Omega Thermometer HH803U User Manual

Input Protection:  
INTRODUCTION  
24V dc or 24V ac rms maximum input voltage on any combination of input  
This instrument is a 5 digit, compact-sized portable digital thermometer de-  
signed to use external K type and J type thermocouples. Temperature indica-  
tion follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175  
Revised to ITS-90) for K-type and J-type thermocouples. One K-type thermo-  
couple is supplied with the thermometer.  
pins.  
Maximum Differential Common Mode Voltage (Maximum Voltage T1  
during measurement): 1volt.  
Reading Rate: 1 sample/second.  
SAFETY INFORMATION  
OPERATING INSTRUCTIONS  
It is recommended that you read the safety and operation instructions before  
using the thermometer.  
OPERATING INSTRUCTIONS  
1.  
Power Switch  
The  
key turns the thermometer on or off. In the SET mode the unit can-  
MODEL HH803U  
WARNING  
not be powered off. Exit SET mode to power off.  
To avoid electrical shock, do not use this instrument when working voltages  
at the measurement surface over 24V AC or DC.  
P
SINGLE IN UT K/J  
APO function mode  
DIGITAL THERMOMETER  
Press  
power key for more than 6 seconds to disable the auto power off  
function. The display will show “APO OFF”.  
WARNING  
To avoid damage or burns, do not make temperature measurement in mi-  
crowave ovens.  
2. °C/°F Selecting the Temperature Scale  
Readings are displayed in either degrees Celsius(°C) or degrees Fahren-  
heit(°F). When the thermometer is turned on, it is set to the temperature  
scale that was in use when the thermometer was last turned off. To change  
the temperature scale, press the °C/°F key.  
CAUTION  
Repeated sharp flexing can break the thermocouple leads. To prolong lead  
life, avoid sharp bends in the leads, especially near the connector.  
3.  
Button (only Main display)  
Press this key to enter the Data Hold mode, the “HOLD” annunciator is dis-  
played. When HOLD mode is selected, the thermometer holds the present  
®
HH803U  
Input Connector:  
Accepts standard miniature thermocouple connectors (flat blades spaced  
7.9mm, center to center).  
readings and stops all further measurements. Press the  
cancel HOLD mode and resume measurement. In the MIN/MAX recording  
key again to  
mode, press  
key to stop recording. Press  
key again to resume  
ENVIRONMENTAL  
cording. (Previously recorded readings are not erased).  
Ambient Operating Ranges:  
Backlight function mode  
0°C to 50°C (32°F to 122°F) <70% R.H.  
Storage Temperature:  
-20°C to 60°C (-4°F to 140°F) <80% R.H.  
Press the  
button for more than two seconds to turn on the backlight  
Press the button again for more than two seconds to turn off the backlight  
The backlight will switch-off automatically after 30 seconds.  
4. Cold Junction Temperature Display mode (second display)  
Press C.J. TEMP key to display the thermocouple input connector cold  
junction temperature on the second display. Press C.J. TEMP key again to  
exit this mode.  
GENERAL  
Display: 5 digit liquid crystal display (LCD).  
Overload: “----.-“ or “OL” is display.  
5. K/J Input Type Selection  
Battery: 1.5V x 4 PCS (AAA SIZE) UM-4 R03.  
Battery Life: 200 hours typical with carbon zinc battery.  
Auto power off: 30 minutes  
Dimensions: 160mm(H) x 83mm(W) x 38mm(D)  
Weight: Approx. 270g including batteries.  
The K/J key selects the thermocouple type. When the thermometer is turned  
on, it is set to the type selected when the thermometer was last turned off.  
6. MIN MAX with Elapsed Time Mode  
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the  
Maximum reading with time, Minimum reading with time and Average  
reading stored in record mode). In the this mode the automatic power-off  
Supplied Thermocouples (1 per input):  
feature is disabled and  
key, °C/°F key, REL key, SET key, Hi/Lo Limits  
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insula-  
tion temperature is 482°C (900°F). Thermocouple accuracy is 1.1°C or 0.4%  
of reading (whichever is greater) from 0°C to 1250°C.  
key, K/J, C.J. TEMP and Recall keys are disabled. The beeper emits a tone  
when a new minimum or maximum value is recorded. The current tem-  
perature reading displayed on second display. Push the MIN MAX key to  
cycle through the MAX, MIN and AVG readings. If an overload is re-  
corded, the averaging function is stopped. In this mode, press the HOLD  
key to stop the recording of readings, all values are held, press again to re-  
start recording.  
SPECIFICATIONS  
ELECTRICAL  
Temperature Scale: Celsius or Fahrenheit user-selectable.  
To prevent accidental loss of MIN, MAX and AVG data, this mode can  
only be cancelled by pressing and holding the MIN MAX key for 2 seconds.  
All recorded readings are erased.  
Measurement Range:  
J-TYPE -200°C to 1050°C, (-328°F to 1922°F)  
K-TYPE -200°C to 1370°C, (-328°F to 2498°F)  
Resolution: 0.1°C or 0.2°F  
7. REL Relative mode  
Press the REL key to enter the Relative mode, zero the display, and store  
the displayed Reading as a reference value. The REL annunciator is dis-  
played. The current temperature reading displayed on the second display  
Press REL key again to exit the relative mode. The relative value can also  
be entered by the user. (See "SET mode" later in this manual.) When the de-  
sired Relative value has been entered, press REL key to enter the Relative  
mode, press SET key use the set Relative value as a reference value. Press  
REL key again to exit the relative mode. In the Relative mode, the val  
(can not > 3000.0 counts) shown on the LCD is always the difference be-  
Accuracy: Accuracy is specified for operating temperatures over the range of  
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.  
(0.05% rdg + 0.3°C) -50°C to 1370°C  
MADE IN TAIWAN  
(0.05% rdg + 0.7°C) -50°C to -200°C  
(0.05% rdg + 0.6°F) -58°F to 2498°F  
(0.05% rdg + 1.4°F) -58°F to -328°F  
Temperature Coefficient:  
0.1 times the applicable accuracy specification per °C from 0°C to 18°C and  
28°C to 50°C (32°F to 64°F and 82°F to122°F).  
tween the stored reference and the present readin  
g.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miller Electric Welding System OM 494 User Manual
Moxa Technologies Surge Protector A51 User Manual
NEC Server set810 User Manual
Nikon Microscope Magnifier LV100D User Manual
Nordyne Heat Pump T3BN User Manual
Olevia Flat Panel Television 242T FHD User Manual
OPTI UPS Surge Protector MG2T MG3T User Manual
Oregon Scientific MP3 Player VR383 User Manual
Panasonic Air Conditioner CS E9DKEW User Manual
Panasonic Cordless Drill EY6780 User Manual