Introduction  
Thank you for purchasing the OMRON Comp Air Elite.  
This product was developed in conjunction with respiratory therapists for the  
successful treatment of asthma, chronic bronchitis, allergies and other respiratory  
disorders. The compressor forces air to the nebuliser. When the air enters the  
nebuliser, it converts the prescribed medication into an aerosol of microscopic  
droplets that can easily be inhaled.  
The Comp Air Elite is a medical device. Operate this device only as instructed by  
your doctor and/or respiratory therapist.  
Intended use  
Medical Purpose  
Intended User  
This product is intended to be used for inhaling medication  
for respiratory disorders.  
• Legally certified medical experts, such as doctor, nurse and  
therapist, or healthcare personnel or patient under the  
guidance of qualified medical experts.  
• The user should also be capable of understanding general  
machine operation and the content of instruction manual.  
Intended Patients  
Environment  
This product should not be used by patients, who are  
unconscious or are not breathing spontaneously.  
This product is intended for use in a medical facility, such as  
Hospital, clinic and doctor’s office, in a room of general  
household, and in open-air environment with a roof.  
Durable period  
Durable periods are as follows, provided the product is used  
to nebulise saline 3 times a day for 10 minutes each time at  
room temperature (23°C).  
Durable period may vary depending on usage environment.  
EN  
Main unit  
5 years  
Nebuliser Kit  
(Cap, Inhalation Air Inlet, Baffle, Medication Cup) 1 year  
Mouthpiece  
Air Tube  
Air Filter  
1 year  
1 year  
60 days  
1 year  
Child Mask  
Adapter for Child Mask 1 year  
Precautions for use Warnings and cautions described in the instruction manual  
should be observed.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Important safety instructions  
Read all the information in the instruction manual and any other literature included  
in the box before using the device.  
When using an electrical product, especially when children are present, basic  
safety precautions should always be followed including the following:  
Warning:  
Indicates a potentially hazardous situation which, if not avoided, could result in  
serious injury.  
(Usage)  
• For type, dose, and regime of medication follow the instructions of your doctor or  
respiratory therapist.  
• Do not use only water in the nebuliser, for nebulising purposes.  
• Clean and disinfect the nebuliser kit, mouthpiece and optional nosepiece or  
masks before using them for the first time after purchase, if the device has not  
been used for a long period of time, or if more than one person uses the same  
device.  
• Be sure to wash the parts after use, and ensure that they are thoroughly  
disinfected and dried, and stored in a clean location.  
• Keep the device out of the reach of unsupervised infants and children. The  
device may contain small pieces that can be swallowed.  
• Store the device and accessories in a clean location.  
• Do not store the air tube while there is moisture or medication remaining inside it.  
• Always rinse the nebulising parts with clean hot tap water after disinfecting them.  
Never leave the cleaning solution in the medication tank or mouthpiece.  
• Do not use or store the device where it may be exposed to noxious fumes or  
volatile substances.  
• Do not use the device where it may be exposed to flammable gas.  
• Make sure that the nebuliser kit is clean before use.  
• Always dispose of any remaining medication after use, and use fresh medication  
each time.  
• Do not use in anaesthetic or ventilator breathing circuits.  
(Risk of electrical shock)  
• When using the optional rechargeable battery, always charge the battery fully  
using the battery charger specially designed for this device. If the battery is  
frequently used without being fully charged, the capacity of the battery will be  
reduced and the device may stop during use. In this case have an alternative  
power source, such as the AC adapter ready.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Important safety instructions  
• Do not use the main unit, AC adapter, optional rechargeable battery, or optional  
battery charger while they are wet, or connect them to a power outlet or other  
devices with wet hands.  
• The compressor, AC adapter, and optional rechargeable battery and charger are  
not waterproof. Do not spill water, or other liquids, on these parts. If liquid does  
spill on these parts, immediately unplug the power cord and wipe off the liquid  
with gauze or other absorbent material.  
• Do not immerse the main unit in water or other liquid.  
• Do not use or store the device in humid locations, such as a bathroom.  
• Do not short circuit the battery terminals on the rechargeable battery. Make sure  
battery terminals do not come in contact with keys, coins or other metal objects.  
• When using the carrying bag to carry the device and optional rechargeable  
battery, do not place hair pins or other metal objects in the bag.  
• Use only original OMRON AC adapter for AC power outlet. Use of unsupported  
AC adapter may damage the device.  
• Do not operate the unit with a damaged cord or plug.  
• Keep the power cord away from heated surfaces.  
Read and follow the “Important information regarding Electro Magnetic  
Compatibility (EMC)” in the Technical Data Section.  
Caution:  
Indicates a potentially hazardous situation which if not avoided, may result in minor  
or moderate injury, or physical damage.  
(Usage)  
• Provide close supervision when this device is used by, on, or near children or  
invalids.  
• Make sure that the parts are attached correctly.  
• Make sure that the vaporiser head is correctly installed before use.  
• Make sure that the air filter is correctly attached.  
EN  
• Make sure that the air filter is clean. If the air filter has changed colour, or has  
been used on average for more than 60 days, replace it with a new one.  
• Do not spill liquid or medication on the compressor.  
• Do not tilt the nebuliser kit so that it is at an angle of greater than 45°, or shake it  
while in use.  
• Do not use or store the device while the air tube is creased.  
• Use only original nebulising parts, air tube, air filter and filter cover.  
• Do not add more than 7 ml of medication to the medication tank.  
• Do not carry or leave the nebuliser kit while the medication tank contains  
medication.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Important safety instructions  
• Do not leave the unit unattended with infants or persons who cannot express  
their consent.  
• Do not subject the device or the parts to any strong shocks such as dropping the  
device on the floor.  
• Do not distort the vaporiser head, or jab the nozzle of the medication tank with a  
pin etc.  
• Do not cover the compressor, optional rechargeable battery and charger with a  
blanket or towel etc., during use.  
• Do not insert fingers or objects inside the compressor.  
• Do not disassemble or attempt to repair the compressor, AC adapter, optional  
rechargeable battery or charger.  
• Do not leave the device or its parts where is will be exposed to extreme  
temperatures or changes in humidity, such as in a vehicle during the summer, or  
where it will be exposed to direct sunlight.  
• Do not block the air filter cover.  
• When sterilising parts by boiling, make sure that the container does not boil dry.  
• Never place, or attempt to dry the compressor, AC adapter or nebulising parts in  
a microwave oven.  
• Do not use the device while sleeping or if drowsy.  
To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back  
of the nose.  
• Do not block the slit between the cap and the inhalation air inlet.  
• Do not use a microwave oven or hair dryer to dry the nebulising parts.  
• Approved for human use only.  
(Risk of electrical shock)  
• Always unplug the AC adapter from the power outlet after use and before  
cleaning.  
• Plug the device into the appropriate voltage outlet. Do not overload power outlets  
or use extension cords.  
• Always read the instructions included with optional parts, such as the  
rechargeable battery and charger, before using them.  
• Always remove the optional battery and AC adapter from the device after use  
and before cleaning.  
• Do not misuse the cord of the AC adapter.  
• Do not wind the cord of the AC adapter around the main unit or AC adapter.  
• Changes or modifications not approved by OMRON Healthcare will void the user  
warranty.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
Important safety instructions  
General Safety Precautions:  
• Inspect the device and parts before using them each time, and check that there  
are no problems. In particular, be sure to check the following:  
- That the vaporiser head, nozzle or air tube are not damaged.  
- That the nozzle is not blocked.  
- That the compressor operates normally.  
• When using this device, there will be some noise and vibration caused by the  
pump in the compressor. There will also be some noise caused by the emission  
of compressed air from the nebuliser kit. This is normal and does not indicate a  
malfunction, or reduce the working life of the device.  
• Limit use to 20 minutes at a time, and allow a 40 minute interval before using the  
device again.  
• Operate the device only as intended. Do not use the device for any other  
purpose.  
• Do not use the device at temperatures greater than +40°C.  
• Do not spill liquid or medication on the battery connector, optional rechargeable  
battery or charger connectors, or allow them to become dusty. Immediately wipe  
off the liquid or dust with gauze or other absorbent material.  
• Make sure that the air tube is securely attached to the compressor (main unit)  
and nebuliser kit, and does not come loose. Twist the airtube plug slightly when  
inserting it into the connectors to avoid the tube disconnecting during use.  
• The AC adapter is provided so that the unit can be completely isolated from the  
power source. To completely isolate the unit from the power source unplug the  
AC adapter from the power source.  
Read and follow the “Correct Disposal of This Product” in the Technical Data  
Section when disposing of the device and any used accessories or optional parts.  
Save these instructions for future reference.  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
1. Know your unit  
D
A
C
A1  
A2  
A3  
D1  
D2  
D3  
D4  
5
3
E
A4  
A5  
A6  
E1  
B1  
E2  
E3  
B2  
B
E4  
Rear View  
Bottom View  
E5  
E7  
E6  
5
3
A7  
A8  
A. Compressor (Main Unit)  
A1. Power Indicator  
A2. Power Switch  
D. Inhalation Attachments  
D1. Mouthpiece  
D2. Adult Mask (PVC)  
D3. Child Mask (PVC)  
D4. Nosepiece  
A3. Charge Indicator  
A4. Air Connector  
A5. Filter Cover (* Air filter inside.)  
A6. Nebuliser Kit Holder  
A7. AC Adapter Jack  
A8. Battery Connector  
B. Air Tube (PVC, 207cm)  
B1. Air Plug  
E. Nebuliser Kit  
E1. Cap  
E2. Inhalation Air Inlet  
E3. Inhalation Top  
E4. Vaporiser Head  
E5. Medication Tank  
E6. Nozzle  
B2. Air Plug  
C. Nebulising Parts  
E7. Air Tube Connector  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
1. Know your unit  
Accessories  
F
G2  
H
G1  
G
I
J
F. Spare Air Filters x 5  
G. AC Adapter  
I. Instruction Manual  
J. Warranty Card  
G1. Power Connector  
G2. Power Cord  
H. Carrying Bag  
Material of attachment  
Nebuliser Kit, Mouthpiece, Nosepiece, Adapter for  
Child Mask (SEBS):  
Polypropylene  
Adult Mask (PVC), Child Mask (PVC):  
Adult Mask (SEBS), Child Mask (SEBS):  
Air Tube:  
PVC  
EN  
SEBS  
PVC and Polypropylene  
Air Filter:  
Polyester  
ABS  
Air Filter Cover:  
• Also applicable Section 10. Optional parts and accessories.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9
2. Guide to using the Comp Air Elite  
Preparing the Nebuliser for Use (Section 3)  
Note: Clean and disinfect the nebuliser kit and applicable inhalation  
attachments before using the device for the first time, or if it has not  
been used for more than one week.  
Taking Medication (Section 4)  
Notes:  
• Do not tilt the nebuliser kit so that it is at an  
angle of greater than 45°.  
5
• Do not crease or fold the air tube.  
Cleaning Parts After Use (Section 5)  
Note: Failure to clean the nebulising parts  
after use may lead to infection.  
How the nebuliser kit works  
Baffle  
The compressed air from the  
Aerosol  
Aerosol  
compressor is forced through the  
nozzle, as it does so, it is mixed with  
the medication in the vaporiser head  
and forced upwards. The medication  
is then turned into a fine aerosol and  
directed upwards by the baffle,  
where it is broken down into many  
smaller aerosol particles.  
Vaporiser  
Head  
Nozzle  
Medication  
Medication  
Compressed Air  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
Operating instructions  
3. Preparing the nebuliser for use  
Warning:  
Clean and disinfect the nebuliser kit, mouthpiece and optional nosepiece or  
masks before using them for the first time after purchase, if the device has  
not been used for a long period of time, or if more than one person uses the  
same device.  
For directions on cleaning and disinfecting the parts of the device refer to Section 5  
in this instruction manual.  
Note: Make sure that the air filter is changed regularly. (See “7. Changing the air  
filter”.)  
3.1 Connecting the AC adapter  
1. Insert the power connector of the AC adapter  
into the AC adapter jack on the compressor.  
Power Connector  
AC Adapter Jack  
2. Plug the AC adapter into a power outlet.  
Power Outlet  
EN  
Note: When disconnecting the AC adapter, unplug the AC adapter from  
the power outlet before removing the power connector on the  
compressor.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11  
3. Preparing the nebuliser for use  
3.2 Adding medication and assembling the parts  
1. Remove the mouthpiece and cap from the  
nebuliser kit.  
2. Remove the inhalation top from the medication  
2
tank.  
1) Rotate the inhalation top anti-clockwise.  
1
2) Lift the inhalation top out of the medication  
tank.  
5
3
3. Remove the vaporiser head from the medication  
tank.  
4. Add the correct amount of prescribed  
medication to the medication tank.  
Caution: The capacity of the medication  
tank is 2 to 7 ml.  
max  
5
3
Note: Scales on the nebuliser kit are for  
reference only. The scale on the outside  
of the tank is with vaporiser head in place.  
The scale on the inside is with no vaporiser head. Please use the  
scale on your syringe or vial for accurate measurement of  
medication.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12  
3. Preparing the nebuliser for use  
5. Reattach the vaporiser head to the medication  
tank.  
6. Put the inhalation top back inside the  
medication tank.  
1
1) Align the protrusion on the inhalation top  
with the indent on the medication tank, as  
shown.  
max  
5
3
2) Rotate the inhalation top clockwise and align  
with the medication tank until they click in  
place.  
2
7. Attach the desired inhalation attachment, as described below.  
Attaching the mouthpiece  
2
Make sure you also attach the cap, when using  
the mouthpiece.  
1
1) Attach the cap to the inhalation air inlet.  
2) Attach the mouthpiece to the inhalation top.  
EN  
Attaching the adult mask (PVC)  
Attach the mask to the inhalation top.  
Note: If the amount of aerosol is too great,  
attach the cap.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
13  
3. Preparing the nebuliser for use  
Attaching the child mask (PVC)  
Attach the mask to the inhalation top.  
1) Attach the cap to the inhalation air inlet.  
2) Attach the child mask to the inhalation top.  
2
1
Attaching the nosepiece  
Attach the cap when using the nosepiece.  
1) Attach the cap to the inhalation air inlet.  
2) Attach the nosepiece to the inhalation top.  
2
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
14  
3. Preparing the nebuliser for use  
8. Attach the air tube.  
1) Attach the air tube to the air connector on the compressor.  
2) Attach the other end of the air tube to the air tube connector on the  
medication tank.  
2
1
Notes:  
• Make sure that the air tube is securely attached to the compressor (main  
unit) and nebuliser kit, and does not come loose. Twist the airtube plug  
slightly when inserting it into the connectors to avoid the tube  
disconnecting during use.  
Take care not to spill the medication when attaching the air tube. Keep  
the nebuliser kit upright.  
• Use the nebuliser kit holder as a temporary  
holder for the nebuliser kit.  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
15  
4. Taking medication  
1. Hold the nebuliser kit as indicated on the right.  
Caution:  
Do not tilt the nebuliser kit so that  
it is at an angle of greater than  
45°. Medication may flow into the  
mouth.  
45°  
2. Press the power switch.  
The compressor starts and nebulisation  
starts.  
Notes:  
• The green light shows that the power is  
on.  
• Observe whether aerosol is generated.  
To stop nebulisation  
Press the power switch again. The light goes out.  
Caution:  
Do not block the air filter cover.  
Notes:  
• Do not leave the compressor on for a long period of time. This could  
result in the compressor overheating or malfunctioning.  
• Limit use to 20 minutes at a time, and allow a 40 minute interval (cool off  
period) before using the device again.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16  
4. Taking medication  
3. Inhale medication according to your doctor’s and/or health care adviser’s  
instructions.  
Using the mouthpiece  
Prepare your mouthpiece according to  
Section 3.2.7.  
Insert the mouthpiece into your mouth and  
inhale the medication breathing normally.  
Exhale as normal through the mouthpiece.  
Using the nosepiece  
Prepare your nosepiece according to  
Section 3.2.7.  
Inhale the medication through the nosepiece, as  
shown and exhale through the mouth.  
Using the adult mask (PVC)  
Attach the mask so that it covers your nose and  
mouth and inhale the medication.  
Exhale through the mask.  
Note: If the amount of aerosol is too great,  
attach the cap.  
Using the child mask (PVC)  
EN  
Prepare your child mask according to  
Section 3.2.7.  
Attach the mask so that it covers your nose and  
mouth and inhale the medication.  
Exhale through the mask.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
17  
4. Taking medication  
4. When treatment is complete, turn the power off. Check that there is no  
condensation or moisture in the air tube and unplug the device from the  
power outlet.  
Power Outlet  
Warning:  
Condensation may occur in the air tube. Do not store the air tube  
while it contains condensation or moisture. This could result in  
infection via bacteria.  
If there is moisture or liquid remaining in the air tube, be sure to follow the  
procedure below to remove any moisture from within the air tube.  
1)Make sure that the air tube is still connected to the air connector on the  
compressor.  
2)Unplug the air tube from the nebuliser kit.  
3)Turn on the compressor and pump air through the air tube to expel the  
moisture.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18  
Care and maintenance  
5. Cleaning and daily disinfecting  
5.1 Cleaning  
Following the cleaning instructions after each use will prevent any remaining  
medication in the tank from drying, resulting in the device not nebulising effectively  
and also for preventing infections.  
Warning:  
Thoroughly rinse the nebuliser and the mask, nosepiece or mouthpiece with  
clean hot tap water after each use, and disinfect the parts of the nebuliser kit  
and applicable inhalation attachments after the last treatment of the day.  
Note: Do not start operation before the cleaned parts are completely dried.  
1. Make sure that the power indicator is out.  
Power Indicator  
2. Disconnect the AC adapter from the wall outlet  
and the compressor.  
3. Remove the nebuliser kit from the air tube and disassemble the nebuliser  
kit. (see Section 3)  
4. Discard any remaining medication in the  
medication tank.  
EN  
5. Wash all the nebulising parts in warm water and mild detergent and rinse in  
clean hot tap water, or wash in a dishwasher.  
For instructions on how to disinfect the parts, see the following section.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
19  
5. Cleaning and daily disinfecting  
6. After cleaning and disinfecting the parts, rinse them thoroughly with clean  
water and allow the parts to air dry in a clean environment.  
5
3
7. If the compressor is dirty, wipe it clean with a soft  
cloth moistened with water or mild detergent.  
Caution: The compressor is not  
waterproof.  
8. If the outside of the air tube is dirty, wipe it clean  
with a soft cloth moistened with water or mild detergent.  
9. Reassemble the nebuliser kit. Put the parts in a dry, sealed bag and place  
in the storage pocket of the carrying bag.  
Note: Nebuliser kit should be replaced after using for 6 months.  
5.2 Disinfecting the nebulising parts  
Always clean before disinfection. (See Section 5.1)  
Always disinfect the medication tank, mouthpiece and other nebulising parts after  
the last treatment of the day. To disinfect the parts, use one of the methods  
described below:  
A.Use a commercially available disinfectant. Follow the instructions provided by  
the disinfectant manufacturer.  
1. Submerge the parts in the cleansing solution for the specified period.  
2. Remove the parts and discard the solution.  
3. Rinse the parts with clean hot tap water, shake off excess water and allow  
to air dry in a clean environment.  
B.Parts may be boiled between 15 to 20 minutes (except for PVC mask).  
After boiling, carefully remove the parts, shake off excess water and allow to air dry  
in a clean environment.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
20  
5. Cleaning and daily disinfecting  
Handling the Vaporiser Head  
The vaporiser head is an important part that is used to vaporise medication.  
Be sure to observe the following precautions when handling it.  
Caution:  
• Always wash the vaporiser head after each  
use. Failure to do so could cause blockages  
and a deterioration in your condition due to a  
failure to vaporise medication effectively.  
Baffle  
• Do not use a brush or pin etc., to clean the  
parts. This could damage the baffle and cause  
a deterioration in your condition due to a fail-  
ure to vaporise medication effectively.  
• When disinfecting the parts by boiling, be  
sure to boil them in plenty of water. If there is  
insufficient water, the parts could become dis-  
torted due to the high temperature near the  
bottom of the pan. Distorted parts could  
Vaporiser head  
cause a deterioration in your condition due to a failure to vaporise medi-  
cation effectively.  
• Do not boil vaporiser head together with other objects, other than appli-  
cable nebuliser accessories. This could damage the baffle and cause a  
deterioration in your condition due to a failure to vaporise medication  
effectively.  
• Do not use a microwave oven, dish dryer or hair dryer to dry parts. The  
parts could become distorted due to the high temperature and cause a  
deterioration in your condition due to a failure to vaporise medication  
effectively.  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
21  
6. Maintenance and storage  
To keep your device in the best condition and protect the device from damage  
follow these directions:  
Caution:  
• Do not leave the unit unattended with infants or persons who cannot  
express their consent.  
• Do not subject the device or the parts to any strong shocks such as  
dropping the device on the floor.  
Notes:  
• Do not store the device in extreme hot or cold temperatures, high humidity or  
under direct sunlight.  
• Do not crease or fold the air tube.  
• Dispose of the device and any used accessories or optional parts according to  
applicable local regulations.  
• Never clean with benzene, thinner or a flammable chemical.  
Always transport the device in its carrying bag, and store the  
device in the bag if it will not be used for a long period.  
Caution:  
Do not carry or leave the nebuliser while the medication tank contains  
medication.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
22  
7. Changing the air filter  
On average, the air filter should be changed every 60 days or so, or when the filter  
has changed colour from white.  
1. Pull the air filter cover off the compressor.  
2. Change the air filter.  
Remove the old air filter with a sharp object,  
such as a toothpick, and insert the new air  
filter.  
3. Put the air filter cover back in place.  
Notes:  
• Use only OMRON air filters designed for this  
unit. Do not operate without a filter.  
• There is no front or back orientation for the  
air filters.  
• Check that air filters are clean and free of dust before inserting them.  
• Do not attempt to wash or clean the filter. If the air filter becomes wet,  
replace it. Damp air filters can cause blockages.  
To prevent the air filter cover from becoming blocked, wash the cover  
regularly. Do not boil the air filter cover.  
EN  
Remove the air filter cover, wash it, and make sure it is dry and  
reattached correctly.  
• Please contact your OMRON dealer for additional air filters, or if the parts  
supplied with the device are damaged or missing.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
23  
8. Troubleshooting  
Check the following if your unit should fail during operation. You can also refer to  
the pages of this manual for complete instructions.  
Symptom  
Cause  
Remedy  
Check that the plug is  
inserted in an electrical  
outlet. Unplug then reinsert  
the plug if necessary.  
Is the AC adapter connected  
correctly to the an electrical  
outlet and the compressor?  
Nothing happens when the  
power switch is pressed.  
(The power indicator does  
not light.)  
Is the optional rechargeable Make sure that the battery is  
battery connected correctly? correctly attached.  
Is the optional rechargeable  
Charge the battery.  
battery charged sufficiently?  
Is the optional DC auto  
Make sure that the adapter  
adapter connected  
is correctly attached.  
correctly?  
Is there medication in the  
Fill the medication tank.  
medication tank?  
Is there too much/little  
Add the correct amount of  
medication in the medication medication to the medication  
tank?  
tank.  
Is the vaporiser head  
missing and not assembled  
correctly?  
Attach the vaporiser head  
correctly.  
Is the nebuliser kit  
assembled correctly?  
Assemble the nebuliser kit  
correctly.  
Make sure that the nozzle is  
free of blockages.  
No nebulisation or low  
nebulisation rate, when the  
power is on. (The green  
power indicator is lit.)  
Is the nozzle blocked?  
Make sure that the nebuliser  
kit is not tilted at an angle of  
more than 45 degrees.  
Is the nebuliser kit tilted at a  
sharp angle?  
Make sure that the air tube is  
correctly connected to the  
compressor and nebuliser.  
Is the air tube connected  
correctly?  
Is the air tube folded or  
damaged?  
Make sure that the air tube  
does not contain kinks.  
Make sure that the air tube is  
free of blockages.  
Is the air tube blocked?  
Is the air filter dirty?  
Replace the air filter with a  
new one.  
The device is abnormally  
loud.  
Is the air filter cover attached Attach the air filter cover  
correctly? correctly.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
24  
8. Troubleshooting  
Symptom  
Cause  
Remedy  
The compressor is covered  
by a cloth.  
The device is very hot.  
Remove the cloth.  
Is the optional rechargeable  
battery charged sufficiently?  
Charge the battery.  
The battery has expired and  
cannot be recharged  
enough.  
Replace with a new battery.  
Charge the battery.  
The device stops soon when  
using the optional  
rechargeable battery, or  
operation time is short.  
Has the battery been left  
unused for a long time?  
Rundown, then recharge the  
battery two or three times.  
(See Section 11)  
Is the battery allowed to fully  
rundown before it is  
recharged?  
Notes:  
• If the suggested remedy does not solve the problem, do not try to repair the  
device - no parts of the unit are user serviceable.  
Return the unit to an authorized OMRON retail outlet or distributor.  
• The device should be serviced once a year by an approved OMRON service  
centre.  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
25  
9. Technical data  
Name:  
Comp Air Elite  
Model:  
OMRON NE-C30-E  
Type:  
Power Supply:  
Non-heating compressor nebuliser  
12VDC 1A (Compressor)  
100 - 240VAC, 50/60Hz (AC adapter)  
12 watts  
Power Consumption:  
Operating Temperature/  
Humidity:  
+10°C to +40°C 30% to 85% RH  
Storage Temperature/  
Humidity/Air Pressure:  
-20°C to +60°C 10% to 95% RH 700 - 1060 hPa  
Weight:  
440 g (compressor only)  
Dimensions:  
Contents:  
Approx. 124 (W) × 52 (H) × 103 (D) mm (compressor only)  
Compressor, Nebuliser Kit, Air Tube (PVC, 207cm),  
5 pcs Replacement Filters, Mouthpiece, Adult Mask (PVC),  
Child Mask (PVC), Nosepiece, AC Adapter, Carrying Bag,  
Instruction Manual, Warranty Card.  
Notes:  
• Subject to technical modification without prior notice.  
• This OMRON product is produced under the strict quality system of OMRON Healthcare Co. Ltd.,  
Japan.  
• The device may not work if the temperature and voltage conditions are different to those defined in  
the specifications.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26  
9. Technical data  
Technical data for the OMRON Comp Air compressor with the OMRON V.V.T.  
(Virtual Valve Technology) Nebuliser kit:  
Particle Size:  
Medication Tank  
Capacity:  
*MMAD **approximately 3µm  
7 ml maximum  
Appropriate Medication  
Quantities:  
2 ml minimum - 7 ml maximum  
Sound:  
Noise level (at 1 m distance) 53 db  
0.35 ml/min (by weight loss)  
**0.3 ml (2ml, 1%NaF)  
Nebulisation Rate:  
Aerosol Output:  
Aerosol Output Rate:  
**0.05 ml/min (2ml, 1%NaF)  
Notes:  
• Performance may vary with drugs such as suspensions or high viscosity. See  
drug supplier’s data sheet for further details.  
• The data above is referenced by Omron Healthcare.  
• See web site of Omron Healthcare Europe to update technical information.  
Result of cascade impactor** measurements for particle size with OMRON  
NE-C30-E and OMRON V.V.T. Nebuliser kit  
Cumulative % particle mass of sodium fluoride under size  
100  
90  
80  
70  
60  
50  
40  
30  
Individual tests  
20  
10  
0
Mean  
MMAD: 3µm  
EN  
0.1  
1
10  
100  
Particle Size Dp (µm)  
*MMAD = Mass Median Aerodynamic Diameter  
**Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to  
EN 13544-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
27  
9. Technical data  
Notes:  
= Type B  
= Type B  
• Do not use the device where it may be exposed to flammable gas.  
• This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device  
Directive) and the European Standard EN13544-1:2007, Respiratory therapy  
equipment - Part1: Nebulizing systems and their components.  
= Class ll  
= Read the instruction manual carefully  
Important information regarding Electro Magnetic Compatibility (EMC)  
With the increased number of electronic devices such as PC’s and mobile  
(cellular) telephones, medical devices in use may be susceptible to  
electromagnetic interference from other devices. Electromagnetic interference  
may result in incorrect operation of the medical device and create a potentially  
unsafe situation.  
Medical devices should also not interfere with other devices.  
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)  
with the aim to prevent unsafe product situations, the EN60601-1-2 standard  
has been implemented. This standard defines the levels of immunity to  
electromagnetic interferences as well as maximum levels of electromagnetic  
emissions for medical devices.  
This medical device manufactured by OMRON Healthcare conforms to this  
EN60601-1-2:2007 standard for both immunity and emissions.  
Nevertheless, special precautions need to be observed:  
• Do not use mobile (cellular) telephones and other devices, which generate  
strong electrical or electromagnetic fields, near the medical device. This may  
result in incorrect operation of the unit and create a potentially unsafe  
situation. Recommendation is to keep a minimum distance of 7 m. Verify  
correct operation of the device in case the distance is shorter.  
Further documentation in accordance with EN60601-1-2:2007 is available at  
OMRON Healthcare Europe at the address mentioned in this instruction  
manual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
28  
9. Technical data  
Correct Disposal of This Product  
(Waste Electrical & Electronic Equipment)  
This marking shown on the product or its literature, indicates that it should not  
be disposed of, with other household wastes at the end of its working life. To  
prevent possible harm to the environment or human health from uncontrolled  
waste disposal, please separate this from other types of wastes and recycle it  
responsibly to promote the sustainable reuse of material resources.  
Household users should contact either the retailer where they purchased this  
product, or their local government office, for details of where and how they can  
take this item for environmentally safe recycling.  
Business users should contact their supplier and check the terms and  
conditions of the purchase contract. This product should not be mixed with  
other commercial wastes for disposal.  
This product does not contain any hazardous substances.  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
29  
10.Optional parts and accessories  
OMRON V.V.T. Nebuliser  
Kit Set  
Air Tube (PVC, 207cm)  
Order No. 9956270-0  
Mouthpiece  
Order No. 9956273-5  
Order No. 9956268-9  
Nebuliser Kit  
Air Tube (Silicon, 100cm)  
Order No. 9956269-7  
Nosepiece  
Order No. 9956274-3  
Mouthpiece  
Child Mask (SEBS) Set  
Adult Mask (SEBS)  
Adult Mask (PVC)  
Order No. 9956281-6  
Order No. 9956312-0  
Order No. 9956275-1  
Child Mask (SEBS)  
Child Mask (PVC)  
Order No. 9956276-0  
Air Filters (x 5)  
Order No. 9956636-6  
Adapter for  
Child Mask (SEBS)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
30  
10. Optional parts and accessories  
Filter Cover  
Order No.9956638-2  
Instruction Manual  
Order No.9956633-1  
Carrying Bag  
Order No.9956631-5  
Battery Set  
Includes the rechargeable battery,  
battery charger.  
Rechargeable Battery  
Order No.9986457-0  
Battery Charger  
Order No.9986461-8  
Order No.9986453-7  
AC Adapter  
Order No.9956628-5  
EN  
Download from Www.Somanuals.com. All Manuals Search And Download.  
31  
11.Using optional accessories  
11.1 Using the optional rechargeable battery  
Use the rechargeable battery as shown in the chart below. For details, refer to the  
following pages.  
When the battery has  
Using the battery for  
not been used for a  
the first time.  
while.  
Fully charge the battery.  
Attach the battery to the  
Attach the battery to the  
main unit and use the  
main unit.  
device.  
Notes:  
• Limit use to 20 minutes at a time.  
• If the battery is frequently used  
without being fully charged, have  
an alternative power source ready.  
Charge indicator lights?  
No  
Yes  
Operate the device until the  
charge indicator lights.  
Charge the battery before  
using it next time.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
32  
11. Using optional accessories  
Important notes on using the optional rechargeable battery  
• Charge the battery before using it for the first time. The battery is not charged  
at the time of shipment.  
• Use the optional charger to charge the battery. The battery takes about 3 hours  
to charge fully.  
• A fully charged battery can be used to operate the device for approximately 35  
minutes, when used within the normal operating temperature range.  
Note: Limit use to 20 minutes at a time, and allow a 40 minute interval before  
using the device again.  
• Use the battery until the charge indicator on the main unit lights.  
• The battery can be charged approximately 300 times. After 300 charges the  
operating capacity of the battery will be reduced to about half its normal  
capacity. We therefore recommend that you replace the rechargeable battery  
with a new one after 300 charges. (If the battery is charged once each day and  
is used for three sessions each day, the battery should last for approximately  
one year.)  
• Always recharge the battery before using it again. Even if the battery has been  
fully charged after the previous use, its operating time can be reduced due to  
self-discharge.  
1. Charge the battery.  
1) Insert the power connector of the AC adapter into the AC adapter jack  
on the charger.  
2) Plug the AC adapter into a power outlet.  
2
EN  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
33  
11. Using optional accessories  
3) Attach the optional battery to the  
Charging Indicator blinks  
charger.  
Align the connectors on the battery and  
charger, and press the battery until it  
clicks into place. Charging starts and the  
green charging indicator blinks.  
3
2. After the battery is fully charged, remove the battery from the charger.  
The green charging indicator changes from blinking to remaining  
continuously lit when the battery is fully charged.  
1) Unplug the AC adapter from the  
ChargingIndicator  
remains lit  
power outlet, then disconnect the  
charger.  
2) While pressing the button on the side  
of the battery, lift the battery off the  
charger.  
Push Button  
3. Mount the compressor on the battery.  
Align the connector on the bottom of the compressor with that on the  
battery and mount the compressor on the battery. Push the compressor  
down until it clicks into place.  
Note: To separate the compressor and battery,  
press the push button on the side of the  
battery and lift the compressor off the battery.  
Push Button  
Download from Www.Somanuals.com. All Manuals Search And Download.  
34  
11. Using optional accessories  
When battery power is low  
The charge indicator on the main unit will light  
when battery power is low. Please note the  
following points about using the device after the  
charge indicator lights:  
• Stop using the device when the charge indicator  
comes on.  
• Battery power is low and the device will cease to  
operate shortly after the charge indicator comes  
on.  
Charge Indicator  
To continue using the device after the charge indicator comes on, use the AC  
adapter or an alternative power source to power the device.  
• Always fully charge the battery before using it again after the charge indicator  
lights.  
• If the operating time for each charge when using the battery become noticeably  
shorter than usual, refer to “About the Memory Effect” below.  
If the battery will not be used for a long time  
• If the battery is not used for approximately 10 days after being charged, the  
battery capacity will be reduced to about half that of normal due to self-discharge  
of the battery power. If the battery will not be used for a long time, use the battery  
until the charge indicator on the main unit lights (to fully discharge the battery)  
before charging it.  
About the Memory Effect  
• If the battery is repeatedly charged before it has fully discharged, the operating  
time possible for each charge will gradually be reduced. This is known as the  
“memory effect” and is normal for nickel metal hydride (NiMH) batteries. It does  
not indicate a battery error or malfunction.  
EN  
• If the operating time for each charge become reduced, fully charge the battery  
and operate the device until the unit stops operating (rather then when the charge  
indicator lights) two or three times in a row. This should restore the battery  
capacity to normal. If the battery capacity does still not return to normal, replace  
the battery with a new one.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
35  
Manufacturer  
OMRON HEALTHCARE Co., Ltd.  
53, Kunotsubo, Terado-cho, Muko, Kyoto,  
617-0002 JAPAN  
EU-representative  
OMRON HEALTHCARE EUROPE B.V.  
Scorpius 33, 2132 LR Hoofddorp  
THE NETHERLANDS  
OMRON (DALIAN) CO., LTD.  
Production facility  
Subsidiary  
Dalian, CHINA  
OMRON HEALTHCARE UK LTD.  
Opal Drive  
Fox Milne, Milton Keynes, MK15 0DG, U.K.  
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH  
John-Deere-Str. 81a, 68163 Mannheim, GERMANY  
OMRON SANTÉ FRANCE SAS  
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE  
Made in China  
Download from Www.Somanuals.com. All Manuals Search And Download.  
36  

Multi Tech Systems Network Card MT5600BAV90 User Manual
National Instruments Network Card VXI MXI Express Series User Manual
NeumannBerlin Microphone KMS 104 User Manual
NeumannBerlin Microphone U 87 i User Manual
Nexstar Computer Drive NST 210S2 BK User Manual
Nikon Binoculars 10x50CF User Manual
NordicTrack Home Gym 831298961 User Manual
Nortel Networks Switch 8691SF User Manual
Olympus Camcorder C 220 ZOOM User Manual
Panasonic CD Player CQ DFX88EUC User Manual