Omega Thermometer CL3512A User Manual

#17000-0513-OME1  
CL3512A  
Digital Thermometer & Calibrator  
OMEGAnetSM On-Line Service  
Internet e-mail  
Servicing North America:  
USA: ISO 9001 Certified  
One Omega Drive, Box 4047  
Stamford, CT 06907-0047  
Tel: (203) 359-1660  
FAX: (203)359-7700  
Canada:  
976 Bergar  
Laval (Quebec) H7L5A1  
Tel: (514) 856-6928  
FAX: (514) 856-6886  
For immediate technical or application assistance:  
USA and Canada:  
Mexico and Latin America:  
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGASM  
Customer Service: 1-800-622-2378 / 1-800-622-BESTSM  
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSM  
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA  
Tel: (95) 800-TC-OMEGASM  
FAX: (95) 203-359-7807  
En Espanol: (203) 359-7803  
Servicing Europe:  
Benelux:  
Germany/Austria:  
Postbus 8034, 1180 LA Amstelveen,  
The Netherlands  
Daimlerstrasse 26, D-75392  
Deckenpfronn, Germany  
Tel: (31) 20 6418405 FAX: (31) 20 6434643  
Toll Free in Benelux: 06 0993344  
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540  
Toll Free in Germany: 0800 82 66342  
Czech Republic:  
United Kingdom: ISO 9002 Certified  
One Omega Drive  
Riverbend Technology Centre Northbank, Irlam,  
Manchester, M44 5EX, England  
ul. Rude armady 1868, 733 01 Karvina-Hranice, Czech Repubic  
Tel: 420 (69) 6311627 FAX: 420 (69) 6311114  
Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622  
France:  
9, rue Denis Papin, 78190 Trappes  
Tel: (33) 130-621-400 FAX: (33)130-699-120  
Toll Free in France: 0800-4-06342  
Toll Free in England: 0800-488-488  
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its  
products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.  
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and  
reserves the right to alter specifications without notice.  
WARNING: These products are not designed for use in, and should not be used for, patient connected application.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.0 Introduction  
The OMEGA® CL3512A calibrator/thermometer is two meters in one. The  
CL3512A simulates type J/K/T/E thermocouple signals. Each signal is  
adjustable by using the coarse and fine dials. The CL3512A can also be used  
as a dual type J/K/T/E thermocouple input thermometer. Features include a  
large 3½ digit display with backlighting and display selections of HOLD, °C/°F,  
and0.1/1°.  
The source mode of the CL3512A simulates the thermocouple output to check  
the operation of  
a
thermocouple meter and make rough calibration  
adjustments. A more accurate calibrator would be required for calibration of  
thermocouple meters to specify tolerances.  
1.1 Safety Information  
It is recommended that you read the safety and operation instructions before  
using the thermometer.  
WARNING  
To avoid electrical shock, do not use this instrument when working  
voltages at the measurement surface over 24V AC or DC.  
WARNING  
To avoid damage or burns, do not make temperature measurement  
in microwave ovens.  
CAUTION  
Repeated sharp flexing can break the thermocouple leads. To prolong  
lead life, avoid sharp bends in the leads, especially near the connector.  
1
1.2 Unpacking  
Remove the packing list and verify that all equipment has been received. If  
there are any questions about the shipment, please call the OMEGA  
Customer Service Department.  
Upon receipt of shipment, inspect the container and equipment for any signs  
of damage. Immediately report any damage to the shipping agent.  
NOTE  
The carrier will not honor any claims unless all shipping  
material is saved for their examination. After examining and  
removing contents, save the packing material and carton in  
the event reshipment is necessary.  
The following is supplied in the box:  
CL3512A  
Rubber boot  
2 K beaded wire thermocouples  
K calibration cable  
9 volt battery  
Operator's manual  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.0 Operation Procedure  
2.1 Drawing of Unit  
Actual Dimensions: 195mm(H) x 92mm(W) x 53mm(D)  
3
2.2 Descriptions of Buttons and Switches  
/
Buttons  
Temperature is displayed in either degrees Celsius(°C) or degrees  
Fahrenheit(°F). When the thermometer is turned on, it is set to the  
temperature scale that was in use when the thermometer was last turned off.  
To change the temperature scale, press the °C or °F key.  
Button  
Press “ ” button to toggle the on and off backlight. The backlight will  
switch-off automatically after 37 seconds.  
Button (HOLD Mode)  
Press the HOLD key to enter the Data Hold mode, the “HOLD” annunciator is  
displayed. When HOLD mode is selected, the thermometer will hold the  
present reading and stop all further measurements.  
Pressing the HOLD key again will cancel HOLD mode causing the  
thermometer to resume taking measurements.  
Button (K/J/T/E Input Thermocouple Type Selection)  
The TYPE key allows for selection of J, K, E or T thermocouple types as either  
the input or simulated output. To select thermocouple type press type key  
once. Unit will briefly show 1888 and beep and then the type of thermocouple  
selected will be indicated on the left hand side of the display. Continue  
pressing the type key to step through the thermocouple types until desired  
type is indicated.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
/
/
Input Selections  
The input selection indicates which input is selected; T1 thermocouple, T2  
thermocouple or the difference between the two thermocouples (T1-T2).  
When the thermometer is turned on, it is set to the temperature input that was  
in use when the thermometer was last turned off.  
Selector Switch  
The circular selector switch is used to turn the unit off () or to select Reading  
Mode (Input) or Simulation Mode (Output). There are nine selector switch  
positions under output which provide rough adjustment of the simulated  
output.  
Adjust  
Course and fine adjustment Dials are used in Simulation Mode to allow a  
particular simulated temperature to be selected once the output has been  
adjusted close to the desired temperature using the selector switch.  
5
2.3 Read Mode Procedure  
1. Plug Thermocouple Sensor into input T1 and/or T2.  
2. Turn selector switch to Input.  
3. Select proper thermocouple type (J, K, T or E) using type button.  
4. Press T1 button to read thermocouple connected to input T1.  
Press T2 button to read thermocouple connected to input T2.  
Press T1-T2 if difference between T1 and T2 is to be read.  
5. To Hold reading press HOLD button. To resume making measurements  
press HOLD again.  
Note: Unit will display -OL if input is out of range or thermocouple is broken  
or not connected to proper input.  
To save battery life, CL3512A will turn off automatically if no is key is pressed  
for 90 minutes.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.4 Source Mode Procedure (Calibration Simulation)  
1. Turn CL3512A on by turning selector switch to any one of the output  
selector positions.  
2. Press T1 button.  
3. Select thermocouple type (J, K, T or E) using type button.  
Select °C or °F using °C or °F buttons.  
4. Plug proper type of thermocouple wire into either one of the two output  
connectors. (SMP male connector required)  
5. Attach the other end of the thermocouple wire to the instrument being  
tested/calibrated.  
6. Move the output range switch to the temperature value close to the desired  
simulated value.  
Note: There are 9 temperature range switch positions arranged from left to  
right representing low to high temperatures.  
The display will read -OL if the output voltage for the meter is higher  
than the maximum voltage for that particular thermocouple type. For  
type T this will be in the highest 4 positions, type K in the highest 3  
positions, type J in the highest 2 positions, and type E in the highest  
position only.  
7. Adjust course and/or fine adjustment dial until desired temperature is  
displayed on CL3512A.  
8. Instrument under test should read same temperature as CL3512A. If not  
adjust instrument under test or have recalibrated.  
9. Repeat steps 6-8 at different temperatures as needed.  
7
Source Mode Diagram  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.0 Operator Maintenance  
WARNING  
To avoid possible electrical shock, disconnect the thermocouple  
connectors from the thermometer before removing the cover.  
Battery Replacement  
Power is supplied by a 9 volt “transistor” battery. The “ ” appears on the  
LCD display when replacement is needed. To replace the battery, remove the  
two screws from the back of the meter and lift off the battery cover. Remove  
the battery from battery contacts.  
9
4.0 Specifications  
ELECTRICAL  
Temperature Scale: Celsius or Fahrenheit user-selectable  
Measurement Range:  
Thermocouple  
K-TYPE(0.1°C)  
J-TYPE(0.1°C)  
T-TYPE(0.1°C)  
E-TYPE(0.1°C)  
Range  
-200°C to 1372°C, -328°F to 1999°F  
-210°C to 1200°C, -346°F to 1999°F  
-200°C to 400°C , -328°F to 752°F  
-220°C to 1000°C, -364°F to 1832°F  
Calibration Range: -210°C to 1372°C, (-364°F to 1999°F)  
Auto range: 0.1°C/1°C, 0.1°F/1°F  
Accuracy: Accuracy is specified for operating temperatures over the range  
of18°C to 28°C (64°F to 82°F), for 1 year, not incl uding thermocouple  
error.(Read mode)  
±(0.1%rdg + 1°C) range -60°C to 1372°C  
±(0.1%rdg + 2°C) range -60°C to -220°C  
±(0.1%rdg + 2°F) range -76°F to 1999°F  
±(0.1%rdg + 4°F) range -76°F to -364°F  
Simulation Accuracy: Measurement accuracy plus 1°C  
ENVIRONMENTAL  
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.  
Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
GENERAL  
Display: 3½ digit liquid crystal display (LCD) with a maximum reading of1999.  
Polarity: Automatic, positive implied, negative polarity indication.  
Overrange: -OL is displayed.  
Zero: Automatic.  
Low battery indication: The “ ” is displayed when the battery voltage  
drops below the operating level.  
Measurement rate: 1 times/second.  
Accuracy: Stated accuracy at 23°C±5°C, <75% R.H.  
Dimensions: 195mm(H) x 92mm(W) x 53mm(D).  
Weight: approx. 9 oz. (250g) including battery.  
Input Connector: Accepts standard miniature thermocouple connectors (flat  
blades spaced 7.9mm, center to center SMP type).  
Battery Life: 100 hours typical with carbon zinc battery.  
Auto power off: The meter key switch inactive for more than 90 minutes.  
Temperature Coefficient: 0.1 times the applicable accuracy specification per  
°C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F  
and 82°F to 122°F).  
Input Protection: 24V dc or 24V ac rms maximum input voltage on any  
combination of input pins.  
Maximum Differential Common Mode Voltage (Maximum Voltage  
betweenT1 and T2 during measurement): 1volt.  
11  
5.0 CL3152A Calibration Procedure  
Note: The following calibration procedure should be performed only by  
qualified technicians who have access to the items as following items:  
Equipment: Thermocouple calibrator had better 10 times greater than the  
measured meter.  
1. Short CAL jumper and turn on the meter to INPUT range.  
2. Press “  
” key display was showed 35°C / T1/ SET / K.  
3. Input 35°C / K-type to the T1, then press “  
display was showed 1300°C / T1 / SET / K.  
4. Input 1300°C / K-type to the T1, then press “  
display was showed 0°C / T1 / SET / K.  
” key and wait for the  
” key and wait for the  
5. Input 0°C/K-type to the T1, then press “  
was showed “-OL” then press “  
35°C / T2 / SET / K.  
” key and wait for the display  
” key and wait for the display was  
6. Input 35°C / K-type to the T2, then press “  
display was showed 1300°C / T2 / SET / K.  
7. Input 1300°C / K-type to the T2, then press “  
display was showed Out / K / °C / SET / T2.  
” key and wait for the  
” key and wait for the  
8. Knob being with clockwise rotation 2 range and two knobs of lower left  
corner are set to the position of the middle (horizontal).  
9. Press “  
” key and wait for the meter auto power off then remove  
jumper.  
10. The calibration was completed.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY / DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from  
date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to  
cover handling and shipping time. This ensures that OMEGA's customers receive maximum coverage on each product. If the unit  
should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an  
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be  
defective it will be repaired or replaced at no charge. OMEGA's WARRANTY does not apply to defects resulting from any action of  
the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or  
unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of  
being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication;  
misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not  
limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes  
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in  
accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts  
manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR  
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED  
WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE  
HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total  
liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict  
liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall  
OMEGA be liable for consequential, incidental or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10  
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any  
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,  
OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will  
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in  
such a manner.  
RETURN REQUESTS / INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY  
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S  
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be  
marked on the outside of the return package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.  
FOR WARRANTY RETURNS, please have the following  
information available BEFORE contacting OMEGA:  
1. P.O. number under which the product was PURCHASED.  
2. Model and serial number of the product under warranty, and  
3. Repair instructions and/or specific problems relative to the  
product.  
FOR NON-WARRANTY REPAIRS, consult OMEGA for current  
repair charges. Have the following information available  
BEFORE contacting OMEGA:  
1. P.O. number to cover the COST of the repair.  
2. Model and serial number of product , and  
3. Repair instructions and/or specific problems relative to the  
product.  
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in  
technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2002 OMEGA ENGINEERING, INC.  
All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or  
machine-readable from, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.  
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
TEMPERATURE  
DATA ACQUISITION  
Data Acquisition & Engineering Software  
Communications-Based  
Acquisition Systems  
Thermocouple, RTD & Thermistor  
Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
Plug-in Cards for Apple, IBM &  
Compatibles  
Datalogging Systems  
Recorders, Printers & Plotters  
PRESSURE/STRAIN AND FORCE  
Transducers & Strain Gauges  
Load Cells & Pressure Gauges  
Displacement Transducers  
HEATERS  
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Instrumentation & Accessories  
FLOW/LEVEL  
Rotameters, Gas Mass Flowmeters &  
Laboratory Heaters  
Flow Computers  
ENVIRONMENTAL  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
MONITORING AND CONTROL  
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
pH/CONDUCTIVITY  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen  
Instruments  
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators  
& Pumps  
Industrial pH & Conductivity Equipment  
M-3806/0610  
Download from Www.Somanuals.com. All Manuals Search And Download.  

NETGEAR Network Router DGN1000 100NAS User Manual
Nikon Camcorder SB1L01 User Manual
Nokia Wireless Office Headset HS 36W User Manual
Nostalgia Electrics Fondue Maker CHM 915 User Manual
Nuvo Home Theater System NV E6GM User Manual
Pacific Digital Automobile PTV3606 User Manual
Palsonic CRT Television 6125TS User Manual
Panasonic Network Card 2SC4691J User Manual
Panasonic Printer CL500 User Manual
Paradyne Network Card 3151 CSU User Manual