Invacare Oxygen Equipment 20EX User Manual

Owner's Operator and  
Maintenance Manual  
DEMAND OXYGEN DELIVERY  
DEVICE AND SYSTEM  
Venture 20EX, 20 PSIG Device  
Venture 50EX, 50 PSIG Device  
Venture Pack 211EX, 20 PSIG Shoulder Bag System  
w/M6 Cylinder  
Venture Pack 212EX, 20 PSIG Shoulder Bag System  
w/M9 Cylinder  
DEALER: THIS MANUAL MUST BE GIVEN TO THE USER OF THIS  
DEMAND OXYGEN DELIVERY DEVICE.  
USER: BEFORE USING THIS DEMAND OXYGEN DELIVERY  
DEVICE, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
SAFETY SUMMARY ............................................. 4  
UNPACKING/HANDLING ..................................... 8  
INTRODUCTION ................................................ 10  
Invacare Venture Demand Oxygen  
Delivery Device ....................................................................................... 10  
FEATURES ......................................................... 11  
Venture Demand Oxygen Delivery Device  
Components and Definition of Terms ................................................. 11  
Venture Demand Oxygen Delivery Device System  
Components and Definition of Terms ................................................. 13  
SPECIFICATIONS ............................................... 14  
OPERATION ...................................................... 16  
The Use of Oxygen.................................................................................... 16  
Set-up and Normal Pulse Operation .................................................... 16  
MAINTENANCE................................................. 20  
Prolonging the Life of Your Battery....................................................... 20  
Care of Your Venture Demand Oxygen Delivery  
Device and System.................................................................................... 21  
Cleaning the Outer Case.......................................................................... 21  
Carrying Case .............................................................................................. 21  
ACCESSORIES AND SPARE PARTS ..................... 22  
TROUBLESHOOTING.......................................... 23  
OXYGEN CONSERVATION................................. 26  
Battery Use Time........................................................................................ 26  
Compatibility ............................................................................................... 26  
LIMITED WARRANTY ........................................ 27  
PART NO. 1093845  
3
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY  
WARNING  
This device should ONLY be used under the supervision  
of a physician. Verify your oxygen flowrate with your  
physician when you begin using your Demand Oxygen  
Delivery System.  
If you have any questions about the oxygen delivery  
system or its use, call your Home Health Care Dealer or  
Service Representative.  
If you suspect that an oxygen delivery system has been  
damaged, contact your Service Representative  
IMMEDIATELY. Your Service Representative MUST  
handle all maintenance.  
Keep all oxygen delivery systems out of the reach of  
children.  
Avoid creation of any spark near medical oxygen  
equipment. This includes sparks from static electricity  
created by any type of friction.  
All electrical equipment near oxygen units MUST be  
properly grounded.  
This system only delivers medical grade (U.S.P.)  
oxygen when used with oxygen cylinders marked as  
such. When using this system with Invacare Venture  
Homefill, oxygen delivery will be USP 93%+3%. Oxygen  
supplied by the system is supplemental and is NOT to  
be considered life-supporting.  
Before using the Demand Oxygen Delivery Device, check  
the contents gauge of the oxygen supply  
equipment. The oxygen supply equipment MUST be filled  
only by an authorized Service Representative, unless  
being filled by the Invacare Venture Homefill system.  
If at anytime the attendant or the end user concludes  
that an insufficient amount of oxygen is being received,  
VENTURE™  
4
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY  
WARNING (CONTINUED)  
the supplier and/or physician should be contacted  
IMMEDIATELY. No adjustments should be made to  
the flowrate unless prescribed by a physician.  
DO NOT attempt to alter the switches in the battery  
compartment. These switches are preset by your  
Home Health Care Dealer and control the breath  
delivery prescription.  
RISK OF FIRE:  
Oxygen does not burn. However, it will accelerate  
burning of flammable materials.  
TO PREVENT FIRE:  
DO NOT SMOKE when operating the oxygen delivery  
system. No Smoking signs should be prominently  
displayed.  
Keep oxygen delivery system at least 5-feet away from  
ANY electrical appliances; such as radios, TVs, air  
conditioners, fans, electric razors, etc.  
Keep the oxygen delivery system away from open  
flames or heating sources such as furnaces or stoves.  
Keep flammable materials away from the oxygen  
delivery system; such as oils, grease and aerosol sprays.  
These will ignite and burn quickly in the presence of  
concentrated oxygen.  
TO PREVENT HIGH CONCENTRATIONS OF OXYGEN:  
Keep the oxygen delivery system in a well ventilated  
area at all times.  
When the oxygen delivery system is not in use, press  
the OFF button to place the Demand Oxygen Delivery  
Device in the Continuous Mode, turn the Cylinder OFF,  
bleed the remaining oxygen from the Regulator, turn  
the Regulator OFF.  
PART NO. 1093845  
5
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY  
WARNING (CONTINUED)  
DO NOT wear the Demand Oxygen Delivery Device  
under any clothing.  
OTHER PRECAUTIONS:  
Never attempt to lubricate the oxygen delivery system.  
DO NOT insert any object (except suitable AC/DC  
Adapter) in external power jack of Demand Oxygen  
Delivery Device. This may cause an electrical shock  
which could cause burns.  
DO NOT subject the Demand Oxygen Delivery Device to  
temperatures below -40°F (-40°C) or above 158°F (70°C).  
To do so could damage the electronics and battery.  
NEVER tamper with the oxygen delivery system.  
Doing so could create a hazardous condition and  
possibly cause the oxygen delivery system to malfunction.  
If you have any questions, call your Home Health Care  
Dealer or Service Representative.  
ALWAYS connect the Demand Oxygen Delivery  
Device to a pressure regulated oxygen source.  
ONLY connect 20 psig Demand Oxygen Delivery  
Device to a 20 psig oxygen source and a 50 psig  
Demand Oxygen Delivery Device to a 50 psig oxygen  
source.  
DO NOT use any additional lengths of tubing between the  
oxygen source and the Demand Oxygen Delivery Device.  
ONLY use Invacare tubing supplied with the Demand  
Oxygen Delivery Device, if any tubing is required.  
DO NOT use a mask when using pulse mode delivery,  
USE ONLY a standard nasal cannula.  
When using the continuous flow delivery, a mask or  
nasal cannula can be used.  
VENTURE™  
6
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY  
WARNING (CONTINUED)  
DO NOT use a humidifier with the Demand Oxygen  
Delivery Device.  
DO NOT expose the Demand Oxygen Delivery Device  
or system to water or any other liquids.  
RADIO FREQUENCY INTERFERENCE  
Most electronic equipment is influenced by Radio Frequency  
Interference (RFI). CAUTION should be exercised with  
regard to the use of portable communications equipment in  
theareaaroundsuchequipment.  
PART NO. 1093845  
7
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
UNPACKING / HANDLING  
UNPACKING  
1. Check for any obvious damage to the carton or its contents.  
(If damage is evident, notify your local Invacare Home Health  
Care Dealer/Carrier.)  
2. Remove all the loose packing from the carton.  
3. Carefully remove all the components from the carton.  
The following is a listing of items included in the device and  
system packages:  
VENTURE50EX,  
MODELIDD50EX  
50 psig Demand Device with  
Threaded Fitting Battery  
Venture Operator's Manual.  
VENTURE20EX,  
MODELIDD20EX  
20 psig Demand Device with Barb  
Fitting and Connection Tubing  
Battery  
Venture Operator's Manual Optional  
Threaded Fitting.  
VENTURE PACK 211EX,  
MODELIDD211EX  
IDD20EX Demand Device With  
Connecting Tube  
AND VENTURE PACK 212EX,  
MODELIDD212EX  
20 psig Piston Type Flow Control  
Regulator (0-6 L/min.)  
Carrying Case: Dual Size Hunter  
Green Shoulder Bag  
Cylinder:  
M6 Cylinder (IDD211EX)  
M9 Cylinder (IDD212EX)  
Nasal Cannula, Cylinder Wrench,  
VENTURE™  
8
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
HANDLING PROCEDURES  
NOTE: Unless the Venture Demand Oxygen Delivery Device is to be  
used immediately, retain boxes, containers and packing materials  
for use in returning or storing the equipment until usage is required.  
INSPECTION  
1. Examine exterior of the equipment for nicks, dents,  
scratches or other damages. Inspect all components.  
STORAGE  
1. Carefully repackage all components in the original packaging.  
2. Store the repackaged equipment in a dry area.  
3. DO NOT place other objects on top of the repackaged  
equipment.  
PART NO. 1093845  
9
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INTRODUCTION  
INVACARE VENTURE DEMAND OXYGEN  
DELIVERY DEVICE  
Ordinary oxygen supply units deliver oxygen continuously.  
The normal breathing pattern is to inhale about one-third of the time,  
and exhale two-thirds of the time. This means that continuous flow  
units may be inefficient during user exhalation.  
In contrast, the Venture Demand Oxygen Delivery Device (Venture  
DODD) senses your inhalation and instantly releases a programmed  
pulse of oxygen at a specified flowrate at the beginning of inhalation.  
The specified flowrate of oxygen is delivered within the first second of  
inhalation. Oxygen is not delivered during exhalation. This means that  
your oxygen use time is extended by the ratio of breath duration over  
the pulse duration. Venture DODD will extend user time an average of  
four (4) times in the Variable duration breath mode.  
Comfort is another advantage of the Venture DODD. Nasal drying caused  
by continuous flow is eliminated, as oxygen is only delivered during  
inhalation. Therefore, additional humidification is not necessary. Since the  
oxygen is delivered at normal flowrates, there is not a high blast of oxygen  
to the end user, which makes the Venture DODD more comfortable than  
other pulse systems.  
The Venture DODD can be used in the traditional continuous flow mode  
simply by turning the Venture DODD device OFF. Effective flow deliveries of  
0.5 to 6 liters per minute (L/min.) can be instantly selected, in pulse or  
continuous mode, by each individual as prescribed by your physician.  
During pulse mode, the Venture DODD runs on a disposable D-cell battery  
which must be replaced after approximately 80 hours of normal use  
(Refer to the OPERATION section in this manual for information on the  
Low Battery Indicator LED). During continuous mode operation, the unit  
does not need the battery to operate.  
NOTE: Actual Battery use time depends on your breathing rate.  
The Venture DODD can be used with compressed gas or liquid  
oxygen systems. Both 20 and 50 psig units were designed for use on  
20 and 50 psig oxygen delivery systems, respectively.  
NOTE: The 50 psig system should ONLY be used with the threaded (DISS)  
fitting, and NOT the 12-inch connecting tube, as the pressure may be too  
great for the tube to stay connected.  
VENTURE™  
10  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FEATURES  
VENTURE DEMAND OXYGEN DELIVERY  
DEVICE COMPONENTS AND DEFINITION  
OF TERMS (FIGURE 1)  
ON/OFF Buttons - Determines flow selection method. The ON  
button turns Power ON for normal pulse operation (  
the OFF button disengages the electronics for continuous flow  
operation ( ).  
) and  
NOTE: The OFF button does NOT stop flow delivery and is NOT  
recommended for unit shutdown. See SETUP AND NORMAL PULSE  
OPERATION in the OPERATION Section of this manual.  
Oxygen OUTLET Fitting - The delivery point of oxygen to the  
end user from the Venture DODD. The fitting is barbed to  
securely hold the cannula for oxygen delivery.  
Oxygen INLET Fitting - The delivery point of oxygen from the  
oxygen system to the Venture DODD.  
External Power Input Jack - AC/DC Power Adapter connection port.  
Status Indicators:  
O2 - GREEN LED will illuminate with each breath indicating  
that the Venture DODD is operating.  
- Attention - YELLOW LED illumination indicates the  
Venture DODD is NOT sensing end user's inhalation or  
that there is an internal malfunction.  
Refer to TROUBLESHOOTING Section.  
- Low Battery - RED LED indicates two (2) low battery  
levels.  
O2 - Low/No Gas - GREEN LED illumination indicates  
Venture DODD is NOT sensing O2 presence in pulse  
delivery.  
PART NO. 1093845  
11  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FEATURES  
External Power  
Input Jack  
ç
é
é
Oxygen OUTLET Fitting Oxygen INLET Fitting  
Low O2 LED (GREEN)  
Battery LED (RED)  
Attention LED (YELLOW)  
Pulse LED (GREEN)  
êêêê  
ON(Pulse ModeOperation)  
è
é
OFF (Continuous Mode Operation)  
NOTE: Battery Compartment located on  
bottom back of Venture Demand Oxygen  
Delivery Device.  
è
FIGURE 1 - VENTURE DODD FEATURES  
VENTURE™  
12  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FEATURES  
VENTURE DODD SYSTEM COMPONENTS AND  
DEFINITION OF TERMS (FIGURE 2)  
Cylinder - Lightweight aluminum cylinder provided.  
Regulator - Provides flowrate control of oxygen delivery with  
compressed gas cylinders.  
DISS Fitting - Connection point for direct connect to oxygen source.  
Barb Fitting - Connection point for tubing used with compressed  
gas cylinders.  
Connection Tubing - Provides oxygen path from regulator to  
the device inlet barb fitting for use with compressed gas cylinders.  
Carrying Case with Padded Strap - Used for convenient  
transportation and is adjustable to ensure comfort.  
Optional AC/DC Adpater - Designed to provide AC current to device.  
Barb Fitting  
Regulator  
DISS Fitting  
M9 Cylinder  
Tubing  
M6 Cylinder  
Case  
Optional AC/DC  
Adapter  
(Cannula and Wrench not shown)  
FIGURE 2 - VENTURE DODD SYSTEM FEATURES  
PART NO. 1093845  
13  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
OPERATING VOLTAGES  
EXTERNAL AC/DC POWER  
ADAPTER:  
Output: 12 + 3 VDC, 100 mA  
Connector: 5.5mm OD x 2.1mm  
ID x 12.0mm No. 2 Connector  
ALARMS  
Low Battery (Two [2] levels)  
Cannula Disconnect (Breath not  
sensed)  
Device Malfunction  
Low / No Gas Indicator  
INDICATORS/SYMBOLS:  
O2 LED:  
Pulse Delivery  
LED:  
LED:  
Cannula Disconnect/Missing  
Breath Indicator, Audible Alarm  
Battery Low Charge Indicator (1/2 to 2  
hours of use time left), Audible Alarm  
O2  
LED:  
Low or No Gas, Audible Alarm  
Class BF  
:
BATTERY:  
Approx. Battery Usage Time:  
Alkaline: (1) "D" size cell  
80 Hours @ 20 bpm  
(Variable Breath Mode)  
OXYGEN DOSE DURATION:  
1-variable duration breathing:  
Auto-adjusts between 0.5 and 1.0  
seconds based on patient breathing rate.  
3-fixed duration breath mode:  
Fixed 1.0 seconds, 0.75 seconds,  
0.50 seconds  
FLOWRATERANGE(L/Min.):  
0 to 6 L/Min.  
OXYGENSOURCEPRESSURE:  
Venture 20:  
Venture 50:  
20 + 3 psig (138 + 20 kPa)  
50 + 5 psig (345 + 35 kPa)  
REGULATOR  
Flow Settings:  
0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,  
5.0, 6.0 L/Min.  
Regulated Pressure:  
20 psig  
VENTURE™  
14  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
ENVIRONMENTAL  
Operating Temperature:  
Storage Temperature:  
40° to 104° F (5° to 40°C)  
- 40° to 158°F (-40° to 70°C)  
DEVICE DIMENSIONS  
Length:  
Thickness:  
Width:  
Weight:  
5.4-inches (13.7cm)  
1.6-inches (4.0cm)  
3.6-inches (9.1cm)  
16.0 ozs. (454gm)  
PART NO. 1093845  
15  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
THE USE OF OXYGEN  
The type of oxygen administrating equipment to be used, as well as the  
flowrate and time of usage, must be prescribed by a physician. Your  
physician should verify your Venture DODD flowrate prescription.  
WARNING  
If at anytime the attendant or the end user concludes  
that an insufficient amount of oxygen is being received,  
the supplier and/or physician should be contacted im-  
mediately. No adjustments should be made to the  
flowrate unless prescribed by a physician.  
NOTE: ONLY a NASAL CANNULA can be used in the PULSE MODE to  
deliver oxygen. An oxygen mask or nasal cannula can be used with  
Continuous Flow to deliver oxygen. Low flow cannula should not be used.  
SET-UP AND NORMAL PULSE OPERATION  
1. Install the battery into the Venture DODD or plug AC/DC  
power adapter into external power jack.  
A. To Install or Replace the Battery:  
Open the battery cover on the back of the Venture  
DODD by applying light inward pressure with your thumb  
to the cover latch and then gently lift the cover open.  
If replacing a battery: Remove the used battery from the  
battery compartment by grasping the top and bottom ends  
and pulling straight out.  
To install the battery: Orient the new battery so that its  
polarity matches the polarity marked in the battery compartment.  
WARNING  
DONOTattempttoaltertheswitchesinthebattery  
compartment.Theseswitchesarepresetbyyourhome  
healthcareproviderandcontrolthebreathdeliveryprescription.  
Press the new battery in place and close the battery cover.  
2. Check Oxygen Source to verify adequate oxygen supply.  
VENTURE™  
16  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
3. If using an oxygen cylinder, check that the regulator is  
properly attached before opening the cylinder valve.  
4. Connect the Venture DODD to the regulator by performing  
one (1) of the following:  
A. If using an Oxygen Cylinder and Regulator:  
On a 20 psig regulator, connect the tubing (supplied with  
the Venture DODD) from the barb fitting of the regulator  
to the barb inlet fitting on the Venture DODD.  
On a 50 psig regulator, connect the threaded (DISS) fitting  
(supplied with the Venture DODD) to the regulator DISS fitting.  
NOTE: The 50 psig system should ONLY be used with the threaded  
(DISS) fitting and not the 12-inch connecting tube, as the pressure may  
be too great for the tube to stay connected.  
B. If using a Liquid Oxygen Reservoir:  
Directly connect the Venture DODD to the oxygen  
source using a DISS fitting.  
5. Connect the nasal cannula tubing to the barbed oxygen outlet  
fitting on the Venture DODD and adjust cannula for comfort.  
WARNING  
When using nasal or masked type administering  
equipment, oxygen or air tubing MUST be routed and  
secured properly to ensure that the tubing does NOT  
become entangled, kinked or severed.  
NOTE: Use ONLY a NASAL CANNULA with PULSE MODE  
delivery, NOT a mask. A mask may NOT fit your face well enough for the  
unit to sense your inhalation.  
NOTE: A special nasal cannula is NOT required. Any standard flow  
crush-proof nasal cannula can be used.  
NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.  
NOTE: The cannula tubing can be up to 50 feet long to function properly  
in the pulse mode. The continuous flow mode is NOT affected by the  
length of the canula tubing.  
PART NO. 1093845  
17  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
6. Set the prescribed flow rate on the oxygen source by performing  
one (1) of the following:  
A. If using an Oxygen Cylinder and Regulator:  
OPEN cylinder valve and adjust regulator to prescribed flow rate.  
B. If using a Liquid Oxygen Reservoir:  
Set flow control knob on reservoir to prescribed flowrate.  
7. Press the ON button to engage pulse mode. The unit will  
pulse each time the user inhales, the GREEN PULSE LED will  
blink with each oxygen pulse delivery.  
8. As battery power depletes, the unit will alarm to inform the end  
user to replace the battery or switch to external power source.  
A. INITIAL BATT. LED - The Red LED will FLASH and the  
audible alarm will BEEP three (3) times. This indicates that  
you have about two (2) hours of use time left on your battery.  
B. SECOND BATT. LED - The Red LED will be SOLID and  
the audible alarm will BEEP six (6) times. This indicates that you  
have between 15 and 30 minutes of use time left on your battery.  
NOTE: Remaining battery use times are dependent on battery type  
being used. Alkaline battery type is recommended for longest life.  
OFF  
(Continuous Mode  
Operation)  
FIGURE 3 - SWITCHING TO CONTINUOUS FLOW  
VENTURE™  
18  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
NOTE: Whenever the pulse mode is used, the battery is monitored.  
If the SECOND BATT. LED illuminates, either replace the battery or  
switch to continuous mode if oxygen supply is adequate. When in  
continuous mode, the battery is NOT used.  
NOTE: When in the continuous flow mode, the oxygen supply will be  
used considerably faster then in the pulse mode. If using a portable  
oxygen supply, make plans for an alternative source of oxygen. Replace  
the battery as soon as possible.  
9. If it is necessary to switch to continuous flow at any time,  
press the OFF button (FIGURE 3).  
10. If turning the system OFF for cylinder replacement or storage,  
press the Venture DODD OFF button. Turn OFF the cylinder  
post valve, bleed the oxygen from the regulator and turn OFF  
the regulator prior to any disassembly.  
PART NO. 1093845  
19  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MAINTENANCE  
PROLONGING THE LIFE OF YOUR BATTERY  
The Venture DODD battery is powered by a single D-cell battery.  
The Alkaline battery type is recommended for longest life.  
NOTE: The Alkaline battery should last for approximately 80 hours with  
normal use, but will last even longer if you follow this procedure:  
1. Use the battery to its fullest capacity prior to replacing. This is  
accomplished by running the unit from the battery until it stops  
working (approximately 15 to 30 minutes after the SECOND  
BATT. indicator illuminates SOLID and the audible alarm  
sounds six [6] times).  
CAUTION  
DO NOT store Venture DODD with the battery installed.  
WARNING  
DO NOT recharge, dispose of in fire, mix with different  
battery type, connect improperly or disassemble  
Alkaline batteries. Doing so may cause battery to  
explode or leak resulting in injury.  
2. Operate your unit from the AC/DC power adapter when at home.  
3. Avoid exposure to extreme temperatures.  
WARNING  
DO NOT operate the unit in temperatures above  
104°F (40°C) or below 40°F (4°C), or place near heat  
sources such as radiators or fires.  
Electrical shock hazard. DO NOT remove outer case  
of this unit. All disassembly and maintenance of this  
unit MUST BE done by an Invacare Dealer. If outer  
case is removed by anyone EXCEPT an Invacare  
Dealer the warranty will be voided.  
VENTURE™  
20  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MAINTENANCE  
CARE OF YOUR VENTURE DEMAND  
OXYGEN DELIVERY DEVICE AND SYSTEM  
CLEANING THE OUTER CASE  
WARNING  
DO NOT submerge in water to clean.  
DO NOT open outer case. If outer case is removed by  
anyone EXCEPT an Invacare Dealer the warranty will  
be voided.  
CAUTION  
Use no oil. Never apply waxes, polishes, or other similar  
compounds to any surface of the Venture Demand  
Oxygen Delivery Device, as these compounds may be  
flammable.  
Clean outer case with damp cloth. If necessary, a mild dish  
detergent may be used. Dry outer case with lint free cloth.  
CARRYING CASE  
The carrying case can be hand washed in warm water. Air dry  
ONLY. NO bleach. DO NOT dryclean.  
PART NO. 1093845  
21  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ACCESSORIES AND SPARE PARTS  
M6 Aluminum Cylinder, 164L Capacity  
M9 Aluminum Cylinder, 248L Capacity  
Piston Regulator, 0.5-6 L/Min., 20 psig, CGA 870  
Shoulder Bag, Hunter Green  
(Dual size for Cylinder diameters of 3 and 4-inches)  
Venture Device Pouch  
12-inch Connection Tubing  
Barb Fitting, 1/4-28 UNF  
7-Foot (2.1m) Nasal Cannula  
AC/DC Adapter  
Venture Operator's Manual  
NOTE: Rechargeable batteries can be used to power Venture DODD,  
however the batteries and charger are a user supply item not provided  
by Invacare.  
NOTE: If you have any questions regarding accessories, please call  
Invacare or your Home Healthcare Dealer.  
VENTURE™  
22  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING  
PROBLEM  
POSSIBLE CAUSE  
SOLUTIONS  
Indicates Oxygen Pulse Flow  
in normal pulse delivery.  
O2 LED - (GREEN)  
Blinking.  
This is normal  
operation.  
Indicates Venture DODD  
not sensing presence of  
oxygen in pulse delivery.  
O2 LED - (GREEN)  
Turn unit OFF, check  
cylinder pressure is above  
refill limit (500 psig) and  
post valve is open, check  
regulator is set to correct  
flowrate. If problem still  
exists, call your Home  
Health Care Dealer or  
Service Representative  
IMMEDIATELY and go to  
continuous mode if  
oxygen supply is adequate.  
Indicates oxygen pulse  
delivery, but Venture DODD  
not sensing inhalation.  
O2 and  
(GREEN and YELLOW)  
Blinking  
LED -  
Breathe normally. Check  
cannula for kinking and  
check tubing connection.  
Indicates no pulse  
delivery, Venture DODD  
not sensing inhalation.  
Breathe normally. Check  
cannula for kinking and  
check tubing connection.  
LED - (YELLOW)  
Blinking and  
Intermittent Beep  
Possible internal problem.  
End user possibly not  
receiving oxygen.  
LED - (YELLOW)  
Solid and Continuous  
Beep  
Turn unit off then on  
again. If problem still  
exists, call your Home  
Health Care Dealer or  
Service Representative  
IMMEDIATELY and go to  
continuous mode if  
oxygen supply is adequate.  
LED - (RED)  
Battery low.  
Replace battery as soon  
as possible.  
Blinking, Three (3) Beeps Approximately two (2) hours  
at Beginning.  
of battery use time left.  
Approximately 30 minutes  
of battery use time left.  
LED - (RED) Solid,  
Six (6) Beeps at Beginning.  
Replace battery or turn  
unit OFF to enable  
continuous mode if  
oxygen supply is adequate.  
PART NO. 1093845  
23  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING  
PROBLEM  
POSSIBLE CAUSE  
SOLUTIONS  
Possible internal sensor  
problem. End user  
possibly not receiving  
oxygen.  
Turn unit off then on  
again. If problem still  
exists, call your Home  
Health Care Dealer or  
Service Representative  
IMMEDIATELY and go to  
continuous mode if  
LED - (RED) Solid,  
Continuous Beep.  
oxygen supply is adequate.  
No lights illuminate or  
RED and YELLOW LEDs  
on together.  
Electronic failure.  
Call your Home Health  
Care Dealer or Service  
Representative  
IMMEDIATELY and go  
to continuous mode if  
oxygen supply is  
adequate.  
Check contents gauge on  
oxygen source. Call your  
Home Health Care Dealer  
or Service Representative  
IMMEDIATELY.  
Oxygen supply exhausted.  
Oxygen NOT being  
delivered.  
Cylinders - OPEN valve.  
Liquid Oxygen Supply -  
Make sure coupling is  
FULLY engaged.  
Oxygen supply NOT  
turned ON.  
Check all cannula  
connections.  
Cannula kinked or  
crushed.  
Turn regulator knob to  
prescribed flowrate.  
Regulator knob NOT  
turned ON.  
Call your Home Health  
Care Dealer or Service  
Representative  
Internal Problem.  
IMMEDIATELY.  
External input will not  
power device.  
Push plug in all the way.  
Connector jack not fully  
inserted into unit.  
Wall outlet bad or no  
power.  
Check that outlet power  
switch is on. Check outlet  
by plugging in another  
appliance or lamp.  
VENTURE™  
24  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING  
PROBLEM  
POSSIBLE CAUSE  
SOLUTIONS  
External input will not  
power device (Cont.).  
Incorrect power adapter.  
Check that power adapter  
used within specification for  
operating voltage of external  
power supply. Call your  
Home Health Care Dealer  
or Service Representative  
IMMEDIATELY.  
Check battery  
"Low" battery light  
Battery life expired.  
manufacturer's  
comes on too soon after  
battery replacement.  
expiration date. See  
Prolonging the Life of the  
Battery in the MAINTE-  
NANCE Section of this  
manual. Call your Home  
Health Care Dealer or  
Service Representative  
IMMEDIATELY.  
External power adapter  
plug does not go into  
jack in unit.  
Jack or plug damaged.  
Call your Home Health  
Care Dealer or Service  
Representative  
IMMEDIATELY.  
Non-compatible adapter  
plug.  
Refer to SPECIFICATION  
Section for external  
power adapter connector  
size.  
Check if YELLOW and  
GREEN O2 LEDS illumi-  
nate. If so, see indicator  
description in this section.  
Call your Home Health  
Care Dealer or Service  
Representative  
Unit pulses at times  
other than inhalation.  
Unit may be in backup  
Pulse Mode.  
IMMEDIATELY.  
Verify compatibility of  
model and pressure.  
Call your Home Health  
Care Dealer or Service  
Representative  
Operating pressure of  
oxygen supply source too  
high.  
Continuous flow out of  
cannula when in pulse  
mode.  
IMMEDIATELY.  
NOTE: If any of the above problems occur and the solutions do not work,  
please contact your Invacare Home Health Care Dealer for service.  
PART NO. 1093845  
25  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OXYGEN CONSERVATION  
OXYGEN CONSERVED AND BATTERY  
USE TIME BASED ON BREATH RATE  
BREATH RATE (BPM)  
% OXYGENCONSERVED  
Variable Breath  
Fixed 1.0 sec. Breath Mode:  
Fixed 0.75  
Fixed 0.50 sec. Breath Mode:  
10  
20  
30  
40  
Mode:  
8277  
70  
63  
75  
83  
Breath  
91  
66  
Mode:  
83  
50  
75  
33  
87  
66  
sec.  
i.e. 77% Oxygen conservation over Continuous Flow achieved  
with Venture DODD in Variable Breath Mode @ 20 BPM Rate.  
Oxygen supply lasts 4.3-times longer. Venture DODDresponds to  
individual breath rates. Therefore, use times stated are only  
approximates. Actual savings depends on individual breathing  
patterns.  
COMPATIBILITY  
Venture DODD requires using a flowrate controlled oxygen  
source. Gaseous or liquid oxygen systems that meet the  
following requirements are compatible to the Venture  
DODD:  
Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source.  
Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source.  
Flowrate Range: 0-6 L/Min.  
VENTURE™  
26  
PART NO. 1093845  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED  
TO COMPLY WITH FEDERAL LAW APPLICABLE TO  
PRODUCTS MANUFACTURED AFTER JULY 4, 1975.  
This warranty is extended only to the original purchaser/user of  
our products.  
This warranty gives you specific legal rights and you may also have  
other legal rights which vary from state to state.  
Invacare warrants the Demand Oxygen Delivery Device to be free  
from defects in materials and workmanship for a period of one (1)  
year from date of purchase. In addition, all internal electronic  
components and switches carry a three (3) year warranty from  
date of purchase. If within such warranty period any such product  
shall be proven to Invacare satisfaction to be defective, product  
shall be repaired or replaced, at Invacare option. This warranty  
does not include normal wear and tear or shipping charges  
incurred in replacement part installation or repair of such  
product. Invacare sole obligation and your exclusive remedy  
under this warranty shall be limited to such repair and/or  
replacement.  
For warranty service, please contact Invacare Service Department  
at the Toll Free Number listed below during normal hours.  
Upon receiving notice of an alleged defect in a product, Invacare  
will issue a serialized return authorization. It shall be the  
responsibility of the purchaser to return, at the purchaser’s cost,  
to either Invacare’s plant or service center as specified by Invacare  
in advance. Defective components part(s) must be returned for  
warranty inspection using the serial number for identification  
within 30 days of the return authorization date. DO NOT return  
products to our factory without prior consent. C.O.D. shipments  
will be refused - please prepay shipping charges.  
LIMITATIONSANDEXCLUSIONS:THEFOREGOING  
WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PROD-  
UCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DE-  
FACED, PRODUCTSSUBJECTEDTONEGLIGENCE,  
ACCIDENT, IMPROPEROPERATION, MAINTENANCEORSTOR-  
AGE, PRODUCTSMODIFIEDWITHOUTINVACARE’SEXPRESS  
WRITTENCONSENT(INCLUDING, BUTNOT  
LIMITEDTO, MODIFICATIONTHROUGHTHEUSEOF  
UNAUTHORIZEDPARTSORATTACHMENTS;PRODUCTS  
DAMAGED BY REASON OF REPAIRS MADE TO ANY  
COMPONENTWITHOUTTHESPECIFICCONSENTOF  
INVACARE, OR TO A PRODUCT DAMAGED BY  
CIRCUMSTANCESBEYONDINVACARERESPIRATORY -  
PART NO. 1093845  
27  
VENTURE™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
CORPORATION’SCONTROL, ANDSUCHEVALUATIONWILLBE  
SOLELYDETERMINEDBYINVACARE. THEWARRANTYSHALL  
NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR  
FAILURETOADHERETOTHEFOLLOWINGINSTRUCTIONS.  
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN  
LIEU OF ANY OTHER WARRANTIES WHATSOEVER,  
WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED  
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A  
PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR  
VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE  
LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE  
PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.  
THE APPLICATION OF ANY IMPLIED WARRANTY  
WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION  
OF THE EXPRESS WARRANTY PROVIDED HEREIN.  
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL  
OR INCIDENTAL DAMAGES WHATSOEVER.  
SOME STATES DO NOT ALLOW THE EXCLUSION OR  
LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE,  
OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY  
LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY  
NOT APPLY TO YOU.  
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH  
STATE/PROVINCIAL LAWS AND REQUIREMENTS.  
Invacare Corporation  
_____________________________________________________________________________________________________________________  
USA  
One Invacare Way  
Elyria, Ohio USA  
44036-2125  
Invacare is a registered trademark  
of Invacare Corporation.  
Yes, you can. is a trademark of  
Invacare Corporation.  
800-333-6900  
©2002 Invacare Corporation  
Canada  
Part No. 1093845  
Rev C 10/02  
570 Matheson Blvd E Unit 8  
Mississauga Ontario  
L4Z 4G4 Canada  
800-668-5324  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Indesit Washer Dryer WA 115 User Manual
Ingersoll Rand Heat Pump AD25A X X User Manual
JBL Speaker SVA1500 User Manual
JVC Computer Monitor DT V17L2D User Manual
Kenmore Water System 6253485400 User Manual
KitchenAid Clothes Dryer KEYE800S User Manual
KitchenAid Ice Maker KUIS155HPB4 User Manual
Kodak Projector E PLUS SLIDE User Manual
Konica Minolta Printer C351 User Manual
Korg Electronic Keyboard LP 350 User Manual