| 	
		 Manual Inflation Blood Pressure Monitor   
					Model M1 Plus   
					• Instruction Manual   
					• Mode d’emploi   
					EN   
					FR   
					DE   
					IT   
					• Gebrauchsanweisung   
					• Manuale di istruzioni   
					• Manual de instrucciones   
					• Gebruiksaanwijzing   
					• ꢀYꢃTHTJWXHT ꢁT ꢂꢃWꢁQYFXF\NN   
					• Instrukcja   
					ES   
					NL   
					RU   
					PL   
					HU   
					• Kezelési Útmutató   
					IM-HEM-4011C-E-03-10/2011   
					1649533-0C   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Introduction   
					Thank you for purchasing the OMRON M1 Plus Upper Arm   
					Blood Pressure Monitor.   
					The OMRON M1 Plus is a manual inflation blood pressure   
					monitor, operating on the oscillometric principle. It measures   
					your blood pressure and pulse rate simply and quickly.   
					The unit also stores up to 21 measurements in memory and   
					calculates an average reading based on the three most recent   
					measurements taken within 10 minutes of the last reading. (If   
					there are only two readings in memory for that period, the   
					average will be based on two readings. If there is only one   
					reading in memory for that period, that reading will be   
					displayed as the average value.)   
					Please read this instruction manual thoroughly before   
					using the unit. For specific information about your own   
					blood pressure, CONSULT YOUR DOCTOR.   
					EN   
					3 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Important Safety Information   
					Consult your doctor during pregnancy, arrhythmia and   
					arteriosclerosis.   
					You should never change the dosage of medication prescribed by   
					your doctor.   
					Use the M1 Plus only for yourself.   
					Caution:   
					• Do not leave the unit unattended with infants or persons who   
					cannot express their consent.   
					• Do not use the unit for any purpose other than measuring   
					blood pressure.   
					• Do not use a cellular phone near the unit. This may result in   
					incorrect operation of the unit.   
					• Do not operate unit in a moving vehicle (car, airplane).   
					• Do not apply strong shocks and vibrations to or drop the main   
					unit and arm cuff.   
					• Only use four “AAA” manganese or alkaline batteries with this   
					unit. Do not use other types of batteries.   
					• When using batteries, the battery liquid may leak and damage   
					the main unit.   
					• Please observe the following points.   
					- Do not insert the batteries with their polarities incorrectly aligned.   
					- Replace old batteries with new ones immediately.   
					- When you are not going to use the unit for a long period of time   
					(approximately three months or more), take out the batteries.   
					- Do not use new and used batteries together.   
					- Do not use batteries of a different type together.   
					• Do not forcibly bend the arm cuff or bend the air tube   
					excessively.   
					• When removing the air tube, pull on the edge of the tube at   
					the connection with the main unit not the middle of the tube.   
					• Do not inflate the arm cuff when it is not wrapped around your   
					arm.   
					4 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				1. Overview   
					Main unit   
					A 
					F 
					E 
					D 
					B 
					C 
					C 
					A. Display   
					F. Battery compartment   
					B. SET button   
					C. Air connector   
					D. MEMORY button   
					E. I/O button   
					EN   
					5 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				1.Overview   
					Arm Cuff   
					G 
					H 
					G. Arm cuff   
					(Medium cuff: arm circumference 22-32 cm)   
					H. Air Tube   
					If air starts to leak from the arm cuff, please replace with a new   
					one. (Refer to Chapter 7 “Optional Parts”)   
					Inflation Bulb   
					J 
					I 
					K 
					I. Air Release Button   
					J. Air Tube   
					K. Air Inflation Bulb   
					6 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				1.Overview   
					Display   
					L 
					M 
					N 
					R 
					T 
					O 
					S 
					U 
					P 
					Q 
					V 
					W 
					L. Systolic blood pressure   
					M. Diastolic blood pressure   
					N. Memory symbol   
					R. Heartbeat symbol   
					1. (Flashes during measurement)   
					2. (If flashing after measurement   
					is complete, indicates blood   
					pressure out of recommended   
					range)   
					(Displayed when viewing   
					values stored in memory)   
					O. Average value symbol   
					(Displayed when viewing   
					average value for the last three   
					measurements)   
					S. Pulse   
					T. Deflation symbol   
					U. Date/Time   
					P. Buzzer symbol   
					Q. Battery low symbol   
					V. Irregular heartbeat symbol   
					W. Reinflation symbol   
					Package contents   
					EN   
					X 
					Y 
					X. Storage case   
					Y. Four “AAA” alkaline (LR03)   
					batteries   
					• Instruction manual   
					• Guarantee card   
					• Blood pressure pass   
					• Blood Pressure Monitor unit   
					7 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2. Preparation   
					2.1 Installing/Replacing the Batteries   
					1. Turn the main unit upside down.   
					2. Slide the battery cover in the   
					direction of the arrow while   
					pressing the ribbed part of the   
					cover.   
					3. Install or replace four “AAA”   
					size batteries so that the   
					+ (positive) and - (negative)   
					polarities match the polarities   
					indicated on the battery   
					compartment.   
					4. Put the battery cover back in   
					place.   
					Slide the battery cover as indicated, until it clicks into place.   
					Note: The measurement values continue to be stored in   
					memory (for maximum 30 seconds) even after the   
					batteries are replaced.   
					8 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.Preparation   
					Battery Life & Replacement   
					If the battery low symbol ( ) appears on the display, replace   
					all four batteries at the same time.   
					- When the battery low symbol ( ) starts to blink, you will still   
					be able to use the unit for a short while. You should replace   
					the batteries with new ones ahead of time.   
					- When the symbol ( ) remains lit, the batteries are   
					exhausted. You should replace the batteries with new ones   
					at once.Turn the unit off before replacing the batteries.   
					• Remove the batteries if the unit will not be used for three   
					months or more.   
					• If the batteries are removed for 30 seconds, the Date/Time   
					setting will need to be reset. See “2.2 Setting the Date and   
					Time” for details.   
					• Dispose of batteries according to applicable local regulations.   
					Four new “AAA” alkaline batteries will last for approximately   
					1500 measurements, when used to take two measurements a   
					day.   
					EN   
					Since the supplied batteries are for monitoring use only, they   
					may have a shorter life and not last for 1500 measurements.   
					9 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.Preparation   
					2.2 Setting the Date and Time   
					Your blood pressure monitor automatically stores up to 21   
					measurements values, including date and time of   
					measurement, in its memory and calculates an average value   
					based on the three most recent measurements taken within 10   
					minutes of the last reading. (If there are only two readings in   
					memory for that period, the average will be based on two   
					readings. If there is only one reading in memory for that period,   
					that reading will be displayed as the average value.)   
					To make use of the memory and average values functions:   
					• Set the unit to the correct date and time before taking a   
					measurement for the first time.   
					• If the batteries have been removed for more than 30 seconds,   
					the date and time setting will need to be reset.   
					1. When the I/O button is   
					pressed to turn the unit on for   
					the first time after inserting   
					batteries, the year digits   
					(2005) will flash on the   
					display.   
					2. Press the MEMORY button to   
					advance the digits one at a   
					time.   
					Notes:   
					• The range for the year   
					setting is 2005 to 2030. If the   
					year reaches 2030, it will   
					return to 2005.   
					• If you hold down the   
					MEMORY button, the digits   
					will advance rapidly.   
					10   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.Preparation   
					3. Press the SET button to confirm the setting when the desired   
					number appears on the display.   
					The year is set and the month digits flash on the display.   
					4. Repeat steps 2 and 3 to set   
					the month.   
					The month is set and the day   
					digits flash on the display.   
					5. Repeat steps 2 and 3 to set   
					the day.   
					The day is set and the hour   
					digits flash on the display.   
					6. Repeat steps 2 and 3 to set   
					the hour.   
					The hour is set and the   
					minutes digits flash on the   
					display.   
					EN   
					11   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.Preparation   
					7. Repeat steps 2 and 3 to set   
					the minutes.   
					The minutes settings is set.   
					The unit automatically   
					switches to standby mode   
					after the minute setting has   
					been set.   
					To adjust the date and time, press the SET button twice while   
					the monitor is in standby mode.   
					Notes:   
					• After the buzzer setup is completed, the time/date setting will   
					flash on the display.   
					• After each setup is completed, please push the I/O button to   
					turn off the monitor.   
					12   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.Preparation   
					2.3 Adjusting the Settings Modes   
					You can modify the options for the various settings of your unit.   
					This is done by pressing the SET button to select a setting,   
					then pressing the MEMORY button to select the options for that   
					setting. After selecting a setting, press the I/O button to confirm   
					the setting and turn the power off.   
					Buzzer Mode   
					When Buzzer mode is on, the unit emits sounds, during   
					measurement (to indicate that pulse waves are detected), and   
					when measurement is complete.   
					1. Press the SET button once   
					while the unit is in standby   
					mode.   
					The buzzer symbol   
					appears on the display.   
					2. Press the MEMORY button to select   
					“on” or “off”.   
					The setting alternates between “on”   
					and “off” with each press of the   
					MEMORY button.   
					EN   
					The default setting is “on”.   
					3. When you have selected the desired setting, press the I/O   
					button to turn the unit off.   
					Notes:   
					• If you want to adjust the date or time, press the SET button   
					after setting the buzzer setting, follow steps for setting date   
					and time (Refer to 2.2) press the I/O button to turn the unit off.   
					• If you forget to turn the unit off, it will automatically shut itself   
					off after five minutes.   
					13   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3. Using the Unit   
					3.1 How to Sit Correctly When Taking a Measurement   
					You can take a measurement on either your left or right arm.   
					Note:   
					• Measurements should be taken in a quiet place and you   
					should be in a relaxed, seated position. Make sure that the   
					room is not too hot or cold.   
					• Do not take measurement after bathing, drinking alcohol, or   
					exercising.   
					• Do not move or talk during measurement.   
					Correct   
					Posture   
					Remove tight fitting clothing from   
					your upper arm, and any thick   
					clothing such as a sweater.   
					Do not place the cuff over thick   
					clothes and do not roll up your sleeve   
					if it is too tight.   
					Sit upright   
					with your   
					back straight.   
					Place your arm on a table so   
					that the cuff will be at the same   
					level as your heart.   
					The gap between the   
					chair and the top of the   
					table should be between   
					25cm to 30cm.   
					14   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					Notes:   
					• Correct posture during measurement is necessary to get   
					accurate results.   
					• You should also try to measure your blood pressure at the   
					same time each day. (Within 1 hour after waking up is   
					recommended.)   
					Incorrect Posture   
					• Arched back (leaning   
					forwards)   
					• Sitting cross-legged   
					• Sitting on a sofa or at a low   
					table so that you lean forward   
					These situations tend to lead to higher blood pressure values due to   
					strain or the arm cuff being lower than the heart.   
					If the arm cuff is at a lower position than your heart use cushions   
					etc., to adjust the height of your arm.   
					EN   
					15   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					3.2 Applying the Arm Cuff   
					Be sure to wrap the arm cuff correctly so that you get accurate   
					results.   
					Measurements can be taken in light clothing. However, please   
					remove thick clothes, such as sweaters, before taking a   
					reading.   
					1. Connect the air tube of the arm cuff and the inflation bulb to   
					their respective air connectors.   
					2. Put your left arm through the cuff   
					loop.   
					The air tube should run down the   
					inside of your forearm and be in line   
					with your middle finger.   
					Note: If the cuff is not assembled,   
					pass the end of the cuff   
					furthest from the tubing   
					through the metal D-ring to   
					form a loop. The smooth cloth   
					should be on the inside of the   
					cuff loop.   
					16   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					3. Position the arm correctly.   
					1) The air tube should run   
					down the inside of your   
					forearm and be in line with   
					your middle finger.   
					2) The bottom of the cuff   
					should be approximately 1   
					to 2 cm above your elbow.   
					3) Apply the cuff to your   
					upper arm so that the   
					coloured marker (blue   
					arrow under tube) is   
					centered on the middle of   
					your inner arm and points   
					down the inside of the   
					arm.   
					Taking measurements on the right arm   
					Apply the cuff so that the   
					air tube is at the side of   
					your elbow.   
					• Be careful not to rest   
					your arm on the air tube,   
					or otherwise restrict the   
					flow of air to the cuff.   
					EN   
					1 to 2 cm   
					• The cuff should be 1 to   
					2 cm above the elbow.   
					4. When the cuff is positioned correctly, close the fabric fastener   
					FIRMLY.   
					17   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					3.3 Taking a Reading   
					1. Turning on the unit.   
					Note: Press the air release button to release any air in the arm cuff.   
					1) Press the I/O button to turn the unit on.   
					2) All items in the display will be displayed briefly.   
					The battery symbol will also be displayed for test purposes.   
					This does not mean the batteries are empty.   
					3) The current date is displayed and the deflation symbol will flash.   
					4) Finally, the heartbeat symbol will appear and the current time   
					and 0 is displayed.   
					Note: If the deflation symbol does not disappear soon, press   
					the air release button to release any air in the arm cuff.   
					2. Pump the inflation bulb to inflate the arm cuff.   
					1) Inflate the cuff until it is 30 to 40   
					mmHg above your expected   
					systolic blood pressure value.   
					ex) If your expected blood   
					pressure is around 140mmHg,   
					inflate the arm cuff to between 170   
					and 180 mmHg. Inflate the cuff   
					rapidly so that the pressure is   
					reached in about five seconds.   
					2) When the desired pressure has been achieved, release the   
					inflation bulb. Remain still and do not talk.   
					18   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					• The current time continues to be displayed while the cuff is   
					being inflated.   
					Note: Do not inflate the arm cuff more than necessary.   
					3. Measurement starts.   
					Measurement starts automatically   
					after you stop inflating the arm cuff.   
					Decreasing numbers appear on the   
					display and the heartbeat symbol   
					flashes.   
					Remain still and do not talk during measurement.   
					During measurement, the buzzer (if set to “on”) will beep in   
					rhythm with your heartbeat.   
					Note: To cancel a measurement, press the I/O button to turn   
					off the unit and press the air release button to release   
					the air in the arm cuff.   
					Common Causes of Inaccurate Results   
					Arm cuff is too loose.   
					Movement or talking   
					Leaning forwards.   
					during measurement.   
					EN   
					Note: If the arm cuff is at a lower position than   
					your heart use cushions etc., to adjust the   
					height of your arm.   
					19   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					The deflation symbol flashes on the   
					display. Blood pressure and pulse   
					rate values are displayed.   
					4. Measurement ends.   
					Press the air release button to   
					release the air in the arm cuff until   
					the deflation symbol is no longer   
					displayed.   
					5. Check the measurement results.   
					Systolic blood   
					pressure   
					The unit automatically stores blood   
					pressure and pulse rate into its   
					memory, including date and time of   
					measurement. Refer to “3.4 Using   
					the Memory Function”.   
					Diastolicblood   
					pressure   
					Pulse   
					The buzzer will sound (if switched   
					on) when measurement is completed.   
					Notes:   
					• Self-diagnosis of measured results and treatment are   
					dangerous. Please follow the instructions of your doctor.   
					• Wait 2-3 minutes before taking another blood pressure   
					measurement. Waiting between readings allows the arteries   
					to return to the condition prior to taking the blood pressure   
					measurement.   
					6. Undo the fastener and remove the arm cuff.   
					7. Press the I/O button to turn the unit off.   
					Note: If you forget to turn the unit off, it will automatically shut   
					itself off after five minutes.   
					20   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					Important:   
					• If your systolic or diastolic pressure is outside   
					the standard range, the heartbeat symbol will   
					blink when the measurement result is   
					displayed.   
					Recent research suggests that the following   
					values can be used as a guide to high blood   
					pressure for measurements taken at home.   
					Systolic Blood Pressure   
					Above 135 mmHg   
					Diastolic Blood Pressure   
					Above 85 mmHg   
					This criteria is for home blood pressure measurement.   
					For professional office blood pressure measurement criteria,   
					please refer to Chapter 9 “Some Useful Information about   
					Blood Pressure”   
					• Your blood pressure monitor includes an   
					irregular heartbeat feature. Irregular heartbeats   
					can influence the results of the measurement.   
					The irregular heartbeat algorithm automatically   
					determines if the measurement is usable or   
					needs to be repeated. If the measurement   
					results are affected by irregular heartbeats but   
					the result is valid, the result is shown together   
					EN   
					with the irregular heartbeat symbol (   
					heartbeats cause the measurement to be invalid, no result is   
					shown. If the irregular heartbeat symbol ( ) is shown after   
					). If the irregular   
					your have taken a measurement, repeat the measurement. If   
					the irregular heartbeat symbol ( ) is shown frequently,   
					please make your doctor aware of it.   
					21   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					What is Irregular Heartbeat?   
					Normal Heartbeat   
					Irregular Heartbeat   
					An irregular heartbeat is a   
					heartbeat rhythm that varies   
					by more than 25% from the   
					average heartbeat rhythm   
					detected while the unit is   
					measuring the systolic and   
					diastolic blood pressure.   
					Pulse   
					Blood pressure   
					Long   
					Short   
					Pulse   
					If such an irregular rhythm is   
					Blood pressure   
					detected more than twice   
					during measurement, the irregular heartbeat symbol (   
					) 
					appears on the symbol when the measurement results are   
					displayed.   
					What is Arrhythmia?   
					A heartbeat is stimulated by electrical signals that cause the   
					heart to contract.   
					Arrhythmia is a condition where the heartbeat rhythm is   
					abnormal due to flaws in the bio-electrical system that drives   
					the heartbeat. Typical symptoms are skipped heartbeats,   
					premature contraction, an abnormally rapid (tachycardia) or   
					slow (bradycardia) pulse. This can be caused by heart disease,   
					aging, physical predisposition, stress, lack of sleep, fatigue etc.   
					Arrhythmia can only be diagnosed by a doctor through a   
					special examination.   
					Whether the appearance of the irregular heartbeat symbol   
					( 
					) in the results indicates arrhythmia or not can only be   
					determined by an examination and diagnosis by your doctor.   
					If the irregular heartbeat symbol (   
					) is shown frequently,   
					please make your doctor aware of it. Conducting self-diagnosis   
					and treatment based on measurement results is dangerous. Be   
					sure to follow the instructions of your doctor.   
					22   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					3.4 Using the Memory Function   
					This unit has a memory capable of storing 21 sets of readings.   
					Every time you complete the measurement, the unit   
					automatically stores the blood pressure and pulse rate.   
					If there are three readings within 10 minutes of the last reading,   
					the average of the three readings is displayed. (If there are only   
					two readings in memory for that period, the average will be based   
					on two readings. If there is only one reading in memory for that   
					period, that reading will be displayed as the average value.)   
					Note: When 21 sets of readings are stored in memory, the   
					oldest set will be deleted to store a new set.   
					1. Press the O/I button to turn the   
					power on.   
					2. When the heartbeat symbol appears   
					on the display, press the MEMORY   
					button.   
					EN   
					If there are three readings stored in   
					memory that were taken within 10   
					minutes of the last reading, an   
					average value for those reading is   
					displayed. (If there are only two   
					readings in memory for that period, the average will be based   
					on two readings. If there is only one reading in memory for that   
					period, that reading will be displayed as the average value.)   
					23   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					Note: If there are no measurements results stored in   
					memory, the screen to the right is displayed.   
					3. Press the MEMORY button again,   
					while the average values screen is   
					displayed.   
					The result from the most recent   
					measurement is displayed.   
					The date and time of each   
					measurement is displayed alternately   
					with the measurement values.   
					Important: If your systolic or diastolic pressure   
					is outside the standard range, the   
					heartbeat symbol will blink when   
					the measurement result is   
					displayed. Refer to 3.3, number 7.   
					4. Press the MEMORY button   
					repeatedly to cycle through the   
					previous measurement results.   
					Keep the button pressed down to   
					cycle rapidly through the previous   
					results.   
					If the irregular heartbeat symbol was displayed at the time of a   
					measurement, this is displayed when that result is displayed.   
					5. Press the I/O button to turn the unit off.   
					If you forget to turn the unit off, it will automatically shut itself off   
					after five minutes.   
					24   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3.Using the Unit   
					To Delete All the Values Stored in Memory   
					You cannot delete individual stored readings, all the readings in   
					the unit will be deleted.   
					1. To delete stored readings, first press   
					the MEMORY button. Then while   
					holding it down, press the I/O button   
					simultaneously for about 2-3   
					seconds. All readings will then be   
					deleted.   
					2 
					1 
					Important: Be careful not to press   
					the I/O button first. If the   
					I/O button is pressed   
					first, the monitor is shut   
					off.   
					2. Press the I/O button to turn the unit off.   
					If you forget to turn the unit off, it will automatically shut itself off   
					after five minutes.   
					EN   
					Notes:   
					• If the date and time are reset to a time before the most recent   
					measurement, the average value will be based on any   
					measurements taken after the date and time were reset.   
					However, you can still view the readings stored in memory.   
					• Please note that all measurements are stored in the memory. If   
					different people measure with the same device, make sure that   
					you are aware of this fact.   
					25   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				4. Quick Reference Guide   
					Use this as a quick reference guide only. If you are using this   
					device for the first time, please read carefully Chapter 3 of this   
					Instruction Manual.   
					To help ensure a reliable reading, avoid eating, smoking, or   
					exercising for at least 30 minutes before taking a measurement.   
					Note: Remove any tight-fitting clothing from your upper arm.   
					1. Sit on a chair with your   
					feet flat on the floor and   
					place your arm on a table   
					so that the arm cuff will be   
					at the same level as your   
					heart.   
					2. Apply the arm cuff to your upper arm. The coloured marker   
					should be centred on the inside of your arm and point down the   
					inside of the arm, so that the air tube runs down the inside of   
					your forearm and is in line with your middle finger.   
					3. Secure the cuff around your arm using the fabric fastener strip.   
					4. Press the I/O button to turn the unit on, then   
					pump the inflation bulb to inflate the cuff.   
					After measurement is complete and the   
					measurement results are displayed, press   
					the air release button to release any air in   
					the cuff.   
					5. Press the I/O button to turn the unit off.   
					Note: Always wait at least 2-3 minutes before taking another   
					blood pressure measurement.   
					26   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5. Handling Errors and Problems   
					5.1 Error Messages   
					Error Display   
					Cause   
					Remedy   
					Press air release button and   
					restart the measurement with a   
					higher inflation level. Carefully   
					read steps under section 3.3.   
					Cuff is under inflated.   
					Repeat measurement. Remain   
					still and do not talk during   
					measurement.   
					Movement during   
					measurement   
					Refer to section 3.3.   
					Insert the air tube securely.   
					Refer to section 3.2.   
					Air tube disconnected.   
					Apply the arm cuff correctly.   
					Refer to section 3.2.   
					Arm cuff not applied   
					correctly.   
					Remove any clothing interfering   
					with the arm cuff.   
					Refer to section 3.2.   
					Clothing is interfering   
					with the arm cuff.   
					Replace cuff with new one.   
					Refer to Chapter 7.   
					Air is leaking from the   
					arm cuff.   
					EN   
					Release your hand from the air   
					inflation bulb before the pressure   
					reaches 299 mmHg.   
					The arm cuff was inflated   
					above 299 mmHg.   
					Refer to section 3.3.   
					Replace all four “AAA” batteries   
					with new ones.   
					Refer to section 2.1.   
					Battery power is low.   
					Blinks or appears   
					continuously   
					27   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5.Handling Errors and Problems   
					Error Display   
					Cause   
					Remedy   
					Contact your OMRON retail outlet   
					or distributor.   
					Refer to Chapter 8.   
					Device error.   
					Press the inflation bulb to inflate   
					the arm cuff until the reinflation   
					symbol goes out. Or, deflate the   
					arm cuff and repeat measurement   
					after checking that the heartbeat   
					symbol has been displayed.   
					Pressure is too low.   
					Refer to section 3.3.   
					Note: The irregular heartbeat symbol may also be displayed with error messages.   
					28   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5.Handling Errors and Problems   
					5.2 Troubleshooting   
					Problem   
					Cause   
					Remedy   
					Apply the arm cuff   
					correctly.   
					Refer to section 3.2.   
					Arm cuff not applied   
					correctly.   
					Remain still and do not talk   
					during measurement.   
					Refer to section 3.3.   
					Movement or talking   
					during measurement.   
					The reading is   
					extremely low (or high).   
					Remove any clothing   
					interfering with the arm   
					cuff.   
					Clothing is interfering   
					with the arm cuff.   
					Refer to section 3.1.   
					Make sure that the air tube   
					is connected securely.   
					Refer to section 3.2.   
					Is the air tube securely   
					connected into the main   
					unit?   
					Arm cuff pressure does   
					not rise.   
					Replace the arm cuff with   
					a new one.   
					Refer to Chapter 7.   
					Air is leaking from the   
					arm cuff.   
					Apply the cuff correctly so   
					that it is firmly wrapped   
					around the arm.   
					Arm cuff deflates too   
					soon.   
					The arm cuff is loose.   
					Refer to section 3.2.   
					EN   
					Inflate the cuff so that it is   
					30 to 40 mmHg above   
					your previous   
					Has the arm cuff been   
					inflated sufficiently?   
					measurement result.   
					Cannot measure or   
					readings are too low or   
					too high.   
					Refer to section 3.3.   
					Is the air release button   
					being pressed during   
					inflation?   
					Be careful not to press the   
					air release button during   
					measurement.   
					Replace the batteries with   
					new ones.   
					Refer to section 2.1.   
					The unit loses power   
					during measurement.   
					The batteries are empty.   
					29   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5.Handling Errors and Problems   
					Problem   
					Cause   
					Remedy   
					Replace the batteries with   
					new ones.   
					Refer to section 2.1.   
					The batteries are empty.   
					Nothing happens when   
					you press the buttons.   
					Insert the batteries with   
					the correct (+/ -) polarity.   
					Refer to section 2.1.   
					The batteries have been   
					inserted incorrectly.   
					Press the I/O button and repeat measurement.   
					If the problem continues, try replacing the batteries   
					Other problems.   
					with new ones.   
					If this still does not solve the problem, contact your   
					OMRON retail outlet or distributor.   
					30   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6. Maintenance and Storage   
					Maintenance   
					To protect your unit from damage, please observe the   
					following:   
					• Do not subject the main unit and cuff to extreme temperatures,   
					humidity, moisture or direct sunlight.   
					• Do not fold the cuff or tubing tightly.   
					• Do not inflate the arm cuff over 299 mmHg.   
					• Do not disassemble the unit.   
					• Do not subject the unit to strong shocks or vibrations (for   
					example, dropping the unit on the floor).   
					• Do not use volatile liquids to clean the main unit. The unit should   
					be cleaned with a soft, dry cloth.   
					• Use a soft, moistened cloth and soap to clean the arm cuff.   
					• Do not wash the arm cuff or immerse it in water.   
					• Do not use petrol, thinners or similar solvents to clean the arm   
					cuff.   
					EN   
					• Do not carry out repairs of any kind yourself. If a defect   
					occurs, consult your OMRON retail outlet or distributor as   
					mentioned on the packaging.   
					31   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6.Maintenance and Storage   
					Calibration and Service   
					• The accuracy of this blood pressure monitor has been   
					carefully tested and is designed for a long service life.   
					• It is generally recommended to have the unit inspected every   
					two years to ensure correct functioning and accuracy. Please   
					consult your authorised OMRON dealer or the OMRON   
					Customer Service at the address given on the packaging or   
					attached literature.   
					• Do not carry out any repairs yourself. If a defect occurs or you   
					have doubts about the correct functioning of the device,   
					consult your authorised OMRON dealer or the OMRON   
					Customer Service.   
					32   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6.Maintenance and Storage   
					Storage   
					Keep the unit in its storage case when not in use.   
					1. Unplug the air tube from the air connector.   
					2. Gently fold the air tube into the arm   
					cuff.   
					Note: Do not bend the air tube   
					excessively.   
					3. Place the arm cuff, inflation bulb   
					and main unit in the storage case.   
					Do not store the unit in the following situations:   
					• If the unit is wet.   
					EN   
					• Locations exposed to extreme temperatures, humidity, direct   
					sunlight, dust or corrosive vapours.   
					• Locations exposed to vibrations, shocks or where it will be at   
					an angle.   
					• Locations exposed to chemicals or corrosive vapours.   
					33   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				7. Optional Parts   
					Medium Arm Cuffs   
					Large Arm Cuff   
					Small Arm Cuff   
					Arm circumference 17 - 22 cm   
					Arm circumference 22 - 32 cm Arm circumference 32 - 42 cm   
					CM-4997086-7   
					CL-4997065-4   
					CS-4997076-0   
					CM1-7935058-8   
					Small cuff and bulb combination   
					Regular bulb   
					Arm circumference 17 - 22 cm   
					4997099-9   
					4997965-1   
					34   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				8. Technical Data   
					Product Description   
					Model   
					Display   
					Manual Inflation Blood Pressure Monitor   
					OMRON M1 Plus (HEM-4011C-E)   
					LCD Digital Display   
					Measurement Method   
					Oscillometric method   
					Pressure: 0 mmHg to 299 mmHg   
					Pulse: 40 to 180/min.   
					21 Measurements with date and time   
					Measurement Range   
					Memory   
					Pressure: ꢀ3 mmHg   
					Pulse: ꢀ 5% of display reading   
					Accuracy   
					Inflation   
					Manual by inflation bulb   
					Deflation   
					Pressure Detection   
					Power Source   
					Automatic pressure release valve   
					Capacitive pressure sensor   
					4 “AAA” batteries 1.5V   
					Capacity of new alkaline batteries is approx. 1500   
					Battery life   
					measurements   
					+10°C to +40°C   
					Maximum: 30 to 85% RH   
					Operating temperature/   
					Humidity   
					-20°C to +60°C   
					Maximum: 10 to 95% RH   
					Storage temperature   
					Humidity   
					Console Weight   
					Cuff Weight   
					Approximately 210g without batteries   
					Approximately 130g   
					Outer Dimensions   
					Approximately 106 (w) mm ꢁ 80 (h) mm ꢁ 136 (l) mm   
					Approximately 146 mm ꢁ 446 mm   
					Cuff Dimensions   
					Package Content   
					(Medium cuff: arm circumference 22 to 32 cm)   
					Medium cuff, instruction manual, storage case, battery set,   
					guarantee card, blood pressure pass, regular bulb.   
					EN   
					Note: Subject to technical modification without prior notice   
					• This OMRON product is produced under the strict quality system of OMRON Healthcare   
					Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the   
					Pressure Sensor, is produced in Japan for assembly.   
					• Disposal of this product and used batteries should be carried out in accordance   
					with the national regulations for the disposal of electronic products.   
					= Type B   
					This device fulfils the previsions of EC directive 93/42/EEC (Medical Device   
					Directive). This blood pressure monitor is designed according to the European Standard   
					EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3:   
					Supplementary requirements for electromechanical blood pressure measuring systems.   
					Caution: Please read the instruction manual carefully   
					before using the device.   
					35   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				8.Technical Data   
					Important information regarding Electro Magnetic Compatibility (EMC)   
					With the increased number of electronic devices such as PC’s and mobile (cellular)   
					telephones, medical devices in use may be susceptible to electromagnetic   
					interference from other devices. Electromagnetic interference may result in   
					incorrect operation of the medical device and create a potentially unsafe situation.   
					Medical devices should also not interfere with other devices.   
					In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with   
					the aim to prevent unsafe product situations, the EN60601-1-2 standard has been   
					implemented. This standard defines the levels of immunity to electromagnetic   
					interferences as well as maximum levels of electromagnetic emissions for medical   
					devices.   
					This medical device manufactured by OMRON Healthcare conforms to this   
					EN60601-1-2:2001 standard for both immunity and emissions.   
					Nevertheless, special precautions need to be observed:   
					• Do not use mobile (cellular) telephones and other devices, which generate strong   
					electrical or electromagnetic fields, near the medical device. This may result in   
					incorrect operation of the unit and create a potentially unsafe situation. Recom-   
					mendation is to keep a minimum distance of 7 m. Verify correct operation of the   
					device in case the distance is shorter.   
					Further documentation in accordance with EN60601-1-2:2001 is available at   
					OMRON Healthcare Europe at the address mentioned in this instruction manual.   
					
					Correct Disposal of This Product   
					(Waste Electrical & Electronic Equipment)   
					This marking shown on the product or its literature, indicates that it should not be   
					disposed of, with other household wastes at the end of its working life. To prevent   
					possible harm to the environment or human health from uncontrolled waste   
					disposal, please separate this from other types of wastes and recycle it responsibly   
					to promote the sustainable reuse of material resources.   
					Household users should contact either the retailer where they purchased this   
					product, or their local government office, for details of where and how they can take   
					this item for environmentally safe recycling.   
					Business users should contact their supplier and check the terms and conditions of   
					the purchase contract. This product should not be mixed with other commercial   
					wastes for disposal.   
					This product does not contain any hazardous substances.   
					36   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				9. Some Useful Information about Blood Pressure   
					What is Blood Pressure?   
					Blood pressure is a measure of the force of blood flowing   
					against the walls of the arteries. Arterial blood pressure is   
					constantly changing during the course of the heart’s cycle.   
					The highest pressure in the cycle is called the Systolic Blood   
					Pressure; the lowest is the Diastolic Blood Pressure.   
					Both pressure readings, the Systolic and Diastolic, are   
					necessary to enable a doctor to evaluate the status of a   
					patient’s blood pressure.   
					Why is it a Good Thing to measure Blood Pressure at   
					Home?   
					Having your blood pressure measured by a doctor can cause   
					anxiety which is itself a cause of high blood pressure. As a   
					variety of conditions affect blood pressure, a single   
					measurement may not be sufficient for an accurate diagnosis.   
					Many factors such as physical activity, anxiety, or the time of   
					day, can influence your blood pressure. Thus it is best to try   
					and measure your blood pressure at the same time each day,   
					to get an accurate indication of any changes in blood pressure.   
					Blood pressure is typically low in the morning and increases   
					from afternoon to evening. It is lower in the summer and higher   
					in the winter.   
					EN   
					Blood pressure is measured in millimetres of mercury (mmHg)   
					and measurements are written with the systolic pressure   
					before the diastolic e.g. A blood pressure written as 135/85, is   
					referred to as 135 over 85 mmHg.   
					37   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				9.Some Useful Information about Blood Pressure   
					Upper curve: systolic blood pressure   
					Lower curve: diastolic blood pressure   
					200   
					150   
					100   
					50   
					6 
					12   
					18   
					24   
					Time of day   
					Example: fluctuation within a day (male, 35 years old)   
					Classification of Blood Pressure by the World Heath   
					Organization   
					The World Health Organization (WHO) and the International   
					Society of Hypertension (ISH) developed the Blood Pressure   
					Classification shown in this figure.   
					Systolic blood pressure   
					(mmHg)   
					Severe hypertension   
					180   
					Moderate hypertension   
					160   
					Mild hypertension   
					140   
					Normal systolic value   
					130   
					Normal blood   
					pressure   
					120   
					Optimal blood   
					pressure   
					(target value)   
					Diastolic   
					blood   
					pressure   
					(mmHg)   
					80   
					85   
					90   
					100 110   
					This classification is based on the blood pressure values   
					measured on people in a sitting position in outpatient   
					departments of hospitals.   
					* There is no universally accepted definition of hypotension.   
					However, those having the systolic pressure below   
					100 mmHg are assumed as hypotensive.   
					38   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				9.Some Useful Information about Blood Pressure   
					Manufacturer   
					OMRON HEALTHCARE Co., Ltd.   
					53, Kunotsubo, Terado-cho, Muko, Kyoto,   
					617-0002 JAPAN   
					EU-representative   
					OMRON HEALTHCARE EUROPE B.V.   
					Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS   
					
					OMRON (DALIAN) CO., LTD.   
					Production facility   
					Subsidiary   
					Dalian, CHINA   
					OMRON HEALTHCARE UK LTD.   
					Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.   
					OMRON MEDIZINTECHNIK   
					HANDELSGESELLSCHAFT mbH   
					John-Deere-Str. 81a, 68163 Mannheim, GERMANY   
					
					OMRON SANTÉ FRANCE SAS   
					14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE   
					Made in China   
					EN   
					39   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 |