Operating and Installation
Instructions
Laboratory Glassware
Washer
G 7883 CD
en - US, CA
To prevent accidents
and machine damage
read these instructions
before
installation or use.
M.-Nr. 07 157 440
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
IMPORTANT SAFETY INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Guide to the lab washer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Description of the machine. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Opening and closing the door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Water softener . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the water softener . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Filling the salt reservoir . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Salt indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Rinse Aid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Rinse aid indicator light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the rinse aid dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Neutralizing agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adding neutralizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Neutralizer indicator light. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Detergent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Setting the liquid detergent dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Loading the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Spring adapter for the water connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Adjusting the upper basket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Turning on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Starting a program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Selecting the (optional) drying function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Program sequence display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
End of a program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Additional drying. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Turning off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Canceling a program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Program Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Programming functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
The following instructions apply to all programming functions:. . . . . . . . . . . . . . . 36
To access the programming mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Saving to memory and exiting the programming mode . . . . . . . . . . . . . . . . . . 36
Exiting the programming mode without saving . . . . . . . . . . . . . . . . . . . . . . . . 36
DOS 1 (liquid detergent) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Setting the detergent dosage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
DOS 3 (neutralizer) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Setting the neutralizer dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Changing program parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Changing the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Changing the holding time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Restoring the factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Setting the time and date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Resetting the coarse and fine filter timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Data transfer (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Routine checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Cleaning and Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Frequently asked questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Thermo switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
After Sales Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
INSTALLATION INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Electrical connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Caring for the environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY INSTRUCTIONS
Please pay attention to the following
This machine conforms to current
notes to maintain safe procedures.
safety requirements. However,
inappropriate use can lead to
personal injury and property
damage. Read the operating
instructions carefully before using
this machine.
The machine should be installed,
~
commissioned and maintained only by
a Miele Service Technician. Repairs by
unqualified persons could be
dangerous.
Do not install the machine in an area
~
where a danger of explosion or freezing
may be present.
The manufacturer cannot be held
responsible for damage caused by
non-observance of these Important
Safety Instructions.
Be certain this appliance is properly
~
installed and grounded by an
authorized technician. To guarantee the
electrical safety of this appliance,
continuity must exist between the
appliance and an effective grounding
system. It is imperative that this basic
safety requirement be met. If there is
any doubt, have the electrical system
checked by a qualified technician. The
manufacturer cannot be held
WARNING - To reduce the risk of fire,
electric shock or injury:
Use
This machine is not intended for any
~
purpose that is not indicated in the
operating instructions. Conversions,
modifications, and other unintended
uses are considered to be dangerous.
The cleaning processes described here
apply only to those medical products
that are specifically designed by the
manufacturer to be re-used. Follow the
product manufacturer's instructions.
Miele will not be held liable for damage
caused by incorrect use of any kind.
responsible for damage or injury
caused by the lack, or inadequacy, of
an effective grounding system
(e.g. electric shock).
A damaged machine is dangerous.
~
Turn off the machine immediately at the
main switch and call the Miele Service
Department.
The installation of this unit in non
~
Personnel operating the machine
~
stationary locations (e.g. ships) must be
performed by a qualified installer or
service agency in strict accordance
with national and local safety
should be trained regularly. Children
and untrained personnel must not be
allowed access to the machine or its
controls.
regulations and standards.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY INSTRUCTIONS
Take care when handling
If you are exposed to toxic vapors or
~
~
processing chemicals (detergents, processing chemicals, consult the
softeners, neutralizers, etc.). These may manufacturer's material safety data
contain irritant or corrosive ingredients.
Do not use organic solvents as these
could cause an explosion. Wear
sheets for emergency procedures.
Always allow mobile units, modules,
~
inserts, and loads to cool before
removing. Be sure to empty any water
from concave items into the wash
cabinet before removing.
protective gloves and goggles. With all
chemical agents the manufacturer’s
safety conditions must be observed.
Do not stand, sit or lean on the
~
After drying open the door to allow
~
opened machine door. This could
cause the machine to tip over, resulting
in injuries and damage.
the items and inserts to cool.
Never clean the machine or near its
~
vicinity with a water or high pressure
hose.
Do not touch the heating elements
~
during or directly after the end of a
program, burns could result. The
Avoid inhaling powder products.
~
elements remain hot for some time after They can cause burning in the mouth
the end of the program.
and throat if swallowed, and can inhibit
breathing.
The water in the machine must not
~
be used as drinking water.
Before servicing, disconnect the
~
power supply by either removing the
fuse, unplugging the unit or manually
"tripping" the circuit breaker.
Be careful when sorting items with
~
sharp, pointed ends. Carefully position
these items in the machine, so that you
will not hurt yourself or create a danger
for other operators.
Whenever work is conducted on the
~
tap water network, the water supply line
to the washer must be cleared of air.
Failure to do so can result in damage to
the water softener.
When using this machine regard the
~
high temperatures and be especially
careful not to scald or burn yourself.
When opening the door bypassing the
electrical lock, a danger of burning,
scalding or corrosion exists or when
using disinfection agents, the inhalation
of toxic fumes. Let baskets and inserts
cool. Empty any water from parts into
the wash cabinet.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY INSTRUCTIONS
The following points should be
The machine must only be operated
~
observed to assist in maintaining
quality standards for critical labware
and to avoid damage to the loads
being cleaned.
with water and the recommended
cleaning agents. Organic solvents must
not be used in the machine as there is
the danger, under certain
circumstances, of explosion and
damage to rubber and synthetic
materials. Follow the dosage
For this type of cleaning, the program
~
must not be interrupted until completion. In
other words, the On/Off Button must not be
used. Interruptions should also be avoided
for all other programs, such as during
performance testing, because this can
negatively affect the cleaning and rinsing
results. After any interruption, the program
should be restarted from the beginning.
recommendations of the cleaning
agent’s manufacturer.
In critical applications where very
~
stringent requirements have to be met,
it is strongly recommended that all the
relevant factors for the process, such
as detergent, quality of water, etc. are
discussed with a Miele Application
Technology specialist.
Only use cleaning agents
~
formulated for special processes and
approved by Miele for use with this
machine. Use of unsuitable cleaning
agents could adversely affect the
components of the machine.
It is the responsibility of the operator
~
to routinely check that the required
cleaning standards are being met.
Therefore, results need to be regularly
tested and documented, both thermo
electrically and through inspection. For
thermo chemical processes, additional
testing is required using chemical or
biological indicators.
Use only Miele approved cleaning
~
agents with this machine. Use of unsuitable
cleaning agents could adversely affect the
components of the machine. Damages
resulting from using unsuitable cleaning
agents are not covered by the warranty.
The performance of the cycles was
~
Pre-treatments with cleaning agents
~
tested using Neodisher products. The
use of other products is not
can create suds, as can certain rinsing
agents. For pre-treatment and / or
cleaning, only use low-sudsing
discouraged but may not yield identical
results. Please be aware that changes
in formulation, storage conditions, etc.
which may not be publicized by the
chemical manufacturer can have a
negative effect on cleaning results. and
are not the responsibility of the machine
manufacturer.
detergents which have been approved
by Miele. Suds can have an adverse
effect on the cleaning process.
If toxic chemicals may be present in
~
the wash cabinet, beware of this risk
when interrupting a program and
opening the washer. The door seal and
steam condensor must always be
checked for soundness.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY INSTRUCTIONS
When using cleaning agents and
Accessories
~
specialized products, it is essential that
the manufacturer’s instructions are
followed. Only use the product for the
application described by the
Only genuine, Miele parts and
~
accessories should be used with this
machine, including Miele mobile units,
baskets and inserts.
manufacturer to avoid any material
damage or the occurrence of strong
chemical reactions.
Adequate cleaning results cannot
~
be guaranteed when non-Miele
accessories are used, or when Miele
accessories are altered. Damages
resulting from the use of unsuitable
accessories are not covered by the
warranty.
The mobile units, baskets and
~
special inserts should only be used for
their specific applications. Instruments
with hollow parts must be flushed
through with wash water.
Empty all containers and utensils
Disposal of an old appliance
~
before arranging them in the machine.
Obsolete appliances should be
~
Do not allow any remains of acids,
made unusable. To do this, first
~
solvents or corroding ferrous material,
and in particular hydrochloric acid or
chloride solutions to get into the wash
disconnect the appliance from the
power supply, then cut off the power
cord. It is important to note that the old
cabinet. The presence in compounds of appliance may be contaminated from
any solvents should be minimal,
exposure to pathogens, genetically
modified material, blood, etc., and
therefore must be decontaminated
before disposal.
(especially those in hazard class A1).
To avoid any corrosion damage
~
ensure that solutions or steam
containing hydrochloric acid do not
come in contact with the stainless steel
casing of the machine.
For safety and environmental
protection, be sure to remove all
chemical residue, in compliance with
safety regulations (wear protective
eyewear and gloves!).
Please follow the installation advice
~
in these instructions and the separate
Installation Instructions.
Remove or destroy the door latch, to
prevent children from locking
themselves inside. The appliance can
Please note the following symbols on now be sent for disposal through the
the machine:
proper resource recovery facilities.
Caution : See Operating
SAVE THESE
Instructions!
INSTRUCTIONS
Caution : Danger of electric
shock!
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Guide to the lab washer
aDisplay
i"Rinse aid" indicator
j"Neutralizer agent" indicator
k"Recharge Softener" indicator
l"Fill/Drain" indicator
b"Display" indicator
c"Drying" indicator
d"Start" indicator
e"Start" 6 button
f"Drying" 0 button
mProgram sequence display
nProgram selector
g"Display" ß toggle button (toggles
between the elapsed time, active
wash blocks and temperature
displays)
o"On/Off" I - 0 button
p"Door" a button
h"8 Dosing pump" indicator
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Guide to the lab washer
aDrying unit
eLevel indicator for rinse aid
bDispensers for neutralizer (red) and
liquid detergent (blue)
fFilter combination
gSalt reservoir
(water softener)
cDispenser for powdered detergent
dRinse aid reservoir
(with dosage selector)
hService panel
iSerial interface
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Description of the machine
Miele laboratory washers are designed
to clean, rinse, and dry laboratory
equipment.
Intended applications: Laboratory
equipment
Lab equipment includes a wide range
of items:
In this operating manual, the general
term "labware" is used to cover a range
of laboratory equipment.
– Glassware, such as beakers,
round-bottomed flasks, Erlenmeyer
flasks, bottles, test tubes, funnels,
Always follow the labware
manufacturer's instructions. Individual
items need to be checked to ensure
that they are designed to withstand the
temperatures and acid / alkaline
environment in the machine.
– measuring vessels, such as
graduated flasks, graduated
cylinders, pipettes
– dishes, such as petri dishes, watch
glasses, mortars, and
Standard model
– hardware, such as stoppers,
spatulas and stirrers.
The labwasher comes as standard with
a:
Labware can be made of glass,
heat-resistant plastic, ceramic,
stainless steel, or coated metal
(e.g., stirrers).
– drying unit
– serial interface
– water softener
– steam condenser
– flow meter
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Opening and closing the door
Electric door lock
Emergency release
The machine is equipped with an
electric door lock.
The emergency release feature
should only be used when it is not
possible to open the door normally,
for example after a power outage.
The door can only be opened when
^ the electrical supply to the machine
is turned on, and the
^ Turn the program selector to f.
^ On-Off switch is pressed.
^ Press the On-Off button to switch off
the machine.
To open the door
^ Press the a button in as far as it will
go. At the same time, grip the handle
and open the door.
,Do not touch the heating
elements when you open the door at
the end of a program. They remain
hot for some time and can cause
burns.
Once a program has started, the door
stays locked until completion and
cannot be opened. However, the two
programs "Rinse" and "Drain" are
exempt from the electric door lock.
The door can be programmed to open
during the "Drying" cycle, if desired.
^ Locate the ring for the emergency
release cable (at the bottom of the
machine, behind the service panel),
and pull the ring downward to open
the door.
To close the door
^ Lift the door upwards and push it
until it clicks shut. Do not press the
release catch while shutting the door.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Water softener
Before using the machine for the first
time, the water softener should be
programmed. The water needs to be
softened to avoid calcium deposits on
the items being cleaned and in the
machine itself.
Setting the water softener
^ Use the I - 0 button to switch off the
machine.
^ Turn the program selector to f.
^ Simultaneously press and hold the
ß and 6 buttons, while turning the
machine on with the On-Off button.
If the water supply is harder or softer
than the factory setting (including
below 4 gr/gal or 70 ppm), the setting
will need to be changed. See "Setting
the water softener".
The current program status P....
will appear in the display, and the
"Fill/Drain" indicator will come on.
– In order to function properly, the
water softener requires a supply of
reactivation salt.
^ Press the 0 button one time.
E01 will appear in the display.
– The cleaner must be correctly
programmed to match the hardness
of the tap water supply.
^ Turn the program selector one switch
position clockwise (to 1 o'clock).
The number 19 will appear in the
display (this indicates the factory
setting in gr/gal).
– Your local water authority can
provide information about the
hardness of the tap water in your
area.
^ Press and hold the ß button until
the required value appears in the
display.
If the water hardness is known to
fluctuate, always program for the
highest value.
Please refer to the "Settings" table for
detailed water hardness tables (water
hardness in gr/gal and ppm).
The built-in water softener can be
programmed at settings from
1 - 60 gr/gal or 20 - 1080 ppm.
Once the 60 setting has been reached,
the counter begins again from 0.
It is useful to make a note of your water
hardness so that you can provide the
service technician with this information
in the event of any service calls.
^ Press the 6 button.
SP will appear in the display.
^ Press the 6 button again.
Please note the water hardness in the
space provided here:
The setting you selected will now be
stored in memory. The display light(s)
will turn off.
gr/gal or ppm
The machine is now ready for use.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Water softener
Settings
Settings
1 - 30
gr/gal
1 - 30
ppm
Settings
31 -60
gr/gal
31 -60
ppm
CaCO
CaCO
3
3
1
1
2
20
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
560
580
590
610
630
650
670
680
700
720
740
760
770
790
810
830
850
860
880
900
920
940
950
970
990
1010
1030
1040
1060
1080
2
40
3
3
50
4
4
70
5
5
90
6
6
110
130
140
160
180
200
220
230
250
270
290
310
320
340
360
380
400
410
430
450
470
490
500
520
540
7
7
8
8
9
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
10
11
12
13
14
15
16
17
18
19 *)
20
21
22
23
24
25
26
27
28
29
30
*) factory setting
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Water softener
Filling the salt reservoir
Only special, coarse-grained
reactivation salt* should be used in
this machine.
*Available from Miele. Please see
back page for more information.
Do not use other types of salt, e.g. table
salt, agricultural or gritting salt. These
could contain components which are
insoluble in water and could damage
the water softener. If in doubt, consult
Miele Technical Service.
^ Place the funnel provided in place.
The salt reservoir holds approximately
2.5 kg salt.
^ Carefully fill the reservoir with the
reactivation salt.
Some displaced water will run out - this
is normal.
,Inadvertently filling the salt
reservoir with detergent will damage
the water softener.
^ Wipe any residual salt off the screw
threads of the reservoir.
Before filling the reservoir, make
sure that you are using reactivation
salt.
^ Screw the cap on securely.
^ Immediately after this step, start the
"Rinse" program, to remove any
traces of salt from the inside of the
machine.
^ Remove the bottom basket from the
machine.
^ Unscrew the salt reservoir cap.
Note: There may be a delay of a few
minutes before the "Rinse" program
starts due to the reactivation of the
water softener.
Before filling the salt reservoir for the
first time fill the reservoir with
approximately 2.5 liters of water, so
that the salt can dissolve properly.
Once the washer has been
operated, there will always be
enough water in the reservoir.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Water softener
Salt indicator
When the salt reservoir is empty, the
"Recharge Softener" indicator b lights
to remind you to fill the reservoir.
Reactivation takes place automatically
as needed during a program. The
"Recharging" indicator a lights while
this is occurring.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Rinse Aid
Adding rinse aid
^ Open the washer door.
Rinse aid is not used in laboratory
glassware cleaning applications.
If rinse aid is required, the rinse aid
dosage must be activated with
assistance from Miele Technical
Service.
Rinse aid (e.g., MieleClear) is filled in
the storage container, and the selected
dose is then dispensed automatically.
The rinse aid reservoir holds
approximately 180 ml.
,Inadvertently filling the rinse aid
reservoir with detergent will damage
the machine. Be sure only to use
rinse aid in the rinse aid reservoir.
^ Unscrew the reservoir cap.
^ Pour enough rinse aid into the
reservoir so that the level indicator
(see arrow) is dark. Use a funnel if
necessary.
^ Replace and tighten the reservoir
cap.
^ Wipe up any spilled rinse aid, to
prevent over sudsing during the next
wash cycle.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Rinse Aid
Rinse aid indicator light
Setting the rinse aid dosage
The dosage selector has settings from
1 to 6 (1 - 6 ml). The factory setting is at
3 (3 ml).
^ Rinse aid does not need to be refilled
until the "Rinse aid" indicator light
comes on in the display.
If the glassware shows spots after
washing,
^ adjust to a higher setting.
If the glassware shows smearing or
cloudy streaks after washing,
^ adjust to a lower setting.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Neutralizing agent
Adding neutralizer
^ Open the door of the storage cabinet
that houses the neutralizer and
detergent dispensers.
Neutralizer has an acidic pH, and
works by neutralizing any traces of
(alkaline) cleaner from the surface of
the glassware.
When a mild alkaline detergent is
used for washing, select a
neutralizer that contains phosphoric
acid.
The neutralizer is automatically
dispensed during the second rinse
after the main wash cycle. To function
properly, the neutralizer container must
be filled and cleared of air.
For specific instructions, please read
the section on "Programming functions".
^ Take out the (red) neutralizer
container and refill it or replace it with
a filled container.
Note: the program "Inorganica"
dispenses neutralizer after the
pre-wash.
^ Insert the siphon tube into the
container opening and screw it into
place. Make sure to observe the
color coding.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Neutralizing agent
Neutralizer indicator light
^ When the "Neutralizer" indicator light
comes on, it is important to promptly
refill or exchange the neutralizer
container.
Make sure to refill / exchange the
container before it becomes
completely empty.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Detergent
Adding detergent
^ Open the door of the storage cabinet
that houses the neutralizer and
detergent dispensers.
,Use only cleaning agents
formulated for this machine. Do not
use detergents formulated for
household dishwashers.
^ Take out the (blue) liquid detergent
container and refill it or replace it with
a filled container.
The factory setting is designed for
liquid detergents. Although powered
detergents can be used, Miele
recommends that you use liquid
detergents instead.
Note also that the program
"Organica" can only be used with
liquid detergents.
Setting the liquid detergent
dosage
The liquid detergent is automatically
dispensed during Main Wash 1 and/or
Main Wash 2 in the program sequence.
The detergent dispenser must be
properly filled and cleared of air.
^ Insert the siphon tube into the
container opening and screw it into
place. Make sure to observe the
color coding.
For specific instructions, please read
the section on "Programming functions".
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Detergent
Liquid detergent refill light
Before starting the program, load the
powdered detergent into the
compartment (except for the programs
"Rinse" and "Drain").
^ When the "8 Dosing Pump"
indicator comes on, it is important to
promptly refill or exchange the
detergent container.
^ Press the button on the dispenser
cover. The cover will pop open.
Make sure to refill / exchange the
container before it becomes
completely empty.
The flap is always open at the end of a
program.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Detergent
^ Load the detergent into the
compartment.
Dosage: To achieve a detergent
concentration of approximately 3 g/l,
use 30 g of detergent.
Always follow the detergent
manufacturer's recommendations for
dosage concentration.
^ Snap the cover closed.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
Depending on the type of labware you
wish to clean, the washer can be used
with an upper basket, lower basket, or
injector wagon, as well as an
After every program completion,
conduct the following visual
inspection:
– Inspect the labware for cleanliness.
assortment of inserts and special
baskets to suit your needs.
– Has any hollowware been dislodged
from its proper position on the
nozzles?
Baskets and inserts should be
properly selected to match the
application desired.
,Any items that were dislodged
from the adapters during a program
cycle must be washed again.
Before every program start, conduct
the following visual inspection:
– Are the cavities and channels in any
hollow vessels cleaned through?
– Is the labware properly loaded and
secured in the machine?
– Are the nozzles and connections
firmly attached to the
baskets/inserts?
– Are the hollow vessels properly
loaded so that all cavities and
channels will be flushed clean?
– Are the spray arms clean, and can
they rotate freely?
– Is the filter combination free of
coarse debris (remove any coarse
material and clean the coarse, fine,
flat, and micro-fine filters as
needed)?
– Is the basket adapter properly
connected to the water supply for the
spray arms and nozzles?
– Are the chemical dispensers filled?
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
Loading the machine
Prepare items before loading
– Always load items in a way that all
surfaces are exposed. This ensures
that the items will be properly
cleaned.
^ Empty all glassware before loading
into the machine. Comply with all
applicable infection-control
regulations.
– Do not nest any items inside other
items. Keep each item separate.
^ Remove all agar residue from petri
dishes.
– Hollow vessels must be arranged for ^ Remove any blood clots and blood
complete inside exposure to the
residue.
cleaning water.
^ Remove all stoppers, corks, labels,
sealing wax residue, etc.
– Hollow vessels should be inverted
and placed in suitable baskets and
inserts so that water will have
,Ensure that no acid or solvent
residues, especially hydrochloric
acid or chlorides, get inside the
machine.
unrestricted access and exit.
– Deep-based items should be
arranged at an angle, so that water
can run off easily.
Special preparations before
operating the machine
– Tall, narrow pieces should be placed
in the center of the basket for best
water coverage.
Items effected by heavy contamination
that resists removal, such as vacuum
grease, paper labels, etc., can affect
the cleaning and rinsing of the whole
load. Pre-treat these items before
loading.
– Baskets and inserts with an adapter
must be properly connected.
– The spray arms must not be blocked
by items that are too tall or hang
through the baskets.
Lab equipment that is contaminated by
microbiological material, pathogens,
propathogens, genetically altered
material, etc., may need to be sterilized
before being washed in the machine.
The user is responsible for this
decision.
– To avoid corrosion, only load
stainless steel instruments in good
condition.
– Do not use this machine to wash
nickel-plated items or discolored
(oxidized) aluminum items.
– Do not load plastics unless they are
heat-resistant.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
Spring adapter for the water
connection
Adjusting the upper basket
The upper basket can be adjusted
above and below the middle position,
by 2 cm.
The spring adapter for the water
connection must be properly engaged
when a basket or the injector wagon is
inserted into the machine. It must be
4 - 5 mm higher than the water inlet in
the machine.
Depending on the height selected and
which inserts are used, labware of
various heights can be arranged in the
baskets.
If this is not the case, the adapter
needs to be adjusted as follows:
To adjust the upper basket:
^ Pull the upper basket out as far as it
will go, lift it up off the runners and
remove.
^ Use a wrench to unscrew the roller
bearings on either side of the basket,
then reposition as required.
^ Loosen the locking ring a.
^ Move the adapter up to the proper
height b.
^ Tighten the locking ring c.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
Laboratory equipment
This section only covers basic
information on how to prepare and load
labware.
Wide-necked labware, e.g., beakers,
wide-necked Erlenmeyer flasks and
Petri dishes, or cylindrical items,
– Petri dishes and similar items should
be loaded in a suitable insert with the
soiled side facing center.
e.g., test tubes, can be cleaned inside
and out by the rotating spray arms.
For proper cleaning, load these items in
whole, half or quarter inserts, and place
them in an empty lower basket or upper
basket with spray arm.
– Pipettes should be placed with the
narrow end facing down.
– Protective netting can be used to
avoid glass breakage.
Narrow-necked items, e.g., narrow
necked Erlenmeyer flasks, round flasks, – Quarter baskets should be loaded at
volumetric flasks and pipettes, require
the use of injector wagons or injector
baskets.
least 3 cm away from the edge of the
wagon.
Separate instructions are included
for the use of injector baskets and
inserts for narrow-necked labware.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
Upper basket O 188
E 106
For various inserts.
Stainless steel half insert with 28 spring
hooks in 2 different heights for various
glassware, e.g. narrow necked flasks,
graduated cylinders, medicine bottles,
etc.
Lower basket U 874
For various inserts.
E 109
Stainless steel half insert to
accommodate 21 beakers up to 250 ml,
Erlenmeyer flasks, round flasks, etc.
28
Download from Www.Somanuals.com. All Manuals Search And Download.
Loading the machine
E 350
O 187
Injector wagon
Injector wagon
For narrow necked glassware,
15 injectors, 160 mm high;
18 injectors, 220 mm high.
(Upper basket) For direct injection of
narrow necked glassware,
34 injectors,160 mm high.
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Use
Turning on
Selecting the (optional) drying
function
^ Close the door.
^ Press the I - 0 button.
The additional function "Drying" can be
selected immediately after a program
has been selected (except for "Rinse"
and "Drain").
If the program selector is in the f
position, an indicator light will appear in
the display.
^ Press the "Drying" 0 button.
Starting a program
The options Drying 1 (if available) and
Drying 2 will now be added - see the
Program Guide for details.
^ Turn the program selector clockwise
or counterclockwise to the desired
program. Refer to the "Program
Guide" for more details.
The selected Drying 2 time appears in
the display. The drying time can be
changed in 5-minute increments, using
the 0 button.
Once the program has started the
changed drying time for the selected
program is stored. The total duration of
the program adjusts accordingly.
The display shows the "Main Wash 1"
temperature of the selected program
(except for "Rinse" and "Drain"). The
"Start" indicator flashes.
^ Press the "Start" 6 button.
The display shows the current
temperature.
The "Start" indicator comes on.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
Use
Once a program cycle has started, all
other programs are locked until
Program sequence display
Whenever a program is running, the
indicator light for that program will
remain on.
completion. If the program selector is
turned to a different program, the
display goes out. The display will
re-appear if the program selector is
turned back to the running program.
Recharging
Pre-Wash 1 + 2
Wash 1 + 2 *
During a program, the "Display" ß
button can be used to toggle between
the:
Rinse 1 - 4 / Final rinse 1 + 2
Dry
– actual temperature
Complete
– elapsed program time, and
(Program parameter achieved)
– active wash block
(depending on model):
1 = Pre-rinse 1
2 = Pre-rinse 2
3 = Main wash 1
4 = Main wash 2
5 = Chemical disinfection
6 = Interim rinse 1
7 = Interim rinse 2
8 = Interim rinse 3
9 = Interim rinse 4
10 = Final rinse 1
11 = Final rinse 2
12 = Drying 1
13 = Drying 2
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Use
End of a program
Turning off
Once the "Start" indicator goes out, and ^ Press and release the I - 0 button.
the "Complete" indicator comes on, the
Because of the electric door lock, the
program has ended. You will see a 0 in
door can only be opened if the I - 0
the middle of the display, and all other
button is pressed (see "Opening and
indicators will show a running light.
closing the door").
A buzzer sounds for a maximum of
30 seconds by default.
To change the buzzer function, see
"Programming functions".
^ Use the "Display" button to view the
total running time or temperature of
the program.
Additional drying
If additional drying is required after the
end of a program, you can use the
"Drain" program with Drying (optional
function). This program runs for
approximately 1 min in addition to the
drying time selected.
32
Download from Www.Somanuals.com. All Manuals Search And Download.
Use
Canceling a program
^ Turn the program selector to f.
Avoid canceling a running program
unless absolutely necessary, for
example if the glassware is moving
around excessively or the wrong
program has been selected.
The program is canceled after
2 seconds.
^ Open the door.
,Caution! Water and items in the
machine may be hot. There is a
danger of burning or scalding.
,In cases of officially recognized
epidemics, contaminated water
must NEVER be drained from the
machine.
^ Rearrange the items securely. Wear
protective gloves as needed, and
comply with all applicable health and
safety regulations for infectious
disease control.
^ Replenish powder detergent if
necessary.
^ Close the door.
^ Start the "Drain" program.
The water will be drained away.
^ Select and start the program.
33
Download from Www.Somanuals.com. All Manuals Search And Download.
Program Guide
Program
Description
Open program available for programming by end user.
Custom
For removing inorganic residue;
INORGANICA
ORGANICA
STANDARD
UNIVERSAL
INTENSIVE
PLASTIC
generally for analytical applications, for light to moderate soiling, with moderate
to high rinse requirements.
For removing organic residue, such as oils, fats, wax, agar, preparation
applications and for some analytical applications, with moderate to heavy soiling,
and moderate rinse requirements. Liquid detergent required, cold, warm, and a
DI water supply is recommended.
For a wide range of soiling types that do not denature or precipitate in alkaline
environments, for some preparation and analytical applications, for light soiling,
and low rinse requirements.
For removing organic residue (such as proteins, oils and some fats) and some
inorganic residues (pH 7, water-soluble metallic salts), for preparation
applications and for some analytical applications, moderate to heavy soiling and
moderate rinse requirements.
For removing organic residue (such as proteins, cell and tissue cultures, oils and
some fats) and some inorganic residues (pH 7, water-soluble metallic salts), for
preparation applications and analytical applications, moderate to heavy soiling,
and moderate to high rinse requirements.
For temperature-sensitive labware, such as plastic bottles (minimum required
temperature resistance 55°C, for preparation applications and for some
analytical applications, for light to moderate soiling, for moderate rinse
requirements.
To pump away water, for example when a program is canceled.
DRAIN
RINSE
To rinse heavily soiled labware, e.g., to pre-treat heavy soiling, remove residue or
to avoid drying and encrusting before a full program is run.
For cleaning and thermal disinfection at 93 °C with 5 min holding time (as per
EN ISO 15883-1).
varioTD
For cleaning and thermal disinfection at 93 °C with 10 min holding time (as per
¶ 18 of Germany's infectious disease control act, section A/B, fungicidal,
bactericidal, and virus-inactivating).
SPECIAL
93°C - 10'
CW = cold water, HW = hot water, DI = distilled water, °F (C°) = Temperature, min = holding time
* The DOS 2 rinse aid dosage must be activated by Miele Technical Service
34
Download from Www.Somanuals.com. All Manuals Search And Download.
Program Guide
Program sequence
x
Pre-Wash
Main Wash
Rinse
Rinse
Final Rinse *
Drying
1
2
1
2
1
2
3
4
1
2
1
2
CW
HW
HW HW
DOS 3
DI
DI 99°C
25 min
70°C
5 min
DOS 1
80°C
75°C
1 min
3 min
CW
HW
DOS 1
75°C
HW DI
DOS 3
DI
DI
DI
DI
DI 99°C
25 min
70°C
5 min
DOS 3
50°C
70°C
1 min
1 min
3 min
HW
HW
DOS 1
85°C
HW HW
DOS 3
DI 99°C
25 min
70°C
5 min
65°C
3 min
80°C
1 min
3 min
HW
DOS 1
75°C
HW
DOS 3
DI 99°C
70°C
5 min
75°C 25 min
1 min
3 min
CW
CW
CW
HW
DOS 1
80°C
HW HW
DI 99°C
25 min
75°C
70°C
5 min
DOS 3
3 min
1 min
HW
DOS 1
85°C
HW
DI DI
DI 99°C
25 min
70°C
5 min
DOS 3
80°C
3 min
1 min
CW
DOS 1
55°C
CW CW
DOS 3
DI DI
65°C
5 min
55°C
35 min
55°C
1 min
5 min
65°C
30 min
CW
CW
CW
HW HW
DOS 3
DI
99°C
DOS 1
55°C
5 min
35 min
93°C
5 min
CW
HW HW
DOS 3
DI
99°C
DOS 1
93°C
10 min
35 min
75°C
3 min
x = Additional function DOS 1 = Detergent dosage from 131°F (55°C)
DOS 3 = Neutralizer dosage after water inflow
35
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
For future servicing, be sure to document any changes made to the factory
settings in the fields provided.
General information:
– Program selector positions that have no assigned function are represented by a
bar ( - ) in the middle of the display.
– To call up the programming level, use the f position.
The status E.. will appear in the display.
The following instructions apply to all programming functions:
To access the programming mode
^ Turn the program selector to f.
^ Turn off the machine.
^ Simultaneously press and hold the ß The current program status P..
and 6 buttons, while turning on the
machine by pressing the On-Off button.
appears in the display. The Fill/Drain
indicator lights.
Saving to memory and exiting the programming mode
^ Press the 6 button.
SP appears in the display.
^ Press the 6 button again.
The change will now be stored in the
memory.
Exiting the programming mode without saving
^ Press the On-Off button to switch off The change is canceled.
the machine.
36
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
DOS 1 (liquid detergent)
Setting the detergent dosage
The dosage concentration only needs to be set for liquid detergents. Set the
concentration according to the detergent manufacturer's recommendations.
Do not change the setting if you are using powdered detergent in the door
dispenser.
Step
Display / Result
^ Press the 0 button until E02 appears in E02
the display.
^ Turn the program selector to
"Inorganica".
Using door dispenser
Using DOS module
10 = 10 sec dosage length
1.00 = 1.0 % dosage concentration
^ Press or hold down the ß button until 0.50 The dosage time / dosage
the required value appears in the
concentration is now set.
display, e.g., 0.5.
Dosage concentration setting:
Priming the dosage system
The dosage system does not need to be primed (cleared of air) unless:
– the dosage concentration is being set for the first time,
– the liquid detergent container was inadvertently completely emptied .
Step
Display / Result
^ Press the 0 button until E02 appears E02
in the display.
^ Turn the program selector to "Custom". d§1
^ Press the ß button.
The dosage system will be primed
automatically. Priming is complete
when 0 appears in the display.
^ Exit the programming mode.
Start "Rinse" program.
This will rinse away any agents that
entered the wash cabinet during
priming.
37
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
DOS 3 (neutralizer)
Setting the neutralizer dosage
Set dosage concentration according to the neutralizer manufacturer's
recommendations.
Step
Display / Result
^ Press the 0 button until E02 appears in E02
the display.
^ Turn the program selector to "Intensive". The programmed dosage
concentration appears in the
display, e.g., 0.10 = 0.10 %
^ Press and hold the ß button until the
required value appears in the display,
e.g., 0.2.
0.20 The dosage concentration is
now set.
Dosage concentration setting:
Priming the dosage system
The dosage system does not need to be primed (cleared of air) unless:
– the dosage concentration is being set for the first time,
– the neutralizer container was inadvertently completely emptied .
Step
Display / Result
^ Press the 0 button until E02 appears in E02
the display.
^ Turn the program selector to "Universal". d§3
^ Press the ß button.
The dosage system will be primed
automatically. Priming is complete
when 0 appears in the display.
^ Exit the programming mode.
Start "Rinse" program.
This will rinse away any agents that
entered the wash cabinet during
priming.
38
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Buzzer
The buzzer can be programmed for the following functions:
– a continuous tone at the end of a program,
– an error signal that beeps in 1-second intervals, and
– a sample-collection signal (allows a sample of the final rinse water to be
collected) that beeps in sets of 3 followed by a 2-sec pause. Contact Miele
Technical Service to program the sample-collection
signal.
The numbers shown in the display correspond to the following settings:
20 Buzzer off
21 End of program (factory setting)
22 Error
23 End of program + Error
24 Sample-collection signal
25 End of program + Sample-collection signal
26 Error + Sample-collection signal
27 All functions (End of program + Error + Sample-collection signal)
Step
^ Press the 0 button until E04 appears in the display. E04
^ Turn the program selector to "Inorganica".
21
Display / Result
^ Press and hold the ß button until the required value 26
appears in the display, e.g., 26.
Value selected:
The buzzer will sound for 30 sec during the selected setting.
To switch it off earlier:
– Turn the program selector to f.
– Press the ß button.
– Interrupt the power supply, e.g., open the door.
39
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Changing program parameters Changing the drying parameters
The factory settings for various
parameters are listed in the "Program
Guide".
The drying parameters within the
various wash programs can be
adjusted to meet specific requirements.
For example, the temperatures and
drying times within the program blocks
Drying 1 (if available) and 2 can be
changed.
Program parameters cannot be
changed unless they are activated
within a given program (see "Program
Guide".
It is important to document any
changes e.g., by recording them in the
corresponding column in the program
guide.
Available temperature range:
50 °C - 99 °C
Available drying time range for:
Drying 1: 1 - 99 min in 1 min
increments
Changing the wash and rinse
parameters
Drying 2: 5 - 95 min in 5 min
increments
The wash programs can be adjusted to
meet various technical requirements.
For example, the temperatures and
cycle times for the program blocks
Main Wash 1 and Main Wash 2, and
Rinse 1 and Rinse 2, can be changed.
The drying time in the Drying 2
program block can also be changed
by using the 0 button (see
"Selecting the drying function").
Available temperature range:
30 °C - 93 °C in all programs except
"Rinse", "Drain" and "Special".
Available time range:
1 - 15 min.
40
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Changing the temperature
Step
Display / Result
E09 for Main Wash 1
E12 for Main Wash 2
E22 for Rinse 1
^ Press and hold the 0 button until the
setting of the desired program block
appears in the display.
E25 for Rinse 2
E28 for Drying 1
E31 for Drying 2
^ Set the program selector to the program
you wish to change, e. g., "Custom".
°C ~75
^ Press and hold the ß button until the
desired value appears in the display,
e.g., 70 °C.
°C ~70
Changing the holding time
Step
Display / Result
E10 for Main Wash 1
E13 for Main Wash 2
E23 for Rinse 1
^ Press and hold the 0 button until the
setting of the desired program block
appears in the display.
E26 for Rinse 2
E29 for Drying 1
E32 for Drying 2
^ Set the program selector to the program
you wish to change, e. g., "Custom".
min _ 1
^ Press and hold the ß button until the
desired value appears in the display,
e.g., 10.
min _ 10
41
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Restoring the factory settings
Step
Display / Result
^ Press and hold the 0 button until E33
appears in the display.
E33
^ Turn the program selector to a program.
Default factory setting(s):
Re-programmed setting(s):
00
PP
^ When PP appears in the display, press the ß 00, the factory settings have
button.
been restored.
The water softener has been restored to the factory settings. Reprogramming is
required (see "Setting the water softener").
Program changes to the open programs "Custom", "Inorganica" and "Organica"
remain unchanged.
42
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Setting the time and date
The date and time are set in the serial interface (optional accessory). This
information appears in the print log.
The date and time can be changed, e.g., from daylight saving to standard time.
Use the program selector to activate the desired time and date:
1 o'clock position: Calendar day 01 - 31
2 o'clock position: Month
3 o'clock position: Year
4 o'clock position: Hours
5 o'clock position: Minutes
01 - 12
00 - 99 = 2000 - 2099
00 - 23
00 - 59
When level E34 is selected without an activated serial interface, a "-" appears in
the display for all program selector positions.
Step
^ Press and hold the 0 button until E34 appears in E34
the display.
Display / Result
Wait approximately 30 sec before the next step.
The control unit retrieves the
currently stored values.
^ Switch the program selector to the position for the 10
value you wish to change, e.g., to the 4 o'clock
position to set the hour.
^ Press and hold the ß button until the required
value appears in the display, e.g., 1 x.
11, the hour is changed.
Wait approximately 30 sec before turning off the
machine.
The updated values are
transmitted to the serial
interface.
Once the values are saved and you have exited the programming mode, the
updated values are transmitted to the interface. Therefore wait about 30
seconds before switching off the machine.
43
Download from Www.Somanuals.com. All Manuals Search And Download.
Programming functions
Resetting the coarse and fine filter timer
Whenever the coarse or fine filter in the drying unit are changed, the service timer
should be reset to zero.
Step
Display / Result
^ Press and hold the 0 button until E04 appears in E04
the display.
Switch the program selector to the position for the The control unit shows the
value you wish to change.
currently stored values, e.g.,
105
512
Coarse filter = 6 o'clock position
Fine filter = 7 o'clock position
^ Press and hold the ß button until the service
timer is erased.
000
44
Download from Www.Somanuals.com. All Manuals Search And Download.
Data transfer (optional)
The machine is equipped with an
interface socket so that data can be
transferred from Novotronic to a log
printer.
During program operation a protocol is
prepared. It can be printed using an
external printer.
The protocol contains the following
Please contact Miele Technical Service information:
to order printer cables.
– date and device number
The serial interface is RS 232
compatible.
– program start time and name
– rinse blocks used
Various printer types with
Epson-compatible fonts can be used
for printing.
– dosage system with dosage
temperature and target dosage
amount
Miele Technical Service maintains a list
of recommended printers.
– target values for temperature and
holding times
Pin assignment 9-pin sub-D connector:
5
GND (ground)
TXD (send)
1-4-6 (bridge)
7-8 (CTS)
– minimal temperature during the
holding time
3
– all errors (e.g., water supply defect)
– program end time
,Printer initialization must only be
performed by Miele. Consult
Technical Service using the contact
information on that back of this
manual.
– warnings, e.g., salt required.
45
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
Periodic maintenance for the machine
is recommended annually.
Routine checks
The routine checks must be done daily
by the user before using the machine.
For the routine checks a checklist is
supplied with the washer.
Check the following points:
– the filters in the base of the wash
cabinet must be empty,
– the spray arms in the machine and at
the baskets should be clear,
– the wash cabinet and the door
gasket should be clear,
– the dispensing systems, and
– the baskets and inserts.
46
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the coarse filter
,Wear protective gloves and
goggles when cleaning this
machine.
Cleaning the filters in the wash
cabinet
,The machine must not be used
without all filters in place. The filters
protect the circulation pump from
damage caused by foreign objects.
The filter combination in the base of
the wash cabinet should be
inspected regularly and cleaned if
necessary.
^ Press the two tabs together, remove
and clean the coarse filter.
^ Put the clean filter back in position
and press until it clicks into place.
Watch out for sharp objects which
could cause injury e.g. glass
splinters.
47
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the fine, flat and micro-fine
filters
^ Remove the coarse filter.
^ Remove the fine filter from between
the flat and micro-fine filters.
^ Then remove together with the flat
filter.
^ Clean the filters.
^ Replace the filter combination in
reverse order. The flat filter must lie
flat in the base of the wash cabinet.
^ To unscrew the micro-fine filter,
grasp the two tabs and turn twice
counterclockwise.
48
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the spray arms
^ Unscrew the upper spray arm.
The spray arms can get clogged.
Check daily and clean if necessary.
^ Loosen the knurled nut of the lower
spray arm and remove.
^ Use a sharp pointed object to push
particles into the spray arm jets and
rinse thoroughly under running water.
Removing the spray arms
^ Remove the baskets.
Spray arm at the upper basket or
mobile unit (if available):
^ Refit the spray arms in reverse order
after cleaning.
^ Loosen the knurled nut and remove
the spray arm.
After replacing the spray arms
check that they can rotate freely.
Metal knurled nuts have a left-hand
thread. Ceramic knurled nuts have a
right-hand thread.
49
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the drain pump and
non-return valve
The drain pump is beneath the
non-return valve (see arrow).
If water has not been pumped away at
the end of a program the drain pump or
the non-return valve may be blocked.
^ Turn off the machine.
^ Remove the filter combination out of
the wash cabinet.
^ Before refitting the non-return valve,
check that the drain pump is not
blocked. Spin the propeller several
times in both directions to check for
obstructions.
^ Carefully refit the non-return valve
and secure it with the locking clamp.
Refit and lock the filter in place.
^ Tilt the locking clamp to the side.
For safety the load should be
washed again.
^ Pull out the non-return valve and
rinse well under running water.
50
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the water inlet filters
Cleaning the filters
To protect the water inlet faucet, filters
are incorporated in the intake hose
attached to the water supply. If the
filters are soiled they must be cleaned
to ensure sufficient water intake to the
wash cabinet.
^ Disconnect the machine from
electricity (unplug it or "trip" the
circuit breaker).
^ Close the water supply and unscrew
the water inlet faucet.
^ Carefully remove the rubber seal.
,The plastic housing of the water
connection contains electrical
components and should not be
submerged or run through water.
^ Take out the filters with needle nose
pliers.
^ Rinse the large area filter 1 and fine
filter 2 under running water, and
replace if necessary.
^ Return filters and seal. Make sure
they are seated correctly.
^ Reconnect the hose to the water
faucet, checking that it is not
cross-threaded.
^ Open the water supply.
^ Check for leaks.
51
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Cleaning the control panel
,Never clean the machine or near
its vicinity with a water or high
pressure hose.
^ The control panel should only be
cleaned with a damp cloth or a
suitable cleaner for use on plastic
materials.
Cleaning the wash cabinet
,Do not use abrasive cleaners,
glass cleaners or all-purpose
cleaners! They will damage the
control panel.
The wash cabinet is mostly self
cleaning.
If deposits have built, contact Miele for
info.
Cleaning the exterior
Cleaning the door seal
^ Stainless steel surfaces can be
cleaned using a non-abrasive
stainless steel cleaner or
dishwashing detergent and warm
water.
Clean the door seal regularly with a
damp cloth to remove soiling.
To replace damaged or leaking door
seals please contact Miele's Technical
Service Dept.
^ To help prevent resoiling, a
conditioner for stainless steel can
also be used. Apply sparingly with
even pressure.
Dispensing systems
Check the consumption of the used
chemicals regularly to notice any
irregularities of the dispensing.
,Do not use cleaners containing
thinner or ammonia. They will
damage the surface.
52
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning and Care
Baskets and inserts
If available:
To ensure the function of baskets and
inserts, they must be checked daily. A
checklist is supplied with the machine.
– do the spray arms rotate freely?
– Are the spray jets clogged? See
"Cleaning and Care - Cleaning the
spray arms"?
Check the following points:
– Are the rollers in the proper condition
and are they securely fixed in the
mobile unit/insert?
– Is the basket connection set to the
correct height and tightened?
– Are all spray jets, spray sleeves and
hose adapters tightly connected to
the basket/insert?
– Are all spray jets, spray sleeves and
hose adapters, unclogged so that
the wash water can flow through?
– Are covers and fasteners tightly
screwed on the spray sleeves?
53
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
The following guide can be used to help address minor problems without a
service call.
,Repairs should only be performed by Miele Technical Service. Work
performed by unqualified persons can place the user at considerable risk of
harm.
To help avoid unnecessary service calls, please be sure to check first whether an
error message is due to operator error.
To do so:
^ Turn the program selector to f. The error message is erased.
^ Press the On-Off button to switch off the machine.
^ Turn the machine back on and re-start the program.
^ If the error occurs again and cannot be corrected, contact Miele Technical
Service for assistance.
^ You will need to provide the error code ( "F...") .
What if...
Possible fault
Solution
The washer will not
start.
The door is not properly
closed.
Close the door firmly.
The machine is not
plugged in.
Plug in the power cord.
The fuses are defective or Make sure the breaker has
the circuit breaker has not tripped.
tripped.
The washer is not turned Press the On-Off button
on. and select a program.
The program does not Error message: F 04 - F Contact Miele Technical
continue.
17, F 20 - F 26, F 28 -
F - -
Service.
Fault code F 31 - 33
Dosing system faulty.
Contact Miele Technical
Service.
54
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
Before the start of a
program, the "8
Dosing Pump"
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
indicator flashes, and
the program can not be
started.
– Press the On-Off button
to switch off the
machine.
Error message: Fdo:
Refill or replace the liquid
detergent container.
The liquid detergent
container is empty.
Then:
– Switch the washer on.
– Restart the program.
A few minutes after the
start of a program, the
"8 Dosing Pump"
indicator flashes, and
the program stops.
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
– Press the On-Off button
to switch off the
machine.
Error message: Fdo:
Refill or replace the liquid
detergent container.
There is a problem
dispensing the liquid
detergent.
Prime the dosage system
(see "Programming
functions").
Then:
– Switch the washer on.
– Restart the program.
55
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
Before the start of a
program, the
Before addressing the
error:
"Neutralizer" indicator
flashes, and the
program can not be
started.
– Turn the program
selector to f. The error
message is erased.
– Press the On-Off button
to switch off the
machine.
Error message: Fdo:
Refill or replace the
neutralizer container.
The neutralizer container is
empty.
Then:
– Switch the washer on.
– Restart the program.
A few minutes after the
start of a program, the
"Neutralizer" indicator
flashes, and the
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
program stops.
– Press the On-Off button
to switch off the
machine.
Error message: Fdo:
Refill or replace the
neutralizer container.
There is a problem
dispensing the neutralizer. Prime the dosage system
(see "Programming
functions").
Then:
– Switch the washer on.
– Restart the program.
56
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
The "Recharge"
indicator flashes.
The water softener has run Refill the salt reservoir.
out of salt.
A few minutes after the
start of a program, the
"Fill/Drain" indicator
flashes, and the
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
program stops.
– Press the On-Off button
to switch off the
machine.
Error message: F..E:
– Open the water valve
completely.
There is a problem with the
water intake.
– Clean the water inlet
filters.
– The flow pressure at the
water connection is too
low. Contact a qualified
plumber for advice.
Then:
– Switch the washer on.
– Restart the program.
57
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
The rinse program is
interrupted and the
"Fill/Drain" indicator
light is flashing.
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
– Press the On/Off button
to switch off the
machine.
Error message: F..A:
– Clean the filter
combination in the wash
cabinet.
There is a problem with the
water out take.
– Clean the drain pump.
– Clean the non-return
valve.
– Remove any kinks in the
drain hose.
Then:
– Switch the washer on.
– Drain the water.
– Restart the program.
The water in the wash Error message: F01 - F03, – Clean the filter
cabinet does not get F18, F19, F27 combination in the wash
hot; the program
cabinet.
The washer is equipped with
sequence lasts too
long.
a temperature sensor that
switches off the heater when
it begins to overheat.
– Rearrange the
glassware.
– Reset the Thermo
switch. See "Frequently
asked questions -
Thermo switch".
Overheating can occur
when, e.g., large items
cover the heating elements,
or when the filters in the
wash cabinet are blocked.
58
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
After a program, there is The dispenser was wet
Make sure the dispenser is
dry before loading it.
The powdered detergent Always arrange glassware
powdered detergent
residue left behind in
the door dispenser.
when it was loaded.
dispenser cover was
blocked by glassware.
so the dispenser cover can
open.
The powdered detergent Detergent residue is
dispenser cover will not blocking the latch.
close.
Wipe away any detergent
residue.
A banging noise can be A spray arm is hitting the Interrupt the program and
heard in the wash
cabinet.
glassware.
re-arrange any glassware
that it blocking the wash
arm, (see "interrupting a
program").
A rattling noise can be Glassware is moving
Interrupt the program and
re-arrange any glassware
that is loose, (see
"interrupting a
heard in the wash
cabinet.
around in the wash
cabinet.
program").
A banging noise is
This may be caused by the Consult a qualified
coming from the water installation or the diameter plumber for advice.
pipes.
of the pipes.
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
Items made of glass
show signs of
corrosion.
The items are not
washer-safe.
Only use washer-safe
glassware.
No neutralization took
place in the program
sequence.
Fill the neutralizer
container.
Excessively high
Select an appropriate
temperatures are causing program or lower the wash
glass corrosion.
temperature.
Too alkaline detergent is
causing glass corrosion.
Choose a milder detergent
or lower the DOS1
concentration.
Items made of stainless The chloride content of the Test your water
steel show signs of
corrosion.
water is too high.
composition. If needed,
condition the water
externally and use DI
water.
The screw cap on the
Properly thread the cap
reactivation salt reservoir is and screw it tightly.
not tightened.
No neutralization took
place in the program
sequence.
Fill the neutralizer
container.
Rust articles or a rust film – Check the piping.
appears in the wash
– Remove all rusty items
cabinet, due to:
from use.
– excessive iron content in
the water, or
– rusty items in the
washer.
The grade of stainless
steel is not washer-safe.
Only use high grade
stainless steel.
60
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
There are white
The water softener setting Test the tap water
deposits and/or spots
on the labware.
is too low.
hardness and program the
water softener according.
Too much powder
Be sure you are using the
correct amount of
detergent.
detergent is being used.
The neutralizer is not
dispensing properly.
Check the neutralizer
dispenser.
There is no reactivation
salt in the reservoir.
Carefully fill the reservoir
with the reactivation salt.
The screw cap on the
Properly thread the cap
reactivation salt reservoir is and screw it tightly.
not tightened.
The rinse water quality was – Use water with a lower
insufficient.
conductivity.
– If the washer is
connected to a
demineralizing
cartridge, check the
status and change as
needed.
61
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
What if...
Possible fault
Solution
The wash result is
unacceptable.
The baskets and inserts
Properly arrange the
were incorrectly loaded or glassware.
overcrowded.
Do not overload baskets
and inserts.
The program used was not Select an appropriate
suited to the type/amount program or modify the
of soiling.
program.
The detergent used was
not suited to the
Use a suitable detergent
for machine washing.
type/amount of soiling.
The labware was left soiled Do not leave soiled
for too long.
labware unwashed for
more than 6 hours.
A spray arm is blocked.
Arrange the items do not
block the spray arms.
Jets are blocked in the
spray arms or inserts.
Check the spray jets and
clean as needed.
The basket or insert is not Use the adapter to
properly connected to the properly connect the water
water supply. supply to the basket or
insert.
62
Download from Www.Somanuals.com. All Manuals Search And Download.
Frequently asked questions
Thermo switch
This machine has a resettable heater
limiter which will shut off the elements in
the event of over-heating. This could be
caused for example, if large articles
cover the heating elements or if the
filters in the wash cabinet are blocked.
If the following fault shows: F01 - F03,
F18, F19, F27 (Water in the wash
cabinet is not heated, the program
cycle takes too long). Please proceed
as follows.
^ Remove the cause of the problem.
^ Remove the service plinth.
^ Press the reset button on the
temperature sensor on the right side.
If this switch trips repeatedly,
contact the Miele Technical Service
Department.
63
Download from Www.Somanuals.com. All Manuals Search And Download.
After Sales Service
In the event of a fault which you cannot
correct yourself please contact the
Miele Technical Service Department
U
V
1-800-991-9380
1-800-565-6435
^ Please quote the model of your
machine. This information can be
found on the machine’s front panel.
64
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLATION INSTRUCTIONS
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation
Please refer to the installation
Leveling the legs
diagram supplied with the machine.
Any unevenness in the floor can be
compensated for and the height of the
machine raised or lowered by adjusting
the four screw feet.
,Furniture and fittings installed
near the machine must be of a
commercial standard, able to
withstand the effects of steam.
When the feet are screwed in, the
machine can be moved back and forth
on its rollers.
Installation options
– Free-standing
Positioning and securing the
machine
– Slot-in, next to other machines or
furniture or in a suitable space. The
space must be at least 35 7/16"
(90 cm) wide and 27 9/16" (70 cm)
deep
To ensure stability, the machine must
be aligned and screwed to the counter.
^ Open the door.
– Undercounter
The machine can be installed under a
continuous counter top or sink drain.
The niche must be at least:
Width 35 7/16" (90 cm)
Depth 29 7/16" (75 cm)
Height 32 5/16" (82 cm).
The labwasher can be built into a
recess of 32 5/16" (82 cm) high by
removing the lab washer top.
^ Secure the machine to the front edge
of the countertop using screws
through each of the 2 holes in the
front trim.
^ When not mounted undercounter,
floor mounting brackets must be used.
Do not use silicone sealant to seal
the gaps between the machine and
any neighboring units. This will hinder
ventilation to the circulation pump.
66
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation
Removing the lid (if necessary) Protecting the countertop
^ Open the door.
Important for machines with steam
condenser (depending on model):
^ Remove the fixing screws on both
sides using a cross slotted
screwdriver.
To prevent steam damage to the
countertop, adhere the protective foil
9 13/16" x 22 13/16" (25 x 58 cm)
underneath the countertop near the
steam condenser.
^ Pull the machine lid approx. 5 mm
forwards, lift upwards and remove.
Depending on the requirements for
installation, the following kit can be
ordered from the Miele Technical
Service Department.
Cover plate (protects the countertop)
The underside of the countertop is
protected from steam damage by a
stainless steel plate.
67
Download from Www.Somanuals.com. All Manuals Search And Download.
Electrical connection
WARNING
,All electrical work must be
carried out by a suitably qualified
electrician in accordance with local
and national safety regulations.
THIS APPLIANCE MUST BE
GROUNDED
Grounding connection
– Connection should be made via a
suitable isolator, with an on-off switch
that is easily accessible for servicing
work.
^ The ground lead must be connected
to the screw connection point
(marked with the ground symbol 8) at
the back of the machine.
– The electrical connection is made
through a receptacle according to
national requirements. The
The machine must only be operated
with the voltage, frequency and
fusing shown on the data plate
located on the rear of the machine,
and on the plinth (behind the service
panel).
receptacle has to be easily
accessible after installation.
– To increase the security it is
recommended to install a protective
switch (30 mA) before the machine.
The conversion diagram and the wiring
diagram is secured to the inner side of
the service panel.
– For technical data see the data plate
or wiring diagram supplied.
The machine must only be operated
with the voltage, frequency and fusing
shown on the data plate.
The machine can be converted
according to the supplied converting
diagram and wiring diagram.
The data plate is located at the rear of
the machine and behind the plinth on
the plastic cover.
The wiring diagram is supplied with the
machine.
See also the supplied installation
diagram.
A damaged power cord must only
be replaced with a genuine Miele
cord by a Miele service technician.
68
Download from Www.Somanuals.com. All Manuals Search And Download.
Plumbing
Connecting the water inlet
– An acceptable water pressure (flow
rate) is 10 - 147 psi. However, the
recommended pressure is
,The water in the machine is not
drinkable.
25 - 60 psi.
If the water pressure is below 30 psi,
the fill time will take longer.
– The machine must be connected to
the water supply in accordance with
local and national regulations.
If the "Fill/Drain" indicator fault code
"F..E" lights, contact the Miele
Technical Service Department.
– The washer is constructed so it may
be connected to a suitable supply
without an extra non-return valve
unless required by code.
– The maximum permissible static
water pressure is 145 psi over
pressure.
– High iron content can leave a rust
film on stainless steel labware and
the machine itself.
– If the water pressure lies outside
these ranges, please contact Miele
Technical Service for advice (the
"Fill/Drain" indicator, may appear in
the display "F E").
When the chloride content in the
domestic water supply is above
100mg/l, the risk of corrosion
damage to stainless steel labware is
greatly increased.
– The machine is designed for
connection to cold water (blue
marking) and hot water (red marking)
up to 158°F (70°C) max.
– In certain regions (e.g., mountainous
areas) the water composition may
cause condensate to form, requiring
the use of softened water in the
steam condenser.
Connect the water intake hoses to
the cold and hot water faucets.
– If there is no hot water supply, the
intake hose with the red marking
must also be connected to the cold
water supply.
69
Download from Www.Somanuals.com. All Manuals Search And Download.
Plumbing
– Connect the intake hose for the
steam condenser (without water
safeguard) to the cold water faucet.
– See the next page for how to
connect the DI water line.
– Water valves with 3/4 inch thread are
to be provided on site. They should
be easily accessible, because the
water supply will need to be turned
off when not in use.
½
– The inlet hoses are approx. 5 ft (1.7
m) long pressure hoses with 3/4 inch
thread. Do not remove the water
inlet filters in the threads.
Do not cut or damage the inlet hose
in any way.
– Install the large surface area filters
(in the kit supplied with the machine)
between the water valve and the
water inlet hose (see illustration in
"Cleaning the water inlet filters"). The
large surface area filter for distilled
water is made of stainless steel and
can be recognized by the matte
surface.
Please also refer to the installation
diagram.
70
Download from Www.Somanuals.com. All Manuals Search And Download.
Plumbing
DI Connection (pressurized)
DI connection (unpressurized)
(optional)
The machine comes factory standard
for connection to a pressure resistant
system. The recommended water
pressure (flow rate) is 25 - 60 psi
however, water pressure that is
10 - 147 psi is acceptable.
For connection to 1.23 - 8.7 psi the
machine must be modified, if this has
not done at the factory. If a feed pump
is to be installed, this must only be
done by Miele Technical Service.
^ The DI pressure tested hose (marked
"H2O pure" with a 3/4" hose thread
must be connected to the on-site
water valve for DI purified water.
The discharge port for any DI
container (unpressurized) must be
at least the height of the upper edge
of the machine (see "Installation
instructions").
If the DI water connection is not
used, the electronic needs to be
reprogrammed by Miele Technical
Service. The water intake hose
remains at the rear side of the
machine.
71
Download from Www.Somanuals.com. All Manuals Search And Download.
Plumbing
Drainage
^ The drainage system is fitted with a
non-return valve which prevents dirty
water from flowing back into the
machine via the drain hose.
^ The machine should preferably be
connected to a separate drainage
system onsite.
If separate drainage is not available
contact Miele Technical Service for
advice.
The onsite drain connection point
should be located between 1 ft.
(0.3 m) and 3 ft. (1 m) above the
lower edge of the machine.
If it is lower than 1 ft. (0.3 m), lay the
hose in a curve at a height of at least
1 ft. (0.3 m).
The drainage system must be able to
take a minimum drainage flow of
16 l/min.
^ The drain hose is approx. 5 ft.
(1.5 m), is flexible and has an internal
diameter of 7/8 in. (22 mm). It must
not be shortened. Hose clamps are
supplied for securing it in position.
^ A longer drain hose (up to 13 ft.
[4 m]) is available to order from
Miele.
^ The drainage system must not
exceed 13 ft. (4 meters).
^ See the supplied installation diagram.
72
Download from Www.Somanuals.com. All Manuals Search And Download.
Technical data
English
Metric
Height:
with lid
without lid
Width:
33.5"
32.3"
35.4"
27.5"
49.8"
85 cm
82 cm
90 cm
Depth:
70 cm
Depth with the door open:
126.5 cm
Net weight:
265 lbs
120 kg
Voltage / Fuse:
Power cord
see data plate
approx. 5.11 ft.
approx. 1.8 m
Water temperature:
Cold water
max. 68°F
20 °C
70 °C
Hot and DI water connection
max. 158° F
Static water pressure
max. 145 psi (10 bar)
Hot, Cold and DI Water Connection:
Acceptable water pressure
(flow rate)
10 - 147 psi
.7 - 10 bar
Recommended water pressure
(flow rate)
25 - 60 psi
1.72 - 4.13 bar
DI Water Connection (pressureless)
optional
1.23 - 8.70 psi
.09 - .6 bar
Delivery head:
min. 1 ft. (0.3 m), max. 3.3 ft. (1 m)
from 40°F (5°C) to 104°F (40°C)
Surrounding temperature:
Max. relative humidity
Linear reduction up to
80% for temperature up to 88°F (31°C)
50% for temperature up to 104°F (40°C)
Height above sea level:
up to 1640 yds
1500 m
Noise level in dB (A):
Sound pressure level LpA
during washing and drying
<70
Test marks:
CSA
!-mark:
MPG-Guidelines 93/42/EWG, Class IIa
73
Download from Www.Somanuals.com. All Manuals Search And Download.
Caring for the environment
Disposal of the packing
materials
Disposal of an old appliance
Old appliances may contain materials
that can be recycled. Please contact
your local authorities about recycling in
your area.
The cardboard box and packing
materials protect the appliance during
shipping. They have been designed to
be biodegradable and recyclable.
Ensure that the appliance presents no
danger to children while being stored
for disposal. See "Important Safety
Instructions".
Ensure that any plastic wrappings,
bags, etc. are disposed of safely and
kept out of the reach of children.
Danger of suffocation! Please recycle.
74
Download from Www.Somanuals.com. All Manuals Search And Download.
75
Download from Www.Somanuals.com. All Manuals Search And Download.
Subject to change without notice / 5207
M.-Nr. 07 157 440 / 00
Download from Www.Somanuals.com. All Manuals Search And Download.
|