SPECTRORADIOMETER   
					CS-2000/CS-2000A   
					Instruction Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
		
			
				 
			 
			
				
					Safety Precautions   
					To ensure correct use of this instrument, read the following points carefully and adhere to   
					them. After you have read this manual, keep it in a safe place where it can be referred to   
					anytime a question arises.   
					(Failure to adhere to the following points may result in   
					Warning   
					death or serious injury.)   
					Do not use this instrument in places where flammable or combustible gases   
					(gasoline etc.) are present.   
					Doing so may cause fire.   
					Always use the AC adapter and power cord supplied as a standard accessory   
					or optional (AC-A3ꢀ2), and connect it to indoor AC outlet of rated voltage and   
					frequency (ꢀ00 - ꢀ20 V or 200 - 240 V , 50/60Hz). Failure to follow either of   
					these may result in damage to unit, fire or electric shock.   
					If this instrument is not used for a long time, disconnect AC adapter from AC   
					outlet. Accumulated dirt or water on prongs of AC adapter plug may cause fire   
					and should be removed.   
					Do not forcibly pull any part on power cord when unplugging since this may   
					cause fire or electric shock. Gently disconnect by holding plug. Also, do not   
					handle power cord with wet hands. Doing so may cause electric shock.   
					Do not forcibly bend, twist or pull power cord. Also, do not place heavy object   
					on power cord, or damage or modify one. Any of these may cause fire or   
					electric shock due to damage to power cord.   
					Do not disassemble or modify this instrument or AC adapter. Doing so may   
					cause fire or electric shock.   
					Do not spill liquid on this instrument or drop metal into this instrument. Should   
					either of these happen, switch power off, unplug AC adapter immediately and   
					contact the nearest KONICA MINOLTA authorized service facility.   
					Should this instrument or AC adapter be damaged or smoke or odd smell be   
					generated, do not keep using such instrument or AC adapter without correction.   
					Doing so may cause fire. In such situations, switch power off immediately,   
					unplug AC adapter and contact the nearest KONICA MINOLTA authorized   
					service facility.   
					Do not look at sun or intense light through finder of this instrument. This may   
					lose your sight.   
					ꢀ 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					(Failure to adhere to following points may result in injury or   
					damage to this instrument or other property.)   
					Caution   
					Use this instrument near AC outlet for easy plugging or unplugging in using AC   
					adapter.   
					Do not place this instrument on unstable or sloping surface which may drop or   
					overturn it. Dropping or overturning may injure someone around. Take care not   
					to drop this instrument when carrying it.   
					Do not move while looking inside finder since this would fall or injure user.   
					Take special care in handling the ND filter or closeup lens included in the   
					optional accessories. Breakage of the ND filter or closeup lens may injure   
					someone around.   
					2 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Introduction   
					This instrument is a high-accuracy spectroradiometer designed to measure luminance and   
					chromaticity up to super-low luminance regions. Carefully read this manual before using one.   
					Packaging material   
					Be sure to save all packaging materials (corrugated cardboard boxes, pads and plastic   
					bags) supplied with the purchase. This is delicate measurement instrument. Use packaging   
					materials supplied in purchasing in case this instrument needs to be transferred for such   
					purpose as maintenance in KONICA MINOLTA's factories. These packaging materials are   
					useful for minimizing shock or vibration to this instrument in such situation.   
					Should any of these packaging materials be lost or broken, please contact the nearest   
					KONICA MINOLTA authorized service facility.   
					Note on Use   
					Operating Environment   
					 The standard AC adapter (AC-A3ꢀ2) of this instrument is designed specifically for use   
					indoors. Do not use it outdoors.   
					 
					Do not disassemble this instrument for being composed of delicate electronic components.   
					 Use this instrument at rated voltage of ꢀ00 V - ꢀ20 V or 200 V - 240 V (50/60Hz).   
					Connect AC power cord to AC outlet with rated voltage and frequency. Connected   
					voltage should not be outside the range of ꢀ0% of nominal.   
					 This instrument is classified into a Pollution Degree 2 as instrument used in mainly in   
					manufacturing plant, laboratory, warehouse or equivalents. Use this instrument in metal   
					dust free and non condensing potential environment.   
					 This instrument is categorized into Installation Category II as equipment connected to   
					commercially available power source.   
					 Connect PC for controlling this instrument to the outlet with protective grounding.   
					Failure to follow this may result in electric shock due to short circuit.   
					 Take care not to enter foreign substance like water or metal in this instrument.   
					Operating in such state cause serious danger.   
					 
					Do not use this instrument under direct sunlight or near heater. The internal temperature of this   
					instrument to becomes much higher than ambient temperature which may break this instrument.   
					Also, use this instrument in a well-ventilated place. To ensure proper heat dissipation,   
					keep the ventilation holes free from obstructions.   
					 Avoid rapid change in ambient temperature which may form dew condensation.   
					 Avoid using this instrument in extremely dusty or humid place.   
					 Use the CS-2000 at ambient temperature between 5 and 35ºC and relative humidity   
					80% or less (at 35ºC) with no condensation. Use the CS-2000A at ambient   
					temperature between 5 and 30ºC and relative humidity 80% or less (at 30ºC) with no   
					condensation. Operating this instrument outside specified temperature and humidity   
					range may not satisfy its original performance.   
					3 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					This Instrument   
					 Do not subject this instrument to strong impact or vibration.   
					 Do not forcibly pull, bend, or apply strong force to power cord for attached AC adapter   
					or USB cable. This may result in snapping.   
					 Connect this unit to power source with minimal noise.   
					 Do not measure a high-luminance light source (including sunlight) beyond the   
					measurement range. The failure to observe this warning could result in damage to the   
					optical system.   
					 Should breakage or abnormality be found during operation, switch power off   
					immediately and unplug. Then refer to “Error Check” on page 82.   
					 Should this instrument break down, do not try to disassemble and repair it by yourself.   
					Please contact the nearest KONICA MINOLTA authorized service facility.   
					 Warm this instrument up for 20 minutes at least after switching power on when the   
					object luminance is 2 cd/m2 or lower (measuring angle ꢀº).   
					Objective lens, ND filter and Closeup lens (Optional Accessories)   
					 Make sure that surfaces of objective lens, ND filter or closeup lens are clear. Correct   
					measurement may not be performed if there is dirt, dust, hand soil or part left unclean.   
					 Do not touch surface of objective lens, ND filter or closeup lens with hand.   
					 Do not change ambient temperature rapidly under high humidity. This may mist   
					objective lens, ND filter or closeup lens, resulting in incorrect measurement.   
					Note on Storage   
					Body   
					 Do not store this instrument under direct sunlight or near heater. The internal   
					temperature of this instrument becomes much higher than ambient temperature which   
					may break this instrument.   
					 Store this instrument at ambient temperature between 0 and 35ºC and relative humidity   
					80% or less (at 30ºC) with no condensation. Storage under high temperature and   
					humidity may deteriorate performance of this instrument. For added safety, we   
					recommend storage with such drying agent at room temperature.   
					 Take care not to form condensation. Avoid rapid change in ambient temperature when   
					transferring body for storage.   
					 Put the body in a packaging box supplied when purchased or the storage case (CS-   
					A30) in the optional accessories to store in safe place.   
					Objective lens   
					 For storage, cover the objective lens with standard accessory lens cap.   
					4 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Cleaning   
					Body   
					 If this unit becomes dirty, wipe with dry and soft cloth. Do not use organic solvent like   
					benzine or thinner and other chemical agent for cleaning. Should none of these   
					methods be helpful, please contact the nearest KONICA MINOLTA authorized service   
					facility.   
					Objective lens   
					 Should it be gotten dirt or dust, wipe off with dry and soft cloth or lens cleaning paper. Do not   
					use organic solvent like benzine or thinner and other chemical agent for cleaning. Should   
					none of these methods be helpful, please contact the nearest KONICA MINOLTA authorized   
					service facility.   
					Notes on Transfer   
					 Use packaging material supplied when purchased to minimize vibration or shock   
					generated during transfer.   
					 Put all material including unit and accessories in original packaging material when   
					returning this instrument for service.   
					Maintenance   
					 Periodical checkup is recommended annually to maintain measurement accuracy of   
					instrument. For details on checkup, please contact the nearest KONICA MINOLTA   
					authorized service facility.   
					Disposal Method   
					 Make sure that the CS-2000/CS-2000A, its accessories and the packing materials are   
					either disposed of or recycled correctly in accordance with local laws and regulations.   
					5 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Contents   
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					6 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					
					7 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Standard Accessories   
					Lens Cap CS-A31   
					 Attached to objective lens for protecting it   
					when not using this instrument.   
					USB Cable (2 m) CS-A32   
					 Used for communication between this   
					instrument and PC.   
					AC Adapter AC-A312   
					 Supplies power from AC outlet.   
					Input: ꢀ00 - ꢀ20 V or 200 - 240 V   
					0.75 - 0.42 A   
					50/60 Hz   
					Output: ꢀ2 V   
					3 A   
					Data Management Software   
					CS-S10w Professional   
					 Software to control this instrument from PC   
					for various data management.   
					 The protect key is attached.   
					Screw for focus ring lock CS-A38   
					 Locks the focus adjustment ring of the CS-   
					2000/CS-2000A so that it does not move   
					unintentionally and change the focus.   
					· Do not use any screw except for the   
					attached screw. Should it be lost or   
					damaged, purchase a new CS-A38.   
					· When storing the CS-2000/CS-2000A in the Storage Case (optional accessory),   
					remove the screw.   
					· When storing the CS-2000/CS-2000A in the packing box for transfer, remove the   
					screw.   
					8 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Optional Accessories   
					Calibration Certificate   
					ND Eyepiece Filter CS-A1   
					 Reduces glare during observation through   
					the finder when a high-luminance object is   
					measured. Be sure to place this filter in front   
					of the finder when measuring high-   
					luminance objects.   
					ND Filter (1/10) CS-A33   
					ND Filter (1/100) CS-A34   
					 Placed in front of objective lens for   
					measurement of high luminance object.   
					Calibration Certificate (for ND filter)   
					 Calibration certificates can be attached to   
					the ND filters (ꢀ/ꢀ0) CS-A33 and (ꢀ/ꢀ00) CS-   
					A34.   
					Closeup Lens CS-A35   
					 Placed in front of objective lens for   
					measurement of small object.   
					Adapter for CCD camera CS-A36   
					 Placed between the finder and the body   
					when a C-mount industrial camera is used.   
					9 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					White Calibration Plate CS-A5 (without data)   
					White Calibration Plate CS-A5 (with data)   
					White Calibration Plate CS-A5 (with data   
					and calibration certificate)   
					 Used for measurement of object colors.   
					Three types (named, not-named, named   
					with calibration certificate) are prepared.   
					Tripod CS-A3   
					Pan Head CS-A4   
					 Used when installing this instrument.   
					Storage Case CS-A30   
					 Used to house the instrument and   
					accessories or to carry them by hand. Never   
					use this as a transport case.   
					ꢀ0   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					System Configuration   
					Standard Accessories   
					Optional Accessories   
					Screw for   
					Lens Cap focus ring lock CS-2000 Main Unit Calibration Certificate   
					CS-A38   
					ND Eyepiece Filter   
					CS-A1   
					ND Filter   
					(1/10) CS-A33   
					(1/100) CS-A34   
					Calibration Certificate   
					(for ND filter)   
					Adapter   
					for CCD camera   
					CS-A36   
					AC Adapter   
					AC-A312   
					Closeup Lens   
					CS-A35   
					USB Cable (2 m)   
					CS-A32   
					White Calibration Plate   
					CS-A5 (w/o data)   
					CS-A5 (w/ data)   
					CS-A5 (w/ data and calibration certificate   
					) 
					PC   
					Pan Head   
					CS-A4   
					(on the market)   
					Tripod   
					CS-A3   
					Data Management   
					Software   
					Storage Case   
					CS-A30   
					CS-S10w Professional   
					ꢀꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Names and Functions of Parts   
					Names of Each Part   
					(7) Focus distance scale   
					Handle   
					(6) Focus   
					adjustment ring   
					(ꢀ) Power switch   
					(2) AC adapter   
					input terminal   
					(5) Objective lens   
					(ꢀ4) Screw holes for fixing   
					(8) LCD screen   
					(ꢀ0) Measurement   
					button   
					(3) USB connector   
					(9) Key panel   
					(ꢀ5) Vertically   
					synchronized   
					signal input   
					terminal   
					(ꢀ2) Diopter   
					adjustment ring   
					(4) Measuring   
					angle selector   
					Inside finder   
					(ꢀ3) Aperture   
					(ꢀꢀ) Finder   
					ꢀ2   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Functions of Each Part   
					(1) Power switch   
					Switches this instrument on/off. ( | ) for ON; (O) for OFF. …(p.23)   
					(2) AC adapter input terminal Connects the attached AC adapter. ………………………(p.2ꢀ)   
					(3) USB connector   
					(4) Measuring angle selector Selects measurement angle among ꢀº, 0.2º and 0.ꢀº. ………(p.52)   
					(5) Objective lens Directed to object for measurement. ………………………(p.53)   
					Connects the USB cable when connecting to PC. ………(p.72)   
					(6)Focus adjustment ring Adjusts focus of objective lens before measurement.……(p.53)   
					(7) Focus distance scale Helps adjusting focus. ………………………………………(p.53)   
					(8) LCD Screen   
					(9) Key Panel   
					Displays various screens like measurement and menu. … (p.ꢀ6)   
					Offers several keys for operation of this instrument. …… (p.ꢀ4)   
					(10)Measurement button For measurement. ……………………………………………(p.53)   
					(11) Finder Used to observe object for measurement. ………… (p.ꢀ5, 53)   
					(12) Diopter adjustment ring Adjusts diopter. ………………………………………… (p.ꢀ5, 53)   
					(13)Aperture   
					Indicates measuring area. …………………………………(p.53)   
					Size of black circle will change depending on measuring angle.   
					1º Aperture   
					0.2º Aperture   
					0.1º Aperture   
					(14) Screw holes for fixing Used to fix this instrument with tripod or jig. …………… (p.20)   
					(15) Vertically synchronized Connects the cable to input the vertically synchronized   
					signal input terminal   
					signal or the external sync measurement. ……………… (p.26)   
					ꢀ3   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Key Panel   
					(1)   
					(4)   
					(7)   
					(2)   
					(3)   
					(5)   
					(6 )   
					Main Functions of Each Key   
					(1) MENU key   
					The MENU screen appears if this key pressed when the   
					measurement value screen appears. (p.ꢀ7)   
					(2) DISPLAY key   
					Selects whether chromaticity is displayed in absolute value (ABS) or difference   
					(DIFF) if this key pressed when the measurement value screen appears. (p.4ꢀ)   
					(3) COLOR MODE key Color space modes are switched in turn as follows, by pressing   
					the key when the measurement value screen appears: Lvxy    
					Lvu’v’  LvTΔuv  XYZ  Dominant wavelength/Excitation purity   
					Spectral graph  Lvxy. (p.39)   
					(4) ESC key   
					If this key is pressed when the MENU screen is displayed, the   
					settings are canceled and the measurement value screen   
					appears again. If pressed during numerical input or when making   
					each setting, the settings are canceled. If pressed during   
					continuous measurement, the measurement ends.   
					(5) BACKLIGHT key   
					(6) MEMORY key   
					Selects backlight ON/OFF on LCD screen. (p.49)   
					Measured data is stored in memory by pressing this key when the   
					measurement value screen appears. (p.55)   
					(7)   
					keys   
					Memory data, target color channels, calibration channels, etc.,   
					are changed by pressing the key when the screen for display of   
					various data appears. The cursor position is moved up and   
					down, or the values and set items are changed, by pressing the   
					key during numerical input or when making each setting.   
					keys   
					The cursor position moves right and left by pressing the key for   
					numerical input or when making each setting.   
					ENTER key   
					Press the key to fix the contents selected in   
					. 
					ꢀ4   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Diopter Adjustment   
					Rotate the diopter adjustment ring for adjustment   
					of diopter.   
					Diopter   
					adjustment ring   
					Adjust so that A or B on aperture or black circle   
					indicating measuring area looks clear when   
					observing object through finder.   
					A 
					B 
					Adjustment would be easy starting with ꢀº   
					aperture where object near aperture looks blur.   
					Make sure to adjust diopter before   
					measurement. Diopter should be adjusted for   
					the eyesight of the person who will be taking   
					measurement. If diopter is not adjusted before   
					focus measurement, correct measurement value   
					may not be expected. This is because the focus   
					is actually off even if you think it is correctly in   
					focus. In addition, if diopter is not correctly   
					adjusted, you may see the aperture moving   
					depending on viewing angle.   
					* You sometimes see small black dots or stripes   
					in internal finder. It gives no effect on   
					measuring performance.   
					ꢀ5   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					LCD Screen   
					MEAS (Measurement value) screen   
					“SNGL” is displayed for the   
					values obtained during single   
					measurement, while “CONT”   
					is displayed for the values   
					obtained during continuous   
					measurement. (p.53)   
					Calibration channel (p.47) Target color channel   
					(p.6ꢀ, 65, 67)   
					Measurement result is displayed in the   
					currently selected color space.   
					(Lvxy, Lvu’v’, LvTΔuv, XYZ, Dominant   
					wavelength/Excitation purity, Spectral   
					graph) (p.39)   
					cd   
					m 
					2 
					Lv   
					x 
					80. 00   
					0 . 0000   
					Display format can be changed. (p.37)   
					y 
					Ob   
					0 . 0000   
					s 
					SPD SYNC LEN ND BL   
					10˚ M-N Non C U 100 on   
					Setting status in this instrument is   
					displayed.   
					<SPD>   
					<SYNC>   
					<Obs>   
					Currently selected   
					measuring time is   
					displayed.   
					“Int” is displayed when   
					the internal sync   
					Currently selected   
					observer angle is   
					displayed.   
					measurement mode is   
					(Nrm, Fst, M-N, M-F,   
					Mnl) (p.29)   
					set. “Ext” is displayed   
					when the external sync   
					measurement mode.   
					“Non” is displayed when   
					the sync measurement   
					mode is not set. (p.26)   
					(2°,ꢀ0°) (p.35)   
					ꢀ6   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					MENU screen   
					The MENU screen appears if MENU key is pressed when the measurement value screen   
					is displayed.   
					 MEAS   
					Used to set measurement time or   
					synchronizing method. (p.26, 29)   
					 MEMORY   
					Used to read or delete the measurement   
					memory data. (p.56, 58)   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					 TARGET   
					Used to register, select or delete the   
					target color. (p.6ꢀ, 65, 67)   
					 OPTION   
					Used to set the closeup lens, ND filter or   
					calibration channel. (p.43, 45, 47)   
					 SETUP   
					Used to set the observer, backlight or   
					display format. (p.35, 49, 37)   
					<LENS>   
					<ND>   
					<BL>   
					“C-U” is displayed, if the   
					closeup lens is attached.   
					If not, “Std” is displayed   
					(p.43).   
					The current ND filter type   
					is displayed.   
					“On” is displayed, if the   
					backlight is set to be   
					turned on during the   
					measurement. “Off” is   
					displayed, if the backlight   
					is set to be turned off   
					during that time. (p.49)   
					(Non, ꢀ0, ꢀ00) (p.45)   
					ꢀ7   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				
					ꢀ8   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Installation   
					ꢀ9   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Installing   
					Use screw holes for fixing at the bottom of this instrument if utilized with the tripod or jig.   
					2-type holes are available.   
					Tripod screw hole: To set on the tripod. Use a tripod screw with top diameter of 3/8 inches   
					and depth of ꢀ0.5 mm.   
					[Note] The tripod screw holes correspond with the 3/8-inch screws of a   
					large camera tripod. ꢀ/4-inch screws cannot be used for fixing   
					this instrument.   
					ISO screw hole: To set on the jig. Use ISO screws with top diameter of 5 mm and depth   
					of 6.5 mm.   
					Measurement distance   
					reference   
					ISO screw Tripod screw   
					ISO screw   
					Optical   
					axis   
					70   
					170   
					190.4   
					For other detailed dimensions, see p.80.   
					20   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Connecting AC Adapter   
					The AC adapter supplied with this instrument is used for the corresponding power   
					source.   
					(Failure to adhere to the following points may result in death or serious injury.)   
					Warning   
					Always use the AC adapter and power cord supplied as a standard accessory or   
					optional accessory (AC-A312), and connect it to indoor AC outlet of rated voltage   
					and frequency (100 - 120 V or 200 - 240 V , 50/60Hz). Failure to follow either of   
					these may result in damage to unit, fire or electric shock.   
					If this instrument is not used for a long time, disconnect AC adapter from AC   
					outlet. Accumulated dirt or water on prongs of AC adapter plug may cause fire and   
					should be removed before use.   
					Do not forcibly pull any part on power cord when unplugging since this may cause   
					fire or electric shock. Gently disconnect by holding plug. Also, do not handle   
					power cord with wet hands. Doing so may cause electric shock.   
					Do not forcibly bend, twist or pull power cord. Also, do not place heavy object on   
					power cord, or damage or modify one. Any of these may cause fire or electric   
					shock due to damage to power cord.   
					Do not disassemble or modify this instrument or AC adapter. Doing so may cause   
					fire or electric shock.   
					Should this instrument or AC adapter be damaged or smoke or odd smell be   
					generated, do not keep using such instrument or AC adapter without correction.   
					Doing so may cause fire. In such situations, switch power off immediately, unplug   
					AC adapter and contact the nearest KONICA MINOLTA authorized service facility.   
					Caution   
					(Failure to adhere to following points may result in injury or damage to this instrument or other property.)   
					Use this instrument near AC outlet for easy plugging or unplugging in using AC   
					adapter.   
					2ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Connection Method   
					1.   
					Make sure that power switch is OFF   
					(slided to [O] side).   
					2.   
					Connect the AC adapter plug to the   
					AC adapter input terminal of the body.   
					3.   
					Plug the AC adapter to outlet (100 - 120 VAC   
					50 Hz/60 Hz).   
					or 200 - 240 VAC   
					, 
					Insert the AC adapter plug all the way seated in AC outlet.   
					22   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Power Switch ON( | )/OFF(O)   
					The warm-up time needed is a minimum of 20 minutes in order to measure the objects   
					with excellent accuracy under the conditions described below. Warm up this instrument for   
					20 or more minutes when the power source is turned off even for a short period, and   
					turned on again.   
					(ꢀ) The object is a low-luminance light source, using 2856 K (standard light source A) as a   
					guide:   
					2 cd/m2 or lower (ꢀ° Aperture)   
					50 cd/m2 or lower (0.2° Aperture)   
					200 cd/m2 or lower (0.ꢀ° Aperture)   
					(2)Outside room temperature and normal humidity ranges   
					Turning power switch ON   
					1.   
					Slide power switch to ON ( | )   
					side.   
					The measurement screen appears 5   
					seconds after the initial screen on the   
					LCD.   
					The model type (CS-2000 or CS-   
					2000A), body version and product   
					serial numbers are displayed on the   
					initial screen. The model type can be   
					also confirmed on the nameplate.   
					Turning power switch OFF   
					2.   
					Slide power switch to OFF (O   
					side) after measurement.   
					23   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				
					24   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Setting   
					25   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Setting of Synchronization   
					The synchronized measurement refers to measurement mode where measurement is   
					made in the same timing as periodical light source pulse frequency, such as vertically   
					synchronized frequency for the display device.   
					[INT SYNC]   
					The internal sync measurement mode is used to measure the display equipment without   
					the input of vertically synchronized signals to the body, or to measure flicker light from a   
					light source such as a fluorescent light. Input the frequency of vertically synchronized   
					signals for the display equipment, or the commercial frequency (50 or 60 Hz) for flicker   
					light from a light source such as a luminescent light. The optimal integration time is   
					automatically set based on the input value and the brightness of the object. For this   
					reason, enter the correct frequency value to two places of decimals.   
					However, if the vertically synchronized frequency of the display equipment is not clear, accurate   
					measurement will not be possible with an inaccurate frequency setting. In this case it is   
					recommended to select [NO SYNC] mode without sync measurement (at 60 Hz of frequency), but to   
					select [MULTI-NORMAL] mode or [MULTI-FAST] mode for the measurement time (refer to p. 29).   
					[EXT SYNC]   
					The external sync measurement mode is used to measure the display equipment after the   
					line input of a vertically synchronized signal through the input terminal for vertically   
					synchronized signals to the body. The optimal integration time is set automatically, based   
					on the frequency of vertically synchronized signals and the brightness of the object.Input   
					CMOS (5V) Level of the input signals.   
					* Range of synchronized frequencies : 20.00 to 200.00 Hz   
					* Factory default setting : NO SYNC   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1.   
					Press   
					MENU   
					key when the MEAS   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					screen is displayed.   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					BACKLIGHT   
					off via   
					key on the MEAS screen,   
					y 
					Ob   
					the backlight is turned on.   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					26   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					2.   
					3.   
					Press either   
					[MEAS] and then press   
					The MENU – MEAS screen appears.   
					or   
					key to select   
					key.   
					MEAS   
					NORMAL   
					NO SYNC   
					ENTER   
					SPEED   
					SYNC   
					The current set contents are displayed in the   
					SYNC item.   
					Press either   
					[SYNC] and then press   
					or   
					key to select   
					key.   
					MEAS   
					SYNC MODE   
					ENTER   
					NO SYNC   
					The MENU - MEAS - SYNC MODE (sync   
					method selection) screen appears.   
					INT SYNC   
					EXT SYNC   
					4. Press either   
					or   
					key to select the   
					MEAS   
					SYNC MODE   
					sync method.   
					NO SYNC   
					INT SYNC   
					EXT SYNC   
					4-a-1. INT SYNC setting:   
					MEAS   
					SYNC MODE   
					INT SYNC   
					Select [INT SYNC] and press   
					key.   
					ENTER   
					060. 00 Hz   
					The MENU - MEAS - SYNC MODE – INT   
					SYNC screen appears. This screen is used   
					for the input of an internal synchronized   
					frequency.   
					4-a-2. Press either   
					value.   
					or   
					key to set a   
					key for larger number.   
					key for smaller number.   
					The range of the internal synchronized   
					frequency is from 20 Hz to 200 Hz.   
					4-a-3. Press either   
					or   
					key to move   
					MEAS   
					SYNC MODE   
					INT SYNC   
					the cursor.   
					050. 00 Hz   
					27   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					4-a-4. Repeat the same procedures 4-a-   
					2. and -3. as necessary.   
					4-a-5. Press   
					key.   
					ENTER   
					After the setting is entered, the MENU -   
					MEAS - SYNC MODE screen appears   
					again on the LCD.   
					5. Press   
					key.   
					ENTER   
					When the synchronizing method is set, the   
					MENU - MEAS screen appears again on the   
					LCD.   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - MEAS screen appears   
					again on the LCD.   
					The setting of the synchronizing method is   
					saved even after the power switch is turned off   
					(O).   
					6. Press   
					key.   
					ENTER   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					2 
					The MEAS screen appears again on the LCD.   
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					I n t S t dNon on   
					28   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Selecting Measurement Time   
					Select measurement time according to the purpose. Five modes are available for   
					measurement time.   
					Measurement   
					time mode   
					Examples of   
					measurement subjects   
					When measuring low luminances, Constant light   
					Operation explanation   
					Feature   
					Cautions   
					In this mode, integration time*ꢀ is adjusted Improved accuracy   
					between 0.005s and ꢀ20s according to the and repeatability for   
					brightness of the measurement subject.   
					This mode emphasizes performance   
					when measuring low luminances.   
					since the total measurement   
					time can be as long as 4   
					sources (halogen   
					lamps, etc.)   
					luminances below   
					approx. 4 cd/m2   
					minutes, if the brightness of the   
					(measurement angle: ꢀ°) measurement subject changes,   
					the measurement results will   
					be the average luminance.   
					Periodic light   
					sources (whose are   
					stable and whose   
					synchronization   
					frequency is   
					In addition, be careful to not   
					move the instrument during   
					NORMAL   
					measurement. When measuring   
					periodic light sources, MULTI   
					INTEG-NORMAL mode or   
					known)   
					MULTI INTEG-FAST mode may   
					be more suitable, for example,   
					when the synchronization   
					frequency is unknown.   
					In this mode, integration time is adjusted   
					between 0.005s and ꢀ6s according to the times for luminances   
					brightness of the measurement subject.   
					This mode emphasizes measurement   
					time when measuring low luminances.   
					Shorter measuring   
					When higher accuracy and   
					repeatability are required   
					at low luminances, the   
					Same as above   
					Periodic light   
					below approx. 4 cd/m2   
					(measurement angle: ꢀ°) NORMAL mode should   
					be used if necessary.   
					FAST   
					In this mode, several cycles of the   
					integration time for NORMAL mode are   
					taken and averaged. Under luminance   
					conditions which require an integration   
					time longer than the set luminance, the   
					integration time will be the same as for   
					normal measurement.   
					This mode can be used when the   
					synchronization time for measurement is   
					unknown or when the synchronization   
					time is known but the frequency is   
					unstable. In such case, the sync mode   
					should be set to NO SYNC.   
					Measurements which do Even for high luminances,   
					not depend on the   
					synchronization   
					frequency of the   
					measurement subject   
					can be taken.   
					the set integration time (ꢀs or sources (whose   
					longer) will be used.   
					synchronization   
					frequency is   
					unknown or   
					unstable)   
					MULTI   
					INTEG-   
					NORMAL   
					Improved accuracy and   
					repeatability for   
					luminances below   
					approx. 4 cd/m2   
					(measurement angle: ꢀ°)   
					In this mode, several cycles of the   
					Measurements which do Even for high luminances,   
					Periodic light   
					integration time for FAST mode are taken not depend on the   
					the set integration time (ꢀs or sources (whose   
					and averaged. Under luminance   
					conditions which require an integration   
					time longer than the set luminance, the   
					integration time will be the same as for   
					FAST measurement.   
					This mode can be used when the   
					synchronization time for measurement is   
					unknown or when the synchronization   
					time is known but the frequency is   
					unstable. In such case, the sync mode   
					should be set to NO SYNC.   
					synchronization   
					frequency of the   
					longer) will be used   
					synchronization   
					frequency is   
					unknown or   
					unstable)   
					measurement subject   
					can be taken.   
					Shorter measuring times   
					for luminances below   
					approx. 4 cd/m2   
					MULTI   
					INTEG-   
					FAST   
					(measurement angle: ꢀ°)   
					This mode can be used when you want to set a   
					fixed integration time for measurements.   
					Integration time: 0.005s to ꢀ20s   
					The desired fixed integration Be careful that the “OVER” error   
					All light sources   
					MANUAL   
					time can be set.   
					message does not occur and that the   
					measurement accuracy is not reduced.   
					*ꢀ Time for sensor to measure light indicating “exposure time”. On the other hand, measurement time shows time for   
					integration × 2 +Time to open/close shutter + Time for calculation), indicating time needed for actual measurement.   
					* Factory default setting : MULTI INTEG-NORMAL, IN-ND : AUTO   
					29   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					key when the MEAS   
					screen is displayed.   
					MENU   
					cd   
					m 
					2 
					Lv   
					x 
					35. 85   
					0 . 3964   
					0 . 4059   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					key.   
					MEAS   
					NORMAL   
					NO SYNC   
					[MEAS] and then press   
					SPEED   
					SYNC   
					ENTER   
					The MENU – MEAS screen appears.   
					The current set contents are displayed in the   
					SPEED item.   
					3. Press either   
					or   
					key to select   
					key.   
					MEAS   
					SPEED MODE   
					[SPEED] and then press   
					ENTER   
					NORMAL   
					The MENU - MEAS - SPEED MODE   
					(measurement speed setting) screen   
					appears.   
					FAST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					30   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					4. Press either   
					or   
					key to select   
					MEAS   
					SPEED MODE   
					measurement speed.   
					NORMAL   
					FAST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					4-a-1. NORMAL or FAST setting: Select   
					[NORMAL] or [FAST] and press   
					MEAS   
					SPEED MODE   
					IN ND   
					AUTO   
					OFF   
					ON   
					key.   
					ENTER   
					The MENU - MEAS - SPEED MODE - IN-   
					ND screen appears.   
					This screen is used to determine whether the   
					ND filter built into the body is used or not.   
					4-a-2. Press either   
					or   
					key to select   
					MEAS   
					SPEED MODE   
					IN ND   
					[AUTO], [OFF] or [ON].   
					AUTO   
					OFF   
					ON   
					Select [ON] when the error message “OVER” appears.   
					When measuring objects having a wide range of luminance   
					under the adjusted measurement conditions (ex. γ   
					measurement), set IN-ND to [OFF] or [ON]. For [OFF] or [ON]   
					selection, select [OFF] when the upper limit luminance is ꢀ00cd/   
					m2 or less, and [ON] when it is ꢀ00cd/m2 or more as a guide.   
					4-a-3. Press   
					key.   
					ENTER   
					MEAS   
					SPEED MODE   
					After the setting is entered, the MENU   
					- MEAS - SPEED MODE screen   
					appears again on the LCD.   
					NORMAL   
					FA ST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					4-b-1. MULTI-NORMAL or MULTI-FAST   
					setting:   
					MEAS   
					SPEED MODE   
					MULTI INTEG   
					INTEG TIME   
					01   
					Select [MULTI-NORMAL] or [MULTI-   
					FAST] and press   
					key.   
					ENTER   
					IN-ND AUTO   
					The MENU - MEAS - SPEED MODE -   
					MULTI INTEG screen appears.   
					This screen is used for input of the integration   
					time in the MULTI INTEG-NORMAL mode or   
					MULTI INTEG-FAST mode.   
					3ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					4-b-2. Press either   
					value.   
					or   
					key to set a   
					MEAS   
					SPEED MODE   
					MULTI INTEG   
					INTEG TIME   
					10   
					key for larger number.   
					key for smaller number.   
					IN-ND AUTO   
					The setting range of integration time is from ꢀ to ꢀ6 s.   
					4-b-3. Press   
					key.   
					ENTER   
					MEAS   
					SPEED MODE   
					MULTI INTEG   
					The cursor moves to the IN-ND item.   
					This screen is used to determine whether   
					the ND filter built into the body is used or   
					not.   
					INTEG TIME   
					10   
					IN-ND AUTO   
					4-b-4. Press either   
					or   
					key to   
					MEAS   
					SPEED MODE   
					MULTI INTEG   
					select [AUTO], [OFF] or [ON].   
					INTEG TIME   
					10   
					Select [ON] when the error message “OVER” appears.   
					When measuring objects having a wide range of luminance   
					under the adjusted measurement conditions (ex. γ   
					IN-ND ON   
					measurement), set IN-ND to [OFF] or [ON]. For [OFF] or [ON]   
					selection, select [OFF] when the upper limit luminance is ꢀ00cd/   
					m2 or less, and [ON] when it is ꢀ00cd/m2 or more as a guide.   
					MEAS   
					SPEED MODE   
					4-b-5. Press   
					key.   
					ENTER   
					NORMAL   
					After the setting is entered, the MENU -   
					MEAS - SPEED MODE screen appears   
					again on the LCD.   
					FAST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					4-c-1. MANUAL setting:   
					MEAS   
					SPEED MODE   
					Select [MANUAL] and press   
					NORMAL   
					key.   
					ENTER   
					FAST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					The MENU - MEAS – SPEED MODE -   
					MANUAL screen appears. This screen is   
					used for input of the integration time in the   
					manual mode.   
					MEAS   
					SPEED MODE   
					MANUAL   
					INTEG TIME   
					000033.333   
					ms   
					IN ND OFF   
					32   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					4-c-2. Press either   
					value.   
					or   
					key to set a   
					MEAS   
					SPEED MODE   
					MANUAL   
					INTEG TIME   
					key for larger number.   
					000033.330   
					m 
					s 
					key for smaller number.   
					IN ND OFF   
					The setting range of integration time is from   
					5 to ꢀ20,000 ms.   
					4-c-3. Press either   
					the cursor.   
					or   
					key to move   
					4-c-4. Repeat the same procedures 4-a-   
					2. and -3. as necessary.   
					MEAS   
					SPEED MODE   
					MANUAL   
					INTEG TIME   
					000500.000   
					m 
					s 
					s 
					s 
					IN ND OFF   
					4-c-5. Press   
					key.   
					ENTER   
					MEAS   
					SPEED MODE   
					MANUAL   
					The cursor moves to the IN-ND item.   
					This screen is used to determine whether   
					the ND filter built into the body is used or   
					not.   
					INTEG TIME   
					000500.000   
					m 
					IN ND OFF   
					4-c-6. Press either   
					[ON] or [OFF].   
					or   
					key to select   
					MEAS   
					SPEED MODE   
					MANUAL   
					INTEG TIME   
					Select [ON] when the error message   
					“OVER” appears.   
					000500.000   
					IN   
					m 
					N 
					D ON   
					4-c-7. Press   
					key.   
					ENTER   
					MEAS   
					SPEED MODE   
					When the setting is entered, the MENU -   
					MEAS - SPEED MODE screen appears   
					again on the LCD.   
					NORMAL   
					FAST   
					MULTI-NORMAL   
					MULTI-FAST   
					MANUAL   
					33   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					5. Press   
					key.   
					ENTER   
					When the measurement time is set, the MENU -   
					MEAS screen appears again on the LCD.   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - MEAS screen appears   
					again on the LCD.   
					Measurement time setting is saved even after   
					switching OFF (O).   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					2 
					The MEAS screen appears again on the LCD.   
					Lv   
					x 
					35. 85   
					0 . 3964   
					0 . 4059   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ Mn   
					l 
					Non S t dNon on   
					34   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Setting Observer   
					Color matching function for chromaticity calculation is selectable between 2°OBS and   
					ꢀ0°OBS.   
					* Observer setting   
					: 2°OBS, ꢀ0°OBS   
					* Factory default setting : 2°OBS   
					Operation Procedure   
					8, 9   
					1 
					2, 3, 4,   
					5, 6, 7   
					1. Press   
					key when the MEAS   
					screen is displayed.   
					MENU   
					cd   
					m 
					2 
					Lv   
					x 
					35. 85   
					The MENU screen appears.   
					0 . 3964   
					When the backlight of the LCD has been turned   
					BACKLIGHT   
					the backlight is turned on.   
					off via   
					key on the MEAS screen,   
					y 
					Ob   
					0 . 4059   
					Non S t dNon on   
					s 
					SPD SYNC LEN ND BL   
					l 
					2 
					˚ Mn   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					SETUP   
					[SETUP] and then press   
					key.   
					ENTER   
					OBSERVER   
					BACKLIGHT   
					2 
					° 
					@ 
					MEAS   
					O 
					The MENU – SETUP screen appears.   
					The current set contents are displayed in   
					N 
					F 
					DATA FORM   
					the [OBSERVER] item.   
					3. Press either   
					or   
					key to select   
					SETUP   
					OBSERVER   
					[OBSERVER] and then press   
					key.   
					ENTER   
					° 
					2 
					O 
					BS   
					° 
					10   
					O 
					BS   
					The MENU - SETUP – OBSERVER (Observer   
					selection) screen appears.   
					35   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					4. Press either   
					or   
					key to select [2º   
					SETUP   
					OBSERVER   
					OBS] or [10º OBS].   
					° 
					2 
					O 
					BS   
					° 
					10   
					O 
					BS   
					5. Press   
					key.   
					ENTER   
					SETUP   
					OBSERVER   
					The confirmation screen appears. [CANCEL] is   
					CHANGE   
					selected.   
					OK   
					CANCEL   
					6. Press   
					key to move the cursor to   
					SETUP   
					OBSERVER   
					[OK].   
					CHANGE   
					OK   
					CANCEL   
					7 
					. Press   
					key.   
					ENTER   
					When the observer angle is set, the MENU -   
					MEAS screen appears again on the LCD.   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - SETUP screen appears   
					again on the LCD.   
					Observer setting is saved even after switching   
					OFF (O).   
					8. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					9. Press   
					key.   
					ESC   
					cd   
					m 
					2 
					Lv   
					x 
					35. 85   
					The MEAS screen appears again on the LCD.   
					When observer angle is set to ꢀ0°, Lv display   
					changes to Y display.   
					0 . 3964   
					y 
					Ob   
					0 . 4059   
					Non S t dNon on   
					s 
					SPD SYNC LEN ND BL   
					l 
					10˚ Mn   
					36   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Selecting Display Format   
					The formats of indicating the luminance and excitation values X, Y and Z can be selected as either   
					normal indication to display the values to four places of decimals, or as index number indication. If   
					the measurement values on the LCD are unreadable, use the index number indication.   
					* Display format setting : Normal, Index   
					* Factory default setting : ****.**** [F]   
					* When the number of displayed digits is six (luminance and X/Y/Z are ꢀ00000 or more) in   
					the normal indication, “******.*****” will be displayed. In this case, the value will be   
					displayed if you set to the index number indication.   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					key when the MEAS   
					screen is displayed.   
					MENU   
					cd   
					m 
					2 
					Lv   
					x 
					35. 85   
					0 . 3964   
					0 . 4059   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					the backlight is turned on.   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ Mn   
					l 
					Non S t dNon on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					key.   
					SETUP   
					[SETUP] and then press   
					ENTER   
					OBSERVER   
					BACKLIGHT   
					2 
					° 
					@ 
					MEAS   
					O 
					The MENU – SETUP screen appears.   
					The current set contents are displayed in the   
					N 
					F 
					DATA FORM   
					[DATA FORM] item.   
					3. Press either   
					or   
					key to select   
					SETUP   
					DATA FORM   
					[DATA FORM] and then press   
					. 
					**** ****   
					key.   
					ENTER   
					E 
					. 
					+ 
					* ****   
					* 
					The MENU - SETUP - DATA FORM (Display   
					format selection) screen appears.   
					37   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					4. Press either   
					or   
					key to select   
					SETUP   
					DATA FORM   
					[****.**** [F]] or [*.****E+* [E]].   
					. 
					**** ****   
					E 
					. 
					+ 
					* ****   
					* 
					5. Press   
					key.   
					ENTER   
					SETUP   
					OBSERVER   
					BACKLIGHT   
					2 
					° 
					When the display format is set, the MENU -   
					@ 
					MEAS   
					O 
					SETUP screen appears again on the LCD.   
					By pressing   
					N 
					E 
					ESC   
					key, after the setting is   
					DATA FORM   
					canceled the MENU - SETUP screen appears   
					again on the LCD.   
					Display format setting is saved even after   
					switching OFF (O).   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					2 
					Lv 3. 5846E+1   
					The MEAS screen appears again on the LCD.   
					m 
					0. 3964   
					x 
					y 
					Ob   
					0. 4059   
					Non S t dNon on   
					s 
					SPD SYNC LEN ND BL   
					l 
					2 
					˚ Mn   
					38   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Selecting Color Space   
					See below table for available color space.   
					* Factory default setting :Lvxy   
					LCD Screen   
					Color Space   
					Display Description   
					(When Normal is selected for display format)   (When Index is selected for display format)    
					Lvxy *1   
					Displays and outputs in   
					luminance Lv and   
					chromaticity coordinates   
					x,y.   
					cd   
					m 
					cd   
					2 
					2 
					m 
					Lv   
					x 
					34. 22   
					0 . 3958   
					0 . 4060   
					Lv 3. 4221E+1   
					0 . 3958   
					x 
					y 
					Ob   
					y 
					Ob   
					0 . 4060   
					Non S t dNon on   
					s 
					SPD SYNC LEN ND BL   
					s 
					SPD SYNC LEN ND BL   
					m 
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					2 
					˚ N   
					r 
					Lvu’v’ *1   
					Lv T ∆uv   
					XYZ   
					Displays and outputs in   
					luminance Lv and u’v’   
					chromaticity diagram   
					(CIE ꢀ976 UCS   
					chromaticity diagram)   
					coordinates u’, v’.   
					cd   
					m 
					cd   
					2 
					2 
					Lv   
					u’   
					v’   
					34. 22   
					0 . 2236   
					0 . 5161   
					Lv 3. 4221E+1   
					m 
					u’   
					v’   
					0 . 2236   
					0 . 5161   
					Non S t dNon on   
					Ob   
					s 
					SPD SYNC LEN ND BL   
					Ob   
					s 
					SPD SYNC LEN ND BL   
					m 
					2˚ N   
					r 
					m 
					Non S t dNon on   
					2˚ N   
					r 
					Displays and outputs in   
					luminance Lv, correlated   
					color temperature T and   
					color difference from   
					blackbody locus ∆uv.   
					cd   
					m 
					cd   
					2 
					2 
					Lv   
					T 
					34. 22   
					3829K   
					Lv 3. 4221E+1   
					m 
					T 
					3829K   
					duv   
					duv   
					Obs SPD SYNC LEN ND BL   
					+0. 009   
					Non S t dNon on   
					+0. 009   
					Non S t dNon on   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					m 
					r 
					2˚ N   
					r 
					m 
					Displays and outputs in   
					tristimulus values X, Y,   
					Z.   
					X 
					Y 
					33. 36   
					34. 22   
					16. 71   
					X 
					Y 
					3. 3365E+1   
					3. 4221E+1   
					Z 
					Ob   
					Z 
					Ob   
					1. 6709E+1   
					s 
					SPD SYNC LEN ND BL   
					s 
					SPD SYNC LEN ND BL   
					m 
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					2 
					˚ N   
					r 
					Non S t dNon on   
					Dominant   
					wavelength/   
					Excitation purity   
					*2   
					Displays and outputs in   
					dominant wavelength λd   
					and excitation purity Pe.   
					+576 . 220n   
					m 
					d 
					λ 
					0 . 407%   
					Pe   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					Spectral graph   
					Displays or outputs   
					spectral radiance Le(λ)   
					in the spectral waveform   
					380n   
					1. 3118E   
					m 
					380   
					Ob   
					780   
					5 
					SPD SYNC LEN ND BL   
					s 
					˚ N   
					2 
					r 
					m 
					Non S t dNon on   
					39   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					*ꢀ Y is displayed instead of Lv when observer angle is ꢀ0°.   
					*2 For non-spectral colors, the complementary wavelength will be displayed. The display   
					indication will remain λd.   
					* 
					If the calculated value does not establish a proper combination with the value in the   
					color space mode, “_ _ _ _ _ _ _ _” will be displayed.   
					Operation Procedure   
					1 
					2 
					1. When the MENU or MEMORY screen   
					cd   
					m 
					2 
					Lv   
					x 
					34. 22   
					0 . 3958   
					0 . 4060   
					is displayed, press   
					key to   
					ESC   
					switch to the MEAS screen.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					2.   
					Press   
					key to display   
					COLOR MODE   
					cd   
					m 
					2 
					Lv   
					u’   
					v’   
					34. 22   
					0 . 2236   
					0 . 5161   
					the desired color space.   
					Measurement screen switches in order of Lvxy   
					 Lvu’v’  LvT∆uv  XYZ  λd/Pe  Spectral   
					graph  Lvxy while   
					COLOR MODE   
					key is   
					Ob   
					s 
					SPD SYNC LEN ND BL   
					pressed.   
					2˚ N   
					r 
					m 
					Non S t dNon on   
					It switches in order of Yxy  Yu’v’  XYZ  λd/   
					Pe  Spectral graph  Yxy when observer   
					angle is ꢀ0°.   
					Color space setting is saved even after   
					switching OFF (O).   
					40   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Selecting Absolute Value (ABS)/Difference (DIFF) Display   
					Whether chromaticity value is shown in absolute (ABS) or difference (DIFF) is selectable.   
					See below table for each case.   
					* Factory default setting: Absolute value (ABS)   
					Color Space   
					Lvxy *1   
					Absolute value (ABS)   
					Difference (DIFF)   
					Lv, x , y   
					Lv, x, y   
					∆Lv, %LV, ∆x, ∆y   
					cd   
					2 
					Lv   
					x 
					34. 22   
					0 . 3958   
					0 . 4060   
					m 
					cd   
					2 
					2 
					Lv   
					x 
					34   
					3958   
					4059   
					. 
					34   
					m 
					0 
					0 
					. 
					. 
					y 
					cd   
					m 
					Lv   
					+ 
					0 
					. 
					. 
					1188   
					35   
					0000   
					0001   
					% 
					Lv   
					100   
					% 
					x 
					y 
					. 
					. 
					0 
					0 
					. 
					. 
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					Obs   
					2˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					Lvu’v’ *1   
					Lv T ∆uv   
					XYZ   
					Lv, u’, v’   
					Lv, u’, v’   
					%Lv, ∆Lv, ∆u’, ∆v’   
					cd   
					m 
					2 
					Lv   
					u’   
					v’   
					34. 22   
					0 . 2236   
					0 . 5161   
					cd   
					2 
					2 
					Lv   
					u’   
					v’   
					34   
					2236   
					5160   
					. 
					34   
					m 
					0 
					0 
					. 
					. 
					cd   
					m 
					Lv   
					Lv   
					+ 
					0 
					. 
					1188   
					35   
					0000   
					0000   
					% 
					100   
					. 
					% 
					u ’   
					v ’   
					+ 
					0 
					. 
					. 
					0 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					r 
					m 
					Non S t dNon on   
					Lv, T, duv   
					Lv   
					Lv, T, duv   
					∆Lv, %Lv, ∆T   
					cd   
					m 
					2 
					34. 22   
					3829K   
					cd   
					m 
					2 
					2 
					Lv   
					T 
					duv   
					34 . 34   
					3830K   
					+ 
					T 
					0 
					. 
					009   
					cd   
					m 
					Lv   
					+ 
					100   
					0 
					. 
					. 
					35   
					OK   
					1188   
					% 
					% 
					Lv   
					duv   
					+0. 009   
					Non S t dNon on   
					T 
					Ob SPD SYNC LEN ND BL   
					r 
					s 
					˚ N   
					2 
					m 
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					X, Y, Z   
					X, Y, Z   
					∆X, ∆Y, ∆Z   
					%X, %Y, %Z   
					X 
					Y 
					33. 36   
					34. 22   
					16. 71   
					X 
					Y 
					Z 
					33   
					34   
					16   
					. 
					. 
					. 
					. 
					. 
					. 
					49   
					34   
					78   
					1204   
					1188   
					07479   
					% 
					% 
					% 
					X 
					Y 
					Z 
					+ 
					+ 
					+ 
					0 
					0 
					0 
					. 
					. 
					. 
					Z 
					Ob   
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					2 
					r 
					m 
					Non S t dNon on   
					X%   
					Y%   
					Z%   
					100   
					100   
					100   
					36   
					35   
					45   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					λ d, Pe   
					λd  Pe   
					∆λd, ∆Pe *2   
					Dominant wavelength   
					+576 . 220n   
					m 
					d 
					λ 
					λ 
					d 
					+ 
					576   
					0 
					. 
					406   
					220n   
					m 
					Fe   
					. 
					0 . 407%   
					Pe   
					λd   
					Pe   
					0 . 000n   
					0 . 000%   
					m 
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					r 
					m 
					Non S t dNon on   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					4ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Color Space   
					Absolute value (ABS)   
					Difference (DIFF)   
					Spectral graph   
					Le (λ) spectral waveform   
					Measurement value and Le (λ)   
					Wavelength at the cursor position spectral waveform of target color   
					and Le (λ) in the wavelength   
					Wavelength of the cursor position   
					and Le (λ) of the measurement value   
					in the wavelength   
					380n   
					1. 3118E   
					SPD SYNC LEN ND BL   
					˚ N Non S t dNon on   
					m 
					380   
					780   
					5 
					Ob   
					2 
					s 
					rm   
					380n   
					1.8704E   
					SPD SYNC LEN ND BL   
					˚ N Non S t dNon on   
					m 
					380   
					780   
					+ 
					7 
					Ob   
					2 
					s 
					rm   
					*ꢀ Y is displayed instead of Lv when observer angle is ꢀ0°.   
					*2 Even if either the measured value, target color, or both are the complementary   
					wavelength, the difference between the two values will be displayed. The display   
					indication will remain ∆λd.   
					* 
					If the calculated value or the registered target color does not establish a proper   
					combination with the value in the color space mode, “_ _ _ _ _ _ _ _” will be displayed.   
					Operation Procedure   
					1 
					2 
					1. When the MENU or MEMORY screen   
					cd   
					m 
					2 
					ESC   
					Lv   
					x 
					34. 22   
					0 . 3958   
					0 . 4060   
					is displayed, press   
					key to   
					switch to the MEAS screen.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					Non S t dNon on   
					2.   
					Press   
					key to show   
					DISPLAY   
					cd   
					2 
					2 
					Lv   
					x 
					34   
					3958   
					4059   
					. 
					34   
					m 
					absolute value (ABS) or color   
					difference (DIFF) to select.   
					0 
					. 
					. 
					y 
					0 
					cd   
					m 
					Lv   
					+ 
					0 
					. 
					. 
					1188   
					35   
					0000   
					0001   
					% 
					Lv   
					100   
					% 
					Measurement value switches between that for   
					x 
					. 
					. 
					0 
					0 
					. 
					. 
					y 
					absolute value (ABS) and difference (DIFF)   
					DISPLAY   
					while   
					key is pressed.   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					Non S t dNon on   
					rm   
					The setting of absolute value (ABS) and difference   
					(DIFF) is saved even after switching OFF (O).   
					42   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					When Using Closeup Lens   
					Use the closeup lens in the optional accessories to measure a fine surface. For the   
					attachment of the closeup lens, refer to the instruction manual for the closeup lens.   
					If the closeup lens is attached, the measurement value must be compensated for the lens   
					transmittance. This compensation coefficient is attached to the closeup lens. Use the data   
					management software CS-Sꢀ0w Professional supplied with the CS-2000 as standard to   
					set the coefficient in this instrument. After that, when the closeup lens is selected as lens   
					type, the measurement value compensated with the compensation coefficient can be   
					obtained. See instruction manual of CS-Sꢀ0w for details.   
					The wrong setting of lens type leads to inaccurate measurement.   
					Do not use the closeup lens with the ND filter. It leads to inaccurate measurement.   
					* Lens type   
					: STANDARD, CLOSE-UP   
					* Factory default setting : STANDARD   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					MENU   
					key when the MEAS   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					screen is displayed.   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					I n t S t dNon on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					OPTION   
					[OPTION] and then press   
					key.   
					ENTER   
					USER. CAL   
					LENS   
					NON   
					STD   
					NON   
					ND FILTER   
					The MENU – OPTION screen appears.   
					The current set contents are displayed in the   
					[LENS] item.   
					43   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					3. Press either   
					or   
					key to select   
					key.   
					OPTION   
					LENS   
					[LENS] and then press   
					ENTER   
					STANDARD   
					CLOSE UP   
					The MENU - SETUP – LENS (Lens type   
					selection) screen appears.   
					4. Press either   
					[CLOSE-UP].   
					or   
					key to select   
					OPTION   
					LENS   
					STANDARD   
					CLOSE UP   
					When you removed the closeup lens, select   
					[STANDARD].   
					5. Press   
					key.   
					ENTER   
					OPTION   
					USER. CAL   
					LENS   
					NON   
					C UP   
					NON   
					When the lens type is set, the MENU - OPTION   
					screen appears again on the LCD.   
					ND FILTER   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - OPTION screen appears   
					again on the LCD.   
					Lens type setting is saved even after switching   
					OFF (O).   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					The MEAS screen appears again on the LCD.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					I n t C UNon on   
					44   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					When Using ND Filter   
					Use the ND filter in the optional accessories when measuring a high-luminance object.   
					If the ND filter is attached, the measurement value must be compensated for the filter   
					transmittance. This compensation coefficient is attached to the ND filter. Use the data management   
					software CS-Sꢀ0w Professional supplied with the CS-2000 as standard to set the coefficient in this   
					instrument. After that, when the ND filter is selected, the measurement value compensated with the   
					compensation coefficient can be obtained. See instruction manual of CS-Sꢀ0w for details.   
					The wrong setting of ND filter leads to inaccurate measurement.   
					Do not use the ND filter with the closeup lens. It leads to inaccurate measurement.   
					Note that an extra ND filter is built into this instrument. There are three setting options for use or   
					non-use of this built-in ND filter: [AUTO] for switching automatically depending on the luminance of   
					the object, [OFF] for constant non-use, and [ON] for constant use (See p.3ꢀ to p.33).   
					* ND-FILTER   
					: NONE, NDꢀ0, NDꢀ00   
					* Factory default setting: NONE   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					MENU   
					key when the MEAS   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					screen is displayed.   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					I n t S t dNon on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					OPTION   
					USER. CAL   
					LENS   
					NON   
					STD   
					NON   
					[OPTION] and then press   
					key.   
					ENTER   
					ND FILTER   
					The MENU – OPTION screen appears.   
					The current set contents are displayed in the   
					[ND-FILTER] item.   
					45   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					3. Press either   
					or   
					key to select   
					OPTION   
					FILTER   
					N 
					D 
					[ND-FILTER] and then press   
					key.   
					ENTER   
					NONE   
					N 
					N 
					D 
					D 
					10   
					The MENU - SETUP – ND-FILTER (ND filter   
					100   
					selection) screen appears.   
					4. Press either   
					or   
					key to select   
					OPTION   
					D 
					N 
					FILTER   
					[NONE], [ND 10] or [ND 100].   
					NONE   
					N 
					N 
					D 
					D 
					10   
					100   
					5. Press   
					key.   
					ENTER   
					OPTION   
					USER. CAL   
					LENS   
					NON   
					STD   
					10   
					When the ND filter is set, the MENU - OPTION   
					screen appears again on the LCD.   
					ND FILTER   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - OPTION screen appears   
					again on the LCD.   
					ND filter setting is saved even after switching   
					OFF (O).   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					The MEAS screen appears again on the LCD.   
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					y 
					Ob   
					0 . 4060   
					I n t S t d 10 on   
					s 
					SPD SYNC LEN ND BL   
					r 
					2 
					˚ N   
					m 
					46   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Calibration   
					Calibration Channel   
					This instrument includes ꢀꢀ calibration channels from Ch00 to Chꢀ0.   
					Ch00 is for measurement based upon KONICA MINOLTA’s calibration standard. Its   
					calibration correction coefficient has been set and is unchangeable.   
					The following contents can be set to Ch0ꢀ to Chꢀ0 respectively using data management   
					software CS-Sꢀ0w Professional supplied with the CS-2000 as standard. See instruction   
					manual of CS-Sꢀ0w for details.   
					 Correction coefficient of user calibration   
					 Correction coefficient ID   
					They are commonly used among each color space of Lvxy, Lvu’v’, LvT∆uv, XYZ, dominant   
					wavelength/excitation purity and spectral graph in one channel.   
					Calibration channels can be changed using the following procedure.   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					MENU   
					key when the MEAS   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					screen is displayed.   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t d 10 on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					OPTION   
					USER. CAL   
					[OPTION] and then press   
					key.   
					NON   
					STD   
					10   
					ENTER   
					LENS   
					ND FILTER   
					The MENU – OPTION screen appears.   
					The current set contents are displayed in the   
					[USER.CAL] item.   
					47   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					3. Press either   
					or   
					key to select   
					OPTION   
					USER. CAL   
					[USER.CAL] and then press   
					key.   
					ENTER   
					CAL NO,   
					NON   
					The MENU - SETUP – USER.CAL (Calibration   
					channel selection) screen appears.   
					The calibration channel number and   
					compensation coefficient ID (with a maximum of   
					ꢀ0 letters) are displayed. In case of Ch00,   
					“NON” is displayed. See instruction manual of   
					CS-Sꢀ0w for details.   
					4. Press either   
					channel.   
					or   
					key to select the   
					OPTION   
					USER. CAL   
					CAL NO,   
					03   
					key for larger number.   
					key for smaller number.   
					The calibration channel numbers can be   
					selected among NON, and 0ꢀ to ꢀ0.   
					5. Press   
					key.   
					ENTER   
					OPTION   
					USER. CAL   
					03   
					STD   
					10   
					When the calibration channel is set, the MENU -   
					OPTION screen appears again on the LCD.   
					If the calibration channel without compensation   
					coefficient setting is selected, the setting will not   
					be possible.   
					LENS   
					ND FILTER   
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - OPTION screen appears   
					again on the LCD.   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					The MEAS screen appears again on the LCD.   
					0 . 3968   
					y 
					Ob   
					0 . 4060   
					s 
					SPD SYNC LEN ND BL   
					r 
					2 
					˚ N   
					m 
					No   
					n 
					S t d 10 on   
					48   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Backlight ON/OFF During Measurement   
					The LCD backlight can be selectively turned on or off during measurement.   
					If the backlight is turned off, LCD backlight reflection on the surrounding area   
					affecting the measurement value during measurement can be avoided.   
					If pressing the BACKLIGHT key to turn off the backlight, while the MEAS screen is   
					displayed, the backlight is forcibly turned off independently of the following setting.   
					* Factory default setting : ON   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					MENU   
					key when the MEAS   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					screen is displayed.   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t d 10 on   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					key.   
					SETUP   
					OBSERVER   
					BACKLIGHT   
					2 
					° 
					[SETUP] and then press   
					ENTER   
					@ 
					MEAS   
					O 
					The MENU – SETUP screen appears.   
					N 
					F 
					DATA FORM   
					The current set contents are displayed in the   
					[BACKLIGHT@MEAS] item.   
					49   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					3. Press either   
					or   
					key to select   
					SETUP   
					BACKLIGHT   
					@ 
					MEAS   
					[BACKLIGHT@MEAS] and then press   
					O 
					O 
					FF   
					N 
					key.   
					ENTER   
					The MENU - SETUP - BACKLIGHT@MEAS   
					(Turning ON/OFF backlight during   
					measurement) screen appears on the LCD.   
					4. Press either   
					[ON] or [OFF].   
					or   
					key to select   
					5. Press   
					ENTER   
					key.   
					SETUP   
					OBSERVER   
					BACKLIGHT   
					2 
					° 
					When the setting is made to turn the backlight   
					on or off during measurement, the MENU -   
					SETUP screen appears again on the LCD.   
					@ 
					MEAS   
					O 
					FF   
					DATA FORM   
					F 
					ESC   
					By pressing   
					key, after the setting is   
					canceled the MENU - SETUP screen appears   
					again on the LCD.   
					Setting of backlight ON/OFF during   
					measurement is saved even after switching OFF   
					(O).   
					6. Press   
					key.   
					ENTER   
					The MENU screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					cd   
					m 
					The MEAS screen appears again on the LCD.   
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t d 10 off   
					50   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Measurement   
					5ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Measurement   
					Operation Procedure   
					5 
					6 
					2 
					1, 4   
					3 
					1.   
					Decide whether you use closeup lens   
					(optional accessory) or not according to   
					measuring object size and distance.   
					See the table below for details on measuring distance   
					and measuring area. If you set the closeup lens, lens   
					type setting is required in this instrument. (See p.43)   
					(Unit: mm)   
					Measuring distance and measuring area   
					Measuring area   
					when the measuring   
					distance is 500 mm ø distance is ꢀ,000 mm ø   
					Measuring area   
					when the measuring   
					Minimum measurement   
					diameter ø   
					Maximum measurement   
					diameter ø   
					Minimum   
					measurement area   
					Maximum   
					measurement area   
					(Measuring   
					angle)   
					ꢀ° 0.2° 0.ꢀ° ꢀ° 0.2° 0.ꢀ° ꢀ° 0.2° 0.ꢀ° ꢀ° 0.2° 0.ꢀ° ꢀ° 0.2° 0.ꢀ°   
					ꢀ° 0.2° 0.ꢀ°   
					Without   
					5.00 ꢀ.00 0.50   
					350   
					7.78 ꢀ.56 0.78 ꢀ6.66 3.33 ꢀ.67   
					closeup lens   
					With closeup   
					lens   
					ꢀ.00 0.20 0.ꢀ0 ꢀ.39 0.28 0.ꢀ4   
					55.0   
					70.9   
					− 
					− 
					− 
					− 
					− 
					− 
					*Measurement distance is the distance from the front edge of the metal lens barrel or closeup lens ring.   
					2.   
					Rotate the measuring angle selector   
					according to size of object and   
					1°   
					0.2°   
					0.1°   
					measurement distance, and set the   
					aperture to 1°, 0.2°, or 0.1°.   
					Do not operate the measuring angle selector during   
					measurement. If measuring angle is switched   
					during measurement, you may fail to perform   
					measurement or to obtain correct measurement   
					value. To rotate the measuring angle selector, move   
					to a position where you can confirm the click. If you   
					stop the selector in a halfway position,   
					measurement may not be performed or wrong   
					measurement values may be obtained.   
					52   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					3.   
					4.   
					Rotate the diopter adjustment ring for   
					adjustment of diopter.   
					Make sure that aperture (black circle indicating   
					measurement area) looks clear when observing   
					object through finder. (See p.ꢀ5)   
					Rotate the focus adjustment ring for   
					objective lens for that purpose.   
					Make sure that object around aperture looks   
					clear when observing object through finder.   
					Only measuring area for measuring object must be   
					placed in aperture. If extra area, which is not   
					measuring object, is included in aperture, you   
					cannot perform correct measurement.   
					5.   
					6.   
					When the MENU or MEMORY screen   
					cd   
					m 
					2 
					Lv   
					x 
					36. 20   
					0 . 3968   
					0 . 4056   
					is displayed, press   
					key to   
					ESC   
					switch to the MEAS screen.   
					The MEAS (Measurement value) screen   
					appears.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					Press the measurement button.   
					When the measurement time is long, the   
					measurement progress bar is displayed on the   
					LCD until measurement is complete.   
					Display during measurement   
					(Single measurement/when   
					measurement time is long)   
					If the measurement time is set to options other   
					than [MANUAL], the measurement time will be   
					determined after the approximate luminance is   
					checked inside the measurement device. For   
					this reason, it may take several seconds until   
					the measurement time appears. The displayed   
					time shows approximately how long it will take   
					from the point of time display to the end of   
					measurement.   
					MEASURING   
					177s   
					ESC   
					: 
					STOP   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					No S t dNo off   
					r 
					m 
					n 
					n 
					(Single measurement/when   
					measurement time is short)   
					cd   
					m 
					2 
					Lv   
					x 
					36. 20   
					0 . 3968   
					0 . 4056   
					If the measurement time determined from the   
					approximate luminance is short, the remaining   
					time will not be displayed.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					53   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Display during measurementt   
					(Continuous measurement/when   
					measurement time is long)   
					CONT   
					When the measuring button is pressed and held   
					for two or more seconds, continuous   
					measurement is conducted.   
					When the measurement time is long, the   
					measurement progress bar appears with the   
					latest measurement value. The displayed time   
					shows the remaining time, same as for the   
					single measurement.   
					MEASURING   
					17s   
					ESC   
					: 
					STOP   
					cd   
					2 
					Lv   
					x 
					13 . 17   
					0 . 3966   
					0 . 3905   
					m 
					y 
					Obs   
					2˚ N   
					SPD SYNC LEN ND BL   
					No S t dNo off   
					When the measurement time is short, the   
					measurement progress bar does not appear, but   
					the measurement value is sequentially updated   
					and displayed.   
					r 
					m 
					n 
					n 
					(Continuous measurement/when   
					measurement time is short)   
					cd   
					2 
					Lv   
					x 
					36. 20   
					0 . 3968   
					0 . 4056   
					ESC   
					When   
					key is pressed during continuous   
					m 
					measurement, measurement is stopped. In this   
					case, the measurement that is ongoing when   
					ESC   
					key is pressed will be canceled, and the   
					last obtained measurement value will be   
					ESC   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					displayed. If   
					key is pressed in the middle   
					of the first measurement, the measurement   
					value will not be displayed.   
					ENTER   
					When   
					key is pressed while the   
					MEASURE   
					measurement value is displayed, the   
					ANGLE   
					SPEED   
					SYNC   
					LENS   
					1. 0   
					˚ 
					RMAL   
					N 
					N 
					O 
					measurement properties are displayed so that   
					the measurement conditions can be confirmed.   
					When the measuring button or a random key is   
					pressed, the MEAS screen appears again.   
					O 
					SYNC   
					STANDARD   
					FILTER NONE   
					IN ND   
					OFF   
					INTG   
					66. 667   
					m 
					s 
					54   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Saving the Measurement Value   
					This instrument can save ꢀ00 measurement values with designated numbers from 00 to 99.   
					Operation Procedure   
					2, 3   
					1 
					1. When the MEAS screen is displayed,   
					MEAS   
					cd   
					2 
					m 
					Lv   
					x 
					36 . 20   
					0 . 3968   
					0 . 4056   
					press   
					key to switch to the   
					MEMORY   
					y 
					MEMORY screen.   
					The MEMORY (Memory data) screen appears.   
					MEMORY 00   
					cd   
					2 
					Lv   
					x 
					34 . 22   
					0 . 3958   
					0 . 4060   
					m 
					y 
					The memory data number oo is displayd.   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					No S t dNo off   
					r 
					m 
					n 
					n 
					2. Press either   
					or   
					key to select the   
					MEAS   
					cd   
					2 
					Lv   
					x 
					36 . 20   
					0 . 3968   
					0 . 4056   
					m 
					m 
					number of memory data.   
					key for larger number.   
					y 
					MEMORY 02   
					cd   
					2 
					Lv   
					x 
					y 
					key for smaller number.   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					No S t dNo off   
					r 
					m 
					n 
					n 
					3. Press   
					key.   
					ENTER   
					MEAS   
					cd   
					2 
					Lv   
					36 . 20   
					0 . 3968   
					0 . 4056   
					m 
					m 
					x 
					y 
					Each measurement value is saved with the   
					selected number.   
					MEMORY 02   
					cd   
					2 
					ESC   
					If   
					key is pressed, saving is canceled and   
					Lv   
					x 
					y 
					36 . 20   
					0 . 3968   
					0 . 4056   
					the MEAS screen appears on the LCD.   
					Ob   
					2 
					s 
					˚ N   
					SPD SYNC LEN ND BL   
					No S t dNo off   
					r 
					m 
					n 
					n 
					55   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Use the following procedure to display the memory data properties (measurement   
					conditions):   
					Operation Procedure   
					5, 6   
					1 
					2, 3, 4   
					cd   
					m 
					MENU   
					screen is displayed.   
					1. Press   
					key when the MEAS   
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					BACKLIGHT   
					the backlight is turned on.   
					y 
					Ob   
					off via   
					key on the MEAS screen,   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					MEMORY   
					DELETE   
					[MEMORY] and then press   
					key.   
					ENTER   
					PROPERTIES   
					The MENU – MEMORY screen appears.   
					3. Press either   
					or   
					key to select   
					MEMORY   
					PROPERTIES   
					[PROPERTIES] and then press   
					MEMORY 00   
					ANGLE   
					SPEED   
					SYNC   
					LENS   
					1. 0   
					RMAL   
					SYNC   
					STANDARD   
					˚ 
					key.   
					ENTER   
					N 
					O 
					N 
					O 
					The MENU – MEMORY – PROPERTIES   
					(Memory data measurement condition   
					confirmation) screen appears.   
					FILTER NONE   
					IN ND   
					OFF   
					INTG   
					66. 667   
					ms   
					The memory data number 00 is displayed.   
					56   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					4. To display the memory data for   
					another number, press either   
					MEMORY   
					PROPERTIES   
					or   
					MEMORY 02   
					ANGLE   
					SPEED   
					SYNC   
					LENS   
					1. 0   
					RMAL   
					SYNC   
					STANDARD   
					˚ 
					key to change the memory data   
					number.   
					The properties of the selected memory data are   
					displayed, and the measurement conditions can   
					be confirmed.   
					N 
					O 
					N 
					O 
					FILTER NONE   
					IN ND   
					OFF   
					INTG   
					66. 667   
					ms   
					key for larger number.   
					If kept pressed, the value continuously changes.   
					key for smaller number.   
					If kept pressed, the value continuously changes.   
					5. Press   
					key.   
					ESC   
					The MENU – MEMORY screen appears again   
					on the LCD.   
					6. Press   
					key.   
					ESC   
					The MEAS screen appears again on the LCD.   
					57   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Deleting the Memory Data   
					Stored measurement values can be deleted in the following procedure.   
					Operation Procedure   
					5, 6, 7   
					1 
					2, 3, 4, 5   
					MENU   
					key when the MEAS   
					screen is displayed.   
					1.   
					Press   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					BACKLIGHT   
					off via   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					the backlight is turned on.   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2.   
					3.   
					Press either   
					[MEMORY] and then press   
					key.   
					or   
					key to select   
					MEMORY   
					DELETE   
					ENTER   
					PROPERTIES   
					The MENU – MEMORY screen appears.   
					Press either   
					[DELETE] and then press   
					key.   
					or   
					key to select   
					MEMORY   
					MEM, DELETE   
					ENTER   
					MEM, DATA N   
					O 
					, 00   
					Lv   
					x 
					y 
					34 . 22   
					0 . 3958   
					0 . 4060   
					The MENU – MEMORY – DELETE (Memory   
					data deletion) screen appears.   
					The memory data number 00 is displayed.   
					58   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					4.   
					Press either   
					number of memory data you want to   
					delete.   
					key for larger number. If kept pressed, the   
					value continuously changes. [ALL] is displayed   
					after No.99.   
					or   
					key to select the   
					MEMORY   
					MEM, DELETE   
					MEM, DATA N   
					O 
					, 02   
					Lv   
					x 
					36 . 20   
					0 . 3968   
					0 . 4056   
					y 
					key for smaller number. If kept pressed, the   
					value continuously changes. [ALL] is displayed   
					after No.00.   
					The deleted memory data can not be recovered.   
					Confirm the number before deleting the memory   
					data.   
					5-a-1. Deletion of saved data, one by one:   
					Select the number of the memory   
					data to be deleted, and press   
					MEMORY   
					MEM, DELETE   
					MEM, DATA NO , 02   
					Lv   
					x 
					y 
					key.   
					ENTER   
					The saved measurement value disappears.   
					ESC   
					By pressing   
					key, the deletion of the   
					memory data is canceled and the MENU -   
					MEMORY screen appears again on the LCD.   
					5-a-2. Repeat the steps 4 through 5-a-1.   
					as necessary.   
					5-a-3. Press   
					key.   
					ESC   
					The MENU – MEMORY screen appears   
					again on the LCD.   
					5-b-1. Collective deletion of all memory   
					data:   
					MEMORY   
					MEM, DELETE   
					MEM, DATA N   
					O 
					, ALL   
					Select [ALL] and press   
					key.   
					ENTER   
					Lv   
					x 
					y 
					36 . 20   
					0 . 3968   
					0 . 4056   
					The confirmation screen for deletion of all   
					memory data is displayed.   
					59   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					5-b-2. Press   
					key to select [OK] and   
					then press ENTER key.   
					MEMORY   
					MEM, DELETE   
					DELETE   
					DELETE   
					ALL DATA   
					All memory data are deleted, and the MENU -   
					? 
					MEMORY screen appears again on the LCD.   
					O 
					K 
					CANCEL   
					ESC   
					By selecting [CANCEL] and pressing   
					ENTER   
					key, or pressing   
					key, the deletion of   
					all memory data is canceled and the MENU -   
					MEMORY screen appears again on the LCD.   
					6. Press ESC key.   
					The MENU – MEMORY screen appears again   
					on the LCD.   
					7 
					. Press ESC key.   
					The MEAS screen appears again on the LCD.   
					60   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Registering Target Color   
					Target color   
					The target color serves as reference for measurement of deviation of measured color from   
					reference.   
					Target colors can be registered up to 20 from Ch 0ꢀ to Ch 20 completely within this   
					instrument.   
					To register the target color:   
					(ꢀ) Register the measurement value   
					(2) Select from the memory data   
					(3) Enter the numerical value   
					To change the previously set target color, select another target color. User calibration   
					correction coefficient will not be influenced despite of target color change.   
					The target color is commonly used among each color space of Lvxy, Lvu’v’, LvTΔuv, XYZ,   
					dominant wavelength/excitation purity and spectral graph.   
					Operation Procedure   
					10, 11   
					1 
					2, 3, 4, 5,   
					6, 7, 8, 9   
					MENU   
					key when the MEAS   
					screen is displayed.   
					1.   
					Press   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					BACKLIGHT   
					off via   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					the backlight is turned on.   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					6ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					2.   
					3.   
					Press either   
					[TARGET] and then press   
					key.   
					or   
					key to select   
					TARGET   
					SELECT   
					ENTER   
					SET   
					DELETE   
					The MENU – TARGET screen appears.   
					Press either   
					[SET] and then press   
					or   
					key to select   
					key.   
					TARGET   
					TARGET SET   
					TARGET Ch 01   
					ENTER   
					The MENU – TARGET – SET screen appears.   
					Lv   
					x 
					y 
					4. Change target color channels by   
					pressing   
					or   
					key, and select the   
					channel used to register a target color.   
					key for larger number. If kept pressed, the   
					number continuously increases.   
					key for smaller number. If kept pressed, the   
					number continuously decreases.   
					Once a target color is registered, the status   
					before registration cannot be restored. If the   
					target colors are changed by overwriting, confirm   
					the channel number before making the change.   
					5. Press   
					key.   
					ENTER   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					The screen used for input of the target color ID   
					appears on the LCD.   
					Lv   
					x 
					y 
					6. Enter the target color ID.   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					When the   
					key is pressed, the characters   
					s 
					ample   
					change among 0 to 9, a to z, A to Z, and space.   
					When the key is pressed, the characters   
					change among space, Z to A, z to a, and 9 to 0.   
					When key is pressed, the cursor moves to   
					the right by one position.   
					Lv   
					x 
					y 
					When   
					key is pressed, the cursor moves to   
					the left by one position.   
					62   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					7 
					. Press   
					key.   
					ENTER   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					The screen used to select a target color   
					s 
					ample   
					SOURCE MES   
					< 
					> 
					appears on the LCD.   
					ESC   
					By pressing   
					key, the input of the target   
					Lv   
					x 
					y 
					36   
					3968   
					4056   
					. 
					20   
					0 
					0 
					. 
					. 
					color is canceled and the MENU - TARGET   
					screen appears again on the LCD.   
					8. Press either   
					or   
					key to select   
					data to be a target color.   
					When the   
					key is pressed, values are   
					changed among, M00 to M99, EDT, and MES. If   
					kept pressed, the value continuously changes.   
					When the   
					key is pressed, values are   
					changed among MES, EDT, and M00 to M99. If   
					kept pressed, the value continuously changes.   
					9-a-1. Selection of a measurement value   
					as a target color:   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					ample   
					SOURCE MES   
					s 
					< 
					> 
					Select [MES] and press   
					key.   
					ENTER   
					Lv   
					x 
					36   
					3968   
					4056   
					. 
					20   
					0 
					0 
					. 
					. 
					y 
					When the target color is registered, the MENU   
					- TARGET screen appears again on the LCD.   
					ESC   
					By pressing   
					key, the registration of the   
					target color is canceled and the MENU -   
					TARGET screen appears again on the LCD.   
					9-b-1. Selection of a memory data as a   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					ample   
					SOURCE M00   
					target color:   
					s 
					< 
					> 
					Select from [M00~M99] and press   
					key.   
					Lv   
					x 
					y 
					35   
					3968   
					4060   
					. 
					93   
					ENTER   
					0 
					0 
					. 
					. 
					When the target color is registered, the MENU   
					- TARGET screen appears again on the LCD.   
					ESC   
					By pressing   
					key, the registration of the   
					target color is canceled and the MENU -   
					TARGET screen appears again on the LCD.   
					9-c-1. Manual input of a target color:   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					ample   
					SOURCE EDT   
					Select [EDT] and press   
					key.   
					ENTER   
					s 
					< 
					> 
					Lv   
					x 
					y 
					0 
					. 
					0000E   
					+0   
					The screen used to input a target color with   
					a numerical value appears on the LCD.   
					0 
					. 
					. 
					0000   
					0000   
					0 
					63   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					9-c-2. Enter the target color.   
					TARGET   
					TARGET SET   
					TARGET Ch 04   
					key for larger number.   
					s 
					ample   
					SOURCE EDT   
					< 
					> 
					If kept pressed, the value continuously   
					increases.   
					key for smaller number.   
					Lv   
					x 
					y 
					0 
					. 
					6200E   
					+1   
					0 
					. 
					. 
					3968   
					4000   
					0 
					If kept pressed, the value continuously   
					becomes small.   
					When   
					key is pressed, the cursor moves   
					to the right by one position.   
					When   
					key is pressed, the cursor moves   
					to the left by one position.   
					9-c-3. Press   
					key.   
					ENTER   
					When the target color is registered, the   
					MENU - TARGET screen appears again on   
					the LCD.   
					When observer angle is ꢀ0°, Y is displayed   
					instead of Lv and registered as Yꢀ0.   
					If the entered value does not establish a   
					proper combination with the value in the   
					color space mode, the cursor will move to   
					the first digit on the first line. Enter another   
					value again.   
					ESC   
					By pressing   
					key, the registration of the   
					target color is canceled and the MENU -   
					TARGET screen appears again on the LCD.   
					10. Press ESC key.   
					The MENU screen appears again on the LCD.   
					11. Press ESC key.   
					The MEAS screen appears again on the LCD.   
					64   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Selecting Target Color   
					Select a target color for color-difference measurement from the channels Ch0ꢀ to Ch20.   
					The new set color is also applied to the measurement value displayed on the current   
					MEAS screen.   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					key when the MEAS   
					screen is displayed.   
					MENU   
					cd   
					m 
					2 
					Lv   
					x 
					36 . 20   
					0 . 3968   
					0 . 4056   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					TARGET   
					SELECT   
					[TARGET] and then press   
					key.   
					ENTER   
					SET   
					DELETE   
					The MENU – TARGET screen appears.   
					3. Press either   
					or   
					[SELECT] and then press   
					key.   
					key to select   
					TARGET   
					TARGET SELECT   
					TARGET Ch01   
					ENTER   
					Lv   
					x 
					y 
					The MENU – TARGET – SELECT screen   
					appears.   
					65   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					4. Press either   
					or   
					key to switch   
					TARGET   
					TARGET SELECT   
					TARGET Ch04   
					the target color channel.   
					key for larger number.   
					s 
					ample   
					Lv   
					x 
					y 
					36   
					3968   
					4056   
					. 
					20   
					If kept pressed, the number continuously   
					increases.   
					0 
					0 
					. 
					. 
					key for smaller number.   
					If kept pressed, the number continuously   
					decreases.   
					5. Press   
					key.   
					ENTER   
					When the target color is set, the MENU -   
					TARGET screen appears again on the LCD.   
					ESC   
					By pressing   
					key, the selection of the   
					target color is canceled and the MENU -   
					TARGET screen appears again on the LCD.   
					6. Press   
					key.   
					ESC   
					The MENU screen appears again on the LCD.   
					cd   
					m 
					7 
					. Press ESC key.   
					2 
					Lv   
					x 
					36 . 20   
					0 . 3968   
					0 . 4056   
					The MEAS screen appears again on the LCD.   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					66   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Deleting Target Color   
					The registered target color can be deleted in the following procedure.   
					Operation Procedure   
					6, 7   
					1 
					2, 3, 4, 5   
					1. Press   
					key when the MEAS   
					screen is displayed.   
					MENU   
					cd   
					m 
					2 
					Lv   
					x 
					35. 93   
					0 . 3968   
					0 . 4060   
					The MENU screen appears.   
					When the backlight of the LCD has been turned   
					off via   
					the backlight is turned on.   
					BACKLIGHT   
					key on the MEAS screen,   
					y 
					Ob   
					s 
					SPD SYNC LEN ND BL   
					2 
					˚ N   
					r 
					m 
					No   
					n 
					S t dNo   
					n 
					off   
					MEAS   
					MEMORY   
					TARGET   
					OPTION   
					SETUP   
					2. Press either   
					or   
					key to select   
					TARGET   
					SELECT   
					[TARGET] and then press   
					key.   
					ENTER   
					SET   
					DELETE   
					The MENU – TARGET screen appears.   
					3. Press either   
					or   
					[DELETE] and then press   
					key.   
					key to select   
					TARGET   
					TARGET DELETE   
					TARGET Ch01   
					ENTER   
					Lv   
					x 
					y 
					The MENU – TARGET – DELETE (Target   
					color deletion) screen appears.   
					67   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					4. Press either   
					or   
					key and select   
					TARGET   
					TARGET DELETE   
					TARGET Ch04   
					the number to delete the desirable   
					s 
					ample   
					memory data or [ALL]. Press   
					key.   
					key for larger number. If kept pressed, the   
					value continuously changes. [ALL] is displayed   
					after Ch 20.   
					ENTER   
					Lv   
					x 
					y 
					36   
					3968   
					4056   
					. 
					20   
					0 
					0 
					. 
					. 
					key for smaller number. If kept pressed, the   
					value continuously changes. [ALL] is displayed   
					after Ch 00.   
					The confirmation screen to delete the target   
					color appears.   
					5.-a-1. Deletion of target colors one by   
					one:   
					TARGET   
					TARGET DELETE   
					TARGET Ch04   
					TARGET DELETE   
					Press   
					then press   
					key to select [OK] and   
					key.   
					TARGET Ch 04   
					ENTER   
					OK   
					CANCEL   
					The display of the registered target color   
					disappears.   
					By selecting [CANCEL] and pressing   
					ENTER   
					ESC   
					key, the   
					key, or pressing   
					deletion of the target color is canceled and   
					the MENU - TARGET screen appears   
					again on the LCD.   
					5.-a-2. To delete another target color   
					continuously, repeat the steps 4   
					through 5-a-1. as necessary.   
					5.-a-3. Press   
					key.   
					ESC   
					The MENU – TARGET screen appears   
					again on the LCD.   
					68   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					TARGET   
					TARGET DELETE   
					TARGET ChALL   
					TARGET DELETE   
					5.-b-1. Collective deletion of all target   
					colors:   
					TARGET Ch ALL   
					key to select [OK] and then   
					OK   
					CANCEL   
					press   
					key.   
					ENTER   
					When all target colors are deleted, the   
					MENU - TARGET screen appears again on   
					the LCD.   
					By selecting [CANCEL] and pressing   
					ENTER   
					ESC   
					key, the   
					key, or pressing   
					deletion of the target color is canceled and   
					the MENU - TARGET screen appears   
					again on the LCD.   
					ESC   
					By pressing   
					key, the selection of the   
					target color is canceled and the MENU -   
					TARGET screen appears again on the   
					LCD.   
					6. Press   
					key.   
					ESC   
					The MEAS screen appears again on the LCD.   
					7 
					. Press   
					key.   
					ESC   
					The MEAS screen appears again on the LCD.   
					69   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				
					70   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Communication   
					7ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Connecting to PC   
					This instrument can be used together with PC for mutual communication. Use the USB   
					cable (2m) CS-A32 supplied as standard accessory for this purpose.   
					The USB cable is allowed to plug/unplug while power is on, but it is recommended to   
					switch power off in this case.   
					Operation Procedure   
					1.   
					2.   
					Switch power OFF (O).   
					Connect the USB cable to the USB   
					connector of this instrument.   
					3.   
					Make sure that USB cable is all seated to   
					the USB connector.   
					Communication interface in this instrument conforms to USBꢀ.ꢀ.   
					Hold the USB cable plug in unplugging. Do not pull the cord.   
					Plug the USB cable to fit the connector entry point.   
					To connect this instrument to PC, install the corresponding USB driver software. It is   
					attached to the data management software CS-Sꢀ0w Professional in the standard   
					accessories. See installation guide of CS-Sꢀ0w Professional for details on installing the   
					USB driver software to PC.   
					72   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Remote Mode   
					Remote mode refers to sending command from PC to this instrument with both connected.   
					If this instrument is controlled with PC, “REMOTE MODE” appears on the LCD. While   
					this message is displayed, key operation of this instrument is not acceptable except   
					for following cases.   
					 If the measurement button is pressed, measurement starts to forward the data to   
					PC. (in case that measurement button is in valid mode by transferring command   
					from PC to this instrument. Use data management software below.)   
					ESC   
					 When   
					key is pressed, the remote mode is canceled.   
					To control this instrument with PC, use the data management software CS-Sꢀ0w   
					Professional in the standard accessories. See instruction manual of CS-Sꢀ0w Professional   
					for details on the specifications and usage.   
					If you want to use an independent program on PC to control this instrument, download   
					Communication Specifications from KONICA MINOLTA’s website at URL below for your   
					reference   
					
					(The above URL is subject to change without notice.)   
					(If the target page will not appear, please search the site by keywords, CS-2000 and   
					download.)   
					Key   
					ESC   
					Measurement button   
					LCD screen   
					73   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				
					74   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
			
				
					Explanation   
					75   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Measurement Principle   
					Light energy passes through the objective lens. The lights from the measurement area pass   
					through the hole in the center of the aperture mirror to the optical fiber, while the remaining   
					light is guided to the finder optics by the aperture mirror. As a result, the part equivalent to   
					the measurement area looks like a black circle when observed through the finder.   
					The light entering the optic fiber is reflected repeatedly so that it is mixed and becomes   
					virtually uniform. It then passes through the collimator lens to the plane diffraction grating.   
					After being dispersed by the grating, the light is focused by the condenser lens according   
					to wavelength. An array sensor is located at this focus point.   
					The amount of detected energy for each wavelength is then converted to a digital value by   
					an A/D converter, based on which the spectral radiant luminance and chromaticity are   
					calculated by the processing section of the CS-2000.   
					Sensor Section   
					The sensor section has a photo diode array consisting of 5ꢀ2 elements. The array is   
					always kept at constant temperature using a Peltier cooler, irrespective of the ambient   
					temperature. This can reduce dark current and improve S/N ratio, thus enabling   
					measurement of low luminance.   
					Objective lens   
					Array sensor   
					Condenser lens   
					Aperture mirror Collimator lens   
					Optical fiber   
					Plane diffraction grating   
					Shutter   
					Dark Measurement   
					Each measurement consists of “light measurement” and “dark measurement”.   
					“Light measurement” is performed with light from the object irradiating the sensor, while   
					“dark measurement” is performed with no light from the object irradiating the sensor, so   
					that detector dark current is measured.   
					When measurement starts, first “light measurement” is performed. When “light measurement”   
					is complete, the shutter will close automatically, then “dark measurement” starts immediately.   
					“Dark measurement” is performed with the same integration time used for “light measurement”.   
					The final measured data is obtained by subtracting the measured data obtained in “dark   
					measurement” from the one obtained in “light measurement”. This method eliminates   
					influences of dark current of the array sensor, resulting in improved measurement accuracy.   
					76   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					LvTΔuv   
					Following factors can be acquired as measurement value with LvTΔuv as color space of this instrument.   
					Lv   
					T 
					: Luminance   
					:Correlated color temperature   
					Δuv :Color difference from blackbody locus   
					While Lvs stands for luminance, T and Δuv for color in LvTΔuv.   
					<Relation between correlated color temperature T and color difference from blackbody locus Δuv>   
					Color temperature refers to the temperature of black body (perfect radiator) which has   
					equal chromaticity coordinates to certain light. However, color temperature only represents   
					colors on blackbody locus.   
					Correlated color temperature, slightly wider interpretation of color temperature, is very   
					useful to eliminate such problem. Here, correlated color temperature covers those which   
					are slightly outside the range of that of blackbody locus.   
					If a certain color positions on isotemperature line, the intersection point of isotemperature   
					line and blackbody locus is indicated as correlated color temperature for the color.   
					Isotemperature line means line on chromaticity coordinates which is a set of colors visually   
					close to color temperature on blackbody locus.   
					However, since all colors on a color-matching temperature line are represented with equal   
					correlated color temperature, it is not possible to describe color only with correlated color   
					temperature. To solve that, Δuv, deviation of correlated color temperature T from   
					blackbody locus, is to apply for that purpose.   
					If Δuv exists above blackbody locus, it is represented with “+”, and below, with “-”.   
					v 
					Blackbody locus   
					0.35   
					Correlated   
					temperature T   
					Δuv   
					0.30   
					u 
					Correlated temperature T and Δuv   
					77   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Dominant wavelength/Excitation purity   
					In the x, y chromaticity diagram shown below, the curve VScSR is the spectrum locus, and   
					point N is the white point.   
					Colors located in the region enclosed by the spectrum locus and the straight lines VN and   
					NR are referred to as spectral colors; colors located in the triangle NVR with the white point   
					N at the apex and the pure purple line VR as the base are referred to as nonspectral colors.   
					<Dominant wavelength and excitation purity (spectral colors)>   
					When the chromaticity point obtained by the measurement is C, the wavelength   
					corresponding to the intersection point S of the extension of NC with the spectrum locus   
					(curve VScSR) is referred to as the dominant wavelength and indicated by the symbol λd.   
					The ratio of the lengths of the straight lines NC and NS is referred to as the excitation   
					purity of color excitation C and indicated by the symbol pe.   
					<Complementary wavelength (non-spectral colors)>   
					When the chromaticity point obtained by measurement is C', the extension of NC' toward C'   
					does not intersect with the spectrum locus but only the pure-purple lines. In this case the   
					wavelength corresponding to the intersection point Sc of the extension of NC' toward N with the   
					spectrum locus is referred to as the complementary wavelength and indicated by the symbol λc.   
					When the intersection point of the extension of the line NC' with the line VR (pure-purple   
					locus) is designated by S', the ratio of the lengths of NC' to NS' is referred to as excitation   
					purity and indicated by the symbol p'v.   
					The following equations are formulated, if each point is designated as the following   
					coordinates: (xn, yn): chromaticity coordinate of point N; (xc, yc): chromaticity coordinate of   
					point C; (xλ, yλ): chromaticity coordinate of point S, (xc', yc): chromaticity coordinate of   
					point C', and (xp, yp): chromaticity coordinate of point:   
					y 
					Excitation purity (spectral colors)   
					Spectrum locus   
					Sc   
					xc – xn   
					x – xn   
					yc – yn   
					y – yn   
					pe =   
					= 
					0.8   
					0.6   
					0.4   
					0.2   
					0 
					550 nm   
					Excitation purity (non-spectral colors)   
					S 
					xc'– xn   
					xp – xn   
					yc' – yn   
					yp – yn   
					pe' =   
					= 
					500 nm   
					C 
					N 
					600 nm   
					White point   
					R 
					780 nm   
					C'   
					S'   
					Pure purple line   
					V 
					380 nm   
					0.2   
					0.4   
					0.6   
					x 
					Dominant wavelength on chromaticity diagram   
					78   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Measurement of Object Color   
					Use of the data management software CS-Sꢀ0w Professional in the standard accessories   
					enables simple measurement of the color of the object. Measured data is evaluated based   
					on luminance which has been stored as light source data in CS-Sꢀ0w. See instruction   
					manual of CS-Sꢀ0w for details.   
					 Set while calibration plate (optional accessory) and object on the same position with   
					the same angle from this instrument. Uniform illumination and measurement conditions   
					of white calibration plate and those of object. If not, measurement data may vary,   
					causing incorrect data.   
					 Keep illumination light source as stable as possible with fixed voltage power source   
					during measurement.   
					Measurement Procedure   
					Necessary Setting for Object Color Measurement   
					1.   
					Set one or more tungsten lumps   
					or equivalent as illumination   
					source toward white calibration   
					plate as in the right illustration.   
					 Set this instrument vertical to white   
					calibration plate.   
					Illumination Light Source   
					White Calibration   
					Plate   
					45°   
					45°   
					 Keep the angle between illumination   
					light source and white calibration plate   
					at 45°.   
					2.   
					Start up this instrument, PC, and   
					software CS-S10w.   
					Illumination Light Source   
					White Calibration   
					3.   
					4.   
					Set the document type to “object color” with CS-S10w.   
					Conduct white calibration with CS-S10w.   
					Measurement of Object   
					5.   
					6.   
					Set object on the same position and at the same angle as white   
					calibration plate.   
					Conduct measurement with CS-S10w.   
					See instruction manual of data management software CS-Sꢀ0w for details.   
					79   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Dimensions   
					(Unit: mm)   
					Filter screw   
					diameter   
					ø52   
					120.5   
					129   
					58.5   
					300   
					392   
					Measurement   
					distance reference   
					Mounting screw M5 Mounting screw M5   
					Depth 6.5mm Depth 6.5mm   
					95.7   
					229.5   
					ø15   
					6 
					Optical   
					axis   
					Positioning pin hole 1   
					ø5mm Depth 4 mm   
					Positioning pin hole 2   
					Depth 4mm   
					70   
					170   
					Tripod screw (3/8 inches screw)   
					Depth 10.5 mm   
					190.4   
					80   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Error Message   
					Error message appears on the LCD when this instrument does not operate normally. Below   
					table shows type of error message, its description and corrective action respectively.   
					Error Message   
					Cause (Description)   
					Corrective action   
					Data stored in ROM is   
					damaged.   
					ꢀ 
					2 
					MEMORY ERROR   
					• Turn on the power switch again.   
					• 
					If the symptoms are not remedied,   
					please contact the nearest   
					KONICA MINOLTA   
					authorized service facility.   
					Luminance of measuring   
					object is higher than the   
					available range.   
					• Use the ND filter and re-measure.   
					• Decrease the measuring area and   
					re-measure.   
					OVER   
					• If the symptoms are not remedied,   
					please contact the nearest   
					KONICA MINOLTA authorized   
					service facility.   
					Flicker of measuring object   
					is high.   
					• Set the flicker cycle in INT SYNC   
					mode, or input the flicker signal in   
					EXT SYNC mode.   
					Input signal in EXT SYNC   
					mode is not at CMOS level.   
					• Input a signal with 3.5 to 5.0 V at   
					high level; 0.0 to ꢀ.5 V at low level.   
					3 
					SYNC ERROR   
					Input signal in EXT SYNC   
					mode exceeds 200 Hz.   
					• Set the frequency value divided by   
					an integer in INT SYNC mode and   
					re-measure.   
					Input signal in EXT SYNC   
					mode is less than 20 Hz.   
					• Set the integral multiple value of   
					the input signal cycle in INT SYNC   
					mode and re-measure using the   
					MULTI INTEG-NORMAL mode or   
					MULTI INTEG-FAST mode.   
					• 
					Set the integral multiple value of the input   
					signal cycle in INT SYNC mode and re-   
					measure using the MANUAL mode.   
					Measurement was   
					performed when the   
					measuring angle selector   
					was in the wrong position,   
					or its position was changed   
					during measurement.   
					• Switch the measuring angle   
					selector and re-measure. Or, do   
					not operate the measuring angle   
					selector during measurement.   
					• If the symptoms are not remedied,   
					please contact the nearest   
					KONICA MINOLTA authorized   
					service facility.   
					4 
					VIEWING-ANGLE   
					SELECTOR ERROR   
					The ambient temperature for the • Decrease the ambient temperature   
					5 
					6 
					TEMPERATURE   
					ERROR   
					measurement device is too high,   
					and an internal temperature of   
					the sensor becomes abnormal.   
					to enable cooling to within the   
					specified temperature.   
					The cooling fan stops, or the • Decrease the ambient temperature   
					FAN ERROR   
					cooling device has an error.   
					to enable cooling to within the   
					specified temperature.   
					• Turn off the power switch once,   
					turn it on again after a while.   
					• 
					If the symptoms are not remedied, the   
					fan must be repaired. Please contact   
					the nearest KONICA MINOLTA   
					authorized service facility.   
					8ꢀ   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Error Check   
					Should any errors be found in this instrument, try the corrective actions shown in the   
					following table. If this does not help, it’s possible the instrument is broken. Please   
					contact the nearest KONICA MINOLTA authorized service facility with the error   
					number and version of your instrument. To confirm the instrument version, see p.23.   
					Error   
					No.   
					Reference   
					Page   
					Symptom   
					Item to Check   
					Corrective Action   
					ꢀ 
					No display on the LCD Has the AC adapter been properly Connect the AC adapter.   
					2ꢀ   
					2ꢀ   
					after power is turned   
					on.   
					plugged into the AC outlet?   
					Has the AC adapter been   
					Connect the AC adapter.   
					connected to this instrument?   
					Has a wrong AC adapter Be sure to use the AC adapter and   
					been connected?   
					power cord supplied as a standard   
					or optional accessory (AC-A3ꢀ2).   
					2ꢀ   
					Is AC power source   
					within rated scale?   
					Use within ꢀ0% of the nominal   
					voltage.   
					2ꢀ   
					8 
					2 
					Nothing is visible   
					through the finder.   
					Is the lens cap still attached Remove the lens cap.   
					to the objective lens?   
					Is the ND filter attached Use the ND filter when the   
					to the objective lens?   
					luminance of that being   
					measured is too high.   
					9, 45   
					Is the ND eyepiece filter Use the ND eyepiece filter when   
					attached to the finder?   
					luminance of that being   
					measured is too high.   
					9, 45   
					ꢀ4, 49   
					49   
					3 
					4 
					Nothing is displayed on Is the backlight set to   
					Press BACKLIGHT key to turn   
					ON the backlight.   
					the LCD.   
					OFF?   
					Has the backlight been In the menu operation, set the   
					set to OFF during   
					measurement?   
					backlight to ON during   
					measurement   
					Does not accept key   
					operation.   
					Has the remote mode   
					been set?   
					Press ESC key to cancel the   
					remote mode.   
					73   
					— 
					Have you pressed a   
					disabled key?   
					Press the correct key.   
					5 
					6 
					Measurement unavailable Has a screen other than Conduct measurement when the   
					even if the measurement MENU been displayed? MEAS screen is displayed.   
					button is pressed.   
					53   
					The entered value for   
					target color is different   
					from the one that will be   
					displayed after setting.   
					A ꢀ-digit difference may be   
					found due to calculation error.   
					6ꢀ   
					52   
					7 
					Measurement values   
					do not appear.   
					Is there data?   
					Conduct measurement.   
					Has the color space   
					mode become color   
					temperature?   
					Color temperature is displayed   
					as "_ _ _" if it is far from the   
					blackbody locus. Change the   
					display using the other color   
					space mode and check.   
					39   
					52   
					Did you interrupt   
					measurement?   
					Conduct measurement again.   
					82   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Error   
					No.   
					Reference   
					Page   
					Symptom   
					Item to Check   
					Corrective Action   
					8 
					Measurement values   
					are inconsistent.   
					Is the measuring object Conduct measurement with the   
					stable? measuring object stable.   
					Is the measuring object Repeatability of x, y worsens if   
					— 
					of low luminance?   
					the measuring object of low   
					luminance is measured.   
					It especially worsens when the   
					measurement angle is 0.2º or   
					0.ꢀº.   
					52, 30   
					It also worsens when the   
					measurement time is short.   
					Make the measurement time   
					longer.   
					Is the measurement   
					sync frequency   
					Set the appropriate   
					measurement sync frequency.   
					appropriate when   
					measuring the display?   
					Use MULTI INTEG-NORMAL   
					mode or MULTI INTEG-FAST   
					mode.   
					26, 30   
					Measure in EXT SYNC mode.   
					Have the ambient   
					temperature and /or   
					humidity changed   
					rapidly?   
					Conduct measurement under an   
					environment free from rapid   
					changes in ambient temperature   
					and humidity.   
					3 
					Did you start   
					measurement   
					immediately after   
					startup?   
					Allow the instrument to warm up   
					for 20 minutes or more when the   
					power switch is turned on.   
					23   
					9 
					Measurement values   
					appear incorrectly.   
					Is the objective lens   
					clean?   
					Wipe off the lens with a dry, soft   
					cloth or lens cleaning paper.   
					52   
					47   
					User calibration may not Check the values without user   
					be performed correctly. calibration (i.e. set the calibration   
					channel to 00 (NON) and check).   
					Is the calibration   
					channel correct?   
					Select the calibration channel   
					according to the light source of   
					luminance and chroma close to   
					the object.   
					47   
					Has the closeup lens   
					been attached?   
					Select the lens type setting   
					according to the attached   
					closeup lens.   
					43, 52   
					45   
					Has the ND filter been   
					attached?   
					Select the ND filter setting   
					according to the attached ND   
					filter.   
					Has the object been   
					focused?   
					Adjust the focus after adjusting   
					the diopter.   
					ꢀ3, ꢀ5   
					52   
					ꢀ0 The measurement   
					Is the measuring object When measuring an object of   
					stops halfway and does of high luminance?   
					not finish in the set   
					measurement time.   
					high luminance, the sensor may   
					be saturated by exceeding the   
					upper limit of the current   
					45   
					measurement setting.   
					Use the ND filter.   
					83   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Error   
					No.   
					Reference   
					Page   
					Symptom   
					Item to Check   
					Corrective Action   
					ꢀꢀ Actual measurement   
					time is different from   
					the displayed   
					The displayed measurement   
					time is the remaining time. The   
					actual measurement time may   
					be different from the displayed   
					time depending on the mode   
					setting of measurement time.   
					30   
					measurement time.   
					ꢀ2 Measurement value on Has the power source   
					Connect to a stable power   
					the LCD disappears.   
					been supplied securely? source and insert the AC adapter   
					plug securely.   
					2ꢀ   
					54   
					Did you interrupt the   
					measurement?   
					When starting continuous   
					measurement, press the   
					measurement button securely.   
					Do not press ESC key.   
					ꢀ3 During USB   
					communication:   
					Has the USB cable   
					been connected   
					Connect this instrument and the   
					PC securely.   
					72   
					— 
					Cannot download data securely?   
					output from this   
					instrument on the PC.   
					Cannot enter   
					Has the USB cable   
					been disconnected?   
					Replace the USB cable.   
					commands or data from   
					the PC to this   
					instrument.   
					Has remote mode been Send the connection command   
					canceled?   
					from the PC to this instrument   
					and switch to remote mode. Use   
					the data management software   
					CS-Sꢀ0w in the standard   
					accessories.   
					73   
					Has the program been   
					prepared correctly?   
					Refer to communication   
					specifications and check the   
					program. Use the data   
					— 
					— 
					management software CS-Sꢀ0w   
					in the standard accessories.   
					ꢀ4 Same error message   
					appears repeatedly.   
					Check the appropriate   
					If the symptoms are not   
					corrective action for the remedied, please contact the   
					error message.   
					nearest KONICA MINOLTA   
					authorized service facility.   
					84   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				 
			 
			
				
					Setting Initialization   
					Preset measurement conditions can be initialized to the factory default settings according   
					to the following procedure.   
					Factory default settings are shown below.   
					* Synchronizing method   
					: NO SYNC   
					* Measurement time   
					* Display format   
					: MULTI INTEG, ꢀs,   
					IN-ND: AUTO   
					* Observer   
					: 2°OBS   
					: Lvxy   
					: ****.**** [F]   
					* Color space mode   
					* Absolute value (ABS)/   
					Difference (DIFF) display   
					: Absolute value (ABS)   
					: NONE   
					* Lens type   
					: STANDARD   
					* ND filter type   
					* Backlight during measurement : ON   
					Operation Procedure   
					1, 2   
					1 
					1. When the power switch is OFF (O),   
					turn it to ON ( I ) while pressing   
					key,   
					key and   
					key   
					MEMORY   
					simultaneously.   
					The INITIALIZE SETTINGS (confirmation of   
					setting initialization) screen appears approx.   
					5 seconds after the initial screen on the LCD.   
					RESET   
					ALL SETTINGS ?   
					MEMORY   
					Continue pressing   
					key until the INITIALIZE SETTINGS screen   
					appears.   
					key,   
					key and   
					OK   
					CANCEL   
					2. Press   
					press   
					key to select [OK] and then   
					key.   
					RESET   
					ENTER   
					ALL SETTINGS ?   
					The preset measurement conditions are   
					initialized and the MEAS screen appears on the   
					LCD.   
					OK   
					CANCEL   
					85   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
			
				
					Switching Luminance Unit   
					You can select [cd/m2] or [ fL ] as the luminance unit.   
					Operation Procedure   
					1 
					1, 2   
					1. When the power switch is OFF (O),   
					turn it to ON ( I ) while pressing   
					key and   
					simultaneously.   
					key   
					MENU   
					fL   
					cd   
					m 
					The Luminance Unit (switching luminance   
					unit) screen appears approx. 5 seconds after   
					the initial screen on the LCD.   
					MENU   
					Continue pressing   
					key and   
					key until   
					the Luminance Unit screen appears.   
					2. Press either   
					or   
					key to select the   
					luminance unit and then press   
					key.   
					ENTER   
					The setting is entered and the MEAS screen   
					appears on the LCD.   
					86   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Main Specifications   
					Model   
					CS-2000   
					380 to 780 nm   
					0.9 nm/pixel   
					Wavelength range   
					Wavelength resolution   
					Display wavelength   
					bandwidth   
					ꢀ.0 nm   
					Wavelength precision   
					Spectral bandwidth   
					0.3 nm (median wavelength: 435.8 nm, 546.ꢀ nm, 643.8 nm Hg-Cd lamp)   
					5 nm or less (half bandwidth)   
					Measuring angle (selectable)   
					ꢀ°   
					0.2°   
					0.ꢀ°   
					Measurement luminance range   
					(Standard light source A)   
					0.003 to 5,000 cd/m2   
					0.075 to ꢀ25,000 cd/m2   
					0.3 to 500,000 cd/m2   
					ø5 mm   
					øꢀ mm   
					ø0.5 mm   
					ø0.ꢀ mm when using closeup lens)   
					Minimum measuring area   
					(øꢀ mm when using closeup lens)   
					ø0.2 mm when using closeup lens)   
					Minimum measuring distance   
					Minimum luminance display   
					350 mm (55 mm when using closeup lens)   
					0.00002 cd/m2   
					Minimum spectral   
					radiance display   
					ꢀ.0×ꢀ0-9 W/sr, m2, nm   
					Accuracy: Luminance   
					(Standard light source A)*ꢀ   
					2%   
					x,y : 0.003 (0.003 to 0.005 cd/m2)   
					x,y : 0.003 (0.075 to 0.ꢀ25 cd/m2)   
					x,y : 0.002 (0.ꢀ25 to ꢀ.25 cd/m2)   
					x,y : 0.003 (0.3 to 0.5 cd/m2)   
					Accuracy: Chromaticity   
					(Standard light source A)*ꢀ   
					x,y : 0.002 (0.005 to 0.05 cd/m2)   
					x,y : 0.002 (0.5 to 5 cd/m2)   
					x 
					y 
					: 
					: 
					0.00ꢀ5 (0.05 cd/m2 or more)   
					0.00ꢀ   
					x 
					y 
					: 
					: 
					0.00ꢀ5(ꢀ.25 cd/m2 or more)   
					0.00ꢀ   
					x 
					y 
					: 
					: 
					0.00ꢀ5(5 cd/m2 or more)   
					0.00ꢀ   
					0.4% (0.003 to 0.05 cd/m2)   
					0.3% (0.05 to 0.ꢀ cd/m2)   
					0.ꢀ5% (0.ꢀ to 5,000 cd/m2)   
					0.4% (0.075 to ꢀ.25 cd/m2)   
					0.3% (ꢀ.25 to 2.5 cd/m2)   
					0.ꢀ5% (2.5 to ꢀ25,000 cd/m2)   
					0.4% (0.3 to 5 cd/m2)   
					0.3% (5 to ꢀ0 cd/m2)   
					Repeatability:   
					Luminance (2 σ )   
					0.ꢀ5% (ꢀ0 to 500,000 cd/m2)   
					0.002 (0.003 to 0.005 cd/m2)   
					0.00ꢀ (0.005 to 0.ꢀ cd/m2)   
					0.0006 (0.ꢀ to 0.2 cd/m2)   
					0.0004 (0.2 to 5,000 cd/m2)   
					0.002 (0.075 to 0.ꢀ25 cd/m2)   
					0.00ꢀ (0.ꢀ25 to 2.5 cd/m2)   
					0.0006 (2.5 to 5 cd/m2)   
					0.002 (0.3 to 0.5 cd/m2)   
					0.00ꢀ (0.5 to ꢀ0 cd/m2)   
					0.0006 (ꢀ0 to 20 cd/m2)   
					0.0004 (20 to 500,000 cd/m2)   
					Repeatability:   
					Chromaticity (2 σ)   
					0.0004 (5 to ꢀ25,000 cd/m2)   
					Polarization error   
					Integration time   
					Measurement time   
					Color space mode   
					Interface   
					ꢀ°: 2% or less (400 to 780 nm); 0.ꢀ° and 0.2°: 3% or less (400 to 780 nm)   
					Fast: 0.005 to ꢀ6 sec.; Normal: 0.005 to ꢀ20 sec.   
					2 sec. min. (Manual mode) to 243 sec. max. (Normal mode)   
					Lv x y, Lvu'v', LvT ∆uv, XYZ, dominant wavelength, spectral graph   
					USB ꢀ.ꢀ   
					Operating temperature/   
					humidity range   
					5 to 35°C, relative humidity 80% or less with no condensation   
					0 to 35°C, relative humidity 80% or less with no condensation   
					Storage temperature/   
					humidity range   
					Size   
					Main unit: ꢀ58 (W) × 200 (H) x 300 (D) mm; Lens: ø70 x 95 mm   
					6.2 kg   
					Weight   
					*ꢀ: Average of ꢀ0 measurements in Normal mode at a temperature   
					of 23 2°C and a relative humidity of 65% or less.   
					87   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				 
			 
		 
		
			
				 
			 
			
				
					Model   
					CS-2000A   
					380 to 780 nm   
					0.9 nm/pixel   
					Wavelength range   
					Wavelength resolution   
					Display wavelength   
					bandwidth   
					ꢀ.0 nm   
					Wavelength precision   
					Spectral bandwidth   
					0.3 nm (median wavelength: 435.8 nm, 546.ꢀ nm, 643.8 nm Hg-Cd lamp)   
					5 nm or less (half bandwidth)   
					Measuring angle (selectable)   
					ꢀ°   
					0.2°   
					0.ꢀ°   
					Measurement luminance range   
					(Standard light source A)   
					0.0005 to 5,000 cd/m2   
					0.0ꢀ25 to ꢀ25,000 cd/m2   
					0.05 to 500,000 cd/m2   
					ø5 mm   
					øꢀ mm   
					ø0.5 mm   
					ø0.ꢀ mm when using closeup lens)   
					Minimum measuring area   
					(øꢀ mm when using closeup lens)   
					ø0.2 mm when using closeup lens)   
					Minimum measuring distance   
					Minimum luminance display   
					350 mm (55 mm when using closeup lens)   
					0.00002 cd/m2   
					Minimum spectral   
					radiance display   
					ꢀ.0×ꢀ0-9 W/sr, m2, nm   
					Accuracy: Luminance   
					(Standard light source A)*ꢀ   
					2%   
					x,y : 0.002 (0.00ꢀ to 0.05 cd/m2)   
					x,y : 0.002 (0.025 to ꢀ.25 cd/m2)   
					x,y : 0.002 (0.ꢀ to 5 cd/m2)   
					Accuracy: Chromaticity   
					(Standard light source A)*ꢀ   
					x 
					y 
					: 
					: 
					0.00ꢀ5 (0.05 cd/m2 or more)   
					0.00ꢀ (0.05 cd/m2 or more)   
					x 
					y 
					: 
					: 
					0.00ꢀ5 (ꢀ.25 cd/m2 or more)   
					0.00ꢀ (ꢀ.25 cd/m2 or more)   
					x 
					y 
					: 
					: 
					0.00ꢀ5 (5 cd/m2 or more)   
					0.00ꢀ (5 cd/m2 or more)   
					ꢀ.5 % (0.0005 to 0.00ꢀ cd/m2)   
					0.7 % (0.00ꢀ to 0.003 cd/m2)   
					0.25 % (0.003 to 0.05 cd/m2   
					0.ꢀ5 % (0.05 to 5,000 cd/m2)   
					ꢀ.5 % (0.0ꢀ25 to 0.025 cd/m2)   
					0.7 % (0.025 to 0.075 cd/m2   
					0.25 % (0.075 to ꢀ.25 cd/m2)   
					0.ꢀ5 % (ꢀ.25 to ꢀ25,000 cd/m2)   
					ꢀ.5 % (0.05 to 0.ꢀ cd/m2   
					) 
					Repeatability:   
					Luminance (2 σ )   
					0.7 % (0.ꢀ to 0.3 cd/m2   
					0.25 % (0.3 to 5 cd/m2   
					) 
					0.ꢀ5 % (5 to 500,000 cd/m2   
					) 
					x : 0.003 y : 0.0035   
					(0.00ꢀ to 0.003 cd/m2)   
					x : 0.00ꢀ y : 0.00ꢀ5   
					(0.003 to 0.ꢀ cd/m2)   
					x, y : 0.0006   
					x : 0.003 y : 0.0035   
					(0.025 to 0.075 cd/m2)   
					x : 0.00ꢀ y : 0.00ꢀ5   
					(0.075 to 2.5 cd/m2)   
					x, y : 0.0006   
					x : 0.003 y : 0.0035   
					(0.ꢀ to 0.3 cd/m2)   
					x : 0.00ꢀ y : 0.00ꢀ5   
					(0.3 to ꢀ0 cd/m2)   
					x, y : 0.0006   
					Repeatability:   
					Chromaticity (2 σ)   
					(0.ꢀ to 0.2 cd/m2)   
					x, y : 0.0004   
					(2.5 to 5 cd/m2)   
					(ꢀ0 to 20 cd/m2)   
					x, y : 0.0004   
					x, y : 0.0004   
					(5 to ꢀ25,000 cd/m2)   
					(0.2 to 5,000 cd/m2)   
					(20 to 500,000 cd/m2)   
					Polarization error   
					Integration time   
					Measurement time   
					Color space mode   
					Interface   
					ꢀ°: 2% or less (400 to 780 nm); 0.ꢀ° and 0.2°: 3% or less (400 to 780 nm)   
					Fast: 0.005 to ꢀ6 sec.; Normal: 0.005 to ꢀ20 sec.   
					Approx. 247 sec. max.   
					Lv x y, Lvu'v', LvT∆uv, XYZ, dominant wavelength, spectral graph   
					USB ꢀ.ꢀ   
					Operating temperature/   
					humidity range   
					5 to 30°C, relative humidity 80% or less with no condensation   
					0 to 35°C, relative humidity 80% or less with no condensation   
					Storage temperature/   
					humidity range   
					Size   
					Main unit: ꢀ58 (W) × 200 (H) x 300 (D) mm; Lens: ø70 x 95 mm   
					6.2 kg   
					Weight   
					*ꢀ: Average of ꢀ0 measurements in Normal mode at a temperature   
					of 23 2°C and a relative humidity of 65% or less.   
					88   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 
			 
		 
		
			
				