
Average Volume Assured Pressure Support (AVAPS) is a feature  
available in the S, S/T, PC, and T modes. It helps patients maintain  
a tidal volume (VT) equal to or greater than the target tidal volume  
(Volume setting in the ventilator) by automatically controlling the  
pressure support (PS) provided to the patient. The AVAPS feature  
adjusts PS by varying the IPAP level between the minimum (IPAP Min)  
and maximum (IPAP Max) settings. AVAPS averages VT and changes  
the PS value gradually. This occurs over several minutes. The rate of  
change is slow, so that the patient is not aware of breath-to-breath  
pressure changes.  
you are using a passive circuit.  
Note: C-Flex, Bi-Flex and AVAPS  
are intended for use by adult  
patients.  
As patient effort decreases, AVAPS automatically increases PS to  
maintain the target tidal volume. The IPAP level will not rise above  
IPAP Max, even if the target tidal volume is not reached. Conversely,  
as patient effort increases, AVAPS will reduce PS. IPAP will not fall  
below IPAP Min, even if the target tidal volume is exceeded.  
If the Ramp function has been activated, it will take precedence over  
the AVAPS feature. Once ramp is complete, AVAPS will resume.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33  
PRESSURE  
IPAP Max  
IPAP Min  
EPAP  
TIME  
15  
15  
30  
30  
45  
45  
60  
VOLUME  
Target  
Volume  
AVAPS Feature  
TIME  
60  
Flow Pattern Types  
Two flow patterns are available in Volume Control ventilation therapy  
modes:  
Square  
Ramp  
Square  
With a square wave pattern, airflow is generally constant throughout  
inspiration of the breath.  
FLOW  
Square Flow Pattern  
TIME  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
34  
Ramp  
With a ramp flow pattern, the airflow starts high and decreases  
throughout inspiration of the breath.  
FLOW  
50% of  
Peak Flow  
Ramp Waveform Pattern  
TIME  
For the active circuit in volume modes, peak flow is required to be  
a minimum of 20 l/min. The wave form may be flattened when the  
combination of Inspiratory Time and Tidal Volume set points would  
result in a flow of less than 20 l/min. Therefore, for some settings,  
a Ramp flow pattern may provide a pattern that more closely  
resembles a Square flow pattern.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
35  
Sigh Feature  
The sigh feature is available for Volume ventilation modes only.  
When the sigh feature is enabled, the ventilator delivers a sigh breath  
in place of every 100th mandatory or assisted breath delivered  
regardless of the mode of operation (i.e., AC, CV, and SIMV). The Sigh  
breath is delivered using a volume equal to 150% of the set volume  
that was in effect when the breath was initiated.  
FLOW  
Example of Sigh Feature  
TIME  
Dual Prescription Feature  
The device provides a dual prescription feature that allows you  
to enter a primary prescription and a secondary prescription for  
the patient if needed. For example, you can set a primary daytime  
prescription and secondary nighttime prescription. See Chapter 5 for  
more information on the dual prescription feature.  
Note: Both prescriptions must  
use the same circuit type.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
36  
Triggering  
The device can be set to trigger breaths using the Auto-Trak or Flow  
Trigger sensitivity features.  
Digital Auto-Trak Sensitivity  
An important characteristic of the device is its ability to recognize  
and compensate for unintentional leaks in the system and to  
automatically adjust its trigger and cycle algorithms to maintain  
optimum performance in the presence of leaks. This feature is known  
as Digital Auto-Trak Sensitivity. The following sections examine this  
function in detail by describing the leak tolerance function and  
sensitivity.  
Leak Tolerance  
A microprocessor monitors the total flow of the patient circuit and  
calculates patient flow values.  
A. Leak Estimation: Average and Parabolic  
The device uses two leak estimation algorithms. A conservation  
of mass algorithm is used to compute the average leak for a given  
pressure support relationship. This average leak is used when large  
leak variations are present in the system. Average leak is a high  
estimate during EPAP pressure and a low estimate during IPAP  
pressure. A better leak estimate, enabled by the digital system, is the  
parabolic leak algorithm. Parabolic leak is proportional to the square  
of the patient pressure; therefore, the leak estimate is correlated to  
the changing patient pressure. Both algorithms include unintentional  
circuit leak and are averaged over several breaths.  
B. Patient Flow  
The total circuit flow is comprised of the circuit leak and the patient  
flow. The calculated patient flow is the total flow minus the circuit  
leak. Patient flow is a primary input into the triggering and cycling  
mechanisms.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
37  
Auto-Trak Sensitivity  
Note: Auto-Trak is only  
available if you are using a  
passive circuit.  
An essential feature of the device while operating in all modes is  
its ability to effectively sense spontaneous breathing efforts, which  
causes the ventilator to trigger to inspiration and cycle to expiration.  
Because no preset sensitivity threshold can assure patient and  
machine synchrony with changing breathing efforts and circuit  
leaks, the device continuously tracks patient breathing patterns  
and automatically adjusts sensitivity thresholds to ensure optimum  
sensitivity as breathing patterns change or as circuit leaks change.  
The algorithms used to ensure optimum sensitivity are the Volume  
Trigger, Shape Signal, Spontaneous Expiratory Threshold (SET), Flow  
Reversal, Maximum IPAP Time, and Volume Control Cycle.  
Volume Trigger (Expiration to Inspiration):  
The volume trigger is one method used to trigger inspiration during  
spontaneous breathing in all modes except T and CV. The volume  
trigger threshold is 6 ml of accumulated patient inspiratory volume.  
When patient effort generates inspiratory ow causing 6 ml of  
volume, inspiration is triggered.  
Shape Trigger/Shape Cycle (Expiration to Inspiration) (Inspiration to  
Expiration):  
The shape trigger/cycle is another method used to trigger inspiration  
and/or cycle from inspiration to expiration during spontaneous  
breathing in all modes except T and CV. This method continuously  
tracks patient inspiratory and expiratory ow and adjusts the  
spontaneous trigger and cycle thresholds for optimum sensitivity.  
The Shape Signal appears as a shadow image of the patient’s actual  
ow. The shape signal functions as a sensitivity threshold at either  
inspiration or expiration. When the patient’s ow rate crosses the  
shape signal the unit changes pressure levels. The following figure  
illustrates how the shape signal is super-imposed onto the actual  
waveform to trigger and cycle off IPAP. The shape signal is created  
by offsetting the signal from the actual patient ow by 15 l/min and  
delaying it for a 300 msec period. This intentional delay causes the  
shape signal to be slightly behind the patient’s ow rate.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
38  
A sudden change in patient ow will cross the shape signal, causing  
the pressure level to change.  
IPAP  
Pressure  
EPAP  
Cycle to  
EPAP  
Crossover  
Point  
Shape  
Signal  
Estimated  
Patient  
Flow  
Flow  
Trigger to  
IPAP  
Crossover  
Point  
Shape Signal  
Tracking the patient’s ow pattern with the Shape Signal provides a  
sensitive mechanism to trigger to inspiration or cycle to expiration in  
response to changing breathing patterns and circuit leaks.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
39  
Spontaneous Expiratory Threshold (Inspiration to Expiration):  
A second method used to cycle to expiration during spontaneous  
breathing in all modes except T, CV, AC, and SIMV, is called  
Spontaneous Expiratory Threshold (SET). The SET rises in proportion  
to the inspiratory ow rate on each breath. When the SET and actual  
patient ow value are equal, the unit cycles to expiration.  
IPAP  
Pressure  
EPAP  
Spontaneous  
Expiratory  
Threshold  
Flow  
Spontaneous Expiratory  
Threshold  
Flow Reversal (Inspiration to Expiration):  
As flow begins to decrease during inspiration, a flow reversal can  
occur due to a large leak around the mask or because the patient’s  
mouth is open. When the device senses this flow reversal, it  
automatically cycles to expiration.  
Maximum IPAP Time (Inspiration to Expiration):  
The maximum inspiratory time is determined by the adjustment of  
the Inspiratory time setting. A maximum IPAP time of 3.0 seconds  
acts as a safety mechanism to limit the time spent in inspiration  
during spontaneous breathing. For mandatory or assisted breaths,  
the maximum inspiratory time will equal the Inspiratory time setting  
up to 5.0 seconds.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
40  
Volume Control Cycle (Inspiration to Expiration) (Only available  
during Volume Control Therapy)  
An Inspiratory Time setpoint limits the time spent in inspiration  
during breathing in all modes. Once the time limit is reached, the unit  
automatically cycles to expiration.  
Flow Trigger  
Flow trigger provides a manual setting that allows for breath  
initiation and termination based on a set flow trigger sensitivity and  
flow cycle sensitivity.  
Flow Trigger Sensitivity (Expiration to Inspiration):  
The flow trigger initiates when the patient’s inspiratory effort creates a flow  
equal to or greater than the flow trigger sensitivity setting. The method of  
the flow trigger is dependent upon the circuit type that is chosen.  
Leak Compensation:  
When using the Passive Circuit configuration, compensation for both the  
intentional and unintentional leak is included in the triggering method.  
When using the Active PAP Circuit configuration, leak compensation  
is not available.  
When using the Active Flow Circuit configuration, flow trigger with  
leak compensation may be enabled. The default setting when using  
the Active Flow Circuit is Leak Compensation On. The clinician has  
the option to turn off leak compensation; however, unintentional  
leak will not be compensated. Both options measure the flow at the  
proximal flow sensor.  
Note: Enabling Leak  
Compensation when using the  
Active Flow Circuit configuration  
only affects triggering and does  
not affect tidal volume delivery  
or Vte measurement.  
Flow Cycle Sensitivity (Inspiration to Expiration):  
This cycling method is only active if the Flow Trigger has been  
selected for the Trigger Type. As flow begins to decrease during  
inspiration, if the patient flow is less than the flow cycle sensitivity  
setting, the device will cycle to expiration.  
For example: if the flow cycle sensitivity setting is set to 75%, when  
the flow has decreased by 25% of the peak flow, the device will cycle  
to the EPAP/PEEP level.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
41  
BTPS Compensation  
All flows and volumes used in Trilogy are expressed in BTPS - Body  
Temperature atmospheric Pressure Saturated with H2O.  
All pressures are expressed relative to atmospheric pressure.  
Ventilator Alarms  
This section describes all of the ventilator alarms and informational  
messages in order of priority, from high priority alarms to low priority  
alarms and finally informational messages. Refer to Chapter 6 for  
more information on alarms.  
Loss of Power Alarm  
This is a high priority alarm. It may occur when a complete power  
failure has occurred and power was lost while the device was  
providing therapy. This may happen if the internal battery was the  
only power source in use and was completely depleted.  
Ventilator Inoperative Alarm  
This is a high priority alarm. It occurs when the ventilator detects an  
internal error or a condition that may affect therapy. The device will  
shut down if the cause of the failure indicates that the device cannot  
deliver therapy safely. If the device can deliver therapy at a limited  
level, the device will continue to deliver limited therapy.  
Ventilator Service Required Alarm  
This is a high priority alarm. It occurs when the device cannot  
perform to specification, a backup safety feature is compromised,  
or the delivery of therapy is compromised. The device continues to  
operate (possibly in a reduced capacity mode). If the problem is not  
corrected, the device will generate a reminder message once per  
hour until the issue is corrected. Additionally, if therapy is stopped, a  
reminder message will immediately appear when therapy is turned  
on again.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
42  
Check Circuit Alarm  
This is a high priority alarm. It occurs when the device detects a  
problem with the patient circuit, such as pinched or detached tubing,  
water condensation in the proximal pressure lines, or problems with  
the active exhalation device.  
Low Circuit Leak Alarm  
This is a high priority alarm that only occurs with the passive circuit.  
It occurs when the system detects a problem with the leak device in  
the passive circuit.  
High Expiratory Pressure Alarm  
This is a high priority alarm. It occurs when the delivered pressure  
exceeds the target patient pressure during the expiratory phase by  
5 cmH2O. This may be due to pinched tubing or the patient having  
a fast breath rate. The device continues to operate. The alarm will  
automatically terminate when the delivered pressure comes within  
5 cmH2O of the target patient pressure during the expiratory phase.  
Low Expiratory Pressure Alarm  
This is a high priority alarm. It occurs when the delivered pressure  
is 5 cmH2O or more below the target patient pressure during the  
expiratory phase. The device continues to operate. The alarm will  
automatically terminate when the delivered pressure comes within  
5 cmH2O of the target patient pressure during the expiratory phase.  
High Internal Oxygen Alarm  
This is a high priority alarm. It occurs when there is a leak in the  
internal air delivery system that allows oxygen to build up inside  
the device. The alarm is generated when the internal oxygen  
concentration reaches 5% above ambient levels.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
43  
Circuit Disconnect Alarm  
WARNING  
This is a high priority alarm. It occurs when the breathing circuit  
is disconnected or has a large leak. The device continues to  
operate. The alarm will automatically terminate when the circuit is  
reconnected or the leak is fixed.  
You should not rely on any  
single alarm to detect a circuit  
disconnect condition. The Low  
Tidal Volume, Low Minute  
Ventilation, Low Respiratory  
Rate, and Apnea alarms should  
be used in conjunction with  
the Circuit Disconnect alarm.  
The Apnea alarm is only  
Apnea Alarm  
This is a high priority alarm. It occurs when the patient has not  
triggered a breath within the time specified in the apnea alarm  
intended for spontaneously  
setting. The device continues to operate. The alarm will automatically breathing patients.  
terminate when two consecutive patient breaths are detected that  
meet the apnea alarm time setting.  
High Vte Alarm  
This is a high priority alarm. It occurs when the estimated exhaled  
tidal volume is greater than the High Vte alarm setting for three  
consecutive breaths. The device continues to operate. The alarm will  
automatically terminate when a breath occurs in which the exhaled  
tidal volume does not reach the High Vte alarm setting.  
Low Vte Alarm  
This is a high priority alarm. It occurs when the estimated exhaled  
tidal volume is lower than the Low Vte alarm setting for three  
consecutive breaths. The device continues to operate. The alarm will  
automatically terminate when a breath occurs in which the exhaled  
tidal volume exceeds the Low Vte alarm setting.  
When AVAPS is on, this alarm will occur when the calculated tidal  
volume is less than 90% of the target tidal volume setting. The alarm  
will automatically terminate when a breath occurs in which the  
exhaled tidal volume is equal or greater to 90% of the target tidal  
volume setting.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
44  
High Vti Alarm  
This is a high priority alarm. It occurs when the delivered tidal  
volume is greater than the High Vti alarm setting for three  
consecutive breaths. The device continues to operate. The alarm will  
automatically terminate when a breath occurs in which the delivered  
tidal volume does not reach the High Vti alarm setting.  
Low Vti Alarm  
This is a high priority alarm. It occurs when the delivered tidal volume  
is less than the Low Vti alarm setting for three consecutive breaths.  
The device continues to operate. The alarm will automatically  
terminate when a breath occurs in which the delivered tidal volume  
exceeds the Low Vti alarm setting.  
High Respiratory Rate Alarm  
This is a high priority alarm. It occurs when the respiratory rate is greater  
than the High Respiratory Rate alarm setting. The device continues to  
operate. The alarm will automatically terminate when the measured  
respiratory rate is less than the High Respiratory Rate alarm setting.  
Low Respiratory Rate Alarm  
This is a high priority alarm. It occurs when the respiratory rate is less  
than the Low Respiratory Rate alarm setting. The device continues to  
operate. The alarm will automatically terminate when the measured  
respiratory rate is greater than the Low Respiratory Rate alarm setting.  
High Inspiratory Pressure Alarm  
This alarm occurs in several stages and escalates from an audible beep for  
the first two occurrences to a high priority alarm if the problem continues.  
It is detected differently for volume and pressure therapy modes.  
For volume modes, the alarm will sound if the measured patient  
pressure exceeds the High Inspiratory Pressure setting specified by  
the clinician. The alarm will automatically terminate when the peak  
inspiratory pressure is less than or equal to the High Inspiratory  
Pressure alarm setting.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
45  
For pressure modes, the alarm occurs when the delivered pressure  
exceeds the target patient pressure by 5 cmH2O or more during  
the inspiratory phase. The device will automatically cycle to  
the expiratory phase and continue to operate. The alarm will  
automatically terminate when the delivered pressure falls within 5  
cmH2O of the target patient pressure during the inspiratory phase.  
Low Inspiratory Pressure Alarm  
This is a high priority alarm. It is detected differently for volume and  
pressure therapy modes.  
For volume modes, the alarm will sound if the measured patient  
pressure is less than the Low Inspiratory Pressure setting specified by  
the clinician. The alarm will automatically terminate when the peak  
pressure at the end of the breath is greater than or equal to the Low  
Inspiratory Pressure alarm setting.  
For pressure modes, the alarm occurs when the delivered pressure  
is 5 cmH2O or more below the target patient pressure during the  
inspiratory phase. The alarm will automatically terminate when  
the delivered pressure comes within 5 cmH2O of the target patient  
pressure during the expiratory phase.  
High Minute Ventilation Alarm  
This alarm is a high priority alarm. It occurs when the patient’s  
minute ventilation is greater than the High Minute Ventilation alarm  
setting. The device continues to operate. The alarm will automatically  
terminate when the calculated minute ventilation is less than the  
High Minute Ventilation alarm setting.  
Low Minute Ventilation Alarm  
This alarm is a high priority alarm. It when the patient’s minute  
ventilation is less than the Low Minute Ventilation alarm setting. The  
device continues to operate. The alarm will automatically terminate  
when the calculated minute ventilation is greater than the Low  
Minute Ventilation alarm setting.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
46  
Low Battery Alarm  
WARNING  
Immediately seek an alternate  
power source when the “Low  
Battery” message appears.  
Complete power failure and loss  
of therapy is imminent.  
The Low Battery alarm occurs when the last battery available is  
low or nearly depleted. This alarm occurs in two stages. When  
approximately 20 minutes of battery run time remains, a medium  
priority alarm is generated, and the device continues to operate. If  
no action is taken and the battery continues to deplete, the alarm  
escalates to a high priority alarm when approximately 10 minutes of  
battery run time remains.  
WARNING  
The ventilator has a two-stage  
low battery alarm. The medium  
priority alarm indicates that  
approximately 20 minutes of  
operation remain, and the high  
priority alarm indicates that less  
than 10 minutes of operation  
remain. Actual run time may be  
more or less than this and varies  
with battery age, environmental  
conditions, and therapy.  
High Temperature Alarm  
This alarm occurs when the estimated patient airstream temperature  
or the ventilator internal temperature is too high. The alarm occurs in  
several stages. The ventilator continues to operate. Internal fans are  
started when the medium priority alarm is generated. If the condition  
causing the high temperature is not corrected and the temperature  
continues to rise, the alarm will escalate to the high priority alarm.  
Replace Detachable Battery Alarm  
The Replace Detachable Battery alarm occurs when the detachable  
battery is nearing the end of its useful life or a failure in the  
detachable battery that prevents it from charging or discharging has  
been detected.  
The alarm occurs in several stages, from low to high priority. The  
device continues to operate when the alarm is the low priority  
alarm. If the alarm is reset without removing the battery, the alarm  
will be regenerated once every hour until the detachable battery is  
removed. The device continues to operate, the detachable battery  
is not used, and the power source is switched to the next available  
power source if the alarm is the high priority alarm.  
Ventilator Service Recommended Alarm  
This is a medium priority alarm. It occurs when the device has  
detected an error, but the error will not affect device performance  
or safety. The device continues to operate. Therapy and safety are not  
compromised.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
47  
If the problem is not corrected, the device will generate a reminder  
message once per day, or whenever power is cycled, until the issue  
is corrected. Additionally, if the device is powered off, a reminder  
message will immediately appear when the device is turned on again.  
AC Power Disconnected Alarm  
This is a medium priority alarm. It occurs when the AC power source  
was lost, and the device has switched to an alternate power source  
(either a detachable or external battery, if connected, or the internal  
battery if no other source is available). The device continues to  
operate. If AC power returns, the ventilator will beep, but no message  
will appear on the display.  
Keypad Stuck Alarm  
This is a low priority alarm. It occurs when a key becomes lodged  
inside the case of the device.  
Battery Discharging Stopped due to Temperature  
Info Message  
This info message occurs when the detachable or internal battery  
becomes overheated while providing power for the device. The  
device continues to operate. The battery is not used and the power  
source is switched to the next available power source.  
Battery Not Charging due to Temperature Info Message  
This info message occurs when the detachable or internal battery  
becomes too hot while charging or the device was in too cold an  
environment before charging started. The device continues to  
operate. Battery charging stops until the battery cools or warms  
sufficiently.  
Battery Not Charging Info Message  
This info message occurs when the device has detected an error  
condition with the battery that prevents it from accepting a charge.  
The device continues to operate. Battery charging stops.  
Chapter 3 Modes, Features, and Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
48  
Check External Battery Info Message  
This info message occurs when a bad connection exists to the  
external battery or the external battery failed. The device continues  
to operate using power from the detachable battery, if available, or  
the internal battery.  
Battery Depleted Info Message  
This info message occurs when the affected battery is fully depleted.  
The device continues to operate using the next available power source.  
External Battery Disconnected Info Message  
This info message occurs when the external battery power source is  
lost and the device has switched to an alternate power source (either  
a detachable battery, if connected, or the internal battery if no other  
source is available). If external battery power returns, the ventilator  
will beep, but no message will appear on the display.  
Detachable Battery Disconnected Info Message  
This info message occurs when the detachable battery power source  
is lost and the device has switched to an alternate power source (the  
internal battery if no other source is available). If detachable battery  
power returns, the ventilator will beep, but no message will appear  
on the display.  
Start On Battery Info Message  
This info message indicates that the ventilator has started on battery  
power and no AC power is available. The device operator should  
verify that this is what is wanted.  
Card Error Info Message  
This info message occurs when an unusable SD card is inserted into  
the ventilator. The device continues to operate but data cannot be  
logged onto the SD card.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
4. Ventilator Setup  
This chapter provides instructions on how to assemble the ventilator.  
It includes the following setup information:  
WARNING  
Ventilator dependent patients  
should be continuously  
monitored by qualified  
personnel. These personnel  
should be prepared to provide  
alternate therapy in the event of  
ventilator failure or inoperative  
equipment.  
Properly positioning the device  
Installing the air filter  
Supplying power to the device  
Connecting the breathing circuit  
Connecting a water trap (optional)  
Connecting a remote alarm (optional)  
WARNING  
Do not use the ventilator on a  
patient until a system checkout  
has been performed. See  
Chapter 10 of this manual.  
WARNING  
For ventilator dependent  
patients, always have alternate  
ventilation equipment, such as  
a back-up ventilator, manual  
resuscitator, or similar device,  
available.  
CAUTION  
Do not operate the ventilator at  
temperatures below 5° C (41° F)  
or above 40° C (104° F).  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
50  
Position the Device  
Place the ventilator on a flat, level surface. Or, as an alternative, you  
can place the device in the optional in-use bag. The in-use bag is  
required for portable use in order to properly secure the ventilator.  
Do not operate the ventilator while on its side, upside-down, or in  
any other orientation. Make sure that the air inlet on the back of the  
device is not blocked. If you block the air flow around the device, the  
ventilator may not work properly.  
CAUTION  
Install the Air Filter  
The reusable foam inlet filter is  
required to protect the ventilator  
from dirt and dust. Wash  
periodically and replace when  
damaged for proper operation.  
The device uses a gray foam lter that is washable and reusable. The  
reusable lter screens out normal household dust and pollens. The  
lter must be in place at all times when the device is operating. Two  
reusable gray foam lters are supplied with the device. If the lter is  
not already installed when you receive the device, you must install  
the lter before using the ventilator.  
To install the lter, insert the gray foam lter into the lter area as  
shown.  
Supply Power to the Device  
Installing the Filter  
The device can operate on AC or DC power. The ventilator accesses  
power from potential sources in the following order:  
Note: See Chapter 7 for  
information on how to clean  
and replace the air filter.  
AC Power  
External Battery  
Detachable Battery Pack  
Internal Battery  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
51  
Note: This device is activated  
(i.e., ‘powered on’) when the AC  
power cord is connected or any  
of the three battery sources is  
available. Pressing the Start/Stop  
button turns the airow on or off.  
Using AC Power  
An AC power cord is provided with the device.  
1. Plug the socket end of the AC power cord into the AC power inlet  
on the side of the device. To prevent accidental removal of the  
AC power cord from the device, you may secure the AC power  
cord to the device using the AC power cord retainer located on  
the back of the device.  
CAUTION  
The device may only be operated  
at temperatures between 5˚ C  
and 40˚ C (41˚ F and 104˚ F).  
2. Plug the pronged end of the cord into a wall outlet not  
connected to a wall switch.  
3. Ensure that all connections are secure. If AC power is connected  
correctly and the device is operating properly, the green AC  
Power LED should be on.  
WARNING  
Periodically inspect the power  
cord for damage or signs of  
wear. Discontinue use and  
replace if damaged.  
Note: To remove AC power,  
disconnect the power supply  
cord from the electrical outlet.  
Connecting the AC Power  
Cord  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
52  
4. If desired, secure the power cord using the cord retainer on the  
back of the device, as shown below. To secure the power cord,  
use a screwdriver to remove the screw on the cord retainer.  
Insert the cord as shown into the cord retainer, and then re-  
attach the cord retainer to the device by re-attaching the screw.  
Securing the Power Cord  
Using the Cord Retainer  
Using DC Power  
You can operate the ventilator using an external battery, detachable  
battery, or the internal battery.  
CAUTION  
External Battery  
Do not use the same external  
battery to operate both the  
ventilator and any other equipment  
such as power chairs.  
The ventilator can operate from a 12 VDC deep cycle marine-type  
(lead acid) battery using the Philips Respironics Trilogy External  
Battery Cable. This cable is pre-wired and properly terminated to  
ensure safe connection of an external battery to the ventilator.  
Battery operating time depends on the characteristics of the battery  
and usage of the device.  
CAUTION  
An external battery should only  
be connected to the ventilator  
using the Philips Respironics  
Trilogy External Battery Cable.  
This cable is fused, pre-wired and  
properly terminated to ensure  
safe connection to a standard  
deep cycle lead acid battery. Use  
of any other adapter or cable  
may cause improper operation  
of the ventilator.  
Due to a variety of factors, including battery chemistry, battery age,  
and use profile, the capacity of the external battery as shown on the  
device display is only an estimate of the actual remaining capacity.  
Refer to the instructions supplied with the Philips Respironics Trilogy  
External Battery Cable for detailed information on how to operate  
the device using an external battery.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
53  
Detachable Battery Pack  
Philips Respironics offers a detachable Lithium-Ion battery pack. To  
use the detachable battery pack, snap the battery into place on the  
back of the ventilator, as shown in the following illustration. When  
the device is not connected to an AC power source or an external  
battery, the detachable battery will power the device, if attached.  
The length of time the ventilator will operate on battery power  
depends on many factors such as device settings, battery charge  
level, and condition or age of the battery. When fully charged, a new  
battery can power the ventilator for approximately three hours under  
typical patient conditions.  
CAUTION  
The detachable and internal  
batteries wear out based on  
the amount of use (hours or full  
charge-discharge cycles). The  
battery capacity and life are also  
reduced by operation at higher  
temperatures.  
Whenever the ventilator is connected to AC power, it will  
automatically recharge the detachable battery pack. A completely  
discharged detachable battery will reach 80% charge status within 8  
hours, when charging at approximately 23˚ C ambient temperature.  
CAUTION  
Only use the Philips Respironics  
Trilogy Detachable Battery with  
the device.  
Insert and securely latch the detachable battery into the device as  
shown below.  
Attaching the Detachable  
Battery  
Insert  
Properly Installed  
Detachable  
Detachable  
Battery  
Battery  
CAUTION  
Prolonged operation or storage  
at elevated temperatures may  
reduce the service life of the  
detachable or internal battery  
and other internal components  
of the ventilator.  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
54  
One side of the detachable battery has a set of LEDs that indicate the  
amount of charge left on the battery. You can press the button below  
the LEDs to view how much charge remains:  
LED  
All 5 LEDs are lit  
Battery Capacity  
80-100% capacity  
4 LEDs are lit  
3 LEDs are lit  
2 LEDs are lit  
1 LED is lit  
60-79% capacity  
40-59% capacity  
20-39% capacity  
10-19% capacity  
1 to 9% capacity  
0% capacity  
1 LED flashes  
0 LEDs are lit  
Detachable Battery LEDs  
WARNING  
Internal Battery  
The device contains an internal battery that can be used as a back-  
up power source. It is intended for use during short periods while  
switching between external power sources, emergency situations, or  
short durations when the user needs to be mobile. The length of time  
the ventilator will operate on internal power depends on many factors  
such as device settings, battery charge level, and condition or age of the  
battery. When fully charged, a new battery can power the ventilator for  
approximately three hours under typical patient conditions.  
The internal battery is NOT  
intended to serve as a primary  
power source. It should only be  
used when other sources are  
not available, or briefly when  
necessary; for example when  
changing power sources.  
Whenever the ventilator is connected to AC power, it will  
automatically recharge the internal battery. A completely discharged  
internal battery will reach 80% charge status within 8 hours when  
charging at approximately 23˚ C ambient temperature.  
Device Power Source Indicators  
There are many power source indicators on the device and the  
display screen. These indicators are described in detail below.  
AC Power Indicator  
When AC power is applied to the device, the green AC LED indicator  
(~) on the front of the device lights.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
55  
DC Power Indicators  
Note: Under normal operating  
conditions, the internal battery  
symbol will always display on  
the Monitoring screen. If the  
symbol appears as an empty  
red battery on your screen,  
contact Philips Respironics  
or an authorized service  
The internal, detachable, and external battery symbols that will  
display on the Monitoring screen are shown below. The detachable  
and external battery symbols will only appear on-screen if a  
detachable or external battery is attached to the device.  
Battery  
Symbol  
representative to have your  
Internal Battery  
device serviced.  
Detachable Battery  
External Battery  
There are several DC power indicators that will display on the  
Monitoring screen to indicate which battery is in use (if applicable), if  
the batteries are low, charging, or discharged, etc. The following table  
explains all of the DC power indicators.  
DC Power Indicator  
Description  
Battery In Use Indicator A black box will appear around the battery that is in use. For instance, if  
the external battery is currently in use, the  
Monitoring screen.  
symbol appears on the  
Green Fully Charged  
Battery Indicator  
When a battery is charged to greater than 90% of its capacity, all of the  
bars in the battery symbol will appear in green.  
Partially Charged  
Battery Indicator  
When a battery is partially charged, some of the bars in the battery  
symbol will appear in green, while others will be clear. For instance, if  
the external battery is 50% charged, the following symbol displays on-  
screen:  
Yellow Low Battery  
Indicator (Medium  
Priority)  
When the device detects that an in-use battery’s charge is low (has  
approximately 20 minutes of charge left), the inside of the box  
surrounding the battery symbol turns yellow. (In addition to the  
battery indicator on the Monitoring screen, a medium priority alarm  
message will display indicating “Low Battery.See Chapter 6 for more  
information. The yellow indicator is for the last available battery source.  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
56  
DC Power Indicator  
Description  
Red Low Battery  
Indicator  
When the device detects that an in-use battery’s charge is nearly  
depleted (has approximately 10 minutes of charge left), the inside  
of the box surrounding the battery symbol turns red. In addition to  
the battery indicator on the Monitoring screen, a high priority alarm  
message will display indicating “Low Battery.See Chapter 6 for more  
information. The red indicator is for the last available battery source.  
Yellow Battery  
Whenever AC power is applied to the device, the internal and  
Recharging Symbol (  
)
detachable batteries will recharge as needed. If the internal battery is  
being recharged, the  
being recharged, the  
symbol displays. If the detachable battery is  
symbol displays.  
Battery Disposal  
Dispose of the batteries in accordance with local regulations.  
First Time Use  
When setting up the device for the first time or after annual service,  
apply AC power to the ventilator before turning on the blower.  
Attempting to use the ventilator without first applying AC power,  
such as installing a detachable battery pack and starting the blower,  
will cause the internal battery to be displayed in red as an empty  
battery. When in this state, the internal battery will not be usable  
until AC power is applied.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
57  
Connect the Breathing Circuit to the Ventilator  
WARNING  
Philips Respironics recommends  
that a main line outlet bacteria  
filter (Part Number 342077) be  
used whenever the device is  
used for invasive therapy or if  
the ventilator may be used on  
multiple patients.  
Complete the following steps to set up your patient circuit.  
1. If you are using a bacteria filter, connect one end of the flexible  
tubing to the outlet of the bacteria filter, and connect the  
bacteria filter inlet to the breathing circuit connection located on  
the side of the ventilator.  
If you are not using a bacteria filter, connect the tubing directly  
to the device’s breathing circuit connection.  
Trilogy Breathing Circuit  
Connection  
Flexible Tubing  
Connecting the Breathing  
Circuit to the Device  
Bacteria Filter  
2. Connect the other end of the flexible tubing to a separate  
exhalation device.  
Note: The device can be used  
with reusable or disposable  
circuits. For detailed instructions  
on how to set up your device  
using a disposable circuit, refer  
to the instructions included with  
the disposable circuit.  
a. If your circuit type is Passive and you are using a Philips  
Respironics Passive Exhalation Device:  
1.Connect the flexible tubing to the rotating end of the  
Passive Exhalation Device.  
2.Connect the other end of the Passive Exhalation Device to  
the patient interface (e.g., the mask).  
Passive  
Exhalation  
Device  
Flexible  
Tubing  
Connecting the Passive  
Exhalation Device  
Flexible  
Bacteria Filter  
Tubing  
Whisper  
Swivel II  
Bacteria Filter  
Note: The device does not  
provide circuit compliance  
compensation in volume modes.  
Refer to the instructions included with the Passive  
Exhalation Device for more detailed setup information.  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
58  
b. If your circuit type is Active PAP:  
Active Exhalation Device  
with Proximal Pressure  
Note: Passive circuit ventilation  
provides leak compensation  
in both volume and pressure  
modes. Active PAP circuit  
ventilation does not provide  
leak compensation. Particular  
care should be used at low tidal  
volumes to ensure adequate  
monitoring of exhaled tidal  
volume.  
1. Connect the flexible tubing to an active exhalation device  
with proximal pressure.  
A. Connect the active exhalation device with proximal  
pressure to the flexible tubing that attaches to the  
ventilator and to the patient interface (e.g., the  
tracheostomy tube).  
2. Connect the proximal pressure line and the exhalation  
valve line to the active exhalation device with proximal  
pressure and the Universal Porting Block on the device as  
described below.  
WARNING  
For ventilator dependent  
patients, always have alternate  
ventilation equipment, such as  
a back-up ventilator, manual  
resuscitator, or similar device,  
available.  
A. Attach the proximal pressure line (.476 cm or 3/16”  
line) to the proximal pressure port as shown in the  
next illustration.  
B. If not attached, connect the other end of the proximal  
pressure line to the Trilogy Universal Active PAP Tube  
Adapter as shown in the next illustration.  
Note: Make sure that  
components marked with an  
arrow are oriented properly.  
When attaching an active  
exhalation device with proximal  
pressure to the patient, make  
sure the proximal pressure port  
faces away from the patient.  
C. Connect the Trilogy Universal Active PAP Tube Adapter  
to the top, striped port on the Universal Porting Block  
on the ventilator.  
D. Connect the exhalation valve line (.317 cm or 1/8line)  
to the exhalation valve port on the top of the active  
exhalation device with proximal pressure.  
Note: The  
symbol  
appears next to the Exhalation  
Valve port on the Universal  
Porting Block to indicate where  
the active exhalation valve line  
connects.  
E. Connect the other end of the exhalation valve line  
to the exhalation valve port on the Universal Porting  
Block.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
59  
Universal Porting Block  
Trilogy Universal Active  
Proximal  
Pressure  
Line  
Exhalation Valve Port  
PAP Tube Adapter  
Proximal  
Pressure  
Port  
Exhalation Valve Line  
Exhalation Valve Port  
on Universal Porting  
Block  
Connecting an Active Device  
with Proximal Pressure  
Refer to the instructions included with the Active Exhalation  
Device for more information.  
c. If your circuit type is Active Flow:  
1. Connect the Flow Sensor to an active exhalation device as  
shown below.  
Connecting the Flow Sensor  
Flow Sensor  
Note: If the active exhalation  
device has the proximal pressure  
line connected when you are  
setting up an active flow circuit  
type, remove the proximal  
pressure line and cap the  
proximal pressure port before  
attaching to the flexible tubing.  
2. Connect the other end of the flexible tubing to the active  
exhalation device.  
A. Connect the active exhalation device to the flexible  
tubing that attaches to the ventilator.  
B. Connect the flow sensor to the patient interface (e.g.,  
the tracheostomy tube).  
Note: Make sure that  
components marked with an  
arrow are oriented properly.  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
60  
C. Connect the flow lines and the exhalation valve line to  
the Universal Porting Block on the ventilator as shown  
below.  
Connect the White Striped Flow Line to  
the top, striped port on the Universal  
Porting Block on the device.  
Note: When using an active  
exhalation device with proximal  
flow, ensure that any additional  
proximal ports are capped.  
Connect the other Flow Line to the  
middle port on the Universal porting  
Block.  
Connect the Exhalation Valve Line to the  
Exhalation Valve port on the top of the  
active exhalation device, and connect the  
other end of the line to the Exhalation  
Valve port on the Universal Porting Block.  
Note: The  
symbol  
appears next to the Exhalation  
Valve port on the Universal  
Porting Block to indicate where  
the active exhalation valve line  
connects.  
Universal Porting Block  
White Striped  
Exhalation Valve Port  
on Trilogy Device  
Flow Line  
Connecting the Active  
Exhalation Device with  
Proximal Flow Sensor  
Flow Line  
Exhalation Valve Line  
Exhalation Valve Port  
on Porting Block  
Refer to the instructions included with the active exhalation  
device for more information.  
Connect a Water Trap  
If you are using an optional water trap, connect it to the patient  
circuit according to the manufacturer’s instructions.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
61  
Connect Supplemental Oxygen (Optional)  
To attach supplemental oxygen to the device:  
1. Connect the oxygen tubing to the Oxygen Inlet Quick Connect.  
Then, connect the Quick Connect to the Oxygen Inlet on the  
back of the ventilator.  
Oxygen Inlet  
Oxygen Inlet  
Quick Connect  
Oxygen Source  
Tubing  
Attaching the O2 Tubing  
2. Make sure you press the Oxygen Inlet Quick Connect firmly into  
the inlet. You should hear a click, and the latch at the top of the  
inlet will pop up.  
Press Tubing in and Latch  
Pops Up  
This device is only compatible with a low flow oxygen source  
providing up to 15 l/min.  
CAUTION  
Do not connect an unregulated  
or high pressure oxygen source  
to this connector on the device.  
Chapter 4 Ventilator Setup  
Download from Www.Somanuals.com. All Manuals Search And Download.  
62  
Connect the Remote Alarm (Optional)  
WARNING  
When using a remote alarm,  
make sure you fully test the  
remote alarm connector and  
cable by verifying that:  
You can use a remote alarm as a nurse call system or in-house remote  
alarm system. It can generate an alarm at a distance of up to  
91 meters (300 feet) from the ventilator. An audible tone sounds  
and a red light blinks to indicate that an alarm condition exists. The  
remote alarm will sound when any of the following conditions exist:  
-
Annunciated alarms on  
the ventilator are also  
annunciated on the remote  
alarm.  
The ventilator is off.  
Any alarm occurs and is not silenced or reset.  
-
Disconnecting the remote  
alarm cable from the  
ventilator or from the remote  
alarm results in an alarm  
notification at the remote  
alarm.  
Refer to the Accessories chapter of this manual for information on  
which remote alarm systems are compatible with the device. Refer to  
the instructions included with the Remote Alarm Adapter Cable for  
information on how to connect a remote alarm to the ventilator.  
WARNING  
The remote alarm should be  
tested daily.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
5. Viewing and Changing Settings  
Note: There is a keypad lock  
inactivity time-out period. After  
you have unlocked the keypad  
as indicated, the keypad will  
re-lock after five minutes of  
inactivity to prevent someone  
from accidentally pressing a  
button and changing any of the  
settings.  
This chapter explains how to scroll through the ventilation screens  
and change ventilation settings. It also describes how to connect the  
ventilator to the patient once the settings are complete.  
Keypad Lock Feature  
There is a Keypad Lock feature that users can enable from the  
Options menu. It is intended to prevent accidental changes to device  
settings. This feature will lock the navigation keys (Up, Down, Stop,  
Left, and Right keys). If the Keypad Lock is enabled, a Keypad Unlock  
message will display on the bottom of the screen any time you press  
one of the navigation keys.  
Note: When Keypad Lock is  
enabled, the Left, Right, and Up/  
Down buttons are locked while  
the ventilator is turned on. The  
Alarm Indicator/Audio Pause  
continues to function normally.  
The Start/Stop button is locked  
only when this button is used to  
stop therapy.  
If the keypad is locked, you must unlock it before you can enter the  
Menu. To unlock the keypad and enter the menu, you must first hold  
the Right button for 5 seconds to unlock the keypad. An audible  
indicator sounds when the keypad is successfully unlocked. Once the  
display is unlocked, you can enter the Menu as you normally would  
by pressing the Up button.  
Note: The keypad will  
automatically unlock if an  
alarm or informational message  
occurs and will remain unlocked  
the entire time alarms are active.  
Note: Pressing the Left (Cancel)  
button will cancel the Keypad  
Unlock action.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
64  
Accessing the Startup and Monitor Screens  
1. After you press the  
button to begin therapy, the Startup  
screen appears momentarily, indicating the device name and the  
software version.  
2. The next screen that appears is the Monitor screen.  
The appearance of this screen will vary, depending on how you  
set up the device. If Detailed View is turned off in the Options  
menu, your screen will look like the screen shown below.  
Note: The symbols that appear  
on the Monitor screen are  
described in detail later in this  
chapter.  
Monitor Panel  
Date and Time  
Panel  
Status Panel  
Monitor Screen – Detailed  
View Off  
The top section of the screen, called the Monitor panel,  
shows the therapy mode and, if you set up a dual  
prescription for the patient, the Prescription indicator  
appears, indicating Primary or Secondary prescription. The  
patient breath symbol also displays during a patient-triggered  
breath, and a bar graph displays the current pressure level.  
Note: No Soft Buttons display  
on the Monitor screen if Keypad  
Lock is enabled.  
The center section of the screen displays the current date  
and time.  
The bottom section, called the Status panel, displays certain  
symbols that indicate features being used, such as Ramp, as  
well as battery status.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
65  
If Detailed View is turned on in the Options menu, the Monitor  
screen will look like the screen shown below.  
Monitor Panel  
Measured Settings  
Panel  
Monitor Screen – Detailed  
View On  
Status Panel  
This screen contains more detailed information about the  
therapy.  
The top Monitor panel contains the Prescription indicator  
if a dual prescription exists, the therapy mode, a graph  
displaying the current pressure, and the current date and  
time. Additionally, this panel also displays patient pressure,  
respiratory rate (RR), exhaled tidal volume (Vte), and leak.  
The second panel in Detailed view is the Measured  
Settings panel. It provides patient-related data including  
Peak Inspiratory Pressure (PIP), Minute Ventilation, Peak  
Inspiratory Flow, Mean Airway Pressure (MAP), and I:E Ratio.  
The third panel is the Status panel and shows the same  
information displayed in the Detailed View Off screen,  
including features in use such as Ramp and battery status.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
66  
Note: Dashes on the display  
screen indicate that the device  
was unable to compute the  
parameters. For example, when  
the device is first connected to a  
patient, the tidal volume, minute  
ventilation, and leak may be  
dashed until the device is able  
to accurately calculate these  
patient parameters.  
Monitor Screen Indicators  
This section describes the following indicators:  
Monitor Panel Indicators  
Measured Settings Panel Indicators  
Status Panel Indicators  
Monitor Panel Indicators  
All of the indicators that may appear on the Monitor Panel are  
described in detail in the following table.  
Indicator  
Prescription  
Description  
If you set up a dual prescription for the patient, the words  
“Primaryor “Secondaryappear in the top left corner of the panel  
to indicate which prescription is active.  
Therapy Mode  
The current therapy mode displays at the top of the panel (for  
example, CPAP, S, S/T, etc.). If a special feature such as Flex, AVAPS,  
or Sigh is active, this feature will appear next to the therapy  
mode.  
Date and Time  
Patient Breath  
If you are in Detailed view, the current date and time appear  
in the top right corner of the panel. (In Detailed View Off, they  
appear in the center panel.)  
This symbol displays during a patient-triggered breath.  
Airway Pressure  
Manometer and  
Peak Pressure  
Symbol  
The manometer (bar graph) displays the airway pressure in the  
patient circuit at all times. The manometer bar moves to the  
right as pressure increases during inhalation, and moves to the  
left as pressure decreases during exhalation. The peak pressure  
is also indicated on this bar. It is positioned according to the  
maximum patient pressure reached during each breath. The Peak  
Pressure symbol appears as a blue bar on the manometer. If a  
High Inspiratory Pressure alarm occurs, the Peak Pressure symbol  
changes from blue to red.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
67  
Indicator  
Description  
Low Pressure  
If you enable a volume therapy mode, this indicator appears  
below the manometer bar, indicating the low pressure alarm  
setting.  
Indicator  
L
High Pressure  
If you enable a volume therapy mode, this indicator appears  
below the manometer bar, indicating the high pressure alarm  
setting.  
Indicator  
H
Pressure  
This indicator displays the current patient pressure. This only  
appears in detailed view.  
Respiratory Rate (RR)  
This indicator displays the measured respiratory rate in Breaths  
Per Minute (BPM). This only appears in detailed view.  
Exhaled Tidal  
Volume (Vte)  
This indicator displays the estimated exhaled tidal volume in  
milliliters and reflects compensation for BTPS. This only appears in  
detailed view when Passive Circuit is selected.  
Inhaled Tidal  
Volume (Vti)  
This indicator displays delivered tidal volume in milliliters and  
reflects compensation for BTPS. This only appears in detailed view  
when the Active with PAP Circuit is selected.  
Leak  
This indicator displays the total leak (non-returned flow), between  
the unit outlet and the patient, averaged over the previous  
breath. This only appears in detailed view when the Passive  
Circuit is selected.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
68  
Measured Settings Panel  
All of the indicators that may appear on the Measured Settings panel  
(available only in Detailed view), are described in the following table.  
Indicator  
Description  
PIP  
Peak Inspiratory Pressure displays the maximum pressure  
delivered to the patient during the previous breath.  
I:E Ratio  
Peak Flow  
MAP  
Displays a comparison of the time spent in inspiration to the time  
spent in expiration during the previous breath.  
Displays the maximum inspiratory flow delivered to the patient  
during the previous breath in l/min BTPS.  
Displays the Mean Airway Pressure, which is the weighted  
average of pressure in the patient’s airway over 6 breaths.  
MinVent  
Minute ventilation displays the amount of air delivered to the  
patient over the last minute in l/min BTPS.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
69  
Status Panel Indicators  
All of the indicators that may appear on the Status Panel are  
described in the following table.  
Indicator  
Description  
Indicates that the device is in Full Menu Access mode, which  
means you can adjust all prescription settings. Philips Respironics  
recommends that you change the device to Limited Menu Access  
mode before giving the device to the patient, so patients cannot  
adjust their prescription settings. Only trained health care  
professionals and clinicians should adjust prescription settings.  
Displays when a Secure Digital (SD) memory card is inserted in  
the ventilator.  
Displays when the ventilator detects an error with the SD card.  
Displays at all times when an external battery is attached to  
the ventilator. The level of green shading shown in the symbol  
indicates the battery capacity and will go down as the battery  
charge level decreases. When the entire symbol is green, the  
battery is fully charged.  
Displays at all times when a detachable battery is attached to  
the ventilator. The level of green shading shown in the symbol  
indicates the battery capacity and will go down as the battery  
charge level decreases. When the entire symbol is green, the  
battery is fully charged.  
Displays at all times, indicating the status of the internal battery. The  
level of green shading shown in the symbol indicates the battery  
capacity and will go down as the battery charge level decreases.  
When the entire symbol is green, the battery is fully charged.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
70  
Indicator  
Description  
A black box displays around the battery that is currently  
supplying power to the ventilator when AC power is not available.  
(In the status panel shown above, the external battery is in use, so  
the  
symbol displays.)  
A yellow lightning bolt symbol displays with the Detachable or  
Internal battery symbol to indicate when the battery is charging.  
(In the status panel shown on the previous page, the detachable  
battery is being charged, so the  
symbol displays.)  
Displays when the Alarm Indicator/Audio Pause button has been  
pressed and Audio Pause is active. The alarm is silenced for one  
minute when the Alarm Indicator/Audio Pause button is pressed.  
Displays when the Ramp feature is active.  
Note: If a battery in use is very low (less than 20 minutes remaining), the inside of the box surrounding the  
battery symbol will change to yellow and all of the bar indicators in the battery will be empty. If a battery in use  
is near depletion (less than 10 minutes remaining), the inside of the box surrounding the battery symbol will  
change to red, and all of the bar indicators in the battery will be empty. These color changes only occur for the  
last available battery source.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
71  
On-Screen Button Panel  
The illustration below shows the on-screen button panel on the Main  
Menu screen, in relation to the buttons on the front of the device.  
On-Screen  
Button  
Panel  
Example On-Screen Button  
Panel  
Note: The on-screen buttons  
will vary depending on which  
screen is displayed and what  
settings are enabled on your  
device.  
At the very bottom of the display screen is the on-screen button  
panel. This panel corresponds with the control buttons on the  
ventilator:  
The left on-screen button specifies the action for the Left  
button on the device.  
The center on-screen button specifies the action for the  
Up/Down buttons on the device.  
The right on-screen button specifies the action for the  
right button on the device.  
Navigating the Menu Screens  
To navigate through all of the menu screens and settings:  
Use the Up/Down button to scroll through the menu  
options and settings.  
Use the Left and Right buttons to perform the actions  
specified on the screensleft and right on-screen buttons.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
72  
Note: For some therapy  
settings, once you reach the  
highest or lowest setting  
available, pressing the Up/  
Down button again will cycle  
back through the settings.  
For the parameters that do  
not wrap, when you reach  
the highest or lowest setting  
possible, a “Limit Reached”  
message appears in the Menu  
Banner on-screen.  
Changing and Viewing Settings in Full Menu  
Access Mode  
Clinicians can view and change settings using the Menu screens  
when the device is in Full Menu Access mode. To enter the Menu  
screens from the Monitor screen, press the Up button on the  
ventilator. The Main Menu screen shown below appears.  
Main Menu Screen Example  
Note: In the example Main  
Menu screen shown, the 2/6  
that appears in the Menu  
banner indicates that item 2 is  
highlighted from a total of 6  
items in the menu.  
Choose from the following selections on the Main Menu screen:  
Safely Remove SD Card - This option will appear if an SD  
card is inserted in the ventilator. Select this option when  
you want to remove the SD card. When the “Remove SD  
Cardconfirmation message appears, remove the card.  
If you press the left (cancel) button or don’t remove the  
card within 30 seconds, the confirmation message will  
close and the ventilator will continue writing to the card.  
Settings and Alarms: View and change prescription  
settings and alarms.  
Note: If you change a setting  
but decide you do not want to  
save it, you can press the Left  
button to cancel the change.  
Options: View and change device settings, such as Full or  
Limited Access mode, Detailed View, Language, etc.  
Alarm Log: View a list of the 20 most recent alarms that  
have occurred.  
Event Log: View a list of all events that have occurred,  
such as ventilator setting changes, ventilator inoperative  
conditions, alarms, etc.  
Information: View detailed information about the device,  
such as the device’s software version and serial number.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
73  
Note: In the Options menu  
described later in this chapter,  
you can specify the Pressure  
Units displayed by the device,  
choosing either cm H2O, hPa, or  
mBar. The default setting is  
cm H2O, so this manual uses cm  
H2O throughout.  
Changing the Device Settings and Alarms  
From the Main Menu screen, use the Up/Down button to highlight  
the Settings and Alarms menu, and press the Right button to select  
the menu.  
Device Settings Common to All Therapy Modes  
Some of the settings on this menu will vary depending on the  
therapy mode you select. The section below describes all of the  
settings that are common to all therapy modes.  
Dual Prescription Setting  
You can turn the dual prescription setting on or off. Enable the  
setting if you want to create two separate prescriptions for the  
patient. For instance, you may want to set up a daytime prescription  
and then a separate nighttime prescription. If you enable this setting,  
then the menu options on the Main Menu will change to include  
three new options:  
Switch to Primary/Secondary Settings  
Primary Settings and Alarms  
Secondary Settings and Alarms  
The Main Menu screen will look like the screen below.  
Main Menu Screen with Dual  
Prescription Enabled  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
74  
Mode Setting  
Note: See Chapter 3 for detailed  
information about each therapy  
mode.  
You can change the Mode setting on the Settings and Alarms screen  
to one of the following therapy modes:  
Note: Refer to the chart in  
Chapter 3 to easily review all of  
the settings available in each  
therapy mode.  
CPAP  
S
S/T  
T
Note: This chapter describes  
how to enable all of the device  
settings, including device  
alarms. For detailed information  
about each alarm, please refer  
to Chapter 6.  
PC  
PC-SIMV  
CV  
AC  
SIMV  
Circuit Type  
Note: To change the circuit  
type, you must be in the  
Setup screen with the airflow  
turned off. See the Full Menu  
Access Mode section for more  
information.  
There are three circuit types you can select:  
Passive  
Active PAP  
Active Flow  
Note: When the Circuit Type  
setting is set to Passive Circuit,  
all Ramp Start Pressure settings  
in all modes will maintain the  
minimum range.  
The Passive circuit type uses the Whisper Swivel II passive exhalation  
device. The Active PAP circuit type uses an active exhalation device  
with a proximal air pressure sensing connection. The Active Flow circuit  
type uses an active exhalation device with a proximal flow sensor.  
Note: When the Circuit Type  
setting is set to Active PAP or  
Active Flow, the Flex and AVAPS  
features are unavailable.  
When using the Passive circuit, the ventilator displays estimated  
patient pressures based on the resistance of the standard patient  
circuit (Whisper Swivel II with 1.8 meter tubing). Adding accessories  
to the patient circuit (humidifier, water trap, etc.) may cause an  
increase in circuit resistance and cause the device to display slightly  
higher pressures than what is actually delivered to the patient.  
With the Active PAP or Active Flow circuit type selected, patient  
pressure is measured directly and is not affected by any change in  
circuit resistance.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
75  
The Passive circuit provides leak compensation. When using the Passive  
circuit in Volume Ventilation, the set Vti is delivered to the patient  
above the calculated circuit and cuff (or mask) leak. This is different  
from traditional active circuit ventilation where the cuff (or mask) leak  
reduces the tidal volume delivered to the patient. Volume ventilation  
with the Passive circuit delivers an inspiratory tidal volume close to  
the device setting regardless of leak; this should be considered when  
transitioning a patient from an active to a passive circuit. With a Passive  
circuit, Vte is estimated based on the calculated sum of circuit and cuff  
(or mask) leak.  
The Active Flow circuit monitors proximal flow and proximal  
pressure. When using the Active Flow circuit configuration, flow  
trigger with leak compensation may be enabled. The default setting  
when using the Active Flow Circuit is Leak Compensation On. The  
clinician has the option to turn off leak compensation; however,  
unintentional leak will not be compensated. Both options measure  
the flow at the proximal flow sensor. Leak compensation is not  
available in the Active with PAP circuit configuration.  
WARNING  
Circuit Disconnect  
You should not rely on any  
single alarm to detect a circuit  
disconnect condition. The Low  
Tidal Volume, Low Minute  
Ventilation, Low Respiratory  
This setting enables or disables the circuit disconnect alarm. If  
enabled, an audible alarm will sound when a large, continuous air  
leak (such as mask removal) has been detected in the circuit.  
Rate, and Apnea alarms should  
be used in conjunction with the  
Circuit Disconnect alarm.  
You can choose Off to disable the alarm. Or, you can increase or  
decrease the setting from 5 to 60 seconds in 5 second increments.  
For example, a setting of 10 means that the alarm will sound after the  
circuit has been disconnected for 10 seconds.  
Apnea  
This setting enables or disables the apnea alarm. If enabled, an  
audible alarm will sound when an apnea is detected.  
You can choose Off to disable the alarm. Or, you can increase or  
decrease the setting from 10 to 60 seconds in 5 second increments.  
For example, a setting of 10 means that the alarm will sound if the  
time between spontaneous breaths exceeds 10 seconds.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
76  
Low Vte  
Note: The High and Low Vte  
alarms are available when the  
Passive or Active Flow Circuit is  
selected.  
This setting enables or disables the Low Vte alarm. The alarm  
activates when the estimated exhaled tidal volume is less than or  
equal to this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 40 ml to 2000 ml in 5 ml  
increments. It cannot be set higher than the High Vte setting.  
When AVAPS is On, the alarm activates when the calculated tidal  
volume is less than 90% of the target tidal volume setting. This alarm  
can be set to on or off.  
High Vte  
This setting enables or disables the High Vte alarm. The alarm  
activates when the estimated exhaled tidal volume is greater than or  
equal to this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 50 ml to 2000 ml in 5 ml  
increments. It cannot be set lower than the Low Tidal Volume setting,  
except to be turned off.  
Note: The High and Low Vti  
alarms are only available when  
the Active PAP Circuit is selected.  
Low Vti  
This setting enables or disables the Low Vti alarm. The alarm activates  
when the measured inhaled tidal volume is less than or equal to this  
setting. You can choose Off to disable this alarm, or you can increase  
or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It  
cannot be set higher than the High Vti setting.  
High Vti  
This setting enables or disables the High Vti alarm. The alarm  
activates when the measured inhaled tidal volume is greater than or  
equal to this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 40 ml to 2000 ml in 5 ml  
increments. It cannot be set lower than the Low Vti setting, except to  
be turned off.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
77  
Low Minute Ventilation  
Note: You can increase or  
decrease the Low Minute  
Ventilation setting in  
0.1 l/min increments from  
0.1 to 0.99 l/min and 1 l/min  
increments from 1 to 99 l/min.  
This setting enables or disables the Low Minute Ventilation alarm. The  
alarm activates when the calculated minute ventilation is less than or  
equal to this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 0.1 l/min to 99 l/min.  
High Minute Ventilation  
This setting enables or disables the High Minute Ventilation alarm.  
The alarm activates when the calculated minute ventilation reaches  
or exceeds this setting. You can choose Off to disable this alarm, or  
you can increase or decrease the setting from 1 l/min to 99 l/min  
in 1 l/min increments. It cannot be set lower than the Low Minute  
Ventilation setting except to be turned off.  
Low Respiratory Rate  
This setting enables or disables the Low Respiratory Rate alarm. The  
alarm activates when the measured respiratory rate is less than or  
equal to this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM  
increments. It cannot be set higher than the High Respiratory Rate.  
High Respiratory Rate  
This setting enables or disables the High Respiratory Rate alarm.  
The alarm activates when the measured respiratory rate reaches or  
exceeds this setting. You can choose Off to disable this alarm, or you  
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM  
increments. It cannot be set lower than the Low Respiratory Rate  
except to be turned off.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
78  
Low Inspiratory Pressure  
This setting configures the Low Inspiratory Pressure alarm. It is only  
user-settable in CV, AC, and SIMV modes. It cannot be set lower than  
PEEP + 2 cm H2O or higher than the High Inspiratory Pressure. For  
passive circuits, you can increase or decrease the Low Inspiratory  
Pressure from 6 to 40 cm H2O in increments of 1 cm H2O. For active  
circuits, you can increase or decrease the setting from 2 to 40 cm H2O  
in increments of 1 cm H2O. For pressure modes, this alarm is not user-  
settable.  
High Inspiratory Pressure  
This setting enables or disables the High Inspiratory Pressure alarm. It  
is only user-settable in CV, AC, and SIMV modes. The High Inspiratory  
Pressure cannot be set lower than the Low Inspiratory Pressure. You  
can increase or decrease the High Inspiratory Pressure from 10 to 80  
cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is  
not user-settable.  
Additional Settings Specific to Therapy Modes  
The Settings and Alarms menu also contains many additional  
settings specific to the various therapy modes. The specific settings  
for each therapy mode are described below.  
Continuous Positive Airway Pressure (CPAP) Mode  
In addition to the general settings described in the previous section  
of this manual, you can also set the following settings in CPAP mode.  
Note: If the CPAP pressure is  
set to 4 (the minimum setting),  
the Ramp Length setting will be  
unavailable.  
1. CPAP  
You can increase or decrease the CPAP pressure setting from 4 to  
20 cm H2O in increments of 1.  
2. Trigger Type  
Note: Trigger Type is not  
The device can be set to trigger breaths based on automatic flow  
thresholds or specific flow settings. You can select Auto-Trak or  
Flow Trigger as the Trigger Type. When Auto-Trak is selected, the  
Auto-Trak trigger initiates based on automatic flow thresholds.  
available when an Active PAP  
or Active Flow circuit is selected.  
Flow trigger is the triggering  
method used for Active PAP and  
Active Flow circuits.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
79  
Note: Auto-Trak is only  
available if the Passive Circuit is  
selected.  
When Trigger Type is set to Flow Trigger, Flow Trigger Sensitivity  
Leak Compensation, and Flow Cycle Sensitivity become active,  
and the trigger initiates based on the Flow Trigger Sensitivity  
setting.  
Note: Flow Trigger with leak  
compensation is only available if  
the Active Flow circuit is selected.  
3. Flow Trigger Sensitivity  
If you set the Trigger Type to Flow Trigger, the Flow Trigger  
Sensitivity setting displays. You can increase or decrease the  
setting from 1 to 9 l/min in 1 l/min increments.  
The flow trigger initiates when the patient’s inspiratory effort  
creates a flow equal to or greater than the flow sensitivity  
setting.  
Note: Enabling Leak  
Compensation when using the  
Active Flow Circuit configuration  
only affects triggering and does  
not affect tidal volume delivery  
or Vte measurement.  
4. Leak Compensation  
If you are using an Active Flow circuit, you can turn Leak  
Compensation On or Off.  
5. Flow Cycle Sensitivity  
Note: A flow cycle sensitivity  
set at 90% will result in the  
most sensitivity. If the flow cycle  
sensitivity is set at 10%, this will  
result in the least sensitivity.  
If you set the Trigger Type to Flow Trigger, the Flow Cycle  
Sensitivity setting displays. You can increase or decrease the  
setting from 10 to 90 percent (%) in 1% increments.  
As flow begins to decrease during inspiration,  
if the patient flow is less than the flow cycle set  
point, the device will cycle to expiration. For  
example: if the flow cycle is set to 75%, when the  
flow has decreased by 25% of the peak flow, the  
device will cycle to the EPAP/PEEP level.  
6. Ramp Length  
The Ramp Length allows you to set the ramp time.  
You can disable Ramp by selecting Off, or you  
can increase or decrease the Ramp Length setting from 5 to 45  
minutes in 5-minute increments.  
Flow Cycle Sensitivity  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
80  
7. Ramp Start Pressure  
Note: The Ramp Start Pressure  
setting will not display if the  
Ramp Length is set to Off or if  
the CPAP pressure is set to  
4 cm H2O.  
You can increase or decrease the ramp start pressure in  
increments of 1 from 4 cm H2O to the CPAP pressure setting. The  
patient also has access to this setting, unless the ramp length is  
set to Off.  
Note: Ramp Start Pressure is  
less than or equal to CPAP - 1 cm  
H2O in CPAP mode.  
8. Flex  
You can enable or disable the Flex setting. Off disables the  
setting and prevents the patient from using Flex. To enable the  
setting, set Flex to 1, 2, or 3. The patient also has access to this  
setting, if Flex is enabled. The Flex feature is not available when  
using an active circuit type.  
Note: In CPAP mode, Flex is only  
available when CPAP is greater  
than 4 cm H2O.  
Note: In S mode, Flex is only  
available when EPAP is greater  
than or equal to 4 cm H2O and  
IPAP is less than or equal to  
25 cm H2O.  
Spontaneous (S) Mode  
The following settings, described in the CPAP mode section of this  
chapter, also are available in S mode:  
Note: Flex is only available  
when Auto-Trak is enabled.  
Note: Ramp is not available in  
the Passive Circuit when IPAP =  
EPAP = 4 cm H2O or when IPAP  
Min = EPAP = 4 cm H2O.  
Flex  
Trigger Type  
Flow Trigger Sensitivity  
Flow Cycle Sensitivity  
Ramp Length  
Note: Ramp Start Pressure is  
less than or equal to EPAP - 1 cm  
H2O in S, S/T, T, and PC modes.  
Ramp Start Pressure  
Note: Ramp Start Pressure is  
greater than or equal to 0 cm  
H2O when the circuit type is  
Active PAP or Active Flow in S,  
S/T, T, and PC modes.  
In addition to those settings, the settings below are also available in  
S mode.  
1. AVAPS  
Note: Ramp Start Pressure  
is greater than or equal to 4  
cm H2O when the circuit type  
is Passive in S, S/T, T, and PC  
modes.  
AVAPS is only available if Flex is not enabled.  
You can disable AVAPS by selecting Off, or you can enable AVAPS  
by selecting On. If you select Off, the IPAP setting displays. If you  
select On, the IPAP Max Pressure and IPAP Min Pressure display.  
Note: AVAPS is only available if  
the Passive Circuit is selected.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
81  
2. IPAP  
The IPAP setting displays if AVAPS is Off. You can increase or  
Note: IPAP, IPAP Max, or IPAP  
Min cannot be set to more than  
30 cm H2O above EPAP.  
decrease the Inspiratory Positive Airway Pressure (IPAP) from 4  
to 50 cm H2O in increments of 1. IPAP is limited to a maximum of  
25 cm H2O when Flex is enabled. You cannot set the IPAP setting  
lower than the EPAP setting.  
3. IPAP Max Pressure  
The IPAP Max Pressure setting displays if AVAPS is enabled.  
You can increase or decrease the setting from 4 to 50 cm H2O  
in increments of 1. The IPAP Max Pressure must be equal to or  
greater than the IPAP Min value.  
4. IPAP Min Pressure  
The IPAP Min Pressure setting displays if AVAPS is enabled.  
You can increase or decrease the setting from 4 to 50 cm H2O  
in increments of 1. The IPAP Min Pressure must be equal to or  
greater than the EPAP value, and it must be less than or equal to  
the IPAP Max Pressure.  
Note: EPAP cannot be set to  
more than 30 cm H2O below  
IPAP, IPAP Max, or IPAP Min.  
5. EPAP  
You can increase or decrease the Expiratory Positive Airway  
Pressure (EPAP) from 4 to 25 cm H2O in increments of 1. For  
active circuits, EPAP can be set to zero.  
When AVAPS is disabled, the EPAP setting must be less than  
or equal to the IPAP setting. When AVAPS is enabled, the EPAP  
pressure must be less than or equal to the IPAP Min Pressure.  
6. Tidal Volume  
Note: In CV, AC, and SIMV  
modes, the tidal volume setting  
is limited by the Inspiratory  
Time, to maintain the system’s  
minimum and maximum peak  
flows.  
The Tidal Volume setting displays if AVAPS is enabled. You can  
increase or decrease the setting from 50 to 2000 ml in 5 ml  
increments. Use this setting to establish the target volume of  
gas which the ventilator will produce and deliver during each  
Spontaneous breath.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
82  
Note: The Rise Time setting only  
displays if Flex is disabled. If Flex  
is enabled, the device will use a  
rise time of 3.  
7. Rise Time  
You can adjust the rise time to find the most comfortable setting  
for the patient. Increase or decrease the setting from 1 to 6  
until you find the right setting. The rise time levels from 1 to 6  
progressively reflect slowed response of the pressure increase  
that will take place at the beginning of inspiration.  
Note: In S/T, T, PC, PC-SIMV,  
SIMV, CV, and AC modes, the  
Apnea Rate is greater than or  
equal to the Breath Rate and is  
limited by the current Inspiratory  
Time setting to maintain a  
minimum 1:1 I:E ratio.  
8. Apnea Rate  
If the Apnea alarm is enabled, you can set the Apnea Rate from  
4 to 60 BPM in 1 BPM increments. In S mode, the Apnea Rate is  
greater than or equal to 1:2 I:E ratio.  
Spontaneous/Timed (S/T) Mode  
All of the settings described in the S Mode section are also available  
in S/T mode, except for the Flex setting. In addition to those settings,  
the settings below are also available in S/T mode.  
Note: In volume modes, the  
Breath Rate range is limited by  
the current Inspiratory Time  
setting to maintain a minimum  
1:1 I:E ratio.  
1. Breath Rate  
In AC mode, you can increase or decrease the Breath Rate setting  
from 0 to 60 BPM, while in all other modes, you can increase  
or decrease the setting from 1 to 60 BPM in 1 BPM increments.  
Use the Breath Rate setting to establish the minimum rate of  
mandatory breaths that the ventilator will deliver per minute.  
Note: In pressure modes, the  
inspiratory time range is limited  
by the current Breath Rate  
setting to maintain a minimum  
1:1 I:E ratio.  
2. Inspiratory Time  
You can adjust the Inspiratory Time setting from 0.3 to 5.0  
seconds in 0.1 second increments. Inspiratory Time is the  
duration for the inspiratory phase of a mandatory breath.  
Note: In volume modes, the  
inspiratory time range is limited  
by the current Tidal Volume and  
Breath Rate settings to maintain  
a minimum 1:1 I:E ratio and  
the system’s minimum and  
maximum peak flow.  
Timed (T) Mode  
All of the settings available in S/T mode are available in T mode,  
except for the Trigger Type setting. Please refer to the descriptions in  
the S and S/T Mode sections of this chapter for detailed information  
on the T mode settings.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
83  
Pressure Control (PC) Mode  
All of the settings available in S/T mode are available in PC mode,  
except for the Flow Cycle Sensitivity setting (when Flow Trigger is  
enabled). Please refer to the descriptions in the S and S/T Mode  
sections of this chapter for detailed information on the PC mode  
settings.  
Pressure Control Synchronized Intermittent Mandatory Ventilation  
(PC-SIMV) Mode  
The following settings, described in the S and S/T mode sections of  
this chapter, also are available in PC-SIMV mode:  
Breath Rate  
Inspiratory Time  
Trigger Type  
Flow Trigger Sensitivity (if Trigger Type is set to Flow  
Trigger)  
Flow Cycle Sensitivity (if Trigger Type is set to Flow  
Trigger)  
Rise Time  
In addition to these, the following settings are also available in PC-  
SIMV mode.  
Note: In PC-SIMV mode, you  
cannot set up Pressure Support  
for Mandatory and Assist  
breaths (Pressure - PEEP) greater  
than 30 cm H2O.  
1. Pressure  
You can increase or decrease the Pressure setting from 4 to 50  
cm H2O in increments of 1. This is the pressure the ventilator will  
deliver during the inspiratory phase of a mandatory or assist  
breath.  
Note: The Pressure Support and  
PEEP settings together cannot  
exceed 50 cm H2O.  
2. Pressure Support  
You can increase or decrease the Pressure Support setting from  
0 to 30 cm H2O in increments of 1. This is the pressure support  
the ventilator will deliver during the inspiratory phase of a  
Spontaneous breath.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
84  
3. PEEP  
The Positive End Expiratory Pressure (PEEP) setting can be  
increased from 0 to 25 cm H2O in active circuits and 4 to 25 cm  
H2O in passive circuits, in increments of 1. PEEP is the positive  
pressure maintained in the patient circuit during exhalation. The  
PEEP must be less than or equal to the pressure setting.  
Control Ventilation (CV) Mode  
The following settings, described in the previous sections of this  
chapter, are also available in CV mode:  
Tidal Volume  
Breath Rate  
Inspiratory Time  
PEEP  
High Inspiratory Pressure  
Low Inspiratory Pressure  
In addition to these, the following settings are also available in CV  
mode.  
Note: The Flow Pattern setting  
might be limited to only Ramp  
or Square based on the Tidal  
Volume, Inspiratory Time, and  
Breath Rate settings to maintain  
the minimum and maximum  
peak flows.  
1. Flow Pattern  
You can choose either Ramp or Square for the Flow Pattern setting.  
2. Sigh  
You can enable or disable the Sigh setting by selecting On or Off.  
A Sigh is a breath that is delivered every 100 breaths at 150% of  
the normal volume.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
85  
Assist Control (AC) Mode  
The AC mode contains the following settings described in the S, S/T,  
PC-SIMV, and CV mode sections in this chapter. Please refer to the  
descriptions in those sections for detailed information.  
Tidal Volume  
Breath Rate  
Inspiratory Time  
Flow Pattern  
Note: The Low Inspiratory  
Pressure is limited to PEEP +2 in  
CV, AC, and SIMV modes.  
PEEP  
Trigger Type  
Flow Trigger Sensitivity  
Sigh  
Note: Flow Cycle Sensitivity is  
not available in AC mode.  
High Inspiratory Pressure  
Low Inspiratory Pressure  
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode  
The SIMV mode contains the following settings described in the S,  
S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the  
descriptions in those sections for detailed information.  
Tidal Volume  
Breath Rate  
Inspiratory Time  
Pressure Support  
Flow Pattern  
PEEP  
Trigger Type  
Sigh  
Rise Time  
High Inspiratory Pressure  
Low Inspiratory Pressure  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
86  
Viewing and Changing Options Menu Items  
From the Main Menu screen, select the Options item.  
The Options menu appears, shown in the screen below.  
Options Menu  
The following settings are available on the Options menu.  
Menu Access – You can select Full or Limited menu  
access. Full menu access allows operators to access all  
ventilator and prescription settings. Limited menu access  
allows operators to access only certain settings and  
does not allow them to change prescription settings. To  
prevent patients from changing prescription settings, do  
not give them Full menu access.  
Detailed View – You can turn Detailed View on or off  
using this setting. Detailed view displays additional  
settings and therapy information on the Monitor screen.  
Language – The next item on the Options menu allows  
you to select the Language that the software will appear  
in (English, French, German, etc.). The information on the  
screens will display in the language selected here.  
Pressure Units – The next item allows you to select the  
pressure units that will display on the screens. You can  
choose either:  
-
-
-
cm H2O  
hPa  
mBar  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
87  
All pressure units on the screens will display in the unit of  
measure selected here.  
Alarm Volume – You can adjust the volume of the device  
alarms using this setting. Select either Loud or Soft as the  
alarm volume options.  
WARNING  
Make sure the alarm volume is  
set loud enough to be heard by  
the caregiver. Consider the use of  
a remote alarm.  
Keypad Lock – You can enable or disable the Keypad  
Lock feature, which is described in detail earlier in this  
chapter. Enabling the Keypad Lock feature can prevent  
someone from accidentally pressing a button and  
changing any of the settings. Select On to enable the  
feature or Off to disable it.  
Keypad Backlight – The next item you can set is the  
Keypad Backlight. You can turn the backlight On or Off  
using this setting. Whenever you press the  
button to  
begin therapy, the keypad backlight temporarily lights  
up. Once therapy is being provided, the keypad will be lit  
according to this Keypad Backlight setting. If the setting  
is On, the backlight remains on while therapy is provided.  
If the setting is Off, the backlight remains off while  
therapy is provided.  
LCD Brightness – The LCD display is lit by a backlight.  
The backlight turns on when the initial Startup screen  
displays. You can adjust the brightness of the LCD  
backlight from 1 – 10, with 1 being the dimmest setting  
and 10 being the brightest.  
Screen Saver – You can change the screen saver to  
reduce power consumption or dim the screen in a dark  
room. You can choose the following settings:  
Note: Setting the screen saver  
to Black allows the device to run  
for a longer period of time on  
battery power.  
-
-
-
Off: No screen saver displays and the LCD backlight remains  
lit at your brightness setting.  
Breath: The display appears as a black screen, with only the  
patient breath indicator and manometer visible.  
Black: The display’s backlight is turned off and the display is  
black with no information visible.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
88  
-
Dim: The display’s backlight is decreased, so that the display  
is still visible but not as bright.  
If enabled, the screen saver will display after 5 minutes of no  
keypad activity. Pressing any button on the device will exit the  
screen saver. And, any alarm or informational message will also  
exit the screen saver.  
Date Format – You can choose either mm/dd/yyyy or  
dd/mm/yyyy as the date format that will display on the  
device screens.  
Time Format – You can choose to display either an AM/  
PM time format or 24 Hour time format (for example,  
2:49 PM or 14:49).  
Month – The month defaults to the current month. The  
adjustable range is from 1 (January) – 12 (December).  
Day – The day defaults to the current day. The adjustable  
range is from 1 – 31. The maximum value is based on the  
selected month.  
Year – The year defaults to the current year. The  
adjustable range is from 2000 – 2099.  
Hour – The hour defaults to the current hour. The  
adjustable range is from 12 AM – 12 PM or 0-23,  
depending on the selected Time Format.  
Minute – The minute defaults to the current minute. The  
adjustable range is from 0 – 59.  
IP Address Mode – You can change the IP address  
mode to either DHCP or Static, depending on the type of  
network you are using (if applicable).  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
89  
Operational Hours – The operational hours displays  
the total number of hours that the device blower has  
been on since the last time this value was reset. You can  
reset this value to zero (0) if desired (e.g., each time you  
give the device to a new patient). This value helps you  
determine how often the patient is using the device. The  
Operational Hours shown here differs from the Blower  
Hours shown on the Information screens. The Blower  
Hours displayed in the Information screen is the total  
number of hours that the blower has been working over  
the life of the device. You cannot reset this value.  
Viewing the Alarm Log  
From the Main Menu screen, you can select Alarm Log to access the  
Alarm Log screen. An example is shown below.  
Note: In the Alarm Log screen,  
the 1/2 shown in the Menu  
banner indicates that page 1  
of 2 alarm log pages is being  
viewed at this time.  
Alarm Log Screen  
The alarm log displays the alarms in chronological order with the  
most recent events displayed first. It lists the 20 most recent alarms  
or messages that appeared on the device display. When the device is  
in Limited Menu access mode, the alarm log cannot be cleared. It can  
Note: In Full Menu access mode,  
you can press the Right (Clear)  
button to clear the alarm log if  
be cleared when in Full Menu access mode. Depending on how many desired.  
alarms have occurred, the alarm log may be several pages long. The  
entries in the alarm log use the same names that displayed when the  
alarm initially occurred and was displayed in the Alarm View.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
90  
Note: In Full Menu access mode,  
you can press the Right (Clear)  
button to clear the event log if  
desired.  
Viewing the Event Log  
From the Main Menu screen, you can select Event Log to access the  
Event Log screen.  
The event log displays a list of all events that have occurred, such as  
ventilator setting changes, ventilator inoperative conditions, alarms,  
etc., in chronological order with the most recent events displayed  
first. When the device is in Limited Menu access mode, the event  
log is not available. It can be viewed and cleared when in Full Menu  
access mode. You can page through the event log if it is multiple  
pages. The number of pages appears in the upper right corner of the  
panel.  
In the event log descriptions, any description beginning with a 1:  
or 2: is a prescription change event. The 1 represents a change to a  
primary setting and the 2 represents a change to a secondary setting.  
This is followed by the setting that was changed.  
The last two columns indicate setting and alarm changes. If the entry  
is a setting change, the first column shows the old setting value and  
the last column shows the new setting. If the entry is an alarm, the  
first column shows the value that triggered the alarm and the last  
column shows the number of seconds that the alarm was active.  
Viewing Device Information  
From the Main Menu screen, you can select Information to access  
the Information screen. You can also view the Information screen by  
holding the Down key for 5 seconds. This causes the detailed view  
of the Monitor Screen and the Information Menu to be displayed  
temporarily. This key sequence is valid from the Monitor Screen while  
in Full or Limited Access.  
The Information screen provides you with a summary of the current  
prescription settings, device settings, and system settings. You can  
use the Up/Down buttons to scroll through the information.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
91  
Updating Prescriptions Using the SD Card  
WARNING  
When you change the device  
prescription, alarms and other  
settings using the SD card,  
Trilogy200 requires that the  
caregiver review and verify the  
changes prior to the changes  
being used by the device.  
The caregiver or health care  
professional is responsible to  
ensure that the prescription  
settings are correct and  
With the Trilogy200, you can update the patient’s prescription using  
the SD Card. This feature lets you update a single prescription or  
both prescriptions if the dual prescription feature is enabled. The  
prescription update can occur either when the ventilator is off or on.  
1. Insert an SD Card with a valid prescription into the device. A  
“Change Prescription?” message appears on the display:  
2. Select No to cancel the prescription update process and return  
to the previous display (the black screen if the airflow was off or  
the Monitor/Standby screen if the airflow was on). Select Yes to  
start the prescription update process. Once the prescription is  
read in and validated, one of the following screens appears on  
the display to allow you to ensure the prescription is correct:  
compatible with the patient  
after using this feature. Installing  
the wrong prescription for a  
particular patient may result  
in improper therapy, lack of  
appropriate safety monitoring,  
and risk of death or injury to the  
patient.  
Ventilator Off Prescription  
Display Screen  
Ventilator On Prescription  
Display Screen  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
92  
3. Select Cancel to cancel the prescription update process and  
Note: Make sure you closely  
review the prescription and  
confirm that all settings are  
correct.  
return the screen to the initial state before the prescription  
update started. Select Page to review the entire prescription.  
The Menu Banner will reflect the prescription being updated.  
4. Once the entire prescription has been reviewed, a screen  
displays with the option to Cancel or OK the changes. Select  
Cancel to cancel the prescription update process and return the  
screen to the initial state before the prescription update started.  
Select OK to complete the prescription update and display the  
Prescription Change confirmation screen.  
Note: If both prescriptions of  
a dual prescription are being  
updated, you must accept the  
primary prescription using  
the OK button before being  
able to review the secondary  
prescription. Both prescriptions  
have to be accepted before any  
changes are made.  
If the SD card is removed at any time during the prescription  
update, the process aborts and the screen returns to the initial  
state before the prescription update started.  
A confirmation screen will appear if errors occur during this process.  
The following table summarizes the prescription errors, their possible  
causes, and the actions to take.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
93  
Message  
Cause  
Action  
Prescription Change Failed  
Displayed when an error  
Remove the card and have the  
occurs during the prescription prescription replaced with a  
update. This is caused by an  
attempt to set a value for a  
prescription parameter that  
is not valid for the therapy  
mode or an attempt to set a  
prescription parameter to an  
invalid value.  
valid prescription.  
Prescription Failed – Circuit  
Type  
Displayed when the circuit  
type in the new prescription  
If the circuit type in the  
prescription is correct, remove  
doesn’t match the circuit type the card, change the circuit  
set for the ventilator.  
type on the ventilator, and  
re-insert the card. If the  
circuit type on the ventilator is  
correct, remove the card and  
replace the prescription on  
the card with a prescription  
containing the correct circuit  
type.  
Prescription Failed – Card is  
Read Only  
Displayed when the  
prescription is write-protected. that the small switch on the  
Remove the card and check  
side of the SD Card is not in the  
Locked position.  
If you continue to receive this  
message, remove the card and  
have the prescription replaced  
with a valid prescription.  
Prescription Failed – Serial  
Number  
Displayed when the device  
serial number on the new  
prescription does not match  
the serial number for the  
device.  
Remove the card and have the  
prescription replaced with the  
prescription with the correct  
serial number.  
Prescription Failed – Version  
Displayed when the version  
of the prescription does not  
Remove the card and have  
the prescription replaced with  
match the version accepted by a prescription in the correct  
the device.  
version.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
94  
Changing and Viewing Settings in Limited  
Menu Access Mode  
Note: If the Keypad Lock feature  
is enabled, a message will  
appear that says “Hold Right Key  
for 5 seconds to unlock. Once  
you hold the key down for 5  
seconds, the keypad will unlock  
and you can enter the Main  
Menu screen. The Keypad Lock  
feature is explained in detail  
later in this chapter.  
After you press the  
button and access the Monitor screen, you  
can view and change settings using the Menu screens.  
To enter the Menu screens from the Monitor screen, press the Up  
button on the ventilator. The Main Menu screen appears, shown  
below.  
Note: In the example Main  
Menu screen shown, the 1/5  
that appears in the Menu  
banner indicates that item 1 is  
highlighted from a total of 5  
items in the menu.  
Main Menu Screen  
You can choose from the following selections on the Main Menu  
screen:  
Switch to Primary (or Secondary) Settings: If your health care  
professional has set up a dual prescription for you, this option  
will appear. You can select it to change to either your Primary or  
Secondary prescription settings.  
Safely Remove SD Card: This option will appear if an SD card  
is inserted in the ventilator. Select this option when you want to  
remove the SD card. When the “Remove SD Cardconfirmation  
message appears, remove the card. If you press the left (cancel)  
button or don’t remove the card within 30 seconds, the  
confirmation message will close and the ventilator will continue  
writing to the card.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
95  
Note: The Options, Alarm Log,  
and Information items are  
discussed in detail earlier in this  
chapter, in the “Changing and  
Viewing Settings in Full Menu  
Access Mode” section.  
My Settings: View and change certain prescription settings,  
such as rise time or ramp starting pressure, if these settings were  
enabled by your health care professional.  
Options: View and change certain device settings,  
such as your alarm volume, keypad lock, or keypad  
backlighting.  
Alarm Log: View a list of the 20 most recent alarms that  
have occurred.  
Information: View detailed information about your  
device, such as the device’s software version and serial  
number.  
Activating Your Primary or Secondary Prescription  
Note: If you are currently using  
the primary prescription, the  
menu option will say “Switch to  
Secondary Settings.If you are  
currently using the secondary  
prescription, the option will say  
“Switch to Primary Settings.”  
If your health care professional has set up a dual prescription for  
you, follow the steps below to change your prescription selection.  
Your health care professional might use this feature if, for instance,  
you need one prescription for daytime use and a second, different  
prescription for nighttime use.  
1. Use the Up/Down (Navigate) button to highlight the “Switch to  
Primary (or Secondary) Settingsoption.  
2. Press the Right (Select) button.  
A screen appears with a confirmation message, saying “Switch to  
Primary Settingsor “Switch to Secondary Settingsdepending  
on which prescription you are changing to.  
3. Press the Right (Yes) button to switch to the new prescription.  
If you decide not to change the prescription, press the Left (No)  
button. The display returns to the Main Menu after you’ve made  
your selection.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
96  
Note: The options on the My  
Settings screen will vary depending  
on how your health care  
Viewing and Changing My Settings Menu Items  
To view or change the therapy settings available on the My Settings  
screen, use the Up/Down (Navigate) button to highlight the My  
Settings option on the Main Menu, and press the Right (Select)  
button. The My Settings Menu screen appears.  
professional has set up your device.  
Note: For some therapy  
settings, once you reach the  
highest or lowest setting  
available, pressing the Up/  
Down button again will cycle  
back through the settings.  
For the parameters that do  
not wrap, when you reach  
the highest or lowest setting  
possible, a “Limit Reached”  
message appears in the Menu  
Banner on-screen.  
Follow the general instructions below to navigate and change any of  
the therapy settings. Detailed information about each setting follows.  
1. From the My Settings screen, use the Up/Down button to  
navigate to the setting you want to change and highlight it.  
2
To modify a setting once it is highlighted, press the Right  
(Modify) button.  
3. Use the Up/Down (Edit) button to scroll through the available  
settings. Press Down to decrease the setting, or press Up to  
increase the setting.  
Note: After you select “OK” to  
save the new setting, the next  
setting in the list is automatically  
highlighted.  
4. Once you have chosen the setting you want, press the Right (OK)  
button to save the new setting. Or, if you decide not to change  
the setting, press the Left (Cancel) button.  
Note: If some features or  
settings are not enabled on your  
ventilator, they will not appear  
on your My Settings screen. For  
instance, if Ramp is not enabled,  
the Ramp Start Pressure setting  
does not appear on the screen.  
5. You can now either navigate to the next setting you want to  
change using the Up/Down (Navigate) button, or exit the My  
Settings menu by pressing the Left (Finish) button to return to  
the Main Menu.  
You can change the following settings in the My Settings menu, if  
they are enabled by your health care professional.  
Rise Time – The Rise Time is the time it takes the  
ventilator to change from expiration to inspiration. If this  
feature is enabled, you can adjust the Rise Time from 1  
to 6 to find the setting that provides you with the most  
comfort. A setting of 1 is the fastest Rise Time, while 6 is  
the slowest.  
Ramp Start Pressure – The ventilator is equipped  
with an optional Ramp feature. Ramp reduces the  
pressure and then gradually increases the pressure to  
your prescription setting so you can fall asleep more  
comfortably.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
97  
Note: If the Flex feature is  
If your health care professional has enabled Ramp on  
your ventilator, you can adjust the Ramp Start Pressure  
setting. The setting can be adjusted from 4.0 to your  
prescription pressure setting in increments of 1.  
enabled, the Rise Time setting  
will not appear on your My  
Settings screen and cannot be  
adjusted. When Flex is enabled,  
the Rise Time is fixed at a setting  
of 3.  
Flex – The optional Flex feature allows you to adjust the  
level of air pressure you feel when you exhale during  
therapy. If this feature is enabled by your health care  
professional, you can adjust the setting from 1 to 3.  
Note: When adjusting the Flex  
setting, it is recommended that  
you start with the minimum  
setting of 1, which provides  
the least relief. Levels 2 and  
3 progressively increase the  
pressure relief.  
Note: If you are using an Active  
PAP circuit, the Flex feature is  
unavailable.  
Note: Flex is only available if  
Auto-Trak is enabled.  
Chapter 5 Viewing and Changing Settings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
98  
Note: Trilogy200 provides  
features to control access to the  
device (prescription) settings  
and to transfer new device  
settings onto the device using  
the SD Card. These features  
are intended to be used as part  
of your clinical/institutional  
procedures governing the use,  
security, and control of this  
medical device.  
Connecting the Ventilator to the Patient  
After you have finished adjusting your ventilator settings, perform  
the following steps to connect the ventilator to the patient.  
1. Perform System Checkout  
Do not connect the ventilator to the patient until you perform the  
system checkout procedures defined in Chapter 10.  
2. Start Therapy  
WARNING  
Press the  
button to begin therapy. When you start therapy, the  
display backlight and the backlights on the buttons turn on, the red  
To make sure the device is  
and yellow alarm LEDs turn on momentarily, and an audible indicator operating properly at start-up,  
always verify that the audible  
tone sounds and the alarm  
LEDs light red and then yellow  
sounds to indicate that therapy has started. The Startup screen  
appears on the display.  
momentarily. Contact Philips  
Respironics or an authorized  
service center for service if these  
3. Connect the Breathing Circuit to the Patient Interface  
After you have assembled the system, started therapy, and adjusted  
your ventilator settings as needed, you can connect the breathing  
circuit to the patient. The illustration below shows the breathing  
circuit connected to a mask. You can also connect the breathing  
circuit to a tracheostomy tube.  
indications do not occur at  
start-up.  
Example of a Fully  
Connected Patient Breathing  
Circuit with Passive  
Exhalation Device  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
6. Ventilator Alarms  
This chapter describes the ventilator alarms and what you should do  
if an alarm occurs.  
There are three types of alarms:  
Note: If multiple alarms occur  
High Priority – Require immediate response by the  
operator  
at the same time, all alarms are  
processed and displayed, but  
the alarms are ordered first by  
priority and then by occurrence,  
with the newest, highest priority  
alarms at the top of the list.  
The alarm precedence is in the  
following order: high priority,  
medium priority, low priority,  
and informational messages.  
Medium Priority – Require prompt response by the  
operator  
Low Priority – Require operator awareness. These alarms  
alert you to a change in the ventilator status.  
Additionally, the ventilator also displays informational messages and  
confirmation alerts that notify you of conditions that need attention  
but do not qualify as alarm conditions.  
Note: Not all alarms are  
available in every therapy  
mode; some alarms are mode-  
dependent.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
100  
Audible and Visual Alarm Indicators  
When an alarm condition occurs:  
The alarm LED indicator on the Alarm Indicator/Audio  
Pause button lights  
The audible alarm sounds  
A message appears on the screen describing the type of  
alarm  
The remote alarm (if applicable) is activated  
Each of these is described in detail below.  
Alarm LED Indicators  
The Alarm Indicator/Audio Pause button on the front of the ventilator  
lights up as follows whenever an alarm is detected:  
Alarm Indicator/Audio Pause  
Button  
Red Flashing Indicator – When the device detects a high  
priority alarm, the Alarm Indicator/Audio Pause button  
flashes red.  
Yellow Flashing Indicator – When the device detects a  
medium priority alarm, the Alarm Indicator/Audio Pause  
button flashes yellow.  
Yellow Solid Indicator – When the device detects a low  
priority alarm, a solid yellow light appears on the Alarm  
Indicator/Audio Pause button.  
The Alarm Indicator/Audio Pause button does not light up when  
informational messages or confirmation alerts display.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
101  
Audible Indicators  
An audible indicator sounds whenever a power failure or a high,  
medium, or low priority alarm is detected. Additionally, an audible  
indicator sounds for informational messages and to confirm that  
certain actions have occurred (for example, when an SD card is  
inserted or removed from the device).  
Ventilator Inoperative Audible Indicator – When a  
ventilator inoperative alarm occurs, a continuous audible  
alarm sounds. The alarm descriptions later in this chapter  
display this indicator as:  
Note: For the alarm indicators  
noted throughout this manual,  
each “dot” represents an audible  
beep.  
Power Failure Audible Indicator – When a power failure  
occurs, a series of beeps sounds in a 1 beep pattern,  
repeating one second on, then one second off. The alarm  
descriptions later in this chapter display this indicator as:  
High Priority Audible Indicator – When a high priority  
alarm is detected, a series of beeps sound in the  
following pattern, which is repeated twice: 3 beeps, a  
pause, and then 2 more beeps. This indicator continues  
until the cause of the alarm is corrected or the audible  
alarm is paused. The alarm descriptions later in this  
chapter display this indicator as:  
• • • • • • • • • •  
Medium Priority Audible Indicator – When a medium  
priority alarm is detected, a series of beeps sound in a  
3-beep pattern. This pattern repeats until the cause of  
the alarm is corrected or the audible alarm is paused.  
The alarm descriptions later in this chapter display this  
indicator as: • • •  
Low Priority Audible Indicator – When a low priority  
alarm is detected, a series of beeps sound in a 2-beep  
pattern. This pattern repeats until the cause of the alarm  
is corrected or the audible alarm is paused. The alarm  
descriptions later in this chapter display this indicator as:  
• •  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
102  
Informational Messages and Confirmation Audible  
Indicators – When an informational message appears  
on screen, a brief, 1- beep audible indicator sounds.  
Additionally, when the device detects that a certain  
action has been completed (for example, when the  
Start/Stop button is pressed to start therapy, or when an  
SD card is inserted or removed from the device) a brief,  
1- beep audible indicator sounds. The alarm descriptions  
later in this chapter display this indicator as: •  
Alarm Messages  
When the ventilator detects an alarm, the Alarms and Messages  
Screen is displayed showing a description of the alarm condition.  
When an alarm message appears, it will be highlighted in red if it  
is a high priority alarm or in yellow if it is a medium or low priority  
alarm. (The highlight color matches the alarm LED color on the Alarm  
Indicator/Audio Pause button.) If an alarm is manually reset by the  
user, the Alarms and Messages screen is removed and the Monitoring  
Screen is re-displayed. If the alarm self-cancels, the Alarms and  
Messages screen remains displayed, but the highlight for the active  
alarm is removed, the LED is unlit, and the audible alarm stops. The  
screen below is an example of a possible alarm message.  
Note: An alarm message will  
also display in the Menu Banner  
if a menu is active when an  
alarm occurs.  
Sample Alarms and  
Messages Screen  
If a menu is displayed on the screen when an alarm occurs, the  
description of the newly generated alarm is displayed in the menu  
banner area. This is done so that the modification to the current  
parameter can be completed before addressing the alarm condition in  
case the modification affects the alarm condition. The screen below is an  
example of an alarm message displayed in the menu banner.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
103  
Alarm in Menu Banner  
Screen  
The Alarms and Messages Screen will automatically display in place  
of the Monitor screen when exiting from the menu system using the  
Exit soft key when an alarm is displayed in the menu banner. If an  
alarm is manually reset by the user or self-cancels, the menu banner  
on-screen before the alarm occurred will reappear.  
If a Ventilator Inoperative alarm occurs, the entire display screen  
turns red and the Ventilator Inoperative message appears on-screen,  
as shown below.  
Ventilator Inoperative Alarm  
Screen  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
104  
To turn the ventilator off from a Ventilator Inoperative condition,  
use the normal power off sequence. When the Start/Stop button is  
selected, the following screen will display.  
Ventilator Inoperative Power  
Off Screen  
Select the Right button (Yes) to turn the ventilator off and stop the  
audible alarm. Selecting the Left button (No) will return the screen to  
the Ventilator Inoperative Alarm Screen without silencing the audible  
alarm.  
Remote Alarm  
When the ventilator detects an alarm condition, if you are using a  
remote alarm system, a signal is sent from the ventilator to activate the  
remote alarm. The Remote Alarm is discussed in Chapter 4.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
105  
Audio Pause and Alarm Reset Features  
This section describes the Audio Pause and Alarm Reset features.  
Audio Pause  
When an alarm occurs, you can temporarily silence the audible  
indicator by pressing the Alarm Indicator/Audio Pause button. The  
alarm is silenced for 60 seconds and then will sound again if the  
cause of the alarm has not been corrected. Each time the Alarm  
Indicator/Audio Pause button is pressed, the alarm silence period  
resets to one minute.  
When Audio Pause is active, the Alarm Indicator/Audio Pause symbol  
(
) appears if you are on the Monitor screen. Additionally, an  
“Audio Pausemessage displays in the menu banner on the Alarm  
Display screen.  
You can ‘pre-silencealarms that have not yet occurred by pressing  
the Alarm Indicator/Audio Pause button while no alarms are active.  
Then, if an alarm occurs, the audible indicator does not sound until  
the Audio Pause time limit has expired. This can be useful during  
patient setup, so caregivers can prevent alarms from sounding that  
they know will occur during the setup process. The LED and display  
will still show the alarm, but the audible alarm will not sound.  
Alarm Reset  
The Reset button is used to clear the currently active alarm(s) from  
the display and stop the LED and audible alarm indicator. This button  
should be selected after the situation causing the alarm(s) has been  
corrected. All active alarms are cancelled and alarm detection is  
restarted when this button is selected.  
The ventilator self-cancels certain alarms if the cause of the alarm is  
corrected, shutting off the alarm LED, the audible alarm, and the alarm  
background color. However, the alarm text remains on the screen. You  
can manually reset an alarm by pressing the Left button (Reset). The  
Audio Pause function is cancelled if the alarm is manually reset.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
106  
Alarm Volume Control  
WARNING  
As explained in Chapter 5, you can adjust the Alarm Volume from  
the Options menu. You can select Loud or Soft, depending on your  
preference.  
Make sure the alarm volume is  
set loud enough to be heard by  
the caregiver. Consider the use of  
a remote alarm.  
What to Do When An Alarm Occurs  
WARNING  
You should not rely on any  
single alarm to detect a circuit  
disconnect condition. The Low  
Tidal Volume, Low Minute  
Ventilation, Low Respiratory  
Rate, and Apnea alarms should  
be used in conjunction with the  
Circuit Disconnect alarm.  
Complete the following steps when an alarm occurs:  
1. Whenever an alarm occurs, first always observe the patient  
and ensure that adequate ventilation and oxygenation (if  
appropriate) are available.  
2. Look at the alarm indicators and listen to the audible alarm  
sound. Note the color of the Alarm Indicator/Audio Pause button  
(red or yellow) and whether the LED is solid or flashing.  
3. Look at the display to check the alarm message that appears on-  
screen and whether it is highlighted in red or yellow.  
4. Press the Alarm Indicator/Audio Pause button to temporarily  
silence the audible alarm. A visual indicator displays if you are on  
the Monitor screen (  
), or an “Audio Pausemessage appears in  
the menu banner on the Alarm Display screen.  
5. Look up the alarm in the alarm descriptions later in this chapter  
to determine the source of the alarm and the appropriate action.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
107  
Alarm Summary Table  
The following table summarizes all of the high, medium, and low priority alarms and informational  
messages.  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Loss of Power  
High  
Red flashing button;  
Blank screen  
Shuts down  
User: Press Start/Stop button  
and then press Right button  
to silence alarm. If using  
AC power, try plugging  
device into alternate AC  
power source. If loss of  
power continues, switch to  
DC power by connecting  
a fully charged detachable  
or external battery to the  
device. If there is still no  
power, connect patient  
to alternate source of  
ventilation and contact your  
health care professional.  
Clinician: Restore AC power.  
If alarm continues, connect a  
fully charged detachable or  
external battery to device to  
restore power. If alarm still  
continues, have the device  
serviced.  
Ventilator  
Inoperative  
High  
Red flashing  
button; “Ventilator  
Inoperativemessage therapy safely.  
Or, continues  
Shuts down if User: Press Start/Stop button.  
If display is operational,  
Power Off confirmation  
screen appears. Select  
Right button to shut off  
device and silence alarm.  
Immediately remove patient  
from ventilator and connect  
them to alternate source of  
ventilation. Contact your  
health care professional for  
service.  
can’t provide  
to operate at a  
limited level.  
Clinician: Restore power. If  
alarm continues, have the  
device serviced.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
108  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Ventilator  
Service  
Required  
High  
Red flashing button;  
“Ventilator Service  
Requiredmessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to silence  
alarm. Or, press Left (Reset)  
button to reset alarm. The  
alarm goes away until next  
reminder appears. Connect  
patient to alternate source of  
ventilation and contact your  
health care professional as  
soon as possible.  
• • • • •  
Clinician: If alarm continues,  
have the device serviced.  
Check Circuit  
High  
Red flashing button;  
“Check Circuit”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
• • • • •  
Left (Reset) button to reset  
alarm. Check patient circuit  
to make sure it is set up as  
your health care professional  
specified. Make sure there  
is no kinked or pinched  
tubing. Make sure tubing is  
properly attached. Make sure  
all tubing and exhalation  
devices are clean and free  
of condensation (refer to  
the Troubleshooting section  
– tubing condensation). If  
using Active PAP or Active  
Flow Exhalation Device, make  
sure diaphragm is inserted  
correctly and is not wrinkled  
or kinked. If alarm continues,  
contact your health care  
professional.  
Clinician: Verify patient  
status. Verify that breathing  
circuit setup is correct.  
Correct any errors. If alarm  
continues, have device  
serviced.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
109  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Low Circuit  
Leak  
High  
Red flashing button;  
“Low Circuit Leak”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Check for occlusions in  
• • • • •  
exhalation devices. Make sure  
exhalation device is clean  
and functioning properly. If  
alarm continues, contact your  
health care professional.  
Clinician: Remove  
obstruction in leak device.  
If alarm continues, have the  
device serviced.  
High Expiratory High  
Pressure  
Red flashing button;  
“High Expiratory  
Pressuremessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Make sure tubing is not  
kinked or pinched. Check  
patient’s breath rate. If alarm  
continues, contact your  
health care professional.  
Clinician: Check patient’s  
tubing to make sure it is not  
kinked or pinched. If alarm  
continues, have device  
serviced.  
Low Expiratory High  
Pressure  
Red flashing button;  
“Low Expiratory  
Pressuremessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to silence  
alarm. Or, press Left (Reset)  
button to reset alarm. Make  
sure tubing is not kinked or  
pinched. If alarm continues,  
contact your health care  
professional.  
• • • • •  
Clinician: Check patient’s  
tubing to make sure it is not  
kinked or pinched. If alarm  
continues, have device  
serviced.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
110  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
High Internal  
Oxygen  
High  
Red flashing button;  
“High Internal  
Oxygenmessage  
Continues  
to operate  
User: Press Alarm Indicator/  
• • • • •  
Audio Pause button to  
when internal silence alarm. Or, press Left  
oxygen  
(Reset) button to reset alarm.  
concentration Disconnect supplemental  
reaches  
5% above  
ambient  
levels.  
oxygen supply from the  
ventilator and contact your  
health care professional.  
Clinician: Disconnect  
supplemental oxygen from  
device. Check external  
oxygen connection. If  
problem continues, have  
device serviced.  
Circuit  
Disconnect  
High  
Red flashing button;  
“Circuit Disconnect”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Check breathing circuit and  
reconnect it if it has become  
disconnected, or fix the leak.  
If alarm continues to occur,  
contact your health care  
professional. If device will not  
exit circuit disconnect, switch  
to an alternate source of  
ventilation.  
Clinician: Reconnect tubing  
or fix leak. If alarm continues,  
have device serviced.  
Apnea  
High  
Red flashing button;  
“Apneamessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
using an Active PAP or Active  
Flow circuit, check proximal  
pressure line to ensure it is  
not pinched or has water  
condensation (refer to the  
Troubleshooting section  
– tubing condensation). If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
111  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
High Vte  
High  
Red flashing button;  
“High Vtemessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Low Vte  
High Vti  
Low Vti  
High  
High  
High  
Red flashing button;  
“Low Vtemessage  
Operates  
Operates  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
• • • • •  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Red flashing button;  
“High Vtimessage  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Red flashing button;  
“Low Vtimessage  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
112  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
High  
Respiratory  
Rate  
High  
Red flashing button;  
“High Respiratory  
Ratemessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Low Respiratory High  
Rate  
Red flashing button;  
“Low Respiratory  
Ratemessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
High  
Inspiratory  
Pressure  
Escalates  
Peak Pressure symbol Operates  
turns red. When  
condition first occurs,  
a beep will sound.  
When condition  
occurs for third time,  
button flashes yellow  
and yellow “High  
Inspiratory Pressure”  
message appears.  
When condition  
occurs for 10th  
time, button flashes  
red and red “High  
User: Press Alarm Indicator/  
Audio Pause button to silence  
alarm. Or, press Left (Reset)  
button to reset alarm. Make  
sure tubing is not pinched or  
blocked. If alarm continues,  
contact your health care  
professional.  
(for first two  
from  
consecutive  
Audible  
occurrences)  
(for 3rd  
indicator  
to Medium  
priority and  
then High  
priority  
consecutive  
occurrence)  
(for 10th  
consecutive  
occurrence)  
Clinician: Verify patient  
status. If problem continues,  
have device serviced.  
Inspiratory Pressure”  
message appears.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
113  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Low Inspiratory High  
Pressure  
Red flashing button;  
“Low Inspiratory  
Pressuremessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
• • • • •  
Left (Reset) button to reset  
alarm. Check patient circuit  
for leaks or disconnects. If  
using an Active PAP or Active  
Flow circuit, check proximal  
lines to ensure they are  
not pinched or have water  
condensation (refer to the  
Troubleshooting section  
– tubing condensation). If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status. If problem continues,  
have device serviced.  
High Minute  
Ventilation  
High  
Red flashing button;  
“High Minute  
Ventilationmessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Low Minute  
Ventilation  
High  
Red flashing button;  
“Low Minute  
Ventilationmessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • • • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Continue using device. If  
alarm continues, contact your  
health care professional.  
Clinician: Verify patient  
status.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
114  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Low Battery  
Escalates  
from  
Medium to  
High  
Medium Priority-  
Yellow flashing  
button. “Low  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to silence  
alarm. Or, press Left button  
(Reset) to reset alarm. Switch  
to an alternate battery. Or,  
switch to AC power source  
and recharge low battery.  
• • •  
(Medium -  
when approx.  
20 minutes  
remains)  
Detachable Battery,“  
“Low External Battery,”  
or “Low Internal  
Batterymessage  
appears in yellow,  
On Status Panel, box  
around battery is  
yellow  
Clinician: Switch to alternate  
battery or AC power while  
you recharge low battery.  
If low battery is recharged  
and alarm continues, replace  
battery.  
• • • • •  
(High - when  
approx. 10  
minutes  
High Priority-Red  
flashing button. “Low  
Detachable Battery,  
“Low External Battery,”  
or “Low Internal  
remains)  
Batterymessage  
appears in red. On  
Status panel, box  
around battery is red  
High  
Temperature  
Escalates  
from  
Medium to  
High  
Yellow flashing  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Make sure device is not  
close to a heat source. Make  
sure tubing is not under any  
bedding. Check inlet filter  
and clean or replace it if  
needed. Make sure cooling  
vents are not blocked. If the  
ventilator is running off of  
the internal or detachable  
battery, move to a cooler  
location and/or power the  
device with AC power or a  
lead-acid battery.  
• • •  
for  
Medium  
button and yellow  
“High Temperature”  
message appears. If  
condition worsens,  
button flashes  
red and red “High  
Temperature”  
message appears.  
• • • • •  
for High  
If alarm continues, place  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: If alarm continues  
after the measures above  
have been taken, have device  
serviced.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
115  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Replace  
Detachable  
Battery  
Low or  
High,  
depending  
on cause of  
alarm  
“Replace Detachable  
Batterymessage  
appears. If battery  
is nearing end of  
useful life, message  
appears with yellow  
background and  
Operates  
User: Press Alarm Indicator/  
• •  
for Low•  
Audio Pause button to silence  
alarm. Or, press Left button  
(Reset) to reset alarm. Switch  
to an alternate battery or AC  
power source while replacing  
current detachable battery.  
for High  
button is solid  
Clinician: Switch to an  
yellow. If battery fails,  
message appears with  
red background and  
button flashes red.  
alternate battery or AC power  
source while replacing the  
current detachable battery.  
Ventilator  
Service  
Recommended  
Medium  
Yellow flashing  
button;  
“Ventilator Service  
Recommended”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
• • •  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Alarm goes away until next  
reminder appears. Contact  
your health care professional  
to report alarm and have  
device serviced.  
Clinician: If alarm continues,  
have device serviced.  
AC Power  
Disconnected  
Medium  
Yellow flashing  
button; “AC Power  
Disconnected”  
message, and a box  
appears around  
battery in use.  
Switches to  
alternate  
User: Press Alarm Indicator/  
Audio Pause button to  
• • •  
power source silence alarm. Or, press  
Left (Reset) button to reset  
alarm. If you were using AC  
power, check power cord and  
reconnect it if it has become  
disconnected. Make sure  
device is not connected to  
an overloaded AC circuit. If  
you were using an external,  
detachable, or internal  
battery, recharge the battery.  
Clinician: Check batteries  
and recharge f necessary.  
Restore AC power if available.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
116  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Keypad Stuck  
Low  
Solid yellow button;  
“Keypad Stuck”  
message.  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Check the keys to determine  
if they are lodged in the case.  
If alarm continues, place  
• •  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: Check keys  
to determine if they are  
lodged in the case. If alarm  
continues, place patient  
on alternate source of  
ventilation and have device  
serviced.  
Battery  
Info  
“Batt Discharge  
Stopped – Temp.”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
Discharging  
Stopped Due to  
Temperature  
silence alarm. Or, press  
Left (Reset) button to reset  
alarm. Move device to a  
cooler location. Make sure  
device is not close to a heat  
source. Make sure cooling  
vents are not blocked. If  
alarm is detected for internal  
battery and continues, place  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: If alarm is detected  
on the internal battery and  
continues after the measures  
above have been taken, place  
patient on alternate source of  
ventilation and have device  
serviced.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
117  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Battery Not  
Charging Due  
to Temperature  
Info  
“Batt Not Charging –  
Temp.message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
Left (Reset) button to reset  
alarm. Make sure device is  
not close to a heat source.  
Make sure cooling vents are  
not blocked. Move device to  
a cooler location. If device  
is too cold, allow it to warm  
up. If alarm continues, place  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: If alarm continues  
after the measures above  
have been taken, place  
patient on alternate source of  
ventilation and have device  
serviced  
Battery Not  
Charging  
Info  
“Detach Battery Not  
Chargingor “Internal  
Battery Not Charging”  
message  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
Left (Reset) button to reset  
alarm. If condition continues  
for internal battery, place  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: If condition  
continues for internal battery,  
place patient on alternate  
source of ventilation and  
have device serviced.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
118  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Check External Info  
Battery  
“Check External  
Batterymessage  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
Left (Reset) button to reset  
alarm. Check connection to  
external battery. Replace  
external battery with another  
external battery, if available.  
If alarm continues, place  
patient on alternate source of  
ventilation and contact your  
health care professional.  
Clinician: If the alarm  
continues after the measures  
above have been taken, place  
patient on alternate source of  
ventilation and have device  
serviced.  
Battery  
Depleted  
Info  
“External Battery  
Depletedor  
“Detachable Battery  
Depletedmessage  
appears  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press Left  
(Reset) button to reset alarm.  
Replace depleted battery  
with another or switch to AC  
power, if available.  
Clinician: Replace depleted  
battery with another or  
switch to AC power.  
External Battery Info  
Disconnected  
“External Batt  
Switches to  
alternate  
power source  
User: Check connection  
of the external battery to  
ventilator. Check the charge  
available on external battery  
and recharge battery if  
necessary.  
Disconnected”  
message, and a box  
appears around  
battery in use.  
Clinician: Check external  
battery and recharge if  
necessary. Restore AC power  
if available.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
119  
Alarm  
Priority  
Audible  
Visual (Alarm  
Indicator Button  
and Display)  
Device  
Action  
User/Clinician Action  
Detachable  
Battery  
Disconnected  
Info  
“Detachable Batt  
Disconnected”  
message, and a box  
appears around  
battery in use.  
Switches to  
alternate  
power source  
User: Check connection of  
the detachable battery to  
ventilator. Check the charge  
available on detachable  
battery and recharge battery  
if necessary.  
Clinician: Check detachable  
battery and recharge if  
necessary. Restore AC power  
if available.  
Start On Battery Info  
“Start On Battery”  
message appears.  
Operates  
User: Check battery status  
and determine how long  
device can operate on  
battery power. Find AC power  
source as soon as possible.  
Clinician: Restore AC power  
when available. If alarm  
continues, replace AC power  
cord to determine if there is a  
problem with the cord.  
Card Error  
Info  
A “Card Error”  
message appears.  
Operates  
User: Press Alarm Indicator/  
Audio Pause button to  
silence alarm. Or, press  
Left button (Reset) to reset  
alarm. Remove SD Card  
and use another card, if  
available. Check the write-  
protect switch on the card.  
If condition persists, contact  
your health care professional.  
Clinician: Remove SD Card  
and use another card, if  
available. Ensure card meets  
specifications. Check write-  
protect switch on card. If  
condition persists, contact an  
authorized representative of  
Philips Respironics.  
Chapter 6 Ventilator Alarms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
120  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
7. Cleaning and Maintenance  
Cleaning the Ventilator  
WARNING  
The ventilator’s exterior surface and the exterior of the detachable  
battery pack (if using) should be cleaned before and after each  
patient use, and more often if needed.  
To avoid electrical shock, always  
unplug the power cord from the  
wall outlet before cleaning the  
ventilator.  
1. Unplug the device and clean the front panel and exterior of the  
enclosure as needed using a clean cloth dampened with any of  
the following cleaning agents:  
CAUTION  
Do not immerse the device or  
allow any liquid to enter the  
enclosure or the inlet filter.  
Water  
Soapy water or a mild detergent  
Hydrogen Peroxide (3%)  
Isopropyl Alcohol (91%)  
CAUTION  
Do not use harsh detergents,  
abrasive cleaners, or brushes to  
clean the ventilator system. Use  
only the cleaning agents and  
methods described in this manual.  
10% bleach solution (10% bleach, 90% water)  
2. Do not allow any liquid to drip into the ventilator case or  
detachable battery pack. After cleaning, use a soft, dry cloth to  
remove any residual cleaner. Use extra care when cleaning the  
display. Abrasive cleaners can scratch the display.  
3. Allow the device to dry completely before plugging in the power  
cord.  
Chapter 7 Cleaning and Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
122  
Cleaning and Replacing the Air Inlet Filter  
Under normal usage, you should clean the gray foam filter at least  
once every two weeks and replace it with a new filter every six  
months or sooner if needed.  
CAUTION  
1. If the device is operating, stop the airflow by pressing the  
button. Disconnect the device from the power source.  
The reusable foam inlet filter is  
required to protect the ventilator  
from dirt and dust. Wash  
periodically and replace when  
damaged for proper operation.  
2. Remove the filter from the enclosure by gently squeezing the  
filter in the center and pulling it away from the device, as shown  
below.  
Removing the Air Inlet Filter  
Note: Never install a wet  
filter into the device. It is  
3. Examine the filter for cleanliness and integrity.  
recommended that you clean  
the filter in the morning and  
alternate using the two foam  
filters provided with the system  
to ensure sufficient drying time  
for the cleaned filter.  
4. Wash the gray foam filter in warm water with a mild detergent.  
Rinse thoroughly to remove all detergent residue. Allow the filter  
to air dry completely before reinstalling it. If the foam filter is torn  
or damaged, replace it. Only Philips Respironics-supplied filters  
should be used as replacement filters.  
5. Reinstall the filter as shown below.  
Reinstalling the Filter  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
123  
Replacing the Air Inlet Path Foam  
Follow your institution’s protocol for how often you replace the air  
inlet path foam. To replace the foam, complete the steps below.  
1. Using a screwdriver, remove the air inlet path cover by  
unscrewing the four screws as shown below.  
Removing the Air Inlet Path  
Cover  
2. Remove the gray foam air inlet path by hand. The removed air  
inlet path is shown below.  
Air Inlet Path  
Compartment  
Air Inlet Path Cover  
Removing the Air Inlet Path  
Foam  
Air Inlet Path Foam  
Chapter 7 Cleaning and Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
124  
3. Insert the new air inlet path foam in the slot, lining the large  
circular cutout on the foam up with the circular connection at  
the top of the air inlet path compartment on the device. Make  
sure the air inlet path foam is securely in place.  
4. Replace the air inlet path cover, using a screwdriver to secure each  
of the four screws. Make sure all of the screws are firmly in place.  
Cleaning the Patient Circuit  
Cleaning the reusable circuit is important in the hospital and in the  
home. Circuits infected with bacteria may infect the user’s lungs.  
Clean the respiratory circuit on a regular basis. If you are using a  
disposable circuit, dispose of and replace it on a regular basis.  
Follow your institution’s protocol for cleaning the circuit. Philips  
Respironics recommends that you perform the cleaning twice a  
week under normal conditions and more frequently as required.  
If the patient uses the device on a 24 hour a day basis, it may be  
convenient to have a second breathing circuit so you can switch  
circuits while one is being cleaned.  
CAUTION  
Exhalation devices, patient  
circuits, and water traps are  
shipped clean, not sterile.  
Cleaning and disinfection  
of these parts should follow  
individual institution processes  
and conform to guidelines  
provided by Philips Respironics  
with each accessory.  
Cleaning Instructions (Reusable Circuits)  
Clean the patient circuit twice a week, or follow your institution’s  
protocol.  
1. Disconnect the circuit from the device, and disassemble the  
circuit for cleaning. Thoroughly wash your hands.  
2. Using a mild detergent, such as liquid dishwashing soap, clean  
all accessible surfaces of the circuit. Do not clean using alcohol.  
3. Rinse the circuit with tap water, removing all remaining detergent.  
4. Prepare a solution of one part white vinegar to three parts  
distilled water. An average beginning quantity is 16 ounces  
vinegar to 48 ounces distilled water. The actual amount will vary  
according to individual needs. Regardless of the quantity the  
ratio must remain 1 part:3 parts. Soak the circuit in this solution  
for one hour. Rinse the circuit completely with tap water.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
125  
5. Place the circuit on a clean towel to dry. Do not wipe dry. The  
circuit must be completely dry before storing.  
6. Reassemble the circuit when dry. Store in a plastic bag or dust-  
free area.  
7. Inspect components for deterioration prior to use.  
Cleaning the Exhalation Devices  
Disassemble the exhalation device from the patient circuit. Follow  
the detailed cleaning instructions included with your exhalation  
device.  
Replacing the Bacteria Filter (Optional)  
To remove a bacteria filter and replace it with a new one:  
1. Disconnect the flexible tubing from the bacteria filter by gently  
pulling straight out from the filter. The proximal tubing (if using  
the Active PAP or Active Flow Exhalation Device) can remain  
connected to the ventilator.  
WARNING  
Philips Respironics recommends  
that a main line outlet bacteria  
filter (Part Number 342077) be  
used whenever the device is  
used for invasive therapy or if  
the ventilator may be used on  
multiple patients.  
2. Remove the used bacteria filter from the ventilator by  
gently pulling straight out from the connector. Follow the  
manufacturer’s labeling and your institution’s infection control  
guidelines for disposal of the bacteria filter.  
3. Attach a new bacteria filter. The openings on either end of the  
bacteria filter are different sizes. Fit the larger opening onto the  
gas outlet port on the side of the ventilator as shown.  
Chapter 7 Cleaning and Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
126  
4. Connect the patient circuit. Fit the ventilator-end of the flexible  
tubing onto the bacteria filter.  
Installing a New Bacteria  
Filter  
Preventive Maintenance  
Your new Philips Respironics Trilogy ventilator includes a blank PM  
Service Due label affixed to the bottom of your device. The label is  
blank, which allows you to document when service is due. Service  
is due every 10,000 hours or 24 months, whichever comes first  
depending on the usage of the device.  
Service Date  
Service Hours  
Preventive Maintenance  
Sticker  
You will need to record both the date and hours on the blank label  
to maximize your in-service interval. Record the service date as 24  
months from the date first placed on a patient. Record the hours as  
10,000 hours for initial usage.  
Use the device blower hours to determine when service is required.  
The device blower hours are listed in the device software in the  
information menu.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
127  
The following table provides information on recommended periodic maintenance for your device.  
Time / Hours of Service  
Recommended Maintenance  
Prior to initial use on a patient  
Charge Internal and Detachable Batteries to 100%  
capacity by plugging ventilator into an AC power  
source for up to 8 hours.  
Perform System Checkout Procedure in Chapter 10.  
Prior to long term storage after  
initial use  
Charge both Internal and Detachable batteries to  
100% capacity prior to storage.  
While in storage after initial  
connection to AC Power, every 3  
months  
Recharge internal and detachable batteries to  
100% capacity by plugging ventilator into an AC  
power source. Batteries should recharge in 8 hours  
or less.  
If in use, every two weeks  
If in use, every 6 months  
Inspect and clean air inlet filter.  
Replace air inlet filter.  
Inspect enclosures and external connections for  
damage and contact Philips Respironics service if  
necessary.  
Inspect power cord for damage and replace if  
necessary.  
Every 10,000 hours or 2 Years,  
whichever comes first  
Perform performance verification test per Trilogy  
Service Manual by Philips Respironics trained  
service technician.  
Every 17,500 hours  
Replace the Motor / Blower Assembly by Philips  
Respironics trained service technician.  
Perform full calibration and verification by Philips  
Respironics trained service technician.  
Refer to the Trilogy Service Manual for additional information.  
Chapter 7 Cleaning and Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
128  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
8. Troubleshooting  
The ventilator is designed to respond to most issues with an appropriate alarm or information  
message. These messages appear on the display screen. Chapter 6 describes many of the alarms  
and what action you should take when the alarms occur.  
This chapter provides additional troubleshooting information for other common issues you may  
have.  
Issue  
Why It Happened  
What To Do  
Device does not turn on.  
Nothing happens when you  
press Start/Stop to begin  
therapy. The audible indicator Note: The device requires AC  
does not sound and the power to charge the internal  
backlights on the buttons do battery when a new device is  
not light. powered up for the first time.  
The AC power cord is not  
plugged in and the internal  
battery is not charged.  
Plug the AC power cord into the  
device and a wall outlet that is  
not controlled by a switch. This  
powers the system and charges  
the internal battery. If the device  
still does not turn on, contact an  
authorized service representative  
or Philips Respironics to have the  
device serviced. Please have the  
model number and serial number  
ready when you call.  
Chapter 8 Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
130  
Issue  
Why It Happened  
What To Do  
The batteries are not  
charging.  
The batteries may not charge Make sure the device is not close  
if the device is too hot or too to a heat source. Ensure the  
The device is plugged in, but cold or is operating at an  
cooling air vents are not blocked.  
the detachable (if installed) ambient temperature outside Bring the ventilator to ambient  
and internal batteries are not of the specified valid range.  
showing full capacity and are The device cooling fan may  
room temperature. If the problem  
continues, contact an authorized  
service representative or Philips  
Respironics to have the device  
serviced. Please have the model  
number and serial number ready  
when you call.  
not charging.  
not be working properly. If  
this continues for more than  
15 minutes, an error will  
occur.  
Unexplained changes in the The device has been dropped Press the Start/Stop button and  
performance of the device  
The device is not working  
properly or is making unusual  
sounds.  
or mishandled, or is in an area then select the Right button  
of high EMI emissions.  
to shut off the device for a few  
seconds. Restart the device by  
pressing the Start/Stop button. If  
the problem continues, relocate  
the device to an area with lower  
EMI emissions (e.g., away from  
electronic equipment such as  
cellular phones, cordless phones,  
computers, TVs, electronic  
games, hair dryers, etc.). Refer to  
Chapter 13. If the problem still  
occurs, contact an authorized  
service representative or Philips  
Respironics to have the device  
serviced. Please have the model  
number and serial number ready  
when you call.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
131  
Issue  
Why It Happened  
What To Do  
Excessive Warm Air Out of  
Mask  
The air filter may be dirty. Or, Clean or replace the air filter as  
the device may be operating described in Chapter 7.  
The air out of the mask is  
much warmer than usual.  
in direct sunlight or near a  
heater.  
The temperature of the air  
may vary based on the room  
temperature. Make sure the  
device is properly ventilated.  
Keep it away from bedding or  
curtains that could block the  
flow of air around the device.  
Make sure it is away from direct  
sunlight and heating equipment.  
Ensure the cooling air vents  
are not blocked. If the problem  
continues, contact an authorized  
service representative or Philips  
Respironics to have the device  
serviced. Please have the model  
number and serial number ready  
when you call.  
Mask Discomfort  
The mask feels  
This could be due to  
improper headgear  
Make sure the patient is properly  
fitted with the correct size mask.  
uncomfortable to wear, there adjustment or improper mask If the problem continues, try  
is significant air leakage  
around the mask, or the  
patient experiences other  
mask-related issues.  
fitting.  
fitting the patient with a different  
mask.  
Runny Nose  
This may be caused by a nasal Call the patient’s health care  
The patient has a runny nose. reaction to the airflow. This  
only applies to non-invasive  
professional.  
therapy and will not occur  
during invasive therapy.  
Chapter 8 Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
132  
Issue  
Why It Happened  
What To Do  
Dryness  
The patient has throat or  
nose dryness.  
The air is too dry. This only  
applies to non-invasive  
therapy and will not occur  
during invasive therapy.  
Increase the room humidity.  
Consult with the patient’s health  
care professional about using  
a humidifier with the device. If  
using a humidifier, refer to the  
instructions included with the  
humidifier to make sure it is  
working properly.  
Sinus or Ear Pain  
The patient may have a sinus Tell the patient to stop using the  
The patient experiences  
nasal, sinus, or ear pain.  
or middle ear infection. This  
only applies to non-invasive  
therapy and will not occur  
during invasive therapy.  
device and contact their health  
care professional.  
Patient Tubing Condensation Moisture added to the circuit  
There is condensation in the from a heated humidifier may  
Use an appropriate method of  
water management, such as a  
circuit tubing or exhalation  
valve lines.  
condense. This condensation is water trap when using a heated  
typically referred to asrain-out. humidifier.  
Exhalation Valve or Flow  
Excessive humidity (from  
Remove tube(s) with  
Tube Condensation (Active the heated humidifier, the  
condensation: the exhalation  
valve line must be removed from  
both ends of the circuit; the flow  
tube lines must be removed  
from the ventilator side. The  
tubes are permanently attached  
to the flow element, so the flow  
element must be removed from  
the patient circuit. If there is  
water built up in the exhalation  
diaphragm, disassemble the  
exhalation valve and clean/dry  
the assembly. Clear the flow  
tubes of water droplets using  
a low flow air source and re-  
assemble the circuit.  
Flow Circuit only)  
patient, or the cleaning  
process) condensed in the  
tubing or exhalation valve  
diaphragm. Droplets of  
water affected the ability of  
the ventilator to measure  
proximal patient flow and  
pressure and trigger on a  
patient breath.  
There is condensation  
in the proximal flow or  
exhalation valve lines, and  
the condensation is causing  
the system to alarm or not  
properly trigger on patient  
effort.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
9. Accessories  
There are several accessories you can use with the ventilator.  
Adding a Humidifier  
Use of a humidifier may reduce nasal dryness and irritation by adding  
moisture to the airflow.  
If you are using a humidifier with the ventilator, refer to the  
instructions provided with the humidifier for complete setup  
information.  
Adding Supplemental Oxygen to the Device  
The delivered oxygen concentration varies with changes in flow  
in the circuit. The following may have an impact on oxygen  
concentration:  
Pressure settings  
Patient Tidal Volume  
Peak Inspiratory Flow  
I:E Ratio  
Respiratory rate  
Circuit leak rate  
Oxygen flow rate  
To add oxygen to the circuit, the oxygen supply must comply with the  
local regulations for medical oxygen. The oxygen flow into the oxygen  
valve cannot exceed 15 l/min and the pressure cannot exceed 50 psi.  
Chapter 9 Accessories  
Download from Www.Somanuals.com. All Manuals Search And Download.  
134  
Supplemental Oxygen Warnings  
When administering fixed-flow supplemental oxygen,  
the oxygen concentration may not be constant. The  
inspired oxygen concentration will vary, depending on  
the pressures, patient flows and circuit leak. Substantial  
leaks may reduce the inspired oxygen concentration  
to less than the expected value. Appropriate patient  
monitoring should be used, as medically indicated, such  
as an alarming pulse oximeter.  
This device DOES NOT alarm for loss of the low flow  
oxygen supply.  
Oxygen supports combustion. Oxygen should not be  
used while smoking or in the presence of an open flame.  
Do not connect the device to an unregulated or high  
pressure oxygen source.  
The device may result in incorrect flow and tidal volume  
measurements and improper operation of related alarms  
if you add low flow oxygen directly into the patient  
circuit or mask instead of directly adding it into the  
oxygen inlet on the back of the ventilator.  
If oxygen is used with the device, the oxygen flow must  
be turned off when the device is not in use. Explanation  
of the Warning: When the device is not in operation  
and the oxygen flow is left on, oxygen delivered into the  
tubing may accumulate within the device’s enclosure.  
WARNING  
The remote alarm should be  
tested daily. When using a  
remote alarm, make sure you  
fully test the remote alarm  
connector and cable by verifying  
that:  
The ventilator should not be operated in the presence of  
flammable gasses. This could cause a fire or explosion.  
Using a Remote Alarm Unit  
-
Annunciated alarms on  
the ventilator are also  
annunciated on the remote  
alarm.  
You can use a Philips Respironics Remote Alarm unit with your  
device. There is a dedicated adapter cable assembly for connecting  
the device to the Remote Alarm unit (REF 1045291). Refer to the  
instructions included with your Remote Alarm unit and adapter cable  
assembly if you have questions on how to use it with device.  
-
Disconnecting the remote  
alarm cable from the  
ventilator or from the remote  
alarm results in an alarm  
notification at the remote  
alarm.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
135  
Using a Nurse Call System  
WARNING  
Ensure that the Nurse Call  
You can use an institutional Nurse Call system with your device.  
There are several Philips Respironics cables available to connect a  
nurse call system to the device. Refer to the instructions included  
with your cable assembly if you have questions on how to connect it  
to the device.  
systems used do not exceed SELV  
(Safety Extra Low Voltage) levels  
as described in IEC 60601-1.  
SELV levels do not exceed a  
nominal value of 25VAC or  
60VDC at rated supply voltage  
on the transformer or converter,  
between conductors in an earth-  
free circuit which is isolated from  
the supply mains by a safety  
extra low voltage transformer  
or by a device with equivalent  
separation.  
WARNING  
The Trilogy’s Nurse Call feature  
is for use only in a medically  
supervised environment.  
WARNING  
Do not rely on the audible  
indicator provided by the Nurse  
Call system as the primary  
indicator of the operating  
state of the device or of patient  
events.  
Using a Secure Digital (SD) Card  
An SD card is provided with the device to record device usage  
information. You can insert the SD card into the SD Data Card slot on  
the left side of the device. Make sure the label on the SD card faces  
the back of the ventilator.  
WARNING  
To insert the SD card:  
The Nurse Call feature should  
be considered a back-up to the  
Trilogy device’s primary alarm  
system. Do not rely solely on the  
Nurse Call feature.  
1. Open the SD card door by sliding the door forward and then  
pulling it out, as shown below.  
Opening the SD Card Door  
Chapter 9 Accessories  
Download from Www.Somanuals.com. All Manuals Search And Download.  
136  
2. Push the SD card completely into the card slot until the card  
Note: The SD card does not  
need to be installed for the  
device to work properly.  
clicks into place.  
Inserting the SD Card  
Note: Use only SD cards  
available from Philips  
Respironics or the type listed in  
the specifications section of this  
manual.  
To remove the SD card, select the “Safely Remove SD Cardoption  
from the main menu. After the “Remove SD Cardconfirmation  
message appears, remove the card by pushing the card in until the  
card clicks to release. The card will be slightly ejected from the device.  
Using the Philips Respironics DirectView  
Software  
You can use the Philips Respironics DirectView software to download  
the prescription data from the SD card to a computer. DirectView  
can be used by clinicians to receive and report stored data from  
the SD card. DirectView does not perform any automatic scoring or  
diagnosing of a patient’s therapy data.  
Note: Remote prescription changes are not permitted.  
Using the Optional In-Use Bag  
An optional, in-use bag is included for use with the ventilator. The  
in-use bag is required for portable use in order to properly secure  
the ventilator. See the instructions included with the in-use bag for  
details on how to use the bag with your device.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
137  
Traveling with the System  
When traveling, the carrying case is for carry-on luggage only. The  
carrying case will not protect the system if it is put through checked  
baggage.  
For your convenience at security stations, there is a note on the  
bottom of the device stating that it is medical equipment and is  
suitable for airline use. It may be helpful to bring this manual along  
with you to help security personnel understand the device.  
If you are traveling to a country with a line voltage different than  
the one you are currently using, a different power cord or an  
international plug adaptor may be required to make your power cord  
compatible with the power outlets of the country to which you are  
traveling.  
Note: The device is not  
suitable for airline use with any  
humidifiers or modems installed  
in the unit.  
Airline Travel  
The device is suitable for use on airlines when the device is operating  
from an internal, detachable, or external battery power source. It is  
not suitable for airline use when operating from AC power or from  
DC power sourced from the airplane.  
Chapter 9 Accessories  
Download from Www.Somanuals.com. All Manuals Search And Download.  
138  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
10. System Checkout Procedures  
This chapter details the test procedures that should be performed by  
the clinician prior to connecting the device to the patient. Test the  
Note: The actual circuit  
configuration to be used on  
the patient should be used to  
perform the system checkout  
procedure.  
Active PAP, Active Flow, and Passive circuit types if you want to do a  
complete checkout on the device. The tests should be performed as  
described in order to verify proper operation of the device. Some of  
the procedures in this chapter require you to change settings on the  
device. If you are not familiar with the procedures for accomplishing  
this, please refer to Chapter 4, Ventilator Setup and Chapter 5, Viewing  
and Changing Settings.  
Tools Required  
Universal Porting Block  
Active PAP Exhalation Device (with Tube Adapter  
connected)  
Active Flow Exhalation Device with Flow Sensor  
Passive Exhalation Device  
Test Lung  
WARNING  
If you notice any unexplained  
changes in the performance  
of the device, if it is making  
unusual sounds, if the device  
or detachable battery are  
Small Flat Head Screwdriver  
dropped, if water is spilled into  
Visual Inspection  
the enclosure, or if the enclosure  
is cracked or broken, discontinue  
use and contact Philips  
Respironics or an authorized  
service center for service.  
1. Verify that the enclosure is not broken and that all applicable  
screws are in place.  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
140  
2. Verify that the device handle, SD Card door, and detachable  
battery are secure and in good working order.  
3. Verify that the rubber feet are on the bottom of the device.  
Initial Setup  
1. Connect the power cord to the device and then to an AC outlet.  
2. Attach the test lung to the patient connection end of the desired  
circuit (Active PAP, Active Flow, or Passive).  
3. Follow the instructions in Chapter 5 to access the Setup Screen.  
Settings and Alarms Tests  
Complete the following steps to set up the settings and alarms tests.  
1. Setup  
A. Settings And Alarms Menu  
Modify the settings in the Settings and Alarms menu to match  
those shown below in Table 1. If necessary, refer to Chapter 5 for  
instructions on modifying ventilator settings.  
Setting  
Dual Prescription  
Circuit Type  
Value  
Off  
Active PAP, Active Flow or  
Passive  
Therapy Mode  
S/T  
AVAPS (passive circuit only) Off  
IPAP  
20 cm H2O  
EPAP  
4 cm H2O  
12 BPM  
Table 1 - Ventilator Settings  
in the Settings and Alarms  
Menu  
Breath Rate  
Inspiratory Time  
1.6 seconds  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
141  
Setting  
Value  
Trigger Type (passive circuit Auto-Trak  
type)  
Flow Trigger Sensitivity  
(Active PAP or Active Flow  
circuit type)  
6.0 l/min  
Leak Compensation (Active On  
Flow circuit type)  
Flow Cycle Sensitivity  
(Active PAP or Active Flow  
circuit type)  
20%  
Rise Time  
1
Ramp Length  
All other alarms  
Off  
Off  
B. Options Menu  
Modify the settings in the Options menu to match those shown  
below in Table 2.  
Setting  
Menu Access  
Value  
Full  
Table 2 - Ventilator Settings  
in the Options Menu  
Detailed View  
On  
All other settings  
Discretionary  
C. Turn Device Power On  
Press the Start/Stop button on the front of the ventilator. The  
system will begin operating using the defined ventilation  
settings.  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
142  
2. Verify the High Tidal Volume Alarm  
This procedure verifies that the High Tidal Volume alarm is working  
properly. For Passive and Active Flow circuits, this will verify the High  
Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It  
assumes that you have attached the test lung, verified the ventilator  
settings, and turned on ventilator power as described in the Initial  
Setup section.  
A. Change Alarm Ventilator Setting  
Modify the High Tidal Volume alarm setting to match the one  
shown below in Table 3.  
Setting  
Value  
High Vte/High Vti  
50 ml  
Table 3 - High Vte/Vti Alarm  
Setting  
B. Verify the Alarm  
Wait up to 40 seconds and verify the following alarm signals:  
Note: Do not use the “Reset”  
button to manually reset the  
alarm. Instead, use the “Modify”  
button to change ventilator  
settings. This applies to all tests.  
The High Priority audible indicator sounds  
A red light flashes on the Alarm Indicator/Audio  
Pause button  
The High Tidal Volume alarm condition appears on  
the screen, highlighted in red  
C. Modify Ventilator Alarm Settings  
Modify the High Tidal Volume alarm setting to match the one  
shown below in Table 4.  
Setting  
Value  
Table 4 - Modify Vte/High Vti  
Alarm Setting  
High Vte/High Vti  
500 ml  
D. Verify Reset  
Wait 40 seconds and verify the following auto-reset conditions:  
The High Priority audible indicator has stopped  
sounding  
The red light on the Alarm Indicator/Audio Pause  
button has stopped flashing  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
143  
E. Restore Ventilator Settings  
Modify the ventilator settings and change the following value  
shown in Table 5.  
Setting  
Value  
High Vte/High Vti  
Off  
Table 5 - Restore Ventilator  
Settings  
3. Verify the Low Tidal Volume Alarm  
This procedure verifies that the Low Tidal Volume alarm is working  
properly. For Passive and Active Flow circuits, this will verify the Low  
Vte alarm. For Active PAP circuits, this will verify the Low Vti alarm. It  
assumes that you have attached the test lung, verified the ventilator  
settings, and turned on ventilator power as described in the Initial  
Setup section.  
A. Change Alarm Ventilator Setting  
Modify the Low Tidal Volume alarm setting to match the one  
shown below in Table 6.  
Setting  
Low Vte/Low Vti  
Value  
500 ml  
Table 6 - Low Vte/Low Vti  
Alarm Setting  
B. Verify the Alarm  
Wait up to 40 seconds and verify the following alarm signals:  
The High Priority audible indicator sounds  
A red light flashes on the Alarm Indicator/Audio  
Pause button  
The Low Tidal Volume alarm condition appears on the  
screen, highlighted in red  
C. Modify Ventilator Alarm Settings  
Modify the Low Tidal Volume alarm setting to match the one  
shown below in Table 7.  
Setting  
Low Vte/Low Vti  
Value  
Table 7 - Modify Low Vte/Low  
Vti Alarm Setting  
50 ml  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
144  
D. Verify Reset  
Wait 40 seconds and verify the following auto-reset conditions:  
The High Priority audible indicator has stopped  
sounding  
The red light on the Alarm Indicator/Audio Pause  
button has stopped flashing  
E. Restore Ventilator Settings  
Modify the ventilator settings and change the following value  
shown in Table 8.  
Setting  
Low Vte/Low Vti  
Value  
Table 8 - Restore Ventilator  
Settings  
Off  
4. Verify Circuit Disconnect Alarm  
This procedure verifies that the Circuit Disconnect alarm is working  
properly. It assumes that you have attached the test lung, verified  
ventilator settings, and turned on ventilator power as described in  
the Initial Setup section.  
A. Change Circuit Disconnect Ventilator Setting  
Modify the Circuit Disconnect ventilator setting to match the  
value shown below in Table 9.  
Setting  
Value  
Table 9 - Ventilator Settings  
Circuit Disconnect  
10 seconds  
Note: The Low Inspiratory or  
Low Expiratory Pressure Alarm  
may also be detected.  
B. Disconnect Test Lung  
Disconnect the test lung from the circuit.  
C. Verify the Alarm  
Wait approximately 10 seconds and verify the following alarm  
signals:  
The High Priority Audible Indicator sounds  
A red light flashes on the Alarm Indicator/Audio  
Pause button  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
145  
The Circuit Disconnect alarm condition appears on  
the screen, highlighted in red  
D. Reconnect Test Lung  
Reconnect the test lung to the circuit.  
E. Verify Reset  
Wait at least 40 seconds and verify the following auto-reset  
conditions:  
The High Priority audible indicator has stopped  
sounding  
The red light on the Alarm Indicator/Audio Pause  
button has stopped flashing  
F. Restore Ventilator Settings  
Modify the ventilator settings and change the following values  
shown below in Table 10.  
Setting  
Value  
Table 10 - Restore Ventilator  
Settings  
Circuit Disconnect  
Off  
5. Verify the High Inspiratory Pressure Alarm  
This procedure verifies that the High Inspiratory Pressure alarm  
is working properly. It assumes that you have attached the test  
lung, verified ventilator settings, and turned on ventilator power as  
described in the Initial Setup section.  
A. Change Ventilator Settings  
Modify the ventilator settings and change the following values  
shown below in Table 11.  
Setting  
Value  
Mode  
CV  
Table 11 - Ventilator Settings  
Tidal Volume  
Breath Rate  
500 ml  
12 BPM  
1.0 seconds  
Ramp  
Note: If this alarm is not reset  
within 3 occurrences, the alarm  
is elevated to High Priority, and  
the High Priority Indicators  
occur.  
Inspiratory Time  
Flow Pattern  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
146  
Setting  
Value  
PEEP  
Sigh  
4 cm H2O  
Off  
Circuit Disconnect  
Low Inspiratory Pressure  
High Inspiratory Pressure  
Apnea  
Off  
6 cm H2O  
10 cm H2O  
Off  
All other alarms  
Off  
B. Verify the Alarm  
Wait up to 40 seconds and verify the following alarm signals:  
The Medium Priority audible indicator sounds  
A yellow light flashes on the Alarm Indicator/Audio  
Pause button  
The High Inspiratory Pressure alarm condition  
appears on the screen, highlighted in yellow  
C. Modify Ventilator Alarm Settings  
Modify the High Inspiratory Pressure setting to match the one  
shown below in Table 12.  
Setting  
Value  
Table 12 - Modify High  
Inspiratory Pressure Alarm  
Setting  
High Inspiratory Pressure  
60 cm H2O  
D. Verify Reset  
Wait 40 seconds and verify the following auto-reset conditions:  
The Medium Priority audible indicator has stopped  
sounding  
The yellow light on the Alarm Indicator/Audio Pause  
button has stopped flashing  
6. Verify the Low Inspiratory Pressure Alarm  
This procedure verifies that the Low Inspiratory Pressure alarm  
is working properly. It assumes that you have attached the test  
lung, verified ventilator settings, and turned on ventilator power as  
described in the Initial Setup section.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
147  
A. Change Ventilator Settings  
Modify the ventilator settings and change the following values  
shown below in Table 13.  
Setting  
Value  
Mode  
CV  
Tidal Volume  
Breath Rate  
500 ml  
12 BPM  
1.0 seconds  
Ramp  
4 cm H2O  
Off  
Inspiratory Time  
Flow Pattern  
PEEP  
Sigh  
Circuit Disconnect  
Low Inspiratory Pressure  
High Inspiratory Pressure  
Apnea  
Off  
40 cm H2O  
60 cm H2O  
Off  
Table 13 - Ventilator Settings  
All other alarms  
Off  
B. Verify the Alarm  
Wait up to 40 seconds and verify the following alarm signals:  
The High Priority audible indicator sounds  
A red light flashes on the Alarm Indicator/Audio  
Pause button  
The Low Inspiratory Pressure alarm condition  
appears on the screen, highlighted in red  
C. Modify Ventilator Alarm Settings  
Modify the Low Inspiratory Pressure setting to match the one  
shown below in Table 14.  
Setting  
Value  
Table 14 - Modify Low  
Inspiratory Pressure Alarm  
Setting  
Low Inspiratory Pressure  
6 cm H2O  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
148  
D. Verify Reset  
Wait 40 seconds and verify the following auto-reset conditions:  
The High Priority audible indicator has stopped  
sounding  
The red light on the Alarm Indicator/Audio Pause  
button has stopped flashing  
Battery Function Verification  
Make sure the batteries are functioning properly and fully charged  
before patient use.  
1. Verify the Detachable and Internal (Lithium-Ion)  
Batteries Function  
A. Connect AC Power to the device and verify that the green AC  
LED on the front panel is lit.  
B. Verify that the detachable battery is properly installed.  
C. Turn the device on and verify that both the detachable and  
internal battery symbols appear on the display. Verify that if  
either battery is less than fully charged, the charge symbol will  
display on the respective battery.  
D. Disconnect the AC Power source from the device.  
Verify that the AC Power Disconnected alarm  
message appears on the display and the green AC  
LED is not lit. Press Reset.  
Verify that the detachable battery symbol shows the  
level of charge noted in the previous step and that  
the device continues to operate.  
Verify that the detachable battery symbol has a black  
box around it to indicate that it is in use.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
149  
E. Disconnect the detachable battery pack from the device.  
Verify that the Detach Batt Disconnected alarm  
message appears on the display. Press Reset.  
Verify that the internal battery symbol shows the  
same level of charge as noted in Step C and the  
device continues to operate.  
Verify that the internal battery symbol has a black  
box around it to indicate that it is in use.  
F. Reconnect the Detachable Battery and AC Power source.  
2. Verify the External Battery Function (If Available)  
A. Connect AC Power to the device and verify that the green AC  
LED is lit.  
B. Connect the external battery cable to the external battery and to  
the ventilator.  
C. Verify that the external battery symbol is shown on the display  
and some level of charge is present.  
D. Disconnect the AC Power source from the device.  
Verify that the AC Power Disconnected alarm  
message appears on the display and the green AC  
LED is not lit. Press Reset.  
Verify that the external battery symbol shows the  
level of charge as noted in the previous step and the  
device continues to operate.  
Verify that the external battery symbol has a black  
box around it to indicate that it is in use.  
E. Reconnect the AC Power source.  
Chapter 10 System Checkout Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
150  
Alarm and Event Log Clean-Up  
1. In the Setup Menu, select Alarm Log.  
A. Press Clear to clear the log file.  
B. Press Yes to confirm.  
C. Press Finish to complete.  
2. In the Setup Menu, select Event Log.  
A. Press Clear to clear the log file.  
B. Press Yes to confirm.  
C. Press Finish to complete.  
Results  
All portions of this checkout procedure should be completed prior  
to connection to the patient. If any of the tests fail to complete as  
indicated, if possible, correct the error, clear the alarm and resume  
testing. If correction of the failed portion is not possible, return the  
device to Philips Respironics or an authorized service center for  
service and repair.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
11. Technical Specifications  
Environmental  
Operating  
Storage  
Temperature  
Relative Humidity  
Atmospheric Pressure  
41° F to 104° F (5° C to 40° C)  
15 to 95% (non-condensing)  
110 kPa to 60 kPa  
-4° F to 140° F (-20° C to 60° C)  
15 to 95% (non-condensing)  
N/A  
The operating range for the charging of the internal and detachable batteries is 10° C to 30° C. The internal and detachable  
batteries will power the ventilator for the full operating range of 5° C to 40° C.  
Accuracies stated in this manual are based on specific environmental conditions. For stated accuracy, the environmental  
conditions are: Temperature: 20-30° C; Humidity: 50% relative; Altitude: nominally 380 meters.  
Physical  
Dimensions:  
Weight:  
6.6L x 11.2”W x 9.3H (16.68 cm L x 28.45 cm W x 23.52 cm H)  
Approximately 11 lbs (5 kg) with the detachable battery installed  
Standards Compliance  
This device is designed to conform to the following standards:  
IEC 60601-1: Medical electrical equipment - Part 1: General requirements for safety  
IEC 60601-1-2: General requirements for safety - Collateral standard: Electromagnetic  
compatibility - Requirements and tests  
ISO 10651-2-Lung ventilators for medical use -- Particular requirements for basic safety and  
essential performance -- Part 2: Home care ventilators for ventilator-dependent patients  
RTCA-D0160F section 21, category M; Emission of Radio Frequency Energy  
Chapter 11 Technical Specifications  
Download from Www.Somanuals.com. All Manuals Search And Download.  
152  
Electrical  
AC Voltage Source:  
Detachable Battery:  
100 to 240 VAC, 50/60 Hz, 2.1 A  
Voltage: 14.4 VDC  
Capacity: 4176 mAh  
Chemistry type: Lithium-Ion  
Internal Battery:  
Voltage: 14.4 VDC  
Capacity: 4176 mAh  
Chemistry type: Lithium-Ion  
Type of Protection Against Electric Shock:  
Degree of Protection Against Electric Shock:  
Degree of Protection against Ingress of Water:  
Mode of Operation:  
Class II/Internally Powered Equipment  
Type BF Applied Part  
Device: Drip Proof, IPX1  
Continuous  
Fuses:  
There are no user-replaceable fuses.  
Pressure  
Output:  
4 to 50 cm H2O  
SD Card and SD Card Reader  
Use only SD cards and SD card readers available from Philips Respironics or the following:  
SanDisk® Standard SD Card - 1.0 GB - REF SDSDJ-1024  
SanDisk ® Card Reader/Writer - SanDisk ImageMate - REF SDDR-99-A15  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
153  
Control Accuracy  
Parameter  
IPAP  
Range  
4 to 501 cm H2O2  
Accuracy  
Greater of 2 cm H2O or 8% of setting  
EPAP  
0 to 25 cm H O for Active Circuits  
4 to 25 cm H22O for Passive Circuits  
Greater of 2 cm H2O or 8% of setting  
CPAP  
PEEP  
4 to 20 cm H2O  
Greater of 2 cm H2O or 8% of setting  
Greater of 2 cm H2O or 8% of setting4  
0 to 25 cm H O for Active Circuits  
4 to 25 cm H22O for Passive Circuits  
Pressure  
4 to 50 cm H2O  
0 to 30 cm H2O3  
50 to 2000 ml5  
Greater of 2 cm H2O or 8% of setting  
Greater of 2 cm H2O or 8% of setting4  
Pressure Support  
Tidal Volume  
Greater of 15ml or 10% of setting (Active  
Circuits)  
Greater of 15ml or 15% of setting (Passive  
Circuits)  
Breath Rate  
0 to 60 BPM for AC mode  
1 to 60 BPM for all other modes  
Greater of 1 BPM or 10% of the setting  
Timed Inspiration  
Rise Time  
0.3 to 5.0 seconds  
1 to 66  
.1 second  
2 cm H O (the device will increase to a  
pressure2of .67* (IPAP - EPAP) 2 cm H2O @ the  
set rise time multiplied by 100 ms for pressure  
supports less than or equal to 25.)  
Ramp Start Pressure  
0 to 25 cm H O for Active Circuits  
4 to 25 cm H2O for Passive Circuits  
4 to 19 cm H22O in CPAP mode  
8% of setting + 2% Full Scale  
Ramp Length  
Flex  
Off, 5 to 45 minutes  
Off, 1 to 37  
2 minutes  
N/A  
Flow Trigger Sensitivity  
Flow Cycle  
1 to 9 l/min  
N/A  
10 to 90%  
N/A  
Apnea Rate  
4 to 60 BPM  
Greater of 1 BPM or 10% of setting  
Specifications listed are based on using a standard circuit consisting of 1.8 meter tubing- REF 622038, one of the following exhalation  
devices: Whisper Swivel II - REF 332113; Active PAP Exhalation Device - REF 1053716; Active Flow Exhalation Device - REF 1049503 or  
Active Exhalation Device – REF 1065658; UPB Active PAP Exhalation Device - REF 1073862 and  
Trilogy Proximal Flow Sensor- REF 1050408 (if required).  
1Limited to 25 cm H2O when using the Bi-Flex feature in S mode.  
2Pressure units may be cm H2O, hPa, or mBar depending on device setup.  
3The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 30 cm H2O.  
4Pressure Support and PEEP not to exceed 50 cm H2O.  
5All flows and volumes are measured at BTPS conditions.  
6The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).  
7Flex is not available when AVAPS is active. Flex is not available with Active Circuits.  
Chapter 11 Technical Specifications  
Download from Www.Somanuals.com. All Manuals Search And Download.  
154  
Measured Patient Parameters  
Parameter  
Range  
Accuracy  
Vte/Vti  
0 to 2000 ml  
0 to 99 l/min  
Greater of 15ml or 15% of reading  
Minute Ventilation  
Calculation based on measured Vte or Vti  
and Respiratory Rate  
Estimated Leak Rate  
Respiratory Rate  
0 to 200 l/min  
0 to 80 BPM  
0 to 200 l/min  
0 to 99 cm H2O  
0 to 99 cm H2O  
0 to 100%  
N/A  
Greater of 1 BPM or 10% of reading  
3 l/min plus 15% of reading  
Greater than 2 cm H2O or 10% of reading  
Greater than 2 cm H2O or 10% of reading  
N/A  
Peak Inspiratory Flow  
Peak Inspiratory Pressure  
Mean Airway Pressure  
% Patient Triggered Breaths  
I:E Ratio  
9.9- 1: 1-9.9  
Calculation based on Inspiratory Time and  
Expiratory Time  
All flows and volumes are measured at BTPS conditions.  
Spontaneous Breathing During Power Failure Conditions  
Flow Set  
Inspiratory Resistance (cm H2O)  
Expiratory Resistance (cm H2O)  
Point (l/min)  
Active Active Circuit with Passive  
Active Active Circuit with Passive  
Circuit  
Proximal Flow  
Sensor  
Circuit  
Circuit  
Proximal Flow  
Sensor  
Circuit  
30  
60  
< 3.0  
< 9.0  
< 3.5  
< 2.0  
< 5.0  
< 2.0  
< 3.0  
< 2.0  
< 4.5  
< 1.5  
< 4.0  
< 10.5  
Disposal  
Dispose of this device and batteries in accordance with local regulations.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
12. Glossary  
The following terms and acronyms appear throughout this manual.  
Term/Acronym Definition  
AC Power  
Alternating Current power  
AC  
Assist Control therapy mode; AC mode delivers Assisted and  
Mandatory breaths with a user-defined inspired tidal volume. This  
is a Volume Control mode.  
Apnea  
A condition marked by the temporary cessation of spontaneous  
breathing.  
Assisted Breath  
Breath type in which the breath is patient-triggered and the  
ventilator controls how the air is delivered based on defined  
settings. An assisted breath is not cycled until the inspiratory  
time setting has been reached.  
AVAPS  
Average Volume Assured Pressure Support feature (available in S,  
S/T, T, and PC therapy modes).  
Blower Hours  
The total number of hours that the blower has been on over the  
life of the device. This value helps determine when the ventilator  
needs to be serviced. You cannot reset this value. It can only be  
reset by an authorized service center.  
BPM  
Breaths Per Minute  
Chapter 12 Glossary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
156  
Term/Acronym  
BTPS  
Definition  
Body Temperature and Pressure Saturated; A standardization  
for lung volumes and flows to barometric pressure at sea level,  
body temperature, and saturated with water vapor reflecting the  
condition of air in the lung.  
CPAP  
CV  
Continuous Positive Airway Pressure  
Control Ventilation therapy mode; CV mode delivers Mandatory  
breaths with a user-defined inspired tidal volume. This is a Volume  
Control mode.  
DC  
Direct Current  
EPAP  
ESD  
Expiratory Positive Airway Pressure  
Electrostatic Discharge  
Flex  
The device provides the Flex comfort feature. The C-Flex™ feature  
provides increased pressure relief during the expiratory phase  
of breathing. The Bi-Flex® feature provides a level of pressure  
relief taking place at the end of inhalation and at the start of  
exhalation.  
I:E Ratio  
IPAP  
The ratio of inspiratory time to expiratory time.  
Inspiratory Positive Airway Pressure  
Light Emitting Diode  
LED  
l/min  
Liters Per Minute  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
157  
Term/Acronym  
Definition  
Mandatory Breath  
A Mandatory Breath is completely controlled by the ventilator.  
Breaths are initiated by the ventilator according to the breath  
Rate (BPM) setting. Breaths are cycled according to the Inspiratory  
time setting.  
MAP  
Mean Airway Pressure. This is the average airway pressure over  
a full breath cycle. The MAP value displayed on the screen  
represents the average MAP value over six breaths and is updated  
at the end of each exhalation.  
Minute Ventilation  
The volume of gas that moves in and out of the lungs in one  
minute. It is calculated by multiplying the tidal volume by the  
respiratory rate. The Minute Ventilation value displayed on the  
screen represents the average Minute Ventilation over six breaths  
and is updated at the end of each exhalation.  
Operational Hours  
This is the total number of hours that the blower has been on  
since this value was last reset. You can reset this value each time  
you give the device to a new patient to help track their device  
usage.  
OSA  
PC  
Obstructive Sleep Apnea  
Pressure Control therapy mode; PC mode delivers Assisted and  
Mandatory breaths with a user-defined pressure. This is a Pressure  
Control mode.  
PC SIMV  
Pressure Control Synchronous Intermittent Mandatory Ventilation  
therapy mode; PC-SIMV mode delivers Spontaneous, Assisted,  
and Mandatory breaths. This is a Pressure Control mode.  
Peak Flow  
Maximum flow rate (in liters per minute) reached during a breath.  
Highest pressure reached during inspiration.  
Peak Inspiratory  
Pressure (PIP)  
Chapter 12 Glossary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
158  
Term/Acronym  
PEEP  
Definition  
Positive End Expiratory Pressure  
Pressure Control  
Ventilation (PCV)  
Ventilation in which breaths are controlled by operator-defined  
Pressure, Inspiratory Time, and Rise Time.  
PS  
Pressure Support  
Ramp  
A feature that may increase patient comfort when therapy is  
started. The ramp feature reduces pressure and then gradually  
increases the pressure to the prescription setting so patients can  
fall asleep more comfortably.  
Rise Time  
The Rise Time is the time it takes the ventilator to change from  
expiration to inspiration.  
RR  
Respiratory Rate (the number of breaths per minute).  
SD Card  
Secure Digital card; This card records usage and therapy data  
from the device.  
Sigh  
A breath that is delivered every 100 mandatory or assisted  
breaths at 150% of the normal volume.  
SIMV  
Synchronous Intermittent Mandatory Ventilation therapy mode;  
SIMV mode delivers Spontaneous, Assisted, and Mandatory  
breaths with a user-defined inspired tidal volume. This is a  
Volume Control mode.  
Sleep State  
A low-power state entered when the ventilator is running on the  
internal or detachable battery and the blower has been off for  
five minutes. This state preserves battery power. The device is  
restored by plugging the unit into AC power, inserting an SD card,  
or by pressing the Start/Stop button.  
Spontaneous Breath  
Breath type in which the breath is patient-triggered.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
159  
Term/Acronym  
Definition  
Spontaneous (S)  
Mode  
Therapy mode in which all breaths are spontaneous. The  
ventilator supports breathing with user-defined Pressure Support  
(PS) and Rise Time values.  
Spontaneous/Timed  
(S/T) Mode  
Therapy mode that is similar to S mode, except that it can also  
deliver a mandatory breath if the patient does not spontaneously  
breathe within a set time.  
Tidal Volume  
The amount of air passing in and out of the lungs for each breath.  
Timed (T) Mode  
Timed pressure support therapy mode. In Timed mode, all  
breaths delivered are mandatory.  
Volume Control  
Ventilation (VCV)  
Ventilation in which breaths are controlled by an operator-  
defined volume, flow pattern, breath rate, and inspiratory time.  
Vte  
Vti  
Exhaled Tidal Volume  
Inhaled Tidal Volume  
Chapter 12 Glossary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
160  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
13. EMC Information  
Guidance and Manufacturers Declaration - Electromagnetic  
Emissions  
This device is intended for use in the electromagnetic environment specified below. The user of this  
device should make sure it is used in such an environment.  
Emissions Test  
Compliance  
Electromagnetic Environment -  
Guidance  
RF emissions  
Group 1  
The device uses RF energy only for its internal  
function. Therefore, its RF emissions are very low and  
are not likely to cause any interference in nearby  
electronic equipment.  
CISPR 11  
RF emissions  
Class B  
The device is suitable for use in all establishments,  
including domestic establishments and those directly  
connected to the public low-voltage power supply  
network that supplies building used for domestic  
purpose.  
CISPR 11  
Harmonic emissions  
IEC 61000-3-2  
Class A  
Voltage fluctuations/Flicker  
emissions  
Complies  
IEC 61000-3-3  
Chapter 13 EMC Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
162  
Guidance and Manufacturers Declaration - Electromagnetic  
Immunity  
This device is intended for use in the electromagnetic environment specified below. The user of this  
device should make sure it is used in such an environment.  
Immunity Test  
IEC 60601 Test  
Level  
Compliance Level  
Electromagnetic  
Environment -  
Guidance  
Electrostatic  
6 kV contact  
8 kV contact  
Floors should be wood, concrete  
or ceramic tile. If floors are covered  
with synthetic material, the relative  
humidity should be at least 30%.  
Mains power quality should be  
that of a typical home or hospital  
environment.  
Discharge (ESD)  
8 kV air  
15 kV air  
IEC 61000-4-2  
Electrical fast  
Transient/burst  
2 kV for power supply  
2 kV for supply mains  
lines  
IEC 61000-4-4  
1 kV for input-output  
lines  
1 kV for input/output  
lines  
1 kV differential mode  
Surge  
1 kV differential mode  
Mains power quality should be  
that of a typical home or hospital  
IEC 61000-4-5  
2 kV common mode  
<5% UT  
2 kV for common mode environment.  
Voltage dips, short  
<5% UT  
Mains power quality should be  
interruptions and voltage (>95% dip in UT) for  
(>95% dip in UT) for  
0.5 cycle  
that of a typical home or hospital  
environment. If the user of the  
variations on power  
supply input lines  
0.5 cycle  
40% UT  
40% UT  
device requires continued operation  
during power mains interruptions,  
it is recommended that the device  
(60% dip in UT) for  
5 cycles  
(60% dip in UT) for 5  
cycles  
IEC 61000-4-11  
70% UT (30% dip in  
UT) for 25 cycles  
<5% UT (>95% dip in UT)  
for 5 sec  
70% UT (30% dip in UT) for be powered from an uninterruptible  
25 cycles  
power supply or a battery.  
<5% UT (>95% dip in UT)  
for 5 sec  
Power frequency (50/60  
Hz) magnetic field  
3 A/m  
3 A/m  
Power frequency magnetic fields  
should be at levels characteristic  
of a typical home or hospital  
environment.  
IEC 61000-4-8  
NOTE: UT is the a.c. mains voltage prior to application of the test level.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
163  
Guidance and Manufacturers Declaration - Electromagnetic  
Immunity  
This device is intended for use in the electromagnetic environment specified below. The user of this  
device should make sure it is used in such an environment.  
Immunity Test IEC 60601 Test Compliance  
Level Level  
Electromagnetic Environment -  
Guidance  
Portable and mobile RF communications equipment  
should be used no closer to any part of the device,  
including cables, than the recommended separation  
distance calculated from the equation applicable to the  
frequency of the transmitter.  
Recommended separation distance:  
Conducted RF  
IEC 61000-4-6  
3 Vrms  
3 V  
d = 1.2  
P
P
P
150 kHz to 80 MHz  
outside ISM bandsa  
10Vrms  
10 V  
d = 1.2  
150 kHz to 80 MHz  
in ISM bandsa  
Radiated RF  
10 V/m  
10 V/m  
d = 1.2  
80 MHz to 800 MHz  
800 MHz to 2.5 GHz  
IEC 61000-4-3  
80 MHz to 2.5 GHz  
26 MHz to 2.5 GHz  
d = 2.3  
P
where P is the maximum output power rating of the  
transmitter in watts (W) according to the transmitter  
manufacturer and d is the recommended separation  
distance in meters (m).  
Field strengths from fixed RF transmitters, as determined  
by an electromagnetic site surveya, should be less than  
the compliance level in each frequency range.b  
Interference may occur in the vicinity of equipment  
marked with the following symbol:  
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.  
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and  
reflection from structures, objects, and people.  
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile  
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should  
be considered. If the measured field strength in the location in which the device is used exceeds the applicable  
RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is  
observed, additional measures may be necessary, such as re-orienting or relocating the device.  
b
Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.  
Chapter 13 EMC Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
164  
Recommended Separation Distances between Portable and  
Mobile RF Communications Equipment and This Device  
The device is intended for use in an electromagnetic environment in which radiated RF  
disturbances are controlled. The customer or the user of this device can help prevent  
electromagnetic interference by maintaining a minimum distance between portable and mobile RF  
communications equipment (transmitters) and this device as recommended below, according to  
the maximum output power of the communications equipment.  
Rated Maximum  
Power Output of  
Transmitter  
Separation Distance According to Frequency of Transmitter  
(meters)  
150 kHz to 80 MHz 150 kHz to 80 MHz 80 MHz to 800 MHz  
800 MHz to 2.5 GHz  
(Watts)  
outside ISM Bands  
in ISM Bands  
d = 1.2  
P
d = 2.3  
P
d = 1.2  
P
d = 1.2  
P
0.01  
0.1  
0.12  
0.12  
0.12  
0.38  
0.23  
0.73  
0.38  
0.38  
1
1.2  
3.8  
12  
1.2  
3.8  
12  
1.2  
3.8  
12  
2.3  
7.3  
23  
10  
100  
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters  
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output  
power of the transmitter manufacturer.  
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.  
Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;  
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.  
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the  
ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80 MHz and 2.5 GHz to decrease the  
likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into  
patient areas.  
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and  
reflection from structures, objects and people.  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
Index  
External Battery Disconnected 48  
System Alarms  
Ventilator Service Recommended 46  
A
AC Power 51  
AC Power Inlet 17  
Air Filter  
B
Installation 50  
Air Inlet  
Air Inlet Filter  
Cleaning 122  
Replacing 123  
Air Inlet Path  
Replacing 123  
Alarm and Event Log Clean-Up 148  
Alarm Log 89  
Bacteria Filter  
Replacing 125  
Battery Function Verification 146  
Breathing Circuit  
Connection 57  
Breathing Circuit Connection 84  
Breath Rate 82  
Breath Types  
Assist 20  
Alarms 102  
C
Alarm Indicators 104  
Alarm Messages 104  
Alarm Reset 105  
Alarm Volume Control 106  
Patient Alarms  
Cautions 8, 15, 133  
Circuit Type Setting 74  
Contraindications 11  
Control Ventilation Mode 27  
High Exhaled Tidal Volume 44  
High Minute Ventilation 45  
High Peak Inspiratory Pressure 46  
Low Exhaled Tidal Volume 44  
Low Minute Ventilation 45  
Low Peak Inspiratory Pressure 44  
Low Respiratory Rate 44  
Power Alarms 107  
D
Date Format 88  
DC Power  
Detachable Battery Pack 52  
External Battery 52  
Detailed View 86  
Index  
Download from Www.Somanuals.com. All Manuals Search And Download.  
166  
DirectView Software 134  
I
Inspiratory Time 82  
In-Use Bag 134  
IP Address Mode Setting 88  
IPAP 81  
IPAP Max Pressure 81  
IPAP Min Pressure 81  
E
EMC Information 159  
EPAP 81  
Ethernet Connector 18  
Event Log 90  
Exhalation Devices  
Cleaning 125  
K
Exhalation Porting Block 18  
Keypad Backlight 87  
Keypad Lock Feature 63, 87  
F
L
Flex 97  
Flex Comfort Feature 30, 80  
Bi-Flex 30  
Language Setting 86  
LCD Brightness Setting 87  
Low Respiratory Rate 78  
Low Tidal Volume 76  
C-Flex 30  
Flow Cycle Sensitivity 79  
Flow Pattern 84  
Ramp 33  
Square 33  
M
Main Menu Screen 94  
Menu Access i, 86  
Full Menu i  
Limited Menu i  
Monitor Screen 64  
Indicators 66  
My Settings Menu 96  
Flex 97  
Ramp Start Pressure 96  
Rise Time 96  
Flow Trigger Sensitivity 79  
Front Panel  
AC Power LED 15  
Audio Pause Button 16  
Keypad Backlight LEDs 16  
Left/Right Buttons 16  
Red Alarm LED 15  
Start/Stop Button 15  
Up/Down Button 16  
H
O
High Minute Ventilation 77  
High Respiratory Rate 77  
High Tidal Volume Alarm 76  
Humidifier 131  
Operational Hours 89  
Options Menu 86  
Oxygen Valve Connector 18  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
167  
Environmental 149  
Pressure 150  
Standards Compliance 149  
P
Passive Exhalation Device 57  
Patient Circuit  
Cleaning 124  
T
PC-SIMV Mode 25, 83  
Tidal Volume 81  
Timed Mode 82  
Time Format 88  
Trigger Type  
Auto-Trak 78  
Flow Trigger 78  
Troubleshooting 129  
Positive End Expiratory Pressure (PEEP) 84  
Pressure Control Mode 24  
Pressure Control Ventilation 85  
Pressure Setting 83  
Pressure Support 83  
Pressure Units 86  
R
V
Ramp Feature 31  
Volume Control Ventilation 19  
Ramp Length 79  
Ramp Start Pressure 80, 96  
Remote Alarm 62, 104  
Remote Alarm Connector 18  
Rise Time 32, 82, 96  
W
Warnings 3  
Warranty 167  
Water Trap 60  
S
Screen Saver 87  
Secondary Prescription Setting 95  
Secure Digital (SD) Card 17, 133  
Sensitivity 36  
Digital Auto-Trak Sensitivity  
Auto-Trak Sensitivity 37  
Leak Tolerance 36  
Flow Trigger 40  
Serial Connector  
Settings and Alarms Tests 138  
Setup Screen i  
Sigh Feature 84  
SIMV Mode 28, 85  
Specifications  
Control Accuracy 150  
Electrical 150  
Index  
Download from Www.Somanuals.com. All Manuals Search And Download.  
168  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trilogy200  
clinical manual  
Limited Warranty  
Respironics, Inc. warrants that the Trilogy200 system shall be free from defects of workmanship and  
materials and will perform in accordance with the product specifications for a period of two (2) years  
from the date of sale by Respironics, Inc. to the dealer. If the product fails to perform in accordance with  
the product specifications, Respironics, Inc.will repair or replace – at its option – the defective material or  
part. Respironics, Inc. will pay customary freight charges from Respironics, Inc. to the dealer location only.  
This warranty does not cover damage caused by accident, misuse, abuse, alteration, and other defects  
not related to material or workmanship.  
Respironics, Inc. disclaims all liability for economic loss, loss of profits,overhead, or consequential  
damages which may be claimed to arise from any sale or use of this product. Some states do not allow  
the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion  
may not apply to you.  
Accessories and replacement parts, including, but not limited to, circuits, tubing, leak devices, exhaust  
valves, filters and fuses, are not covered under this warranty. However, Respironics warrants that the  
Product’s internal and detachable battery (where supplied) will be free from defects in material and  
workmanship, under normal and proper use and when correctly maintained in accordance with  
applicable instructions, for a period of 90 days from the date of shipment by Respironics to the original  
purchaser. This warranty does not apply to batteries that are dropped, misused, altered or otherwise  
damaged after they are shipped.  
This warranty is given in lieu of all other express warranties. In addition, any implied warranties –  
including any warranty of merchantability or fitness for the particular purpose – are limited to two years.  
Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may  
not apply to you. This warranty gives you specific legal rights, and you may also have other rights which  
vary from state to state.  
Limited Warranty  
Download from Www.Somanuals.com. All Manuals Search And Download.  
170  
To exercise your rights under this warranty, contact your local authorized  
Respironics, Inc. dealer or contact Respironics, Inc. at:  
1001 Murry Ridge Lane  
Murrysville, Pennsylvania 15668-8550  
1-724-387-4000  
Trilogy200 clinical manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Pelco Power Supply WCSI 4 User Manual
Philips DVD Player 721VR User Manual
Philips DVR LFH0652 User Manual
Philips Network Card SPE3000 Series User Manual
Philips Portable CD Player AX2100 User Manual
Phoenix Gold Speaker ATCLCR6 User Manual
Planar Projector PR9030 User Manual
Poulan Tiller PRRT65A User Manual
Proxima ASA Digital Camera COOLPIX 950 User Manual
PYLE Audio Computer Monitor PM15LCD User Manual