National Instruments Car Stereo System SCXI 1125 User Manual |
CALIBRATION PROCEDURE
™
SCXI -1125
Introduction
This document contains information and step-by-step instructions for
verifying and calibrating the National Instruments SCXI-1125 signal
conditioning module.
What Is Calibration?
Calibration is a procedure of reading offset and gain errors from a device
and updating special analog calibration circuitry that corrects these errors.
National Instruments calibrates every SCXI-1125 module at the factory.
During the factory calibration process, the calibration constants are stored
in the non-volatile memory of the device. These values are loaded from
memory and used as needed by the device.
Why Should You Calibrate?
Offset and gain errors drift with time and temperature changes, which could
invalidate the factory-set calibration of a device and increase measurement
uncertainty. Calibration restores the device to its specified accuracy.
How Often Should You Calibrate?
The measurement accuracy requirements of your application
determine how often you should calibrate your SCXI-1125 module.
National Instruments recommends you perform a complete calibration
at least once every year. You can shorten this interval to six months or
90 days, based on the demands of your application.
CVI™, LabVIEW™, National Instruments™, ni.com™, NI-DAQ™, and SCXI™ are trademarks of National Instruments Corporation. Product and company
names mentioned herein are trademarks or trade names of their respective companies.
370271A-01
© Copyright 2000 National Instruments Corp. All rights reserved.
December 2000
Download from Www.Somanuals.com. All Manuals Search And Download.
Documentation
The NI-DAQ Function Reference Online Help file and NI-DAQ User
Manual for PC Compatibles contain detailed information on using the
NI-DAQ driver. The function reference online help file includes detailed
information on the driver functions. You can access the online help by
clicking Start»Programs»National Instruments DAQ»NI-DAQ Help.
The user manual provides instructions on installing and configuring
National Instruments DAQ devices. This manual also includes detailed
information on creating applications that use the NI-DAQ driver. These are
your primary references for writing your calibration utility. For further
information on the products you are calibrating, refer to your device user
manuals.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
•
Keep connections to the SCXI module short. Long cables and wires act
as antennae, picking up extra noise and thermal offsets that can affect
measurements.
•
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
•
•
•
Maintain the temperature between 18 and 28 °C.
Keep relative humidity below 80%.
Allow a warm-up time of at least 15 minutes for the SCXI module and
30 minutes for the E Series device to ensure the measurement circuitry
is at a stable operating temperature.
Calibration
The calibration process consists of the following steps:
1. Setting up the device for testing.
2. Verifying the existing operation of the device to determine whether it
is operating within its specifications.
3. Adjusting the device with respect to a known voltage source.
4. Verifying that the device is operating within its specifications after
adjustments.
© National Instruments Corporation
3
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
Setting Up Your Module
Complete the following steps to set up your module for testing:
1. Install the SCXI-1125 in slot 1 of the SCXI chassis.
2. Connect a 68-to-68-pin cable between the SCXI module and the
E Series DAQ device installed in the host computer.
3. Configure the hardware with Measurement & Automation Explorer
(MAX). For more information on configuring the system, refer to the
E Series device and SCXI module documentation.
Verifying Gain and Offset Errors
meeting its specifications. You can use this information to select the
appropriate calibration interval for your application.
Complete the following steps to verify gain and offset errors in the
SCXI-1125:
1. Read the Test Conditions section earlier in this document.
2. Refer to the specifications in Table 1 at the end of this document.
3. Connect the E Series device to the SCXI-1125.
4. Call Calibrate_E_Seriesto minimize the uncertainty associated
with the E Series device measurements. Set the following parameters:
•
•
•
•
deviceNumber—The value assigned by MAX
calOp—ND_SELF_CALIBRATE
setOfCalConstants—ND_USER_EEPROM_AREA
calRefVolts—0.0
5. Call AI_Configureto put the E Series device in bipolar mode. Set the
following parameters:
•
•
•
•
•
•
deviceNumber—The value assigned by MAX
chan—0
inputMode—0
inputRange—0
polarity—0
driveAIS—0
6. Call MIO_Configto enable dithering on the E Series device. Set the
following parameters:
•
•
•
deviceNumber—The value assigned by MAX
dither—1
useAMUX—0
SCXI-1125 Calibration Procedure
4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Connect the calibrator to the appropriate analog input, starting
with channel 0. Refer to Figure 1 to determine which pins on the
96-pin front connector correspond to the positive and negative inputs
of the specified channel. For example, the positive input for channel 0
(CH0+) is pin A32. The negative input for channel 0 (CH0–) is pin
C32. If you are using an SCXI-1320 terminal block connected to the
SCXI-1125, connect the calibrator to the CH0+ and CH0– inputs.
8. Call SCXI_Single_Chan_Setupto configure the module for
single-channel measurements. Set the following parameters:
•
•
•
•
SCXIchassisID—The value assigned by MAX
moduleSlot—1
moduleChan—The same channel number from step 7
DAQdeviceNumber—The device number assigned by MAX for
the E Series DAQ device
9. Call SCXI_Set_Gainto set the gain you want to test on the module.
Set the following parameters:
•
with MAX
•
•
•
moduleSlot—1
channel—The channel number from step 7
gain—The first gain value you want to test—you can find this
gain value in Table 1.
10. Call SCXI_Configure_Filterto configure the filter setting on the
module. Set the following parameters:
•
SCXIchassisID—The ID obtained when the chassis is configured
with MAX
•
•
•
•
moduleSlot—1
channel—The channel number from step 7
filterMode—1
cutoffFrequency—10,000
11. Set the calibrator to output the test point voltage from Table 1.
12. Call DAQ_Op. Set the following parameters:
•
•
•
deviceNumber—The device number of the DAQ device
channel—0
gain—1 for 16-bit E Series devices or –1 for 12-bit E Series
devices
•
•
count—5,000
sampleRate—10,000
© National Instruments Corporation
5
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
13. Call SCXI_Scale. Set the following parameters:
•
•
•
•
•
•
•
•
•
•
SCXIChassisID—The value assigned by MAX
moduleSlot—1
channel—The same channel number from step 7
SCXIgain—The same gain value from step 9
TBgain—1
DAQboard—The device number for the DAQ device
DAQChannel—0
DAQgain—1 for 16-bit devices, –1 for 12-bit devices
numPoints—5.000
binArray—The array returned from the DAQ_Opfunction
The result is a set of scaled voltages that you read from the SCXI
module.
14. Average the results returned from the SCXI_Scalefunction. Compare
the averaged result to the upper and lower limits in Table 1. If the result
falls within these limits, your module has passed this portion of the
test. If it does not, your module needs to be adjusted using the steps in
the next section, Adjusting for Gain and Offset Errors.
15. Repeat steps 8 through 14 for the remaining gains.
16. Repeat steps 7 through 15 for the remaining channels.
You have now finished verifying your SCXI-1125.
Note If all channels fall within the upper and lower limits, your device does not need to
be adjusted.
Adjusting for Gain and Offset Errors
The following two sections explain how to determine the gain and offset
errors of your SCXI-1125 and how to adjust the calibration constants to
correct for these errors.
Determining Gain and Offset Errors
Complete the following steps to determine the gain and offset errors of the
SCXI-1125:
1. Read the Test Conditions section.
2. Connect the E Series device to the SCXI-1125.
3. Call Calibrate_E_Seriesto minimize the uncertainty associated
with the E series device measurements. Set the following parameters:
•
deviceNumber—The value assigned by MAX
SCXI-1125 Calibration Procedure
6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
•
•
•
calOp—ND_SELF_CALIBRATE
setOfCalConstants—ND_USER_EEPROM_AREA
calRefVolts—0.0
4. Call AI_Configureto put the E Series device in bipolar mode. Set the
following parameters:
•
•
•
•
•
•
deviceNumber—The value assigned by MAX
chan—0
inputMode—0
inputRange—0
polarity—0
driveAIS—0
5. Call MIO_Configto enable dithering on the E Series device. Set the
following parameters:
•
•
•
deviceNumber—The value assigned by MAX
dither—1
useAMUX—0
6. Call SCXI_Single_Chan_Setupto configure the module for
single-channel measurements. Set the following parameters:
•
•
•
•
SCXIchassisID—The value assigned by MAX
moduleSlot—1
moduleChan—The channel you want to adjust
DAQdeviceNumber—The device number assigned by MAX for
the E Series DAQ device
7. Call SCXI_Set_Gain. Set the following parameters:
•
SCXIchassisID—The ID of the SCXI chassis obtained when you
configure the chassis with MAX
•
moduleSlot—1, because the SCXI-1125 should be installed in
slot 1
•
•
channel—The same channel number from step 6
gain—The gain value you want to adjust. Refer to Table 1 for
acceptable gain values
8. Call SCXI_Configure_Filterto configure the filter setting on the
module. Set the following parameters:
•
SCXIchassisID—The ID of the SCXI chassis obtained when you
configure the chassis with MAX
•
•
moduleSlot—1
channel—The same channel number from step 7
© National Instruments Corporation
7
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
•
•
filterMode—1
cutoffFrequency—10,000
9. Connect the calibrator to the analog input you want to adjust. Refer
to Figure 1 to determine the pins on the 96-pin front connector that
correspond to the positive and negative inputs of the specified channel.
For example, the positive input for channel 0 is pin A32, which is
is labeled CH0–. If you are using an SCXI-1320 terminal block
connected to the SCXI-1125, connect the calibrator to the CH0+
and CH0– inputs.
10. Connect the DMM to the positive output on pin 3 (MCH0+) and
negative output on pin 4 (MCH0–) on the rear panel connector.
Figure 2 shows the 50-pin rear panel connector.
Tip For easy access to individual pins, use a screw terminal adapter connected to a
50-pin terminal block.
11. Set the calibrator voltage to the positive test point value in Table 1.
12. Read the voltage from the DMM. Record the actual DMM reading,
which is output1, and the calibrator output voltage, which is volt1, for
later use.
13. Set the calibrator to the negative test point value for the same gain.
Skip any input limits that are specified as 0.0 V. You need only upper
and lower limits for adjustment.
14. Read the voltage from the DMM. Record the actual DMM reading,
output2, and the calibrator output voltage, voltage2.
You now have two pairs of data points, (volt1, output1) and (volt2,
output2), where volt is the calibrator reading and output is the DMM
reading.
15. Convert the DMM voltage readings (output1 and output2) to binary
readings (binary1 and binary2) by using one of the following
equations:
output
20
× 212
----------------
binary reading for 12-bit E Series products (binary) =
output
----------------
20
binary reading for 16-bit E Series products (binary)=
× 216
For example, using a 12-bit product such as a PCI-MIO-16E-1 and
obtaining a DMM reading of –9.9 V, you would have the following
results:
binary reading = (–9.9/20) × 212 = –2027.52
SCXI-1125 Calibration Procedure
8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
16. Record binary1 with volt1 and binary2 with volt2 for later use.
adjust on this channel.
18. Repeat steps 6 through 17 for the remaining channels you want to
adjust.
You have determined the new calibration constants. Use these values in the
the following section, Adjusting Calibration Constants.
Adjusting Calibration Constants
Complete the following steps to adjust the calibration constants on the
SCXI-1125:
1. Call SCXI_Cal_Constantsto create and store the new calibration
constants in the SCXI-1125 memory. Set the following parameters:
•
•
SCXIchassisID—The ID assigned by MAX
ModuleSlot—1, because the SCXI-1125 should be installed in
slot 1
•
•
•
•
•
•
•
•
•
•
channel—The channel on the SCXI-1125 you want to adjust
OpCode—2
calibrationArea—0
rangeCode—0 (not used for the SCXI-1125)
SCXIgain—The gain you want to adjust
DAQboard—The device number assigned by MAX
DAQChan—0
DAQGain—1 for 16-bit devices, –1 for 12-bit devices
Tbgain—1.0
volt1 and volt2—The first and second voltage readings for the
gain values that you are about to save on the SCXI-1125
•
•
binary1 and binary2—The first and second binary readings for
the gain values that you are about to save on the SCXI-1125
calConst1 and calConst2—These are return values
2. Repeat step 1, changing calibrationArea to 1.
3. Repeat steps 1 and 2 for the next channel.
You have now finished adjusting the calibration constants of your
SCXI-1125.
© National Instruments Corporation
9
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
Verifying Adjusted Values
After completing the adjustments to your SCXI-1125 module, it is
important that you verify the analog input operation by repeating the steps
listed in the Verifying Gain and Offset Errors section of this document.
Reverifying after adjustments ensures that your SCXI-1125 module is now
operating within its specifications.
SCXI-1125 Calibration Procedure
10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Front and Rear Signal Connector Diagrams
A
B
C
32
32
CH0+
CH1+
CH0–
CH1–
NC
30
28
26
24
22
20
18
16
14
12
10
8
30
28
NC
CH2+
CH3+
26 CH2–
24 CH3–
22
20 CH4–
NC
NC
CH4+
CH5+
18
16
CH5–
NC
NC
CH6+
14 CH6–
12 CH7–
CH7+
NC
NC
10
8
RSVD
RSVD
CJTEMP
RSVD
RSVD
RSVD
+5 V
6
6
4
4
CHSGND
2
2
This symbol indicates that no physical pin is present
Figure 1. Front Signal Connector Pin Assignments
© National Instruments Corporation
11
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
AOGND
MCH0+
MCH1+
MCH2+
MCH3+
1
3
5
7
9
2
4
6
8
AOGND
MCH0–
MCH1–
MCH2–
10 MCH3–
MCH4+ 11 12 MCH4–
MCH5+ 13 14 MCH5–
15 16
MCH6–
MCH6+
MCH7+ 17 18 MCH7–
OUTREF 19 20
21 22
23 24
DIGGND
SERDATIN 25 26 SERDATOUT
DAQD*/A 27 28
SLOT0SEL* 29 30
31 32
DIGGND 33 34
SCANCLK
35 36
SERCLK 37 38
39 40
41 42
43 44
45 46
SYNC
47 48
49 50
Figure 2. Rear Signal Connector Pin Assignments
SCXI-1125 Calibration Procedure
12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
The following table contains the specifications you should use when
verifying and calibrating your SCXI-1125 module. If your device has been
calibrated within the last year, the test point value should fall between the
Upper Limit and Lower Limit values.
Table 1. SCXI-1125 Specifications at 25 °C 10 °C
Gain
1
Range
5
Test Point
4.950000
0.000000
–4.950000
2.475000
0.000000
–2.475000
0.990000
0.000000
–0.990000
0.495000
0.000000
–0.495000
0.247500
0.000000
–0.247500
0.099000
0.000000
–0.099000
0.0495000
0.000000
–0.049500
0.024750
0.000000
Upper Limit (V)
4.9623181
0.005002
Lower Limit (V)
4.9376819
–0.005002
–4.9623181
2.4688400
–0.002502
–2.4811601
0.9875348
–0.001002
–0.9924652
0.4937664
–0.000502
–0.4962336
0.24688220
–0.000252
–0.24811781
0.09875168
–0.000102
–0.09924832
0.04937484
–0.000052
–0.04962516
0.02468642
–0.000027
1
0
1
–5
–4.9376819
2.4811601
0.002502
2
2.5
0
2
2
–2.5
1
–2.4688400
0.9924652
0.001002
5
5
0
5
–1
–0.9875348
0.4962336
0.000502
10
10
10
20
20
20
50
50
50
100
100
100
200
200
0.5
0
–0.5
0.25
0
–0.4937664
0.24811781
0.000252
–0.25
0.1
0
–0.24688220
0.09924832
0.000102
–0.1
0.05
0
–0.09875168
0.04962516
0.000052
–0.05
0.025
0
–0.04937484
0.02481358
0.000027
© National Instruments Corporation
13
SCXI-1125 Calibration Procedure
Download from Www.Somanuals.com. All Manuals Search And Download.
Table 1. SCXI-1125 Specifications at 25 °C 10 °C (Continued)
Gain
200
Range
–0.025
0.02
0
Test Point
–0.024750
0.019800
0.000000
–0.019800
0.009900
0.000000
–0.009900
0.00495
Upper Limit (V)
–0.02468642
0.01985126
0.000022
Lower Limit (V)
–0.02481358
0.01974874
–0.000022
250
250
250
–0.02
0.01
0
–0.01974874
0.00992663
0.000012
–0.01985126
0.00987337
–0.000012
500
500
500
–0.01
0.005
0
–0.00987337
0.00496432
0.000007
–0.00992663
0.00493568
–0.000007
1000
1000
1000
2000
2000
2000
0.000000
–0.004950
0.002475
0.000000
–0.002475
–0.005
0.0025
0
–0.00493568
0.00248316
0.000005
–0.00496432
0.00246684
–0.000005
0.0025
–0.00246684
–0.00248316
Limits Valid for 4 Hz or 10 kHz Filter Settings
Valid when using an MIO device after performing an MIO internal calibration.
Valid for both filter settings with 5.000 point averaging at 10,000 samples per second.
SCXI-1125 Calibration Procedure
14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
|