R E F 1 0 5 4 7 2 8
1 0 7 2 9 1 9
J H 6 / 7 / 1 0
CLINICAL MANUAL
Download from Www.Somanuals.com. All Manuals Search And Download.
FOR CLINICIAN’S USE ONLY
Accessing Prescription Setting Screens
WARNING
The information on this page is ONLY for health care professionals. Remove this page from the manual
before giving the manual to the patient!
Full Menu Access Mode
The ventilator has two levels of menu access, Full and Limited. Full
Menu Access allows you to alter all available settings. Limited Menu
access permits the user to alter only those prescription parameters
Note: The Full Menu Access key
sequence can be performed
either from the Power Off screen
that affect patient comfort, such as Rise Time, Flex, and Ramp Start
Pressure, if those parameters are available as part of the prescription.
The ventilator defaults to Full Menu Access mode.
or from the Monitor screen.
Note: When the airflow is off
and AC power is removed from
the device for more than 5
minutes, the device will enter
a low power mode to save
battery life. When the device is
in low power mode, the Setup
key sequence will be ignored.
Press the Start/Stop button, or
connect AC power, or insert an
SD card to exit the low power
mode.
After accessing the Main Menu, if the device is in Limited Menu
Access mode, you can use the following Setup key sequence to enter
Full Menu Access mode and be able to change prescription settings:
•
Press the Down button and the Alarm Indicator/Audio
Pause button simultaneously for several seconds. This will
temporarily place the device in Full Menu Access mode.
When you perform this key sequence from the Monitor screen,
the Main Menu screen appears and an audible indicator sounds
indicating you are now in Full Menu Access mode.
When you perform this key sequence when the airflow is off, the
Setup screen appears and an audible indicator sounds.
Note: Philips Respironics
recommends that after you are
finished changing prescription
settings, and you give the
device to the patient, you set
the device back to Limited
Menu Access mode so
You can go into the Options menu and permanently change the
Menu Access setting to Full Menu Access. Otherwise, the device
will return to the Menu Access mode stored in the setting once you
exit the menu screens or if one minute passes without pressing any
device buttons. If you are in the Setup mode, and an SD card is in the
device, “Write Event Log to SD Card”will appear in the menu.
patients cannot change their
prescription settings.
Download from Www.Somanuals.com. All Manuals Search And Download.
R E F 1 0 5 4 7 2 8
1 0 7 2 9 1 9
J H 6 / 7 / 1 0
Download from Www.Somanuals.com. All Manuals Search And Download.
© 2010 Koninklijke Philips Electronics N.V. All rights reserved.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
Table of Contents
FOR CLINICIAN’S USE ONLY Accessing Prescription Setting Screens................................................... i
Chapter 1. Introduction....................................................................................................................................... 1
Package Contents.................................................................................................................1
Intended Use..........................................................................................................................2
Warnings and Cautions ......................................................................................................3
Warnings.........................................................................................................................3
Cautions ..........................................................................................................................8
Notes ..............................................................................................................................10
Contraindications...............................................................................................................11
System Overview................................................................................................................11
Symbols..................................................................................................................................12
Front Panel ...................................................................................................................12
Rear and Side Panels.................................................................................................12
How to Contact Philips Respironics.............................................................................13
Chapter 2. System Description........................................................................................................................15
Front Panel Features..........................................................................................................15
Buttons ..........................................................................................................................15
Visual Indicators .........................................................................................................16
Display Screen.............................................................................................................16
Side and Rear Panel Features .........................................................................................17
Table of Contents
Download from Www.Somanuals.com. All Manuals Search And Download.
ii
Chapter 3. Modes, Features, and Alarms......................................................................................................19
Therapy Modes....................................................................................................................19
Breath Types.................................................................................................................20
Therapy Mode Table .................................................................................................21
Pressure Control Ventilation Therapy Modes...................................................22
Volume Control Ventilation Therapy Modes....................................................27
Therapy Mode Features...................................................................................................30
Flex Comfort Feature................................................................................................30
Ramp ..............................................................................................................................31
Rise Time .......................................................................................................................32
AVAPS Feature.............................................................................................................32
Flow Pattern Types ....................................................................................................33
Sigh Feature.................................................................................................................35
Dual Prescription Feature .......................................................................................35
Triggering ....................................................................................................................36
BTPS Compensation..................................................................................................41
Ventilator Alarms ................................................................................................................41
Loss of Power Alarm..................................................................................................41
Ventilator Inoperative Alarm .................................................................................41
Ventilator Service Required Alarm.......................................................................41
Check Circuit Alarm ..................................................................................................42
Low Circuit Leak Alarm ...........................................................................................42
High Expiratory Pressure Alarm ...........................................................................42
Low Expiratory Pressure Alarm ............................................................................42
High Internal Oxygen Alarm .................................................................................42
Circuit Disconnect Alarm ........................................................................................43
Apnea Alarm ...............................................................................................................43
High Vte Alarm............................................................................................................43
Low Vte Alarm ............................................................................................................43
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
iii
High Vti Alarm.............................................................................................................44
Low Vti Alarm .............................................................................................................44
High Respiratory Rate Alarm .................................................................................44
Low Respiratory Rate Alarm ..................................................................................44
High Inspiratory Pressure Alarm ..........................................................................44
Low Inspiratory Pressure Alarm ...........................................................................45
High Minute Ventilation Alarm ............................................................................45
Low Minute Ventilation Alarm ..............................................................................45
Low Battery Alarm.....................................................................................................46
High Temperature Alarm .......................................................................................46
Replace Detachable Battery Alarm......................................................................46
Ventilator Service Recommended Alarm..........................................................46
AC Power Disconnected Alarm.............................................................................47
Keypad Stuck Alarm .................................................................................................47
Battery Discharging Stopped due to Temperature Info Message............47
Battery Not Charging due to Temperature Info Message ..........................47
Battery Not Charging Info Message....................................................................47
Check External Battery Info Message .................................................................48
Battery Depleted Info Message ...........................................................................48
External Battery Disconnected Info Message..................................................48
Detachable Battery Disconnected Info Message...........................................48
Start On Battery Info Message ..............................................................................48
Card Error Info Message .........................................................................................48
Chapter 4. Ventilator Setup...............................................................................................................................49
Position the Device............................................................................................................50
Install the Air Filter .............................................................................................................50
Supply Power to the Device............................................................................................50
Using AC Power ..........................................................................................................51
Using DC Power..........................................................................................................52
Device Power Source Indicators...........................................................................54
Table of Contents
Download from Www.Somanuals.com. All Manuals Search And Download.
iv
Battery Disposal..........................................................................................................56
First Time Use ..............................................................................................................56
Connect the Breathing Circuit to the Ventilator......................................................57
Connect a Water Trap .......................................................................................................60
Connect Supplemental Oxygen (Optional) ..............................................................61
Connect the Remote Alarm (Optional).......................................................................62
Chapter 5. Viewing and Changing Settings................................................................................................63
Keypad Lock Feature.........................................................................................................63
Accessing the Startup and Monitor Screens ............................................................64
Monitor Screen Indicators ......................................................................................66
On-Screen Button Panel..........................................................................................71
Navigating the Menu Screens........................................................................................71
Changing and Viewing Settings in Full Menu Access Mode...............................72
Changing the Device Settings and Alarms.......................................................73
Device Settings Common to All Therapy Modes...................................73
Additional Settings Specific to Therapy Modes .....................................78
Continuous Positive Airway Pressure (CPAP) Mode..............................78
Spontaneous (S) Mode....................................................................................80
Spontaneous/Timed (S/T) Mode.................................................................82
Timed (T) Mode .................................................................................................82
Pressure Control (PC) Mode...........................................................................83
Pressure Control Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode..........................................................................83
Control Ventilation (CV) Mode .....................................................................84
Assist Control (AC) Mode................................................................................85
Synchronized Intermittent Mandatory Ventilation
(SIMV) Mode........................................................................................................85
Viewing and Changing Options Menu Items ..................................................86
Viewing the Alarm Log ............................................................................................89
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
v
Viewing the Event Log .............................................................................................90
Viewing Device Information ..................................................................................90
Updating Prescriptions Using the SD Card ...............................................................91
Changing and Viewing Settings in Limited Menu Access Mode.......................94
Activating Your Primary or Secondary Prescription ......................................95
Viewing and Changing My Settings Menu Items...........................................96
Connecting the Ventilator to the Patient...................................................................98
Chapter 6. Ventilator Alarms ............................................................................................................................99
Audible and Visual Alarm Indicators.........................................................................100
Audio Pause and Alarm Reset Features...................................................................105
What to Do When An Alarm Occurs..........................................................................106
Alarm Summary Table....................................................................................................107
Chapter 7. Cleaning and Maintenance.......................................................................................................121
Cleaning the Ventilator..................................................................................................121
Cleaning and Replacing the Air Inlet Filter.............................................................122
Replacing the Air Inlet Path Foam.............................................................................123
Cleaning the Patient Circuit.........................................................................................124
Cleaning Instructions (Reusable Circuits) ......................................................124
Preventive Maintenance ...............................................................................................126
Chapter 8. Troubleshooting............................................................................................................................129
Chapter 9. Accessories......................................................................................................................................133
Adding a Humidifier .......................................................................................................133
Adding Supplemental Oxygen to the Device .......................................................133
Using a Remote Alarm Unit..........................................................................................134
Using a Nurse Call System............................................................................................135
Using a Secure Digital (SD) Card ................................................................................135
Using the Philips Respironics DirectView Software ............................................136
Using the Optional In-Use Bag ...................................................................................136
Traveling with the System............................................................................................137
Airline Travel......................................................................................................................137
Table of Contents
Download from Www.Somanuals.com. All Manuals Search And Download.
vi
Chapter 10. System Checkout Procedures................................................................................................139
Tools Required..................................................................................................................139
Visual Inspection..............................................................................................................139
Initial Setup........................................................................................................................140
Settings and Alarms Tests.............................................................................................140
Battery Function Verification.......................................................................................148
Alarm and Event Log Clean-Up ..................................................................................150
Results .................................................................................................................................150
Chapter 11. Technical Specifications...........................................................................................................151
Chapter 12. Glossary .........................................................................................................................................155
Chapter 13. EMC Information ........................................................................................................................161
Index ........................................................................................................................................................................165
Limited Warranty.................................................................................................................................................169
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
1. Introduction
This chapter provides an overview of the Trilogy200 device.
Package Contents
The Trilogy200 system may include the following components.
Some components are optional accessories that may not be
packaged with the device.
Package Contents
Reusable Gray Foam Filters
In-Use Bag
Trilogy200
Clinical Manual
Disposable Active Circuit, No Water Trap
Secure Digital
(SD) Card
(pre-installed)
Disposable Passive Circuit, No Water Trap
Detachable
Battery
Flexible Tubing
Flexible Trach Adapter
(Not Shown)
Universal Porting Block
Exhalation Valve
(pre-installed)
Line (pre-assembled
O2 Inlet
Quick Connects
to AED)
AC Power Cord
Active Exhalation
Device (AED)
Proximal
Pressure Line
(pre-assembled
to AED)
Flow Sensor
Whisper Swivel II
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Intended Use
The Philips Respironics Trilogy200 system provides continuous
or intermittent ventilatory support for the care of individuals who
require mechanical ventilation. Trilogy200 is intended for pediatric
through adult patients weighing at least 5 kg (11 lbs.).
The device is intended to be used in home, institution/hospital, and
portable applications such as wheelchairs and gurneys, and may be
used for both invasive and non-invasive ventilation. It is not intended
to be used as a transport ventilator.
The system is recommended to be used only with various
combinations of Philips Respironics-approved patient circuit
accessories, such as patient interface devices, humidifiers, water
traps, and circuit tubing.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Warnings and Cautions
Caution: U.S. federal law restricts this device to sale by or on the order of a physician.
Warnings
A warning indicates the possibility of injury to the user or operator.
Patient Monitoring Prior to placing a patient on the ventilator, a clinical assessment should
be performed to determine:
•
•
•
The device alarm settings
Needed alternative ventilation equipment
If an alternative monitor (i.e., an alarming Pulse Oximeter or
Respiratory Monitor) should be used
Alternative
Ventilation
For ventilator dependent patients, always have alternate ventilation
equipment, such as a back-up ventilator, manual resuscitator, or similar
device, available.
Ventilator dependent patients should be continuously monitored by
qualified personnel. These personnel should be prepared to provide
alternate therapy in the event of ventilator failure or inoperative
equipment.
Patient Disconnect For ventilator dependent patients, do not rely on any single alarm
Protection
to detect a circuit disconnect condition. The Low Tidal Volume, Low
Minute Ventilation, Low Respiratory Rate, and Apnea alarms should
be used in conjunction with the Circuit Disconnect and Low Peak
Inspiratory Pressure alarms.
Test the operation of the circuit disconnect function daily and
whenever a change is made to the patient circuit. An increase in circuit
resistance can prevent proper operation of some alarms.
Speaking valves, Heat Moisture Exchangers (HMEs), and filters create
additional circuit resistance and may affect the performance of alarms
chosen for circuit disconnect protection.
Do not set the Low Peak Inspiratory Pressure alarm too low, or the
system may not detect large circuit leaks or a patient disconnect.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Personnel
Qualifications
Trilogy200 is a restricted medical device designed for use by
Respiratory Therapists or other trained and qualified caregivers under
the supervision of a physician.
The prescription and other device settings should only be changed on
the order of the supervising physician.
The operator of the ventilator is responsible to read and understand
this manual before use.
Modes of
Ventilation
The device can provide therapies typically associated with both
ventilator dependent and non-dependent patients. The mode of
ventilation, circuit type, and alarm strategies should be chosen after a
clinical evaluation of each patient’s needs.
C-Flex, Bi-Flex, and AVAPS are intended for use by adult patients.
SD Card
Prescription
Changes
When you change the device prescription, alarms, and other settings
using the SD card, Trilogy200 requires that the caregiver review and
verify the changes prior to the changes being used by the device. The
caregiver or health care professional is responsible to ensure that the
prescription settings are correct and compatible with the patient after
using this feature. Installing the wrong prescription for a particular
patient may result in improper therapy, lack of appropriate safety
monitoring, and risk of death or injury to the patient.
Electrical
Interference
This device is intended for use in the electromagnetic environment
specified in Chapter 13 of this manual. The user of this device should
make sure it is used in a compatible environment.
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than the
recommended separation distance calculated using the information
provided in Chapter 13 of this manual.
Battery Back-up
Power
The internal battery is NOT intended to serve as a primary power
source. It should only be used when other sources are not available or
briefly when necessary; for example, when changing power sources.
The ventilator has a two-stage low battery alarm. The medium priority
alarm indicates that approximately 20 minutes of operation remain,
and the high priority alarm indicates that less than 10 minutes of
operation remain. Actual run time may be more or less than this and
varies with battery age, environmental conditions, and therapy.
Immediately seek an alternate power source when the “Low Battery”
alarm appears. Complete power failure and loss of therapy is imminent.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Operating
and Storage
Temperatures
Do not use this device if the ambient temperature is warmer than
40˚ C (104˚ F). If the device is used at room temperatures warmer than
40˚ C, the temperature of the airflow may exceed 43˚ C. This could
cause system alarms, thermal irritation, or injury to the patient’s airway.
Bacteria Filter
Philips Respironics recommends that a main line outlet bacteria filter
(Part Number 342077) be used whenever the device is used for invasive
therapy or if the ventilator may be used on multiple patients.
Patient Circuits
(General)
The ventilator should only be used with patient interfaces (e.g., masks,
circuits and exhalation ports) recommended by Philips Respironics.
Proper operation of the device, including alarms, with other circuits has
not been verified by Philips Respironics and is the responsibility of the
health care professional or respiratory therapist.
When adding any components to the breathing system, the flow
resistance and dead space of the added components such as
humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and
filters should be carefully considered in relation to the potential for
adverse effects on the patient’s ventilatory management and device
alarms.
Passive Circuits
Active Circuits
An exhalation port is required when using a passive circuit.
For the passive circuit, at low expiratory pressures, the flow through
the exhalation port may be inadequate to clear all exhaled gas from
the tubing – some rebreathing may occur. Rebreathing of exhaled air
for longer than several minutes can in some circumstances lead to
suffocation.
Only use the active exhalation devices designed for Trilogy200. Philips
Respironics has not verified proper operation of other active exhalation
devices, and their use may result in improper or unsafe device
operation.
With active exhalation circuits, the exhalation device must be operating
properly for the ventilator to deliver therapy. The exhalation device
should be inspected on a daily basis and replaced whenever necessary.
System Checkout
Do not use the ventilator on a patient until a system checkout has been
performed. See Chapter 10 of this manual.
To make sure the device is operating properly at start-up, always verify
that the audible tone sounds and the alarm LEDs light red and then
yellow momentarily. Contact Philips Respironics or an authorized
service center for service if these indications do not occur at start-up.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
6
Remote Alarms
When using a remote alarm, make sure you fully test the remote alarm
connector and cable by verifying that:
– Annunciated alarms on the ventilator are also
annunciated on the remote alarm.
– Disconnecting the remote alarm cable from the
ventilator or from the remote alarm results in an alarm
notification at the remote alarm.
The remote alarm should be tested daily.
Oxygen
When administering fixed-flow supplemental oxygen, the oxygen
concentration may not be constant. The inspired oxygen concentration
will vary, depending on the pressures, patient flows and circuit leak.
Substantial leaks may reduce the inspired oxygen concentration to less
than the expected value. Appropriate patient monitoring should be
used, as medically indicated, such as an alarming pulse oximeter.
This device DOES NOT alarm for loss of the low flow oxygen supply.
This device is intended to be connected to a low flow (0-15 l/min)
oxygen source such as an oxygen concentrator or other oxygen source
equipped with a pressure regulator (set to 50 PSI or less) and a flow
regulator/meter.
Do not connect the device to an unregulated or high pressure oxygen
source.
The device may result in incorrect flow and tidal volume measurements
and improper operation of related alarms if you add low flow oxygen
directly into the patient circuit or mask instead of using the oxygen
inlet on the back of the ventilator.
Oxygen supports combustion. Oxygen should not be used while
smoking or in the presence of an open flame.
If oxygen is used with the device, the oxygen flow must be turned
off when the device is not in use. Explanation of the Warning:
When the device is not in operation and the oxygen flow is left on,
oxygen delivered into the tubing may accumulate within the device’s
enclosure.
Fire or Explosion
The ventilator should not be operated in the presence of flammable
gasses. This could cause a fire or explosion.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Alarms
Respond immediately to any alarm. It may indicate a potentially
life-threatening condition. Refer to the Alarms and Troubleshooting
chapters for more information.
Visually monitor the patient and ventilator at all times during an Alarm
Silence period. Allowing alarms to continue without intervention may
result in harm to the patient.
If the high priority “Low Internal Battery”message appears,
immediately connect the ventilator to an alternate power source. If no
alternate power source is available, immediately place the patient on
an alternate source of ventilation.
If the “Ventilator Inoperable”alarm occurs, immediately place the
patient on an alternate source of ventilation.
You should not rely on any single alarm to detect a circuit disconnect
condition. The Low Tidal Volume, Low Minute Ventilation, Low
Respiratory Rate, and Apnea alarms should be used in conjunction with
the Circuit Disconnect alarm.
Make sure the alarm volume is set loud enough to be heard by the
caregiver. Consider the use of a remote alarm.
Trilogy200 offers the following circuit type selections:
•
•
•
Passive
Active Flow
Active PAP (Proximal Airway Pressure )
The Passive circuit type provides an ESTIMATE of Vte.
Only the Active Flow circuit type directly measures exhaled tidal
volume (Vte).
The Active PAP circuit type DOES NOT measure Vte and only provides
for an indication of the delivered tidal volume (Vti).
Improperly
Functioning
Ventilator
If you notice any unexplained changes in the performance of the
device, if it is making unusual sounds, if the device or detachable
battery are dropped, if water is spilled into the enclosure, or if the
enclosure is cracked or broken, discontinue use and contact Philips
Respironics or an authorized service center for service.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Maintenance
Follow the service recommendations provided in Chapter 7 of this
manual.
Periodically inspect electrical cords, cables, and the detachable battery
pack for damage or signs of wear. Discontinue use and replace if
damaged.
Repairs and adjustments must be performed by Philips Respironics-
authorized service personnel only. Unauthorized service could cause
death or injury, invalidate the warranty, or result in costly device
damage.
Cleaning
To avoid electrical shock, always unplug the power cord from the wall
outlet before cleaning the ventilator.
(Refer to Chapter 7
for detailed cleaning
instructions.)
Do not immerse the device in any fluids or spray the device with water
or cleaners. Clean the device with a cloth dampened with an approved
cleaner.
If the device has been exposed to rain or dampness, dry the device
including the area around the power cord connection with the power
cord disconnected from the device before applying AC power.
Cautions
A caution indicates the possibility of damage to the device.
Storage
The internal and detachable batteries will self-discharge in storage. If it
is desired to keep the batteries fully charged (for example, as a back-
up ventilator), plug the device into AC power for about eight hours
every 16 days. Alternatively, the ventilator may be left continuously
connected to AC power without battery degradation.
Allowing the batteries to fully discharge will not harm the batteries or
lose device settings, but may require a longer battery charge time prior
to use.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
9
Operating
and Storage
Temperatures
The device may only be operated at temperatures between 5˚ C and
40˚ C (41˚ F and 104˚ F).
Do not operate the device in direct sunlight or near a heating
appliance because these conditions can increase the temperature of
the airflow delivered to the patient.
Prolonged operation or storage at elevated temperatures may reduce
the service life of the battery and other internal components of the
ventilator.
The ventilator has an internal and detachable Lithium-Ion Battery. Do
not expose the device or detachable battery to temperatures above
40˚ C (104˚ F) during use, or above 60˚ C (140˚ F) during storage. This
will reduce battery life and may increase the risk of fire or damage the
battery.
Condensation
Condensation may affect operation or accuracy of the device. If the
device has been exposed to either very hot or very cold temperatures
during storage, allow it to adjust to ambient temperature before
starting therapy.
Air Filter
The reusable foam inlet filter is required to protect the ventilator from
dirt and dust. Wash periodically and replace when damaged for proper
operation.
Cooling Air Vents
Battery Life
Do not block the cooling air vents located on the base and the rear
of the device. This may cause the device to overheat in high ambient
temperatures or at high therapy settings.
The internal and detachable batteries wear out based on the amount
of use (hours or full charge-discharge cycles). The battery capacity and
life are also reduced by operation at higher temperatures.
Detachable Battery Only use the Philips Respironics Trilogy Detachable Battery with the
ventilator.
Cleaning
Do not steam autoclave the ventilator. Doing so will destroy the
ventilator.
Do not immerse the device in liquid or allow any liquid to enter the
enclosure or inlet filter.
Do not spray water or any other solutions directly onto the ventilator.
Do not use harsh detergents, abrasive cleaners, or brushes to clean the
ventilator system. Use only cleaning agents and methods listed in this
manual.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
10
Patient Circuit
Exhalation valves, patient circuits, and water traps are shipped clean,
not sterile. Cleaning and disinfection of these parts should follow
individual institution processes and conform to guidelines provided by
Philips Respironics with each accessory.
External DC Power
Do not use the same external battery to operate both the ventilator
and any other equipment such as power chairs.
An external battery should only be connected to the ventilator using
the Philips Respironics Trilogy External Battery Cable. This cable is
fused, pre-wired, and properly terminated to ensure safe connection
to a standard deep-cycle, lead acid battery. Use of any other adapter or
cable may cause improper operation of the ventilator.
The ventilator should only be connected to an automotive electrical
system using the Philips Respironics Trilogy Automotive Adapter
(when available). This adapter is fused, filtered, and designed for safe
connection to a standard automotive electrical system. Use of any
other adapter or cable may cause improper operation of the ventilator.
Do not operate the ventilator from a car electrical system when
starting the vehicle or jump-starting the vehicle. Electrical transients
during starting may cause improper operation of the ventilator.
Electrostatic
Discharge (ESD)
Do not use antistatic or conductive hoses or conductive patient tubing
with the device.
Notes
•
This product does not contain natural latex rubber or dry
natural rubber in patient or operator accessible areas or
in the air path or breathing circuit.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
11
Contraindications
If the patient has any of the following conditions, consult their health
care professional before using the device in a non-invasive mode:
•
Inability to maintain a patent airway or adequately clear
secretions
•
•
•
•
At risk for aspiration of gastric contents
Diagnosed with acute sinusitis or otitis media
Epistaxis, causing pulmonary aspiration of blood
Hypotension
System Overview
This ventilator provides both pressure control and volume modes of
therapy. The device can provide non-invasive or invasive ventilation.
It can be used to provide total therapy to patients as they progress
from non-invasive to invasive ventilation.
When prescribed, the device provides numerous special features to
help make patient therapy more comfortable. For example, the ramp
function allows you to lower the pressure when trying to fall asleep.
The air pressure will gradually increase until the prescription pressure
is reached. Additionally, the Flex comfort feature provides increased
pressure relief during the expiratory phase of breathing.
The ventilator can be operated using several different power
sources, including an internal Lithium-Ion battery. This battery is
automatically used when the detachable Lithium-Ion battery pack,
external Lead Acid battery, or AC power are not available.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
12
Symbols
The following symbols appear on the device.
Front Panel
Symbol
Description
Therapy Start/Stop
Alarm Indicator/Audio Pause
AC Power Indicator
~
Rear and Side Panels
Symbol
Description
AC Power Connector
~
Secure Digital (SD) Card Slot
Serial Port Connector
Remote Alarm Connector
Ethernet Connector
DC Power Connector
Oxygen Inlet
O2
Consult accompanying instructions for use.
Type BF Applied Part
Class II (Double Insulated)
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
13
Drip Proof Equipment
For Airline Use. Complies with RTCA-D0160F section 21, category M.
How to Contact Philips Respironics
To have your device serviced, contact Philips Respironics Customer
Service department at 1-724-387-4000 or 1-800-345-6443.
Chapter 1 Introduction
Download from Www.Somanuals.com. All Manuals Search And Download.
14
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
2. System Description
This chapter describes the front and rear panel device controls and
features.
2
1
Front Panel Features
The front panel contains the control buttons, visual indicators, and
display screen.
3
4
5
Buttons
Front Panel Controls and
Display Screen
The following buttons are included on the front panel of the device.
Note: When you start therapy,
the display backlight and the
backlights on the buttons turn
on, the red and yellow alarm
LEDs turn on momentarily, and
an audible indicator sounds
to indicate that therapy has
started. The Startup screen
appears on the display.
1.
Start/Stop Button
This button turns the airflow on or off, starting or stopping
therapy.
2.
Alarm Indicator and Audio Pause Button
This button serves two purposes: it temporarily silences
the audible portion of an alarm, and it also acts as an alarm
indicator. When silencing an alarm, if the cause of the alarm is
not corrected, the alarm sounds again after one minute. Each
time the button is pressed, the alarm silence period resets to one
minute. See Chapter 6 for more information.
WARNING
To make sure the device is operating
properly at start-up, always verify
that the audible tone sounds and
the alarm LEDs light red and then
yellow momentarily. Contact Philips
Respironics or an authorized service
center for service if these indications
do not occur at start-up.
3.
Up/Down Button
This button allows you to navigate the display menu and edit
device settings.
Chapter 2 System Description
Download from Www.Somanuals.com. All Manuals Search And Download.
16
4.
Left and Right Buttons
These buttons allow you to select display options or perform
certain actions specified on-screen.
Visual Indicators
Several power and alarm indicators appear on the front panel.
5. AC Power LED
In the lower right corner of the front panel, a green LED (~)
indicates that AC power is applied to the device. This light
remains on as long as adequate AC power is available.
6. Keypad Backlight LEDs
The Start/Stop, Up/Down, and Left/Right buttons all have a white
LED that lights up if the keypad backlight is turned on in the
device Options menu. See Chapter 5 for more information.
7. Red Alarm LED
On the Alarm Indicator/Audio Pause button, a red light flashes to
indicate a high priority alarm.
8. Yellow Alarm LED
Note: See Chapter 6 for more
information about high,
medium, and low priority
alarms.
On the Alarm Indicator/Audio Pause button, a yellow light
flashes to indicate a medium priority alarm. A solid yellow light
indicates a low priority alarm.
Display Screen
The display screen allows you to view settings, system status
information, real-time patient data, alarms, and logs. You can also
modify certain settings on the display screen.
See Chapter 5 for more information on viewing and modifying
device settings.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
17
Side and Rear Panel Features
2
The ventilator’s side and rear panels contain the following connectors
and features, shown at right.
3
1. AC Power Inlet
You can plug the AC power cord into this connector, located on
the right side of the ventilator.
2. Breathing Circuit Connection
1
The breathing circuit connector is located on the right side of
the device. You can connect your circuit tubing system here. See
Chapter 4 for details.
Right Side Panel
3. Exhalation Porting Block
The porting block used here depends on the type of exhalation
device you are using. The Universal Exhalation Porting Block is
shown here. See Chapter 4 for more information.
4. Secure Digital (SD) Card Slot
On the left side of the device is a slot for the optional SD Card.
You can have the patient record usage and therapy information
from the device on the SD card.
4
Left Side Panel
Chapter 2 System Description
Download from Www.Somanuals.com. All Manuals Search And Download.
18
5. Serial Connector
You can use this connector to connect the device to a computer
running PC Direct or Sleepware software or to other Philips
Respironics devices such as Alice 5 and AOM. Use the Trilogy
RS232 Serial Cable to connect the ventilator to the external
device or computer.
11
10
12
8
9
5
6
7
6. Remote Alarm/Nurse Call Connector
If you are using an optional remote alarm or nurse call system
with the ventilator, you can connect the Philips Respironics
remote alarm adapter cable or nurse call adapter cable to this
connector.
Rear Panel
7. Ethernet Connector (when available)
You can connect a PC or router to this connector to upload
therapy information to a secure web site so you can review
therapy information remotely or remotely troubleshoot and
service the device.
8. External Battery Connector (DC Power Inlet)
You can connect an external, stand-alone lead acid battery here,
using the Trilogy External Battery cable.
9. Oxygen (O2) Inlet Connector
If using low flow supplemental oxygen, connect the oxygen
source to this connector using one of the O2 Inlet Quick Connects
provided with the device.
10. Air Inlet and Filter
Insert the filter supplied with the device into the air inlet.
11. Detachable Battery Pack Slot
If you are using the Philips Respironics Lithium-Ion detachable
battery pack to power the device, attach it here.
12. Cord Retainer
Secure the power cord using the cord retainer to prevent
someone from accidentally disconnecting the power cord. See
Chapter 4 for more information.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
3. Modes, Features, and Alarms
Therapy Modes
The device provides Pressure Control Ventilation (PCV) and Volume
Control Ventilation (VCV) for non-invasive and invasive patients.
Pressure Control ventilation delivers a prescribed pressure to
the patient according to set breath rate and set inspiration time
parameters. This means that each breath is controlled so that a
prescribed amount of pressure is delivered to the patient. The device
offers six different Pressure Control modes of operation:
•
•
•
•
•
•
CPAP – Continuous Positive Airway Pressure
S – Spontaneous Ventilation
S/T – Spontaneous/Timed Ventilation
T – Timed Ventilation
PC – Pressure Control Ventilation
PC-SIMV – Pressure Controlled Synchronized Intermittent
Mandatory Ventilation
Volume Control ventilation delivers a prescribed inspired tidal
volume to the patient according to set breath rate and set inspiratory
time parameters. This means that each breath is controlled so that a
prescribed tidal volume is delivered to the patient. The device offers
three different Volume Control modes of operation:
•
•
•
AC – Assist Control Ventilation
CV – Control Ventilation
SIMV – Synchronized Intermittent Mandatory Ventilation
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
20
Breath Types
There are four breath types that apply to the Volume Control and
Pressure Control ventilation therapy modes:
•
•
•
•
Spontaneous
Mandatory
Assisted
Sigh
Spontaneous Breath
A Spontaneous breath is triggered by the patient. Breaths are
initiated by the patient’s inhalation effort, and air delivery is
controlled based on the current pressure or volume setting. Breaths
are terminated by either the ventilator settings or by the patient’s
exhalation effort, depending on the mode selected.
Mandatory Breath
A Mandatory breath (or machine breath) is completely controlled by
the ventilator. The ventilator controls both the beginning (triggering)
and end (cycling) of the inspiratory phase.
Assisted Breath
An Assisted breath is controlled by both the patient and the ventilator.
Breaths are initiated by the patient’s effort and air delivery is controlled
by the current pressure or volume settings. Volume Assisted breaths
will deliver the prescribed Tidal Volume within the prescribed
Inspiratory Time. Pressure Assisted breaths will deliver the prescribed
Inspiratory Pressure for the prescribed Inspiratory Time. Breaths are
terminated when the Inspiratory Time setting has been reached.
Sigh
A Sigh breath is a breath where 150% of the prescribed volume
is delivered. The device will deliver this breath once every 100
Mandatory or Assist breaths when the Sigh setting is enabled. Sigh
breaths are only available in volume modes of ventilation.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
21
Therapy Mode Table
The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table
are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity,
Leak Compensation, and Flow Cycle settings will display.
Note: Pressure Support, referred to in the table below and later in this manual, is defined as IPAP - EPAP or Pressure -
PEEP (PC-SIMV).
Therapy Modes
CPAP
S
√
√
S/T
√
T
√
√
PC
√
PC-SIMV
CV
√
AC
√
SIMV
Dual Prescription
√
√
√
√
√
Circuit Type
CPAP
√
√
√
√
√
√
IPAP
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
AVAPS (On, Off)*
IPAP Max Pressure
IPAP Min Pressure
EPAP
Pressure
Pressure Support (PS)
PEEP
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
Tidal Volume
Breath Rate
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
Inspiratory Time
Trigger Type*
Flow Trigger Sensitivity
Leak Compensation
Flow Cycle Sensitivity
Ramp Length
Ramp Start Pressure
Flex *
Rise Time
Flow Pattern
Sigh
Circuit Disconnect
Apnea
Apnea Rate
High Vte*
Low Vte*
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
High Vti*
Low Vti*
High Minute Ventilation
Low Minute Ventilation
High Respiratory Rate
Low Respiratory Rate
High Inspiratory Pressure
Low Inspiratory Pressure
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only
available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit
type. Leak compensation is only available with the Active Flow circuit type.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
22
Pressure Control Ventilation Therapy Modes
Pressure Control ventilation modes deliver a prescribed pressure to
the patient.
Continuous Positive Airway Pressure (CPAP) Mode
In the Continuous Positive Airway Pressure (CPAP) mode, the device
delivers a continuous pressure to the patient at all times. All breaths
in this mode are Spontaneous breaths.
Spontaneous (S) Mode
In Spontaneous (S) mode, the device delivers bi-level pressure
support. This mode provides only spontaneous breaths. In this
mode, an Inspiratory Positive Airway Pressure (IPAP) is delivered
during inhalation and a lower Expiratory Positive Airway Pressure
(EPAP) is delivered during exhalation. The following illustration
describes these concepts.
= Pressure Supported
Spontaneous Breath
IPAP = 26 cm H2O
1
EPAP = 6 cm H2O
PS = IPAP-EPAP = 20 cm H2O
PRESSURE
5 seconds
4 seconds
8 seconds
26
20
10
6
1
1
1
1
Enter S mode
S Mode
TIME
5
10
15
20
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
23
Spontaneous/Timed (S/T) Mode
In Spontaneous/Timed (S/T) mode, the device delivers bi-level
pressure support. This mode provides Spontaneous and Mandatory
breaths. A Mandatory breath is delivered if the patient does not
spontaneously breathe within the prescribed Breath Rate (BPM)
setting. This ensures that the patient receives a minimum number
of breaths per minute. In this mode, an IPAP is delivered during
inhalation and a lower EPAP is delivered during exhalation. The
duration of a Spontaneous breath is determined by the patient effort.
The duration of a Mandatory breath is determined by the inspiratory
time setting. The following illustration describes these concepts.
= Pressure Supported
Spontaneous Breath
= Mandatory Breath
(note longer inspiratory
time)
IPAP = 26 cm H2O
1
2
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
PRESSURE
6 seconds
6 seconds
6 seconds
26
20
S/T Mode
10
6
1
1
2
1
Enter S/T mode
TIME
5
10
15
20
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
24
Timed (T) Mode
In Timed (T) mode, the device delivers bi-level pressure support.
This mode delivers Mandatory breaths only. A Mandatory breath
is delivered according to the prescribed BPM and inspiratory time
settings. This also means that the ventilator will not respond to
patient effort. In this mode, an IPAP is delivered during inhalation and
a lower EPAP is delivered during exhalation. The following illustration
describes these concepts.
IPAP = 26 cm H2O
= Mandatory Breath
2
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
PRESSURE
6 seconds
6 seconds
6 seconds
6 seconds
26
20
10
6
2
2
2
2
Enter T mode
T Mode
TIME
5
10
15
20
25
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
25
Pressure Control (PC) Mode
In Pressure Control (PC) mode, the device delivers bi-level pressure
support. This mode delivers Assist and Mandatory breaths. This
mode is identical to S/T mode, except that all breaths have a fixed
inspiratory time.
IPAP = 26 cm H2O
= Mandatory Breath
= Assist Breath
2
3
EPAP = 6 cm H2O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H2O
PRESSURE
6 seconds
6 seconds
6 seconds
26
20
PC Mode
10
6
3
3
2
3
Enter PC mode
TIME
5
10
15
20
Pressure Control - Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode
The Pressure Control – Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and
Mandatory breaths. This mode uses a time window to decide
what type of breaths should be delivered. This time window is the
maximum time between breaths according to the set Breath Rate.
When you enter PC-SIMV mode, the time window is started. If
the patient does not trigger a breath, the ventilator will provide
a Mandatory breath when the time expires and then begin the
process again. Mandatory breaths deliver the Pressure setting during
inhalation and a lower Peak End Expiratory Pressure (PEEP)
during exhalation.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
26
During the time window, if patient effort is detected, either a
Spontaneous or an Assist breath is delivered for the first effort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the first patient effort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the first patient effort of the window. Assist breaths
deliver the Pressure setting during inhalation and the PEEP setting
during exhalation.
During the time window, if patient effort continues to be detected
(after the first patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. The figure below provides
example breath patterns in PC-SIMV mode.
Pressure = 26 cm H2O
Pressure Support = 10 cm H2O
PEEP = 6 cm H2O
= Spontaneous
Breath
= Mandatory Window
M
S
1
= Spontaneous Window
= Mandatory Breath
= Assist Breath
2
3
Rate = 5 BPM
PRESSURE
12 seconds
12 seconds
12 seconds
12 seconds
M
S
M
S
M
M
26
20
Sample Breath Patterns in
PC-SIMV Mode
10
6
2
1
1
3
1
2
Enter PC-SIMV mode
TIME
5
10
15
20
25
30
35
40
45
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
27
Volume Control Ventilation Therapy Modes
Volume Control ventilation modes deliver a prescribed inspired tidal
volume according to a set Breath Rate and a set Inspiratory Time.
Control Ventilation (CV) Mode
In Control Ventilation (CV) mode, the device delivers volume control
therapy. This mode provides only Mandatory breaths. A Mandatory
breath is delivered according to the prescribed BPM setting for the
prescribed Inspiratory Time. This also means that the ventilator
will not respond to patient effort. In this mode, the Tidal Volume is
delivered during inhalation and PEEP is delivered during exhalation.
The following illustration describes these concepts.
Tidal Volume = 500 mL
Rate = 10 BPM
= Mandatory Breath
2
Inspiratory Time = 1.5 seconds
Flow Pattern = Square
FLOW
6 seconds
6 seconds
6 seconds
6 seconds
50
Enter CV mode
Control Mode
2
2
2
10
15
0
TIME
5
20
-50
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
28
Assist Control (AC) Mode
In Assist Control (AC) mode, the device delivers volume control
therapy. This mode provides Assist and Mandatory breaths. An Assist
breath is started when there is patient effort, but it is ended when
the Inspiratory Time setting has been met. A Mandatory breath
is delivered if the patient does not spontaneously breathe within
the prescribed BPM setting. This ensures that the patient receives
a minimum number of breaths per minute. In this mode, the Tidal
Volume is delivered during inhalation and PEEP is delivered during
exhalation. The following illustration describes these concepts.
= Mandatory Breath
= Assist Breath
Tidal Volume = 500 mL
Rate = 10 BPM
2
3
Inspiratory Time = 2.0 seconds
Flow Pattern = Square
FLOW
6 seconds
5 seconds
6 seconds
50
Enter A/C mode
2
3
2
10
15
0
TIME
5
20
Assist Control Mode
-50
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
In Synchronized Intermittent Mandatory Ventilation (SIMV) mode, the
device delivers both volume control and pressure control therapy.
This mode provides Spontaneous, Assist, and Mandatory breaths.
This mode uses a time window to decide what type of breaths should
be delivered. This time window is the maximum time between
breaths according to the set Breath Rate.
When you enter SIMV mode, the time window is started. If the
patient does not provide any effort, the ventilator will provide a
Mandatory Volume breath when the time expires and then begin the
process again. Mandatory breaths deliver the Tidal Volume setting
during inhalation and provide a lower Peak End Expiratory Pressure
(PEEP) during exhalation.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
29
During the time window, if patient effort is detected, either a
Spontaneous or an Assist breath is delivered for the first effort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the first patient effort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the first patient effort of the window. Assist breaths
deliver the prescribed Tidal Volume during inhalation and the PEEP
setting during exhalation.
During the time window, if patient effort continues to be detected
(after the first patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. Spontaneous breaths
deliver the prescribed Pressure Support setting above PEEP during
inhalation and PEEP during exhalation. The figure below provides
example breath patterns in SIMV mode.
Tidal Volume = 500 mL
Breath Rate = 5 BPM
= Pressure Supported
Breath
= Mandatory Window
M
S
1
= Spontaneous Window
Inspiratory Time = 3.0 seconds
Pressure Support = 10 cm H2O
PEEP= 6 cm H2O
= Mandatory Breath
= Assist Breath
2
3
PRESSURE
12 seconds
12 seconds
12 seconds
12 seconds
M
S
M
S
M
M
26
20
10
6
2
1
1
3
1
2
SIMV Mode
Enter SIMV mode
TIME
5
10
15
20
25
30
35
40
45
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
30
Therapy Mode Features
The device has several additional features that enhance patient
comfort.
Flex Comfort Feature
Note: Flex is not available if
AVAPS is enabled or if an active
circuit is used.
The device consists of a special comfort feature called Flex. The
device provides the Flex feature in CPAP mode and S mode. This
feature is only available when Auto-Trak is enabled.
Note: The patient has access to
this setting, if Flex is enabled.
C-Flex
When in CPAP mode, if C-Flex is enabled, it enhances patient comfort
by providing pressure relief during the expiratory phase of breathing.
In the following diagram, the dashed lines represent normal CPAP
therapy in comparison to the bold line representing C-Flex. C-Flex
levels of 1, 2, or 3 progressively reflect increased pressure relief.
Note: C-Flex is not available if
CPAP is set to 4 cm H2O.
Note: C-Flex, Bi-Flex and AVAPS
are intended for use by adult
patients.
PRESSURE
Inhalation
Exhalation
CPAP
1
2
C-Flex in Comparison to
Traditional CPAP Therapy
3
TIME
C-Flex pressure relief is determined by the C-Flex setting and the
amount of patient flow. C-Flex returns to the set pressure by the end
of exhalation, when the airway is most vulnerable to closure.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
31
Bi-Flex
Note: Bi-Flex is only available
up to 25 cm H2O in S mode.
In S mode, the Bi-Flex attribute adjusts therapy by inserting a small
amount of pressure relief during the latter stages of inspiration and
during the beginning part of exhalation. In the following diagram,
the bold lines represent Bi-Flex in comparison to the dashed line
representing normal BiPAP therapy. Bi-Flex levels of 1, 2, or 3
progressively reflect increased pressure relief that will take place at
the end of inspiration and at the beginning of expiration.
PRESSURE
BiPAP
IPAP
Bi-Flex in Comparison to
Traditional Bi-Level Therapy
EPAP
1
Bi-Flex
2
3
TIME
Ramp
Note: IPAP will not ramp below
4 cm H2O.
The device is equipped with a linear ramp function. In CPAP, S, S/T, T,
and PC modes, the Ramp feature will reduce the pressure and then
gradually increase (ramp) the pressure to the prescription pressure
setting so patients can fall asleep more comfortably. The figure below
illustrates how the ramp function works.
The Ramp Function
TIME
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
32
Rise Time
In S, S/T, PC, T, PC-SIMV, and SIMV modes, rise time is the amount
of time it takes the device to change from the expiratory pressure
setting to the inspiratory pressure setting. Rise time levels of 1, 2,
3, 4, 5, or 6 progressively reflect slowed response of the pressure
increase that will take place at the beginning of inspiration. Adjust
the rise time to find the most comfortable setting for the patient.
PRESSURE
Rise Time
IPAP
EPAP
Rise Time
TIME
AVAPS Feature
Note: AVAPS is only available if
Average Volume Assured Pressure Support (AVAPS) is a feature
available in the S, S/T, PC, and T modes. It helps patients maintain
a tidal volume (VT) equal to or greater than the target tidal volume
(Volume setting in the ventilator) by automatically controlling the
pressure support (PS) provided to the patient. The AVAPS feature
adjusts PS by varying the IPAP level between the minimum (IPAP Min)
and maximum (IPAP Max) settings. AVAPS averages VT and changes
the PS value gradually. This occurs over several minutes. The rate of
change is slow, so that the patient is not aware of breath-to-breath
pressure changes.
you are using a passive circuit.
Note: C-Flex, Bi-Flex and AVAPS
are intended for use by adult
patients.
As patient effort decreases, AVAPS automatically increases PS to
maintain the target tidal volume. The IPAP level will not rise above
IPAP Max, even if the target tidal volume is not reached. Conversely,
as patient effort increases, AVAPS will reduce PS. IPAP will not fall
below IPAP Min, even if the target tidal volume is exceeded.
If the Ramp function has been activated, it will take precedence over
the AVAPS feature. Once ramp is complete, AVAPS will resume.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
33
PRESSURE
IPAP Max
IPAP Min
EPAP
TIME
15
15
30
30
45
45
60
VOLUME
Target
Volume
AVAPS Feature
TIME
60
Flow Pattern Types
Two flow patterns are available in Volume Control ventilation therapy
modes:
•
•
Square
Ramp
Square
With a square wave pattern, airflow is generally constant throughout
inspiration of the breath.
FLOW
Square Flow Pattern
TIME
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
34
Ramp
With a ramp flow pattern, the airflow starts high and decreases
throughout inspiration of the breath.
FLOW
50% of
Peak Flow
Ramp Waveform Pattern
TIME
For the active circuit in volume modes, peak flow is required to be
a minimum of 20 l/min. The wave form may be flattened when the
combination of Inspiratory Time and Tidal Volume set points would
result in a flow of less than 20 l/min. Therefore, for some settings,
a Ramp flow pattern may provide a pattern that more closely
resembles a Square flow pattern.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
35
Sigh Feature
The sigh feature is available for Volume ventilation modes only.
When the sigh feature is enabled, the ventilator delivers a sigh breath
in place of every 100th mandatory or assisted breath delivered
regardless of the mode of operation (i.e., AC, CV, and SIMV). The Sigh
breath is delivered using a volume equal to 150% of the set volume
that was in effect when the breath was initiated.
FLOW
Example of Sigh Feature
TIME
Dual Prescription Feature
The device provides a dual prescription feature that allows you
to enter a primary prescription and a secondary prescription for
the patient if needed. For example, you can set a primary daytime
prescription and secondary nighttime prescription. See Chapter 5 for
more information on the dual prescription feature.
Note: Both prescriptions must
use the same circuit type.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
36
Triggering
The device can be set to trigger breaths using the Auto-Trak or Flow
Trigger sensitivity features.
Digital Auto-Trak Sensitivity
An important characteristic of the device is its ability to recognize
and compensate for unintentional leaks in the system and to
automatically adjust its trigger and cycle algorithms to maintain
optimum performance in the presence of leaks. This feature is known
as Digital Auto-Trak Sensitivity. The following sections examine this
function in detail by describing the leak tolerance function and
sensitivity.
Leak Tolerance
A microprocessor monitors the total flow of the patient circuit and
calculates patient flow values.
A. Leak Estimation: Average and Parabolic
The device uses two leak estimation algorithms. A conservation
of mass algorithm is used to compute the average leak for a given
pressure support relationship. This average leak is used when large
leak variations are present in the system. Average leak is a high
estimate during EPAP pressure and a low estimate during IPAP
pressure. A better leak estimate, enabled by the digital system, is the
parabolic leak algorithm. Parabolic leak is proportional to the square
of the patient pressure; therefore, the leak estimate is correlated to
the changing patient pressure. Both algorithms include unintentional
circuit leak and are averaged over several breaths.
B. Patient Flow
The total circuit flow is comprised of the circuit leak and the patient
flow. The calculated patient flow is the total flow minus the circuit
leak. Patient flow is a primary input into the triggering and cycling
mechanisms.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
37
Auto-Trak Sensitivity
Note: Auto-Trak is only
available if you are using a
passive circuit.
An essential feature of the device while operating in all modes is
its ability to effectively sense spontaneous breathing efforts, which
causes the ventilator to trigger to inspiration and cycle to expiration.
Because no preset sensitivity threshold can assure patient and
machine synchrony with changing breathing efforts and circuit
leaks, the device continuously tracks patient breathing patterns
and automatically adjusts sensitivity thresholds to ensure optimum
sensitivity as breathing patterns change or as circuit leaks change.
The algorithms used to ensure optimum sensitivity are the Volume
Trigger, Shape Signal, Spontaneous Expiratory Threshold (SET), Flow
Reversal, Maximum IPAP Time, and Volume Control Cycle.
Volume Trigger (Expiration to Inspiration):
The volume trigger is one method used to trigger inspiration during
spontaneous breathing in all modes except T and CV. The volume
trigger threshold is 6 ml of accumulated patient inspiratory volume.
When patient effort generates inspiratory flow causing 6 ml of
volume, inspiration is triggered.
Shape Trigger/Shape Cycle (Expiration to Inspiration) (Inspiration to
Expiration):
The shape trigger/cycle is another method used to trigger inspiration
and/or cycle from inspiration to expiration during spontaneous
breathing in all modes except T and CV. This method continuously
tracks patient inspiratory and expiratory flow and adjusts the
spontaneous trigger and cycle thresholds for optimum sensitivity.
The Shape Signal appears as a shadow image of the patient’s actual
flow. The shape signal functions as a sensitivity threshold at either
inspiration or expiration. When the patient’s flow rate crosses the
shape signal the unit changes pressure levels. The following figure
illustrates how the shape signal is super-imposed onto the actual
waveform to trigger and cycle off IPAP. The shape signal is created
by offsetting the signal from the actual patient flow by 15 l/min and
delaying it for a 300 msec period. This intentional delay causes the
shape signal to be slightly behind the patient’s flow rate.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
38
A sudden change in patient flow will cross the shape signal, causing
the pressure level to change.
IPAP
Pressure
EPAP
Cycle to
EPAP
Crossover
Point
Shape
Signal
Estimated
Patient
Flow
Flow
Trigger to
IPAP
Crossover
Point
Shape Signal
Tracking the patient’s flow pattern with the Shape Signal provides a
sensitive mechanism to trigger to inspiration or cycle to expiration in
response to changing breathing patterns and circuit leaks.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
39
Spontaneous Expiratory Threshold (Inspiration to Expiration):
A second method used to cycle to expiration during spontaneous
breathing in all modes except T, CV, AC, and SIMV, is called
Spontaneous Expiratory Threshold (SET). The SET rises in proportion
to the inspiratory flow rate on each breath. When the SET and actual
patient flow value are equal, the unit cycles to expiration.
IPAP
Pressure
EPAP
Spontaneous
Expiratory
Threshold
Flow
Spontaneous Expiratory
Threshold
Flow Reversal (Inspiration to Expiration):
As flow begins to decrease during inspiration, a flow reversal can
occur due to a large leak around the mask or because the patient’s
mouth is open. When the device senses this flow reversal, it
automatically cycles to expiration.
Maximum IPAP Time (Inspiration to Expiration):
The maximum inspiratory time is determined by the adjustment of
the Inspiratory time setting. A maximum IPAP time of 3.0 seconds
acts as a safety mechanism to limit the time spent in inspiration
during spontaneous breathing. For mandatory or assisted breaths,
the maximum inspiratory time will equal the Inspiratory time setting
up to 5.0 seconds.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
40
Volume Control Cycle (Inspiration to Expiration) (Only available
during Volume Control Therapy)
An Inspiratory Time setpoint limits the time spent in inspiration
during breathing in all modes. Once the time limit is reached, the unit
automatically cycles to expiration.
Flow Trigger
Flow trigger provides a manual setting that allows for breath
initiation and termination based on a set flow trigger sensitivity and
flow cycle sensitivity.
Flow Trigger Sensitivity (Expiration to Inspiration):
The flow trigger initiates when the patient’s inspiratory effort creates a flow
equal to or greater than the flow trigger sensitivity setting. The method of
the flow trigger is dependent upon the circuit type that is chosen.
Leak Compensation:
When using the Passive Circuit configuration, compensation for both the
intentional and unintentional leak is included in the triggering method.
When using the Active PAP Circuit configuration, leak compensation
is not available.
When using the Active Flow Circuit configuration, flow trigger with
leak compensation may be enabled. The default setting when using
the Active Flow Circuit is Leak Compensation On. The clinician has
the option to turn off leak compensation; however, unintentional
leak will not be compensated. Both options measure the flow at the
proximal flow sensor.
Note: Enabling Leak
Compensation when using the
Active Flow Circuit configuration
only affects triggering and does
not affect tidal volume delivery
or Vte measurement.
Flow Cycle Sensitivity (Inspiration to Expiration):
This cycling method is only active if the Flow Trigger has been
selected for the Trigger Type. As flow begins to decrease during
inspiration, if the patient flow is less than the flow cycle sensitivity
setting, the device will cycle to expiration.
For example: if the flow cycle sensitivity setting is set to 75%, when
the flow has decreased by 25% of the peak flow, the device will cycle
to the EPAP/PEEP level.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
41
BTPS Compensation
All flows and volumes used in Trilogy are expressed in BTPS - Body
Temperature atmospheric Pressure Saturated with H2O.
All pressures are expressed relative to atmospheric pressure.
Ventilator Alarms
This section describes all of the ventilator alarms and informational
messages in order of priority, from high priority alarms to low priority
alarms and finally informational messages. Refer to Chapter 6 for
more information on alarms.
Loss of Power Alarm
This is a high priority alarm. It may occur when a complete power
failure has occurred and power was lost while the device was
providing therapy. This may happen if the internal battery was the
only power source in use and was completely depleted.
Ventilator Inoperative Alarm
This is a high priority alarm. It occurs when the ventilator detects an
internal error or a condition that may affect therapy. The device will
shut down if the cause of the failure indicates that the device cannot
deliver therapy safely. If the device can deliver therapy at a limited
level, the device will continue to deliver limited therapy.
Ventilator Service Required Alarm
This is a high priority alarm. It occurs when the device cannot
perform to specification, a backup safety feature is compromised,
or the delivery of therapy is compromised. The device continues to
operate (possibly in a reduced capacity mode). If the problem is not
corrected, the device will generate a reminder message once per
hour until the issue is corrected. Additionally, if therapy is stopped, a
reminder message will immediately appear when therapy is turned
on again.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
42
Check Circuit Alarm
This is a high priority alarm. It occurs when the device detects a
problem with the patient circuit, such as pinched or detached tubing,
water condensation in the proximal pressure lines, or problems with
the active exhalation device.
Low Circuit Leak Alarm
This is a high priority alarm that only occurs with the passive circuit.
It occurs when the system detects a problem with the leak device in
the passive circuit.
High Expiratory Pressure Alarm
This is a high priority alarm. It occurs when the delivered pressure
exceeds the target patient pressure during the expiratory phase by
5 cmH2O. This may be due to pinched tubing or the patient having
a fast breath rate. The device continues to operate. The alarm will
automatically terminate when the delivered pressure comes within
5 cmH2O of the target patient pressure during the expiratory phase.
Low Expiratory Pressure Alarm
This is a high priority alarm. It occurs when the delivered pressure
is 5 cmH2O or more below the target patient pressure during the
expiratory phase. The device continues to operate. The alarm will
automatically terminate when the delivered pressure comes within
5 cmH2O of the target patient pressure during the expiratory phase.
High Internal Oxygen Alarm
This is a high priority alarm. It occurs when there is a leak in the
internal air delivery system that allows oxygen to build up inside
the device. The alarm is generated when the internal oxygen
concentration reaches 5% above ambient levels.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
43
Circuit Disconnect Alarm
WARNING
This is a high priority alarm. It occurs when the breathing circuit
is disconnected or has a large leak. The device continues to
operate. The alarm will automatically terminate when the circuit is
reconnected or the leak is fixed.
You should not rely on any
single alarm to detect a circuit
disconnect condition. The Low
Tidal Volume, Low Minute
Ventilation, Low Respiratory
Rate, and Apnea alarms should
be used in conjunction with
the Circuit Disconnect alarm.
The Apnea alarm is only
Apnea Alarm
This is a high priority alarm. It occurs when the patient has not
triggered a breath within the time specified in the apnea alarm
intended for spontaneously
setting. The device continues to operate. The alarm will automatically breathing patients.
terminate when two consecutive patient breaths are detected that
meet the apnea alarm time setting.
High Vte Alarm
This is a high priority alarm. It occurs when the estimated exhaled
tidal volume is greater than the High Vte alarm setting for three
consecutive breaths. The device continues to operate. The alarm will
automatically terminate when a breath occurs in which the exhaled
tidal volume does not reach the High Vte alarm setting.
Low Vte Alarm
This is a high priority alarm. It occurs when the estimated exhaled
tidal volume is lower than the Low Vte alarm setting for three
consecutive breaths. The device continues to operate. The alarm will
automatically terminate when a breath occurs in which the exhaled
tidal volume exceeds the Low Vte alarm setting.
When AVAPS is on, this alarm will occur when the calculated tidal
volume is less than 90% of the target tidal volume setting. The alarm
will automatically terminate when a breath occurs in which the
exhaled tidal volume is equal or greater to 90% of the target tidal
volume setting.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
44
High Vti Alarm
This is a high priority alarm. It occurs when the delivered tidal
volume is greater than the High Vti alarm setting for three
consecutive breaths. The device continues to operate. The alarm will
automatically terminate when a breath occurs in which the delivered
tidal volume does not reach the High Vti alarm setting.
Low Vti Alarm
This is a high priority alarm. It occurs when the delivered tidal volume
is less than the Low Vti alarm setting for three consecutive breaths.
The device continues to operate. The alarm will automatically
terminate when a breath occurs in which the delivered tidal volume
exceeds the Low Vti alarm setting.
High Respiratory Rate Alarm
This is a high priority alarm. It occurs when the respiratory rate is greater
than the High Respiratory Rate alarm setting. The device continues to
operate. The alarm will automatically terminate when the measured
respiratory rate is less than the High Respiratory Rate alarm setting.
Low Respiratory Rate Alarm
This is a high priority alarm. It occurs when the respiratory rate is less
than the Low Respiratory Rate alarm setting. The device continues to
operate. The alarm will automatically terminate when the measured
respiratory rate is greater than the Low Respiratory Rate alarm setting.
High Inspiratory Pressure Alarm
This alarm occurs in several stages and escalates from an audible beep for
the first two occurrences to a high priority alarm if the problem continues.
It is detected differently for volume and pressure therapy modes.
For volume modes, the alarm will sound if the measured patient
pressure exceeds the High Inspiratory Pressure setting specified by
the clinician. The alarm will automatically terminate when the peak
inspiratory pressure is less than or equal to the High Inspiratory
Pressure alarm setting.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
45
For pressure modes, the alarm occurs when the delivered pressure
exceeds the target patient pressure by 5 cmH2O or more during
the inspiratory phase. The device will automatically cycle to
the expiratory phase and continue to operate. The alarm will
automatically terminate when the delivered pressure falls within 5
cmH2O of the target patient pressure during the inspiratory phase.
Low Inspiratory Pressure Alarm
This is a high priority alarm. It is detected differently for volume and
pressure therapy modes.
For volume modes, the alarm will sound if the measured patient
pressure is less than the Low Inspiratory Pressure setting specified by
the clinician. The alarm will automatically terminate when the peak
pressure at the end of the breath is greater than or equal to the Low
Inspiratory Pressure alarm setting.
For pressure modes, the alarm occurs when the delivered pressure
is 5 cmH2O or more below the target patient pressure during the
inspiratory phase. The alarm will automatically terminate when
the delivered pressure comes within 5 cmH2O of the target patient
pressure during the expiratory phase.
High Minute Ventilation Alarm
This alarm is a high priority alarm. It occurs when the patient’s
minute ventilation is greater than the High Minute Ventilation alarm
setting. The device continues to operate. The alarm will automatically
terminate when the calculated minute ventilation is less than the
High Minute Ventilation alarm setting.
Low Minute Ventilation Alarm
This alarm is a high priority alarm. It when the patient’s minute
ventilation is less than the Low Minute Ventilation alarm setting. The
device continues to operate. The alarm will automatically terminate
when the calculated minute ventilation is greater than the Low
Minute Ventilation alarm setting.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
46
Low Battery Alarm
WARNING
Immediately seek an alternate
power source when the “Low
Battery” message appears.
Complete power failure and loss
of therapy is imminent.
The Low Battery alarm occurs when the last battery available is
low or nearly depleted. This alarm occurs in two stages. When
approximately 20 minutes of battery run time remains, a medium
priority alarm is generated, and the device continues to operate. If
no action is taken and the battery continues to deplete, the alarm
escalates to a high priority alarm when approximately 10 minutes of
battery run time remains.
WARNING
The ventilator has a two-stage
low battery alarm. The medium
priority alarm indicates that
approximately 20 minutes of
operation remain, and the high
priority alarm indicates that less
than 10 minutes of operation
remain. Actual run time may be
more or less than this and varies
with battery age, environmental
conditions, and therapy.
High Temperature Alarm
This alarm occurs when the estimated patient airstream temperature
or the ventilator internal temperature is too high. The alarm occurs in
several stages. The ventilator continues to operate. Internal fans are
started when the medium priority alarm is generated. If the condition
causing the high temperature is not corrected and the temperature
continues to rise, the alarm will escalate to the high priority alarm.
Replace Detachable Battery Alarm
The Replace Detachable Battery alarm occurs when the detachable
battery is nearing the end of its useful life or a failure in the
detachable battery that prevents it from charging or discharging has
been detected.
The alarm occurs in several stages, from low to high priority. The
device continues to operate when the alarm is the low priority
alarm. If the alarm is reset without removing the battery, the alarm
will be regenerated once every hour until the detachable battery is
removed. The device continues to operate, the detachable battery
is not used, and the power source is switched to the next available
power source if the alarm is the high priority alarm.
Ventilator Service Recommended Alarm
This is a medium priority alarm. It occurs when the device has
detected an error, but the error will not affect device performance
or safety. The device continues to operate. Therapy and safety are not
compromised.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
47
If the problem is not corrected, the device will generate a reminder
message once per day, or whenever power is cycled, until the issue
is corrected. Additionally, if the device is powered off, a reminder
message will immediately appear when the device is turned on again.
AC Power Disconnected Alarm
This is a medium priority alarm. It occurs when the AC power source
was lost, and the device has switched to an alternate power source
(either a detachable or external battery, if connected, or the internal
battery if no other source is available). The device continues to
operate. If AC power returns, the ventilator will beep, but no message
will appear on the display.
Keypad Stuck Alarm
This is a low priority alarm. It occurs when a key becomes lodged
inside the case of the device.
Battery Discharging Stopped due to Temperature
Info Message
This info message occurs when the detachable or internal battery
becomes overheated while providing power for the device. The
device continues to operate. The battery is not used and the power
source is switched to the next available power source.
Battery Not Charging due to Temperature Info Message
This info message occurs when the detachable or internal battery
becomes too hot while charging or the device was in too cold an
environment before charging started. The device continues to
operate. Battery charging stops until the battery cools or warms
sufficiently.
Battery Not Charging Info Message
This info message occurs when the device has detected an error
condition with the battery that prevents it from accepting a charge.
The device continues to operate. Battery charging stops.
Chapter 3 Modes, Features, and Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
48
Check External Battery Info Message
This info message occurs when a bad connection exists to the
external battery or the external battery failed. The device continues
to operate using power from the detachable battery, if available, or
the internal battery.
Battery Depleted Info Message
This info message occurs when the affected battery is fully depleted.
The device continues to operate using the next available power source.
External Battery Disconnected Info Message
This info message occurs when the external battery power source is
lost and the device has switched to an alternate power source (either
a detachable battery, if connected, or the internal battery if no other
source is available). If external battery power returns, the ventilator
will beep, but no message will appear on the display.
Detachable Battery Disconnected Info Message
This info message occurs when the detachable battery power source
is lost and the device has switched to an alternate power source (the
internal battery if no other source is available). If detachable battery
power returns, the ventilator will beep, but no message will appear
on the display.
Start On Battery Info Message
This info message indicates that the ventilator has started on battery
power and no AC power is available. The device operator should
verify that this is what is wanted.
Card Error Info Message
This info message occurs when an unusable SD card is inserted into
the ventilator. The device continues to operate but data cannot be
logged onto the SD card.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
4. Ventilator Setup
This chapter provides instructions on how to assemble the ventilator.
It includes the following setup information:
WARNING
Ventilator dependent patients
should be continuously
monitored by qualified
personnel. These personnel
should be prepared to provide
alternate therapy in the event of
ventilator failure or inoperative
equipment.
•
•
•
•
•
•
Properly positioning the device
Installing the air filter
Supplying power to the device
Connecting the breathing circuit
Connecting a water trap (optional)
Connecting a remote alarm (optional)
WARNING
Do not use the ventilator on a
patient until a system checkout
has been performed. See
Chapter 10 of this manual.
WARNING
For ventilator dependent
patients, always have alternate
ventilation equipment, such as
a back-up ventilator, manual
resuscitator, or similar device,
available.
CAUTION
Do not operate the ventilator at
temperatures below 5° C (41° F)
or above 40° C (104° F).
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
50
Position the Device
Place the ventilator on a flat, level surface. Or, as an alternative, you
can place the device in the optional in-use bag. The in-use bag is
required for portable use in order to properly secure the ventilator.
Do not operate the ventilator while on its side, upside-down, or in
any other orientation. Make sure that the air inlet on the back of the
device is not blocked. If you block the air flow around the device, the
ventilator may not work properly.
CAUTION
Install the Air Filter
The reusable foam inlet filter is
required to protect the ventilator
from dirt and dust. Wash
periodically and replace when
damaged for proper operation.
The device uses a gray foam filter that is washable and reusable. The
reusable filter screens out normal household dust and pollens. The
filter must be in place at all times when the device is operating. Two
reusable gray foam filters are supplied with the device. If the filter is
not already installed when you receive the device, you must install
the filter before using the ventilator.
To install the filter, insert the gray foam filter into the filter area as
shown.
Supply Power to the Device
Installing the Filter
The device can operate on AC or DC power. The ventilator accesses
power from potential sources in the following order:
Note: See Chapter 7 for
information on how to clean
and replace the air filter.
•
•
•
•
AC Power
External Battery
Detachable Battery Pack
Internal Battery
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
51
Note: This device is activated
(i.e., ‘powered on’) when the AC
power cord is connected or any
of the three battery sources is
available. Pressing the Start/Stop
button turns the airflow on or off.
Using AC Power
An AC power cord is provided with the device.
1. Plug the socket end of the AC power cord into the AC power inlet
on the side of the device. To prevent accidental removal of the
AC power cord from the device, you may secure the AC power
cord to the device using the AC power cord retainer located on
the back of the device.
CAUTION
The device may only be operated
at temperatures between 5˚ C
and 40˚ C (41˚ F and 104˚ F).
2. Plug the pronged end of the cord into a wall outlet not
connected to a wall switch.
3. Ensure that all connections are secure. If AC power is connected
correctly and the device is operating properly, the green AC
Power LED should be on.
WARNING
Periodically inspect the power
cord for damage or signs of
wear. Discontinue use and
replace if damaged.
Note: To remove AC power,
disconnect the power supply
cord from the electrical outlet.
Connecting the AC Power
Cord
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
52
4. If desired, secure the power cord using the cord retainer on the
back of the device, as shown below. To secure the power cord,
use a screwdriver to remove the screw on the cord retainer.
Insert the cord as shown into the cord retainer, and then re-
attach the cord retainer to the device by re-attaching the screw.
Securing the Power Cord
Using the Cord Retainer
Using DC Power
You can operate the ventilator using an external battery, detachable
battery, or the internal battery.
CAUTION
External Battery
Do not use the same external
battery to operate both the
ventilator and any other equipment
such as power chairs.
The ventilator can operate from a 12 VDC deep cycle marine-type
(lead acid) battery using the Philips Respironics Trilogy External
Battery Cable. This cable is pre-wired and properly terminated to
ensure safe connection of an external battery to the ventilator.
Battery operating time depends on the characteristics of the battery
and usage of the device.
CAUTION
An external battery should only
be connected to the ventilator
using the Philips Respironics
Trilogy External Battery Cable.
This cable is fused, pre-wired and
properly terminated to ensure
safe connection to a standard
deep cycle lead acid battery. Use
of any other adapter or cable
may cause improper operation
of the ventilator.
Due to a variety of factors, including battery chemistry, battery age,
and use profile, the capacity of the external battery as shown on the
device display is only an estimate of the actual remaining capacity.
Refer to the instructions supplied with the Philips Respironics Trilogy
External Battery Cable for detailed information on how to operate
the device using an external battery.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
53
Detachable Battery Pack
Philips Respironics offers a detachable Lithium-Ion battery pack. To
use the detachable battery pack, snap the battery into place on the
back of the ventilator, as shown in the following illustration. When
the device is not connected to an AC power source or an external
battery, the detachable battery will power the device, if attached.
The length of time the ventilator will operate on battery power
depends on many factors such as device settings, battery charge
level, and condition or age of the battery. When fully charged, a new
battery can power the ventilator for approximately three hours under
typical patient conditions.
CAUTION
The detachable and internal
batteries wear out based on
the amount of use (hours or full
charge-discharge cycles). The
battery capacity and life are also
reduced by operation at higher
temperatures.
Whenever the ventilator is connected to AC power, it will
automatically recharge the detachable battery pack. A completely
discharged detachable battery will reach 80% charge status within 8
hours, when charging at approximately 23˚ C ambient temperature.
CAUTION
Only use the Philips Respironics
Trilogy Detachable Battery with
the device.
Insert and securely latch the detachable battery into the device as
shown below.
Attaching the Detachable
Battery
Insert
Properly Installed
Detachable
Detachable
Battery
Battery
CAUTION
Prolonged operation or storage
at elevated temperatures may
reduce the service life of the
detachable or internal battery
and other internal components
of the ventilator.
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
54
One side of the detachable battery has a set of LEDs that indicate the
amount of charge left on the battery. You can press the button below
the LEDs to view how much charge remains:
LED
All 5 LEDs are lit
Battery Capacity
80-100% capacity
4 LEDs are lit
3 LEDs are lit
2 LEDs are lit
1 LED is lit
60-79% capacity
40-59% capacity
20-39% capacity
10-19% capacity
1 to 9% capacity
0% capacity
1 LED flashes
0 LEDs are lit
Detachable Battery LEDs
WARNING
Internal Battery
The device contains an internal battery that can be used as a back-
up power source. It is intended for use during short periods while
switching between external power sources, emergency situations, or
short durations when the user needs to be mobile. The length of time
the ventilator will operate on internal power depends on many factors
such as device settings, battery charge level, and condition or age of the
battery. When fully charged, a new battery can power the ventilator for
approximately three hours under typical patient conditions.
The internal battery is NOT
intended to serve as a primary
power source. It should only be
used when other sources are
not available, or briefly when
necessary; for example when
changing power sources.
Whenever the ventilator is connected to AC power, it will
automatically recharge the internal battery. A completely discharged
internal battery will reach 80% charge status within 8 hours when
charging at approximately 23˚ C ambient temperature.
Device Power Source Indicators
There are many power source indicators on the device and the
display screen. These indicators are described in detail below.
AC Power Indicator
When AC power is applied to the device, the green AC LED indicator
(~) on the front of the device lights.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
55
DC Power Indicators
Note: Under normal operating
conditions, the internal battery
symbol will always display on
the Monitoring screen. If the
symbol appears as an empty
red battery on your screen,
contact Philips Respironics
or an authorized service
The internal, detachable, and external battery symbols that will
display on the Monitoring screen are shown below. The detachable
and external battery symbols will only appear on-screen if a
detachable or external battery is attached to the device.
Battery
Symbol
representative to have your
Internal Battery
device serviced.
Detachable Battery
External Battery
There are several DC power indicators that will display on the
Monitoring screen to indicate which battery is in use (if applicable), if
the batteries are low, charging, or discharged, etc. The following table
explains all of the DC power indicators.
DC Power Indicator
Description
Battery In Use Indicator A black box will appear around the battery that is in use. For instance, if
the external battery is currently in use, the
Monitoring screen.
symbol appears on the
Green Fully Charged
Battery Indicator
When a battery is charged to greater than 90% of its capacity, all of the
bars in the battery symbol will appear in green.
Partially Charged
Battery Indicator
When a battery is partially charged, some of the bars in the battery
symbol will appear in green, while others will be clear. For instance, if
the external battery is 50% charged, the following symbol displays on-
screen:
Yellow Low Battery
Indicator (Medium
Priority)
When the device detects that an in-use battery’s charge is low (has
approximately 20 minutes of charge left), the inside of the box
surrounding the battery symbol turns yellow. (In addition to the
battery indicator on the Monitoring screen, a medium priority alarm
message will display indicating “Low Battery.”See Chapter 6 for more
information. The yellow indicator is for the last available battery source.
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
56
DC Power Indicator
Description
Red Low Battery
Indicator
When the device detects that an in-use battery’s charge is nearly
depleted (has approximately 10 minutes of charge left), the inside
of the box surrounding the battery symbol turns red. In addition to
the battery indicator on the Monitoring screen, a high priority alarm
message will display indicating “Low Battery.”See Chapter 6 for more
information. The red indicator is for the last available battery source.
Yellow Battery
Whenever AC power is applied to the device, the internal and
Recharging Symbol (
)
detachable batteries will recharge as needed. If the internal battery is
being recharged, the
being recharged, the
symbol displays. If the detachable battery is
symbol displays.
Battery Disposal
Dispose of the batteries in accordance with local regulations.
First Time Use
When setting up the device for the first time or after annual service,
apply AC power to the ventilator before turning on the blower.
Attempting to use the ventilator without first applying AC power,
such as installing a detachable battery pack and starting the blower,
will cause the internal battery to be displayed in red as an empty
battery. When in this state, the internal battery will not be usable
until AC power is applied.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
57
Connect the Breathing Circuit to the Ventilator
WARNING
Philips Respironics recommends
that a main line outlet bacteria
filter (Part Number 342077) be
used whenever the device is
used for invasive therapy or if
the ventilator may be used on
multiple patients.
Complete the following steps to set up your patient circuit.
1. If you are using a bacteria filter, connect one end of the flexible
tubing to the outlet of the bacteria filter, and connect the
bacteria filter inlet to the breathing circuit connection located on
the side of the ventilator.
If you are not using a bacteria filter, connect the tubing directly
to the device’s breathing circuit connection.
Trilogy Breathing Circuit
Connection
Flexible Tubing
Connecting the Breathing
Circuit to the Device
Bacteria Filter
2. Connect the other end of the flexible tubing to a separate
exhalation device.
Note: The device can be used
with reusable or disposable
circuits. For detailed instructions
on how to set up your device
using a disposable circuit, refer
to the instructions included with
the disposable circuit.
a. If your circuit type is Passive and you are using a Philips
Respironics Passive Exhalation Device:
1.Connect the flexible tubing to the rotating end of the
Passive Exhalation Device.
2.Connect the other end of the Passive Exhalation Device to
the patient interface (e.g., the mask).
Passive
Exhalation
Device
Flexible
Tubing
Connecting the Passive
Exhalation Device
Flexible
Bacteria Filter
Tubing
Whisper
Swivel II
Bacteria Filter
Note: The device does not
provide circuit compliance
compensation in volume modes.
Refer to the instructions included with the Passive
Exhalation Device for more detailed setup information.
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
58
b. If your circuit type is Active PAP:
Active Exhalation Device
with Proximal Pressure
Note: Passive circuit ventilation
provides leak compensation
in both volume and pressure
modes. Active PAP circuit
ventilation does not provide
leak compensation. Particular
care should be used at low tidal
volumes to ensure adequate
monitoring of exhaled tidal
volume.
1. Connect the flexible tubing to an active exhalation device
with proximal pressure.
A. Connect the active exhalation device with proximal
pressure to the flexible tubing that attaches to the
ventilator and to the patient interface (e.g., the
tracheostomy tube).
2. Connect the proximal pressure line and the exhalation
valve line to the active exhalation device with proximal
pressure and the Universal Porting Block on the device as
described below.
WARNING
For ventilator dependent
patients, always have alternate
ventilation equipment, such as
a back-up ventilator, manual
resuscitator, or similar device,
available.
A. Attach the proximal pressure line (.476 cm or 3/16”
line) to the proximal pressure port as shown in the
next illustration.
B. If not attached, connect the other end of the proximal
pressure line to the Trilogy Universal Active PAP Tube
Adapter as shown in the next illustration.
Note: Make sure that
components marked with an
arrow are oriented properly.
When attaching an active
exhalation device with proximal
pressure to the patient, make
sure the proximal pressure port
faces away from the patient.
C. Connect the Trilogy Universal Active PAP Tube Adapter
to the top, striped port on the Universal Porting Block
on the ventilator.
D. Connect the exhalation valve line (.317 cm or 1/8”line)
to the exhalation valve port on the top of the active
exhalation device with proximal pressure.
Note: The
symbol
appears next to the Exhalation
Valve port on the Universal
Porting Block to indicate where
the active exhalation valve line
connects.
E. Connect the other end of the exhalation valve line
to the exhalation valve port on the Universal Porting
Block.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
59
Universal Porting Block
Trilogy Universal Active
Proximal
Pressure
Line
Exhalation Valve Port
PAP Tube Adapter
Proximal
Pressure
Port
Exhalation Valve Line
Exhalation Valve Port
on Universal Porting
Block
Connecting an Active Device
with Proximal Pressure
Refer to the instructions included with the Active Exhalation
Device for more information.
c. If your circuit type is Active Flow:
1. Connect the Flow Sensor to an active exhalation device as
shown below.
Connecting the Flow Sensor
Flow Sensor
Note: If the active exhalation
device has the proximal pressure
line connected when you are
setting up an active flow circuit
type, remove the proximal
pressure line and cap the
proximal pressure port before
attaching to the flexible tubing.
2. Connect the other end of the flexible tubing to the active
exhalation device.
A. Connect the active exhalation device to the flexible
tubing that attaches to the ventilator.
B. Connect the flow sensor to the patient interface (e.g.,
the tracheostomy tube).
Note: Make sure that
components marked with an
arrow are oriented properly.
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
60
C. Connect the flow lines and the exhalation valve line to
the Universal Porting Block on the ventilator as shown
below.
•
•
•
Connect the White Striped Flow Line to
the top, striped port on the Universal
Porting Block on the device.
Note: When using an active
exhalation device with proximal
flow, ensure that any additional
proximal ports are capped.
Connect the other Flow Line to the
middle port on the Universal porting
Block.
Connect the Exhalation Valve Line to the
Exhalation Valve port on the top of the
active exhalation device, and connect the
other end of the line to the Exhalation
Valve port on the Universal Porting Block.
Note: The
symbol
appears next to the Exhalation
Valve port on the Universal
Porting Block to indicate where
the active exhalation valve line
connects.
Universal Porting Block
White Striped
Exhalation Valve Port
on Trilogy Device
Flow Line
Connecting the Active
Exhalation Device with
Proximal Flow Sensor
Flow Line
Exhalation Valve Line
Exhalation Valve Port
on Porting Block
Refer to the instructions included with the active exhalation
device for more information.
Connect a Water Trap
If you are using an optional water trap, connect it to the patient
circuit according to the manufacturer’s instructions.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
61
Connect Supplemental Oxygen (Optional)
To attach supplemental oxygen to the device:
1. Connect the oxygen tubing to the Oxygen Inlet Quick Connect.
Then, connect the Quick Connect to the Oxygen Inlet on the
back of the ventilator.
Oxygen Inlet
Oxygen Inlet
Quick Connect
Oxygen Source
Tubing
Attaching the O2 Tubing
2. Make sure you press the Oxygen Inlet Quick Connect firmly into
the inlet. You should hear a click, and the latch at the top of the
inlet will pop up.
Press Tubing in and Latch
Pops Up
This device is only compatible with a low flow oxygen source
providing up to 15 l/min.
CAUTION
Do not connect an unregulated
or high pressure oxygen source
to this connector on the device.
Chapter 4 Ventilator Setup
Download from Www.Somanuals.com. All Manuals Search And Download.
62
Connect the Remote Alarm (Optional)
WARNING
When using a remote alarm,
make sure you fully test the
remote alarm connector and
cable by verifying that:
You can use a remote alarm as a nurse call system or in-house remote
alarm system. It can generate an alarm at a distance of up to
91 meters (300 feet) from the ventilator. An audible tone sounds
and a red light blinks to indicate that an alarm condition exists. The
remote alarm will sound when any of the following conditions exist:
-
Annunciated alarms on
the ventilator are also
annunciated on the remote
alarm.
•
•
The ventilator is off.
Any alarm occurs and is not silenced or reset.
-
Disconnecting the remote
alarm cable from the
ventilator or from the remote
alarm results in an alarm
notification at the remote
alarm.
Refer to the Accessories chapter of this manual for information on
which remote alarm systems are compatible with the device. Refer to
the instructions included with the Remote Alarm Adapter Cable for
information on how to connect a remote alarm to the ventilator.
WARNING
The remote alarm should be
tested daily.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
5. Viewing and Changing Settings
Note: There is a keypad lock
inactivity time-out period. After
you have unlocked the keypad
as indicated, the keypad will
re-lock after five minutes of
inactivity to prevent someone
from accidentally pressing a
button and changing any of the
settings.
This chapter explains how to scroll through the ventilation screens
and change ventilation settings. It also describes how to connect the
ventilator to the patient once the settings are complete.
Keypad Lock Feature
There is a Keypad Lock feature that users can enable from the
Options menu. It is intended to prevent accidental changes to device
settings. This feature will lock the navigation keys (Up, Down, Stop,
Left, and Right keys). If the Keypad Lock is enabled, a Keypad Unlock
message will display on the bottom of the screen any time you press
one of the navigation keys.
Note: When Keypad Lock is
enabled, the Left, Right, and Up/
Down buttons are locked while
the ventilator is turned on. The
Alarm Indicator/Audio Pause
continues to function normally.
The Start/Stop button is locked
only when this button is used to
stop therapy.
If the keypad is locked, you must unlock it before you can enter the
Menu. To unlock the keypad and enter the menu, you must first hold
the Right button for 5 seconds to unlock the keypad. An audible
indicator sounds when the keypad is successfully unlocked. Once the
display is unlocked, you can enter the Menu as you normally would
by pressing the Up button.
Note: The keypad will
automatically unlock if an
alarm or informational message
occurs and will remain unlocked
the entire time alarms are active.
Note: Pressing the Left (Cancel)
button will cancel the Keypad
Unlock action.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
64
Accessing the Startup and Monitor Screens
1. After you press the
button to begin therapy, the Startup
screen appears momentarily, indicating the device name and the
software version.
2. The next screen that appears is the Monitor screen.
The appearance of this screen will vary, depending on how you
set up the device. If Detailed View is turned off in the Options
menu, your screen will look like the screen shown below.
Note: The symbols that appear
on the Monitor screen are
described in detail later in this
chapter.
Monitor Panel
Date and Time
Panel
Status Panel
Monitor Screen – Detailed
View Off
–
The top section of the screen, called the Monitor panel,
shows the therapy mode and, if you set up a dual
prescription for the patient, the Prescription indicator
appears, indicating Primary or Secondary prescription. The
patient breath symbol also displays during a patient-triggered
breath, and a bar graph displays the current pressure level.
Note: No Soft Buttons display
on the Monitor screen if Keypad
Lock is enabled.
–
–
The center section of the screen displays the current date
and time.
The bottom section, called the Status panel, displays certain
symbols that indicate features being used, such as Ramp, as
well as battery status.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
65
If Detailed View is turned on in the Options menu, the Monitor
screen will look like the screen shown below.
Monitor Panel
Measured Settings
Panel
Monitor Screen – Detailed
View On
Status Panel
This screen contains more detailed information about the
therapy.
–
The top Monitor panel contains the Prescription indicator
if a dual prescription exists, the therapy mode, a graph
displaying the current pressure, and the current date and
time. Additionally, this panel also displays patient pressure,
respiratory rate (RR), exhaled tidal volume (Vte), and leak.
–
–
The second panel in Detailed view is the Measured
Settings panel. It provides patient-related data including
Peak Inspiratory Pressure (PIP), Minute Ventilation, Peak
Inspiratory Flow, Mean Airway Pressure (MAP), and I:E Ratio.
The third panel is the Status panel and shows the same
information displayed in the Detailed View Off screen,
including features in use such as Ramp and battery status.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
66
Note: Dashes on the display
screen indicate that the device
was unable to compute the
parameters. For example, when
the device is first connected to a
patient, the tidal volume, minute
ventilation, and leak may be
dashed until the device is able
to accurately calculate these
patient parameters.
Monitor Screen Indicators
This section describes the following indicators:
•
•
•
Monitor Panel Indicators
Measured Settings Panel Indicators
Status Panel Indicators
Monitor Panel Indicators
All of the indicators that may appear on the Monitor Panel are
described in detail in the following table.
Indicator
Prescription
Description
If you set up a dual prescription for the patient, the words
“Primary”or “Secondary”appear in the top left corner of the panel
to indicate which prescription is active.
Therapy Mode
The current therapy mode displays at the top of the panel (for
example, CPAP, S, S/T, etc.). If a special feature such as Flex, AVAPS,
or Sigh is active, this feature will appear next to the therapy
mode.
Date and Time
Patient Breath
If you are in Detailed view, the current date and time appear
in the top right corner of the panel. (In Detailed View Off, they
appear in the center panel.)
This symbol displays during a patient-triggered breath.
Airway Pressure
Manometer and
Peak Pressure
Symbol
The manometer (bar graph) displays the airway pressure in the
patient circuit at all times. The manometer bar moves to the
right as pressure increases during inhalation, and moves to the
left as pressure decreases during exhalation. The peak pressure
is also indicated on this bar. It is positioned according to the
maximum patient pressure reached during each breath. The Peak
Pressure symbol appears as a blue bar on the manometer. If a
High Inspiratory Pressure alarm occurs, the Peak Pressure symbol
changes from blue to red.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
67
Indicator
Description
Low Pressure
If you enable a volume therapy mode, this indicator appears
below the manometer bar, indicating the low pressure alarm
setting.
Indicator
L
High Pressure
If you enable a volume therapy mode, this indicator appears
below the manometer bar, indicating the high pressure alarm
setting.
Indicator
H
Pressure
This indicator displays the current patient pressure. This only
appears in detailed view.
Respiratory Rate (RR)
This indicator displays the measured respiratory rate in Breaths
Per Minute (BPM). This only appears in detailed view.
Exhaled Tidal
Volume (Vte)
This indicator displays the estimated exhaled tidal volume in
milliliters and reflects compensation for BTPS. This only appears in
detailed view when Passive Circuit is selected.
Inhaled Tidal
Volume (Vti)
This indicator displays delivered tidal volume in milliliters and
reflects compensation for BTPS. This only appears in detailed view
when the Active with PAP Circuit is selected.
Leak
This indicator displays the total leak (non-returned flow), between
the unit outlet and the patient, averaged over the previous
breath. This only appears in detailed view when the Passive
Circuit is selected.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
68
Measured Settings Panel
All of the indicators that may appear on the Measured Settings panel
(available only in Detailed view), are described in the following table.
Indicator
Description
PIP
Peak Inspiratory Pressure displays the maximum pressure
delivered to the patient during the previous breath.
I:E Ratio
Peak Flow
MAP
Displays a comparison of the time spent in inspiration to the time
spent in expiration during the previous breath.
Displays the maximum inspiratory flow delivered to the patient
during the previous breath in l/min BTPS.
Displays the Mean Airway Pressure, which is the weighted
average of pressure in the patient’s airway over 6 breaths.
MinVent
Minute ventilation displays the amount of air delivered to the
patient over the last minute in l/min BTPS.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
69
Status Panel Indicators
All of the indicators that may appear on the Status Panel are
described in the following table.
Indicator
Description
Indicates that the device is in Full Menu Access mode, which
means you can adjust all prescription settings. Philips Respironics
recommends that you change the device to Limited Menu Access
mode before giving the device to the patient, so patients cannot
adjust their prescription settings. Only trained health care
professionals and clinicians should adjust prescription settings.
Displays when a Secure Digital (SD) memory card is inserted in
the ventilator.
Displays when the ventilator detects an error with the SD card.
Displays at all times when an external battery is attached to
the ventilator. The level of green shading shown in the symbol
indicates the battery capacity and will go down as the battery
charge level decreases. When the entire symbol is green, the
battery is fully charged.
Displays at all times when a detachable battery is attached to
the ventilator. The level of green shading shown in the symbol
indicates the battery capacity and will go down as the battery
charge level decreases. When the entire symbol is green, the
battery is fully charged.
Displays at all times, indicating the status of the internal battery. The
level of green shading shown in the symbol indicates the battery
capacity and will go down as the battery charge level decreases.
When the entire symbol is green, the battery is fully charged.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
70
Indicator
Description
A black box displays around the battery that is currently
supplying power to the ventilator when AC power is not available.
(In the status panel shown above, the external battery is in use, so
the
symbol displays.)
A yellow lightning bolt symbol displays with the Detachable or
Internal battery symbol to indicate when the battery is charging.
(In the status panel shown on the previous page, the detachable
battery is being charged, so the
symbol displays.)
Displays when the Alarm Indicator/Audio Pause button has been
pressed and Audio Pause is active. The alarm is silenced for one
minute when the Alarm Indicator/Audio Pause button is pressed.
Displays when the Ramp feature is active.
Note: If a battery in use is very low (less than 20 minutes remaining), the inside of the box surrounding the
battery symbol will change to yellow and all of the bar indicators in the battery will be empty. If a battery in use
is near depletion (less than 10 minutes remaining), the inside of the box surrounding the battery symbol will
change to red, and all of the bar indicators in the battery will be empty. These color changes only occur for the
last available battery source.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
71
On-Screen Button Panel
The illustration below shows the on-screen button panel on the Main
Menu screen, in relation to the buttons on the front of the device.
On-Screen
Button
Panel
Example On-Screen Button
Panel
Note: The on-screen buttons
will vary depending on which
screen is displayed and what
settings are enabled on your
device.
At the very bottom of the display screen is the on-screen button
panel. This panel corresponds with the control buttons on the
ventilator:
•
•
•
The left on-screen button specifies the action for the Left
button on the device.
The center on-screen button specifies the action for the
Up/Down buttons on the device.
The right on-screen button specifies the action for the
right button on the device.
Navigating the Menu Screens
To navigate through all of the menu screens and settings:
•
Use the Up/Down button to scroll through the menu
options and settings.
•
Use the Left and Right buttons to perform the actions
specified on the screens’left and right on-screen buttons.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
72
Note: For some therapy
settings, once you reach the
highest or lowest setting
available, pressing the Up/
Down button again will cycle
back through the settings.
For the parameters that do
not wrap, when you reach
the highest or lowest setting
possible, a “Limit Reached”
message appears in the Menu
Banner on-screen.
Changing and Viewing Settings in Full Menu
Access Mode
Clinicians can view and change settings using the Menu screens
when the device is in Full Menu Access mode. To enter the Menu
screens from the Monitor screen, press the Up button on the
ventilator. The Main Menu screen shown below appears.
Main Menu Screen Example
Note: In the example Main
Menu screen shown, the 2/6
that appears in the Menu
banner indicates that item 2 is
highlighted from a total of 6
items in the menu.
Choose from the following selections on the Main Menu screen:
•
Safely Remove SD Card - This option will appear if an SD
card is inserted in the ventilator. Select this option when
you want to remove the SD card. When the “Remove SD
Card”confirmation message appears, remove the card.
If you press the left (cancel) button or don’t remove the
card within 30 seconds, the confirmation message will
close and the ventilator will continue writing to the card.
•
•
•
•
Settings and Alarms: View and change prescription
settings and alarms.
Note: If you change a setting
but decide you do not want to
save it, you can press the Left
button to cancel the change.
Options: View and change device settings, such as Full or
Limited Access mode, Detailed View, Language, etc.
Alarm Log: View a list of the 20 most recent alarms that
have occurred.
Event Log: View a list of all events that have occurred,
such as ventilator setting changes, ventilator inoperative
conditions, alarms, etc.
•
Information: View detailed information about the device,
such as the device’s software version and serial number.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
73
Note: In the Options menu
described later in this chapter,
you can specify the Pressure
Units displayed by the device,
choosing either cm H2O, hPa, or
mBar. The default setting is
cm H2O, so this manual uses cm
H2O throughout.
Changing the Device Settings and Alarms
From the Main Menu screen, use the Up/Down button to highlight
the Settings and Alarms menu, and press the Right button to select
the menu.
Device Settings Common to All Therapy Modes
Some of the settings on this menu will vary depending on the
therapy mode you select. The section below describes all of the
settings that are common to all therapy modes.
Dual Prescription Setting
You can turn the dual prescription setting on or off. Enable the
setting if you want to create two separate prescriptions for the
patient. For instance, you may want to set up a daytime prescription
and then a separate nighttime prescription. If you enable this setting,
then the menu options on the Main Menu will change to include
three new options:
•
•
•
Switch to Primary/Secondary Settings
Primary Settings and Alarms
Secondary Settings and Alarms
The Main Menu screen will look like the screen below.
Main Menu Screen with Dual
Prescription Enabled
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
74
Mode Setting
Note: See Chapter 3 for detailed
information about each therapy
mode.
You can change the Mode setting on the Settings and Alarms screen
to one of the following therapy modes:
Note: Refer to the chart in
Chapter 3 to easily review all of
the settings available in each
therapy mode.
•
•
•
•
•
•
•
•
•
CPAP
S
S/T
T
Note: This chapter describes
how to enable all of the device
settings, including device
alarms. For detailed information
about each alarm, please refer
to Chapter 6.
PC
PC-SIMV
CV
AC
SIMV
Circuit Type
Note: To change the circuit
type, you must be in the
Setup screen with the airflow
turned off. See the Full Menu
Access Mode section for more
information.
There are three circuit types you can select:
•
•
•
Passive
Active PAP
Active Flow
Note: When the Circuit Type
setting is set to Passive Circuit,
all Ramp Start Pressure settings
in all modes will maintain the
minimum range.
The Passive circuit type uses the Whisper Swivel II passive exhalation
device. The Active PAP circuit type uses an active exhalation device
with a proximal air pressure sensing connection. The Active Flow circuit
type uses an active exhalation device with a proximal flow sensor.
Note: When the Circuit Type
setting is set to Active PAP or
Active Flow, the Flex and AVAPS
features are unavailable.
When using the Passive circuit, the ventilator displays estimated
patient pressures based on the resistance of the standard patient
circuit (Whisper Swivel II with 1.8 meter tubing). Adding accessories
to the patient circuit (humidifier, water trap, etc.) may cause an
increase in circuit resistance and cause the device to display slightly
higher pressures than what is actually delivered to the patient.
With the Active PAP or Active Flow circuit type selected, patient
pressure is measured directly and is not affected by any change in
circuit resistance.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
75
The Passive circuit provides leak compensation. When using the Passive
circuit in Volume Ventilation, the set Vti is delivered to the patient
above the calculated circuit and cuff (or mask) leak. This is different
from traditional active circuit ventilation where the cuff (or mask) leak
reduces the tidal volume delivered to the patient. Volume ventilation
with the Passive circuit delivers an inspiratory tidal volume close to
the device setting regardless of leak; this should be considered when
transitioning a patient from an active to a passive circuit. With a Passive
circuit, Vte is estimated based on the calculated sum of circuit and cuff
(or mask) leak.
The Active Flow circuit monitors proximal flow and proximal
pressure. When using the Active Flow circuit configuration, flow
trigger with leak compensation may be enabled. The default setting
when using the Active Flow Circuit is Leak Compensation On. The
clinician has the option to turn off leak compensation; however,
unintentional leak will not be compensated. Both options measure
the flow at the proximal flow sensor. Leak compensation is not
available in the Active with PAP circuit configuration.
WARNING
Circuit Disconnect
You should not rely on any
single alarm to detect a circuit
disconnect condition. The Low
Tidal Volume, Low Minute
Ventilation, Low Respiratory
This setting enables or disables the circuit disconnect alarm. If
enabled, an audible alarm will sound when a large, continuous air
leak (such as mask removal) has been detected in the circuit.
Rate, and Apnea alarms should
be used in conjunction with the
Circuit Disconnect alarm.
You can choose Off to disable the alarm. Or, you can increase or
decrease the setting from 5 to 60 seconds in 5 second increments.
For example, a setting of 10 means that the alarm will sound after the
circuit has been disconnected for 10 seconds.
Apnea
This setting enables or disables the apnea alarm. If enabled, an
audible alarm will sound when an apnea is detected.
You can choose Off to disable the alarm. Or, you can increase or
decrease the setting from 10 to 60 seconds in 5 second increments.
For example, a setting of 10 means that the alarm will sound if the
time between spontaneous breaths exceeds 10 seconds.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
76
Low Vte
Note: The High and Low Vte
alarms are available when the
Passive or Active Flow Circuit is
selected.
This setting enables or disables the Low Vte alarm. The alarm
activates when the estimated exhaled tidal volume is less than or
equal to this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 40 ml to 2000 ml in 5 ml
increments. It cannot be set higher than the High Vte setting.
When AVAPS is On, the alarm activates when the calculated tidal
volume is less than 90% of the target tidal volume setting. This alarm
can be set to on or off.
High Vte
This setting enables or disables the High Vte alarm. The alarm
activates when the estimated exhaled tidal volume is greater than or
equal to this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 50 ml to 2000 ml in 5 ml
increments. It cannot be set lower than the Low Tidal Volume setting,
except to be turned off.
Note: The High and Low Vti
alarms are only available when
the Active PAP Circuit is selected.
Low Vti
This setting enables or disables the Low Vti alarm. The alarm activates
when the measured inhaled tidal volume is less than or equal to this
setting. You can choose Off to disable this alarm, or you can increase
or decrease the setting from 40 ml to 2000 ml in 5 ml increments. It
cannot be set higher than the High Vti setting.
High Vti
This setting enables or disables the High Vti alarm. The alarm
activates when the measured inhaled tidal volume is greater than or
equal to this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 40 ml to 2000 ml in 5 ml
increments. It cannot be set lower than the Low Vti setting, except to
be turned off.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
77
Low Minute Ventilation
Note: You can increase or
decrease the Low Minute
Ventilation setting in
0.1 l/min increments from
0.1 to 0.99 l/min and 1 l/min
increments from 1 to 99 l/min.
This setting enables or disables the Low Minute Ventilation alarm. The
alarm activates when the calculated minute ventilation is less than or
equal to this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 0.1 l/min to 99 l/min.
High Minute Ventilation
This setting enables or disables the High Minute Ventilation alarm.
The alarm activates when the calculated minute ventilation reaches
or exceeds this setting. You can choose Off to disable this alarm, or
you can increase or decrease the setting from 1 l/min to 99 l/min
in 1 l/min increments. It cannot be set lower than the Low Minute
Ventilation setting except to be turned off.
Low Respiratory Rate
This setting enables or disables the Low Respiratory Rate alarm. The
alarm activates when the measured respiratory rate is less than or
equal to this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM
increments. It cannot be set higher than the High Respiratory Rate.
High Respiratory Rate
This setting enables or disables the High Respiratory Rate alarm.
The alarm activates when the measured respiratory rate reaches or
exceeds this setting. You can choose Off to disable this alarm, or you
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM
increments. It cannot be set lower than the Low Respiratory Rate
except to be turned off.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
78
Low Inspiratory Pressure
This setting configures the Low Inspiratory Pressure alarm. It is only
user-settable in CV, AC, and SIMV modes. It cannot be set lower than
PEEP + 2 cm H2O or higher than the High Inspiratory Pressure. For
passive circuits, you can increase or decrease the Low Inspiratory
Pressure from 6 to 40 cm H2O in increments of 1 cm H2O. For active
circuits, you can increase or decrease the setting from 2 to 40 cm H2O
in increments of 1 cm H2O. For pressure modes, this alarm is not user-
settable.
High Inspiratory Pressure
This setting enables or disables the High Inspiratory Pressure alarm. It
is only user-settable in CV, AC, and SIMV modes. The High Inspiratory
Pressure cannot be set lower than the Low Inspiratory Pressure. You
can increase or decrease the High Inspiratory Pressure from 10 to 80
cm H2O in increments of 1 cm H2O. For pressure modes, this alarm is
not user-settable.
Additional Settings Specific to Therapy Modes
The Settings and Alarms menu also contains many additional
settings specific to the various therapy modes. The specific settings
for each therapy mode are described below.
Continuous Positive Airway Pressure (CPAP) Mode
In addition to the general settings described in the previous section
of this manual, you can also set the following settings in CPAP mode.
Note: If the CPAP pressure is
set to 4 (the minimum setting),
the Ramp Length setting will be
unavailable.
1. CPAP
You can increase or decrease the CPAP pressure setting from 4 to
20 cm H2O in increments of 1.
2. Trigger Type
Note: Trigger Type is not
The device can be set to trigger breaths based on automatic flow
thresholds or specific flow settings. You can select Auto-Trak or
Flow Trigger as the Trigger Type. When Auto-Trak is selected, the
Auto-Trak trigger initiates based on automatic flow thresholds.
available when an Active PAP
or Active Flow circuit is selected.
Flow trigger is the triggering
method used for Active PAP and
Active Flow circuits.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
79
Note: Auto-Trak is only
available if the Passive Circuit is
selected.
When Trigger Type is set to Flow Trigger, Flow Trigger Sensitivity
Leak Compensation, and Flow Cycle Sensitivity become active,
and the trigger initiates based on the Flow Trigger Sensitivity
setting.
Note: Flow Trigger with leak
compensation is only available if
the Active Flow circuit is selected.
3. Flow Trigger Sensitivity
If you set the Trigger Type to Flow Trigger, the Flow Trigger
Sensitivity setting displays. You can increase or decrease the
setting from 1 to 9 l/min in 1 l/min increments.
The flow trigger initiates when the patient’s inspiratory effort
creates a flow equal to or greater than the flow sensitivity
setting.
Note: Enabling Leak
Compensation when using the
Active Flow Circuit configuration
only affects triggering and does
not affect tidal volume delivery
or Vte measurement.
4. Leak Compensation
If you are using an Active Flow circuit, you can turn Leak
Compensation On or Off.
5. Flow Cycle Sensitivity
Note: A flow cycle sensitivity
set at 90% will result in the
most sensitivity. If the flow cycle
sensitivity is set at 10%, this will
result in the least sensitivity.
If you set the Trigger Type to Flow Trigger, the Flow Cycle
Sensitivity setting displays. You can increase or decrease the
setting from 10 to 90 percent (%) in 1% increments.
As flow begins to decrease during inspiration,
if the patient flow is less than the flow cycle set
point, the device will cycle to expiration. For
example: if the flow cycle is set to 75%, when the
flow has decreased by 25% of the peak flow, the
device will cycle to the EPAP/PEEP level.
6. Ramp Length
The Ramp Length allows you to set the ramp time.
You can disable Ramp by selecting Off, or you
can increase or decrease the Ramp Length setting from 5 to 45
minutes in 5-minute increments.
Flow Cycle Sensitivity
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
80
7. Ramp Start Pressure
Note: The Ramp Start Pressure
setting will not display if the
Ramp Length is set to Off or if
the CPAP pressure is set to
4 cm H2O.
You can increase or decrease the ramp start pressure in
increments of 1 from 4 cm H2O to the CPAP pressure setting. The
patient also has access to this setting, unless the ramp length is
set to Off.
Note: Ramp Start Pressure is
less than or equal to CPAP - 1 cm
H2O in CPAP mode.
8. Flex
You can enable or disable the Flex setting. Off disables the
setting and prevents the patient from using Flex. To enable the
setting, set Flex to 1, 2, or 3. The patient also has access to this
setting, if Flex is enabled. The Flex feature is not available when
using an active circuit type.
Note: In CPAP mode, Flex is only
available when CPAP is greater
than 4 cm H2O.
Note: In S mode, Flex is only
available when EPAP is greater
than or equal to 4 cm H2O and
IPAP is less than or equal to
25 cm H2O.
Spontaneous (S) Mode
The following settings, described in the CPAP mode section of this
chapter, also are available in S mode:
Note: Flex is only available
when Auto-Trak is enabled.
Note: Ramp is not available in
the Passive Circuit when IPAP =
EPAP = 4 cm H2O or when IPAP
Min = EPAP = 4 cm H2O.
•
•
•
•
•
•
Flex
Trigger Type
Flow Trigger Sensitivity
Flow Cycle Sensitivity
Ramp Length
Note: Ramp Start Pressure is
less than or equal to EPAP - 1 cm
H2O in S, S/T, T, and PC modes.
Ramp Start Pressure
Note: Ramp Start Pressure is
greater than or equal to 0 cm
H2O when the circuit type is
Active PAP or Active Flow in S,
S/T, T, and PC modes.
In addition to those settings, the settings below are also available in
S mode.
1. AVAPS
Note: Ramp Start Pressure
is greater than or equal to 4
cm H2O when the circuit type
is Passive in S, S/T, T, and PC
modes.
AVAPS is only available if Flex is not enabled.
You can disable AVAPS by selecting Off, or you can enable AVAPS
by selecting On. If you select Off, the IPAP setting displays. If you
select On, the IPAP Max Pressure and IPAP Min Pressure display.
Note: AVAPS is only available if
the Passive Circuit is selected.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
81
2. IPAP
The IPAP setting displays if AVAPS is Off. You can increase or
Note: IPAP, IPAP Max, or IPAP
Min cannot be set to more than
30 cm H2O above EPAP.
decrease the Inspiratory Positive Airway Pressure (IPAP) from 4
to 50 cm H2O in increments of 1. IPAP is limited to a maximum of
25 cm H2O when Flex is enabled. You cannot set the IPAP setting
lower than the EPAP setting.
3. IPAP Max Pressure
The IPAP Max Pressure setting displays if AVAPS is enabled.
You can increase or decrease the setting from 4 to 50 cm H2O
in increments of 1. The IPAP Max Pressure must be equal to or
greater than the IPAP Min value.
4. IPAP Min Pressure
The IPAP Min Pressure setting displays if AVAPS is enabled.
You can increase or decrease the setting from 4 to 50 cm H2O
in increments of 1. The IPAP Min Pressure must be equal to or
greater than the EPAP value, and it must be less than or equal to
the IPAP Max Pressure.
Note: EPAP cannot be set to
more than 30 cm H2O below
IPAP, IPAP Max, or IPAP Min.
5. EPAP
You can increase or decrease the Expiratory Positive Airway
Pressure (EPAP) from 4 to 25 cm H2O in increments of 1. For
active circuits, EPAP can be set to zero.
When AVAPS is disabled, the EPAP setting must be less than
or equal to the IPAP setting. When AVAPS is enabled, the EPAP
pressure must be less than or equal to the IPAP Min Pressure.
6. Tidal Volume
Note: In CV, AC, and SIMV
modes, the tidal volume setting
is limited by the Inspiratory
Time, to maintain the system’s
minimum and maximum peak
flows.
The Tidal Volume setting displays if AVAPS is enabled. You can
increase or decrease the setting from 50 to 2000 ml in 5 ml
increments. Use this setting to establish the target volume of
gas which the ventilator will produce and deliver during each
Spontaneous breath.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
82
Note: The Rise Time setting only
displays if Flex is disabled. If Flex
is enabled, the device will use a
rise time of 3.
7. Rise Time
You can adjust the rise time to find the most comfortable setting
for the patient. Increase or decrease the setting from 1 to 6
until you find the right setting. The rise time levels from 1 to 6
progressively reflect slowed response of the pressure increase
that will take place at the beginning of inspiration.
Note: In S/T, T, PC, PC-SIMV,
SIMV, CV, and AC modes, the
Apnea Rate is greater than or
equal to the Breath Rate and is
limited by the current Inspiratory
Time setting to maintain a
minimum 1:1 I:E ratio.
8. Apnea Rate
If the Apnea alarm is enabled, you can set the Apnea Rate from
4 to 60 BPM in 1 BPM increments. In S mode, the Apnea Rate is
greater than or equal to 1:2 I:E ratio.
Spontaneous/Timed (S/T) Mode
All of the settings described in the S Mode section are also available
in S/T mode, except for the Flex setting. In addition to those settings,
the settings below are also available in S/T mode.
Note: In volume modes, the
Breath Rate range is limited by
the current Inspiratory Time
setting to maintain a minimum
1:1 I:E ratio.
1. Breath Rate
In AC mode, you can increase or decrease the Breath Rate setting
from 0 to 60 BPM, while in all other modes, you can increase
or decrease the setting from 1 to 60 BPM in 1 BPM increments.
Use the Breath Rate setting to establish the minimum rate of
mandatory breaths that the ventilator will deliver per minute.
Note: In pressure modes, the
inspiratory time range is limited
by the current Breath Rate
setting to maintain a minimum
1:1 I:E ratio.
2. Inspiratory Time
You can adjust the Inspiratory Time setting from 0.3 to 5.0
seconds in 0.1 second increments. Inspiratory Time is the
duration for the inspiratory phase of a mandatory breath.
Note: In volume modes, the
inspiratory time range is limited
by the current Tidal Volume and
Breath Rate settings to maintain
a minimum 1:1 I:E ratio and
the system’s minimum and
maximum peak flow.
Timed (T) Mode
All of the settings available in S/T mode are available in T mode,
except for the Trigger Type setting. Please refer to the descriptions in
the S and S/T Mode sections of this chapter for detailed information
on the T mode settings.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
83
Pressure Control (PC) Mode
All of the settings available in S/T mode are available in PC mode,
except for the Flow Cycle Sensitivity setting (when Flow Trigger is
enabled). Please refer to the descriptions in the S and S/T Mode
sections of this chapter for detailed information on the PC mode
settings.
Pressure Control Synchronized Intermittent Mandatory Ventilation
(PC-SIMV) Mode
The following settings, described in the S and S/T mode sections of
this chapter, also are available in PC-SIMV mode:
•
•
•
•
Breath Rate
Inspiratory Time
Trigger Type
Flow Trigger Sensitivity (if Trigger Type is set to Flow
Trigger)
•
•
Flow Cycle Sensitivity (if Trigger Type is set to Flow
Trigger)
Rise Time
In addition to these, the following settings are also available in PC-
SIMV mode.
Note: In PC-SIMV mode, you
cannot set up Pressure Support
for Mandatory and Assist
breaths (Pressure - PEEP) greater
than 30 cm H2O.
1. Pressure
You can increase or decrease the Pressure setting from 4 to 50
cm H2O in increments of 1. This is the pressure the ventilator will
deliver during the inspiratory phase of a mandatory or assist
breath.
Note: The Pressure Support and
PEEP settings together cannot
exceed 50 cm H2O.
2. Pressure Support
You can increase or decrease the Pressure Support setting from
0 to 30 cm H2O in increments of 1. This is the pressure support
the ventilator will deliver during the inspiratory phase of a
Spontaneous breath.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
84
3. PEEP
The Positive End Expiratory Pressure (PEEP) setting can be
increased from 0 to 25 cm H2O in active circuits and 4 to 25 cm
H2O in passive circuits, in increments of 1. PEEP is the positive
pressure maintained in the patient circuit during exhalation. The
PEEP must be less than or equal to the pressure setting.
Control Ventilation (CV) Mode
The following settings, described in the previous sections of this
chapter, are also available in CV mode:
•
•
•
•
•
•
Tidal Volume
Breath Rate
Inspiratory Time
PEEP
High Inspiratory Pressure
Low Inspiratory Pressure
In addition to these, the following settings are also available in CV
mode.
Note: The Flow Pattern setting
might be limited to only Ramp
or Square based on the Tidal
Volume, Inspiratory Time, and
Breath Rate settings to maintain
the minimum and maximum
peak flows.
1. Flow Pattern
You can choose either Ramp or Square for the Flow Pattern setting.
2. Sigh
You can enable or disable the Sigh setting by selecting On or Off.
A Sigh is a breath that is delivered every 100 breaths at 150% of
the normal volume.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
85
Assist Control (AC) Mode
The AC mode contains the following settings described in the S, S/T,
PC-SIMV, and CV mode sections in this chapter. Please refer to the
descriptions in those sections for detailed information.
•
•
•
•
•
•
•
•
•
•
Tidal Volume
Breath Rate
Inspiratory Time
Flow Pattern
Note: The Low Inspiratory
Pressure is limited to PEEP +2 in
CV, AC, and SIMV modes.
PEEP
Trigger Type
Flow Trigger Sensitivity
Sigh
Note: Flow Cycle Sensitivity is
not available in AC mode.
High Inspiratory Pressure
Low Inspiratory Pressure
Synchronized Intermittent Mandatory Ventilation (SIMV) Mode
The SIMV mode contains the following settings described in the S,
S/T, PC-SIMV, and CV mode sections in this chapter. Please refer to the
descriptions in those sections for detailed information.
•
•
•
•
•
•
•
•
•
•
•
Tidal Volume
Breath Rate
Inspiratory Time
Pressure Support
Flow Pattern
PEEP
Trigger Type
Sigh
Rise Time
High Inspiratory Pressure
Low Inspiratory Pressure
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
86
Viewing and Changing Options Menu Items
From the Main Menu screen, select the Options item.
The Options menu appears, shown in the screen below.
Options Menu
The following settings are available on the Options menu.
•
Menu Access – You can select Full or Limited menu
access. Full menu access allows operators to access all
ventilator and prescription settings. Limited menu access
allows operators to access only certain settings and
does not allow them to change prescription settings. To
prevent patients from changing prescription settings, do
not give them Full menu access.
•
•
Detailed View – You can turn Detailed View on or off
using this setting. Detailed view displays additional
settings and therapy information on the Monitor screen.
Language – The next item on the Options menu allows
you to select the Language that the software will appear
in (English, French, German, etc.). The information on the
screens will display in the language selected here.
•
Pressure Units – The next item allows you to select the
pressure units that will display on the screens. You can
choose either:
-
-
-
cm H2O
hPa
mBar
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
87
All pressure units on the screens will display in the unit of
measure selected here.
•
•
Alarm Volume – You can adjust the volume of the device
alarms using this setting. Select either Loud or Soft as the
alarm volume options.
WARNING
Make sure the alarm volume is
set loud enough to be heard by
the caregiver. Consider the use of
a remote alarm.
Keypad Lock – You can enable or disable the Keypad
Lock feature, which is described in detail earlier in this
chapter. Enabling the Keypad Lock feature can prevent
someone from accidentally pressing a button and
changing any of the settings. Select On to enable the
feature or Off to disable it.
•
Keypad Backlight – The next item you can set is the
Keypad Backlight. You can turn the backlight On or Off
using this setting. Whenever you press the
button to
begin therapy, the keypad backlight temporarily lights
up. Once therapy is being provided, the keypad will be lit
according to this Keypad Backlight setting. If the setting
is On, the backlight remains on while therapy is provided.
If the setting is Off, the backlight remains off while
therapy is provided.
•
•
LCD Brightness – The LCD display is lit by a backlight.
The backlight turns on when the initial Startup screen
displays. You can adjust the brightness of the LCD
backlight from 1 – 10, with 1 being the dimmest setting
and 10 being the brightest.
Screen Saver – You can change the screen saver to
reduce power consumption or dim the screen in a dark
room. You can choose the following settings:
Note: Setting the screen saver
to Black allows the device to run
for a longer period of time on
battery power.
-
-
-
Off: No screen saver displays and the LCD backlight remains
lit at your brightness setting.
Breath: The display appears as a black screen, with only the
patient breath indicator and manometer visible.
Black: The display’s backlight is turned off and the display is
black with no information visible.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
88
-
Dim: The display’s backlight is decreased, so that the display
is still visible but not as bright.
If enabled, the screen saver will display after 5 minutes of no
keypad activity. Pressing any button on the device will exit the
screen saver. And, any alarm or informational message will also
exit the screen saver.
•
•
Date Format – You can choose either mm/dd/yyyy or
dd/mm/yyyy as the date format that will display on the
device screens.
Time Format – You can choose to display either an AM/
PM time format or 24 Hour time format (for example,
2:49 PM or 14:49).
•
•
Month – The month defaults to the current month. The
adjustable range is from 1 (January) – 12 (December).
Day – The day defaults to the current day. The adjustable
range is from 1 – 31. The maximum value is based on the
selected month.
•
•
Year – The year defaults to the current year. The
adjustable range is from 2000 – 2099.
Hour – The hour defaults to the current hour. The
adjustable range is from 12 AM – 12 PM or 0-23,
depending on the selected Time Format.
•
•
Minute – The minute defaults to the current minute. The
adjustable range is from 0 – 59.
IP Address Mode – You can change the IP address
mode to either DHCP or Static, depending on the type of
network you are using (if applicable).
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
89
•
Operational Hours – The operational hours displays
the total number of hours that the device blower has
been on since the last time this value was reset. You can
reset this value to zero (0) if desired (e.g., each time you
give the device to a new patient). This value helps you
determine how often the patient is using the device. The
Operational Hours shown here differs from the Blower
Hours shown on the Information screens. The Blower
Hours displayed in the Information screen is the total
number of hours that the blower has been working over
the life of the device. You cannot reset this value.
Viewing the Alarm Log
From the Main Menu screen, you can select Alarm Log to access the
Alarm Log screen. An example is shown below.
Note: In the Alarm Log screen,
the 1/2 shown in the Menu
banner indicates that page 1
of 2 alarm log pages is being
viewed at this time.
Alarm Log Screen
The alarm log displays the alarms in chronological order with the
most recent events displayed first. It lists the 20 most recent alarms
or messages that appeared on the device display. When the device is
in Limited Menu access mode, the alarm log cannot be cleared. It can
Note: In Full Menu access mode,
you can press the Right (Clear)
button to clear the alarm log if
be cleared when in Full Menu access mode. Depending on how many desired.
alarms have occurred, the alarm log may be several pages long. The
entries in the alarm log use the same names that displayed when the
alarm initially occurred and was displayed in the Alarm View.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
90
Note: In Full Menu access mode,
you can press the Right (Clear)
button to clear the event log if
desired.
Viewing the Event Log
From the Main Menu screen, you can select Event Log to access the
Event Log screen.
The event log displays a list of all events that have occurred, such as
ventilator setting changes, ventilator inoperative conditions, alarms,
etc., in chronological order with the most recent events displayed
first. When the device is in Limited Menu access mode, the event
log is not available. It can be viewed and cleared when in Full Menu
access mode. You can page through the event log if it is multiple
pages. The number of pages appears in the upper right corner of the
panel.
In the event log descriptions, any description beginning with a 1:
or 2: is a prescription change event. The 1 represents a change to a
primary setting and the 2 represents a change to a secondary setting.
This is followed by the setting that was changed.
The last two columns indicate setting and alarm changes. If the entry
is a setting change, the first column shows the old setting value and
the last column shows the new setting. If the entry is an alarm, the
first column shows the value that triggered the alarm and the last
column shows the number of seconds that the alarm was active.
Viewing Device Information
From the Main Menu screen, you can select Information to access
the Information screen. You can also view the Information screen by
holding the Down key for 5 seconds. This causes the detailed view
of the Monitor Screen and the Information Menu to be displayed
temporarily. This key sequence is valid from the Monitor Screen while
in Full or Limited Access.
The Information screen provides you with a summary of the current
prescription settings, device settings, and system settings. You can
use the Up/Down buttons to scroll through the information.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
91
Updating Prescriptions Using the SD Card
WARNING
When you change the device
prescription, alarms and other
settings using the SD card,
Trilogy200 requires that the
caregiver review and verify the
changes prior to the changes
being used by the device.
The caregiver or health care
professional is responsible to
ensure that the prescription
settings are correct and
With the Trilogy200, you can update the patient’s prescription using
the SD Card. This feature lets you update a single prescription or
both prescriptions if the dual prescription feature is enabled. The
prescription update can occur either when the ventilator is off or on.
1. Insert an SD Card with a valid prescription into the device. A
“Change Prescription?” message appears on the display:
2. Select No to cancel the prescription update process and return
to the previous display (the black screen if the airflow was off or
the Monitor/Standby screen if the airflow was on). Select Yes to
start the prescription update process. Once the prescription is
read in and validated, one of the following screens appears on
the display to allow you to ensure the prescription is correct:
compatible with the patient
after using this feature. Installing
the wrong prescription for a
particular patient may result
in improper therapy, lack of
appropriate safety monitoring,
and risk of death or injury to the
patient.
Ventilator Off Prescription
Display Screen
Ventilator On Prescription
Display Screen
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
92
3. Select Cancel to cancel the prescription update process and
Note: Make sure you closely
review the prescription and
confirm that all settings are
correct.
return the screen to the initial state before the prescription
update started. Select Page to review the entire prescription.
The Menu Banner will reflect the prescription being updated.
4. Once the entire prescription has been reviewed, a screen
displays with the option to Cancel or OK the changes. Select
Cancel to cancel the prescription update process and return the
screen to the initial state before the prescription update started.
Select OK to complete the prescription update and display the
Prescription Change confirmation screen.
Note: If both prescriptions of
a dual prescription are being
updated, you must accept the
primary prescription using
the OK button before being
able to review the secondary
prescription. Both prescriptions
have to be accepted before any
changes are made.
If the SD card is removed at any time during the prescription
update, the process aborts and the screen returns to the initial
state before the prescription update started.
A confirmation screen will appear if errors occur during this process.
The following table summarizes the prescription errors, their possible
causes, and the actions to take.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
93
Message
Cause
Action
Prescription Change Failed
Displayed when an error
Remove the card and have the
occurs during the prescription prescription replaced with a
update. This is caused by an
attempt to set a value for a
prescription parameter that
is not valid for the therapy
mode or an attempt to set a
prescription parameter to an
invalid value.
valid prescription.
Prescription Failed – Circuit
Type
Displayed when the circuit
type in the new prescription
If the circuit type in the
prescription is correct, remove
doesn’t match the circuit type the card, change the circuit
set for the ventilator.
type on the ventilator, and
re-insert the card. If the
circuit type on the ventilator is
correct, remove the card and
replace the prescription on
the card with a prescription
containing the correct circuit
type.
Prescription Failed – Card is
Read Only
Displayed when the
prescription is write-protected. that the small switch on the
Remove the card and check
side of the SD Card is not in the
Locked position.
If you continue to receive this
message, remove the card and
have the prescription replaced
with a valid prescription.
Prescription Failed – Serial
Number
Displayed when the device
serial number on the new
prescription does not match
the serial number for the
device.
Remove the card and have the
prescription replaced with the
prescription with the correct
serial number.
Prescription Failed – Version
Displayed when the version
of the prescription does not
Remove the card and have
the prescription replaced with
match the version accepted by a prescription in the correct
the device.
version.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
94
Changing and Viewing Settings in Limited
Menu Access Mode
Note: If the Keypad Lock feature
is enabled, a message will
appear that says “Hold Right Key
for 5 seconds to unlock.” Once
you hold the key down for 5
seconds, the keypad will unlock
and you can enter the Main
Menu screen. The Keypad Lock
feature is explained in detail
later in this chapter.
After you press the
button and access the Monitor screen, you
can view and change settings using the Menu screens.
To enter the Menu screens from the Monitor screen, press the Up
button on the ventilator. The Main Menu screen appears, shown
below.
Note: In the example Main
Menu screen shown, the 1/5
that appears in the Menu
banner indicates that item 1 is
highlighted from a total of 5
items in the menu.
Main Menu Screen
You can choose from the following selections on the Main Menu
screen:
•
Switch to Primary (or Secondary) Settings: If your health care
professional has set up a dual prescription for you, this option
will appear. You can select it to change to either your Primary or
Secondary prescription settings.
•
Safely Remove SD Card: This option will appear if an SD card
is inserted in the ventilator. Select this option when you want to
remove the SD card. When the “Remove SD Card”confirmation
message appears, remove the card. If you press the left (cancel)
button or don’t remove the card within 30 seconds, the
confirmation message will close and the ventilator will continue
writing to the card.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
95
Note: The Options, Alarm Log,
and Information items are
discussed in detail earlier in this
chapter, in the “Changing and
Viewing Settings in Full Menu
Access Mode” section.
•
•
My Settings: View and change certain prescription settings,
such as rise time or ramp starting pressure, if these settings were
enabled by your health care professional.
Options: View and change certain device settings,
such as your alarm volume, keypad lock, or keypad
backlighting.
•
•
Alarm Log: View a list of the 20 most recent alarms that
have occurred.
Information: View detailed information about your
device, such as the device’s software version and serial
number.
Activating Your Primary or Secondary Prescription
Note: If you are currently using
the primary prescription, the
menu option will say “Switch to
Secondary Settings.” If you are
currently using the secondary
prescription, the option will say
“Switch to Primary Settings.”
If your health care professional has set up a dual prescription for
you, follow the steps below to change your prescription selection.
Your health care professional might use this feature if, for instance,
you need one prescription for daytime use and a second, different
prescription for nighttime use.
1. Use the Up/Down (Navigate) button to highlight the “Switch to
Primary (or Secondary) Settings”option.
2. Press the Right (Select) button.
A screen appears with a confirmation message, saying “Switch to
Primary Settings”or “Switch to Secondary Settings”depending
on which prescription you are changing to.
3. Press the Right (Yes) button to switch to the new prescription.
If you decide not to change the prescription, press the Left (No)
button. The display returns to the Main Menu after you’ve made
your selection.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
96
Note: The options on the My
Settings screen will vary depending
on how your health care
Viewing and Changing My Settings Menu Items
To view or change the therapy settings available on the My Settings
screen, use the Up/Down (Navigate) button to highlight the My
Settings option on the Main Menu, and press the Right (Select)
button. The My Settings Menu screen appears.
professional has set up your device.
Note: For some therapy
settings, once you reach the
highest or lowest setting
available, pressing the Up/
Down button again will cycle
back through the settings.
For the parameters that do
not wrap, when you reach
the highest or lowest setting
possible, a “Limit Reached”
message appears in the Menu
Banner on-screen.
Follow the general instructions below to navigate and change any of
the therapy settings. Detailed information about each setting follows.
1. From the My Settings screen, use the Up/Down button to
navigate to the setting you want to change and highlight it.
2
To modify a setting once it is highlighted, press the Right
(Modify) button.
3. Use the Up/Down (Edit) button to scroll through the available
settings. Press Down to decrease the setting, or press Up to
increase the setting.
Note: After you select “OK” to
save the new setting, the next
setting in the list is automatically
highlighted.
4. Once you have chosen the setting you want, press the Right (OK)
button to save the new setting. Or, if you decide not to change
the setting, press the Left (Cancel) button.
Note: If some features or
settings are not enabled on your
ventilator, they will not appear
on your My Settings screen. For
instance, if Ramp is not enabled,
the Ramp Start Pressure setting
does not appear on the screen.
5. You can now either navigate to the next setting you want to
change using the Up/Down (Navigate) button, or exit the My
Settings menu by pressing the Left (Finish) button to return to
the Main Menu.
You can change the following settings in the My Settings menu, if
they are enabled by your health care professional.
•
Rise Time – The Rise Time is the time it takes the
ventilator to change from expiration to inspiration. If this
feature is enabled, you can adjust the Rise Time from 1
to 6 to find the setting that provides you with the most
comfort. A setting of 1 is the fastest Rise Time, while 6 is
the slowest.
•
Ramp Start Pressure – The ventilator is equipped
with an optional Ramp feature. Ramp reduces the
pressure and then gradually increases the pressure to
your prescription setting so you can fall asleep more
comfortably.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
97
Note: If the Flex feature is
If your health care professional has enabled Ramp on
your ventilator, you can adjust the Ramp Start Pressure
setting. The setting can be adjusted from 4.0 to your
prescription pressure setting in increments of 1.
enabled, the Rise Time setting
will not appear on your My
Settings screen and cannot be
adjusted. When Flex is enabled,
the Rise Time is fixed at a setting
of 3.
•
Flex – The optional Flex feature allows you to adjust the
level of air pressure you feel when you exhale during
therapy. If this feature is enabled by your health care
professional, you can adjust the setting from 1 to 3.
Note: When adjusting the Flex
setting, it is recommended that
you start with the minimum
setting of 1, which provides
the least relief. Levels 2 and
3 progressively increase the
pressure relief.
Note: If you are using an Active
PAP circuit, the Flex feature is
unavailable.
Note: Flex is only available if
Auto-Trak is enabled.
Chapter 5 Viewing and Changing Settings
Download from Www.Somanuals.com. All Manuals Search And Download.
98
Note: Trilogy200 provides
features to control access to the
device (prescription) settings
and to transfer new device
settings onto the device using
the SD Card. These features
are intended to be used as part
of your clinical/institutional
procedures governing the use,
security, and control of this
medical device.
Connecting the Ventilator to the Patient
After you have finished adjusting your ventilator settings, perform
the following steps to connect the ventilator to the patient.
1. Perform System Checkout
Do not connect the ventilator to the patient until you perform the
system checkout procedures defined in Chapter 10.
2. Start Therapy
WARNING
Press the
button to begin therapy. When you start therapy, the
display backlight and the backlights on the buttons turn on, the red
To make sure the device is
and yellow alarm LEDs turn on momentarily, and an audible indicator operating properly at start-up,
always verify that the audible
tone sounds and the alarm
LEDs light red and then yellow
sounds to indicate that therapy has started. The Startup screen
appears on the display.
momentarily. Contact Philips
Respironics or an authorized
service center for service if these
3. Connect the Breathing Circuit to the Patient Interface
After you have assembled the system, started therapy, and adjusted
your ventilator settings as needed, you can connect the breathing
circuit to the patient. The illustration below shows the breathing
circuit connected to a mask. You can also connect the breathing
circuit to a tracheostomy tube.
indications do not occur at
start-up.
Example of a Fully
Connected Patient Breathing
Circuit with Passive
Exhalation Device
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
6. Ventilator Alarms
This chapter describes the ventilator alarms and what you should do
if an alarm occurs.
There are three types of alarms:
Note: If multiple alarms occur
•
•
•
High Priority – Require immediate response by the
operator
at the same time, all alarms are
processed and displayed, but
the alarms are ordered first by
priority and then by occurrence,
with the newest, highest priority
alarms at the top of the list.
The alarm precedence is in the
following order: high priority,
medium priority, low priority,
and informational messages.
Medium Priority – Require prompt response by the
operator
Low Priority – Require operator awareness. These alarms
alert you to a change in the ventilator status.
Additionally, the ventilator also displays informational messages and
confirmation alerts that notify you of conditions that need attention
but do not qualify as alarm conditions.
Note: Not all alarms are
available in every therapy
mode; some alarms are mode-
dependent.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
100
Audible and Visual Alarm Indicators
When an alarm condition occurs:
•
The alarm LED indicator on the Alarm Indicator/Audio
Pause button lights
•
•
The audible alarm sounds
A message appears on the screen describing the type of
alarm
•
The remote alarm (if applicable) is activated
Each of these is described in detail below.
Alarm LED Indicators
The Alarm Indicator/Audio Pause button on the front of the ventilator
lights up as follows whenever an alarm is detected:
Alarm Indicator/Audio Pause
Button
•
•
•
Red Flashing Indicator – When the device detects a high
priority alarm, the Alarm Indicator/Audio Pause button
flashes red.
Yellow Flashing Indicator – When the device detects a
medium priority alarm, the Alarm Indicator/Audio Pause
button flashes yellow.
Yellow Solid Indicator – When the device detects a low
priority alarm, a solid yellow light appears on the Alarm
Indicator/Audio Pause button.
The Alarm Indicator/Audio Pause button does not light up when
informational messages or confirmation alerts display.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
101
Audible Indicators
An audible indicator sounds whenever a power failure or a high,
medium, or low priority alarm is detected. Additionally, an audible
indicator sounds for informational messages and to confirm that
certain actions have occurred (for example, when an SD card is
inserted or removed from the device).
•
Ventilator Inoperative Audible Indicator – When a
ventilator inoperative alarm occurs, a continuous audible
alarm sounds. The alarm descriptions later in this chapter
display this indicator as:
Note: For the alarm indicators
noted throughout this manual,
each “dot” represents an audible
beep.
•
Power Failure Audible Indicator – When a power failure
occurs, a series of beeps sounds in a 1 beep pattern,
repeating one second on, then one second off. The alarm
descriptions later in this chapter display this indicator as:
•
•
•
High Priority Audible Indicator – When a high priority
alarm is detected, a series of beeps sound in the
following pattern, which is repeated twice: 3 beeps, a
pause, and then 2 more beeps. This indicator continues
until the cause of the alarm is corrected or the audible
alarm is paused. The alarm descriptions later in this
chapter display this indicator as:
• • • • • • • • • •
•
•
Medium Priority Audible Indicator – When a medium
priority alarm is detected, a series of beeps sound in a
3-beep pattern. This pattern repeats until the cause of
the alarm is corrected or the audible alarm is paused.
The alarm descriptions later in this chapter display this
indicator as: • • •
Low Priority Audible Indicator – When a low priority
alarm is detected, a series of beeps sound in a 2-beep
pattern. This pattern repeats until the cause of the alarm
is corrected or the audible alarm is paused. The alarm
descriptions later in this chapter display this indicator as:
• •
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
102
•
Informational Messages and Confirmation Audible
Indicators – When an informational message appears
on screen, a brief, 1- beep audible indicator sounds.
Additionally, when the device detects that a certain
action has been completed (for example, when the
Start/Stop button is pressed to start therapy, or when an
SD card is inserted or removed from the device) a brief,
1- beep audible indicator sounds. The alarm descriptions
later in this chapter display this indicator as: •
Alarm Messages
When the ventilator detects an alarm, the Alarms and Messages
Screen is displayed showing a description of the alarm condition.
When an alarm message appears, it will be highlighted in red if it
is a high priority alarm or in yellow if it is a medium or low priority
alarm. (The highlight color matches the alarm LED color on the Alarm
Indicator/Audio Pause button.) If an alarm is manually reset by the
user, the Alarms and Messages screen is removed and the Monitoring
Screen is re-displayed. If the alarm self-cancels, the Alarms and
Messages screen remains displayed, but the highlight for the active
alarm is removed, the LED is unlit, and the audible alarm stops. The
screen below is an example of a possible alarm message.
Note: An alarm message will
also display in the Menu Banner
if a menu is active when an
alarm occurs.
Sample Alarms and
Messages Screen
If a menu is displayed on the screen when an alarm occurs, the
description of the newly generated alarm is displayed in the menu
banner area. This is done so that the modification to the current
parameter can be completed before addressing the alarm condition in
case the modification affects the alarm condition. The screen below is an
example of an alarm message displayed in the menu banner.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
103
Alarm in Menu Banner
Screen
The Alarms and Messages Screen will automatically display in place
of the Monitor screen when exiting from the menu system using the
Exit soft key when an alarm is displayed in the menu banner. If an
alarm is manually reset by the user or self-cancels, the menu banner
on-screen before the alarm occurred will reappear.
If a Ventilator Inoperative alarm occurs, the entire display screen
turns red and the Ventilator Inoperative message appears on-screen,
as shown below.
Ventilator Inoperative Alarm
Screen
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
104
To turn the ventilator off from a Ventilator Inoperative condition,
use the normal power off sequence. When the Start/Stop button is
selected, the following screen will display.
Ventilator Inoperative Power
Off Screen
Select the Right button (Yes) to turn the ventilator off and stop the
audible alarm. Selecting the Left button (No) will return the screen to
the Ventilator Inoperative Alarm Screen without silencing the audible
alarm.
Remote Alarm
When the ventilator detects an alarm condition, if you are using a
remote alarm system, a signal is sent from the ventilator to activate the
remote alarm. The Remote Alarm is discussed in Chapter 4.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
105
Audio Pause and Alarm Reset Features
This section describes the Audio Pause and Alarm Reset features.
Audio Pause
When an alarm occurs, you can temporarily silence the audible
indicator by pressing the Alarm Indicator/Audio Pause button. The
alarm is silenced for 60 seconds and then will sound again if the
cause of the alarm has not been corrected. Each time the Alarm
Indicator/Audio Pause button is pressed, the alarm silence period
resets to one minute.
When Audio Pause is active, the Alarm Indicator/Audio Pause symbol
(
) appears if you are on the Monitor screen. Additionally, an
“Audio Pause”message displays in the menu banner on the Alarm
Display screen.
You can ‘pre-silence’alarms that have not yet occurred by pressing
the Alarm Indicator/Audio Pause button while no alarms are active.
Then, if an alarm occurs, the audible indicator does not sound until
the Audio Pause time limit has expired. This can be useful during
patient setup, so caregivers can prevent alarms from sounding that
they know will occur during the setup process. The LED and display
will still show the alarm, but the audible alarm will not sound.
Alarm Reset
The Reset button is used to clear the currently active alarm(s) from
the display and stop the LED and audible alarm indicator. This button
should be selected after the situation causing the alarm(s) has been
corrected. All active alarms are cancelled and alarm detection is
restarted when this button is selected.
The ventilator self-cancels certain alarms if the cause of the alarm is
corrected, shutting off the alarm LED, the audible alarm, and the alarm
background color. However, the alarm text remains on the screen. You
can manually reset an alarm by pressing the Left button (Reset). The
Audio Pause function is cancelled if the alarm is manually reset.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
106
Alarm Volume Control
WARNING
As explained in Chapter 5, you can adjust the Alarm Volume from
the Options menu. You can select Loud or Soft, depending on your
preference.
Make sure the alarm volume is
set loud enough to be heard by
the caregiver. Consider the use of
a remote alarm.
What to Do When An Alarm Occurs
WARNING
You should not rely on any
single alarm to detect a circuit
disconnect condition. The Low
Tidal Volume, Low Minute
Ventilation, Low Respiratory
Rate, and Apnea alarms should
be used in conjunction with the
Circuit Disconnect alarm.
Complete the following steps when an alarm occurs:
1. Whenever an alarm occurs, first always observe the patient
and ensure that adequate ventilation and oxygenation (if
appropriate) are available.
2. Look at the alarm indicators and listen to the audible alarm
sound. Note the color of the Alarm Indicator/Audio Pause button
(red or yellow) and whether the LED is solid or flashing.
3. Look at the display to check the alarm message that appears on-
screen and whether it is highlighted in red or yellow.
4. Press the Alarm Indicator/Audio Pause button to temporarily
silence the audible alarm. A visual indicator displays if you are on
the Monitor screen (
), or an “Audio Pause”message appears in
the menu banner on the Alarm Display screen.
5. Look up the alarm in the alarm descriptions later in this chapter
to determine the source of the alarm and the appropriate action.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
107
Alarm Summary Table
The following table summarizes all of the high, medium, and low priority alarms and informational
messages.
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Loss of Power
High
Red flashing button;
Blank screen
Shuts down
User: Press Start/Stop button
and then press Right button
to silence alarm. If using
AC power, try plugging
device into alternate AC
power source. If loss of
•
•
power continues, switch to
DC power by connecting
a fully charged detachable
or external battery to the
device. If there is still no
power, connect patient
to alternate source of
ventilation and contact your
health care professional.
Clinician: Restore AC power.
If alarm continues, connect a
fully charged detachable or
external battery to device to
restore power. If alarm still
continues, have the device
serviced.
Ventilator
Inoperative
High
Red flashing
button; “Ventilator
Inoperative”message therapy safely.
Or, continues
Shuts down if User: Press Start/Stop button.
If display is operational,
Power Off confirmation
screen appears. Select
Right button to shut off
device and silence alarm.
Immediately remove patient
from ventilator and connect
them to alternate source of
ventilation. Contact your
health care professional for
service.
can’t provide
to operate at a
limited level.
Clinician: Restore power. If
alarm continues, have the
device serviced.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
108
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Ventilator
Service
Required
High
Red flashing button;
“Ventilator Service
Required”message
Operates
User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. The
alarm goes away until next
reminder appears. Connect
patient to alternate source of
ventilation and contact your
health care professional as
soon as possible.
• • • • •
Clinician: If alarm continues,
have the device serviced.
Check Circuit
High
Red flashing button;
“Check Circuit”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
• • • • •
Left (Reset) button to reset
alarm. Check patient circuit
to make sure it is set up as
your health care professional
specified. Make sure there
is no kinked or pinched
tubing. Make sure tubing is
properly attached. Make sure
all tubing and exhalation
devices are clean and free
of condensation (refer to
the Troubleshooting section
– tubing condensation). If
using Active PAP or Active
Flow Exhalation Device, make
sure diaphragm is inserted
correctly and is not wrinkled
or kinked. If alarm continues,
contact your health care
professional.
Clinician: Verify patient
status. Verify that breathing
circuit setup is correct.
Correct any errors. If alarm
continues, have device
serviced.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
109
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Low Circuit
Leak
High
Red flashing button;
“Low Circuit Leak”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Check for occlusions in
• • • • •
exhalation devices. Make sure
exhalation device is clean
and functioning properly. If
alarm continues, contact your
health care professional.
Clinician: Remove
obstruction in leak device.
If alarm continues, have the
device serviced.
High Expiratory High
Pressure
Red flashing button;
“High Expiratory
Pressure”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Make sure tubing is not
kinked or pinched. Check
patient’s breath rate. If alarm
continues, contact your
health care professional.
Clinician: Check patient’s
tubing to make sure it is not
kinked or pinched. If alarm
continues, have device
serviced.
Low Expiratory High
Pressure
Red flashing button;
“Low Expiratory
Pressure”message
Operates
User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. Make
sure tubing is not kinked or
pinched. If alarm continues,
contact your health care
professional.
• • • • •
Clinician: Check patient’s
tubing to make sure it is not
kinked or pinched. If alarm
continues, have device
serviced.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
110
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
High Internal
Oxygen
High
Red flashing button;
“High Internal
Oxygen”message
Continues
to operate
User: Press Alarm Indicator/
• • • • •
Audio Pause button to
when internal silence alarm. Or, press Left
oxygen
(Reset) button to reset alarm.
concentration Disconnect supplemental
reaches
5% above
ambient
levels.
oxygen supply from the
ventilator and contact your
health care professional.
Clinician: Disconnect
supplemental oxygen from
device. Check external
oxygen connection. If
problem continues, have
device serviced.
Circuit
Disconnect
High
Red flashing button;
“Circuit Disconnect”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Check breathing circuit and
reconnect it if it has become
disconnected, or fix the leak.
If alarm continues to occur,
contact your health care
professional. If device will not
exit circuit disconnect, switch
to an alternate source of
ventilation.
Clinician: Reconnect tubing
or fix leak. If alarm continues,
have device serviced.
Apnea
High
Red flashing button;
“Apnea”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
using an Active PAP or Active
Flow circuit, check proximal
pressure line to ensure it is
not pinched or has water
condensation (refer to the
Troubleshooting section
– tubing condensation). If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
111
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
High Vte
High
Red flashing button;
“High Vte”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Low Vte
High Vti
Low Vti
High
High
High
Red flashing button;
“Low Vte”message
Operates
Operates
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
• • • • •
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Red flashing button;
“High Vti”message
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Red flashing button;
“Low Vti”message
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
112
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
High
Respiratory
Rate
High
Red flashing button;
“High Respiratory
Rate”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Low Respiratory High
Rate
Red flashing button;
“Low Respiratory
Rate”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
High
Inspiratory
Pressure
Escalates
Peak Pressure symbol Operates
turns red. When
condition first occurs,
a beep will sound.
When condition
occurs for third time,
button flashes yellow
and yellow “High
Inspiratory Pressure”
message appears.
When condition
occurs for 10th
time, button flashes
red and red “High
User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. Make
sure tubing is not pinched or
blocked. If alarm continues,
contact your health care
professional.
•
(for first two
from
consecutive
Audible
o•c•cu•rrences)
(for 3rd
indicator
to Medium
priority and
then High
priority
consecutive
o•c•cu•rren•ce•)
(for 10th
consecutive
occurrence)
Clinician: Verify patient
status. If problem continues,
have device serviced.
Inspiratory Pressure”
message appears.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
113
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Low Inspiratory High
Pressure
Red flashing button;
“Low Inspiratory
Pressure”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
• • • • •
Left (Reset) button to reset
alarm. Check patient circuit
for leaks or disconnects. If
using an Active PAP or Active
Flow circuit, check proximal
lines to ensure they are
not pinched or have water
condensation (refer to the
Troubleshooting section
– tubing condensation). If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status. If problem continues,
have device serviced.
High Minute
Ventilation
High
Red flashing button;
“High Minute
Ventilation”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Low Minute
Ventilation
High
Red flashing button;
“Low Minute
Ventilation”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • • • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
114
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Low Battery
Escalates
from
Medium to
High
Medium Priority-
Yellow flashing
button. “Low
Operates
User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left button
(Reset) to reset alarm. Switch
to an alternate battery. Or,
switch to AC power source
and recharge low battery.
• • •
(Medium -
when approx.
20 minutes
remains)
Detachable Battery,“
“Low External Battery,”
or “Low Internal
Battery”message
appears in yellow,
On Status Panel, box
around battery is
yellow
Clinician: Switch to alternate
battery or AC power while
you recharge low battery.
If low battery is recharged
and alarm continues, replace
battery.
• • • • •
(High - when
approx. 10
minutes
High Priority-Red
flashing button. “Low
Detachable Battery”,
“Low External Battery,”
or “Low Internal
remains)
Battery”message
appears in red. On
Status panel, box
around battery is red
High
Temperature
Escalates
from
Medium to
High
Yellow flashing
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Make sure device is not
close to a heat source. Make
sure tubing is not under any
bedding. Check inlet filter
and clean or replace it if
needed. Make sure cooling
vents are not blocked. If the
ventilator is running off of
the internal or detachable
battery, move to a cooler
location and/or power the
device with AC power or a
lead-acid battery.
• • •
for
Medium
button and yellow
“High Temperature”
message appears. If
condition worsens,
button flashes
red and red “High
Temperature”
message appears.
• • • • •
for High
If alarm continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If alarm continues
after the measures above
have been taken, have device
serviced.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
115
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Replace
Detachable
Battery
Low or
High,
depending
on cause of
alarm
“Replace Detachable
Battery”message
appears. If battery
is nearing end of
useful life, message
appears with yellow
background and
Operates
User: Press Alarm Indicator/
• •
• •f•or L•ow•
Audio Pause button to silence
alarm. Or, press Left button
(Reset) to reset alarm. Switch
to an alternate battery or AC
power source while replacing
current detachable battery.
for High
button is solid
Clinician: Switch to an
yellow. If battery fails,
message appears with
red background and
button flashes red.
alternate battery or AC power
source while replacing the
current detachable battery.
Ventilator
Service
Recommended
Medium
Yellow flashing
button;
“Ventilator Service
Recommended”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
• • •
silence alarm. Or, press Left
(Reset) button to reset alarm.
Alarm goes away until next
reminder appears. Contact
your health care professional
to report alarm and have
device serviced.
Clinician: If alarm continues,
have device serviced.
AC Power
Disconnected
Medium
Yellow flashing
button; “AC Power
Disconnected”
message, and a box
appears around
battery in use.
Switches to
alternate
User: Press Alarm Indicator/
Audio Pause button to
• • •
power source silence alarm. Or, press
Left (Reset) button to reset
alarm. If you were using AC
power, check power cord and
reconnect it if it has become
disconnected. Make sure
device is not connected to
an overloaded AC circuit. If
you were using an external,
detachable, or internal
battery, recharge the battery.
Clinician: Check batteries
and recharge f necessary.
Restore AC power if available.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
116
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Keypad Stuck
Low
Solid yellow button;
“Keypad Stuck”
message.
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Check the keys to determine
if they are lodged in the case.
If alarm continues, place
• •
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: Check keys
to determine if they are
lodged in the case. If alarm
continues, place patient
on alternate source of
ventilation and have device
serviced.
Battery
Info
“Batt Discharge
Stopped – Temp.”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
•
Discharging
Stopped Due to
Temperature
silence alarm. Or, press
Left (Reset) button to reset
alarm. Move device to a
cooler location. Make sure
device is not close to a heat
source. Make sure cooling
vents are not blocked. If
alarm is detected for internal
battery and continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If alarm is detected
on the internal battery and
continues after the measures
above have been taken, place
patient on alternate source of
ventilation and have device
serviced.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
117
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Battery Not
Charging Due
to Temperature
Info
“Batt Not Charging –
Temp.”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
•
Left (Reset) button to reset
alarm. Make sure device is
not close to a heat source.
Make sure cooling vents are
not blocked. Move device to
a cooler location. If device
is too cold, allow it to warm
up. If alarm continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If alarm continues
after the measures above
have been taken, place
patient on alternate source of
ventilation and have device
serviced
Battery Not
Charging
Info
“Detach Battery Not
Charging”or “Internal
Battery Not Charging”
message
Operates
User: Press Alarm Indicator/
Audio Pause button to
•
silence alarm. Or, press
Left (Reset) button to reset
alarm. If condition continues
for internal battery, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If condition
continues for internal battery,
place patient on alternate
source of ventilation and
have device serviced.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
118
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Check External Info
Battery
“Check External
Battery”message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
•
Left (Reset) button to reset
alarm. Check connection to
external battery. Replace
external battery with another
external battery, if available.
If alarm continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If the alarm
continues after the measures
above have been taken, place
patient on alternate source of
ventilation and have device
serviced.
Battery
Depleted
Info
“External Battery
Depleted”or
“Detachable Battery
Depleted”message
appears
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Replace depleted battery
with another or switch to AC
power, if available.
•
•
Clinician: Replace depleted
battery with another or
switch to AC power.
External Battery Info
Disconnected
“External Batt
Switches to
alternate
power source
User: Check connection
of the external battery to
ventilator. Check the charge
available on external battery
and recharge battery if
necessary.
Disconnected”
message, and a box
appears around
battery in use.
Clinician: Check external
battery and recharge if
necessary. Restore AC power
if available.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
119
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Detachable
Battery
Disconnected
Info
“Detachable Batt
Disconnected”
message, and a box
appears around
battery in use.
Switches to
alternate
power source
User: Check connection of
the detachable battery to
ventilator. Check the charge
available on detachable
battery and recharge battery
if necessary.
•
•
Clinician: Check detachable
battery and recharge if
necessary. Restore AC power
if available.
Start On Battery Info
“Start On Battery”
message appears.
Operates
User: Check battery status
and determine how long
device can operate on
battery power. Find AC power
source as soon as possible.
Clinician: Restore AC power
when available. If alarm
continues, replace AC power
cord to determine if there is a
problem with the cord.
Card Error
Info
A “Card Error”
message appears.
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
•
Left button (Reset) to reset
alarm. Remove SD Card
and use another card, if
available. Check the write-
protect switch on the card.
If condition persists, contact
your health care professional.
Clinician: Remove SD Card
and use another card, if
available. Ensure card meets
specifications. Check write-
protect switch on card. If
condition persists, contact an
authorized representative of
Philips Respironics.
Chapter 6 Ventilator Alarms
Download from Www.Somanuals.com. All Manuals Search And Download.
120
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
7. Cleaning and Maintenance
Cleaning the Ventilator
WARNING
The ventilator’s exterior surface and the exterior of the detachable
battery pack (if using) should be cleaned before and after each
patient use, and more often if needed.
To avoid electrical shock, always
unplug the power cord from the
wall outlet before cleaning the
ventilator.
1. Unplug the device and clean the front panel and exterior of the
enclosure as needed using a clean cloth dampened with any of
the following cleaning agents:
CAUTION
Do not immerse the device or
allow any liquid to enter the
enclosure or the inlet filter.
–
–
–
–
–
Water
Soapy water or a mild detergent
Hydrogen Peroxide (3%)
Isopropyl Alcohol (91%)
CAUTION
Do not use harsh detergents,
abrasive cleaners, or brushes to
clean the ventilator system. Use
only the cleaning agents and
methods described in this manual.
10% bleach solution (10% bleach, 90% water)
2. Do not allow any liquid to drip into the ventilator case or
detachable battery pack. After cleaning, use a soft, dry cloth to
remove any residual cleaner. Use extra care when cleaning the
display. Abrasive cleaners can scratch the display.
3. Allow the device to dry completely before plugging in the power
cord.
Chapter 7 Cleaning and Maintenance
Download from Www.Somanuals.com. All Manuals Search And Download.
122
Cleaning and Replacing the Air Inlet Filter
Under normal usage, you should clean the gray foam filter at least
once every two weeks and replace it with a new filter every six
months or sooner if needed.
CAUTION
1. If the device is operating, stop the airflow by pressing the
button. Disconnect the device from the power source.
The reusable foam inlet filter is
required to protect the ventilator
from dirt and dust. Wash
periodically and replace when
damaged for proper operation.
2. Remove the filter from the enclosure by gently squeezing the
filter in the center and pulling it away from the device, as shown
below.
Removing the Air Inlet Filter
Note: Never install a wet
filter into the device. It is
3. Examine the filter for cleanliness and integrity.
recommended that you clean
the filter in the morning and
alternate using the two foam
filters provided with the system
to ensure sufficient drying time
for the cleaned filter.
4. Wash the gray foam filter in warm water with a mild detergent.
Rinse thoroughly to remove all detergent residue. Allow the filter
to air dry completely before reinstalling it. If the foam filter is torn
or damaged, replace it. Only Philips Respironics-supplied filters
should be used as replacement filters.
5. Reinstall the filter as shown below.
Reinstalling the Filter
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
123
Replacing the Air Inlet Path Foam
Follow your institution’s protocol for how often you replace the air
inlet path foam. To replace the foam, complete the steps below.
1. Using a screwdriver, remove the air inlet path cover by
unscrewing the four screws as shown below.
Removing the Air Inlet Path
Cover
2. Remove the gray foam air inlet path by hand. The removed air
inlet path is shown below.
Air Inlet Path
Compartment
Air Inlet Path Cover
Removing the Air Inlet Path
Foam
Air Inlet Path Foam
Chapter 7 Cleaning and Maintenance
Download from Www.Somanuals.com. All Manuals Search And Download.
124
3. Insert the new air inlet path foam in the slot, lining the large
circular cutout on the foam up with the circular connection at
the top of the air inlet path compartment on the device. Make
sure the air inlet path foam is securely in place.
4. Replace the air inlet path cover, using a screwdriver to secure each
of the four screws. Make sure all of the screws are firmly in place.
Cleaning the Patient Circuit
Cleaning the reusable circuit is important in the hospital and in the
home. Circuits infected with bacteria may infect the user’s lungs.
Clean the respiratory circuit on a regular basis. If you are using a
disposable circuit, dispose of and replace it on a regular basis.
Follow your institution’s protocol for cleaning the circuit. Philips
Respironics recommends that you perform the cleaning twice a
week under normal conditions and more frequently as required.
If the patient uses the device on a 24 hour a day basis, it may be
convenient to have a second breathing circuit so you can switch
circuits while one is being cleaned.
CAUTION
Exhalation devices, patient
circuits, and water traps are
shipped clean, not sterile.
Cleaning and disinfection
of these parts should follow
individual institution processes
and conform to guidelines
provided by Philips Respironics
with each accessory.
Cleaning Instructions (Reusable Circuits)
Clean the patient circuit twice a week, or follow your institution’s
protocol.
1. Disconnect the circuit from the device, and disassemble the
circuit for cleaning. Thoroughly wash your hands.
2. Using a mild detergent, such as liquid dishwashing soap, clean
all accessible surfaces of the circuit. Do not clean using alcohol.
3. Rinse the circuit with tap water, removing all remaining detergent.
4. Prepare a solution of one part white vinegar to three parts
distilled water. An average beginning quantity is 16 ounces
vinegar to 48 ounces distilled water. The actual amount will vary
according to individual needs. Regardless of the quantity the
ratio must remain 1 part:3 parts. Soak the circuit in this solution
for one hour. Rinse the circuit completely with tap water.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
125
5. Place the circuit on a clean towel to dry. Do not wipe dry. The
circuit must be completely dry before storing.
6. Reassemble the circuit when dry. Store in a plastic bag or dust-
free area.
7. Inspect components for deterioration prior to use.
Cleaning the Exhalation Devices
Disassemble the exhalation device from the patient circuit. Follow
the detailed cleaning instructions included with your exhalation
device.
Replacing the Bacteria Filter (Optional)
To remove a bacteria filter and replace it with a new one:
1. Disconnect the flexible tubing from the bacteria filter by gently
pulling straight out from the filter. The proximal tubing (if using
the Active PAP or Active Flow Exhalation Device) can remain
connected to the ventilator.
WARNING
Philips Respironics recommends
that a main line outlet bacteria
filter (Part Number 342077) be
used whenever the device is
used for invasive therapy or if
the ventilator may be used on
multiple patients.
2. Remove the used bacteria filter from the ventilator by
gently pulling straight out from the connector. Follow the
manufacturer’s labeling and your institution’s infection control
guidelines for disposal of the bacteria filter.
3. Attach a new bacteria filter. The openings on either end of the
bacteria filter are different sizes. Fit the larger opening onto the
gas outlet port on the side of the ventilator as shown.
Chapter 7 Cleaning and Maintenance
Download from Www.Somanuals.com. All Manuals Search And Download.
126
4. Connect the patient circuit. Fit the ventilator-end of the flexible
tubing onto the bacteria filter.
Installing a New Bacteria
Filter
Preventive Maintenance
Your new Philips Respironics Trilogy ventilator includes a blank PM
Service Due label affixed to the bottom of your device. The label is
blank, which allows you to document when service is due. Service
is due every 10,000 hours or 24 months, whichever comes first
depending on the usage of the device.
Service Date
Service Hours
Preventive Maintenance
Sticker
You will need to record both the date and hours on the blank label
to maximize your in-service interval. Record the service date as 24
months from the date first placed on a patient. Record the hours as
10,000 hours for initial usage.
Use the device blower hours to determine when service is required.
The device blower hours are listed in the device software in the
information menu.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
127
The following table provides information on recommended periodic maintenance for your device.
Time / Hours of Service
Recommended Maintenance
Prior to initial use on a patient
•
Charge Internal and Detachable Batteries to 100%
capacity by plugging ventilator into an AC power
source for up to 8 hours.
•
•
Perform System Checkout Procedure in Chapter 10.
Prior to long term storage after
initial use
Charge both Internal and Detachable batteries to
100% capacity prior to storage.
While in storage after initial
connection to AC Power, every 3
months
•
Recharge internal and detachable batteries to
100% capacity by plugging ventilator into an AC
power source. Batteries should recharge in 8 hours
or less.
If in use, every two weeks
If in use, every 6 months
•
•
•
Inspect and clean air inlet filter.
Replace air inlet filter.
Inspect enclosures and external connections for
damage and contact Philips Respironics service if
necessary.
•
•
Inspect power cord for damage and replace if
necessary.
Every 10,000 hours or 2 Years,
whichever comes first
Perform performance verification test per Trilogy
Service Manual by Philips Respironics trained
service technician.
Every 17,500 hours
•
•
Replace the Motor / Blower Assembly by Philips
Respironics trained service technician.
Perform full calibration and verification by Philips
Respironics trained service technician.
Refer to the Trilogy Service Manual for additional information.
Chapter 7 Cleaning and Maintenance
Download from Www.Somanuals.com. All Manuals Search And Download.
128
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
8. Troubleshooting
The ventilator is designed to respond to most issues with an appropriate alarm or information
message. These messages appear on the display screen. Chapter 6 describes many of the alarms
and what action you should take when the alarms occur.
This chapter provides additional troubleshooting information for other common issues you may
have.
Issue
Why It Happened
What To Do
Device does not turn on.
Nothing happens when you
press Start/Stop to begin
therapy. The audible indicator Note: The device requires AC
does not sound and the power to charge the internal
backlights on the buttons do battery when a new device is
not light. powered up for the first time.
The AC power cord is not
plugged in and the internal
battery is not charged.
Plug the AC power cord into the
device and a wall outlet that is
not controlled by a switch. This
powers the system and charges
the internal battery. If the device
still does not turn on, contact an
authorized service representative
or Philips Respironics to have the
device serviced. Please have the
model number and serial number
ready when you call.
Chapter 8 Troubleshooting
Download from Www.Somanuals.com. All Manuals Search And Download.
130
Issue
Why It Happened
What To Do
The batteries are not
charging.
The batteries may not charge Make sure the device is not close
if the device is too hot or too to a heat source. Ensure the
The device is plugged in, but cold or is operating at an
cooling air vents are not blocked.
the detachable (if installed) ambient temperature outside Bring the ventilator to ambient
and internal batteries are not of the specified valid range.
showing full capacity and are The device cooling fan may
room temperature. If the problem
continues, contact an authorized
service representative or Philips
Respironics to have the device
serviced. Please have the model
number and serial number ready
when you call.
not charging.
not be working properly. If
this continues for more than
15 minutes, an error will
occur.
Unexplained changes in the The device has been dropped Press the Start/Stop button and
performance of the device
The device is not working
properly or is making unusual
sounds.
or mishandled, or is in an area then select the Right button
of high EMI emissions.
to shut off the device for a few
seconds. Restart the device by
pressing the Start/Stop button. If
the problem continues, relocate
the device to an area with lower
EMI emissions (e.g., away from
electronic equipment such as
cellular phones, cordless phones,
computers, TVs, electronic
games, hair dryers, etc.). Refer to
Chapter 13. If the problem still
occurs, contact an authorized
service representative or Philips
Respironics to have the device
serviced. Please have the model
number and serial number ready
when you call.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
131
Issue
Why It Happened
What To Do
Excessive Warm Air Out of
Mask
The air filter may be dirty. Or, Clean or replace the air filter as
the device may be operating described in Chapter 7.
The air out of the mask is
much warmer than usual.
in direct sunlight or near a
heater.
The temperature of the air
may vary based on the room
temperature. Make sure the
device is properly ventilated.
Keep it away from bedding or
curtains that could block the
flow of air around the device.
Make sure it is away from direct
sunlight and heating equipment.
Ensure the cooling air vents
are not blocked. If the problem
continues, contact an authorized
service representative or Philips
Respironics to have the device
serviced. Please have the model
number and serial number ready
when you call.
Mask Discomfort
The mask feels
This could be due to
improper headgear
Make sure the patient is properly
fitted with the correct size mask.
uncomfortable to wear, there adjustment or improper mask If the problem continues, try
is significant air leakage
around the mask, or the
patient experiences other
mask-related issues.
fitting.
fitting the patient with a different
mask.
Runny Nose
This may be caused by a nasal Call the patient’s health care
The patient has a runny nose. reaction to the airflow. This
only applies to non-invasive
professional.
therapy and will not occur
during invasive therapy.
Chapter 8 Troubleshooting
Download from Www.Somanuals.com. All Manuals Search And Download.
132
Issue
Why It Happened
What To Do
Dryness
The patient has throat or
nose dryness.
The air is too dry. This only
applies to non-invasive
therapy and will not occur
during invasive therapy.
Increase the room humidity.
Consult with the patient’s health
care professional about using
a humidifier with the device. If
using a humidifier, refer to the
instructions included with the
humidifier to make sure it is
working properly.
Sinus or Ear Pain
The patient may have a sinus Tell the patient to stop using the
The patient experiences
nasal, sinus, or ear pain.
or middle ear infection. This
only applies to non-invasive
therapy and will not occur
during invasive therapy.
device and contact their health
care professional.
Patient Tubing Condensation Moisture added to the circuit
There is condensation in the from a heated humidifier may
Use an appropriate method of
water management, such as a
circuit tubing or exhalation
valve lines.
condense. This condensation is water trap when using a heated
typically referred to as“rain-out.” humidifier.
Exhalation Valve or Flow
Excessive humidity (from
Remove tube(s) with
Tube Condensation (Active the heated humidifier, the
condensation: the exhalation
valve line must be removed from
both ends of the circuit; the flow
tube lines must be removed
from the ventilator side. The
tubes are permanently attached
to the flow element, so the flow
element must be removed from
the patient circuit. If there is
water built up in the exhalation
diaphragm, disassemble the
exhalation valve and clean/dry
the assembly. Clear the flow
tubes of water droplets using
a low flow air source and re-
assemble the circuit.
Flow Circuit only)
patient, or the cleaning
process) condensed in the
tubing or exhalation valve
diaphragm. Droplets of
water affected the ability of
the ventilator to measure
proximal patient flow and
pressure and trigger on a
patient breath.
There is condensation
in the proximal flow or
exhalation valve lines, and
the condensation is causing
the system to alarm or not
properly trigger on patient
effort.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
9. Accessories
There are several accessories you can use with the ventilator.
Adding a Humidifier
Use of a humidifier may reduce nasal dryness and irritation by adding
moisture to the airflow.
If you are using a humidifier with the ventilator, refer to the
instructions provided with the humidifier for complete setup
information.
Adding Supplemental Oxygen to the Device
The delivered oxygen concentration varies with changes in flow
in the circuit. The following may have an impact on oxygen
concentration:
•
•
•
•
•
•
•
Pressure settings
Patient Tidal Volume
Peak Inspiratory Flow
I:E Ratio
Respiratory rate
Circuit leak rate
Oxygen flow rate
To add oxygen to the circuit, the oxygen supply must comply with the
local regulations for medical oxygen. The oxygen flow into the oxygen
valve cannot exceed 15 l/min and the pressure cannot exceed 50 psi.
Chapter 9 Accessories
Download from Www.Somanuals.com. All Manuals Search And Download.
134
Supplemental Oxygen Warnings
•
When administering fixed-flow supplemental oxygen,
the oxygen concentration may not be constant. The
inspired oxygen concentration will vary, depending on
the pressures, patient flows and circuit leak. Substantial
leaks may reduce the inspired oxygen concentration
to less than the expected value. Appropriate patient
monitoring should be used, as medically indicated, such
as an alarming pulse oximeter.
•
•
•
•
This device DOES NOT alarm for loss of the low flow
oxygen supply.
Oxygen supports combustion. Oxygen should not be
used while smoking or in the presence of an open flame.
Do not connect the device to an unregulated or high
pressure oxygen source.
The device may result in incorrect flow and tidal volume
measurements and improper operation of related alarms
if you add low flow oxygen directly into the patient
circuit or mask instead of directly adding it into the
oxygen inlet on the back of the ventilator.
•
•
If oxygen is used with the device, the oxygen flow must
be turned off when the device is not in use. Explanation
of the Warning: When the device is not in operation
and the oxygen flow is left on, oxygen delivered into the
tubing may accumulate within the device’s enclosure.
WARNING
The remote alarm should be
tested daily. When using a
remote alarm, make sure you
fully test the remote alarm
connector and cable by verifying
that:
The ventilator should not be operated in the presence of
flammable gasses. This could cause a fire or explosion.
Using a Remote Alarm Unit
-
Annunciated alarms on
the ventilator are also
annunciated on the remote
alarm.
You can use a Philips Respironics Remote Alarm unit with your
device. There is a dedicated adapter cable assembly for connecting
the device to the Remote Alarm unit (REF 1045291). Refer to the
instructions included with your Remote Alarm unit and adapter cable
assembly if you have questions on how to use it with device.
-
Disconnecting the remote
alarm cable from the
ventilator or from the remote
alarm results in an alarm
notification at the remote
alarm.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
135
Using a Nurse Call System
WARNING
Ensure that the Nurse Call
You can use an institutional Nurse Call system with your device.
There are several Philips Respironics cables available to connect a
nurse call system to the device. Refer to the instructions included
with your cable assembly if you have questions on how to connect it
to the device.
systems used do not exceed SELV
(Safety Extra Low Voltage) levels
as described in IEC 60601-1.
SELV levels do not exceed a
nominal value of 25VAC or
60VDC at rated supply voltage
on the transformer or converter,
between conductors in an earth-
free circuit which is isolated from
the supply mains by a safety
extra low voltage transformer
or by a device with equivalent
separation.
WARNING
The Trilogy’s Nurse Call feature
is for use only in a medically
supervised environment.
WARNING
Do not rely on the audible
indicator provided by the Nurse
Call system as the primary
indicator of the operating
state of the device or of patient
events.
Using a Secure Digital (SD) Card
An SD card is provided with the device to record device usage
information. You can insert the SD card into the SD Data Card slot on
the left side of the device. Make sure the label on the SD card faces
the back of the ventilator.
WARNING
To insert the SD card:
The Nurse Call feature should
be considered a back-up to the
Trilogy device’s primary alarm
system. Do not rely solely on the
Nurse Call feature.
1. Open the SD card door by sliding the door forward and then
pulling it out, as shown below.
Opening the SD Card Door
Chapter 9 Accessories
Download from Www.Somanuals.com. All Manuals Search And Download.
136
2. Push the SD card completely into the card slot until the card
Note: The SD card does not
need to be installed for the
device to work properly.
clicks into place.
Inserting the SD Card
Note: Use only SD cards
available from Philips
Respironics or the type listed in
the specifications section of this
manual.
To remove the SD card, select the “Safely Remove SD Card”option
from the main menu. After the “Remove SD Card”confirmation
message appears, remove the card by pushing the card in until the
card clicks to release. The card will be slightly ejected from the device.
Using the Philips Respironics DirectView
Software
You can use the Philips Respironics DirectView software to download
the prescription data from the SD card to a computer. DirectView
can be used by clinicians to receive and report stored data from
the SD card. DirectView does not perform any automatic scoring or
diagnosing of a patient’s therapy data.
Note: Remote prescription changes are not permitted.
Using the Optional In-Use Bag
An optional, in-use bag is included for use with the ventilator. The
in-use bag is required for portable use in order to properly secure
the ventilator. See the instructions included with the in-use bag for
details on how to use the bag with your device.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
137
Traveling with the System
When traveling, the carrying case is for carry-on luggage only. The
carrying case will not protect the system if it is put through checked
baggage.
For your convenience at security stations, there is a note on the
bottom of the device stating that it is medical equipment and is
suitable for airline use. It may be helpful to bring this manual along
with you to help security personnel understand the device.
If you are traveling to a country with a line voltage different than
the one you are currently using, a different power cord or an
international plug adaptor may be required to make your power cord
compatible with the power outlets of the country to which you are
traveling.
Note: The device is not
suitable for airline use with any
humidifiers or modems installed
in the unit.
Airline Travel
The device is suitable for use on airlines when the device is operating
from an internal, detachable, or external battery power source. It is
not suitable for airline use when operating from AC power or from
DC power sourced from the airplane.
Chapter 9 Accessories
Download from Www.Somanuals.com. All Manuals Search And Download.
138
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
10. System Checkout Procedures
This chapter details the test procedures that should be performed by
the clinician prior to connecting the device to the patient. Test the
Note: The actual circuit
configuration to be used on
the patient should be used to
perform the system checkout
procedure.
Active PAP, Active Flow, and Passive circuit types if you want to do a
complete checkout on the device. The tests should be performed as
described in order to verify proper operation of the device. Some of
the procedures in this chapter require you to change settings on the
device. If you are not familiar with the procedures for accomplishing
this, please refer to Chapter 4, Ventilator Setup and Chapter 5, Viewing
and Changing Settings.
Tools Required
•
•
Universal Porting Block
Active PAP Exhalation Device (with Tube Adapter
connected)
•
•
•
•
Active Flow Exhalation Device with Flow Sensor
Passive Exhalation Device
Test Lung
WARNING
If you notice any unexplained
changes in the performance
of the device, if it is making
unusual sounds, if the device
or detachable battery are
Small Flat Head Screwdriver
dropped, if water is spilled into
Visual Inspection
the enclosure, or if the enclosure
is cracked or broken, discontinue
use and contact Philips
Respironics or an authorized
service center for service.
1. Verify that the enclosure is not broken and that all applicable
screws are in place.
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
140
2. Verify that the device handle, SD Card door, and detachable
battery are secure and in good working order.
3. Verify that the rubber feet are on the bottom of the device.
Initial Setup
1. Connect the power cord to the device and then to an AC outlet.
2. Attach the test lung to the patient connection end of the desired
circuit (Active PAP, Active Flow, or Passive).
3. Follow the instructions in Chapter 5 to access the Setup Screen.
Settings and Alarms Tests
Complete the following steps to set up the settings and alarms tests.
1. Setup
A. Settings And Alarms Menu
Modify the settings in the Settings and Alarms menu to match
those shown below in Table 1. If necessary, refer to Chapter 5 for
instructions on modifying ventilator settings.
Setting
Dual Prescription
Circuit Type
Value
Off
Active PAP, Active Flow or
Passive
Therapy Mode
S/T
AVAPS (passive circuit only) Off
IPAP
20 cm H2O
EPAP
4 cm H2O
12 BPM
Table 1 - Ventilator Settings
in the Settings and Alarms
Menu
Breath Rate
Inspiratory Time
1.6 seconds
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
141
Setting
Value
Trigger Type (passive circuit Auto-Trak
type)
Flow Trigger Sensitivity
(Active PAP or Active Flow
circuit type)
6.0 l/min
Leak Compensation (Active On
Flow circuit type)
Flow Cycle Sensitivity
(Active PAP or Active Flow
circuit type)
20%
Rise Time
1
Ramp Length
All other alarms
Off
Off
B. Options Menu
Modify the settings in the Options menu to match those shown
below in Table 2.
Setting
Menu Access
Value
Full
Table 2 - Ventilator Settings
in the Options Menu
Detailed View
On
All other settings
Discretionary
C. Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The
system will begin operating using the defined ventilation
settings.
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
142
2. Verify the High Tidal Volume Alarm
This procedure verifies that the High Tidal Volume alarm is working
properly. For Passive and Active Flow circuits, this will verify the High
Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It
assumes that you have attached the test lung, verified the ventilator
settings, and turned on ventilator power as described in the Initial
Setup section.
A. Change Alarm Ventilator Setting
Modify the High Tidal Volume alarm setting to match the one
shown below in Table 3.
Setting
Value
High Vte/High Vti
50 ml
Table 3 - High Vte/Vti Alarm
Setting
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
Note: Do not use the “Reset”
button to manually reset the
alarm. Instead, use the “Modify”
button to change ventilator
settings. This applies to all tests.
•
•
The High Priority audible indicator sounds
A red light flashes on the Alarm Indicator/Audio
Pause button
•
The High Tidal Volume alarm condition appears on
the screen, highlighted in red
C. Modify Ventilator Alarm Settings
Modify the High Tidal Volume alarm setting to match the one
shown below in Table 4.
Setting
Value
Table 4 - Modify Vte/High Vti
Alarm Setting
High Vte/High Vti
500 ml
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
•
The High Priority audible indicator has stopped
sounding
•
The red light on the Alarm Indicator/Audio Pause
button has stopped flashing
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
143
E. Restore Ventilator Settings
Modify the ventilator settings and change the following value
shown in Table 5.
Setting
Value
High Vte/High Vti
Off
Table 5 - Restore Ventilator
Settings
3. Verify the Low Tidal Volume Alarm
This procedure verifies that the Low Tidal Volume alarm is working
properly. For Passive and Active Flow circuits, this will verify the Low
Vte alarm. For Active PAP circuits, this will verify the Low Vti alarm. It
assumes that you have attached the test lung, verified the ventilator
settings, and turned on ventilator power as described in the Initial
Setup section.
A. Change Alarm Ventilator Setting
Modify the Low Tidal Volume alarm setting to match the one
shown below in Table 6.
Setting
Low Vte/Low Vti
Value
500 ml
Table 6 - Low Vte/Low Vti
Alarm Setting
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
•
The High Priority audible indicator sounds
•
A red light flashes on the Alarm Indicator/Audio
Pause button
•
The Low Tidal Volume alarm condition appears on the
screen, highlighted in red
C. Modify Ventilator Alarm Settings
Modify the Low Tidal Volume alarm setting to match the one
shown below in Table 7.
Setting
Low Vte/Low Vti
Value
Table 7 - Modify Low Vte/Low
Vti Alarm Setting
50 ml
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
144
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
•
The High Priority audible indicator has stopped
sounding
•
The red light on the Alarm Indicator/Audio Pause
button has stopped flashing
E. Restore Ventilator Settings
Modify the ventilator settings and change the following value
shown in Table 8.
Setting
Low Vte/Low Vti
Value
Table 8 - Restore Ventilator
Settings
Off
4. Verify Circuit Disconnect Alarm
This procedure verifies that the Circuit Disconnect alarm is working
properly. It assumes that you have attached the test lung, verified
ventilator settings, and turned on ventilator power as described in
the Initial Setup section.
A. Change Circuit Disconnect Ventilator Setting
Modify the Circuit Disconnect ventilator setting to match the
value shown below in Table 9.
Setting
Value
Table 9 - Ventilator Settings
Circuit Disconnect
10 seconds
Note: The Low Inspiratory or
Low Expiratory Pressure Alarm
may also be detected.
B. Disconnect Test Lung
Disconnect the test lung from the circuit.
C. Verify the Alarm
Wait approximately 10 seconds and verify the following alarm
signals:
•
The High Priority Audible Indicator sounds
•
A red light flashes on the Alarm Indicator/Audio
Pause button
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
145
•
The Circuit Disconnect alarm condition appears on
the screen, highlighted in red
D. Reconnect Test Lung
Reconnect the test lung to the circuit.
E. Verify Reset
Wait at least 40 seconds and verify the following auto-reset
conditions:
•
The High Priority audible indicator has stopped
sounding
•
The red light on the Alarm Indicator/Audio Pause
button has stopped flashing
F. Restore Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 10.
Setting
Value
Table 10 - Restore Ventilator
Settings
Circuit Disconnect
Off
5. Verify the High Inspiratory Pressure Alarm
This procedure verifies that the High Inspiratory Pressure alarm
is working properly. It assumes that you have attached the test
lung, verified ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
A. Change Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 11.
Setting
Value
Mode
CV
Table 11 - Ventilator Settings
Tidal Volume
Breath Rate
500 ml
12 BPM
1.0 seconds
Ramp
Note: If this alarm is not reset
within 3 occurrences, the alarm
is elevated to High Priority, and
the High Priority Indicators
occur.
Inspiratory Time
Flow Pattern
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
146
Setting
Value
PEEP
Sigh
4 cm H2O
Off
Circuit Disconnect
Low Inspiratory Pressure
High Inspiratory Pressure
Apnea
Off
6 cm H2O
10 cm H2O
Off
All other alarms
Off
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
•
•
The Medium Priority audible indicator sounds
A yellow light flashes on the Alarm Indicator/Audio
Pause button
•
The High Inspiratory Pressure alarm condition
appears on the screen, highlighted in yellow
C. Modify Ventilator Alarm Settings
Modify the High Inspiratory Pressure setting to match the one
shown below in Table 12.
Setting
Value
Table 12 - Modify High
Inspiratory Pressure Alarm
Setting
High Inspiratory Pressure
60 cm H2O
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
• The Medium Priority audible indicator has stopped
sounding
• The yellow light on the Alarm Indicator/Audio Pause
button has stopped flashing
6. Verify the Low Inspiratory Pressure Alarm
This procedure verifies that the Low Inspiratory Pressure alarm
is working properly. It assumes that you have attached the test
lung, verified ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
147
A. Change Ventilator Settings
Modify the ventilator settings and change the following values
shown below in Table 13.
Setting
Value
Mode
CV
Tidal Volume
Breath Rate
500 ml
12 BPM
1.0 seconds
Ramp
4 cm H2O
Off
Inspiratory Time
Flow Pattern
PEEP
Sigh
Circuit Disconnect
Low Inspiratory Pressure
High Inspiratory Pressure
Apnea
Off
40 cm H2O
60 cm H2O
Off
Table 13 - Ventilator Settings
All other alarms
Off
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
•
•
The High Priority audible indicator sounds
A red light flashes on the Alarm Indicator/Audio
Pause button
•
The Low Inspiratory Pressure alarm condition
appears on the screen, highlighted in red
C. Modify Ventilator Alarm Settings
Modify the Low Inspiratory Pressure setting to match the one
shown below in Table 14.
Setting
Value
Table 14 - Modify Low
Inspiratory Pressure Alarm
Setting
Low Inspiratory Pressure
6 cm H2O
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
148
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
•
The High Priority audible indicator has stopped
sounding
•
The red light on the Alarm Indicator/Audio Pause
button has stopped flashing
Battery Function Verification
Make sure the batteries are functioning properly and fully charged
before patient use.
1. Verify the Detachable and Internal (Lithium-Ion)
Batteries Function
A. Connect AC Power to the device and verify that the green AC
LED on the front panel is lit.
B. Verify that the detachable battery is properly installed.
C. Turn the device on and verify that both the detachable and
internal battery symbols appear on the display. Verify that if
either battery is less than fully charged, the charge symbol will
display on the respective battery.
D. Disconnect the AC Power source from the device.
•
•
•
Verify that the AC Power Disconnected alarm
message appears on the display and the green AC
LED is not lit. Press Reset.
Verify that the detachable battery symbol shows the
level of charge noted in the previous step and that
the device continues to operate.
Verify that the detachable battery symbol has a black
box around it to indicate that it is in use.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
149
E. Disconnect the detachable battery pack from the device.
•
Verify that the Detach Batt Disconnected alarm
message appears on the display. Press Reset.
•
Verify that the internal battery symbol shows the
same level of charge as noted in Step C and the
device continues to operate.
•
Verify that the internal battery symbol has a black
box around it to indicate that it is in use.
F. Reconnect the Detachable Battery and AC Power source.
2. Verify the External Battery Function (If Available)
A. Connect AC Power to the device and verify that the green AC
LED is lit.
B. Connect the external battery cable to the external battery and to
the ventilator.
C. Verify that the external battery symbol is shown on the display
and some level of charge is present.
D. Disconnect the AC Power source from the device.
•
•
•
Verify that the AC Power Disconnected alarm
message appears on the display and the green AC
LED is not lit. Press Reset.
Verify that the external battery symbol shows the
level of charge as noted in the previous step and the
device continues to operate.
Verify that the external battery symbol has a black
box around it to indicate that it is in use.
E. Reconnect the AC Power source.
Chapter 10 System Checkout Procedures
Download from Www.Somanuals.com. All Manuals Search And Download.
150
Alarm and Event Log Clean-Up
1. In the Setup Menu, select Alarm Log.
A. Press Clear to clear the log file.
B. Press Yes to confirm.
C. Press Finish to complete.
2. In the Setup Menu, select Event Log.
A. Press Clear to clear the log file.
B. Press Yes to confirm.
C. Press Finish to complete.
Results
All portions of this checkout procedure should be completed prior
to connection to the patient. If any of the tests fail to complete as
indicated, if possible, correct the error, clear the alarm and resume
testing. If correction of the failed portion is not possible, return the
device to Philips Respironics or an authorized service center for
service and repair.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
11. Technical Specifications
Environmental
Operating
Storage
Temperature
Relative Humidity
Atmospheric Pressure
41° F to 104° F (5° C to 40° C)
15 to 95% (non-condensing)
110 kPa to 60 kPa
-4° F to 140° F (-20° C to 60° C)
15 to 95% (non-condensing)
N/A
The operating range for the charging of the internal and detachable batteries is 10° C to 30° C. The internal and detachable
batteries will power the ventilator for the full operating range of 5° C to 40° C.
Accuracies stated in this manual are based on specific environmental conditions. For stated accuracy, the environmental
conditions are: Temperature: 20-30° C; Humidity: 50% relative; Altitude: nominally 380 meters.
Physical
Dimensions:
Weight:
6.6”L x 11.2”W x 9.3”H (16.68 cm L x 28.45 cm W x 23.52 cm H)
Approximately 11 lbs (5 kg) with the detachable battery installed
Standards Compliance
This device is designed to conform to the following standards:
•
•
IEC 60601-1: Medical electrical equipment - Part 1: General requirements for safety
IEC 60601-1-2: General requirements for safety - Collateral standard: Electromagnetic
compatibility - Requirements and tests
•
•
ISO 10651-2-Lung ventilators for medical use -- Particular requirements for basic safety and
essential performance -- Part 2: Home care ventilators for ventilator-dependent patients
RTCA-D0160F section 21, category M; Emission of Radio Frequency Energy
Chapter 11 Technical Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
152
Electrical
AC Voltage Source:
Detachable Battery:
100 to 240 VAC, 50/60 Hz, 2.1 A
Voltage: 14.4 VDC
Capacity: 4176 mAh
Chemistry type: Lithium-Ion
Internal Battery:
Voltage: 14.4 VDC
Capacity: 4176 mAh
Chemistry type: Lithium-Ion
Type of Protection Against Electric Shock:
Degree of Protection Against Electric Shock:
Degree of Protection against Ingress of Water:
Mode of Operation:
Class II/Internally Powered Equipment
Type BF Applied Part
Device: Drip Proof, IPX1
Continuous
Fuses:
There are no user-replaceable fuses.
Pressure
Output:
4 to 50 cm H2O
SD Card and SD Card Reader
Use only SD cards and SD card readers available from Philips Respironics or the following:
SanDisk® Standard SD Card - 1.0 GB - REF SDSDJ-1024
SanDisk ® Card Reader/Writer - SanDisk ImageMate - REF SDDR-99-A15
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
153
Control Accuracy
Parameter
IPAP
Range
4 to 501 cm H2O2
Accuracy
Greater of 2 cm H2O or 8% of setting
EPAP
0 to 25 cm H O for Active Circuits
4 to 25 cm H22O for Passive Circuits
Greater of 2 cm H2O or 8% of setting
CPAP
PEEP
4 to 20 cm H2O
Greater of 2 cm H2O or 8% of setting
Greater of 2 cm H2O or 8% of setting4
0 to 25 cm H O for Active Circuits
4 to 25 cm H22O for Passive Circuits
Pressure
4 to 50 cm H2O
0 to 30 cm H2O3
50 to 2000 ml5
Greater of 2 cm H2O or 8% of setting
Greater of 2 cm H2O or 8% of setting4
Pressure Support
Tidal Volume
Greater of 15ml or 10% of setting (Active
Circuits)
Greater of 15ml or 15% of setting (Passive
Circuits)
Breath Rate
0 to 60 BPM for AC mode
1 to 60 BPM for all other modes
Greater of 1 BPM or 10% of the setting
Timed Inspiration
Rise Time
0.3 to 5.0 seconds
1 to 66
.1 second
2 cm H O (the device will increase to a
pressure2of .67* (IPAP - EPAP) 2 cm H2O @ the
set rise time multiplied by 100 ms for pressure
supports less than or equal to 25.)
Ramp Start Pressure
0 to 25 cm H O for Active Circuits
4 to 25 cm H2O for Passive Circuits
4 to 19 cm H22O in CPAP mode
8% of setting + 2% Full Scale
Ramp Length
Flex
Off, 5 to 45 minutes
Off, 1 to 37
2 minutes
N/A
Flow Trigger Sensitivity
Flow Cycle
1 to 9 l/min
N/A
10 to 90%
N/A
Apnea Rate
4 to 60 BPM
Greater of 1 BPM or 10% of setting
Specifications listed are based on using a standard circuit consisting of 1.8 meter tubing- REF 622038, one of the following exhalation
devices: Whisper Swivel II - REF 332113; Active PAP Exhalation Device - REF 1053716; Active Flow Exhalation Device - REF 1049503 or
Active Exhalation Device – REF 1065658; UPB Active PAP Exhalation Device - REF 1073862 and
Trilogy Proximal Flow Sensor- REF 1050408 (if required).
1Limited to 25 cm H2O when using the Bi-Flex feature in S mode.
2Pressure units may be cm H2O, hPa, or mBar depending on device setup.
3The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 30 cm H2O.
4Pressure Support and PEEP not to exceed 50 cm H2O.
5All flows and volumes are measured at BTPS conditions.
6The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).
7Flex is not available when AVAPS is active. Flex is not available with Active Circuits.
Chapter 11 Technical Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
154
Measured Patient Parameters
Parameter
Range
Accuracy
Vte/Vti
0 to 2000 ml
0 to 99 l/min
Greater of 15ml or 15% of reading
Minute Ventilation
Calculation based on measured Vte or Vti
and Respiratory Rate
Estimated Leak Rate
Respiratory Rate
0 to 200 l/min
0 to 80 BPM
0 to 200 l/min
0 to 99 cm H2O
0 to 99 cm H2O
0 to 100%
N/A
Greater of 1 BPM or 10% of reading
3 l/min plus 15% of reading
Greater than 2 cm H2O or 10% of reading
Greater than 2 cm H2O or 10% of reading
N/A
Peak Inspiratory Flow
Peak Inspiratory Pressure
Mean Airway Pressure
% Patient Triggered Breaths
I:E Ratio
9.9- 1: 1-9.9
Calculation based on Inspiratory Time and
Expiratory Time
All flows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set
Inspiratory Resistance (cm H2O)
Expiratory Resistance (cm H2O)
Point (l/min)
Active Active Circuit with Passive
Active Active Circuit with Passive
Circuit
Proximal Flow
Sensor
Circuit
Circuit
Proximal Flow
Sensor
Circuit
30
60
< 3.0
< 9.0
< 3.5
< 2.0
< 5.0
< 2.0
< 3.0
< 2.0
< 4.5
< 1.5
< 4.0
< 10.5
Disposal
Dispose of this device and batteries in accordance with local regulations.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
12. Glossary
The following terms and acronyms appear throughout this manual.
Term/Acronym Definition
AC Power
Alternating Current power
AC
Assist Control therapy mode; AC mode delivers Assisted and
Mandatory breaths with a user-defined inspired tidal volume. This
is a Volume Control mode.
Apnea
A condition marked by the temporary cessation of spontaneous
breathing.
Assisted Breath
Breath type in which the breath is patient-triggered and the
ventilator controls how the air is delivered based on defined
settings. An assisted breath is not cycled until the inspiratory
time setting has been reached.
AVAPS
Average Volume Assured Pressure Support feature (available in S,
S/T, T, and PC therapy modes).
Blower Hours
The total number of hours that the blower has been on over the
life of the device. This value helps determine when the ventilator
needs to be serviced. You cannot reset this value. It can only be
reset by an authorized service center.
BPM
Breaths Per Minute
Chapter 12 Glossary
Download from Www.Somanuals.com. All Manuals Search And Download.
156
Term/Acronym
BTPS
Definition
Body Temperature and Pressure Saturated; A standardization
for lung volumes and flows to barometric pressure at sea level,
body temperature, and saturated with water vapor reflecting the
condition of air in the lung.
CPAP
CV
Continuous Positive Airway Pressure
Control Ventilation therapy mode; CV mode delivers Mandatory
breaths with a user-defined inspired tidal volume. This is a Volume
Control mode.
DC
Direct Current
EPAP
ESD
Expiratory Positive Airway Pressure
Electrostatic Discharge
Flex
The device provides the Flex comfort feature. The C-Flex™ feature
provides increased pressure relief during the expiratory phase
of breathing. The Bi-Flex® feature provides a level of pressure
relief taking place at the end of inhalation and at the start of
exhalation.
I:E Ratio
IPAP
The ratio of inspiratory time to expiratory time.
Inspiratory Positive Airway Pressure
Light Emitting Diode
LED
l/min
Liters Per Minute
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
157
Term/Acronym
Definition
Mandatory Breath
A Mandatory Breath is completely controlled by the ventilator.
Breaths are initiated by the ventilator according to the breath
Rate (BPM) setting. Breaths are cycled according to the Inspiratory
time setting.
MAP
Mean Airway Pressure. This is the average airway pressure over
a full breath cycle. The MAP value displayed on the screen
represents the average MAP value over six breaths and is updated
at the end of each exhalation.
Minute Ventilation
The volume of gas that moves in and out of the lungs in one
minute. It is calculated by multiplying the tidal volume by the
respiratory rate. The Minute Ventilation value displayed on the
screen represents the average Minute Ventilation over six breaths
and is updated at the end of each exhalation.
Operational Hours
This is the total number of hours that the blower has been on
since this value was last reset. You can reset this value each time
you give the device to a new patient to help track their device
usage.
OSA
PC
Obstructive Sleep Apnea
Pressure Control therapy mode; PC mode delivers Assisted and
Mandatory breaths with a user-defined pressure. This is a Pressure
Control mode.
PC SIMV
Pressure Control Synchronous Intermittent Mandatory Ventilation
therapy mode; PC-SIMV mode delivers Spontaneous, Assisted,
and Mandatory breaths. This is a Pressure Control mode.
Peak Flow
Maximum flow rate (in liters per minute) reached during a breath.
Highest pressure reached during inspiration.
Peak Inspiratory
Pressure (PIP)
Chapter 12 Glossary
Download from Www.Somanuals.com. All Manuals Search And Download.
158
Term/Acronym
PEEP
Definition
Positive End Expiratory Pressure
Pressure Control
Ventilation (PCV)
Ventilation in which breaths are controlled by operator-defined
Pressure, Inspiratory Time, and Rise Time.
PS
Pressure Support
Ramp
A feature that may increase patient comfort when therapy is
started. The ramp feature reduces pressure and then gradually
increases the pressure to the prescription setting so patients can
fall asleep more comfortably.
Rise Time
The Rise Time is the time it takes the ventilator to change from
expiration to inspiration.
RR
Respiratory Rate (the number of breaths per minute).
SD Card
Secure Digital card; This card records usage and therapy data
from the device.
Sigh
A breath that is delivered every 100 mandatory or assisted
breaths at 150% of the normal volume.
SIMV
Synchronous Intermittent Mandatory Ventilation therapy mode;
SIMV mode delivers Spontaneous, Assisted, and Mandatory
breaths with a user-defined inspired tidal volume. This is a
Volume Control mode.
Sleep State
A low-power state entered when the ventilator is running on the
internal or detachable battery and the blower has been off for
five minutes. This state preserves battery power. The device is
restored by plugging the unit into AC power, inserting an SD card,
or by pressing the Start/Stop button.
Spontaneous Breath
Breath type in which the breath is patient-triggered.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
159
Term/Acronym
Definition
Spontaneous (S)
Mode
Therapy mode in which all breaths are spontaneous. The
ventilator supports breathing with user-defined Pressure Support
(PS) and Rise Time values.
Spontaneous/Timed
(S/T) Mode
Therapy mode that is similar to S mode, except that it can also
deliver a mandatory breath if the patient does not spontaneously
breathe within a set time.
Tidal Volume
The amount of air passing in and out of the lungs for each breath.
Timed (T) Mode
Timed pressure support therapy mode. In Timed mode, all
breaths delivered are mandatory.
Volume Control
Ventilation (VCV)
Ventilation in which breaths are controlled by an operator-
defined volume, flow pattern, breath rate, and inspiratory time.
Vte
Vti
Exhaled Tidal Volume
Inhaled Tidal Volume
Chapter 12 Glossary
Download from Www.Somanuals.com. All Manuals Search And Download.
160
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
13. EMC Information
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF emissions
Group 1
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
CISPR 11
RF emissions
Class B
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies building used for domestic
purpose.
CISPR 11
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions
Complies
IEC 61000-3-3
Chapter 13 EMC Information
Download from Www.Somanuals.com. All Manuals Search And Download.
162
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment -
Guidance
Electrostatic
6 kV contact
8 kV contact
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical home or hospital
environment.
Discharge (ESD)
8 kV air
15 kV air
IEC 61000-4-2
Electrical fast
Transient/burst
2 kV for power supply
2 kV for supply mains
lines
IEC 61000-4-4
1 kV for input-output
lines
1 kV for input/output
lines
1 kV differential mode
Surge
1 kV differential mode
Mains power quality should be
that of a typical home or hospital
IEC 61000-4-5
2 kV common mode
<5% UT
2 kV for common mode environment.
Voltage dips, short
<5% UT
Mains power quality should be
interruptions and voltage (>95% dip in UT) for
(>95% dip in UT) for
0.5 cycle
that of a typical home or hospital
environment. If the user of the
variations on power
supply input lines
0.5 cycle
40% UT
40% UT
device requires continued operation
during power mains interruptions,
it is recommended that the device
(60% dip in UT) for
5 cycles
(60% dip in UT) for 5
cycles
IEC 61000-4-11
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
70% UT (30% dip in UT) for be powered from an uninterruptible
25 cycles
power supply or a battery.
<5% UT (>95% dip in UT)
for 5 sec
Power frequency (50/60
Hz) magnetic field
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical home or hospital
environment.
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
163
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test IEC 60601 Test Compliance
Level Level
Electromagnetic Environment -
Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
3 V
d = 1.2
P
P
P
150 kHz to 80 MHz
outside ISM bandsa
10Vrms
10 V
d = 1.2
150 kHz to 80 MHz
in ISM bandsa
Radiated RF
10 V/m
10 V/m
d = 1.2
80 MHz to 800 MHz
800 MHz to 2.5 GHz
IEC 61000-4-3
80 MHz to 2.5 GHz
26 MHz to 2.5 GHz
d = 2.3
P
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site surveya, should be less than
the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the device is used exceeds the applicable
RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Chapter 13 EMC Information
Download from Www.Somanuals.com. All Manuals Search And Download.
164
Recommended Separation Distances between Portable and
Mobile RF Communications Equipment and This Device
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of this device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and this device as recommended below, according to
the maximum output power of the communications equipment.
Rated Maximum
Power Output of
Transmitter
Separation Distance According to Frequency of Transmitter
(meters)
150 kHz to 80 MHz 150 kHz to 80 MHz 80 MHz to 800 MHz
800 MHz to 2.5 GHz
(Watts)
outside ISM Bands
in ISM Bands
d = 1.2
P
d = 2.3
P
d = 1.2
P
d = 1.2
P
0.01
0.1
0.12
0.12
0.12
0.38
0.23
0.73
0.38
0.38
1
1.2
3.8
12
1.2
3.8
12
1.2
3.8
12
2.3
7.3
23
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power of the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the
ISM frequency bands between 150 kHz and 80 MHz and in the frequency range of 80 MHz and 2.5 GHz to decrease the
likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into
patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
Index
External Battery Disconnected 48
System Alarms
Ventilator Service Recommended 46
A
AC Power 51
AC Power Inlet 17
Air Filter
B
Installation 50
Air Inlet
Air Inlet Filter
Cleaning 122
Replacing 123
Air Inlet Path
Replacing 123
Alarm and Event Log Clean-Up 148
Alarm Log 89
Bacteria Filter
Replacing 125
Battery Function Verification 146
Breathing Circuit
Connection 57
Breathing Circuit Connection 84
Breath Rate 82
Breath Types
Assist 20
Alarms 102
C
Alarm Indicators 104
Alarm Messages 104
Alarm Reset 105
Alarm Volume Control 106
Patient Alarms
Cautions 8, 15, 133
Circuit Type Setting 74
Contraindications 11
Control Ventilation Mode 27
High Exhaled Tidal Volume 44
High Minute Ventilation 45
High Peak Inspiratory Pressure 46
Low Exhaled Tidal Volume 44
Low Minute Ventilation 45
Low Peak Inspiratory Pressure 44
Low Respiratory Rate 44
Power Alarms 107
D
Date Format 88
DC Power
Detachable Battery Pack 52
External Battery 52
Detailed View 86
Index
Download from Www.Somanuals.com. All Manuals Search And Download.
166
DirectView Software 134
I
Inspiratory Time 82
In-Use Bag 134
IP Address Mode Setting 88
IPAP 81
IPAP Max Pressure 81
IPAP Min Pressure 81
E
EMC Information 159
EPAP 81
Ethernet Connector 18
Event Log 90
Exhalation Devices
Cleaning 125
K
Exhalation Porting Block 18
Keypad Backlight 87
Keypad Lock Feature 63, 87
F
L
Flex 97
Flex Comfort Feature 30, 80
Bi-Flex 30
Language Setting 86
LCD Brightness Setting 87
Low Respiratory Rate 78
Low Tidal Volume 76
C-Flex 30
Flow Cycle Sensitivity 79
Flow Pattern 84
Ramp 33
Square 33
M
Main Menu Screen 94
Menu Access i, 86
Full Menu i
Limited Menu i
Monitor Screen 64
Indicators 66
My Settings Menu 96
Flex 97
Ramp Start Pressure 96
Rise Time 96
Flow Trigger Sensitivity 79
Front Panel
AC Power LED 15
Audio Pause Button 16
Keypad Backlight LEDs 16
Left/Right Buttons 16
Red Alarm LED 15
Start/Stop Button 15
Up/Down Button 16
H
O
High Minute Ventilation 77
High Respiratory Rate 77
High Tidal Volume Alarm 76
Humidifier 131
Operational Hours 89
Options Menu 86
Oxygen Valve Connector 18
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
167
Environmental 149
Pressure 150
Standards Compliance 149
P
Passive Exhalation Device 57
Patient Circuit
Cleaning 124
T
PC-SIMV Mode 25, 83
Tidal Volume 81
Timed Mode 82
Time Format 88
Trigger Type
Auto-Trak 78
Flow Trigger 78
Troubleshooting 129
Positive End Expiratory Pressure (PEEP) 84
Pressure Control Mode 24
Pressure Control Ventilation 85
Pressure Setting 83
Pressure Support 83
Pressure Units 86
R
V
Ramp Feature 31
Volume Control Ventilation 19
Ramp Length 79
Ramp Start Pressure 80, 96
Remote Alarm 62, 104
Remote Alarm Connector 18
Rise Time 32, 82, 96
W
Warnings 3
Warranty 167
Water Trap 60
S
Screen Saver 87
Secondary Prescription Setting 95
Secure Digital (SD) Card 17, 133
Sensitivity 36
Digital Auto-Trak Sensitivity
Auto-Trak Sensitivity 37
Leak Tolerance 36
Flow Trigger 40
Serial Connector
Settings and Alarms Tests 138
Setup Screen i
Sigh Feature 84
SIMV Mode 28, 85
Specifications
Control Accuracy 150
Electrical 150
Index
Download from Www.Somanuals.com. All Manuals Search And Download.
168
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Trilogy200
clinical manual
Limited Warranty
Respironics, Inc. warrants that the Trilogy200 system shall be free from defects of workmanship and
materials and will perform in accordance with the product specifications for a period of two (2) years
from the date of sale by Respironics, Inc. to the dealer. If the product fails to perform in accordance with
the product specifications, Respironics, Inc.will repair or replace – at its option – the defective material or
part. Respironics, Inc. will pay customary freight charges from Respironics, Inc. to the dealer location only.
This warranty does not cover damage caused by accident, misuse, abuse, alteration, and other defects
not related to material or workmanship.
Respironics, Inc. disclaims all liability for economic loss, loss of profits,overhead, or consequential
damages which may be claimed to arise from any sale or use of this product. Some states do not allow
the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion
may not apply to you.
Accessories and replacement parts, including, but not limited to, circuits, tubing, leak devices, exhaust
valves, filters and fuses, are not covered under this warranty. However, Respironics warrants that the
Product’s internal and detachable battery (where supplied) will be free from defects in material and
workmanship, under normal and proper use and when correctly maintained in accordance with
applicable instructions, for a period of 90 days from the date of shipment by Respironics to the original
purchaser. This warranty does not apply to batteries that are dropped, misused, altered or otherwise
damaged after they are shipped.
This warranty is given in lieu of all other express warranties. In addition, any implied warranties –
including any warranty of merchantability or fitness for the particular purpose – are limited to two years.
Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may
not apply to you. This warranty gives you specific legal rights, and you may also have other rights which
vary from state to state.
Limited Warranty
Download from Www.Somanuals.com. All Manuals Search And Download.
170
To exercise your rights under this warranty, contact your local authorized
Respironics, Inc. dealer or contact Respironics, Inc. at:
1001 Murry Ridge Lane
Murrysville, Pennsylvania 15668-8550
1-724-387-4000
Trilogy200 clinical manual
Download from Www.Somanuals.com. All Manuals Search And Download.
|