Lumex Syatems Car Seat FR587W User Manual

Series FR587W Deluxe Clinical Care Recliner,  
Extra Wide  
Series FR587WH Deluxe Clinical Care Recliner,  
Extra Wide, with Heat and Massage  
Assembly and Operating Instructions  
IMPORTANT: READ THIS MANUAL BEFORE OPERATING YOUR LUMEX  
FR587W / FR587WH DELUXE EXTRA-WIDE CLINICAL CARE RECLINER.  
SAVE THIS MANUAL FOR FUTURE USE.  
THE MOST CURRENT VERSION OF THIS MANUAL CAN BE FOUND ONLINE  
FR587W-INS-LAB-RevC13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INTRODUCTION  
This manual contains assembly, operation, and maintenance instructions for the Lumex  
Clinical Care Recliners Series FR587W and FR587WH. Read the entire manual carefully  
before using your recliner, and refer to it during use if you have questions. If you have  
further questions, please call Graham-Field / Lumex Technical Support at 770-447-1609.  
INTENDED USE  
Lumex Clinical Care Recliners are intended to make the care, treatment and recovery of  
patients easier and more comfortable. Lumex Clinical Care Recliners are used in a wide  
range of healthcare applications including Dialysis, Oncology, Drug Delivery, Same Day  
Surgery and Post-Operative Recovery. Lumex Clinical Care Recliners are ideal for use in  
hospitals, clinics and rehabilitation centers.  
Lumex Series FR587W and FR587WH supports patient weights up to 450 lb (204 kg) when  
evenly distributed.  
IMPORTANT SAFETY PRECAUTIONS—PLEASE READ BEFORE USE  
The safety statements presented in this chapter refer to the basic safety information that  
the FR587W / FR587WH user must pay attention to and abide by. There are additional  
safety statements in other chapters or sections, which may be the same as or similar to the  
following, or specific to the operations. Please note the following special statements, used  
throughout this manual, and their significance:  
. WARNING:.Indicates.a.potential.hazard.situation.or.unsafe.practice.that,.if.not.  
avoided,.could.result.in.death.or.serious.personal.injury.  
. CAUTION:.Indicates.a.potential.hazard.or.unsafe.practice.that,.if.not.avoided,.could.  
result.in.minor.or.moderate.personal.injury.  
s. NOTICE:.Indicates.a.potential.hazard.or.unsafe.practice.that,.if.not.avoided,.could.  
result.in.product.or.property.damage.  
Info: Provides application recommendations or other useful information to  
ensure that you get the most from your product.  
WARNINGS  
. WARNING:.Read.this.entire.manual.before.assembly.or.operation.  
. WARNING:.Periodically.inspect.all.parts.and.replace.as.needed.  
. WARNING:.Do.not.use.near.an.open.flame,.as.the.chair.could.be.a.combustion.  
source,.resulting.in.fire,.property.damage,.and.severe.physical.injury.  
. WARNING:.Service.and.repair.of.the.Lumex.FR587W./.FR587WH.Deluxe.Extra.Wide.  
Clinical.Care.Recliner.must.be.performed.by.a.Graham-Field.authorized.distributor..  
GF.Health.Products,.Inc..is.not.responsible.for.any.consequence.resulting.from.any.  
unauthorized.service.or.repair.  
. WARNING:.The.electronics.and./.or.actuator.contain.no.serviceable.components..  
If.service.is.required,.the.unit.must.be.returned.to.the.factory.for.repair..DO.NOT.  
attempt.to.open.the.electronics.or.actuator.or.obtain.local.service;.this.will.void.  
the.warranty.and.may.result.in.damage.to.the.unit..Consult.your.Graham-Field.  
authorized.distributor.for.further.information.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
. WARNING:.This.product.has.a.maximum.weight.capacity.of.450.lb.(204.kg),.EVENLY.  
DISTRIBUTED.  
. WARNING:.Casters.must.be.locked.at.all.times,.except.during.transport.  
. WARNING:.Unauthorized.modification.of.your.recliner.or.the.use.of.non-Lumex.  
replacement.parts.may.change.the.structure.of.the.recliner.and.could.create.a.  
hazardous.condition,.which.may.result.in.serious.injury.and.will.void.the.warranty.  
. WARNING:.Notice.for.California.Customers-.California.Proposition.65.WARNING:  
This.product.contains.a.chemical.known.to.the.State.of.California.to.cause.cancer.  
and.reproductive.or.developmental.harm.  
. WARNING:.GF.Health.Products,.Inc..specifically.disclaims.responsibility.for.any.  
bodily.injury.or.property.damage.which.may.occur.during.any.use.which.does.not.  
comply.with.federal,.state.or.local.laws.or.ordinances.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
ASSEMBLY  
UNPACKING  
1. Check for obvious damage to the carton or its contents. If damage is evident, please  
notify the carrier and your Graham-Field authorized distributor.  
2. Remove all packaging materials from recliner base.  
PRODUCT DESCRIPTION  
upholstered back  
upholstered seat  
upholstered armrest  
padded  
push handle  
gas spring  
actuating lever  
upholstered  
legrest / skirt  
upholstered  
footrest  
fold down side table  
frame  
total locking caster  
(total of 3 on recliner; black locking tab)  
rear directional (swivel) locking caster  
(right rear leg only; green locking tab)  
removable side panel  
Clinical Care Recliner Series FR587W  
Upholstered Back Assembly  
3
/
8” diameter  
large cotter pin  
1. After removing all components from the  
carton, engage all four locks on the casters  
before beginning assembly.  
clevis pin  
2. See the location of three clevis pins, two ring  
cotter pins and two large cotter pins at right.  
5
Info: The center clevis pins are /16" in  
back brace with  
U-brackets  
5/16” diameter  
clevis pin  
3
diameter. The outside clevis pins are /8  
"
in diameter.  
Back Assembly  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Remove cardboard shipping retainers and  
tape from around gas spring cylinders.  
Rotate the hinge section on the gas spring  
so that it is perpendicular to the upholstered  
back, as shown at right.  
gas spring hinge  
Rotate HInge  
panel tabs  
side panel  
4. Remove side panels from frame:  
Depress the two push buttons  
located on the lower sides of frame.  
Pull side panel gently away and  
down from frame. There are tabs  
on the frame underneath the  
armrest that hold the side panel  
in place.  
push button  
Remove Side Panel  
5. Lift the upholstered back from the  
sides and place it horizontally with  
the upright back tubes resting on  
the recliner frame.  
clevis pin  
cotter pin  
Clevis Pin and Cotter Pin  
6. Align the holes on the  
upright back tubes of the  
upholstered back with the  
holes on the seat frame.  
3
Place the /8" diameter  
clevis pins through the  
mounting holes in the frame  
and back as shown at right.  
Align Holes  
Place Clevis Pins  
7. Secure the two clevis pins  
by placing a large cotter  
pin through the hole in the  
clevis pin (one for each  
side) as shown at right.  
Place Cotter Pin  
Cotter Pin  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Raise the upholstered back to vertical position.  
Tuck the plastic skirt at rear of upholstered  
back behind seat frame crossbrace. Align the  
hinges at the bottom of the gas springs with  
the holes in the U-brackets on the seat frame  
crossbrace. While holding the back in position  
with one hand, secure both gas spring hinges  
5
to the U-brackets by placing two /16"diameter  
clevis pins through the U-brackets and hinges  
with the other hand.  
Place Clevis Pins  
9. Secure the clevis pins by placing the ring cotter  
pins through the hole in the clevis pins as  
shown at right.  
Place Cotter Pins  
10. Replace side panels:  
panel tabs  
Insert side panel into recliner  
frame by locating the tabs  
side panel  
underneath the armrest.  
Make sure side panel is  
properly aligned and that the  
top edge is behind the locking  
tabs. Press side panel into  
frame sides. Make sure that  
both push buttons are  
protruding into side panel to  
ensure side panel is locked in  
place.  
push button  
Replace Side Panel  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
. WARNING:.Disconnection.of.the.device:.In.case.of.emergency,.always.disconnect.  
the.device.by.removing.the.power.cord.plug.from.the.power.outlet.  
TENTE® CASTERS  
. WARNING:.Casters.must.be.locked.at.all.times,.except.during.transport.  
black actuator tab  
green actuator tab  
All Lumex Clinical Care Recliners now feature  
the latest design, all-plastic casters from TENTE.  
Specifically designed for use in healthcare  
environments, the new casters are easier to clean,  
easier to move, and tolerate a wide array of cleaners  
and chemicals. Each caster is made from high-  
grade synthetic materials and features a double-  
ballbearing swivel head. The wheel center is made  
of polypropylene and the tire tread of TENTEprene  
(thermoplastic rubber). The gray non-marking tires  
feature precision ball bearings.  
total  
locking caster  
directional (swivel)  
locking caster  
(right rear caster)  
Locking casters: Each Lumex Clinical Care Recliner  
features three total locking casters, identified by  
black actuator tabs, that lock both the wheel and the swivel. To activate the total lock,  
simply press down on the BLACK actuator tab until it locks. To release the total lock, press  
on the raised portion of the actuator tab until the lever is back in the normal operating  
position.  
Directional (swivel) caster: The fourth caster, located on the right rear position of the  
recliner, is a directional (swivel) locking caster only. Identified by a green actuator tab, the  
directional lock caster will lock in one of two 180-degree positions relative to the frame  
(forward and backward) and allows the recliner to track straight when pushed. To activate  
the directional lock, simply press down on the GREEN actuator tab until it locks — you may  
need to move the recliner slightly until the caster locks in a straight line with the recliner  
frame. To release the directional lock, press on the raised portion of the actuator tab until  
the lever is back in the normal operating position.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
side table  
(disengaged position)  
SIDE TABLE OPERATION  
armrest  
To raise table: Raise the side  
table all the way up and slide  
in toward the armrest until  
tabs engage slots in mount  
bracket. Press down lightly on  
table to ensure locking tabs  
are engaged.  
3/8" clevis pin  
attachment  
side table  
(engaged position)  
side table  
mount bracket  
tabs engage into  
slots on side table  
mount bracket  
To lower table: Lift the side  
table up slightly until tabs  
disengage slots and pull table  
away from armrest. Lower side  
table to self-storing position.  
side table  
(stored position)  
Side Table  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
SIDE PANEL OPERATION  
For easy cleaning and access to the side and underside of the recliner, Lumex Clinical Care  
Recliners feature easy-to-remove side panels.  
To remove side panel:  
Depress the two push buttons  
panel tabs  
located on the lower sides of  
side panel  
frame. Pull side panel gently away  
and down from frame. There are  
tabs on the frame underneath the  
armrest that hold the side panel  
in place.  
To install side panel:  
Insert side panel into recliner  
frame by locating the tabs  
push button  
underneath the armrest. Make  
sure side panel is properly  
Removable Side Panel  
aligned and that the top edge is  
behind the locking tabs. Press  
side panel into frame sides. Make  
sure that both push buttons are  
protruding into side panel to  
ensure side panel is locked into  
place.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RECLINER OPERATION  
The Lumex Series FR587W and FR587WH Recliners feature three positions for patient  
comfort. The fourth position is a Trendelenburg (or Shock Position) that must be activated  
by a caregiver. The Trendelenburg Position CANNOT be activated by a patient while in the  
recliner. In addition, the Series FR587W features a back that can be repositioned via the  
adjustment lever located on the push handle on the back of the recliner.  
First Position (Upright Position)  
The first position is the Upright Position. This position  
is ideal for patient transport and can be used to initiate  
patient treatment / therapy.  
Second Position (First Recline Position)  
The second position is the First Recline Position and is  
also known as the “TV” position. This position is ideal for  
patient treatment, rest and reading.  
Upright Position  
To Achieve Second Position / First Recline Position:  
• Reclining by patient: Place hands on armrests and push  
body against recliner back to recline yourself to desired  
position. To bring yourself up, place hands on armrests  
and pull yourself up while leaning your body forward,  
away from the recliner back, while at the same time  
pushing down evenly on the footrest with your legs.  
First Recline Position  
(TV Position)  
• Reclining by attendant: Standing on either side of the  
recliner, grasp the armrest with one hand and the push  
handle with the other hand. Push back on push handle  
to recline. To return, grasp push handle and pull forward  
to raise the recliner’s back.  
s. NOTICE:.DO.NOT.use.the.footrest.by.itself.to.  
return.the.recliner.to.an.upright.position.from.any.  
recline.position..Doing.so.may.damage.the.recliner.  
Second Recline Position  
(Full Recline Position)  
mechanism..Please.follow.the.instructions.above.to.  
operate.the.recliner.  
Third Position (Second Recline Position)  
The third position is the Second Recline Position and is also known as the “Full Recline”  
position. This position is ideal for patient treatment, rest and reading.  
To Achieve Third Position / Second Recline Position:  
• By the patient: From the First Recline Position, continue to push against the back by  
placing your hands on armrests and continue to recline yourself to the Second Recline  
Position. To bring yourself up, place hands on armrest and pull yourself up while leaning  
your body forward away from the recliner back while at the same time pushing down  
evenly on the footrest with your legs.  
• Reclining by attendant: Standing on either side of the recliner, grasp the armrest with  
one hand and push handle with the other hand. Push back on push handle to recline. To  
return, grasp push handle and pull forward to raise the recliner back.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Info: Due to the 450 pound weight capacity (when evenly distributed) the FR587W  
Recliner has two gas pistons to assist in the recline and return of the back. One  
piston assists during the rearward movement of the back, the other piston assists  
during the return of the back to an upright position. For this reason, in order for  
the gas piston to operate properly, a person of at least 150 lb (68 kg) should be  
sitting in the recliner.  
Back Angle Adjustment  
For added comfort, the back angle of the  
Series FR577RG and FR577RGH can be adjusted  
independently of the seat position. From any of  
the previous positions, actuate the lever located  
on the rear push handle and position the back  
to desired angle. To return back to standard  
position, actuate the lever and move back to  
desired position.  
Lever  
Trendelenburg Position (Also Known as  
Shock Position)  
Info: This position can only be achieved by a caregiver.  
Standing at the back of the recliner, grasp the  
push handle with one hand while actuating the  
Lever located on the rear push handle with  
other. Slowly guide the back down fully until it  
stops and release the lever. If the articulating  
headrest is in the forward position, push back  
Trendelenburg Position  
so that it is even with the back of the recliner.  
Trendelenburg Position can be obtained from  
(Shock Position)  
any position.  
To bring recliner out of Trendelenburg, stand  
at the back of the recliner and actuate the lever  
shown in upper right picture. Slowly guide the  
back upright and release the lever. The recliner  
can then be placed in the desired position.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
GAS SPRING TROUBLESHOOTING GUIDE  
Info: gas springs have been pre-adjusted at factory  
The gas spring on your recliner provides counterbalancing forces, and is self-locking for  
infinite, precise positioning. Unlocking the gas spring is controlled by actuating the lever  
located on the padded push handle on the back of the recliner. When the lever is actuated  
it pulls on the cable which in turn pulls on a hinge, depressing a button at the end of the  
gas spring rod. If the gas spring does not unlock when the lever is actuated or if it remains  
permanently unlocked when the lever is actuated, the cable must be adjusted.  
Adjusting Cable Tension  
1. Remove the one screw and cable clip which secure the  
cable to the upholstered back. Remove the gas spring  
from the recliner by disassembling the two ring cotter  
pins and the two clevis pins that secure the gas spring to  
the back / frame.  
2. Lower the back until it rests on the stop bars on the  
recliner frame. Loosen the lock nut on the gas spring.  
3. Hold the hinge assembly while turning the gas spring until  
the cable just becomes taut. Back off the gas spring an  
additional half turn.  
Info: If the gas spring was not unlocking when the  
lever was actuated, the gas spring must be turned  
Adjusting Cable Tension 1  
clockwise. If the gas spring was not locking when the  
lever was released, the gas spring must be turned  
counter-clockwise.  
4. Retighten the lock nut.  
5. Reassemble the gas spring to the back of the recliner by  
replacing the clevis pins, bridge pins, and re-attach cable  
to back of recliner with the screw and cable clip.  
6. Actuate the recliner in accordance with the operating  
instructions to verify correct adjustment.  
Adjusting Cable Tension 2  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LUMEX CLINICAL CARE RECLINERS WITH HEAT AND  
MASSAGE FEATURE  
INSTALLATION AND OPERATING GUIDE  
Series FR587WH  
1
Tools required: /4" hex wrench and Phillips Screwdriver  
. WARNING:.Important!.Read.and.understand.these.instructions.before.assembling.  
or using.the.Specialty.Healthcare.Seating.Heat.and.Massage.Unit..If.you.do.  
not.understand.any.part.of.these.warnings,.cautions.or.instructions,.contact.a.  
healthcare.professional.for.direction.in.the.use.of.this.product..If.the.Heat.and.  
Massage.Unit.is.not.properly.assembled,.personal.injury.and.damage.to.the.Heat.and.  
Massage.Unit.could.result.  
. WARNING:.The.recliner.seat.and.back.heating.elements.are.wired.to.operate.in.  
series.only..Do.not.change.the.system.wiring.in.any.way.  
. WARNING:.If.components.are.damaged.or.missing,.contact.your.Graham-Field.  
authorized.distributor.immediately..DO.NOT.use.substitute.parts..Use.only.Lumex.  
replacement.parts..The.use.of.non-Lumex.replacement.parts.could.cause.personal.  
injury,.property.damage,.and.void.the.warranty.  
. WARNING:.GF.Health.Products,.Inc..assumes.no.responsibility.for.any.damage.or.  
injury.caused.by.improper.assembly.or.use.of.this.product.  
Description  
The Lumex Specialty Healthcare Seating Heat and Massage Unit is a low-voltage (12V)  
Heat and Massage Unit, controllable by the user and / or caregiver, designed to enhance  
the comfort of the patient / resident during the use of Lumex Clinical Care Recliners. The  
Heat and Massage Unit offers three levels of seat and back heat, as well as a gentle or firm  
massage function for just lower back or both lower and upper back.  
Connection  
The Heat and Massage Unit comes pre-installed  
in Lumex Clinical Care Recliners with the Heat  
and Massage Option (FR587WH). During recliner  
shipment, the Heat and Selector Switch connections  
are disconnected. Several easy connections are all  
that is required to make the unit functional.  
Main Components  
1. The Selector Switch (shown at right), in the  
switch box mounted on the right side of the  
recliner underneath the armrest, is a single  
multi-position switch that activates both the  
Heat and Massage functions.  
selector switch  
Selector Switch (Mounted  
Underneath Right Armrest)  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. The control module, the  
brains of the system, is inside  
the junction box. Both are  
shown at right.  
3. The power supply, which has  
an 8-foot power cord that  
plugs into the wall power  
outlet, is shown at right.  
Control Module (Junction  
Box Cover Removed)  
Junction Box  
Power  
Supply  
ball lock  
pin  
left  
side table  
Heat and Massage Unit Setup Instructions  
. CAUTION:.To.prevent.personal.injury.and.damage.to.the.recliner.  
always.obtain.assistance.when.positioning.the.recliner.as.  
instructed.in.the.following.sections.  
1. Install the back assembly and gas spring on the seat frame with  
the clevis pins and clips supplied with the recliner as indicated  
previously in the recliner Assembly and Operating Manual.  
Remove  
Left Side Table  
2. Remove the left side table (on the side opposite the mounted  
switch box) from the recliner to prevent damage to it during setup:  
Withdraw the ball lock pin, shown at right, and set ball lock pin and  
table aside.  
3. Place a protective cover or drop cloth on your workbench or floor  
to protect the recliner frame finish and upholstery. With assistance,  
carefully roll the recliner and rest it on its left side to provide  
access to the seat bottom, as shown at right. If desired, for better  
access, you may roll the recliner completely upside down. Carefully  
extend the footrest partway for access to the components, as  
shown at right.  
Roll Recliner  
onto its Left Side  
. WARNING:.DO.NOT.place.your.hands.in.the.path.of.the.moving.  
recliner.mechanism..Moving.parts.could.cause.personal.injury.  
1
4. Use a /4" hex wrench to remove the two hex head screws that  
secure the junction box cover (see picture of junction box at top  
of page). Remove the junction box cover to expose the control  
module.  
Insert Cable  
Connector Labeled  
SWITCH  
5. Insert the cable connector labeled SWITCH through the bushing in  
the plastic skirt at the bottom of the back cushion as shown at right.  
Through Bushing  
switch box cable  
clamp tabs  
6. Insert the cable connector labeled SWITCH into the control module  
connector also labeled SWITCH (see picture of control module at  
top of page) until its locking tab snaps into place.  
screw  
clamp  
7. See picture at right. Use Phillips screwdriver to remove the screw  
that secures the cushioned cable clamp. Remove the clamp and  
place it around the switch box cable. Slide the clamp over the cable  
until the holes in the clamp tabs are in line with the screw, then  
pinch them closed. Insert the screw through the clamp tabs, then  
reinstall the screw and tighten it with a Phillips screwdriver.  
Secure Switch Box  
Cable Clamp  
8. Route the cables from the back cushion between the rear seat  
frame crossbar and seat bottom.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Insert the Y-cable connector labeled UPPER into the back cushion  
cable connector also labeled UPPER until its locking tabs snap into  
place, as shown at right.  
10. Insert the Y-cable connector labeled LOWER into the back cushion  
cable connector also labeled LOWER until its locking tabs snap into  
place, as shown at right.  
Connect Upper  
and Lower Cables  
11. Insert the Y-cable connector labeled MASSAGE into the control  
module connector also labeled MASSAGE (see picture of control  
module at top of previous page) until its locking tab snaps into  
place.  
12. Insert the seat cushion cable connector labeled HEAT A into the  
back cushion cable connector also labeled HEAT A until its locking  
tabs snap into place, as shown at right.  
Connect HEAT A  
Cables  
13. Insert the cable connector labeled HEAT into the control module  
connector also labeled HEAT (see picture of control module at top of previous page)  
until its locking tab snaps into place.  
14. Place the cable labeled HEAT A inside the junction box. Replace the junction box cover;  
replace the two hex head screws that secure the junction box cover (see picture of  
1
junction box at top of previous page). Use a /4" hex wrench to secure the screws.  
15. Gather loose cable sections and secure them with the plastic cable ties furnished (see  
picture of junction box at top of previous page).  
16. With assistance, carefully return the recliner to its upright position.  
17. Reinstall the left side table with the ball lock pin.  
18. The Heat and Massage Unit is now ready for operation. Instructions for use follow.  
s. NOTICE:.Before.rolling.the.recliner,.ensure.the.power.cord.is.unplugged.from.the.AC.  
outlet.  
Heat and Massage Unit Operation Instructions  
1. Plug the power supply cord into any grounded outlet.  
2. The Heat and Massage Unit is operated  
using the single multi-position selector  
switch located on the right side of the  
recliner underneath the armrest. The  
Heat and Massage Functions can be  
operated independently. The selector  
switch faceplate is shown at right.  
3. Press the selector switch UP to activate  
the HEAT FUNCTION.  
4. Press the selector switch DOWN to  
activate the MASSAGE FUNCTION.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. The selector switch has THREE HEAT FUNCTION SETTINGS and FOUR MASSAGE  
FUNCTION SETTINGS as follows:  
a. HEAT FUNCTION SETTINGS:  
1) Press selector switch UP for Setting 1: High Heat, 106°F (41°C), brightest LED.  
2) Press selector switch UP AGAIN for Setting 2: Medium Heat, 104°F (40°C), bright  
LED.  
3) Press selector switch UP AGAIN for Setting 3: Low Heat, 100°F (37°C), dim LED.  
4) Press selector switch UP AGAIN: Heat function shuts off.  
5) There is a small red indicator light, which illuminates only during the Heat  
function, in the selector switch. The light's brightness is determined by the Heat  
level being used: Brightest = High Heat; Dimmest = Low Heat.  
b. MASSAGE FUNCTION SETTINGS:  
1) Press selector switch DOWN for Setting 1: Gentle Lower Back Massage.  
2) Press selector switch DOWN AGAIN for Setting 2: Firm Lower Back Massage.  
3) Press selector switch DOWN AGAIN for Setting 3: Gentle Upper and Lower Back  
Massage.  
4) Press selector switch DOWN AGAIN for Setting 4: Firm Upper and Lower Back  
Massage.  
5) Press selector switch DOWN AGAIN: Massage function Shuts Off.  
6. The Heat and Massage Unit automatically shuts off at the following times:  
a. HEAT FUNCTION: After one hour of continuous use at the same level.  
b. MASSAGE FUNCTION: After fifteen minutes of continuous use at the same level.  
c. To restart the Heat and Massage Unit, follow the instructions above.  
Heat and Massage Unit Troubleshooting Guide  
The cables and connections need to be validated. This will require the services of a  
maintenance person and a clear working area.  
1. INSPECTION: Perform the following checks. If a working recliner of the same model is  
available, it should be used for comparison.  
2. Is the AC cord properly connected to the power supply? Is the cable from the power  
supply to control module junction box in place?  
3. With the power supply AC cord plugged into a live wall outlet, does the small light to the  
left of the white seat symbol (located on the switch) light up when the switch is pressed  
repeatedly or is there any indication of massage operation?  
. WARNING:.Disconnect.the.power.cord.from.the.wall.outlet.before.performing.the.  
following.steps!  
a. Is the switch seat symbol oriented so that the back is vertical with the three arrows  
pointing up? Is the switch loose in the box?  
b. Is there any sign of external damage to the switch, switch box or the surrounding  
area?  
c. Is there any sign of external damage to the recliner?  
d. Do any of the cables show any sign of damage or pinch points?  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
e. Are all cable connections to the control module (in the junction box) in place per the  
instructions? Are the connectors fully engaged?  
f. Are any of the connectors damaged?  
g. Are the wires within the connectors attached to the pins? Each wire may have to be  
tugged lightly to check if the wire is still crimped to the pin.  
4. POST-INSPECTION:  
a. If improper or loose connections are found, reconnect them firmly (unless a loose  
wire has been found). Plug the power supply AC cord into a live wall socket and  
operate the unit through all functions.  
b. If the switch seat symbol is rotated or there is evidence of damage to the box and  
cable notify Graham-Field so that a switch box and cable can be sent for replacement.  
c. If damage to any other electrical cable or component, or disconnected wires are  
found, please notify Graham-Field.  
5. ACTIONS:  
a. If full operation is restored, please advise Graham-Field of your successful outcome.  
b. If the fault cannot be corrected by simple re-connection, please inform Graham-Field  
of the results of your inspection in detail so that we may determine the appropriate  
action to be taken.  
If you have any questions about the setup and / or operation of the Heat and Massage  
Unit, please call the Graham-Field Technical Support Group at 770-447-1609.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CARE AND MAINTENANCE  
. WARNING:.Inspect.all.components.periodically.and.replace.as.needed.  
EVERY THREE MONTHS  
• Check and clean casters.  
• Tighten bolts in each leg that secure caster into frame.  
• Check brakes on casters.  
• Check and clean lock recline mechanism.  
• Check recliner mechanism for operation in all positions, including Trendelenburg  
Position, with assistant seated in recliner.  
UPHOLSTERY MAINTENANCE AND STAIN REMOVAL  
• Ordinary dirt: wash with warm water and a mild soap or detergent.  
• Stubborn dirt: using a soft bristle brush, wash with warm water and a mild soap or  
detergent.  
• Ground-in dirt: scrub with powdered cleanser or similar detergent.  
• Chewing gum: scrape carefully; remove with kerosene or naphtha.  
• Nail polish and remover: Blot immediately to prevent damage.  
• Ball point pen: blot immediately with a white cloth dampened in water or mineral oil.  
• Paint, shoe polish, heel marks: remove immediately! Use a white cloth dampened in  
kerosene, naphtha or turpentine. Do not use paint remover or liquid brush cleaners.  
• Tars, asphalt, creosote: remove to prevent staining. Clean area with kerosene or  
naphtha.  
. WARNING:.When.using.kerosene.or.naphtha,.do.so.in.a.well.ventilated.area.and.keep.  
it.away.from.fire,.flame,.and.recliner.occupant.  
• Waxing and refinishing: improves wearability. Use any quality paste wax.  
• A diluted solution of 10% (or less) of bleach or Clorox may be used when cleaning this  
recliner.  
PERMABLOK3® Care and Cleaning Guide — Advanced Vinyl Protection  
• Step 1 (eye shadow, mascara, grease, black felt tip pen, crayon, chocolate, suntan lotion,  
hair and body oils): remove excess spill with a damp cloth. Clean with a 1:1 mix of  
Ivory® liquid soap and water. Next, rinse with clean water and dry.  
• Step 2 (red lipstick): use a straight application of concentrated cleaners such as Formula  
409® or Fantastik® spray cleaner. Wipe with a clean cloth, rinse with water, and dry.  
• Step 3 (blood, urine, feces, oil base paint, tar, and asphalt): use a 1:1 mix of ammonia and  
water, or a 1:4 mix of bleach and water. Rinse with clean water and dry.  
• Step 4 (all other tough stains): use a straight application of naphtha (lighter fluid). Rinse  
thoroughly with clean water and pat surface dry. If the stain persists, go to step 5.  
• Step 5: use a 1:1 mix of isopropyl alcohol and water. If the stain persists, use straight  
alcohol. Rinse thoroughly with clean water and pat dry.  
Info: For cleaning that requires step 4 or 5, use a soft cotton cloth saturated with  
the cleaning material and rub the stain in circles 10 times. Pat dry with another  
soft cotton cloth and check the results.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The information published in this care guide refers to the performance of PERMABLOK3  
products in specific tests conducted under laboratory conditions. Results may vary under  
actual conditions.  
This information is not a guarantee and does not relieve the user from the responsibility of  
the proper and safe use of the product and all cleaning agents. The use of certain agents can  
be harmful to the surface appearance and lifespan of vinyl. Spradling, its agents, and assigns  
assume no responsibility resulting from the use of such cleaning agents to the vinyl.  
Please check compatibility when using this product in combination with painted or  
varnished surfaces.  
PERMABLOK3 is a registered trademark of Spradling Holdings, Inc.  
Ivory is a registered trademark of Procter and Gamble.  
Formula 409 is a registered trademark of The Clorox Company.  
Fantastik is a registered trademark of S.C. Johnson & Son, Inc.  
Spradling International is not sponsored by, affiliated with, or endorsed by such trademark  
owners.  
CDC RECOMMENDATIONS  
CLEANING AND DISINFECTING STRATEGIES FOR ENVIRONMENTAL  
SURFACES IN PATIENT-CARE AREAS FOR VINYL UPHOLSTERY, WALL  
COVERINGS AND FLOORING  
• Select EPA-registered disinfectants, if available, and use them in accordance with the  
manufacturer’s instructions.  
• Do not use high-level disinfectants / liquid chemical sterilants for disinfection of either  
noncritical instruments and devices or any environmental surfaces; such use is counter  
to label instructions for these toxic chemicals.  
• Follow manufacturers’ instructions for cleaning and maintaining noncritical medical  
equipment.  
• In the absence of a manufacturer’s cleaning instructions, follow certain procedures:  
• Clean noncritical medical equipment surfaces with a detergent / disinfectant.  
• Do not use alcohol to disinfect large environmental surfaces.  
• Use barrier protective coverings as appropriate for noncritical surfaces that are:  
• Touched frequently with gloved hands during the delivery of patient care.  
• Likely to become contaminated with blood or body substances.  
• Difficult to clean (e.g., computer keyboards).  
• Keep housekeeping surfaces (e.g., floors, walls, tabletops) visibly clean on a regular  
basis and clean up spills promptly.  
• Use a one-step process and an EPA-registered hospital detergent / disinfectant designed  
for general housekeeping purposes in patient-care areas where:  
• Uncertainty exists as to the nature of the soil on the surfaces (whether it is blood or  
body fluid or just normal dust or dirt).  
• Uncertainty exists regarding the presence of multi drug resistant organisms on such  
surfaces.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
• Detergent and water are adequate for cleaning surfaces in nonpatient-care areas (e.g.,  
administrative offices). Clean and disinfect high-touch surfaces on a more frequent  
schedule than minimal-touch housekeeping surfaces. Clean walls, blinds, and window  
curtains in patient-care areas when they are visibly dusty or soiled.  
• Do not perform disinfectant fogging in patient-care areas.  
• Avoid large-surface cleaning methods that produce mists or aerosols, or disperse dust in  
patient-care areas.  
• Follow proper procedures for effective uses of mops, cloths, and solutions.  
• Prepare cleaning solutions daily or as needed, and replace with fresh solution  
frequently according to facility policies and procedures.  
• Change the mop head at the beginning of each day and also as required by facility  
policy, or after cleaning up large spills of blood or other body substances.  
• Clean mops and cloths after use and allow to dry before reuse; or use single-use,  
disposable mop heads and cloths.  
• When performing low- or intermediate-level disinfection of environmental surfaces in  
nurseries and neonatal units, avoid unnecessary exposure of neonates to disinfectant  
residues on these surfaces by using EPA-registered germicides in accordance with  
manufacturers’ instructions and safety advisories.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TECHNICAL SPECIFICATIONS  
Overall.Height  
50"  
33"  
36"  
23"  
Overall.Width  
Overall.Depth  
Width.Between.Arms  
Arm.Height.from.Floor  
Arm.Height.from.Seat  
Seat.Height  
1
26 /2"  
8"  
1
21 /2"  
1
Seat.Width  
23 /2"  
1
Seat.Depth  
20 /2"  
Depth.Fully.Reclined.in.Trendelenburg  
Width.with.Both.Side.Tables.in.Upright.Position  
Shipping.Weight  
76"  
53"  
1
138 /2 lb  
Maximum.Weight.Capacity  
450 lb, EVENLY DISTRIBUTED  
For complete technical specifications, please call the Graham-Field Technical Support  
Group at 770-447-1609.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
GF Health Products, Inc. (“Graham-Field”) warrants the FR587W Deluxe Extra Wide Clinical  
Care Recliner and FR587WH Deluxe Extra Wide Clinical Care Recliner with Heat and Massage  
against manufacturing defects in materials and workmanship as listed below:  
Welded Frame.................... Lifetime limited (ten years)  
Heat and Massage Unit..... Two year limited  
Upholstery ......................... One year limited  
All Other Components...... Three year limited  
If a product is deemed to be under warranty, GF Health Products, Inc. shall provide, at its  
option, (1) replacement of any defective part or product or (2) a credit of the original selling  
price made to GF Health Products, Inc.’s initial customer. The warranty does not include  
any labor charges incurred in replacement part(s) installation or any associated freight or  
shipping charges to GF Health Products, Inc.  
The warranties contained herein contain all the representations and warranties with  
respect to the subject matter of this document, and supersede all prior negotiations,  
agreements and understandings with respect thereto. The recipient of this document  
hereby acknowledges and represents that it has not relied on any representation, assertion,  
guarantee, warranty, collateral contract or other assurance, except those set out in this  
document.  
FR587W-INS-LAB-RevC13: Deluxe Extra-Wide Clinical Care Recliner Assembly and Operating Instructions  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
GF Health Products, Inc.  
2935 Northeast Parkway  
Atlanta, Georgia 30360 USA  
Telephone: 770-368-4700  
Fax: 770-368-2386  
Technical Support: 770-447-1609  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Leviton Stereo System 1x6 Passive Audio Module User Manual
LG Electronics Projection Television Z44SZ80 User Manual
Liebherr Beverage Dispenser WS 40 User Manual
Louisville Tin and Stove Furnace W251F User Manual
Lyngdorf Audio Speaker MH 1 User Manual
Mackie Musical Instrument Compact Mixer User Manual
Magnavox Flat Panel Television 15MF User Manual
Melissa Fan 271 001 User Manual
Melissa Scale 231 004 User Manual
Metra Electronics Automobile Accessories 99 3301 User Manual