Important Information
Warranty
The NI PXI-7831R is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced
by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the
warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are
covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,
flood, accident, actions of third parties, or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
Trademarks
LabVIEW™, MXI™, National Instruments™, NI™, ni.com™, NI Developer Zone™, and RTSI™ are trademarks of National Instruments
Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
Patents
For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txtfile
on your CD, or ni.com/patents.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Download from Www.Somanuals.com. All Manuals Search And Download.
Compliance
FCC/Canada Radio Frequency Interference Compliance
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)
or Class B (for use in residential or commercial locations). All National Instruments (NI) products are FCC Class A products.
Depending on where it is operated, this Class A product could be subject to restrictions in the FCC rules. (In Canada, the
Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.) Digital
electronics emit weak signals during normal operation that can affect radio, television, or other wireless products.
All Class A products display a simple warning statement of one paragraph in length regarding interference and undesired
operation. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.
FCC/DOC Warnings
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions
in this manual and the CE marking Declaration of Conformity*, may cause interference to radio and television reception.
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department of
Communications (DOC).
Changes or modifications not expressly approved by NI could void the user’s authority to operate the equipment under the FCC
Rules.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user is required to correct the interference
at their own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Compliance to EU Directives
Readers in the European Union (EU) must refer to the manufacturer’s Declaration of Conformity (DoC) for information*
pertaining to the CE marking compliance scheme. The manufacturer includes a DoC for most hardware products except for those
bought from OEMs. In addition, DoCs are usually not provided if compliance is not required, for example electrically benign
apparatus or cables.
To obtain the DoC for this product, click Declaration of Conformity at ni.com/hardref.nsf/. This Web site lists the DoCs
by product family. Select the appropriate product family, followed by your product, and a link to the DoC appears in Adobe
Acrobat format. Click the Acrobat icon to download or read the DoC.
*
The CE marking Declaration of Conformity contains important supplementary information and instructions for the user or
installer.
Download from Www.Somanuals.com. All Manuals Search And Download.
About This Manual
Conventions ...................................................................................................................vii
Chapter 1
Reconfigurable I/O Architecture.....................................................................1-6
FPGA Module .................................................................................................1-9
RT Module.......................................................................................................1-9
Custom Cabling .............................................................................................................1-10
Unpacking......................................................................................................................1-11
Connecting Analog Input Signals..................................................................................2-4
Types of Signal Sources ................................................................................................2-5
Floating Signal Sources...................................................................................2-6
Ground-Referenced Signal Sources ................................................................2-6
Input Modes ...................................................................................................................2-6
Differential Connection Considerations (DIFF Input Mode)..........................2-8
© National Instruments Corporation
v
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Differential Connections for Ground-Referenced Signal Sources ... 2-8
Differential Connections for Nonreferenced or
Connecting Analog Output Signals............................................................................... 2-14
Connecting Digital I/O Signals ..................................................................................... 2-15
PXI Trigger Bus ............................................................................................................ 2-18
Switch Settings.............................................................................................................. 2-20
Chapter 3
Loading Calibration Constants...................................................................................... 3-1
Internal Calibration........................................................................................................ 3-1
External Calibration....................................................................................................... 3-2
Appendix A
Specifications
Appendix B
Connecting I/O Signals
Appendix C
Appendix D
Technical Support and Professional Services
Glossary
NI PXI-7831R User Manual
vi
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
About This Manual
This manual describes the electrical and mechanical aspects of the
National Instruments PXI-7831R device and contains information
concerning its operation and programming.
The NI PXI-7831R device is a Reconfigurable I/O (RIO) device.
The NI PXI-7831R contains eight independent, 16-bit analog input (AI)
channels, eight independent, 16-bit analog output (AO) channels, and
96 digital I/O (DIO) lines.
Conventions
The following conventions appear in this manual:
<>
Angle brackets that contain numbers separated by an ellipsis represent a
range of values associated with a bit or signal name—for example,
DIO<3..0>.
»
The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash. When this symbol is marked on
the device, refer to the Safety Information section of Chapter 1,
Introduction, for precautions to take.
bold
Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter
names and hardware labels.
italic
Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
© National Instruments Corporation
vii
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
About This Manual
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames and extensions, and code excerpts.
Reconfigurable I/O Documentation
The NI PXI-7831R User Manual is one piece of the documentation set for
your RIO system and application. Depending on the hardware and software
you use for your application, you could have any of several types of
documentation. Use the documentation you have as follows:
•
Where to Start with the NI PXI-7831R—This document lists what you
need to get started, describes how to unpack and install the hardware,
and contains information about connecting signals to the
NI PXI-7831R.
•
•
NI PXI-7831R User Manual—This manual contains detailed
information about the NI PXI-7831R hardware.
LabVIEW FPGA Module Release Notes—This document contains
information about installing and getting started with the FPGA
Module.
•
•
LabVIEW FPGA Module User Manual—This manual describes how
to use the FPGA Module.
LabVIEW Help—This help contains information about using various
virtual instruments (VIs) with the NI PXI-7831R and using the FPGA
Module and the LabVIEW Real-Time (RT) Module.
•
LabVIEW Real-Time Module User Manual—This manual contains
information about how to install and use the RT Module.
Related Documentation
The following documents contain information you might find helpful:
•
NI Developer Zone tutorial, Field Wiring and Noise Considerations
for Analog Signals, at ni.com/zone
•
•
•
PICMG CompactPCI 2.0 R3.0
PXI Hardware Specification Revision 2.1
PXI Software Specification Revision 2.1
NI PXI-7831R User Manual
viii
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Introduction
This chapter describes the NI PXI-7831R, describes the concept of the
Reconfigurable I/O (RIO) device, lists what you need to get started,
describes the optional software and optional equipment, explains how
to unpack the hardware, and contains safety information about the
NI PXI-7831R.
About the Reconfigurable I/O Devices
Thank you for purchasing the NI PXI-7831R. This RIO device has 96
digital I/O (DIO) lines, 8 independent, 16-bit analog output (AO) channels,
and 8 independent, 16-bit analog input (AI) channels.
A user-reconfigurable field-programmable gate array (FPGA) controls the
digital and analog I/O on the NI PXI-7831R. The FPGA on the RIO device
allows you to define the functionality and timing of the device, whereas
traditional multifunction I/O (MIO) devices have a fixed functionality
provided by an application-specific integrated circuit (ASIC). You can
change the functionality of the FPGA on the RIO device by using
LabVIEW, a graphical programming environment, and the LabVIEW
FPGA Module to create and download a custom virtual instrument (VI) to
the FPGA. You can reconfigure the RIO device with a new VI at any time.
Using LabVIEW, you can graphically design the timing and functionality
of the RIO device without having to learn the low-level programming
language or hardware description language (HDL) that is traditionally used
for FPGA design. If you only have LabVIEW and do not have the FPGA
Module, you cannot create new FPGA VIs but you can create VIs that run
in LabVIEW to control existing FPGA VIs.
Some applications require tasks such as real-time, floating-point
processing or data logging while performing I/O and logic on the RIO
device. You can use the LabVIEW Real-Time (RT) Module to perform
these additional applications while also communicating with and
controlling the RIO device.
The RIO device contains flash memory to store VIs for instant loading of
the FPGA when the system is powered on.
© National Instruments Corporation
1-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
synchronize several measurement functions to a common trigger or timing
event. The RTSI bus is implemented on the PXI trigger bus on the PXI
backplane. The RTSI bus can route timing and trigger signals between as
many as seven PXI devices in your system.
Refer to Appendix A, Specifications, for detailed specifications of the RIO
device.
Using PXI with CompactPCI
Using PXI compatible products with standard CompactPCI products is an
important feature provided by PXI Hardware Specification Revision 2.1
and PXI Software Specification Revision 2.1. If you use a PXI-compatible
plug-in card in a standard CompactPCI chassis, you cannot use
PXI-specific functions, but you can still use the basic plug-in card
functions. For example, the RTSI bus on the RIO device is available in a
PXI chassis, but not in a CompactPCI chassis.
The CompactPCI specification permits vendors to develop sub-buses that
coexist with the basic PCI interface on the CompactPCI bus. Compatible
operation is not guaranteed between CompactPCI devices with different
The standard implementation for CompactPCI does not include these
sub-buses. The RIO device works in any standard CompactPCI chassis
adhering to PICMG CompactPCI 2.0 R3.0.
PXI-specific features are implemented on the J2 connector of the
CompactPCI bus. Table 1-1 lists the J2 pins used by the NI PXI-7831R.
The NI PXI-7831R is compatible with any CompactPCI chassis with a
sub-bus that does not drive these lines. Even if the sub-bus is capable of
driving these lines, the RIO device is still compatible as long as those pins
on the sub-bus are disabled by default and are never enabled.
Caution Damage can result if the J2 lines are driven by the sub-bus.
NI PXI-7831R User Manual
1-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
Table 1-1. Pins Used by the NI PXI-7831R
NI PXI-7831R Signal
PXI Pin Name
PXI J2 Pin Number
PXI Trigger<0..7>
PXI Trigger<0..7>
A16, A17, A18, B15, B18, C18,
E16, E18
PXI Clock 10 MHz
PXI Star Trigger
LBLSTAR<0..12>
PXI Clock 10 MHz
PXI Star Trigger
LBL<0..12>
E17
D17
A1, A19, C1, C19, C20, D1, D2,
D15, D19, E1, E2, E19, E20
LBR<0..12>
LBR<0..12>
A2, A3, A20, A21, B2, B20, C3,
C21, D3, D21, E3, E15, E21
What You Need to Get Started
This section contains two lists that detail what you need to get started using
the NI PXI-7831R with Windows 2000/XP or the RT Module.
Getting Started with Windows 2000/XP
To set up and use the NI PXI-7831R with Windows 2000/XP, you need the
following items:
❑ NI PXI-7831R
❑ The following software packages:
–
–
–
LabVIEW version 7.0 or later
NI Device Drivers CD
FPGA Module version 7.0 or later (required to develop custom
FPGA VIs for the RIO device)
❑ PXI/CompactPCI chassis and a PXI/CompactPCI embedded
controller, running Windows 2000/XP (or any computer running
Windows 2000/XP and an MXI-3 link to a PXI/CompactPCI chassis)
❑ At least one cable and terminal block for connecting signals to the
NI PXI-7831R
© National Instruments Corporation
1-3
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
❑ The following documents are included on the NI Device Drivers CD
and are also available at ni.com/manuals(optional):
–
–
–
LabVIEW FPGA Module Release Notes
LabVIEW FPGA Module User Manual
Where to Start with the NI PXI-7831R
❑ The LabVIEW Help, which is available by selecting Help»VI,
Function, & How-To Help from LabVIEW.
Getting Started with the RT Module
To set up and use the NI PXI-7831R with the FPGA Module and the
RT Module, you need the following items:
❑ NI PXI-7831R
❑ The following software packages:
–
–
–
LabVIEW version 7.0 or later
NI Device Drivers CD
FPGA Module version 7.0 or later (required to develop custom
FPGA VIs for the RIO device)
–
RT Module version 7.0 or later
❑ PXI/CompactPCI chassis and real-time PXI controller
❑ One of the following host computers, depending upon your
application, running Windows 2000/XP:
–
–
–
PC
Laptop computer
PXI/CompactPCI embedded controller
❑ At least one cable and terminal block for connecting signals to the
NI PXI-7831R
❑ Category 5 (Cat-5) crossover cable (if the real-time PXI system is not
configured on a network). You need a regular network cable if you are
configured on a network.
NI PXI-7831R User Manual
1-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
❑ The following documents are included on the NI Device Drivers CD
and are also available at ni.com/manuals(optional):
–
–
–
–
LabVIEW FPGA Module Release Notes
LabVIEW FPGA Module User Manual
LabVIEW Real-Time Module User Manual
Where to Start with the NI PXI-7831R
❑ The LabVIEW Help, which is available by selecting Help»VI,
Function, & How-To Help from LabVIEW.
Overview of Reconfigurable I/O
This section introduces the concept of RIO and describes how to use
the reconfigurable FPGA to build high-level functions in hardware.
Refer to Chapter 2, Hardware Overview of the NI PXI-7831R, for
descriptions of the physical I/O resources available on the NI PXI-7831R.
Reconfigurable I/O Concept
The NI PXI-7831R device is based on a reconfigurable FPGA core
surrounded by fixed I/O resources. The behavior of the reconfigurable core
can be configured to better match the requirements of the measurement and
control system. The behavior can be fully user defined and implemented as
a VI, creating an application-specific I/O device. In contrast, a traditional
data acquisition (DAQ) device uses a fixed core with predetermined
functionality.
Flexible Functionality
Flexible functionality allows the RIO device to match individual
application requirements and to mimic the functionality of fixed I/O
devices, including I/O combinations not available in standard products. For
example, you can configure a RIO device in one application for three 32-bit
quadrature decoders and then reconfigure the RIO device in another
application for eight 16-bit event counters.
In timing and triggering applications, the flexible functionality of the RIO
device makes it an ideal complement to applications based on the RT
module, such as control and hardware-in-the-loop (HIL) simulations. For
example, you can configure the RIO device for a single timed loop in one
application and then reconfigure the device in another application for four
independent timed loops with separate I/O resources.
© National Instruments Corporation
1-5
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
User-Defined I/O Resources
With the RIO device, you can define both the combination of I/O resources
and the I/O resources themselves. You can also create new building blocks
on top of fixed I/O resources. For example, one application might require
an event counter that increments when a rising edge appears on any of three
digital input lines. Another application might require a digital line to be
asserted once an analog input exceeds a programmable threshold. You can
implement these user-defined behaviors in the hardware for fast,
deterministic performance.
Device-Embedded Logic and Processing
You can embed logic and processing in the FPGA of the RIO device.
Typical logic functions include Boolean operations, comparisons, and
basic mathematical operations. You can implement multiple functions
possible to implement more complex algorithms such as control loops,
but the size of the FPGA limits the scope of these algorithms.
Reconfigurable I/O Architecture
Figure 1-1, which illustrates a generic representation of RIO device, shows
an FPGA connected to fixed I/O resources and a bus interface.
Fixed I/O Resource
Fixed I/O Resource
Fixed I/O Resource
FPGA
Fixed I/O Resource
Bus Interface
Figure 1-1. High-Level FPGA Functional Overview
The fixed I/O resources include A/D converters (ADCs), D/A converters
(DACs), digital input or output lines, or other I/O resources. Software
accesses the RIO device through the bus interface, and the FPGA provides
NI PXI-7831R User Manual
1-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
the connectivity between the bus interface and the fixed I/O, including any
timing, triggering, processing, and custom I/O required by the application.
Timing, triggering, processing, and custom I/O is provided by consuming
logic in the FPGA. Each fixed I/O resource used by the application
consumes a small portion of the FPGA logic, which is used to perform
basic control of the fixed I/O resource. The bus interface also consumes a
amounts of logic. For example, a typical 32-bit counter consumes 20 times
more logic than a DIO resource, while an 8-bit counter consumes five times
more logic than a DIO resource. Figures 1-2 and 1-3 illustrate the logic
used by the FPGA in two different applications. The application shown in
Figure 1-2 requires many fixed I/O resources, leaving little logic left over
for higher-level functions. The application in Figure 1-3 uses relatively few
I/O resources and has enough logic left over for several large functions.
AI0
AI1
AI2
AI3
DIO<0..7>
Bus Interface
DIO<8..15>
AO3
AO2
AO1
AO0
Figure 1-2. FPGA Logic Use in an Application with Many Fixed I/O Resources
© National Instruments Corporation
1-7
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
Counter
DIO<0..7>
Bus Interface
PID
AO0
Figure 1-3. FPGA Logic Use in an Application with Higher-Level Functions
The FPGA is volatile and does not retain the VI when it is powered off.
Therefore, the VI must be reloaded every time power is turned on. The VI
comes from onboard flash memory or from the software over the bus
interface. One advantage to using flash memory is that the VI can start
executing almost immediately after power up, instead of waiting for the
computer to completely boot and load the FPGA. Refer to the LabVIEW
FPGA User Manual for more information about how to store your VI in
flash memory.
Reconfigurable I/O Applications
To create or obtain new VIs for your application, you can use the FPGA
Module, which allows the application to be specified using a subset of
LabVIEW. Arbitrary functionality can be defined for the RIO device. If
you are using the FPGA Module, refer to the FPGA Module examples
located in LabVIEW 7.0\examples\FPGA.
Software Development
You can use LabVIEW with the FPGA Module to program the
NI PXI-7831R. To develop real-time applications that control the
NI PXI-7831R, you can use the RT Module with LabVIEW and the
FPGA Module.
NI PXI-7831R User Manual
1-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
FPGA Module
The FPGA Module enables you to use LabVIEW to create VIs that run on
the RIO device, which contains a reconfigurable FPGA. The FPGA
Module includes a new function palette, which contains functions that run
on the FPGA on the RIO device. These functions can control the I/O,
timing, and logic of the RIO device and can generate interrupts for
synchronization. The FPGA Module synthesizes a VI into a form that can
be downloaded to the FPGA on the RIO device. The Interactive Front Panel
Communication with the FPGA Module allows you to interact with the VI
running on the FPGA. The FPGA Module also includes a palette of
functions for use in LabVIEW for Windows, or when targeting an RT
Module device, that create applications that wait for interrupts and that
control the FPGA by programmatically reading and writing to the device.
Note A software utility installed with the NI-RIO Device Drivers CD allows users without
the FPGA module to configure the NI PXI-7831R analog input mode, synchronize to the
PXI clock, and configure the device to automatically load FPGA VIs when powered on.
RT Module
The RT Module extends the LabVIEW development environment to
deliver deterministic, real-time performance.
You can develop your RT Module application on a host computer
with graphical programming and then download the program to run on
an independent hardware target with a real-time operating system. The
RT Module allows you to use the NI PXI-7831R in PXI systems being
controlled in real time by a LabVIEW VI.
The NI PXI-7831R plug-in device is designed as a single-point AI, AO, and
DIO complement to the RT Module. Refer to ni.com/labviewrtfor
more information about the RT Module.
Cables and Optional Equipment
NI offers a variety of products to use with your device, including cables,
connector blocks, and other accessories as follows.
© National Instruments Corporation
1-9
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
Table 1-2. Cables and Accessories
Cable Description
Cable
Accessories
SH68-C68-S
Shielded 68-pin VHDCI male
connector to female 0.050 series
D-type connector. The cable is
constructed with 34 twisted wire
pairs plus an overall shield.
Connects to the following standard
68-pin screw terminal blocks:
• SCB-68
• CB-68LP
• CB-68LPR
• TBX-68
NSC68-262650
Non-shielded cable connects from
68-pin VHDCI male connector to
26-pin headers can connect to the
following 5B backplanes for analog
two 26-pin female headers plus one signal conditioning:
50-pin female header. The pinout of
these headers allows for direct
connection to 5B backplanes for
• 5B08 (8-channel)
• 5B01 (16-channel)
analog signal conditioning and SSR
backplanes for digital signal
conditioning.
50-pin header can connect to the
following SSR backplanes for digital
signal conditioning:
• 8-channel backplane
• 16-channel backplane
• 32-channel backplane
NSC68-5050
Non-shielded cable connects from
68-pin VHDCI male connector to
two 50-pin female headers. The
pinout of these headers allows for
direct connection to SSR
50-pin headers can connect to the
following SSR backplanes for digital
signal conditioning:
• 8-channel backplane
• 16-channel backplane
• 32-channel backplane
backplanes for digital signal
conditioning.
Refer to Appendix B, Connecting I/O Signals, for more information on
using these cables and accessories to connect I/O signals to the PXI-7831R.
For the most up-to-date cabling options, refer to ni.com/catalogor call
the sales office nearest to you.
Custom Cabling
NI offers a variety of cables that you can use to connect signals to the
NI PXI-7831R. If you need to develop a custom cable, NI provides a
generic un-terminated shielded cable that makes this task easier. The
NI PXI-7831R User Manual
1-10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
SHC68-NT-S (NI part #189041-02) connects to the NI PXI-7831R VHDCI
connectors on one end of the cable. The other end of the cable is not
terminated. This cable ships with a wire list identifying which wire
corresponds to which NI PXI-7831R pin. Using this cable, you can quickly
connect the NI PXI-7831R signals that you need to the connector of your
choice without having to connect these signals to the VHDCI connector
end of the cable. Refer to Appendix B, Connecting I/O Signals for the
NI PXI-7831R connector pinouts.
Unpacking
The RIO device is shipped in an antistatic package to prevent electrostatic
damage (ESD) to the device. ESD can damage several components on the
device.
Caution Never touch the exposed pins of connectors.
To avoid such damage in handling the device, take the following
precautions:
•
Ground yourself using a grounding strap or by holding a grounded
object.
•
Touch the antistatic package to a metal part of the computer chassis
before removing the device from the package.
Remove the device from the package and inspect the device for loose
components or any sign of damage. Notify NI if the device appears
damaged in any way. Do not install a damaged device into the computer.
Store the RIO device in the antistatic envelope when not in use.
Safety Information
The following section contains important safety information that you must
follow when installing and using the NI PXI-7831R.
Do not operate the NI PXI-7831R in a manner not specified in this
document. Misuse of the NI PXI-7831R can result in a hazard. You can
compromise the safety protection built into the NI PXI-7831R if the
NI PXI-7831R is damaged in any way. If the NI PXI-7831R is damaged,
return it to NI for repair.
© National Instruments Corporation
1-11
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
Do not substitute parts or modify the NI PXI-7831R except as described in
this document. Use the NI PXI-7831R only with the chassis, modules,
accessories, and cables specified in the installation instructions. You must
have all covers and filler panels installed during operation of the
NI PXI-7831R.
Do not operate the NI PXI-7831R in an explosive atmosphere or where
there may be flammable gases or fumes. If you must operate the
NI PXI-7831R in such an environment, it must be in a suitably rated
enclosure.
If you need to clean the NI PXI-7831R, use a soft, nonmetallic brush. Make
sure that the NI PXI-7831R is completely dry and free from contaminants
before returning it to service.
Operate the NI PXI-7831R only at or below Pollution Degree 2. Pollution
is foreign matter in a solid, liquid, or gaseous state that can reduce dielectric
strength or surface resistivity. The following is a description of pollution
degrees:
•
Pollution Degree 1 means no pollution or only dry, nonconductive
pollution occurs. The pollution has no influence.
•
Pollution Degree 2 means that only nonconductive pollution occurs in
most cases. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
•
Pollution Degree 3 means that conductive pollution occurs, or dry,
nonconductive pollution occurs that becomes conductive due to
condensation.
You must insulate signal connections for the maximum voltage for which
the NI PXI-7831R is rated. Do not exceed the maximum ratings for the
NI PXI-7831R. Do not install wiring while the NI PXI-7831R is live with
electrical signals. Do not remove or add connector blocks when power is
connected to the system. Remove power from signal lines before
connecting them to or disconnecting them from the NI PXI-7831R.
Operate the NI PXI-7831R at or below the installation category1 marked
on the hardware label. Measurement circuits are subjected to working
voltages2 and transient stresses (overvoltage) from the circuit to which they
are connected during measurement or test. Installation categories establish
1
Installation categories, also referred to as measurement categories, are defined in electrical safety standard IEC 61010-1.
2
Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.
NI PXI-7831R User Manual
1-12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 1
Introduction
standard impulse withstand voltage levels that commonly occur in
electrical distribution systems. The following is a description of installation
categories:
•
Installation Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS1 voltage. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements
include signal levels, special equipment, limited-energy parts of
equipment, circuits powered by regulated low-voltage sources, and
electronics.
•
•
Installation Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category
refers to local-level electrical distribution, such as that provided by a
standard wall outlet (for example, 115 V for U.S. or 230 V for Europe).
Examples of Installation Category II are measurements performed on
household appliances, portable tools, and similar products.
Installation Category III is for measurements performed in the building
installation at the distribution level. This category refers to
measurements on hard-wired equipment such as equipment in fixed
installations, distribution boards, and circuit breakers. Other examples
are wiring, including cables, bus-bars, junction boxes, switches,
socket-outlets in the fixed installation, and stationary motors with
permanent connections to fixed installations.
•
Installation Category IV is for measurements performed at the primary
electrical supply installation (<1,000V). Examples include electricity
meters and measurements on primary overcurrent protection devices
and on ripple control units.
1
MAINS is defined as a hazardous live electrical supply system that powers equipment. Suitably rated measuring circuits may
be connected to the MAINS for measuring purposes.
© National Instruments Corporation
1-13
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Hardware Overview
of the NI PXI-7831R
This chapter presents an overview of the hardware functions and
I/O connectors on the NI PXI-7831R.
Figure 2-1 shows a block diagram for the NI PXI-7831R, and Figure 2-2
shows the parts locator diagrams for the NI PXI-7831R.
Calibration
DACs
Input Mux
Flash
Memory
Configuration
Control
AI+
AI–
+
16-Bit
ADC
Instrumentation
Amplifier
–
x8 Channels
Input Mode Mux
AISENSE
AIGND
User-
Configurable
FPGA
Temperature
Sensor
Voltage
Reference
Control
Bus
Interface
Data/Address/
Control
Calibration
Mux
on RIO
Devices
Address/Data
2
Calibration
DACs
16-Bit
DAC
PXI Local Bus
RTSI Bus
x8 Channels
Digital I/O (16)
Digital I/O (40)
Digital I/O (40)
Figure 2-1. NI PXI-7831R Block Diagram
© National Instruments Corporation
2-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
SW1
Figure 2-2. Parts Locator Diagram for the NI PXI-7831R
Analog Input
The NI PXI-7831R has eight independent, 16-bit AI channels that can be
simultaneously sampled or sampled at different rates. The input mode is
software configurable, and the input range is fixed at 10 V. The converters
return data in two’s complement format. Table 2-1 shows the ideal output
code returned for a given AI voltage.
Table 2-1. Ideal Output Code and AI Voltage Mapping
Output Code (Hex)
Input Description
Full-scale range –2 LSB
AI Voltage
9.999695
9.999390
(Two’s Complement)
7FFF
7FFE
NI PXI-7831R User Manual
2-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Table 2-1. Ideal Output Code and AI Voltage Mapping (Continued)
Output Code (Hex)
Input Description
AI Voltage
(Two’s Complement)
Midscale
0.000000
0000
8001
8000
—
Negative full-scale range +1 LSB
Negative full-scale range
Any input voltage
–9.999695
–10.000000
Output Code
---------------------------------
× 10.0 V
32,768
Input Modes
The NI PXI-7831R input mode is software configurable. The input
channels support three input modes—differential (DIFF) input, referenced
single-ended (RSE) input, and nonreferenced single-ended (NRSE) input.
The selected input mode applies to all the input channels. Table 2-2
describes the three input modes.
Table 2-2. Available Input Modes for the NI PXI-7831R
Input Mode
Description
DIFF
When the NI PXI-7831R is configured in DIFF input mode, each channel uses
two AI lines. The positive input pin connects to the positive terminal of the
onboard instrumentation amplifier, and the negative input pin connects to the
negative input of the instrumentation amplifier.
RSE
When the NI PXI-7831R is configured in RSE input mode, each channel uses
only its positive AI pin. This pin connects to the positive terminal of the onboard
instrumentation amplifier. The negative input of the instrumentation amplifier is
internally tied to the AI ground (AIGND).
NRSE
When the NI PXI-7831R is configured in NRSE input mode, each channel uses
only its positive AI pin. This pin connects to the positive terminal of the onboard
instrumentation amplifier. The negative input of the instrumentation amplifier on
each AI channel is internally connected to the AI sense (AISENSE) input pin.
Input Range
The NI PXI-7831R AI range is fixed at 10 V.
© National Instruments Corporation
2-3
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Connecting Analog Input Signals
The AI signals for the NI PXI-7831R are AI<0..7>+, AI<0..7>–, AIGND,
and AISENSE. The AI<0..7>+ and AI<0..7>– signals are tied to the eight
AI channels of the NI PXI-7831R. For all input modes, the AI<0..7>+
signals are connected to the positive input of the instrumentation amplifier
on each channel. The signal connected to the negative input of the
instrumentation amplifier depends on the input mode for which the
NI PXI-7831R is configured.
In differential input mode, signals connected to AI<0..7>– are routed to the
negative input of the instrumentation amplifier for each channel. In RSE
input mode, the negative input of the instrumentation amplifier for each
channel is internally connected to AIGND. In NRSE input mode, the
AISENSE signal is connected internally to the negative input of the
instrumentation amplifier for each channel. In DIFF and RSE input modes,
AISENSE is not used and can be left unconnected.
Caution Exceeding the differential and common-mode input ranges distorts the input
signals. Exceeding the maximum input voltage rating can damage the NI PXI-7831R and
the computer. NI is not liable for any damage resulting from such signal connections. The
maximum input voltage ratings are listed in Table B-2, NI PXI-7831R I/O Signal
Summary.
AIGND is a common AI signal that is routed directly to the ground tie point
tie point to the NI PXI-7831R, if necessary.
Connection of AI signals to the NI PXI-7831R depends on the input mode
of the AI channels you are using and the type of input signal source. With
different input modes, you can use the instrumentation amplifier in
different ways. Figure 2-3 shows a diagram of the NI PXI-7831R
instrumentation amplifier.
NI PXI-7831R User Manual
2-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Vin+
+
–
Instrumentation
Amplifier
+
Vm
Measured
Voltage
Vin–
–
Vm = [Vin+ – Vin–]
Figure 2-3. NI PXI-7831R Instrumentation Amplifier
The instrumentation amplifier applies common-mode voltage rejection
and presents high input impedance to the AI signals connected to the
NI PXI-7831R. Signals are routed to the positive and negative inputs of
the instrumentation amplifier through input multiplexers on the device.
The instrumentation amplifier converts two input signals to a signal that is
the difference between the two input signals. The amplifier output voltage
is referenced to the device ground. The NI PXI-7831R ADC measures this
You must reference all signals to ground either at the source device or at the
NI PXI-7831R. If you have a floating source, you should reference the
signal to ground by using RSE input mode or the DIFF input mode with
bias resistors. Refer to the Differential Connections for Nonreferenced or
Floating Signal Sources section for more information about these input
modes. If you have a grounded source, you should not reference the signal
to AIGND. You can avoid this reference by using DIFF or NRSE input
modes.
Types of Signal Sources
When configuring the input channels and making signal connections,
you must first determine whether the signal sources are floating or ground
referenced. The following sections describe these two signal types.
© National Instruments Corporation
2-5
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Floating Signal Sources
A floating signal source is in no way connected to the building ground
system but instead has an isolated ground-reference point. Some examples
of floating signal sources are outputs of transformers, thermocouples,
battery-powered devices, optical isolator outputs, and isolation amplifiers.
An instrument or device that has an isolated output is a floating signal
source. You must tie the ground reference of a floating signal to the
NI PXI-7831R AIGND through a bias resistor to establish a local or
onboard reference for the signal. Otherwise, the measured input signal
varies as the source floats out of the common-mode input range.
Ground-Referenced Signal Sources
A ground-referenced signal source is connected in some way to the
building system ground and is, therefore, already connected to a common
ground point with respect to the NI PXI-7831R, assuming that the
computer is plugged into the same power system. Nonisolated outputs of
instruments and devices that plug into the building power system fall into
this category.
The difference in ground potential between two instruments connected to
the same building power system is typically between 1 and 100 mV but can
be much higher if power distribution circuits are improperly connected. If a
grounded signal source is improperly measured, this difference may appear
as a measurement error. The connection instructions for grounded signal
sources are designed to eliminate this ground potential difference from the
measured signal.
Input Modes
You can configure the NI PXI-7831R for one of three input modes—DIFF,
RSE, or NRSE. The following sections discuss the use of single-ended and
differential measurements and considerations for measuring both floating
and ground-referenced signal sources.
Figure 2-4 summarizes the recommended input mode for both types of
signal sources.
NI PXI-7831R User Manual
2-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Signal Source Type
Floating Signal Source
Grounded Signal Source
(Not Connected to Building Ground)
Examples
Examples
• Ungrounded Thermocouples
• Signal Conditioning with
Isolated Outputs
• Plug-in Instruments with
Nonisolated Outputs
Input
• Battery Devices
AI<i>(+)
AI<i>(+)
+
+
+
–
+
–
V1
V1
AI<i>(–)
AI<i>(–)
–
–
Differential
(DIFF)
AIGND<i>
AIGND<i>
See text for information on bias resistors.
NOT RECOMMENDED
AI<i>
AI
+
+
+
–
+
–
V1
V1
AIGND<i>
–
–
Single-Ended —
Ground
+
V
–
g
Referenced
(RSE)
AIGND
Ground-loop losses, Vg, are added to
measured signal.
AI<i>
AI<i>
+
+
+
–
+
–
V1
V1
AISENSE
AISENSE
–
–
Single-Ended —
Nonreferenced
(NRSE)
AIGND<i>
AIGND<i>
See text for information on bias resistors.
Figure 2-4. Summary of Analog Input Connections
© National Instruments Corporation
2-7
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Differential Connection Considerations (DIFF Input Mode)
In DIFF input mode, the NI PXI-7831R measures the difference between
the positive and negative inputs. DIFF input mode is ideal for measuring
ground-referenced signals from other devices. When using DIFF input
mode, the input signal is tied to the positive input of the instrumentation
amplifier, and its reference signal, or return, is tied to the negative input of
the instrumentation amplifier.
Use differential input connections for any channel that meets any of the
following conditions:
•
•
The input signal is low-level (less than 1 V).
The leads connecting the signal to the NI PXI-7831R are greater than
3 m (10 ft).
•
•
The input signal requires a separate ground-reference point or return
signal.
The signal leads travel through noisy environments.
Differential signal connections reduce noise pickup and increase
common-mode noise rejection. Differential signal connections also allow
instrumentation amplifier.
Differential Connections for Ground-Referenced
Signal Sources
Figure 2-5 shows how to connect a ground-referenced signal source to a
channel on the NI PXI-7831R configured in DIFF input mode.
NI PXI-7831R User Manual
2-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
AI+
AI–
+
–
Ground-
Referenced
Signal
+
–
Instrumentation
Vs
Amplifier
+
–
Source
Measured
Voltage
Vm
Common-
Mode
Noise and
Ground
+
–
Vcm
x8 Channels
AISENSE
AIGND
Potential
I/O Connector
DIFF Input Mode Selected
Figure 2-5. Differential Input Connections for Ground-Referenced Signals
With this connection type, the instrumentation amplifier rejects both the
between the signal source and the NI PXI-7831R ground, shown as Vcm
in Figure 2-5. In addition, the instrumentation amplifier can reject
common-mode noise pickup in the leads connecting the signal sources to
the device. The instrumentation amplifier can reject common-mode signals
as long as V+in and V–in (input signals) are both within their specified input
input ranges.
Differential Connections for Nonreferenced or
Floating Signal Sources
Figure 2-6 shows how to connect a floating signal source to a channel on
the NI PXI-7831R configured in DIFF input mode.
© National Instruments Corporation
2-9
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
AI+
+
Bias
+
–
Resistors
(see text)
AI–
Floating
Signal
Source
Instrumentation
Amplifier
Vs
+
–
Measured
Voltage
Vm
–
Bias
Current
Return
Paths
x8 Channels
AISENSE
AIGND
I/O Connector
DIFF Input Mode Selected
Figure 2-6. Differential Input Connections for Nonreferenced Signals
Figure 2-6 shows two bias resistors connected in parallel with the signal
leads of a floating signal source. If you do not use the resistors and the
source is truly floating, the source is not likely to remain within the
common-mode signal range of the instrumentation amplifier, and the
instrumentation amplifier will saturate, causing erroneous readings. You
must reference the source to AIGND, which you can do by connecting the
positive side of the signal to the positive input of the instrumentation
amplifier and connecting the negative side of the signal to AIGND and to
the negative input of the instrumentation amplifier, without any resistors at
all. This connection works well for DC-coupled sources with low source
impedance (less than 100 Ω).
However, for larger source impedances, this connection leaves the
differential signal path significantly out of balance. Noise that couples
electrostatically onto the positive line does not couple onto the negative
line because it is connected to ground. Hence, this noise appears as a
differential-mode signal instead of a common-mode signal, and the
instrumentation amplifier does not reject it. In this case, instead of directly
connecting the negative line to AIGND, connect it to AIGND through a
resistor that is about 100 times the equivalent source impedance. The
resistor puts the signal path nearly in balance, so about the same amount
of noise couples onto both connections, which yields better rejection of
electrostatically coupled noise. Also, this input mode does not load down
NI PXI-7831R User Manual
2-10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
the source, other than the very high-input impedance of the instrumentation
amplifier.
You can fully balance the signal path by connecting another resistor of the
same value between the positive input and AIGND, as shown in Figure 2-6.
This fully balanced input mode offers slightly better noise rejection but has
the disadvantage of loading the source down with the series combination
(sum) of the two resistors. If, for example, the source impedance is 2 kΩ
and each of the two resistors is 100 kΩ, the resistors load down the source
with 200 kΩ and produce a –1% gain error.
Both inputs of the instrumentation amplifier require a DC path to ground in
order for the instrumentation amplifier to work. If the source is AC coupled
(capacitively coupled), the instrumentation amplifier needs a resistor
between the positive input and AIGND. If the source has low-impedance,
choose a resistor that is large enough not to significantly load the source but
small enough not to produce significant input offset voltage as a result of
input bias current (typically 100 kΩ to 1 MΩ). In this case, you can tie the
negative input directly to AIGND. If the source has high output impedance,
you should balance the signal path as previously described using the same
value resistor on both the positive and negative inputs; you should be aware
that there is some gain error from loading down the source.
Single-Ended Connection Considerations
A single-ended connection is one in which the NI PXI-7831R AI signal is
referenced to a ground that can be shared with other input signals. The input
signal is tied to the positive input of the instrumentation amplifier, and the
ground is tied to the negative input of the instrumentation amplifier.
You can use single-ended input connections for any input signal that meets
the following conditions:
•
•
The input signal is high-level (>1 V).
The leads connecting the signal to the NI PXI-7831R are less than
3 m (10 ft).
•
The input signal can share a common reference point with other
signals.
DIFF input connections are recommended for greater signal integrity for
any input signal that does not meet the preceding conditions.
You can configure in software the NI PXI-7831R channels for two different
types of single-ended connections—RSE input mode and NRSE input
mode. The RSE input mode is used for floating signal sources; in this case,
© National Instruments Corporation
2-11
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
the NI PXI-7831R provides the reference ground point for the external
signal. The NRSE input mode is used for ground-referenced signal sources;
in this case, the external signal supplies its own reference ground point and
the NI PXI-7831R should not supply one.
In single-ended input modes, more electrostatic and magnetic noise couples
into the signal connections than in differential input modes. The coupling
is the result of differences in the signal path. Magnetic coupling
is proportional to the area between the two signal conductors. Electrical
two conductors.
Single-Ended Connections for Floating Signal
Sources (RSE Input Mode)
Figure 2-7 shows how to connect a floating signal source to a channel on
the NI PXI-7831R configured for RSE input mode.
AI+
AI–
+
Instrumentation
Amplifier
+
–
Measured
Voltage
–
Vm
+
–
Floating
Signal
Source
Vs
x8 Channels
AISENSE
AIGND
I/O Connector
RSE Input Mode Selected
Figure 2-7. Single-Ended Input Connections for Nonreferenced or Floating Signals
Single-Ended Connections for Grounded Signal
Sources (NRSE Input Mode)
To measure a grounded signal source with a single-ended input mode, you
must configure the NI PXI-7831R in the NRSE input mode. The signal is
then connected to the positive input of the NI PXI-7831R instrumentation
NI PXI-7831R User Manual
2-12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
amplifier, and the signal local ground reference is connected to the negative
input of the instrumentation amplifier. The ground point of the signal
should, therefore, be connected to AISENSE. Any potential difference
between the NI PXI-7831R ground and the signal ground appears as a
instrumentation amplifier, and this difference is rejected by the amplifier.
If the input circuitry of a NI PXI-7831R were referenced to ground, in this
situation as in RSE input mode, this difference in ground potentials would
appear as an error in the measured voltage.
Figure 2-8 shows how to connect a grounded signal source to a channel on
the NI PXI-7831R configured for NRSE input mode.
AI+
AI–
+
Ground-
Referenced
Signal
+
–
Instrumentation
Amplifier
Vs
+
–
Source
Measured
Voltage
–
Vm
Common-
Mode
Noise and
Ground
+
–
x8 Channels
Vcm
AISENSE
AIGND
Potential
I/O Connector
NRSE Input Mode Selected
Figure 2-8. Single-Ended Input Connections for Ground-Referenced Signals
Common-Mode Signal Rejection Considerations
Figures 2-5 and 2-8 show connections for signal sources that are already
referenced to some ground point with respect to the NI PXI-7831R.
In these cases, the instrumentation amplifier can reject any voltage caused
by ground potential differences between the signal source and the device.
In addition, with differential input connections, the instrumentation
amplifier can reject common-mode noise pickup in the leads connecting the
signal sources to the device. The instrumentation amplifier can reject
© National Instruments Corporation
2-13
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
common-mode signals as long as V+in and V–in (input signals) are both
within their specified input ranges. Refer to Appendix A, Specifications,
for more information about input ranges.
Analog Output
is fixed at 10 V. Some applications require that the AO channels power-on
to known voltage levels. To set the power-on levels, you can configure the
NI PXI-7831R to automatically load and run your VI when the system
powers on. This VI can then set the AO channels to the desired voltage
levels. Data written to the DAC is interpreted in two’s complement format.
Table 2-3 shows the ideal AO voltage generated for a given input code.
Table 2-3. Ideal Output Voltage and Input Code Mapping
Input Code (Hex)
Output Description
Full-scale range –1 LSB
Full-scale range –2 LSB
Midscale
AO Voltage
9.999695
9.999390
0.000000
–9.999695
(Two’s Complement)
7FFF
7FFE
0000
8001
Negative full-scale range,
+1 LSB
Negative full-scale range
Any output voltage
–10.000000
—
8000
AO Voltage
------------------------------
× 32,768
10.0 V
Note If the output value for an AO channel is not specifically set by your VI then the AO
channel voltage output will be undefined.
Connecting Analog Output Signals
The AO signals are AO<0..7> and AOGND.
AO<0..7> are the eight available AO channels. AOGND is the ground
reference signal for the AO channels.
Figure 2-9 shows how to make AO connections to the NI PXI-7831R.
NI PXI-7831R User Manual
2-14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
AO0
Channel 0
+
–
Load
VOUT 0
AOGND0
x8 Channels
NI PXI-7831R
Figure 2-9. Analog Output Connections
Digital I/O
The NI PXI-7831R has 96 bidirectional DIO lines that can be individually
configured for either input or output. When the system powers on, the DIO
lines are all high-impedance. To set another power-on state, you can
configure the NI PXI-7831R to automatically load a VI when the system
powers on. This VI can then set the DIO lines to any desired power-on
state.
Connecting Digital I/O Signals
The DIO signals on the NI PXI-7831R MIO connector are DGND and
port, and DGND is the ground reference signal for the DIO port. The
NI PXI-7831R has one MIO and two DIO connectors for a total of 96 DIO
lines.
Refer to Figure B-1, NI PXI-7831R Connector Locations, and Figure B-2,
NI PXI-7831R I/O Connector Pin Assignments, for the connector locations
and the I/O connector pin assignments on the NI PXI-7831R.
The DIO lines on the NI PXI-7831R are TTL compatible. When configured
as inputs, they can receive signals from 5 V TTL, 3.3 V LVTTL,
5 V CMOS, and 3.3 V LVCMOS devices. When configured as outputs,
they can send signals to 5 V TTL, 3.3 V LVTTL, and 3.3 V LVCMOS
© National Instruments Corporation
2-15
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
devices. Because the NI PXI-7831R digital outputs provide a nominal
output swing of 0 to 3.3 V (3.3 V TTL), the NI PXI-7831R DIO lines
cannot drive 5 V CMOS logic levels. To interface to 5 V CMOS devices,
you must provide an external pull-up resistor to 5 V. This resistor pulls up
the 3.3 V digital output from the NI PXI-7831R to 5 V CMOS logic levels.
For detailed DIO specifications, refer to Appendix A, Specifications.
NI PXI-7831R I/O Signal Summary, can damage the NI PXI-7831R and the computer.
NI is not liable for any damage resulting from such signal connections.
Do not short the DIO lines of the NI PXI-7831R directly to power or to ground. Doing so
can damage the NI PXI-7831R by causing excessive current to flow through the DIO lines.
Refer to Appendix A, Specifications, for more information. NI is not liable for any damage
resulting from such signal connections.
If required by your application, you can connect multiple NI PXI-7831R digital output
lines in parallel to provide higher current sourcing or sinking capability. If you connect
multiple digital output lines in parallel, your application must drive all of these lines
simultaneously to the same value. If you connect digital lines together and drive them to
different values, excessive current may flow through the DIO lines and damage the
NI PXI-7831R. Refer to Appendix A, Specifications, for more information. NI is not liable
for any damage resulting from such signal connections.
NI PXI-7831R User Manual
2-16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Figure 2-10 shows signal connections for three typical DIO applications.
LED
TTL or
LVCMOS
Compatible
Devices
+5 V
DGND
†
*
DIO<4..7>
DIO<0..3>
5 V CMOS
TTL, LVTTL, CMOS, or LVCMOS Signal
+5 V
Switch
DGND
I/O Connector
NI PXI-7831R
*
3.3 V CMOS
†
Use a pull-up resistor when driving 5 V CMOS devices.
Figure 2-10. Example Digital I/O Connections
Figure 2-10 shows DIO<0..3> configured for digital input and DIO<4..7>
configured for digital output. Digital input applications include receiving
TTL, LVTTL, CMOS, or LVCMOS signals and sensing external device
states, such as the state of the switch shown in the figure. Digital output
applications include sending TTL or LVCMOS signals and driving external
devices, such as the LED shown in the figure.
The NI PXI-7831R SH68-C68-S shielded cable contains 34 twisted pairs
of conductors. To maximize the digital I/O available on the NI PXI-7831R,
some of the DIO lines are twisted with power or ground as they are run
through the cable, and some DIO lines are twisted with other DIO lines as
they are run through the cable. To obtain maximum signal integrity, place
© National Instruments Corporation
2-17
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
edge-sensitive or high-frequency digital signals on the DIO lines that are
other DIO lines can couple noise onto each other, these lines should be used
for static signals or for non-edge-sensitive, low-frequency digital signals.
Examples of high-frequency or edge-sensitive signals include clock,
trigger, pulse-width modulation (PWM), encoder, and counter signals.
Examples of static signals or non-edge-sensitive, low-frequency signals
include LEDs, switches, and relays. Table 2-4 summarizes these
guidelines.
Table 2-4. DIO Signal Guidelines for the NI PXI-7831R
SH68-C68-S Shielded Cable
Signal Pairing
Recommended Types
of Digital Signals
Digital Lines
Connector 0, DIO<0..7>;
Connector 1, DIO<0..27>;
Connector 2, DIO<0..27>
DIO line paired with power
or ground
All types (high frequency or low
frequency signals,
edge-sensitive or
non-edge-sensitive signals)
Connector 0, DIO<8..15>;
Connector 1, DIO<28..39>;
Connector 2, DIO<28..39>
DIO line paired with another
DIO line
Static signals or
non-edge-sensitive,
low-frequency signals
PXI Trigger Bus
The NI PXI-7831R can send and receive triggers through the PXI trigger
bus, which provides eight trigger lines that link all PXI slots in a bus
segment. These trigger lines connect to the FPGA on the NI PXI-7831R
and can be used just like any of the other NI PXI-7831R DIO lines.
The PXI trigger lines can be used to synchronize an NI PXI-7831R to any
other device that supports PXI triggers. The PXI trigger lines on the
NI PXI-7831R are PXI/TRIG<0..7>. In addition, the NI PXI-7831R can
use the PXI star trigger line to send or receive triggers from a device
plugged into slot 2 of the PXI chassis. The PXI star trigger line on the
NI PXI-7831R is PXI/STAR.
The PXI-7831R can configure each PXI trigger line either as an input or an
output signal. Since each PXI trigger line in the PXI trigger bus is
connected in parallel to all the PXI slots in a bus segment, only one PXI
device can drive a particular PXI trigger line at a time. For example, if one
NI PXI-7831R is configured to send out a trigger pulse on PXI/TRIG<0>,
the remaining devices on that PXI bus segment must have PXI/TRIG<0>
configured as an input.
NI PXI-7831R User Manual
2-18
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
Caution Do not drive the same PXI trigger bus line on the same PXI bus segment with the
NI PXI-7831R and another device simultaneously. Such signal driving can damage both
devices. NI is not liable for any damage resulting from such signal driving.
Refer to the PXI Hardware Specification Revision 2.1 and PXI Software
PXI triggers.
PXI Local Bus
The NI PXI-7831R can communicate with other PXI devices using the PXI
local bus. The PXI local bus is a daisy-chained bus that connects each PXI
peripheral slot with its adjacent peripheral slot on either side. For example,
the right local bus lines from a given PXI peripheral slot connect to the left
local bus lines of the adjacent slot. Each local bus is 13 lines wide. All of
these lines connect to the FPGA on the NI PXI-7831R and can be used like
any of the other NI PXI-7831R DIO lines. The PXI local bus right lines on
the NI PXI-7831R are PXI/LBR<0..12>. The PXI local bus left lines on the
NI PXI-7831R are PXI/LBLSTAR<0..12>.
The NI PXI-7831R can configure each PXI local bus line either as an input
or an output signal. Only one device can drive the same physical local bus
line at a given time. For example, if an NI PXI-7831R is configured to drive
a signal on PXI/LBR<0>, the device in the slot immediately to the right
must have its PXI/LBLSTAR<0> line configured as an input.
Caution Do not drive the same PXI local bus line with the NI PXI-7831R and another
device simultaneously. Such signal driving can damage both devices. NI is not liable for
any damage resulting from such signal driving.
The NI PXI-7831R local bus lines are only compatible with 3.3 V signaling
LVTTL and LVCMOS levels.
Caution Do not enable the local bus lines on an adjacent device if the device drives
anything other than 0–3.3V LVTTL signal levels on the NI PXI-7831R. Enabling the lines
in this way can damage the NI PXI-7831R. NI is not liable for any damage resulting from
enabling such lines.
The left local bus lines from the left peripheral slot of a PXI backplane
(slot 2) are routed to the star trigger lines of up to 13 other peripheral slots
in a two-segment PXI system. This configuration provides a dedicated,
delay-matched trigger signal between the first peripheral slot and the
© National Instruments Corporation
2-19
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
other peripheral slots and results in very precise trigger timing signals.
For example, an NI PXI-7831R in slot 2 can send out an independent
trigger signal to each device plugged into slots <3..15> using the
PXI/LBLSTAR<0..12>. Each device receives its trigger signal on its own
dedicated star trigger line.
Caution Do not configure the NI PXI-7831R and another device to drive the same physical
star trigger line simultaneously. Such signal driving can damage the NI PXI-7831R and the
other device. NI is not liable for any damage resulting from such signal driving.
Refer to the PXI Hardware Specification Revision 2.1 and PXI Software
PXI triggers.
Switch Settings
Refer to Figure 2-2 for the location of switch SW1. For normal operation,
switch 1 is in the OFF position. To prevent a VI stored in flash memory
from loading to the FPGA upon power up, you can move switch 1 to the
ON position, as shown in Figure 2-11.
ON
ON
1 2 3
1 2 3
a. Normal Operation (Default)
b. Prevent VI From Loading
Figure 2-11. Switch Settings on Switch SW1
To move switch 1 to the ON position, complete the following steps:
1. Power off and unplug the PXI/CompactPCI chassis.
2. Remove the NI PXI-7831R.
3. Move switch 1 to the ON position, as shown in Figure 2-11.
4. Refer to the Installing the Hardware section of the Where to Start with
the NI PXI-7831R document for installation instructions for
reinserting the NI PXI-7831R into the PXI/CompactPCI chassis.
5. Plug in and power on the PXI/CompactPCI chassis.
After completing this procedure, a VI stored in flash memory does not load
to the FPGA on power up. You can use software to reconfigure the
NI PXI-7831R if necessary. To return to the default mode of loading from
NI PXI-7831R User Manual
2-20
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
flash memory, repeat the procedure above but return switch 1 to the OFF
position in step 3.
Note When the NI PXI-7831R is powered on with switch 1 in the ON position, the analog
circuitry does not return properly calibrated data. For this reason, the switch should only
be switched to the ON position while you are using software to reconfigure the
NI PXI-7831R for the desired power-up behavior. Afterwards, you should return switch 1
to the OFF position.
Power Connections
Two pins on each I/O connector supply +5 V from the computer power
supply using a self-resetting fuse. The fuse resets automatically within a
few seconds after the overcurrent condition is removed. The +5 V pins are
referenced to DGND and can be used to power external digital circuitry.
Power rating ........................................... +4.65 to +5.25 VDC at 1 A
(250 mA max per 5 V pin,
1 A max total for all +5 V lines
on the device)
Caution Do not connect the +5 V power pins directly to analog or digital ground or to any
other voltage source on the NI PXI-7831R or any other device under any circumstance.
Doing so can damage the NI PXI-7831R and the computer. NI is not liable for damage
resulting from such a connection.
Field Wiring Considerations
Environmental noise can seriously affect the accuracy of measurements
made with the NI PXI-7831R if you do not take proper care when running
signal wires between signal sources and the device. The following
recommendations mainly apply to AI signal routing to the device, although
they also apply to signal routing in general.
Minimize noise pickup and maximize measurement accuracy by taking the
following precautions:
•
•
Use differential AI connections to reject common-mode noise.
Use individually shielded, twisted-pair wires to connect AI signals to
the device. With this type of wire, the signals attached to the AI+ and
AI– inputs are twisted together and then covered with a shield.
You then connect this shield only at one point to the signal source
ground. This kind of connection is required for signals traveling
© National Instruments Corporation
2-21
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 2
Hardware Overview of the NI PXI-7831R
through areas with large magnetic fields or high electromagnetic
interference.
•
Route signals to the device carefully. Keep cabling away from noise
sources. The most common noise source in a PXI DAQ system is the
video monitor. Separate the monitor from the analog signals as much
as possible.
The following recommendations apply for all signal connections to the
NI PXI-7831R:
•
Separate NI PXI-7831R signal lines from high-current or high-voltage
lines. These lines can induce currents in or voltages on the
NI PXI-7831R signal lines if they run in parallel paths at a close
distance. To reduce the magnetic coupling between lines, separate
them by a reasonable distance if they run in parallel, or run the lines at
right angles to each other.
•
•
Do not run signal lines through conduits that also contain power lines.
Protect signal lines from magnetic fields caused by electric motors,
welding equipment, breakers, or transformers by running them through
special metal conduits.
Refer to the NI Developer Zone tutorial, Field Wiring and Noise
Considerations for Analog Signals, at ni.com/zonefor more information.
NI PXI-7831R User Manual
2-22
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Calibration
Calibration refers to the process of minimizing measurement and output
voltage errors. On the NI PXI-7831R, these errors are corrected in the
analog circuitry by onboard calibration DACs (CalDACs). Because
calibration is handled by the analog circuitry, the data read from the AI
channels or written to the AO channels in the FPGA VI is already
calibrated.
Three levels of calibration are available for the NI PXI-7831R to ensure the
accuracy of its analog circuitry. The first level, loading calibration
constants, is the fastest, easiest, and least accurate. The intermediate level,
internal calibration, is the preferred method of assuring accuracy in your
application. The last level, external calibration, is the slowest, most
difficult, and most accurate.
Loading Calibration Constants
The NI PXI-7831R is factory calibrated before shipment at approximately
25 °C to the levels indicated in Appendix A, Specifications. The associated
calibration constants (the values that were written to the CalDACs to
achieve calibration in the factory) are stored in the onboard nonvolatile
flash memory. These constants are automatically read from the flash
memory and loaded into the CalDACs by the NI PXI-7831R hardware on
power-up. This occurs before a VI is loaded into the FPGA.
Internal Calibration
The NI PXI-7831R can measure and correct for almost all of its
calibration-related errors without any external signal connections. This
calibration method is referred to as internal calibration. NI provides
software to perform an internal calibration. This internal calibration
process, which generally takes less than two minutes, is the preferred
method of assuring accuracy in your application. Initiate an internal
calibration to minimize the effects of any offset and gain drifts, particularly
those due to changes in temperature. During the internal calibration
process, the AI and AO channels are compared to the NI PXI-7831R
© National Instruments Corporation
3-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Chapter 3
Calibration
onboard voltage reference. The offset and gain errors in the analog circuitry
are calibrated out by adjusting the CalDACs to minimize these errors.
Immediately after internal calibration, the only significant residual
calibration error should be gain error due to time and temperature drift of
the onboard voltage reference. This error is addressed by external
calibration, which is discussed in the External Calibration section. If you
are interested primarily in relative measurements, you can ignore a small
amount of gain error, and self-calibration should be sufficient.
The results of an internal calibration can be stored in the flash memory on
the NI PXI-7831R so that the CalDACs are automatically loaded with the
newly calculated calibration constants the next time the NI PXI-7831R is
powered on.
External Calibration
The NI PXI-7831R has an onboard calibration reference to ensure the
accuracy of self-calibration. Its specifications are listed in Appendix A,
Specifications. The reference voltage is measured at the factory and stored
in the flash memory for subsequent internal calibrations. This voltage is
stable enough for most applications, but if you are using your device at an
extreme temperature or if the onboard reference has not been measured for
a year or more, you may want to externally calibrate your device.
An external calibration refers to calibrating your device with a known
external reference rather than relying on the onboard reference. During the
external calibration process, the onboard reference value is re-calculated.
This compensates for any time or temperature drift related errors in the
onboard reference, which may have resulted since the last calibration. You
can save the results of the external calibration process to flash memory so
that the new calibration constants are automatically loaded the next time the
NI PXI-7831R is powered on and so that the newly measured onboard
reference level is used for subsequent internal calibrations.
To externally calibrate your device, be sure to use a very accurate external
reference. The reference should be several times more accurate than the
device itself.
For a detailed calibration procedure for the NI PXI-7831R, refer to the
Procedures at ni.com/calibration.
NI PXI-7831R User Manual
3-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
A
Specifications
This appendix lists the specifications of the NI PXI-7831R. These
specifications are typical at 25 °C unless otherwise noted.
Analog Input
Input Characteristics
Number of channels ............................... 8
Input modes............................................ DIFF, RSE, NRSE
(software-selectable; selection
applies to all 8 channels)
Type of ADC.......................................... Successive approximation
Resolution .............................................. 16 bits, 1 in 65,536
Conversion time ..................................... 4 µs
Maximum sampling rate ........................ 200 kS/s (per channel)
Input impedance
Powered on ..................................... 10 GΩ in parallel with 100 pF
Powered off..................................... 4 kΩ min
Overload.......................................... 4 kΩ min
Input signal range................................... 10 V
Input bias current ................................... 2 nA
Input offset current................................. 1 nA
Input coupling ........................................ DC
Maximum working voltage
(signal + common mode) ....................... Inputs should remain
within 12 V of ground
© National Instruments Corporation
A-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Overvoltage protection ........................... 42 V
Data transfers..........................................Interrupts, programmed I/O
Accuracy Information
Relative
Absolute Accuracy
Accuracy
Noise +
Quantization
(µV)
Absolute
Accuracy
at Full
Scale
( mV)
Nominal Range (V)
Positive Negative
% of Reading
24
Resolution (µV)
Temp
Drift
Full
Full
Offset Single
Single
Pt.
Scale
Scale
Hours
1 Year
(µV)
Pt.
Averaged (%/ °C)
Averaged
10.0
–10.0
0.0496 0.0507
2542
1779
165
0.0005
7.78
2170
217
Note: Accuracies are valid for measurements following an internal calibration. Measurement accuracies are listed for
operational temperatures within 1 °C of internal calibration temperature and 10 °C of external or factory-calibration
temperature. Temp drift applies only if ambient is greater than 10 °C of previous external calibration.
DC Transfer Characteristics
INL.......................................................... 3 LSB typ, 6 LSB max
DNL........................................................–1.0 to +2.0 LSB max
No missing codes resolution...................16 bits typ, 15 bits min
CMRR, DC to 60 Hz ..............................86 dB
Dynamic Characteristics
Bandwidth
Small signal (–3 dB)........................820 kHz
Large signal (1% THD)...................55 kHz
System noise...........................................1.8 LSBrms
(including quantization)
NI PXI-7831R User Manual
A-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Settling time
Accuracy
Step Size
20.0 V
2.0 V
16 LSB
7.5 µs
2.7 µs
1.7 µs
4 LSB
10.3 µs
4.1 µs
2.9 µs
2 LSB
40 µs
5.1 µs
3.6 µs
0.2 V
Crosstalk................................................. –80 dB, DC to 100 kHz
Analog Output
Output Characteristics
Number of channels ............................... 8 single-ended, voltage output
Resolution .............................................. 16 bits, 1 in 65,536
Update time............................................ 1.0 µs
Max update rate...................................... 1 MS/s
Type of DAC.......................................... Enhanced R-2R
Data transfers ......................................... Interrupts, programmed I/O
Accuracy Information
Absolute Accuracy
Absolute
Accuracy at
Nominal Range (V)
% of Reading
Positive Full
Scale
Negative Full
Scale
Temp Drift
(%/ °C)
Full Scale
(mV)
24 Hours
1 Year
Offset (µV)
10.0
–10.0
0.0335
0.0351
2366
0.0005
5.88
Note: Accuracies are valid for analog output following an internal calibration. Analog output accuracies are listed for
operation temperatures within 1 °C of internal calibration temperature and 10 °C of external or factory calibration
temperature. Temp Drift applies only if ambient is greater than 10 °C of previous external calibration.
© National Instruments Corporation
A-3
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
DC Transfer Characteristics
INL.......................................................... 0.5 LSB typ, 4.0 LSB max
DNL........................................................ 0.5 LSB typ, 1 LSB max
Monotonicity ..........................................16 bits, guaranteed
Voltage Output
Range...................................................... 10 V
Output coupling ......................................DC
Output impedance...................................1.25 Ω max
Current drive........................................... 5 mA
Protection ...............................................Short-circuit to ground
Power-on state ........................................User configurable
Dynamic Characteristics
Settling time
Accuracy
Step Size
20.0 V
2.0 V
16 LSB
6.0 µs
2.2 µs
1.5 µs
4 LSB
6.2 µs
2.9 µs
2.6 µs
2 LSB
7.2 µs
3.8 µs
3.6 µs
0.2 V
Slew rate .................................................10 V/µs
Noise.......................................................150 µVrms, DC to 1 MHz
Glitch energy
at midscale transition.............................. 100 mV for 3 µs
NI PXI-7831R User Manual
A-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Digital I/O
Number of channels
NI PXI-7831R................................. 96 input/output
Compatibility ......................................... TTL
Digital logic levels
Level
Input low voltage (VIL)
Min
0.0 V
2.0 V
—
Max
0.8 V
5.5 V
0.4 V
Input high voltage (VIH)
Output low voltage (VOL),
where IOUT = –Imax (sink)
Output high voltage (VOH),
where IOUT = Imax (source)
2.4 V
—
Maximum output current
Driver Type (Software Selectable)
Imax (Source)
5.4 mA
Imax (Sink)
5.0 mA
Default
Slow
Fast
1.9 mA
1.9 mA
16 mA
14 mA
Power-on state........................................ Programmable, by line
Data transfers ......................................... Interrupts, programmed I/O
Protection
Input................................................ –0.5 to 7.0 V
Output ............................................. Short-circuit (up to eight lines
may be shorted at a time)
Reconfigurable FPGA
Number of logic slices ........................... 5, 120
Equivalent number of logic cells .... 11, 520
Available embedded RAM..................... 16, 384 KB
Timebase ................................................ 40 MHz
© National Instruments Corporation
A-5
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Timebase accuracy
With onboard base clock ................. 100 ppm
Phase locked
to PXI 10 MHz clock....................... 350 ps jitter, 300 ps skew (max)
Calibration
Recommended warm-up time.................15 minutes
Calibration interval.................................1 year
Onboard calibration reference
DC level...........................................5.000 V ( 3.5 mV)
(actual value stored
in flash memory)
Temperature coefficient................... 5 ppm/°C max
Long-term stability.......................... 20 ppm/ 1,000 h
Note To generate a calibration certificate for the NI PXI-7831R, click On-line
Calibration Certificates at ni.com/calibration.
Bus Interface
PXI..........................................................Master, slave
+5 VDC ( 5%)
Power Requirement
NI PXI-7831R .................................450 mA (typ), 700 mA (max)
(does not include current drawn
from the +5 V line on the
I/O connectors)
+3.3 VDC ( 5%)
NI PXI-7831R .................................335 mA (typ), 730 mA (max)
Power available at I/O connectors..........+4.65 to +5.25 VDC at 1 A total,
250 mA per I/O connector pin
NI PXI-7831R User Manual
A-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Physical
Dimensions
(not including connectors) .................... 16.0 by 10.0 cm (6.3 by 3.9 in.)
I/O connectors
NI PXI-7831R................................. Three 68-pin female high-density
VHDCI type
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth..................................... 12 V, Installation Category I
Channel-to-channel ................................ 24 V, Installation Category I
Environmental
Operating temperature............................ – 40 to 70 °C
Storage temperature ............................... –55 to 85 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude.................................. 2,000 meters
Pollution Degree (indoor use only)........ 2
Safety
The NI PXI-7831R devices meet the requirements of the following
standards for safety and electrical equipment for measurement, control, and
laboratory use:
•
•
•
IEC 61010-1, EN 61010-1
UL 3111-1
CAN/CSA C22.2 No. 1010.1
Note For UL and other safety certifications, refer to the product label or to ni.com.
© National Instruments Corporation
A-7
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A
Specifications
Electromagnetic Compatibility
Emissions................................................EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Immunity ................................................EN 61326-1:1997 + A2:2001,
Table 1
EMC/EMI ...............................................CE, C-Tick, and FCC Part 15
(Class A) Compliant
Note For EMC compliance, you must operate this device with shielded cabling.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
Low-Voltage Directive (safety)..............73/23/EEC
Electromagnetic Compatibility
Directive (EMC).....................................89/336/EEC
Note Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click Declaration
of Conformity Information at ni.com/hardref.nsf/.
NI PXI-7831R User Manual
A-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
B
Connecting I/O Signals
This appendix describes how to make input and output signal connections
The NI PXI-7831R has two DIO connectors with 40 DIO lines per
connector, and one MIO connector with eight AI lines, eight AO lines, and
16 DIO lines.
Figure B-1 shows the I/O connector locations for the NI PXI-7831R.
The I/O connectors are numbered starting at zero. The text in parentheses
indicates whether each I/O connector is an MIO connector or a DIO
connector.
© National Instruments Corporation
B-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
NI PXI-7831R
Reconfigurable I/O
Figure B-1. NI PXI-7831R Connector Locations
Figure B-2 shows the I/O connector pin assignments for the I/O connectors
on the NI PXI-7831R. The DIO connector pin assignment applies to
connectors<1..2> on the NI PXI-7831R. The MIO connector pin
assignment applies to connector 0 on the NI PXI-7831R.
NI PXI-7831R User Manual
B-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
34 68
34 68
DIO38
AI0-
DIO39
DIO37
DIO35
DIO33
DIO31
DIO29
DIO27
DIO26
DIO25
DIO24
DIO23
DIO22
DIO21
DIO20
DIO19
DIO18
DIO17
AI0+
DIO36 33 67
DIO34 32 66
AIGND1 33 67
AI1- 32 66
AIGND0
AI1+
31 65
30 64
31 65
30 64
DIO32
DIO30
AI2-
AI2+
AIGND3
AIGND2
AI3+
DIO28 29 63
+5V 28 62
+5V 27 61
DGND 26 60
AI3- 29 63
AI4- 28 62
AI4+
AIGND5 27 61
AI5- 26 60
AIGND4
AI5+
DGND
DGND 24 58
AI6-
AIGND7 24 58
25 59
25 59
AI6+
AIGND6
AI7+
23 57
22 56
21 55
23 57
22 56
21 55
DGND
DGND
DGND
AI7-
No Connect
AOGND0
AISENSE
AO0
DGND 20 54
19 53
AOGND1 20 54
19 53
AO1
DGND
AOGND2
AO2
DGND 18 52
DGND 17 51
AOGND3 18 52
AOGND4 17 51
AO3
DIO16
AO4
16 50
15 49
16 50
15 49
DGND
DGND
AOGND5
AOGND6
DIO15
DIO14
AO5
AO6
DGND 14 48
DGND 13 47
DGND 12 46
DGND 11 45
DGND 10 44
AOGND7 14 48
DIO14 13 47
DIO12 12 46
DIO10 11 45
DIO8 10 44
DIO13
DIO12
DIO11
DIO10
DIO9
DIO8
DIO7
DIO6
DIO5
DIO4
DIO3
DIO2
DIO1
DIO0
AO7
DIO15
DIO13
DIO11
DIO9
DIO7
DIO6
DIO5
DIO4
DIO3
DIO2
DIO1
DIO0
+5V
DGND
DGND
9
8
7
6
5
4
3
2
1
43
42
41
40
39
38
37
36
35
DGND
DGND
9
8
7
6
5
4
3
2
1
43
42
41
40
39
38
37
36
35
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
+5V
DIO Connector Pin Assignment
MIO Connector Pin Assignment
Figure B-2. NI PXI-7831R I/O Connector Pin Assignments
To access the signals on the I/O connectors, you must connect a cable from
the I/O connector to a signal accessory. Plug the small VHDCI connector
end of the cable into the appropriate I/O connector, and connect the other
end of the cable to the appropriate signal accessory.
© National Instruments Corporation
B-3
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
.
Table B-1. I/O Connector Signal Descriptions
Signal Name
Reference
DGND
Direction
Description
+5V
Output
+5 VDC Source—These pins supply +5 V from the computer
power supply using a self-resetting 1 A fuse. No more than
250 mA should be pulled from a single pin.
AI<0..7>+
AI<0..7>–
AIGND
AIGND
AIGND
—
Input
Input
—
Positive Input for Analog Channels 0 through 7.
Negative Input for Analog Channels 0 through 7.
Analog Input Ground—These pins are the reference point for
single-ended measurements in RSE configuration and the
bias current return point for differential measurements.
All three ground references—AIGND, AOGND, and
DGND—are connected together on the NI PXI-7831R.
AISENSE
AIGND
Input
Analog Input Sense—This pin serves as the reference node
for channels AI<0..7> when the device is configured for
NRSE mode.
AO<0..7>
AOGND
AOGND
—
Output
—
Analog Output Channels 0 through 7. Each channel can
source or sink up to 5 mA.
Analog Output Ground—The analog output voltages
are referenced to this node. All three ground
references—AIGND, AOGND, and DGND—are
connected together on the NI PXI-7831R.
DGND
—
—
Digital Ground—These pins supply the reference for the
digital signals at the I/O connector as well as the +5 V supply.
All three ground references—AIGND, AOGND, and
DGND—are connected together on the NI PXI-7831R.
DIO<0..15>
Connector 0
DGND
Input or
Output
Digital I/O signals.
DIO<0..39>
Connector<1..2>
Caution Connections that exceed any of the maximum ratings of input or output signals
on the NI PXI-7831R can damage the NI PXI-7831R and the computer. Maximum input
ratings for each signal are given in the Protection column of Table B-2. NI is not liable for
any damage resulting from such signal connections
NI PXI-7831R User Manual
B-4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
Table B-2. NI PXI-7831R I/O Signal Summary
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Driver
Type
Source
(mA at V)
Sink
(mA at V)
Rise
Time
Signal Name
+5V
Bias
—
—
—
DO
AI
—
—
—
—
—
—
—
AI<0..7>+
10 GΩ in
parallel
with
42/35
—
2 nA
100 pF
AI<0..7>–
—
AI
10 GΩ in
parallel
with
42/35
—
—
—
2 nA
100 pF
AIGND
—
—
AO
AI
—
—
—
—
—
—
—
—
—
AISENSE
10 GΩ in
parallel
with
42/35
2 nA
100 pF
AO<0..7>
—
AO
1.25 Ω
Short-
circuit to
ground
5 at 10
5 at –10
10 V/µs
—
AOGND
DGND
—
—
AO
DO
—
—
—
—
—
—
—
—
—
—
—
—
—
—
DIO<0..15>
Connector 0
DIO<0..39>
Connector<1..2>
Default
DIO
–0.5
to +7.0
5.4 at 2.4
5.0 at 0.4
12 ns
Slow
Fast
DIO
DIO
—
—
–0.5
to +7.0
1.9 at 0.4
16 at 2.4
1.9 at 0.4
14 at 0.4
75 ns
6 ns
—
—
–0.5
to +7.0
AI = Analog Input
AO = Analog Output
DIO = Digital Input/Output
DO = Digital Output
Connecting to 5B and SSR Signal Conditioning
NI provides cables that allow you to connect signals from the
NI PXI-7831R directly to 5B backplanes for analog signal conditioning
and SSR backplanes for digital signal conditioning.
© National Instruments Corporation
B-5
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
The NSC68-262650 cable is designed to connect the signals on the
NI PXI-7831R MIO connector directly to 5B and SSR backplanes. This
cable has a 68-pin male VHDCI connector on one end that plugs into the
NI PXI-7831R MIO connector. The other end of this cable provides two
26-pin female headers plus one 50-pin female header.
One of the 26-pin headers contains all the NI PXI-7831R analog input
signals. This connector can be plugged directly into a 5B backplane for
analog input signal conditioning. The NI PXI-7831R AI channels <0..7>
are mapped to the 5B backplane channels <0..7> in sequential order. The
AI channels should be configured to use the NRSE input mode when using
5B signal conditioning.
The other 26-pin header contains all the NI PXI-7831R analog output
signals. This connector can be plugged directly into a 5B backplane for AO
signal conditioning. The NI PXI-7831R AO channels <0..7> are mapped to
the 5B backplane channels <0..7> in sequential order.
The 50-pin header contains the 16 DIO lines available on the
NI PXI-7831R MIO connector. This header can be plugged directly into an
SSR backplane for digital signal conditioning. DIO lines <0..15> are
mapped to the 5B backplane slots <0..15> in sequential order.
The 5B connector pinouts are compatible with 8-channel 5B08 backplanes
and 16-channel 5B01 backplanes, but since the NI PXI-7831R only
provides 8 AI channels, you only have access to the first 8 channels in a
16-channel backplane. The SSR connector pinout is compatible with 8, 16,
24, and 32-channel SSR backplanes. You can connect to an SSR backplane
containing a number of channels that does not equal the 16 DIO lines
available on the 50-pin header. In this case, you only have access to the
channels that exist on both the SSR backplane and the NSC68-262650
cable 50-pin header.
Figure B-3 shows the connector pinouts when using the NSC68-262650
cable.
NI PXI-7831R User Manual
B-6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
,
NC
NC
NC
NC
NC
1
3
5
7
9
2
4
6
8
10
NC
NC
NC
NC
NC
NC
NC
NC
11 12
13 14
15 16
17 18
19 20
21 22
23 24
25 26
27 28
29 30
31 32
33 34
35 36
37 38
39 40
41 42
43 44
45 46
47 48
49 50
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DIO15
DIO14
DIO13
DIO12
DIO11
DIO10
DIO9
DIO8
DIO7
DIO6
DIO5
DIO4
DIO3
DIO2
DIO1
DIO0
+5V
AO0
AOGND0
AO1
AO2
AOGND2
AO3
AO4
AOGND4
AO5
AO6
AOGND6
AO7
1
3
5
7
9
2
4
6
8
10
NC
NC
AOGND1
NC
NC
AOGND3
NC
NC
AOGND5
NC
NC
AI0+
AIGND0
AI1+
AI2+
AIGND2
AI3+
AI4+
AIGND4
AI5+
AI6+
AIGND6
AI7+
1
3
5
7
9
2
4
6
8
10
AI0–
AI1–
AIGND1
AI2–
AI3–
AOGND3
AI4–
AI5–
AOGND5
AI6–
AI7–
AOGND7
NC
11 12
13 14
15 16
17 18
19 20
21 22
23 24
25 26
11 12
13 14
15 16
17 18
19 20
21 22
23 24
25 26
AOGND7
NC
NC
AISENSE
AO 0–7 Connector
Pin Assignment
AI 0–7 Connector
Pin Assignment
DIO 0–15 Connector
Pin Assignment
Figure B-3. Connector Pinouts When Using NSC68-262650 Cable
The NSC68-5050 cable is designed to connect the signals on the
NI PXI-7831R DIO connectors directly to SSR backplanes for digital
signal conditioning. This cable has a 68-pin male VHDCI connector on one
end that plugs into the NI PXI-7831R DIO connectors. The other end of
this cable provides two 50-pin female headers.
Each of these 50-pin headers can be plugged directly into an 8-, 16-, 24-, or
32-channel SSR backplane for digital signal conditioning. One of the
50-pin headers contains DIO lines 0–23 from the NI PXI-7831R DIO
connector. These lines are mapped to slots 0–23 on an SSR backplane in
sequential order. The other 50-pin header contains DIO lines 24–39 from
the NI PXI-7831R DIO connector. These lines are mapped to slots 0–15 on
an SSR backplane in sequential order. You can connect to an SSR
backplane containing a number channels that does not equal the number of
© National Instruments Corporation
B-7
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B
Connecting I/O Signals
lines on the NSC68-5050 cable header. In this case, you only have access
to the channels that exist on both the SSR backplane and the NSC68-5050
cable header you are using.
Figure B-4 shows the connector pinouts when using the NSC68-5050
cable.
DIO23
DIO22
DIO21
DIO20
DIO19
DIO18
DIO17
DIO16
DIO15
DIO14
DIO13
DIO12
DIO11
DIO10
DIO9
DIO8
DIO7
DIO6
DIO5
DIO4
DIO3
DIO2
DIO1
1
3
5
7
9
2
4
6
8
10
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
1
3
5
7
9
2
4
6
8
10
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
11 12
13 14
15 16
17 18
19 20
21 22
23 24
25 26
27 28
29 30
31 32
33 34
35 36
37 38
39 40
41 42
43 44
45 46
47 48
49 50
11 12
13 14
15 16
17 18
19 20
21 22
23 24
25 26
27 28
29 30
31 32
33 34
35 36
37 38
39 40
41 42
43 44
45 46
47 48
49 50
NC
DIO39
DIO38
DIO37
DIO36
DIO35
DIO34
DIO33
DIO32
DIO31
DIO30
DIO29
DIO28
DIO27
DIO26
DIO25
DIO24
+5V
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DGND
DIO0
+5V
DIO 0–23 Connector
Pin Assignment
DIO 24–39 Connector
Pin Assignment
Figure B-4. Connector Pinouts When Using the NSC68-5050 Cable
NI PXI-7831R User Manual
B-8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
C
Using the SCB-68
Shielded Connector Block
This appendix describes how to connect input and output signals to the
NI PXI-7831R with the SCB-68 shielded connector block.
The SCB-68 has 68 screw terminals for I/O signal connections. To use the
SCB-68 with the NI PXI-7831R, you must configure the SCB-68 as a
general-purpose connector block. Refer to Figure C-1 for the
general-purpose switch configuration.
S5 S4 S3
S1
S2
Figure C-1. General-Purpose Switch Configuration for the SCB-68 Terminal Block
After configuring the SCB-68 switches, you can connect the I/O signals to
the SCB-68 screw terminals. Refer to Appendix B, Connecting I/O Signals,
for the connector pin assignments for the NI PXI-7831R. After connecting
I/O signals to the SCB-68 screw terminals, you can connect the SCB-68 to
the NI PXI-7831R with the SH68-C68-S shielded cable.
© National Instruments Corporation
C-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix C
Using the SCB-68 Shielded Connector Block
Quick Reference Label
Figure C-2 shows the pinout that appears on the SCB-68 quick reference
label that ships with the NI PXI-7831R.
SCB-68 Quick Reference Label
NI 7811R/7831R DEVICES1
NATIONAL
INSTRUMENTS
PIN#
68
34
67
33
66
32
65
31
64
30
63
29
62
28
61
27
60
26
59
25
58
24
23
MIO
AI0+
DIO
DIO39
DIO38
DIO37
DIO36
DIO35
DIO34
DIO33
DIO32
DIO31
DIO30
DIO29
DIO28
DIO27
+5V
PIN#
12
46
13
47
14
48
15
49
16
50
17
51
18
52
19
53
20
54
21
55
22
56
PIN#
1
AI0-
MIO
DIO12
DIO13
DIO14
DIO15
AOGND7
AO7
DIO
MIO
+5V
DIO
DGND
DIO0
AIGND0
AIGND1
AI1+
DGND
DIO11
DGND
DIO12
DGND
DIO13
DGND
DIO14
DGND
DIO15
DGND
DIO16
DGND
DIO17
DGND
DIO18
DGND
DIO19
DGND
DIO20
DGND
35
2
+5V
DGND
DIO0
DGND
DIO1
36
3
AI1-
AI2+
DGND
DIO1
DGND
DIO2
37
4
AI2-
AIGND2
AIGND3
AI3+
AOGND6
AO6
DGND
DIO2
DGND
DIO3
38
5
AOGND5
AO5
DGND
DIO3
DGND
DIO4
39
6
AI3-
AI4+
AOGND4
AO4
DGND
DIO4
DGND
DIO5
40
7
AI4-
1
THE MIO COLUMN CORRESPONDS
AIGND4
AIGND5
AI5+
DIO26
+5V
AOGND3
AO3
DGND
DIO5
DGND
DIO6
TO THE MIO CONNECTOR ON THE
NI 7831R, AND THE DIO COLUMN
CORRESPONDS TO THE DIO
CONNECTORS ON THE
41
8
NI 7811R / 7831R.
DIO25
DGND
DIO24
DGND
DIO23
DGND
DIO22
DGND
AOGND2
AO2
DGND
DIO6
DGND
DIO7
NC = No Connect
42
9
AI5-
SET SWITCHES IN
THIS CONFIGURATION
TO USE THE SCB-68
WITH THE
AI6+
AOGND0
AO1
DGND
DIO7
DGND
DIO8
43
10
44
11
45
AI6-
AIGND6
AIGND7
AOGND0
AO0
DIO8
DGND
DIO9
NI 7811R/7831R
DIO9
S1
NC
DIO10
DIO11
DGND
DIO10
S5 S4 S3
AI7-
AISENSE DIO21
Figure C-2. SCB-68 Quick Reference Label for the NI PXI-7831R
NI PXI-7831R User Manual
C-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
D
Technical Support and
Professional Services
Visit the following sections of the National Instruments Web site at
ni.comfor technical support and professional services:
•
Support—Online technical support resources include the following:
–
Self-Help Resources—For immediate answers and solutions,
visit our extensive library of technical support resources available
in English, Japanese, and Spanish at ni.com/support. These
resources are available for most products at no cost to registered
users and include software drivers and updates, a KnowledgeBase,
product manuals, step-by-step troubleshooting wizards,
conformity documentation, example code, tutorials and
application notes, instrument drivers, discussion forums,
a measurement glossary, and so on.
–
Assisted Support Options—Contact NI engineers and other
measurement and automation professionals by visiting
ni.com/support. Our online system helps you define your
question and connects you to the experts by phone, discussion
forum, or email.
•
•
Training—Visit ni.com/custedfor self-paced tutorials, videos, and
interactive CDs. You also can register for instructor-led, hands-on
courses at locations around the world.
System Integration—If you have time constraints, limited in-house
technical resources, or other project challenges, NI Alliance Program
members can help. To learn more, call your local NI office or visit
ni.com/alliance.
•
Declaration of Conformity (DoC)—A DoC is our claim of
compliance with the Council of the European Communities using the
manufacturer’s declaration of conformity. This system affords the user
protection for electronic compatibility (EMC) and product safety. You
can obtain the DoC for your product by visiting
ni.com/hardref.nsf.
© National Instruments Corporation
D-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix D
Technical Support and Professional Services
•
Calibration Certificate—If your product supports calibration, you
can obtain the calibration certificate for your product at
ni.com/calibration.
If you searched ni.comand could not find the answers you need, contact
your local office or NI corporate headquarters. Phone numbers for our
worldwide offices are listed at the front of this manual. You also can visit
the Worldwide Offices section of ni.com/niglobalto access the branch
office Web sites, which provide up-to-date contact information, support
phone numbers, email addresses, and current events.
NI PXI-7831R User Manual
D-2
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
Symbol
Prefix
pico
Value
10–12
10–9
10– 6
10–3
103
p
n
nano
micro
milli
kilo
µ
m
k
M
G
mega
giga
106
109
Numbers/Symbols
°
Degrees.
>
≥
<
≤
–
Greater than.
Greater than or equal to.
Less than.
Less than or equal to.
Negative of, or minus.
Ohms.
Ω
/
Per.
%
Percent.
Plus or minus.
Positive of, or plus.
+
© National Instruments Corporation
G-1
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
Square root of.
+5V
+5 VDC source signal.
A
A
Amperes.
A/D
AC
ADC
Analog-to-digital.
Alternating current.
Analog-to-digital converter—an electronic device, often an integrated
circuit, that converts an analog voltage to a digital number.
AI
Analog input.
AI<i>
AIGND
AISENSE
AO
Analog input channel signal.
Analog input ground signal.
Analog input sense signal.
Analog output.
AO<i>
AOGND
ASIC
Analog output channel signal.
Analog output ground signal.
Application-Specific Integrated Circuit—a proprietary semiconductor
component designed and manufactured to perform a set of specific
functions.
B
bipolar
A signal range that includes both positive and negative values
(for example, –5 to +5 V).
NI PXI-7831R User Manual
G-2
© National Instruments Corporation
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
C
C
Celsius.
CalDAC
CH
Calibration DAC.
Channel—pin or wire lead to which you apply or from which you read the
analog or digital signal. Analog signals can be single-ended or differential.
For digital signals, you group channels to form ports. Ports usually consist
of either four or eight digital channels.
cm
Centimeter.
CMOS
CMRR
Complementary metal-oxide semiconductor.
Common-mode rejection ratio—a measure of an instrument’s ability to
reject interference from a common-mode signal, usually expressed in
decibels (dB).
common-mode voltage
CompactPCI
Any voltage present at the instrumentation amplifier inputs with respect to
amplifier ground.
Refers to the core specification defined by the PCI Industrial Computer
Manufacturer’s Group (PICMG).
D
D/A
Digital-to-analog.
DAC
Digital-to-analog converter—an electronic device, often an integrated
circuit, that converts a digital number into a corresponding analog voltage
or current.
DAQ
dB
Data acquisition—a system that uses the computer to collect, receive, and
generate electrical signals.
Decibel—the unit for expressing a logarithmic measure of the ratio of
two signal levels: dB = 20log10 V1/V2, for signals in volts.
DC
Direct current.
DGND
DIFF
Digital ground signal.
Differential mode.
© National Instruments Corporation
G-3
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
DIO
Digital input/output.
DIO<i>
DMA
Digital input/output channel signal.
Direct memory access—a method by which data can be transferred to/from
computer memory from/to a device or memory on the bus while the
processor does something else. DMA is the fastest method of transferring
data to/from computer memory.
DNL
DO
Differential nonlinearity—a measure in LSB of the worst-case deviation of
code widths from their ideal value of 1 LSB.
Digital output.
E
EEPROM
Electrically erasable programmable read-only memory—ROM that can be
erased with an electrical signal and reprogrammed.
F
FPGA
Field-programmable gate array.
FPGA VI
A configuration that is downloaded to the FPGA and that determines the
functionality of the hardware.
G
glitch
An unwanted signal excursion of short duration that is usually unavoidable.
H
h
Hour.
HIL
Hz
Hardware-in-the-loop.
Hertz.
NI PXI-7831R User Manual
G-4
© National Instruments Corporation
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
I
I/O
Input/output—the transfer of data to/from a computer system involving
communications channels, operator interface devices, and/or data
acquisition and control interfaces.
INL
Relative accuracy.
L
LabVIEW
Laboratory Virtual Instrument Engineering Workbench. LabVIEW is a
graphical programming language that uses icons instead of lines of text to
create programs.
LSB
Least significant bit.
M
m
Meter.
max
Maximum.
MIMO
min
Multiple input, multiple output.
Minimum.
MIO
Multifunction I/O.
monotonicity
A characteristic of a DAC in which the analog output always increases as
the values of the digital code input to it increase.
mux
Multiplexer—a switching device with multiple inputs that sequentially
connects each of its inputs to its output, typically at high speeds, in order to
measure several signals with a single analog input channel.
© National Instruments Corporation
G-5
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
N
noise
An undesirable electrical signal—noise comes from external sources such
as the AC power line, motors, generators, transformers, fluorescent lights,
CRT displays, computers, electrical storms, welders, radio transmitters,
and internal sources such as semiconductors, resistors, and capacitors.
Noise corrupts signals you are trying to send or receive.
NRSE
Nonreferenced single-ended mode—all measurements are made with
respect to a common (NRSE) measurement system reference, but the
voltage at this reference can vary with respect to the measurement system
ground.
O
OUT
Output pin—a counter output pin where the counter can generate various
TTL pulse waveforms.
P
PCI
Peripheral Component Interconnect—a high-performance expansion bus
architecture originally developed by Intel to replace ISA and EISA. It is
achieving widespread acceptance as a standard for PCs and work-stations;
it offers a theoretical maximum transfer rate of 132 MB/s.
port
(1) A communications connection on a computer or a remote controller.
(2) A digital port, consisting of four or eight lines of digital input and/or
output.
ppm
pu
Parts per million.
Pull-up.
PWM
PXI
Pulse-width modulation.
Stands for PCI eXtensions for Instrumentation. PXI is an open specification
that builds off the CompactPCI specification by adding
instrumentation-specific features.
NI PXI-7831R User Manual
G-6
© National Instruments Corporation
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
R
RAM
Random-access memory—the generic term for the read/write memory that
is used in computers. RAM allows bits and bytes to be written to it as well
as read from. Various types of RAM are DRAM, EDO RAM, SRAM, and
VRAM.
resolution
The smallest signal increment that can be detected by a measurement
system. Resolution can be expressed in bits, in proportions, or in percent of
full scale. For example, a system has 12-bit resolution, one part in 4,096
resolution, and 0.0244% of full scale.
RIO
rms
Reconfigurable I/O.
Root mean square.
RSE
Referenced single-ended mode—all measurements are made with respect
to a common reference measurement system or a ground. Also called a
grounded measurement system.
S
s
Seconds.
Samples.
S
S/s
Samples per second—used to express the rate at which a DAQ board
samples an analog signal.
signal conditioning
slew rate
The manipulation of signals to prepare them for digitizing.
The voltage rate of change as a function of time. The maximum slew rate
of an amplifier is often a key specification to its performance. Slew rate
limitations are first seen as distortion at higher signal frequencies.
© National Instruments Corporation
G-7
NI PXI-7831R User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Glossary
T
THD
Total harmonic distortion—the ratio of the total rms signal due to harmonic
distortion to the overall rms signal, in decibel or a percentage.
thermocouple
A temperature sensor created by joining two dissimilar metals. The
junction produces a small voltage as a function of the temperature.
TTL
Transistor-transistor logic.
two’s complement
Given a number x expressed in base 2 with n digits to the left of the radix
point, the (base 2) number 2n – x.
V
V
Volts.
VDC
VHDCI
VI
Volts direct current.
Very high density cabled interconnect.
Virtual instrument—program in LabVIEW that models the appearance and
function of a physical instrument.
VIH
VIL
Volts, input high.
Volts, input low.
VOH
VOL
Vrms
Volts, output high.
Volts, output low.
Volts, root mean square.
W
waveform
Multiple voltage readings taken at a specific sampling rate.
NI PXI-7831R User Manual
G-8
© National Instruments Corporation
Download from Www.Somanuals.com. All Manuals Search And Download.
|