| INSTRUCTION MANUAL   MODEL 212R   THERMAL CONDUCTIVITY   ANALYZER   DANGER   HIGHLYTOXICANDORFLAMMABLELIQUIDSORGASESMAYBEPRESENTINTHISMONITORINGSYSTEM.   PERSONALPROTECTIVEEQUIPMENTMAYBEREQUIREDWHENSERVICINGTHISSYSTEM.   HAZARDOUSVOLTAGESEXISTONCERTAINCOMPONENTSINTERNALLYWHICHMAYPERSISTFORA   TIMEEVENAFTERTHEPOWERISTURNEDOFFANDDISCONNECTED.   P/NM73212   11/29/07   ECO#07-0182   ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE   CONDUCTINGANYMAINTENANCEORSERVICINGCONSULTWITHAUTHORIZEDSUPERVISOR/MANAGER.   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Table of Contents   Model 212R   1 Introduction   1.1 Method of Analysis.......................................................5   1.2 Sensitivity ....................................................................6   1.3 Stability ........................................................................6   1.4 Special Consideration .................................................6   1.5 Physical Configuration.................................................6   2 Installation   2.1 Location .......................................................................7   2.2 Electrical Requirements & Connections ......................7   2.2.1 Primary Power ...............................................8   2.2.2 Signal Output .................................................8   2.2.3 Regulating Transformer..................................8   2.2.4 Completion & Inspection................................9   2.3 Gas Requirements & Connections...............................9   2.3.1 Reference Gas...............................................10   2.3.2 Zero Gas ........................................................10   2.3.3 Span Gas.......................................................10   2.3.4 Installation of Cylinder Supplies ....................11   2.3.5 Sample Pressure ...........................................11   2.3.6 Interconnecting Lines.....................................11   2.3.7 Vent Lines ......................................................12   3 Startup   3.1 Preliminary...................................................................12   3.2 Reference Gas Flow ....................................................13   3.3 Zero Gas Flow .............................................................13   3.4 Warmup........................................................................13   3.5 Zero Standardization ...................................................13   3.6 Span Standardization ..................................................14   3.7 Bypass .........................................................................14   3.8 Sample Mode...............................................................15   4 Routine Operation   4.1 Flowrate .......................................................................15   4.2 Supporting Gas Supplies.............................................16   3 Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   4.3 Standardization............................................................16   5 Periodic Maintenance   5.1 Heater Fan ...................................................................17   6 Troubleshooting   6.1 Preliminary...................................................................17   6.1.1 Electrical Checks ...........................................17   6.1.2 Sampling System Checks..............................18   6.1.3 Summary of Preliminary Checks....................18   6.2 Loss of Zero Control.....................................................18   6.2.1 Dynamic Balance Procedure .........................19   6.3 Correct Operation.........................................................20   6.4 Incorrect Operation.......................................................21   6.4.1 Analyzer Leak Check.....................................21   6.4.2 Temperature Control Check ...........................22   7 Calibration Data   7.1 Ranges ........................................................................24   7.2 Output Signal ...............................................................25   7.3 Span Setting ................................................................25   7.4 Recommended Accessory Gases................................25   7.4.1 Reference Gas...............................................25   7.4.2 Zero Gas ........................................................25   7.4.3 Span Gas.......................................................26   Appendix   A Recommended 2-Year Spare Parts List .......................26   A Drawing List .................................................................27   A Specifications...................................................................... 28   Teledyne Analytical Instruments   4 Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   1.   INTRODUCTION   1.1   Method of Analysis.   The Model 212R compares the thermal conductivity of a sample gas to   that of a fixed composition reference gas and produces an electrical output   signal that is calibrated to represent the difference between the two gases.   Due to the nonspecific nature of thermal conductivity measurement,   standard gases of known composition will be required to calibrate the ana-   lyzer. The accuracy of the analysis will be dependent on the accuracy to   which the composition of the standard gases is known.   The measuring unit is a four element hot wire cell that forms one-half of   an alternating current bridge circuit. Two of the hot wire elements are ex-   posed to the sample gas, and two to the reference gas. The other half of the   bridge circuit is formed by the center tapped secondary winding of a trans-   former.   With reference and zero standardization gas flowing, the bridge circuit   is balanced at one end of the measurement range. A span standardization gas   containing a known concentration of the component of interest is then   introduced into the sample path, and the resulting error signal generated by   the now unbalanced bridge circuit is calibrated to represent the span gas   mixture. The concentration of the component of interest in the span gas is   predicated by the specified ranges of the analysis. After the instrument has   been standardized the electrical error signal is directly related to the compo-   nent of interest content of the sample gas.   The magnitude of the measuring bridge error signal is much too small to   drive an indicating or recording instrument. A 100:1 step-up transformer,   followed by an electronic amplifier stage, is utilized to amplify the error   signal to an acceptable amplitude for demodulation. The signal is then   conditionedappropriatelytodriverecordingequipment.   5 Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   1.2   Sensitivity.   Differencesinthermalconductivitythatproduceameasuringbridgeerror   signalof 0.25microvoltswillbesensedbytheanalyzer.   1.3   Stability.   Internalvariables,otherthanthesamplegas,thatcouldproducevariations   intheoutputsignalarecarefullycontrolledbytheModel212R. Zerodriftis   approximately2%ofthefinerangeina24hourperiodofoperation;further-   more,thedriftisbidirectionalwithaproperlyinstalledanalyzerandisrelatedto   grosschangesinambienttemperature. Thevariables,bothinternalandexternal,   thatcanaffectthestabilityoftheanalyzerwillbedealtwithindetailintheinstal-   lationandtroubleshootingsectionsofthismanual.   1.4   Special Consideration.   ConsiderationofUsingthe212RtoMeasureH2,ormeasureothercompounds   in H2. H2 in the gas state can be assume of one tow states, para H2, or ortho   H2. EachhasitsownThermalConductivityValue. Thesedifferbyabout10%   fromoneanother.   Sothismustbetakenintoconsiderationwhenattemptingtousethe212RonH2   streams.   WewouldrecommendoneconsultWikipedia.comorothersourcestolearn   moreaboutthesetwostatesofH2andhowthethermalconductivityofthegas   streamvarieswiththestate.   IfforinstanceoneisusingpureH2asareferencegas,andthisgasisinthepara   stateassociatedwithcryogenicH2recentlyvaporized,andthisisthencompared   tocylinderH2whichmaybeinadifferentstate,thenconsiderablemeasurement   errorscanresult.   ContactTeledyneforfurtherguidanceorinformationonyourspecificapplica-   tion.   1.5   Physical Configuration.   Theanalyzerishousedinasheetsteelcasethatisdesignedtoflushmount   withinanequipmentpanel. Electricalcontrols,aswellasanintegralgascontrol   Teledyne Analytical Instruments   6 Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   panel,arelocatedimmediatelybehindahingedfrontaccessdoor.Theanalyzer   issuitableforinstallationinashelterednon-hazardousarea.   Arecordingand/orindicatingdevicewillberequiredtotransducethe   electricaloutputsignalintoreadableinformation.   2.   INSTALLATION   2.1   Location.   Theanalyzershouldbepanelmountedinanuprightpositioninanarea that   is notexposedtothefollowingconditions:   1. Directsunlight.   2. Drafts of air.   3. Shock and vibration.   4. Temperatures other than that one would expect to see in   airconditioned, temperature controlled office of lab enviroment.   The 212Rshouldnotbemountedoutdoorsorsubjecttemperature fluc-   tuationsbeyond2or3degrees.   Theanalyzershouldbeplacedascloseaspossible,subjecttotheafore-   mentionedconditions,tothesamplepoint.   Outlinediagramsofbothunitswillbefoundinthedrawingsection.After   thecutouthasbeenmadeintheequipmentpanel TAIrecommendsthatthe   analyzerbeusedasatemplatetolayoutthefourmountingholes.Suchaproce-   durewillcompensateforsheetmetaltoleranceerrors.   2.2   Electrical Requirements and Connections.   Provisionshavebeenmadetoaccommodatethethreeexternalcircuit   connectionsrequiredbytheanalyzer. Accessholesononesideoftheanalyzer   case(seeOutlineDiagram)areprovidedfortheinstallationoftheconduitand   electricalwiring. Allthreecustomerconnectedcircuitsaretobeterminatedon   thebarrierterminalstripidentified“TS1”.   Toinstalltheconduitandwiring,theinnerhorizontallyhingedpanelmustbe   opened (to open the panel, turn the fastener screw at the top of the Panel a 1/4   turnccw).Whileinstallingtheconduitandwiringbecarefulnottodisturbthe   foaminsulationliningtheinteriorofthecaseanymorethanisabsolutelyneces-   sary.   7 Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   IMPORTANT: The foam lining the interior of the analyzer is an   integral part of the environmental temperature   control system. Removal or destruction of this   lining will result in erratic instrument perfor-   mance.   2.2.1 Primary Power.   Asourceofsinglephase,105to125volt,60cyclepowerwillberequired   tooperatetheanalyzer. Themaximumpowerconsumptionoftheanalyzeris   500watts.   NOTE: The analyzer is also available for 50 cycle operation   with special modifications and accessories.   RefertotheInterconnectionDiagraminthedrawingsectionofthemanual   andconnectthepowerandgroundwiringasshown. Besuretopolarizethe   powerserviceconnectionsasindicated. Whenconnectingthewires,donot   leaveanexcessiveamountofslackwithintheanalyzer. Twovacuumtubesare   locatedjustbelowthewiringareaandthewiringshouldbeinstalledtobewell   clearofthem.   2.2.2 Signal Output.   Connectatwoconductorshieldedcablebetweentheanalyzerandrecord-   ingequipment. Besuretoobservetheproperpolarityatbothinstruments.   Connecttheshieldofthecableontheindicatedterminalattheanalyzeronly,and   cutbackandinsulatetheshieldattherecorder.   NOTE: Connecting the shield at both ends of a cable when   dealing with low level circuits can create a ground   loop between two instruments. Improperly installed   shielding can produce more noise in a low level circuit   than no shielding at all.   2.2.3 Regulating Transformer.   Teledyne Analytical Instruments   8 Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Runthecableattachedtothetransformerunitinthroughtheaccesshole   that is equipped with the cable clamp and terminate it as specified on the   interconnectiondiagram.   2.2.4. Completion and Inspection.   After the electrical connections have been completed, slide any excess   slack back into the conduits so that the installed wiring is not in contact with   the components mounted on the analyzer chassis.   Removethethreefoamstripsthataretapedtotheinsideoftheanalyzer   doorandusethemtostufftheconduitopenings. Itisimportantthatthese   openings be as well sealed as possible.   Checktoseethatthetemperaturecontrolprintedcircuitboard,andall   vacuumtubesarefirmlyseatedintheirrespectivesockets.   Closeandlatchthecontrolpanel. Thereshouldbenofurtherneedtohave   accesstotheinterioroftheanalyzer. Allcontrolsandadjustmentsarearranged   sothattheycanbemanipulatedwithoutdisturbingthedelicatetemperature   equilibriumoftheinstrumentinterior.   2.3   Gas Requirements and Connections.   Beforeattemptinginstallationofthesampleandsupportinggaslinesand   accessoriesgivecarefulconsiderationtothefollowingimportantinstallation   notes.   Note #1: Itisabsolutelynecessarythatallconnectionsandcomponents   inthegascontrolsystemaheadofthemeasuringcellbeleak   free. TowardthatendTAIhastestedtheintegralsampling   systemunderpressurewithasensitiveleakdetectorand   certifiesthattheanalyzerisleakfree.   Note #2: Usenosolderconnectionsinthesystem. Solderingfluxes   outgas into the sample lines and produce erratic output   readings. Acid type soldering fluxes actually attack and   permanently change the characteristics of the detector cell   measuringelements.   Note #3: All sample system tubing should be new and clean. Many   gases and vapors are absorbed by dirt or oxide coatings on   tubing walls. These gases and vapors are released as the   ambient temperature rises. Because of the high sensitivity of   the analyzer, this absorption-desorption phenomenon can   9 Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   causealargefluctuationinoutputsignal.Forbestresults,use   electropolished SS tubing   Note #4: Becauseofthediffusionofairthroughcompositionmaterials,   only dual stage SS regulators with metallic diaphragms should   be used in conjunction with the sample and supporting gas   supplies.   Note #5: Regulatorsmustbethoroughlypurged(burped)priortouse.   2.3.1 Reference Gas.   Acylinderofgasoffixedcompositionisrequiredasthereferencefor   comparisonwiththesamplegas. Atotallackofimpuritiesinthereference   would be ideal but is not necessary. An impurity concentration of up to 50%   ofthenarrowestrangeofinterestistolerable. Theexactcompositionofthe   referencegasisacademicaslongastheuseriscertainthatitfallswithinthelimits   specifiedinSection7ofthemanual. Theimportantconsiderationisthatthe   compositionofthereferencegasremainunchangedwheninuse.   NOTE: When it becomes necessary to replace the reference   gas, the analyzer will have to be recalibrated.   2.3.2 Zero Gas.   A supply of gas, composed of the background gas of the analysis and   the lowest attainable concentration of impurity, will be required to standard-   ize one end of the ranges of interest. The composition of the zero gas must   be known to the same exactitude expected of the analysis. TAI suggests that   the gas be obtained from a supplier that will certify its composition. Recom-   mendations as to the composition of the gas will be found in Section 7 of the   manual.   2.3.3 Span Gas.   A supply of gas, composed of the background of the analysis plus a   known concentration of the component of interest, will be required to stan-   dardize the sensitivity of the analyzer. Again, the composition of the gas   must be known to the same order of accuracy expected of the analyzer.   Analysis and certification of the composition is desirable. Specific recom-   mendations governing the composition of the span gas will be found in   Section 7 of the manual.   Teledyne Analytical Instruments   10   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   2.3.4 Installation of Cylinder Supplies.   Thereferenceandstandardgascylindersshouldbeinstalledasclosetothe   analyzeraspossible. Eachcylindershouldbeprovidedwithadualstage,   metallicdiaphragm,pressurereducingregulator. Wheninstallingtheregulators,   crackthecylindervalveopensothatgasisflowing. Thisprocedurewillprevent   airfrombeingentrappedintheregulator,andeliminateacommonsourceof   supportinggascontamination. Improperinstallationofacylinderregulatorcan   appreciablychangethecompositionofastandardgas.Airtrappedintheregula-   torwilldiffusebackintothecylinder,andtheimpurityconcentrationofthe   compositionwillbealtered. Whendealingwiththezerogas,thealterationofthe   compositioncanbesignificant.   Onceinstalled,priortouse,theregulatorsforreference,spanandzero   gasesmustbepurgedbyalterativelypressuringthefirststage,withthesecond   stageoff,thenbleedingoutthefirststagebyopeningthesecondstage,backand   forth severaltimes.   Note: Make sure to close the cylinder prior to bleeding the first stage   through the second.   2.3.5 Sample Pressure.   Thesamplepointshouldbeequippedwithametallicdiaphragmpressure   regulatorandthepressurereducedtobetween10and50psig. Theregulator   shouldbeinstalledasclosetothesamplepointaspossibletominimizesample   linelagtime.   2.3.6 Interconnecting Lines.   Theinletandventconnectionsareidentifiedontheanalyzeroutlinedia-   gram. Theconnectionsarestainlesssteel1/8"femalepipecouplingsthatare   braisedintoagasconnectorbarmountedwithintheanalyzer. Thebraisedpipe   couplingwasselectedasatransitionsothatthetorquegeneratedduringinstalla-   tionoftheexternalsystemwouldbeisolatedfromtheinternalsystem. Bulkhead   typefittingscanbeaccidentlytwisted,andleakspromotedinhardtoreachareas   withintheanalyzer. TAIsuggeststhat1/4"tubingandadapterfittingsbeused   throughouttheexternalsamplingandsupportinggassystem.   11   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   2.3.7 Vent Lines.   Thesamplepath,referencepath,andanintegralbypasssystemventfrom   separateportsattherearoftheanalyzer. Whereverpossible,TAIrecommends   thatthegasesbepermittedtoventdirectlytotheatmosphere.Ifitisnecessary   tocarrythesegasestoaremoteareathefollowingprecautionsmustbeob-   servedwheninstallingtheventlines:   1) Theventlinesmustbe1/4"tubingorlargersothatnobackpressure   resultingfromrestrictedflowisexperiencedbythemeasuringcell.   2) Thesampleandreferencepathsmustventintoanareathatexperi-   ences the same ambient pressure conditions.   3) Theambientpressureattheventlocationshouldundergonomore   thannormalbarometricpressurefluctuations.   4) Theventlinesmustbeinstalledsothatwateranddustcannotaccu-   mulate in them.   Apressuredifferentialexistingatthecellbetweenthereferenceandsample   willresultinacorrespondingchangeinoutputsignal. Thereferenceandsample   pathflowmetersarelocatedupstreamfrom thecellsothatbothcellpathscan   ventdirectlytoatmosphere. Therandombounceofthefloatsintheflowmeters,   whenlocateddownstreamfromthemeasuringcell,canproduceupto5%noise   ontheoutputsignal.   3.0   3.1   STARTUP   Preliminary.   The following preliminary steps should be accomplished before apply-   ing power or starting gas flow.   1. Check the integral gas control panel and be sure that all valves are   closed (fully cw). Do not jam the sample, reference, and bypass   meteringvalves.   2. Place the “RANGE” switch on the #2 position.   3. Set the “SPAN” control to the reading recorded in Section 7 of the   manual.   Teledyne Analytical Instruments   12   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   4.Determinewhatthesamplepressureisandsettheoutputpressureof   the cylinder gas supplies to agree with the sample pressure. This   procedure, although not absolutely necessary, will minimize flow   adjustments when the sample path is switched to the various inputs.   5. Turn the analyzer and recorder power switches to “ON”.   3.2 Reference Gas Flow.   Openthereferencegasthrottlevalveadjacenttothe “Ref.Flow”meter,   andsetthevalveforanindicatedflowrateofbetween0.1and0.3scfh. Since   theactualsettingisnotcritical,TAIsuggeststhattheflowbesetat0.1scfhto   conservethereferencegassupply.   3.3   Zero Gas Flow.   Openthe“ZERO”valve(ccwuntilareleaseintensionisfelt),andadjust   thethrottlevalvetotherightofthe“SAMPLEFLOW’flowmeterforanindi-   catedflowrateof0.3scfh.   3.4   Warmup.   Awarm-upperiodoftwenty-four(24)hoursisrecommendedtostabilize   theinterioroftheanalyzeratthecontroltemperaturelevel. Untilthemeasuring   celltemperaturehasequilibratedwiththecontroltemperature,theoutputsignal   willdrift.   Duringthewarmupperiod,checktherecorderperiodically. Asthedrift   rate decreases, increase the sensitivity by moving the range switch towards   Range #1.   IMPORTANT:   ThezerocontrolwillhavelittleornoeffectuntilRange#1isreached.Ifthe   recorderisoffscaletowardselectricalzero(negativesignal)after24hoursof   runningonzerogas,usetheprocedureoutlinedinSection6.2. Thedriftratewill   decreaseexponentially. Whentherateofchangeslowstoapproximately1%   perhour,placethe“RANGE”switchonposition#1andreadjustthe“ZERO”   controlsothattherecorderisreadingmidscale. Donotattempttocalibratethe   analyzeruntiltherecorderindicationstabilizes.   3.5   Zero Standardization.   13   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Aftertheanalyzerstabilizesadjustthe"ZERO"controluntiltherecorder   indicatestheimpurityconcentrationofthezerogascomposition. Besurethat   therangeswitchisonposition#1andthatthesamplepathflowrateis0.3scfh.   NOTE if the"ZERO" or "FINE ZERO" pot runs out of adjustment and   the analyzer can not be calibrated, set this pot back to mid-scale. Then use   the "COURSE ZERO" pot to readjust the zero calibration as close as pos-   sible to zero calibration. And last use the "ZERO" ot "FINE ZERO" pot to   fine adjust the zero calibration.   3.6   Span Standardization.   Afterthezerohasbeenstandardized,switchtheselectorvalveto"Span",   andresetthesamplepathflowrateto0.3scfh. Therecordershouldcometo   balanceon,orcloseto, thecompositionofthespangas.   Theanalyzerisfactorycalibratedtomakesurethattheoutputislinearover   allthreerangesofinterest. Calibrationisachievedwithmixingblocktechnique   andthespansettingrecordedinSection7isderivedatthattime.   Iftheanalyzerperformedasdescribedinthewarm-upprocedure(Section   3.4),andfailstocloselyapproximatethecompositionofthespangas,thereis   groundsfordoubtingthespangasmixture. Becauseofthedifficultyinvolvedin   obtainingpreciseanalysisofsmallamountsofimpuritiesincylindergas,and   becauseoftheeasewithwhichthegascanbecontaminatedsubsequentto   analysis,anylargeerrorinresponsetothespangasshouldbesuspect. Insucha   caseTAIrecommendsthattheanalyzerbeoperatedattherecommendedspan   settinguntilthespangasisreanalyzed.   Ifadjustmentofthespancontrolisnecessarytocompensateforminor   discrepanciesbetweentherecordedsettingandthespangasreading,TAI   suggeststhatthezeroandspan proceduresberepeateduntilnofurtheradjust-   mentisrequired.   3.7   Bypass.   Theintegralgascontrolpanelfeaturesabypassflowmeterandthrottle   valve that is located downstream from the input manifold (see the analyzer   piping schematic in the drawing section). The bypass system can be used to   speed the response of the analyzer to changes in the process. The bypass   flowmeter will indicate flowrates that are a factor of ten greater than the   sample path flowmeter. TAI recommends that the bypass system be used   whenever the sample path is switched and particularly after using the span   gas. The time required for the analyzer to stabilize on an impurity concentra-   tion within the limits of Range #1, after having been exposed to a concentra-   tion within the limits of Range #3, is not a function of cell response which is   Teledyne Analytical Instruments   14   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   virtuallyinstantaneous. Recoveryisafunctionofhowlongittakestopurgethe   samplingsystemupstreamfromthecell.Acceleratingthesamplepathflowrate   willspeedtheresponse.   NOTE:   TAI recommends that some bypass flow be permitted   at all times. The bypass throttle meter is a part of the   input manifold. Being so located means that the   throttle valve lies between the sample side of the cell   and atmosphere.   Without some bypass flow it is possible that atmo-   sphere could diffuse into the sample line. Do not   attempt to use the throttle valve as a shut off valve.   Jamming the valve will damage the metering needle.   3.8Sample Mode.   Aftertheanalyzerhasbeenstandardized,switchtheselectorvalveto   "Sample". Set the bypass flow so that the float is visibly elevated from the   bottom of the tube and then readjust the sample flowrate to 0.3 scfh. No   furtheradjustmentsarerequired.   NOTE: Although the bellows type valves used in the input   manifold are among the best available, TAI suggests   that the span gas be turned OFF at the cylinder when   not in use, and the span line relieved of pressure   before closing the analyzer span valve. Such a proce-   dure will eliminate any possibility of the span gas   diffusing into the sample path of the analyzer. Nor-   mally, the recommended span gas composition con-   tains a component of interest content of 3/4 the coars-   est range of analysis. The component of interest   content of the sample gas, on the other hand, is usu-   ally well within the finest range of analysis. Instrument   sensitivity in Range #1 is 100 times greater than in   Range #3. The user can readily see what even a   minute rate of span gas diffusion would do to the   validity of a Range #1 analysis.   4.   ROUTINE OPERATION   Flowrates.   4.1   15   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Thereferenceandsampleflowratesshouldbecheckeddaily. Theinstru-   mentissomewhatflowsensitive. OperatingonRange#1achangeinflowrateof   0.1scfhwillcauseacorrespondingchangeinsensitivityoffrom1to20%of   scale. Onthecoarserranges,thechangeinsensitivitywouldbeundetectable.   Thesamplepathflowrateshouldbemaintainedat0.3scfh,thereference   pathflowrateat0.1scfh,andthebypassflowmeterfloatshouldbeslightlyabove   thebottomoftheindicatorglass.   4.2 Supporting Gas Supplies.   Supportingcylindergassuppliesshouldbecheckedfrequentlyonaroutine   basiswithparticularattentionfocusedonthereferencegas. Asparecylinderof   referencegasshouldbeavailableatalltimes. Whenthecylinderpressuredrops   below100psigthereferencesupplyshouldbereplacedastheoperationofthe   regulatorisquestionableatpressuresbelowthispoint.   Whenreplacingsupplycylinders,besuretobleedthegasthroughthe   cylinder valve while installing the pressure regulator (see Section 2.3.4). It is   also advisable to check the connections with soap water whenever a supply   cylinder is changed.   4.3   Standardization.   Theanalyzermustbecompletelyrestandardizedwheneverthereference   gassupplyisreplaced. Barringunforeseendifficultieswiththeanalyzer,re-   standardizationshouldnotbenecessarybetweenreferencecylinderreplacement   periodsiftheanalyzerisruncontinuously. Iftheanalyzerisshutdownforlong   periodoftimeusethestartupprocedure(initsentirety)whenoperationistobe   resumed.   NOTE: TAI strongly recommends that the analyzer run con-   tinuously with gas flowing in both the sample and   reference paths. During inactive periods arrange the   input manifold so that zero gas is flowing. Gas sup-   plies can be conserved by reducing the sample and   reference path flowrates to less than 0.1 scfh and   closing off the bypass flow completely.   Teledyne Analytical Instruments   16   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Duetothedelicatebalanceofthethermalconductivitycellemployedinthe   analyzer,exposuretotheatmosphereinanuncontrolledtemperatureenviron-   mentisundesirable. Afterextendedshutdownperiodsmanydaysofoperation   mayberequiredfortheanalyzertore-stabilize.   5.   PERIODIC MAINTENANCE   Heater Fan.   5.1   Theheaterfanistheonlycomponentwithintheanalyzerthatrequires   periodic attention. Whenever the reference gas is replaced the motor bear-   ings will require a few drops of light machine oil. Since oiling the fan motor   necessitatesopeningtheinnerdooroftheanalyzer,theinstrumentwillrequirethe   necessarytemperatureequilibratingtimeafterthedoorisclosed. Withzerogas   flowingbesurethattheanalyzerhasstabilizedontheRange#1positionbefore   re-standardizationisattempted.   6.   TROUBLE SHOOTING   6.1   Preliminary.   Iftheanalyzerissuspectedofincorrectoperation,asapreliminaryto   evaluation, always arrange the input manifold so that zero gas is flowing   through the sample path of the analyzer. Never attempt to evaluate the   performance of the instrument with sample gas flowing.   Analysisbythermalconductivityisnonspecificinnature. Athermalcon-   ductivityanalyzerwiththefinerangesensitivityoftheModel212Rwill respond   tomanyinfluencingfactorsotherthanthecomponentofinterest;particularly   whenoperatedatRange#1sensitivity.   It is necessary, therefore, to eliminate as many external variables as   possible if the performance of the analyzer is to be assessed. Programming   the analyzer for zero gas places both paths of the measuring cell on relatively   reliablecylindergassources.   6.1.1   Electrical Checks.   17   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Checktoseethatboththeanalyzerandrecorderarebeingfurnished   electrical power. Check to see that the analyzer and recorder fuses are intact.   Check all electrical connections outside of the analyzer.   6.1.2 Sampling System Checks.   Besurethatthereisanadequatesupplyofreferencegasandthatthe   sample and reference path flowrates are correct. Check to see that there are   no obstructions in the vent paths from the analyzer. Check all external   plumbing connections for leaks with soap water.   6.1.3 Summary of Preliminary Checks.   Theanalyzerinnerdoorshouldnotbeopenedandnoadjustments,beyond   manipulationofthenormallyusedcontrols,shouldbemadeuntilalltheafore-   mentionedpreliminarieshavebeencompletedandanynecessaryremedial   repairseffected.   6.2   Loss of Zero Control.   IflossofzerocontrolonRange#1isexperiencedduringinitialstartupor   afterthereferencegassupplyhasbeenchanged,thedynamicbalanceprocedure   mustberepeated.   Lossofzerocontrol,undersuchcircumstances,indicatesthattheimpurity   concentrationwithinthereferencegasisdifferentthanthatofthegasusedduring   factorycheckout(orthepreviouscylinder)anddoesnotinitselfindicatea   defectiveanalyzer.   Theanalyzerwasadjustedatthefactorytohaveazerobalanceatcloseto   themidpointofthezerocontrolpotentiometer(dualdialreadingof500)witha   commonsourceofhighpuritycylindergassupplyingboththesampleand   referencepathsofthecell. Iftherecorderisoffscaleandcannotbereturned   withthezerocontrol,thedynamicbalanceproceduremustbeemployedto   restorecontrolbeforefurtherconclusionsastoinstrumentperformancecanbe   made.   Teledyne Analytical Instruments   18   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   6.2.1   Dynamic Balance Procedure.   Themeasuringbridgecircuit(seeschematic)incorporatestwo 10 turn   potentiometersthatareutilizedtocompensateforminutedifferencesinthe   resistive and reactive properties of the hot wire elements and the center   tapped secondary of the bridge transformer in the reference and zero gas   relatedtotheapplication.   Thepotentiometersareofthescrewdriveradjustmenttypeandare   equippedwithshaftlockingassembliesthathavebeenfactoryadjustedtocreate   enoughfrictionsothattheshaftwillnotchangepositioninshipment.Thepotenti-   ometersarelocatedontheamplifierchassis,andarecontrolledthroughtwo   holesidentified“RESISTIVE”and“REACTIVE”ontheinnercontrolpanel.   DO NOT OPEN THE INNER PANEL OF THE ANALYZER TO   GAIN ACCESS TO THESE CONTROLS. The balance adjustments   should not be disturbed if the analyzer is not up to operating temperature.   Correct bridge balance can only be achieved at operating temperature be-   cause of the inherent thermal junctions within the potentiometers themselves.   The minute voltages produced by contact between the dissimilar materials   within the potentiometers are effectively balanced out in the procedure when   it is accomplished at operating temperature. Because the voltage generated   by these junctions varies with temperature, and because the analyzer will   resolve a 0.25 microvolt change in bridge voltage, temperature stability is a   mandatory prerequisite to the dynamic balance procedure.   To restore zero control within the limits of Range # 1, employ the   followingProcedure:   1) Withzerogasflowing,adjustthesampleandreferencegasflowme-   terstothesamevaluebetween0.1and0.3scfh.   2) Set the ZERO potentiometer to 500.   3) Connectavoltmetertotheinstrumentsignaloutput. Connectan   oscilloscope to the probe point an d set the sweep rate to 5   milliseconds. Set the voltage level so ± 15 volts can be read.   4) If a reading of +15 or –15 volts is indicated on the oscilloscope   with the RANGE switch set to Range # 1 the instrument is   significantly unbalanced. Set to Range #2 and proceed to step 4).   If the voltage is lower in magnitude than + or – 15 skip down to   step 6).   19   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   5) Usingascrewdriverthathasashaftdiameterof1/8”andashaft   length of at least 6”, turn the “Reactive” balance adjustment all   the way counterclockwise (CCW) until the potentiometer stop   pointisreached. Turntheadjustmentbackclockwise(CW)5full   turnstocentertheadjustment.   6) Turnthe“Resistive”balanceadjustmentallthewayclockwise(CW)   untilthepotentiometerstoppointisreached.   7) Slowly turn the “Resistive” adjustment until the output reaches its   lowest value. If the reading is negative adjust to the most nega-   tivevaluepossible.   8) Turnthe“Reactive”balanceadjustmentslowlyCWandCCWto   obtain the lowest possible output value. Again, if the reading is   negative adjust to the most negative value possible.   9) Repeat steps 6) and 7) until the output value is as low as these   adjustments can make it.   10) Detunetheinstrumentbyturningthe“Resistive”balanceadjust   ment ½ turn CW. This will bring the signal out of the baseline   noise, which could otherwise cause non-linear response at the   low concentration end.   NOTE: Due to minute differences between measuring cells   and the electronics in the bridge circuit found in vari-   ous instruments, some analyzers will have to be bal-   anced at a point higher than 10% of scale while others   can be balanced even closer to electrical zero. The   10% of scale recommended in step 6 of the procedure   is an arbitrary starting point. If the bridge circuit can   be easily balanced at this signal level, the procedure   should be repeated at a point even closer to electrical   zero. Conversely, if balance cannot be accomplished   at 10% of scale, the operator will have to select a   higher recorder set point for the procedure. In any   case, the object is to accomplish dynamic balance   with the lowest magnitude of analyzer cutout signal   attainable with the analyzer at Range #1 sensitivity.   6.3   Correct Operation.   If a stable recording is achieved when the analyzer is operated on zero   gas, erratic performance must be attributed to the customers sampling system   Teledyne Analytical Instruments   20   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   orincorrectinstrumentinstallation. Thefollowingconditionscanproducean   erratic output from the analyzer:   1) Installationinanareathatissubjecttolargechangesinambient   temperature, direct sunlight, drafts from wind, blowers, or air condi-   tioners. Any of the foregoing conditions will produce a bidirectional   (diurnal)variationinoutputsignal.   2) Sample gas with a moisture content of 10 ppm or greater. Again, the   outputwillvarybi-directionalitywithambienttemperaturechanges.   Ifthesamplegasisknowntohaveafairlyhighmoisturecontenta   dryershouldbeinstalledinthesampleline.   3) Eventhesmallestleakanywhereinthesamplingsystemupstream   from the analyzer, a small leak at a fitting, valve, or regulator, will   cause almost constant variation in analyzer output. The resulting   signaldriftwouldbeerraticandcouldbeeitherbi-directionalor   unidirectional.   4) Acidinthesamplingsystem(solderingfluxes)willproducecontinu-   ous, uncontrollable, unidirectional output signal drift. Sample tubing   that has been treated with descaling chemicals, if subject to above   ambient temperatures, will outgas continuously. Any form of acid in   the sampling system will outgas at high temperatures and attack the   hot wire elements of the cell.   5) Composition materials (diaphragms, tubing, etc.) in the sampling   system will produce erratic, unpredictable changes in output signal   level because of their permeability to gases other than the sample.   The nylon tubing normally used in quick disconnect type sampling   systems will produce diurnal output signal excursions in the order of   10 to 20% of the fine range of analysis.   6.4   IncorrectOperation.   Iftheanalyzerisnotstableonzerogas,andthesupportinggasinstallation   hasbeenmadeandcheckedinaccordancewithsections2and3,thefollowing   proceduresshouldbeemployed.   6.4.1 Analyzer Leak   21   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Checkaspreviouslystated,theanalyzersentiresamplingsystemhasbeen   checkedunderpressurewithaleakdetector. However,iftheexternalsystem   checksoutandtheinstrumentisdisplayingalargediurnaloruni-directionaldrift,   theinteriorsamplingsystemshouldbecheckedforleaks.Donotopenthe   analyzerinnerdoor. Usethefollowingprocedure:   1)   2)   Acquireasourceofcylinderheliumandthroughadualstage   regulator connect it to a length of 1/8" tubing.   Open the outer door of the analyzer and insert the tubing be-   tween the edge of the sample control panel and the outer case.   CAUTION: Do not insert the tubing along the edges of the   larger swing-down control panel as there is danger of   obstructing the fan or short circuiting the electronics.   The end of the tubing should be shaped into a curve   so that the foam lining is not overly disturbed and the   tubing should only be inserted a couple of inches into   theinterior.   3)   Start the helium flowing. Keep the flowrate low so as not to   disturbthetemperatureequilibriumoftheanalyzerinterior.   4)   Observe the recorder. If any leaks are present in the interior   sampling system, the recorder will be driven off scale in one   direction or the other. NOTE: The analyzer should be on Range   #1 and the zero control adjusted so that the recorder is reading   mid-scale. The direction the recorder indication moves is a   function of the type of zero and reference gas being used (which   varies per application) and which path of the sampling system   (sample or reference) the leak is located in. If the recorder   responds to an atmosphere of helium within the analyzer, all   connections inside the analyzer must be checked.   6.4.2 Temperature Control Check.   Theinteriortemperatureoftheanalyzerisregulatedtoaconstantvalueto   within.002degreescentigrade. Theactualtemperaturecontrolpointvaries   slightlyperapplicationandisafunctionofthecomponentsusedinthepropor-   tionaltemperaturecontrolcircuit.   Iftherecorderisdisplayingalargediurnalrecordingthatisrelatedtothe   timeofdayintermsofthedifferentialbetweendayandnightambienttempera-   ture,troubleinthetemperaturecontrolcircuitryisindicated. Again,thissymp-   Teledyne Analytical Instruments   22   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   tommustbeevidentwithzerogasflowing. Ifthissymptomisdisplayedonly   whenthesampleisflowing,theproblemisrelatedtoone(ormore)ofthefive   (5)pointsoutlinedinsection7.3.   Tocheckthetemperaturecontrolcircuit,usethefollowingprocedures:   1)   WiththepowerswitchON,openthecontrolpaneldoorand   check to see if the fan is running. If it is not, check the left hand   fuse. This fuse protects the entire temperature control circuit   (see schematic). If the fuse will not hold, a short circuit is   indicated. Disconnecttheprintedcircuitcardandreplacethe   fuse. If the fuse holds (as indicated by the fan running) the   printed circuit card will have to be replaced.   NOTE:Unlesscompetentelectronictechniciansareavailable,TAI   recommendsthatareplacementprintedcircuitcardbeordered   and the existing board returned to the factory for repair. Repair   charges (out of warrantee) will be based on time and material.   A schematic of the proportional temperature control card is   included among the drawings at the rear of the manual.   2)   Ifthefusewillnotholdwiththeboardremoved,a short isindi   cated in the fan, the heater resistors, or the interconnecting   wiring. Withtheboardremoved,checkwiringandcomponents   for short circuits.   3)   Ifthefanisrunning,connectanACvoltmeteracrossanyoneof   the heater resistors on the fan assembly (the meter should be set   to read 50 VAC). In a properly operating circuit, the meter   should read approximately 47 volts (heater control full on) as the   control panel is open and the thermistor should be demanding   maximum heater voltage. If no voltage is present, either the   circuit card or the thermistor probe is faulty. Isolate the problem   byfirstdisconnectingthecircuitboardfromitssocketandwith   anohmmetercheckbetweenterminals5and6ofterminalstrip   TS2. Ifareadingisobtained(disregardactualresistance)the   thermistorisintactandanewcircuitcardwillbenecessary. If   no reading is obtained, a new probe assembly will be required.   4)   Ifapproximately47VACismeasuredacrosstheheaterresis-   tors, a runaway, or correct, circuit is indicated. To determine   which is the case, hold a soldering iron in close proximity (1/2")   totheendofthethermistorprobeandobservethemeter.   23   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   CAUTION: Do not touch the thermistor probe with the iron. The   end of the probe is tied and taped to the sample tubing   below the cell cover. If the meter reading suddenly   drops to zero, remove the soldering iron immediately.   After the area around the thermistor cools, the voltage   across the heater resistors should again climb to 47   volts. Such action indicates a properly functioning   heater circuit. If the meter stays in the 47 volt region,   the circuit card will have to be replaced.   7.0 CALIBRATION DATA FOR MODEL 212R   SERIAL NUMBER: 195929   7.1   Ranges.   Therangesoftheanalyzerare:   RangeSwitchPosition#1:   Range Switch Position #2:   0-50,000 ppm N2 in H2   0-5000 ppm N2 in H2   Range Switch Position #3:   0-500 ppm N2 in H2   Teledyne Analytical Instruments   24   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   7.2   Output Signal:   Outputcoincideswith100%analysis.   4mA   = = = = 0 ppm N2 in H2   20mA   50,000 ppm N2 in H2 - Range #1   5000 ppm   500 ppm   N2 in H2 - Range #2   N2 in H2 - Range #3   7.3   7.4   Span Setting: 555   Recommended Accessory Gases.   H2 Purity   H2 Purity   ZeroGas   99.999   Reference Gas   SpanGas   99.999   350-450 ppm N2 in H2   7.4.1 Reference Gas.   Withaequivalentimpurityoflessthan3%ofrange1. Theimpurityratio   mustremainconstant.Whenthereferencegassupplyisreplaced,theanalyzer   mustbere-standardized.   7.4.2 Zero Gas.   Becauseofthedifficultiesinvolvedinobtainingagasinpureform,thezero   gasmustbepurchasedfromasupplierwhowillcertifythecylinderastoits   content. Theequivalentimpurityconcentrationofthegasmustbelessthan3%   ofrange1. Whenthezerogasisintroducedforstandardization,therange   switchmustbeinRange1,andthezerocontrolsetsuchthattherecorder   indicatesthezerogasimpurity.   IMPORTANT:Atstartup, besurethatthespancontroldialissettothe   numberrecordedinSection7.3.   25   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   7.4.3 Span Gas.   Containinganequivalentimpurity. Again,thespangasmustbeprocured   fromasupplierwhowillcertifyitscomposition. Therangeswitchmustbein   range3whenthespangasisintroduced.   IMPORTANT: Theaccuracyoftheanalysisisdirectlyrelatedtothe   usersknowledgeofthespangascomposition.   ReferenceandSampleFlowrates 0.3SCFH   Bypass Flowrate 3.0 SCFH   NOTE:   Cylinder and sample stream pressure settings must   be adjusted in order to maintain the above flowrates   when switching from one sample to another.   RECOMMENDED SPARE PARTS LIST   QUANTITY PART NO.   DESCRIPTION   1 1 1 1 A-6251   A-6981   F-45   MeasuringCellBlockAssembly   ProportionalTemperatureControl   SampleFlowmeterTube   F-22   SampleFlowmeterBypassTube   Aminimumchargeappliestoallsparepartorders.   IMPORTANT: Ordersforreplacementpartsshouldincludethepart   number,themodelandserialnumberoftheanalyzerinwhichtheyaretobe   used.   Teledyne Analytical Instruments   26   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   Ordersshouldbesentto:   TELEDYNE Analytical Instruments   16830ChestnutStreet   City of Industry, CA 91749-1580   Phone (626) 934-1500, Fax (626) 961-2538   TWX (910) 584-1887 TDYANYL COID   Web:   Drawing List   B-73356   D-73212   A-72035   C-73085   C-73088   Outline Diagram   Final Assembly   Piping Diagram   Main PCB Schematic   Power Supply PCB Schematic   Replacement Parts Order Information   A minimum charge of $150.00 is application to all spare parts   orders.   Important:Ordersforreplacementpartsshouldincludetheart   number(ifavailable)andthemodelandserialnumberoftheinstrument   forwhichthepartisintended.   TeledyneAnalyticalInstruments   16830Chestnutstreet   CityofIndustry,Ca91748-1580   Telephone:(626)961-2538   (626) 934-1500   Fax:   (626) 961-2538   (626) 934-1651   27   Teledyne Analytical Instruments   Download from Www.Somanuals.com. All Manuals Search And Download.   Thermal Conductivity Analyzer   Model 212R   APPENDIX   Specifications:   Ranges: Three decade ranges: Minimum Full scale(ppm)   20ppmHydrogenbalanceArgon   25 ppm Hydrogen balance Nitrogen, Air   100ppm NitrogenbalanceArgon,Helium   100ppmHeliumbalanceAir   150ppmNitrogen,CarbonMonoxidebalanceHydrogen   200ppmOxygenbalanceArgon   Range ID:   3 range ID contacts   Local Range ID LED’S   Detector: Thermal conductivity sensor   Signal Output: 0-1VDC negative ground   4-20maDC Isolated ground   Accuracy:   ±2%fullscaleformost binarymixture   atconstanttemperature.±5%ofFull   Scaleoveroperatingtemperature(once   temperatureequilibriumhasbeen achieved)   Operating   Temperature:   20 – 30°C (68° F to 85° F)   Sample   Requirement:   Sample:0.3SCFH   Reference:0.1SCFH   Display:   Mounting:   DigitalLEDreadout(3½Digit)   Flushpanelmount.   Response Time:   90%offullscalelessthen60seconds.   Formostapplication.   Stability:   Lessthen2%fullscaledriftover24   hours   Power   requirement:   110VAC,50/60Hz (220VACoptional)   Teledyne Analytical Instruments   28   Download from Www.Somanuals.com. All Manuals Search And Download.   |