ResMed Oxygen Equipment VPAP ST User Guide

VPAPST  
NONINVASIVE VENTILATOR  
User Guide  
English  
A
Respiratory Care Solutions  
Making quality of care easy  
Download from Www.Somanuals.com. All Manuals Search And Download.  
English  
Indications for use  
The VPAP ST is indicated to provide noninvasive ventilation for patients weighing more than 13 kg or  
more than 30 kg in iVAPS mode with respiratory insufficiency or obstructive sleep apnoea (OSA). The  
VPAP ST is intended for home and hospital use.  
Contraindications  
Positive airway pressure therapy may be contraindicated in some patients with the following pre-  
existing conditions:  
severe bullous lung disease  
pneumothorax or pneumomediastinum  
pathologically low blood pressure, particularly if associated with intravascular volume depletion  
dehydration  
cerebrospinal fluid leak, recent cranial surgery, or trauma.  
Adverse effects  
Patients should report unusual chest pain, severe headache, or increased breathlessness to their  
prescribing physician. An acute upper respiratory tract infection may require temporary  
discontinuation of treatment.  
The following side effects may arise during the course of therapy with the device:  
drying of the nose, mouth, or throat  
nosebleed  
bloating  
ear or sinus discomfort  
eye irritation  
skin rashes.  
Masks and humidifiers  
under Service & Support. For more information on using your mask or humidifier, refer to the  
manual supplied with your mask or humidifier.  
English  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setup  
Refer to illustration A.  
1.  
2.  
3.  
4.  
5.  
Connect the DC plug of the power supply unit to the rear of the device.  
Connect the power cord to the power supply unit.  
Plug the other end of the power cord into the power outlet.  
Connect one end of the air tubing firmly onto the air outlet.  
Connect the assembled mask system to the free end of the air tubing.  
Control panel  
Refer to illustration B.  
The control panel of your device includes the following items:  
1.  
2.  
3.  
4.  
5.  
Start/Stop button: Starts or stops treatment  
LCD screen: Displays the menu, treatment and reminder screens  
Info menu button*: Allows you to view your sleep statistics  
Setup menu button*: Allows you to make changes to settings  
Push dial: Turning the dial allows you to scroll through the menu and change settings. Pushing  
the dial allows you to enter into a menu and confirm your choice.  
*The Info and Setup menus are disabled if S9 Essentials has been enabled by your clinician.  
Navigating the menus  
Refer to illustration C.  
In general, to navigate the menus:  
1.  
2.  
3.  
4.  
Turn  
until the parameter you require is displayed in blue.  
Press . The selection is highlighted in orange.  
Turn  
Press  
until you see the setting that you require.  
to confirm your choice. The screen returns to blue.  
Getting started  
1.  
2.  
3.  
4.  
5.  
6.  
Make sure the power is connected.  
Adjust the ramp time if required.  
Fit your mask as described in your mask user guide.  
To start therapy, simply breathe into the mask and/or press  
Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.  
To stop treatment at any time, remove your mask and/or press  
.
.
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notes:  
If your clinician has enabled SmartStart your device will start automatically when you breathe  
into the mask and stop automatically when you remove the mask.  
If power is interrupted during treatment, the device automatically restarts therapy when power  
is restored.  
Cleaning and maintenance  
You should regularly carry out cleaning and maintenance as described in this section. Refer to your  
mask and humidifier user guides for detailed instructions regarding the care of your device.  
Daily:  
Remove the air tubing by pulling on the finger grips on the cuff. Hang it in a clean, dry place until next  
use.  
Notes:  
Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack.  
Do not wash the air tubing in a washing machine or dishwasher.  
Weekly:  
1.  
2.  
3.  
4.  
Remove the air tubing from the device and the mask.  
Wash the air tubing in warm water using mild detergent.  
Rinse thoroughly, hang, and allow to dry.  
Before next use, reconnect the air tubing to the air outlet and mask.  
Monthly:  
1.  
2.  
Wipe the exterior of the device with a damp cloth and mild detergent.  
Check the air filter for holes and blockage by dirt or dust. Replace the air filter if necessary.  
Replacing the air filter:  
Replace the air filter every six months (or more often if necessary).  
1.  
2.  
3.  
4.  
Remove the air filter cover from the back of the device.  
Remove and discard the old air filter.  
Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.  
Replace the air filter cover.  
Notes:  
Ensure the air filter and air filter cover are fitted at all times.  
Do not wash the air filter. The air filter is not washable or reusable.  
English  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
SD card  
An SD card has been supplied to gather therapy data from your device and provide settings updates  
from your clinician. When instructed to do so, disconnect the device from the power outlet, remove  
your SD card, insert it in the protective folder and send it to your clinician. For more information on  
removing and inserting your card refer to the S9 SD Card Protective Folder provided with your device.  
Please retain the S9 SD Card Protective Folder for future use.  
Troubleshooting  
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your  
equipment supplier or ResMed. Do not attempt to open the device enclosure.  
Problem/Possible cause  
Solution  
No display  
Power is not connected.  
Ensure the power cord is connected and the power outlet (if  
available) is on.  
The DC plug is partially inserted Fully insert the DC plug.  
into the back of the device or  
inserted too slowly.  
Insufficient air delivered from the device  
Ramp time is in use.  
Air filter is dirty.  
Wait for air pressure to build up or change ramp time.  
Replace air filter.  
Check air tubing.  
Air tubing is not connected  
properly.  
Air tubing is blocked, pinched or Unblock or free the air tubing. Check the air tubing for punctures.  
punctured.  
Mask and headgear are not  
positioned correctly.  
Adjust position of mask and headgear.  
Incorrect air tubing selected.  
If you are using the SlimLine, Standard or 3 m air tubing ensure  
that you have the correct air tubing selected via the menu.  
Non-vented mask is used.  
Only use a vented mask.  
Mask vents might be blocked.  
Check if you have sufficient venting. Unblock mask vents if  
necessary.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem/Possible cause  
Solution  
EPAP may be set too low.  
Talk to your clinician about your settings.  
Device does not start when you breathe into the mask  
Breath is not deep enough to  
trigger SmartStart/Stop.  
Take a deep breath in and out through the mask.  
SmartStart/Stop is disabled  
because Leak Alert is enabled.  
Press Start/Stop to start therapy.  
SmartStart/Stop is disabled  
because Confirm Stop is  
enabled.  
A message appears on the screen. To stop therapy, select Yes and  
press the Push Dial.  
SmartStart/Stop is disabled.  
There is excessive leak.  
Talk to your clinician about enabling the SmartStart/Stop feature.  
Adjust position of mask and headgear.  
Connect the air tubing firmly at both ends.  
Device does not stop when you remove your mask  
SmartStart/Stop is disabled  
because Leak Alert is enabled.  
Press Start/Stop to stop therapy.  
SmartStart/Stop is disabled  
because Confirm Stop is  
enabled.  
A message appears on the screen. To stop therapy, select Yes and  
press the Push Dial.  
SmartStart/Stop is disabled.  
Talk to your clinician about enabling the SmartStart/Stop feature.  
SmartStart/Stop is enabled but the device does not stop automatically when you remove your  
mask  
Incompatible mask system  
being used.  
Only use equipment recommended by ResMed.  
Incorrect mask setting being  
used.  
Check the selected mask type in the Setup menu. Change it if  
necessary.  
The patient is using a nasal  
pillows mask with a set  
Disable SmartStart/Stop.  
pressure less than 6 cm H2O.  
English  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem/Possible cause  
Solution  
The patient is using a paediatric Disable SmartStart/Stop.  
mask with a set pressure less  
than 8 cm H2O.  
Pressure rises inappropriately  
Talking, coughing or breathing in Avoid talking with a nasal mask on, and breathe as normally as  
an unusual manner.  
possible.  
Mask cushion is buzzing against Adjust the headgear.  
the skin.  
Cushion seated incorrectly  
causing excessive leak.  
Adjust headgear or re-fit cushion.  
Displays message: High temperature fault, refer to user manual  
Device has been left in a hot  
environment.  
Allow to cool before re-use. Disconnect the power cord and then  
reconnect it to restart the device.  
Air filter is blocked.  
Replace your air filter. Disconnect the power cord and then  
reconnect it to restart the device.  
Air tubing is blocked.  
Check your air tubing and remove any blockages. Disconnect the  
power cord and then reconnect it to restart the device.  
Humidity level setting is too  
high, resulting in accumulation  
of water in the air tubing.  
Turn the humidity level setting down and empty the water from the  
air tubing.  
Displays message: Check ResMed 30/90W Power Supply Unit and fully insert the connector  
The DC plug is partially inserted Fully insert the DC plug.  
into the back of the device or  
inserted too slowly.  
A non-ResMed power supply  
Remove the power supply unit and replace with a ResMed power  
unit is connected to the device. supply unit.  
The power supply unit is being  
covered by bedding.  
Make sure that the power supply unit is free from bedding, clothes  
or other objects that could cover it.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem/Possible cause  
Solution  
Displays message: No tube, please check your tube is connected  
Flow is high because air tubing  
is not connected properly.  
Connect the air tubing firmly at both ends.  
Note: The tube disconnection  
check may not operate when an  
antibacterial filter is used.  
Displays message: Tube blocked, please check your tube  
Air tubing is blocked.  
Check your air tubing and remove any blockages. Disconnect the  
power cord and then reconnect it to restart the device.  
Displays message: High leak, please check system setup and all connections  
There is excessive leak.  
Note: If Leak Alert is enabled,  
an audible alert is activated and  
a high leak message is  
displayed.  
Adjust position of mask and headgear.  
Connect the air tubing firmly at both ends.  
The following message is displayed on the LCD after you try to update settings or copy data to  
the SD card: Card error, please remove SD card and contact service provider  
SD card is not inserted correctly. Ensure that the SD card is inserted correctly.  
You may have removed the  
SD card before settings were  
copied to the device.  
Reinsert the SD card and wait for the Home screen or the  
"Settings updated successfully, press any key" message to  
appear on the LCD.  
Note: This message only appears once. If you re-insert the SD card  
after you have updated your settings, the message will not be re-  
displayed.  
The following message is NOT displayed on the LCD after you try to update the settings using  
the SD card: Settings updated successfully, press any key  
The settings were not updated. Contact your clinician/service provider immediately.  
English  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
General technical specifications  
Power supply  
90W power supply unit  
Input range: 100–240V, 50–60Hz, 115V, 400Hz nominal for aircraft use  
Typical power consumption: 70W (80VA)  
Maximum power consumption: 110W (120VA)  
30W power supply unit  
Input range: 100–240V, 50–60Hz, 115V, 400Hz nominal for aircraft use  
Typical power consumption: 20W (40VA)  
Maximum power consumption: 36W (75VA)  
90W DC/DC converter  
Nominal inputs: 12V, 24V  
Typical power consumption: 70W  
Maximum power consumption: 110W  
Environmental  
conditions  
Operating temperature: +5°C to +35°C  
Note: The air flow for breathing produced by this therapy device can be higher  
than the temperature of the room. Under extreme ambient temperature  
conditions (40ºC) the device remains safe.  
Operating humidity: 10 to 95% non-condensing  
Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa  
Storage and transport temperature: -20°C to +60°C  
Storage and transport humidity: 10 to 95% non-condensing  
ResMed confirms that the device/s meets the Federal Aviation Administration  
(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air  
travel.  
Aircraft use  
Electromagnetic Product complies with all applicable electromagnetic compatibility requirements  
compatibility  
(EMC) according to IEC60601-1-2, for residential, commercial and light industry  
environments.  
It is recommended that mobile communication devices are kept at least 1 m  
away from the device.  
Information regarding the electromagnetic emissions and immunity of this  
ResMed device can be found on www.resmed.com, on the Products page  
under Service and Support. Click on the PDF file for your language.  
Class II (double insulation), Type BF, Ingress protection IP21  
IEC 60601-1  
classification  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
VPAP ST technical specifications  
Mode pressure  
ranges  
CPAP mode  
Set Pressure: 4–20 cm H2O  
S, ST, T and PAC modes  
IPAP: 4–25 cm H2O; EPAP: 2–25 cm H2O  
iVAPS mode  
PS: 0–23 cm H2O; EPAP: 2–25 cm H2O  
Maximum single  
fault pressure  
Maximum single fault steady state pressure: 30 cm H2O—if pressure exceeded  
for > 6 sec; 40 cm H2O—if pressure exceeded for >1 sec  
Nominal dimensions (L x W x H): 153 mm x 140 mm x 86 mm  
Weight: 835 g  
Physical  
Housing construction: Flame retardant engineering thermoplastic  
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)  
Hypoallergenic air filter: Acrylic and polypropylene fibers in a polypropylene  
carrier  
Air filter  
Standard air filter: Polyester non-woven fiber  
Sound  
Pressure level (CPAP mode)  
DECLARED  
With SlimLine air  
tubing:  
With Standard air  
tubing:  
26 dBA with uncertainty of 2 dBA as measured according  
to EN ISO 17510-1:2009  
27 dBA with uncertainty of 2 dBA as measured according  
to EN ISO 17510-1:2009  
DUAL-NUMBER  
NOISE EMISSION  
VALUES in  
accordance with  
ISO 4871:1996  
With either SlimLine 28 dBA with uncertainty of 2 dBA as measured according  
or Standard air  
tubing and H5i:  
to EN ISO 17510-1:2009  
Power level (CPAP mode)  
With SlimLine air  
tubing:  
34 dBA with uncertainty of 2 dBA as measured according  
to EN ISO 17510-1:2009  
With Standard air  
tubing:  
35 dBA with uncertainty of 2 dBA as measured according  
to EN ISO 17510-1:2009  
With either SlimLine 36 dBA with uncertainty of 2 dBA as measured according  
or Standard air  
tubing and H5i:  
to EN ISO 17510-1:2009  
Supplemental  
oxygen  
Recommended maximum supplemental oxygen flow: 15 L/min (CPAP, S, ST, T,  
PAC); 4 L/min (iVAPS)  
English  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Air tubing technical specifications  
Air tubing  
Material  
Length  
Inner diameter  
ClimateLine heated air tubing  
Flexible plastic and electrical 2 m  
components  
Flexible plastic and electrical 1.9 m  
components  
15 mm  
ClimateLineMAX heated air  
tubing  
19 mm  
SlimLine air tubing  
Standard air tubing  
3 m air tubing  
Flexible plastic  
Flexible plastic  
Flexible plastic  
1.8 m  
2 m  
3 m  
15 mm  
19 mm  
19 mm  
Heated air tubing temperature cut-out: 41°C  
Notes:  
The manufacturer reserves the right to change these specifications without notice.  
The temperature and relative humidity settings displayed for Climate Control are not measured  
values.  
Check with your clinician/service provider before using the SlimLine air tubing with devices other  
than the S9 or H5i.  
The electrical connector end of the heated air tubing is only compatible with the H5i air outlet  
and should not be fitted to the device or mask.  
When using the SlimLine or ClimateLine above 20 cm H2O, the device optimum performance  
may not be reached if used with an antibacterial filter. The device performance must be checked  
prior to prescribing the SlimLine for use with an antibacterial filter.  
The ClimateLine or ClimateLineMAX is designed only for use with the H5i.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Humidifier performance  
The following settings have been tested at 22°C ambient temperature:  
Mask pressure  
RH output %  
Nominal system output AHa, BTPSb  
cm H2O  
Setting 3  
Setting 6  
Setting 3  
Setting 6  
3
90  
95  
95  
100  
100  
100  
100  
10  
11.5  
11  
18  
21  
18  
13.5  
10  
20  
25  
100  
12  
a. AH - Absolute Humidity in mg/L.  
b. BTPS - Body Temperature Pressure Saturated.  
Pneumatic flow path  
1. Flow sensor  
2. Blower  
3. Pressure sensor  
4. Mask  
5. Air tubing  
6. H5i  
7. Device  
8. Inlet filter  
Flow (maximum) at set pressures  
The following are measured at the end of the specified air tubing:  
Pressure, cm H2O  
VPAP ST and  
VPAP ST, H5i and  
VPAP ST and  
VPAP ST, H5i and  
ClimateLine,  
L/min  
Standard, L/min Standard, L/min  
SlimLine, L/min  
4
200  
200  
200  
200  
190  
180  
170  
170  
170  
170  
170  
161  
195  
190  
184  
175  
168  
144  
170  
170  
170  
170  
161  
125  
8
12  
16  
20  
25  
English  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Displayed values  
Value  
Range  
Display resolution  
Pressure sensor at air outlet  
Mask pressure  
4-20 cm H2O (CPAP); 2-25 cm  
H2O (S, ST, T, PAC, iVAPS)  
0.1 cm H2O  
Flow derived values  
Leak  
Tidal volume  
Respiratory rate  
Minute ventilation  
Ti  
0–200 L/min  
0–4000 mL  
0–50 BPM  
0–30 L/min  
0.1–4.0 sec  
1:50–2:1  
1 L/min  
1 mL  
1 BPM  
0.1 L/min  
0.1 sec  
0.1  
I:E ratio  
Value  
Accuracya  
Pressure measurementa  
Mask pressure  
0.5 cm H2O (+4% of measured value)  
Flow measurementsa  
Leakb  
12 L/min or 20% of reading, whichever is greater, at 0 to  
60 L/min  
20%  
1 BPM  
20%  
Tidal volumeb.c  
Respiratory rateb,c  
Minute ventilationb, c  
a. Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).  
b. Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL  
or minute ventilation <3 L/min.  
c. Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support  
Devices (Figure 101 and Table 101) using nominal ResMed mask vent flows.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Pressure accuracy  
Maximum static pressure variation at 10 cm H2O according to EN ISO 17510-1:2009  
Standard air tubing  
SlimLine air tubing  
Without H5i  
With H5i  
9.89 cm H2O to 9.97 cm H2O  
9.82 cm H2O to 9.98 cm H2O  
9.76 cm H2O to 9.87 cm H2O  
9.78 cm H2O to 9.88 cm H2O  
Maximum dynamic pressure variation according to EN ISO 17510-1:2009  
Pressure (cm H2O)  
10 BPM  
15 BPM  
20 BPM  
VPAP ST and Standard air tubing without H5i / VPAP ST and Standard air  
tubing with H5i  
4
8
12  
16  
20  
25  
0.18 / 0.18  
0.21 / 0.20  
0.21 / 0.20  
0.22 / 0.21  
0.23 / 0.22  
0.30 / 0.31  
10 BPM  
0.30 / 0.30  
0.26 / 0.24  
0.26 / 0.23  
0.27 / 0.26  
0.26 / 0.28  
0.54 / 0.50  
15 BPM  
0.51 / 0.51  
0.38 / 0.36  
0.34 / 0.31  
0.36 / 0.33  
0.38 / 0.35  
0.74 / 0.71  
20 BPM  
Pressure (cm H2O)  
VPAP ST and SlimLine air tubing without H5i / VPAP ST and SlimLine air  
tubing with H5i  
4
8
12  
16  
20  
25  
0.22 / 0.20  
0.23 / 0.19  
0.22 / 0.21  
0.22 / 0.23  
0.24 / 0.27  
0.31 / 0.31  
0.28 / 0.29  
0.32 / 0.29  
0.35 / 0.29  
0.41 / 0.33  
0.37 / 0.34  
0.50 / 0.54  
0.47 / 0.53  
0.41 / 0.42  
0.41 / 0.45  
0.44 / 0.50  
0.48 / 0.50  
0.78 / 0.84  
Symbols  
The following symbols may appear on your product or packaging.  
Caution; Read instructions before use; Protection against insertion of fingers and against  
vertically dripping water; Type BF equipment; Class II equipment; Start/Stop;  
Manufacturer; European RoHS; Batch code; Catalogue number;  
Serial number;  
English 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Direct current;  
control logo 2;  
Lock/unlock;  
China pollution control logo 1;  
Not drip proof;  
China pollution  
Keep dry;  
European Authorised Representative;  
Environmental information  
WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and  
electronic equipment. This device should be disposed of separately, not as unsorted municipal waste.  
To dispose of your device, you should use appropriate collection, reuse and recycling systems  
available in your region. The use of these collection, reuse and recycling systems is designed to  
reduce pressure on natural resources and prevent hazardous substances from damaging the  
environment.  
If you need information on these disposal systems, please contact your local waste administration.  
The crossed-bin symbol invites you to use these disposal systems. If you require information on  
collection and disposal of your ResMed device please contact your ResMed office, local distributor or  
Servicing  
The VPAP ST device is intended to provide safe and reliable operation when operated in accordance  
with the instructions provided by ResMed. ResMed recommends that the VPAP ST device be  
inspected and serviced by an authorised ResMed Service Centre if there is any sign of wear or  
concern with device function. Otherwise, service and inspection of the devices generally should not  
be required during the five year design life of the device.  
Limited warranty  
ResMed Ltd (hereafter 'ResMed') warrants that your ResMed product shall be free from defects in  
material and workmanship from the date of purchase for the period specified below.  
Product  
Warranty period  
Mask systems (including mask frame, cushion, headgear and 90 days  
tubing)—excluding single-use devices  
Accessories—excluding single-use devices  
Flex-type finger pulse sensors  
Humidifier water tubs  
Batteries for use in ResMed internal and external battery  
systems  
6 months  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Product  
Warranty period  
Clip-type finger pulse sensors  
1 year  
CPAP and bilevel device data modules  
Oximeters and CPAP and bilevel device oximeter adapters  
Humidifiers and humidifier cleanable water tubs  
Titration control devices  
CPAP, bilevel and ventilation devices (including external  
power supply units)  
2 years  
Battery accessories  
Portable diagnostic/screening devices  
This warranty is only available to the initial consumer. It is not transferable.  
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the  
defective product or any of its components.  
This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse,  
modification or alteration of the product; b) repairs carried out by any service organization that has not  
been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due  
to cigarette, pipe, cigar or other smoke; and d) any damage caused by water being spilled on or into  
an electronic device.  
Warranty is void on product sold, or resold, outside the region of original purchase.  
Warranty claims on defective product must be made by the initial consumer at the point of purchase.  
This warranty replaces all other expressed or implied warranties, including any implied warranty of  
merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on  
how long an implied warranty lasts, so the above limitation may not apply to you.  
ResMed shall not be responsible for any incidental or consequential damages claimed to have  
resulted from the sale, installation or use of any ResMed product. Some regions or states do not  
allow the exclusion or limitation of incidental or consequential damages, so the above limitation may  
not apply to you.  
This warranty gives you specific legal rights, and you may also have other rights which vary from  
region to region. For further information on your warranty rights, contact your local ResMed dealer or  
ResMed office.  
English  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNINGS  
Read the entire manual before using the device.  
Use the device only as directed by your physician or healthcare provider.  
Use the device only for the intended use as described in this manual. Advice contained in this  
manual should not supersede instructions given by the prescribing physician.  
If you notice any unexplained changes in the performance of the device, if it is making unusual  
or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled  
into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed  
Service Center.  
Beware of electrocution. Do not immerse the device, humidifier, power supply or power cord in  
water. In the event of a spill, disconnect the device from the power supply and let the parts dry.  
Always unplug the device before cleaning and make sure that all parts are dry before plugging in  
the device.  
Explosion hazard—do not use in the vicinity of flammable anesthetics.  
Make sure the power cord and plug are in good condition and the equipment is not damaged.  
Keep the power cord away from hot surfaces.  
The device should only be used with masks (and connectors1) recommended by ResMed, or by  
a physician or respiratory therapist. A mask should not be used unless the device is turned on.  
Once the mask is fitted, ensure that the device is blowing air. The vent hole or holes associated  
with the mask should never be blocked.  
Explanation: The device is intended to be used with special masks (or connectors) which have  
vent holes to allow continuous flow of air out of the mask. When the device is turned on and  
functioning properly, new air from the device flushes the exhaled air out through the mask vent  
holes. However, when the device is not operating, insufficient fresh air will be provided through  
the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than  
several minutes can, in some circumstances, lead to suffocation. This applies to most models of  
CPAP or bilevel devices.  
Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an  
open flame.  
Always ensure that the device is turned on and airflow generated before the oxygen supply is  
turned on. Always turn the oxygen supply off before the device is turned off, so that unused  
oxygen does not accumulate within the device enclosure and create a risk of fire.  
1 Ports may be incorporated into the mask or in connectors that are near the mask.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Do not leave long lengths of air tubing around the top of your bed. It could twist around your  
head or neck while you are sleeping.  
Do not use electrically conductive or antistatic air tubings.  
Do not use the air tubing if there are any visible signs of damage.  
Only ResMed air tubing and accessories should be used with the device. A different type of air  
tubing or accessory may alter the pressure you actually receive, reducing the effectiveness of  
the treatment.  
Only use the ResMed 90W or 30W power supply unit. Use the 90W power supply unit to power  
the system comprising the device, H5i, air tubing, DC/DC converter and battery pack. The 30W  
power supply unit is designed to power the device only and recommended for travelling.  
Only ResMed products are designed to be connected to the module connector port. Connecting  
other devices could damage the device.  
Blocking the air tubing and/or air inlet of the device while in operation could lead to overheating  
of the device.  
CAUTIONS  
Do not open the device enclosure. There are no user serviceable parts inside. Repairs and  
servicing should only be performed by an authorised ResMed service agent.  
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or antibacterial  
soaps or scented oils to clean the device, humidifier or air tubing. These solutions may cause  
damage and reduce the life of these products.  
Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is  
correctly set up.  
Be careful not to place the device where it can be bumped or where someone is likely to trip  
over the power cord.  
Make sure that the area around the device is dry and clean and clear of bedding, clothes or other  
objects that could block the air inlet or cover the power supply unit.  
Ensure that the device is protected against water if used outdoors. Enclose the device in the S9  
travel bag for transport.  
English  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Respiratory Care Solutions  
Making quality of care easy  
Download from Www.Somanuals.com. All Manuals Search And Download.  
368551/2 2013-02  
VPAP ST  
User  
EUR1 EUR3  
B
1
2
3
4
5
C
1
2
3
4
Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia. Distributed by: ResMed  
Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA.  
ResMed (UK) Ltd 96 Milton Park Abingdon  
ResMed Ltd and S9, ClimateLine, SlimLine, SmartStart and VPAP are registered in U.S. Patent and Trademark Office.  
© 2013 ResMed Ltd.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Global leaders in sleep and respiratory medicine www.resmed.com  

Rheem Water Heater 22DV40S 36N User Guide
Roadmaster Car Stereo System VRCD220S User Guide
Rockford Fosgate Car Stereo System RFX9110 User Guide
Roland Electronic Keyboard HP 203 User Guide
Samsung DVD Player DVD V85K User Guide
Samsung Satellite TV System CT 15K9MJ User Guide
Sanus Systems TV Mount 6901 100068 User Guide
Schlage Door BE365F User Guide
Sennheiser Headphones E 180 Stereo User Guide
Sennheiser Headphones HD 250 User Guide