Micro Technic Medical Alarms 81781 User Manual

ENTERPRISES  
TM  
Informer-Mate Model 81781  
TM  
Informer-Mate Deluxe Model 81791  
User Manual  
800-206-1045  
0163-198-G  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNING  
Read these installation instructions thoroughly before using the Informer-Mate™  
system. Failure to do so may result in injury or death to a person in your care.  
Test the Informer-Matecontrol unit and sensor pad before each use and inspect  
the cords and sensor pads for signs of damage. Replace any components with signs  
of wear or damage immediately.  
The Informer-Matesystem is only one part of your facility's fall management  
program. The Informer-Matesystem is not a substitute for proper nursing care or  
routine visual monitoring by caregivers. The effectiveness of the Informer-Mate™  
system relies entirely on an immediate response by the caregiver to the Informer-  
Matesystem alarm.  
The Informer-Matesystem will not stop a person from leaving a bed or chair. It is  
intended only to alert a caregiver that a patient or resident may need assistance.  
The Informer-Matesystem can be defeated by a cognitively aware person, a  
person with only a few moments of lucidity, or an uncooperative person. Properly  
assess each individual before the Informer-Matesystem is used.  
Resident weight may affect proper operation. Models 82700, 82708, 82710, 82715,  
83700 and 83705 pressure pads are not recommended for persons weighing less  
than 60 lbs.  
Keep the sensor pad flat at all times. Folding the pad will damage it. Do not use the  
pad if it has been folded.  
Sensor pads were designed to withstand normal wear and tear, but are more likely  
to fail and fail without warning due to prolonged use and other factors, e.g.  
bending, exposure to moisture, punctures, repeated cord pulls, connector damage,  
etc. Test each pad before each use and remove it from service at any sign of  
damage or at the end of the stated life of the pad (as labeled on it), whichever  
comes first.  
CAUTION  
Use care when installing the Informer-Matesystem. Gently remove or connect  
cords. Pulling on cords may damage them and/or result in system failure.  
When putting your sensor pads into position, keep the pad face up. Failure to do  
so will result in reduced sensitivity and improper operation of your Informer-  
MateTM system.  
Replace the battery immediately if the red low battery light on the control unit is  
on. The control unit will not function properly with a low battery.  
Do not immerse the sensor pads in liquids. The pad will not operate properly if the  
sensor pad is exposed to excessive liquids. If the pad is immersed in liquid, discard  
it immediately.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
®
Description of the Micro-Tech Informer-  
Mate Series Monitoring Systems  
WARNING  
Before installing this system, please read and follow these instructions carefully.  
Failure to do so could result in injury or death to a person in your care. No  
device is a substitute for proper nursing care.  
®
The Micro-Tech Informer-Mate system is designed to help reduce the risk  
of falls by alerting caregivers when a resident is leaving a bed, chair, or  
wheelchair. Mount the monitor on the bed, chair or wheelchair using one  
®
of the Micro-Tech mounting brackets according to the included  
®
instructions. Place a Micro-Tech pressure pad on the bed, chair or  
wheelchair according to the instructions included with the pad and attach  
the cord to the monitor. When a resident moves from the pad, an alarm  
sounds, alerting you of a potential fall risk.  
Monitor Design  
®
The Micro-Tech Informer-Mate series incorporates several important  
features, allowing you to focus on resident care and safety while  
maintaining protocols. The monitors feature a dual button reset feature  
which helps discourage resident tampering. Multiple alarm notification  
options are available, allowing you to customize alert notification for each  
resident, each wing, or any other configuration you desire. Lights have  
been added to help you identify the monitor when in alarm condition in a  
crowded or dark room. To help protect the monitor, an integrated cushion  
protector and pad strain relief system have been added to protect the  
monitor and pad from accidental damage due to dropping or wheelchair  
entanglement.  
Alert Notifications  
®
The Micro-Tech Informer-Mate series can be set up in multiple alarm  
configurations to allow you to customize the monitor to your specific  
resident and staff needs. The monitors feature Hi/Low volume control, 0 or  
2-second alarm delay to help prevent false alarms from resident movement,  
and lights to alert staff to an alarm condition. Additionally, the Informer-  
Mate Deluxe monitor offers multiple alarm notification tones and a nurse  
call jack for nurse station notification.  
Multiple Pad Options to Fit your Specific Needs  
Micro-Techoffers multiple pad options to allow you to customize each  
system to your specific patient needs. Pads are available for beds, chairs,  
and wheelchairs in a variety of sizes and time-length options.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
2-Button Reset  
DC Adapter  
(81791 only)  
Pad Jack  
Nurse call  
receptacle  
(81791 only)  
Strain  
Relief  
Figure 2  
Figure 1  
Battery  
Compartment  
Hi / Lo/ NC  
Switch  
Alarm tone  
selector  
(81791 only)  
0 or 2-  
second  
delay switch  
Figure 3  
Alarm Mode Switch Settings  
Volume Setting  
81781  
81791  
Local alarm only  
Local alarm only  
Local alarm only *  
High  
Local alarm only  
Low  
Local alarm only  
Alert at nurse station**  
Nurse Call  
*If the Informer-MateModel 81781 is set to Nurse call the monitor defaults to Low setting.  
**If the Informer-MateDeluxe Model 81791monitor is set to Nurse Call and the nurse  
station is not connected the monitor defaults to local alarm.  
Battery Installation  
Slide open the battery cover on the back of the monitor. The cover will  
remain attached to the monitor. Insert a new 9V alkaline battery, making  
sure to observe proper battery polarity (positive to positive, negative to  
negative).  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installing the Informer-Mate System on a Bed  
WARNING  
Keep the sensor pad flat at all times. Folding the pad may damage it. Do  
not use the pad if it has been folded.  
Check for proper operation of the Informer-Matesystem with each  
mattress type used.  
Above the mattress set-up  
1. Place the bed sensor pad across the width of the bed and under any  
sheets, overlays or incontinence pads (Figure 4). Placement should  
be under the patient or resident's buttocks area.  
Pad  
overlay  
Informer-Mate  
control unit  
Sensor  
Sensor wire  
Figure 4  
2. Using the band and clip package provided with the sensor pad,  
fasten the sensor pad to the mattress (Figure 5). Make sure the  
sensor pad is fastened securely and lies flat against the mattress.  
Sensor  
Rubber band  
Clip  
Side of  
mattress  
Bead of  
mattress  
Figure 5  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Route the sensor pad cable from the sensor pad to the control unit.  
The cable will need to run from the top of the mattress to the  
underside of the bed and up the headboard, footboard, or rail of the  
bed (Figure 4). Make sure the cable lies freely and does not  
interfere with handrails or moving parts of the bed.  
4. Mount the control unit on the bed headboard, footboard, or rail in a  
convenient location using the mounting strap or mounting bracket.  
Optional wide mounting brackets are available for extra wide  
headboards and footboards.  
5. Insert the sensor pad cable into the sensor pad receptacle on the  
control unit.  
6. If you are using the Informer-Mate Deluxe system, set the desired  
alarm tone by sliding the volume switch to position 1, 2, 3 or 4. If  
connecting a nurse call cable to the deluxe unit, the alarm will,  
when sounding, automatically be sent to the nurse call unit and the  
local alarm will be disabled.  
7. Test the Informer-Matemonitoring system before each use and  
inspect the pads and monitor regularly to make sure they are not  
damaged. Test the system by pressing firmly on the pad for 3  
seconds. When you remove pressure, the alarm should sound. Reset  
the alarm using the 2-button reset.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installing the Informer-Mate System On a Chair  
Informer-Mate™  
control unit  
Chair setup  
1. Place the chair sensor pad on  
the seat portion of the chair  
(Figure 6). An additional  
cushion may be used on top of  
the pad for patient or resident  
comfort. If you use an  
additional cushion, test the  
system by pressing firmly on  
the top of both cushions.  
Sensor  
pad  
2. Mount the control unit on the  
chair in a convenient location  
with the mounting clip or  
Figure 6  
mounting strap provided.  
3. Insert the sensor pad cable into the sensor pad receptacle on the  
control unit.  
4. If you are using the Informer-Mate Deluxe system, set the desired  
alarm tone by sliding the volume switch to position 1, 2, 3 or 4. If  
connecting a nurse call cable to the deluxe unit, the alarm, when  
sounding, will automatically be sent to the nurse call unit and the  
local alarm will be disabled.  
5. Test the Informer-Matemonitoring system before each use and  
inspect the pads and monitor regularly to make sure they are not  
damaged. Test the system by pressing firmly on the pad for 3  
seconds. When you remove pressure, the alarm should sound. Reset  
the alarm using the 2-button reset.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
To keep your Informer-Mate control unit in good condition:  
• Do not immerse the control unit in liquid of any type. The control  
unit will not operate properly if exposed to excessive liquids. If the  
control unit is immersed in liquid, discard it immediately.  
• Gently clean the Informer-Mate control unit with non-caustic  
cleaning agents.  
• Test your Informer-Mate control unit and sensor pad before each  
use and inspect the cords, controls and sensor pads for signs of  
damage. Replace any components with signs of wear or damage  
immediately. Never use equipment that has been damaged.  
• Call your distributor or Micro-Tech at 800-206-1045 to order  
replacement parts.  
To keep your sensor pads in good condition:  
• Do not immerse the pressure pad in liquid of any type. The  
pressure pads will not operate properly if exposed to excessive  
liquids. If the pad is immersed in liquid, discard it immediately.  
• Clean soiled areas of sensor pad with a clean cloth and water or a  
vinyl-tolerant solution.  
• Keep the pressure pad flat at all times. Folding the pad will damage  
it. Do not use the pad if it has been folded.  
• Call your distributor or Micro-Tech at 800-206-1045 to order  
replacement parts.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Warranty and Disclaimer  
WARRANTY AND DISCLAIMER: Micro-Tech® products are warranted to be free of  
manufacturing defects for a period of one year from the date of purchase. If any defects  
covered by this Warranty appear within the above stated one year period, Micro-Tech will,  
at its option, repair or replace the defective component on an exchange basis with new or  
rebuilt parts at its expense, without charge for parts or labor upon delivery of such product  
to Micro-Tech at Buyer’s expense. EXCEPTION: MICRO-TECH® PADS AND  
ACCESSORIES ARE WARRANTED TO BE FREE OF MANUFACTURING  
DEFECTS FOR PERIODS RANGING FROM 30 DAYS TO 1 YEAR AS  
SPECIFIED TO BUYER. Micro-Tech is not responsible for warranty service should the  
Micro-Tech® label or logo or the serial number be removed or should the Micro-Tech®  
product fail to be properly maintained or fail to function properly as a result of misuse,  
abuse, neglect, improper shipping or installation, incorrect wiring not our own, damage  
caused by disasters, such as fire, flood, and lightning, damage caused by faulty or leaking  
batteries not supplied by Micro-Tech, service other than by Micro-Tech or units in use in  
violation of instructions furnished by Micro-Tech. Postage, insurance, or shipping costs  
incurred in presenting your Micro-Tech® product for warranty service are your  
responsibility.  
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND BUYER HEREBY  
WAIVES, RELEASE AND RENOUNCES, ALL OTHER WARRANTIES,  
OBLIGATION AND LIABILITIES OF MICRO-TECH, AND ALL OTHER  
RIGHTS, CLAIMS AND REMEDIES OF BUYER AGAINST MICRO-TECH  
EXPRESS OR IMPLIED, ARISING BY LAW OR OTHERWISE WITH RESPECT  
TO ANY NONCONFORMANCE OR DEFECT IN ITEMS OR SERVICES  
PURCHASED HEREUNDER, INCLUDING, BUT NOT LIMITED TO: (A) ANY  
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A  
PARTICULAR PURPOSE; (B) ANY IMPLIED WARRANTY ARISING FROM  
COURSE OF PERFORMANCE, COURSE OF DEALING OR USAGE OF  
TRADE; (C) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN  
TORT, WHETHER OR NOT ARISING FROM THE NEGLIGENCE OF MICRO-  
TECH (WHETHER ACTIVE, PASSIVE OR IMPUTED); AND (D) ANY  
OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY FOR LOSS OF, OR  
DAMAGE TO, ANY EQUIPMENT.  
MICRO-TECH SHALL HAVE NO OBLIGATION OR LIABILITY, WHETHER  
ARISING IN CONTRACT, TORT OR OTHERWISE, FOR LOSS OF USE,  
REVENUE OR PROFIT, COST OF CAPITAL, COST OF SUBSTITUTE  
EQUIPMENT, ADDITIONAL COSTS INCURRED BY BUYER, BUYER’S  
CUSTOMERS OR OTHER THIRD PARTIES, FOR DAMAGES RESULTING  
FROM PERSONAL INJURY OR PROPERTY DAMAGE OR ANY OTHER  
INCIDENTAL OR CONSEQUENTIAL DAMAGES WITH RESPECT TO ANY  
NONCONFORMANCE OR DEFECT IN ANY ITEMS PROVIDED HEREUNDER.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
®
Micro-Tech Accessory & Replacement Items  
MONITORS  
81781  
81791  
Informer-Mate Monitor  
Informer-Mate Deluxe Monitor (Includes multi-tone  
alarm and is nurse call system compatible)  
SENSOR PAD - BED  
82700  
82708  
82710  
82715  
30-day warranty  
90-day warranty  
1-year warranty  
Dark blue durable nylon cover with 1-year warranty  
SENSOR PAD - CHAIR or WHEELCHAIR  
83700  
83705  
1-year warranty  
Dark blue durable nylon cover with 1-year warranty  
MISCELLANEOUS  
0707-396  
0034-051  
0141-514  
0707-773  
0120-142  
0120-143  
0707-774  
Nurse Call Cord (for use with Deluxe model only)  
Interlock Peel & Stick Fastener (4 pc/package)  
Battery Cover  
Chair Mounting Strap  
Mounting Bracket (included)  
Wide Mounting Bracket  
Monitor Mounting strap (top of unit)  
Contact your Micro-Tech® distributor to order any of these products.  
Senior Technologies, a division of Stanley Security Solutions, Inc.  
Lincoln, NE 68503 USA  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

KTI Networks Switch KSD 541 User Manual
LG Electronics IP Phone IP8540 User Manual
Logitech Webcam c200 User Manual
Lucent Technologies IP Phone CIB 3020 User Manual
Manitowoc Ice Ice Maker K00349 User Manual
Marathon Computer Laptop V 4224 User Manual
Maverick Ventures Thermometer ET 73 User Manual
Maytag Refrigerator 2007 8 User Manual
Maytag Refrigerator M1TXEGMYB User Manual
MB QUART Speaker ALX2PS User Manual