Konica Minolta Graphics Tablet VI 9i User Manual

NON-CONTACT 3D DIGITIZER  
VIVID 9i/VI-9i  
Instruction Manual (HARDWARE)  
NOTE  
The VI-9i is model name for Europe and the VIVID 9i is model name for other countries.  
Please note that the VIVID 9i model name is intended only for reference with this manual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety Precautions  
When using this hardware, the following points must be strictly observed to ensure correct and safe use.  
After you have read this manual, keep it in a safe place so that it can be referred to easily whenever it is needed.  
Failure to adhere to the following points may result  
in death or serious injury.  
WARNING  
Do not use the VIVID 9i in places where flam-  
mable or combustible gases (gasoline etc.) are  
present. Doing so may cause a fire.  
Do not disassemble or modify the VIVID 9i.  
Doing so may cause a fire or electric shock.  
Always use the AC power cord supplied as a  
standard accessory with the VIVID 9i, and con-  
nect it to an AC outlet (100-240 V , 50-  
60 Hz). Failure to do so may damage the VIVID  
9i, causing a fire or electric shock.  
Do not remove the cover as doing so may cause  
electric shocks.  
Do not bend, twist or pull the AC power cord ex-  
cessively. In addition, do not place heavy items  
on the AC power cord, or damage or modify it in  
any way. Doing so may cause damage to the AC  
power cord, resulting in fire or electric shock.  
Take special care not to allow liquid or metal ob-  
jects to enter the VIVID 9i. Doing so may cause  
a fire or electric shock. Should liquid or metal  
objects enter the VIVID 9i, turn the power OFF  
immediately, disconnect the AC power cord  
from the AC outlet, and contact the nearest  
KONICA MINOLTA SENSING-authorized  
service facility.  
If the VIVID 9i will not be used for a long time,  
disconnect the AC power cord from the AC out-  
let. Accumulated dirt or water on the prongs of  
the AC power cord’s plug may cause a fire. If  
there is any dirt or water on the prongs of the AC  
power cord’s plug, remove it.  
The VIVID 9i should not be operated if dirt or  
dust has entered through the vent holes. Doing  
so may result in a fire. For periodic inspection,  
contact the nearest KONICA MINOLTA SENS-  
ING-authorized service facility.  
When disconnecting the AC power cord’s plug,  
always hold the plug and pull it to remove it.  
Never pull the AC power cord itself. Doing so  
may damage the AC power cord, causing a fire  
or electric shock. In addition, do not insert or  
disconnect the AC power cord’s plug with wet  
hands. Doing so may cause electric shock.  
Never stare into the laser emitting  
window.  
The VIVID 9i should not be operated if it is  
damaged, or smoke or odd smells are detected.  
Doing so may result in a fire. In such situations,  
turn the power OFF immediately, disconnect the  
AC power cord from the AC outlet, and contact  
the nearest KONICA MINOLTA SENSING-au-  
thorized service facility.  
Do not place a lens, mirror or optical  
element in the passage of the laser  
beam. Doing so may converge the la-  
ser beam, resulting in damage to your  
eyes, burns or fire. To prevent the  
above accidents, make sure that a wall  
or similar which can block the laser  
beam is located behind the object.  
Falling to adhere to the following points may result in  
injury or damage to the instrument or other property.  
CAUTION  
Be sure to connect the AC power cord’s plug to  
an AC outlet that has a grounding terminal.  
Do not place the instrument on an unstable or  
sloping surface. Doing so may result in its drop-  
ping or overturning, causing injury. Take care  
not to drop the instrument when carrying it.  
Make sure that the AC outlet is located near the  
VIVID 9i and that the AC power cord’s plug can  
be easily connected and disconnected.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Laser Caution and Identification Label  
Laser Caution and Identification Label  
CAUTION  
LASER RADIATION  
DO NOT STARE INTO BEAM  
LASER STRAHLUNG  
NICHT IN DEN STRAHL SEHEN  
MAX 30mW 690nm  
CLASS  
2
LASER PRODUCT  
Complied with IEC Publication 60825-1:1993, Amendment-2:2001  
AVOID  
EXPOSURE  
Laser radiation is emitted  
from this aperture.  
CLASS 1 LASER PRODUCT  
Complies with 21 CFR Chapter 1. Subchopter J.  
KONICA MINOLTA SENSING, INC. 3-91, Daisennishimachi, Sakai, Osaka 590-8551, Japan  
manufactured  
Warning Label  
WARNING  
WARNUNG  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notes On Use  
• The VIVID 9i is designed for indoor use only, and should never be used outside.  
• This digitized has been calibrated at 20°C. We recommend that you use it environments where ambient tem-  
perature is 20°C (68°F).  
• Use an AC power source which is within ±10% of the rated voltage.  
• The VIVID 9i should be used within a temperature range of 10 to 40°C at a relative humidity of 65% or less.  
• Do not use the VIVID 9i in direct sunlight or near sources of heat such as stoves. Doing so will cause the  
temperature of the VIVID 9i to rise considerably higher than room temperature and may result in malfunc-  
tions. Use the VIVID 9i in a well ventilated area and make sure that the ventilation holes of the VIVID 9i are  
not blocked.  
• Do not use the VIVID 9i in extremely dusty or humid areas. Doing so may result in malfunctions.  
• Do not subject the VIVID 9i to strong vibration or shocks. Doing so may result in malfunctions.  
• When using the VIVID 9i on upper floors of buildings or near high-traffic streets, the instrument and/or  
measurement subject may vibrate, adversely affecting measurement results.  
• Do not allow the digitizer to fall on its side, as it may sustain damaged if knocked over.  
• Do not disconnect cords and cables with the POWER switch of the VIVID 9i set to ON “I”. Doing so may  
result in malfunctions.  
• The VIVID 9i has been tested and found to comply with the limits for a Class A digital device, pursuant to  
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter-  
ference when the equipment is operated in a commercial environment. This equipment generates, uses, and  
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual,  
may cause harmful interference to radio communications. Operation of this equipment in a residential area is  
likely to cause harmful interference in which case the user will be required to correct the interference at his  
own expense.  
• The VIVID 9i is a class II instrument specified in IEC Publication 825. Use it according to the instructions  
given in this manual.  
• Do not use the VIVID 9i at altitudes of higher than 2000 m.  
• Before using this unit for the first time, and before reusing it following transport, check that the lens is se-  
curely in place. If the lens is loose, tighten it in accordance with the tightening instructions included in the  
lens replacement procedure.  
Care On Storage  
• The VIVID 9i should be stored in areas with temperatures of between 0 and +40°C. Do not store it in areas  
subject to high temperature or high humidity or where sudden changes in temperature or condensation are  
likely to occur. We recommend storing the VIVID 9i around room temperature (20 ± 3°C) with a desiccant  
(silica gel etc.).  
• Do not leave the VIVID 9i inside a closed car or in the trunk of a car. Under direct sunlight, the increase in  
temperature can be extreme and may result in malfunctions.  
• The VIVID 9i should not be stored in areas where there is an excessive amount of dust, cigarette smoke or  
chemical gas. Failure to adhere to this may result in performance degradation or break-down.  
• When shipping the product, use the original packing materials in which the product was shipped. The mate-  
rials will provide protection against vibrations and impact, and also provide some protection against sudden  
changes in temperature.  
• Lenses that are not in use should be capped and kept in the lens case.  
• Before storing this equipment, be sure to attach the receiving lens cap and the laser barrier.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notes On Cleaning  
• If the VIVID 9i needs cleaning, wipe with a soft dry cloth. Never use solvents such as thinner or benzene.  
• If the lens or laser emitting window is soiled with sand or dust, blow off the dirt using a blower, and wipe  
them gently with a piece of cleaning paper dampened with cleaning agent.  
• In cases of malfunction, do not disassemble the VIVID 9i or attempt to repair it yourself. Contact the nearest  
KONICA MINOLTA SENSING-authorized service facility.  
Notes On Maintenance  
To maintain the accuracy of the high-precision Galvano mirror, that plays the most important role in measure-  
ment accuracy, a drying agent is used inside the VIVID 9i. Normally, the dehumidification effect of the drying  
agent lasts more than one year. However, its life varies with the operating and storage environment.  
To maintain the designed performance of the VIVID 9i, you are recommended to replace the drying agent once  
a year. Contact the nearest KONICA MINOLTA SENSING-authorized service facility.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Safety Precautions........................................................................................................................................................................ 1  
Laser Caution and Identification Label........................................................................................................................................ 2  
Warning Label.............................................................................................................................................................................. 2  
Notes On Use ............................................................................................................................................................................... 3  
Care On Storage........................................................................................................................................................................... 3  
Notes On Cleaning....................................................................................................................................................................... 4  
Notes On Maintenance................................................................................................................................................................. 4  
Chapter 1 Before Using the Instrument  
Package Contents (Standard Accessories) ........................................................................................................................................... 8  
Optional Accessories............................................................................................................................................................................. 10  
System Configuration........................................................................................................................................................................... 11  
Names and Functions of Parts............................................................................................................................................................. 12  
Main Body.................................................................................................................................................................................. 12  
Operation Panel.......................................................................................................................................................................... 13  
Chapter 2 Preparation  
Connecting the AC Power Cord.......................................................................................................................................................... 16  
Connecting the VIVID 9i to a Computer............................................................................................................................................ 17  
Setting the SCSI ID No.............................................................................................................................................................. 18  
Setting the Terminator ............................................................................................................................................................... 19  
Starting and Quitting ........................................................................................................................................................................... 21  
Starting the VIVID 9i................................................................................................................................................................. 21  
Quitting the VIVID 9i................................................................................................................................................................ 23  
Mounting to the Tripod........................................................................................................................................................................ 24  
Chapter 3 Operation  
Replacing the Lens and Field Calibration.......................................................................................................................................... 28  
Relationship between Measurement Distance and Object......................................................................................................... 28  
Replacing Procedure .................................................................................................................................................................. 29  
Field Calibration................................................................................................................................................................................... 32  
Field Calibration System............................................................................................................................................................ 32  
Executing Field Calibration ....................................................................................................................................................... 34  
Adjusting the White Balance............................................................................................................................................................... 42  
Displaying the Status Information ...................................................................................................................................................... 44  
Adjusting the Laser Power and CCD Gain Manually ...................................................................................................................... 45  
Setting the Scan Mode.......................................................................................................................................................................... 46  
Chapter 4 Appendix  
Error Messages ..................................................................................................................................................................................... 48  
Explanation of Measuring Principle................................................................................................................................................... 49  
Measuring Principle................................................................................................................................................................... 49  
High-Speed Image Processing Circuit ....................................................................................................................................... 50  
Time center of gravity and Space center of gravity................................................................................................................... 50  
Dimension Diagram.............................................................................................................................................................................. 51  
Specifications......................................................................................................................................................................................... 52  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1  
Before Using the Instrument  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Package Contents (Standard Accessories)  
Check that the following standard accessories are present.  
VIVID 9i Main Body  
Polygon Editing Tool  
(with USB protect key)  
White Balance Cap VI-A10  
AC Power Cord  
SCSI Cable VI-A20  
Laser Barrier  
(Half-pitch, D-sub, 50-pin male plug  
– 50-pin male plug)  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Package Contents (Standard Accessories)  
Lens  
Field Calibration System  
(3 types, TELE, MIDDLE, WIDE/  
with Lens Cap)  
(Field Calibration System, Frame for Calibration  
System, Hex wrench (M5), 4 screws)  
Tripod Attachment Set  
(Tripod attachment, Hex wrench (M3), 4  
screws)  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Accessories  
When you need to purchase the following optional items, contact a KONICA MINOLTA SENSING-author-  
ized service facility.  
Rotary Stage Set  
Tripod/Panhead VI-B25  
(Pan head, Tripod, Ball Adapter)  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
System Configuration  
Standard Accessories  
Optional Accessories  
(A commercially available computer can be used)  
White Balance Cap  
VI-A10  
AC Power Cord  
Polygon Editing Tool  
SCSI Cable VI-A20  
VIVID 9i Main Body  
Computer  
Tripod Attachment Set  
Laser Barrier  
3 types of lens  
Tripod/Panhead  
VI-B25  
Field Calibration System  
Rotary Stage Set  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Names and Functions of Parts  
Main Body  
Handle  
Light-Receiving Lens  
Note  
Please keep the receiving  
lens cap and the laser barri-  
er place in place when you  
are not carrying out scan-  
ning. Dust and scratches on  
the laser and lens can ad-  
versely affect scanning pre-  
cision.  
Power Switch  
Laser Emitting Window  
WARNING  
AC Power  
Connector  
A laser beam is emit-  
ted from this window.  
SCSI Port  
Do not stare into this  
window.  
Rubber supports  
Use when necessary to lean the device  
on its side for installing the tripod attach-  
ment, etc.  
Viewfinder (5.7" LCD)  
Displays the lens image and various set-  
up menus.  
Note  
In rare cases some dot breakup  
may be visible on the LCD, but  
this does not indicate a problem  
with the scan data itself.  
Operation Panel  
Used for adjusting the focus and making  
various settings.  
Rubber supports  
Support the device.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Names and Functions of Parts  
Operation Panel  
2
FOCUS  
AUTO  
1
MENU  
CANCEL  
SELECT  
ENTER  
3
4
5
6
7
AUTO  
1 FOCUS  
..........................Focuses the image automatically.  
2
3
(Near) /  
(Far)...........Focuses the image manually.  
indicates “far” and  
indicates “near”.  
.........................................Press once to display the menu on the viewfinder. Press again to close the  
menu and return to the monitor image.  
CANCEL  
4
5
6
........................................Cancels the current operation and returns to the preceding level.  
...................................Selects an item at the viewfinder.  
ENTER  
........................................Sets the current selection, and moves to the next level (if any).  
SELECT  
7
........................................Moves to the next level for the item currently selected in the viewfinder.  
Memo  
If you wish to carry out a local operation while the unit is in REMOTE mode (while REMOTE is displayed in the view-  
finder): Press any key to cancel REMOTE mode, and then press the appropriate keys for the operation you wish to carry  
out.  
(The MENU view may not be displayed even if  
is pressed depending on the view currently displayed.)  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Names and Functions of Parts  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2  
Preparation  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the AC Power Cord  
To ensure correct connection of the AC power cord, read the points given in WARNING and CAUTION care-  
fully.  
WARNING  
Always use the AC power cord supplied as a standard accessory with the VIVID 9i, and connect it to  
an AC outlet (100-240 V , 50-60 Hz). Failure to do so may damage the VIVID 9i, causing a fire or  
electric shock.  
Do not bend, twist or pull the AC power cord excessively. Do not place heavy items on it, scratch or  
modify it. Doing so may damage it, resulting in fire or electric shock.  
If you are not going to use the VIVID 9i for a long time, disconnect the AC power cord from the AC  
outlet. If dirt or water accumulates on the prongs of the AC power cord’s plug, it may cause a fire. If  
there is any dirt or water on the prongs, it must be removed.  
When disconnecting the AC power cord’s plug, always hold the plug and pull it to remove it. Never  
pull the AC power cord itself as it may be damaged, resulting in fire or electric shock. Also do not insert  
or disconnect the AC power cord’s plug with wet hands. Doing so may cause electric shock.  
CAUTION  
Plug the power cord into the outlet on the tripod or into a standard outlet. If using a standard outlet, be  
sure that the ground terminal on the plug is correctly grounded. Improper connection may result in fire  
or electric shock.  
Make sure that the AC outlet is located near the VIVID 9i and the AC power cord’s plug can be easily  
connected and disconnected.  
[Connecting Procedure]  
Set the power switch of the VIVID 9i to OFF  
(“ ”).  
1
POWER  
Note  
SCSI  
AC IN  
If the AC power cord is connected to an AC outlet with the  
power switch set to ON (“ ”), damage to the VIVID 9i or  
computer may result.  
Before connecting the AC power cord, always make sure  
that the power switch is set to OFF.  
Plug the AC power cord to the AC power con-  
nector on the VIVID 9i.  
2
Plug the other end of the AC power cord to an  
AC outlet.  
3
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the VIVID 9i to a Computer  
To operate the VIVID 9i from a computer using the Polygon Editing Tool, the VIVID 9i must be connected to  
the computer with an standard SCSI cable VI-A20.  
WARNING  
The computer must be operated correctly and safely according to its instruction manual.  
Memo  
• The standard SCSI cable has a 50-pin male plug (half-pitch, D-Sub) on both its ends.  
• It is also possible to use a commercially available USB adapter. For details, please contact a KONICA MINOLTA  
SENSING authorized service facility.  
Ref.  
For a description of the standard accessories, refer to page 8.  
[Connecting Procedure]  
Set the power switch of both VIVID 9i and  
computer to OFF (“ ”).  
1
POWER  
AC IN  
Note  
SCSI  
If the AC power cord is connected to an AC outlet with the  
power switch set to ON (“I”), damage to the VIVID 9i or  
computer may result.  
Always be sure that the digitizer and the computer are both  
OFF before making connections.  
Plug the SCSI cable to a SCSI port of the VIVID  
9i.  
2
Plug the other end of the SCSI cable to the  
SCSI connector of the computer.  
Connecting procedure is now complete  
3
Note  
Before turning ON the power to the computer, always turn  
ON the power to the VIVID 9i and wait until it is ready for  
operation (see page 21).  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the VIVID 9i to a Computer  
Setting the SCSI ID No.  
In order for the computer to recognize that the VIVID 9i is connected to the computer via the SCSI interface,  
a SCSI ID No. (0 to 6) must be set for the VIVID 9i.  
Memo  
SCSI ID No. 5 has been set as the default setting before shipment.  
Note  
If other devices are connected to the computer via the SCSI interface, make sure that the SCSI ID No. of the VIVID 9i  
differs from those set for the devices.  
[Setting Procedure]  
Press  
.
STANDARD 3 HQ ON NF ON  
COLOR OFF  
1
2
The MENU view will appear.  
W.B. DEFAULT  
SCSI 5 ENABLED  
Press  
SI SETTINGS”.  
to locate the arrow cursor to “SC-  
WHITE BALANCE  
INITIAL MODE  
INITIAL CONDITION  
SCSI SETTINGS  
INFO  
SELECT  
ENTER  
Press  
(or  
).  
3
4
The SCSI SETTINGS view will appear.  
SCSI SETTINGS  
SCSI ID  
5
TERMINATE  
ENABLED  
Press  
to locate the arrow cursor to “SC-  
SI ID”.  
SELECT  
ENTER  
Press  
(or  
).  
5
6
SCSI SETTINGS  
SCSI ID  
The currently selected SCSI ID No. will turn to pink.  
6
TERMINATE  
ENABLED  
Press  
to select the desired SCSI ID No.  
<Settings>  
0 to 6: SCSI ID No.  
ENTER  
Press  
.
7
Setting is now complete.  
Note  
To cancel the setting and return the SCSI SETTINGS  
CANCEL  
screen to its original state, press  
.
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the VIVID 9i to a Computer  
Setting the Terminator  
The VIVID 9i has a built-in SCSI terminator. The terminator is required if the VIVID 9i is the last device of  
those connected in series to the SCSI interface (i.e., if only one of the SCSI ports on the VIVID 9i is used).  
If the terminator is not set correctly, the VIVID 9i or devices connected to the SCSI interface may mal-  
function.  
Memo  
“ENABLED” has been selected as the default setting before shipment.  
Note  
“DISABLE” must be selected if the VIVID 9i is connected via the SCSI interface but is not the last device of those  
connected in series, and “ENABLE” must be selected if it is not connected via the SCSI interface.  
[Setting Procedure]  
Press  
.
STANDARD 3 HQ ON NF ON  
COLOR OFF  
1
2
The MENU view will appear.  
W.B. DEFAULT  
SCSI 5 ENABLED  
Press  
SI SETTINGS”.  
to locate the arrow cursor to “SC-  
WHITE BALANCE  
INITIAL MODE  
INITIAL CONDITION  
SCSI SETTINGS  
INFO  
SELECT  
ENTER  
Press  
(or  
).  
3
4
SCSI SETTINGS  
SCSI ID  
The SCSI SETTINGS view will appear.  
5
TERMINATE  
ENABLED  
Press  
to locate the arrow cursor to  
“TERMINATE”.  
SELECT  
ENTER  
Press  
(or  
).  
5
6
SCSI SETTINGS  
SCSI ID  
The currently selected option (ENABLED or DISA-  
BLED) will turn to pink.  
5
TERMINATE  
ENABLED  
Press  
to select “ENABLED” or “DISA-  
BLED”.  
<Settings>  
ENABLED: Select this option if the terminator is to be  
used.  
DISABLED: Select this option if the terminator is not to  
be used.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the VIVID 9i to a Computer  
Press  
Setting is now complete.  
.
7
Note  
CANCEL  
To cancel the setting and return the SCSI SETTINGS screen to its original state, press  
.
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Starting and Quitting  
Starting the VIVID 9i  
[Operating Procedure]  
Set the power switch to ON (“ ”).  
Power will be supplied to the VIVID 9i and initial set-  
1
POWER  
AC IN  
up is performed. After several dozen seconds,  
“PLEASE OPEN LASER BARRIER AND PRESS  
ANY KEY” will be displayed on the viewfinder.  
SCSI  
Note  
When turning power ON again after it has been turned  
OFF, wait at least 5 seconds before doing so. Failure to ob-  
serve this may result in malfunction or breakdown of the  
VIVID 9i.  
Memo  
To ensure precise measurement, allow the unit to warm up  
for at least 10 minutes.  
Remove the lens cap from the receiving lens.  
2
Remove the laser barrier.  
3
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Starting and Quitting  
Press any key.  
The MONITOR view will appear, indicating that the  
VIVID 9i is now ready for operation.  
EXTEND  
4
720MM  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Starting and Quitting  
Quitting the VIVID 9i  
[Operating Procedure]  
Set the power switch to OFF (“ ”).  
Power will be turned OFF.  
1
POWER  
AC IN  
SCSI  
Attach the lens barrier.  
2
WARNING  
To ensure safety, please keep the lens bar-  
rier in place at all times when the unit is not  
in use.  
Fit the lens cap onto the receiving lens.  
3
Note  
Please keep the receiving lens cap and the laser barrier  
place in place when you are not carrying out scanning.  
Dust and scratches on the laser and lens can adversely af-  
fect scanning precision.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mounting to the Tripod  
When using the instrument with the tripod, first attach the tripod attachment to the instrument, and then attach  
the instrument to the panhead/tripod.  
CAUTION  
• Make sure that the tripod attachment, panhead and tripod are secured to the instrument and each fas-  
tener on the tripod is firmly tightened. Failure to observe this point may result in the tripod overturning  
or the instrument dropping.  
• When placing the tripod on a floor, examine the floor conditions and flatness to make sure that the tri-  
pod is stably placed.  
• After measurement is completed, the instrument must be removed from the tripod before storing them.  
If the instrument is stored with the tripod attached, the tripod may overturn or the instrument may drop,  
resulting in accidents.  
1. Mounting the Tripod Attachment to the Instrument  
Before mounting the instrument to the tripod, mount the tripod attachment to the instrument as explained be-  
low.  
Memo  
Four screw holes for the tripod attachment are provided on base of the instrument.  
[Operating Procedure]  
After checking to ensure that you have turned off the power to the both the digitizer and  
computer, disconnect the digitizer’s power cord and the SCSI cable.  
1
Note  
Note that disconnection of cables while power is ON may result in damage to the digitizer and/or the computer.  
Be sure to turn off power at both sizes before disconnecting cables.  
Slowly lean the digitizer over on a flat and sta-  
ble surface, so that it is lying on the rubber  
supports on its rear side.  
2
Tripod  
Attachment  
Place the tripod attachment on the base of the  
instrument, and secure it with the four screws  
(supplied with the tripod attachment).  
3
Note  
Align the tripod attachment and digitizer so that the pro-  
jecting area of the attachment fits into the receding area on  
the base of the digitizer.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mounting to the Tripod  
2. Mounting the Instrument to the Panhead  
After you have attached the tripod attachment to the instrument, they need to be mounted to the recommended  
panhead as explained below.  
First, remove the accessory plate from the panhead, attach it to the instrument, then attach both to the panhead.  
Memo  
Three screw holes are provided on the back of the tripod attachment, to allow mounting the tripod attachment to the pan-  
head.  
[Operating Procedure]  
Remove the accessory plate from the panhead.  
1
Place the top of the accessory plate on the tri-  
2
pod attachment (attached to the instrument),  
and secure them with two screws (U3/8, sup-  
plied with the accessory plate).  
Accessory Plate  
U3/8 Screw  
Make sure that the tripod/panhead is secured in place.  
3
4
Slide the accessory plate (attached to the tripod attachment on the instrument) into the  
panhead.  
Memo  
Slide and attach the plate from front of the panhead.  
Tighten the slide fixing knob to firmly secure the accessory plate to the panhead.  
5
Memo  
Refer to the tripod/panhead instruction manual for details on using the accessory plate locking knob.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mounting to the Tripod  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3  
Operation  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replacing the Lens and Field Calibration  
With the VIVID 9i, three types of lens are available: WIDE lens, MIDDLE lens and TELE lens. Be sure to  
mount the lens that best matches the subject’s size and distance. Before taking scans of a subject, most suitable  
lens for the size of the object and distance must be used.  
Note  
After changing the lens, you need to recalibrate using the field calibration system. For Field Calibration, refer to page  
32.  
Relationship between Measurement Distance and Object  
Select the most suitable lens by referring to the table given below.  
Memo  
The closer the object to the VIVID 9i, the higher the measurement accuracy is. On the other hand, influences of shades  
produced due to roughness on the surface of the object can be reduced if the object is positioned far from the VIVID 9i.  
WIDE Lens  
Measurement Distance  
500  
299  
224  
600  
359  
269  
800  
478  
359  
1000  
598  
2500  
1495  
Horizontal  
Object Size  
Vertical  
449  
1121  
Unit (mm)  
MIDDLE Lens  
Measurement Distance  
500  
165  
124  
600  
198  
148  
800  
263  
198  
1000  
329  
2500  
823  
Horizontal  
Object Size  
Vertical  
247  
618  
Unit (mm)  
TELE Lens  
Measurement Distance  
500  
93  
600  
111  
83  
800  
148  
111  
1000  
185  
2500  
463  
Horizontal  
Object Size  
Vertical  
69  
139  
347  
Unit (mm)  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replacing the Lens and Field Calibration  
Replacing Procedure  
You can use the status information to see what type of lens is currently attached. The lens type is also written  
on the lens itself.  
The explanation below assume that the power to the digitizer is ON, but note that it is also possible to change  
the lens while the power is OFF. If the power is OFF, however, you cannot use the digitizer’s viewfinder or  
control keys.  
Ref. For display of the status information, refer to page 44.  
Note  
• Only the lenses supplied as standard accessories with the VIVID 9i can be used. (They are not compatible with other  
VIVID models since they have been calibrated with the VIVID 9i only.)  
• Since the VIVID 9i is a precision instrument, do not leave it without lens when replacing the lens. Replace the lens  
as soon as possible.  
• Do not touch the lens surface with fingers.  
• Dust and scratches on the lens surface may adversely affect precision. When not using the digitizer, keep the lens  
cap on the lens. If the lens is dusty, clean it with a camera blower.  
• Dust or dirt on the lens mount surface may produce incorrect measurement data. In this case, wipe the lens mount  
surface with a clean, dry cloth and then blow it with a camera air blower etc.  
• Dust and dirt collected on the lens mount surface on the body side must also be removed with a camera air blower etc.  
• SCSI communication is not available while the lens is being replaced. Please close the Polygon Editing Tool soft-  
ware’s control dialog before starting replacement.  
[Operating Procedure]  
Remove the lens hood.  
Turn counterclockwise to align with the cutouts.  
1
Memo  
The lens hood can be fixed with its bayonet mount.  
Turn the lens fixing ring counter-clockwise to  
loose it completely.  
2
Memo  
The lens fixing ring can be fixed with its screw mount.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replacing the Lens and Field Calibration  
Hold the external edge of the fixing ring and  
pull it out.  
3
Memo  
Hold the ring with your finger located near the red mark on  
the lens. This will facilitate pulling the lens.  
Note  
Attach the lens cap to the removed lens, place it in the lens  
case and keep it in a safe place.  
The following information will appear in the viewfind-  
er as shown on the right.  
CHANGE LENS  
S/N  
2001005  
PRESS ANY KEY  
AND SCAN FIELD  
CALIBRATION SYSTEM  
Attach the desired lens.  
• Remove the lens cap.  
4
5
Red  
mark  
• Align the red mark on the lens with the one on the  
lens mount socket, and insert the lens until a click is  
heard.  
Memo  
Hold the ring with your finger located near the red mark on  
the lens. This will facilitate inserting the lens.  
Turn the lens fixing ring clockwise to secure  
the lens.  
Note  
After tightening the lens, make sure that the red mark on  
the ring is aligned with the red mark on the body.  
Tightening ring  
Red mark  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replacing the Lens and Field Calibration  
Attach the lens hood.  
Align the cutouts, and then turn clockwise.  
6
Memo  
The hood must be turned until it contacts the bayonet mount  
socket.  
Press any key.  
7
8
After several dozen seconds, “PLEASE OPEN LASER BARRIER AND PRESS ANY KEY” will be  
displayed on the viewfinder.  
Press any key.  
The viewfinder displays the monitor screen, and the digitizer is ready for use.  
Note  
Place a lens cap on the lens you removed, put it in its case, and store in a safe place.  
Use the field calibration system and the Polygon Editing Tool software to calibrate the  
system.  
9
Ref.  
For the field calibration procedure, refer to “Field Calibration” on the next page.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
The field calibration on this unit is done using the Field Calibration System.  
Periodic field calibration is recommended to maintain measurement precision.  
Note  
• Always carry out field calibration when changing lenses.  
• The Field Calibration System that corresponds to the type of lens the unit is equipped with must be used when con-  
ducting field calibration.  
• This unit and the Field Calibration System are paired, so it is not possible to correctly calibrate it when the serial  
numbers differ. Use the unit and Field Calibration System combination that have the same serial number.  
Memo  
The type of lens the unit is equipped with and the serial number of the unit can be confirmed in the Status information.  
Ref.  
For details on displaying the status information, refer to page 41.  
Field Calibration System  
The Field Calibration System is comprised of the Unit Section which contains the Field Calibration System,  
which corresponds to the type of lens on the unit, and the Frame Section, which maintains a facing position at  
a set distance from the unit.  
• Parts Nomenclature  
Handle A  
Unit Section (Field Calibration System)  
Handle B  
Equipped with field calibration systems for the  
WIDE lens type and MIDDLE/TELE lens type.  
*
Lock Lever B  
Unit Section Cover  
Unit Attachment Section  
Frame Section  
(Frame for Calibration System)  
Support  
Section  
Lock Lever A  
Main Unit Attachment  
Section  
Rubber Foot (×6): With Height Adjustment Mechanism  
The rubber feet adjustment mechanism has a structure  
*
shown in the drawing below.  
(Frame section)  
Stopper  
Lock nut  
Rubber foot block  
[Specifications]  
• Lowest height: 43mm  
• Adjustable range: Max. 10mm (with no preventive structure)  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
• Installation Space  
A secure installation location that meets the dimensional requirements as shown in the drawing below is nec-  
essary to install the Field Calibration System. Before using the Field Calibration System, a “Calibration Area”  
for stable and high quality calibration is required, in addition to the place that is necessary to position the Field  
Calibration System.  
Approx. 73 cm  
Installation Space  
Approx. 145 cm  
Approx. 46 cm  
• Calibration Area  
The “Calibration Area” is the box shaped area shown in the drawing below.  
The dimensions are:  
• 30 cm or less in front of or behind the Chart unit  
• 20 cm or less right or left to the Chart unit  
• 20 cm or less above the Chart unit  
Note  
• Prevent foreign matter from getting into the “Calibration Area”. When executing Field Calibration, do not allow  
workers into the area. It may result in incorrect calibration.  
• Be careful to block incident light from the outside. Execute Field Calibration under fluorescent light at less than  
500lx.  
• Prevent light from windows.  
Calibration Area  
In 20 cm above the Chart unit  
In 30 cm front of  
or behind the Chart unit  
In 20 cm right or left to  
the Chart unit  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Executing Field Calibration  
The Field Calibration System is made up of the Field Calibration System Section and the Frame Section. It is  
necessary to assemble it when it is to be used for the first time or after it has been temporarily disassembled to  
move the Field Calibration System, etc. In addition, it is necessary to attach the Field Calibration System to the  
main unit in order to carry out field calibration.  
Memo  
VIVID 9i/VI-9i has been calibrated at 20ºC. The allowable operating environment is 10ºC to 40ºC, but we recommend  
that you execute Field Calibration or scan the objects in environments where ambient temperature is 20ºC (68ºF).  
Assembling the Field Calibration System  
[Procedure]  
Install the Frame Section.  
Install the Frame Section in a secure location with suf-  
1
ficient space.  
Note  
For the place that is necessary to install the Frame of the  
Field Calibration System, see page 33 of this instruction  
manual. A “Calibration Area” is also required for execut-  
ing the calibration.  
Adjust the Height of Rubber Feet to Stabilize the Frame Section.  
The Frame of the Field Calibration System is designed to be used on a sufficiently wide, stable, flat sur-  
2
face. However, even if a sufficiently wide space is available, the surface may not be completely flat, re-  
sulting in one or more of the rubber feet of the frame not being in contact with the surface. To solve this  
problem, the feet have been equipped evenly with a mechanism to allow their height to be adjusted.  
Memo  
All 6 rubber feet have height adjustment mechanisms.  
Note  
• Adjust a height of the rubber feet without any item on the frame during the procedure for setting up the frame  
of this system.  
• When transporting this system, set the rubber feet to their lowest height to avoid damage to the shopping box.  
1Turn the lock nut counterclockwise to unlock the  
rubber foot block.  
Unlock the rubber  
foot block by turn-  
ing the lock nut  
counterclockwise.  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
2Turn the lock nut counterclockwise to unlock the  
rubber foot block.  
Memo  
Turn clockwise for lower heights and counterclockwise for  
higher heights.  
Adjust the height  
by turning the Rub-  
ber foot block.  
• The lock nut  
Note  
Because there is no preventive structure, moving the rub-  
ber foot block more than adjustable range may cause the  
rubber foot block to fall off. In this case, screw the rubber  
foot block onto the stopper unit clockwise.  
moves in the  
same height di-  
rection as the  
rubber foot block.  
3After adjusting the height correctly, move the lock  
nut to the stopper unit by turning it clockwise and  
lock the rubber foot block.  
After adjustment of  
the height, move  
the lock nut to the  
stopper unit by  
turning it clockwise  
and lock the rub-  
ber foot block in  
place.  
Attach the Unit Attachment Section to the  
Field Calibration System Unit Section.  
Use the 4 screws and tighten them firmly with the in-  
cluded hex wrench.  
3
Attach the VIVID 9i Attachment Section to the  
main unit.  
Make sure the footings at the base of the main unit are  
inserted into the 4 footing seats.  
4
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Field Calibration System Preparation  
There are two types of field calibration systems: one for WIDE lens type use and the other for MIDDLE/TELE  
lens type use. Use the Field Calibration System that corresponds to the type of lens the main unit is equipped  
with.  
[Procedure]  
• Preparing the Wide Lens Field Calibration System  
Open the cover of the Field Calibration Sys-  
tem Unit Section.  
1Pull Lock Lever A down to the UNLOCK posi-  
1
tion.  
2Grasp Handle A and Handle B with both hands  
and pull to spread open.  
Note  
There may be instances where the Field Calibration Sys-  
tem for the WIDE lens will slightly jut out when the Unit  
Section Cover is opened.  
When using the Field Calibration System for the WIDE  
lens, gently push on it to return the Field Calibration Sys-  
tem to the unit side.  
Pull Lock Lever A down to  
the UNLOCK position.  
Grasp firmly with  
both hands and pull  
to open.  
1
2
* There may be cases in which  
the Field Calibration System  
for the WIDE lens type will jut  
out slightly when the Unit  
Section Cover is opened, so  
gently push on it to return it  
to its position.  
• Preparing the MIDDLE/TELE Lens Field Calibration Sys-  
tem  
Open the Field Calibration System Unit for the  
WIDE lens.  
1Pull Lock Lever B down to the UNLOCK posi-  
1
tion.  
2Grasp Handle B and the Support Section of the  
Field Calibration System Section using both  
hands and pull to spread open.  
Memo  
The Field Calibration System for the MIDDLE/TELE lens  
may be opened using the procedure for opening the Field  
Calibration System unit for the WIDE lens or it may be  
opened by opening the Unit Section Cover and then opening  
the Field Calibration System for the WIDE lens.  
Pull Lock Lever B down  
to the UNLOCK posi-  
tion.  
Grasp firmly with  
both hands and pull  
to open.  
1
2
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Executing Field Calibration  
Field Calibration on this unit is carried out using the Polygon Editing Tool Software. After the assembly of the  
Field Calibration System has been completed and the unit connected to a PC and powered on, it will be neces-  
sary to start up the Polygon Editing Tool software.  
Note  
• Prevent foreign matter from getting into the “Calibration Area”. When executing Field Calibration, do not allow  
workers into the area. It may result in incorrect calibration.  
• Be careful to block incident light from the outside. Execute Field Calibration under fluorescent light at less than  
500lx.  
• Prevent light from windows.  
Ref. For details, refer to page 13, in the Polygon Editing Tool Instruction Manual.  
[Procedure]  
At the [File] menu, click [Select Digitizer].  
1
Select [VIVID9i (VI-9i)], and click [OK].  
2
Note  
If you change the selection from one digitizer to another at  
this dialog, the newly selected digitizer cannot be used until  
you start the software again.  
At the [File] menu, click [Exit] to close the software. Then start the software again.  
3
Note  
You can skip this step if “none” was originally selected when you opened the Select Digitizer dialog at Step 2  
above.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
At the [File] menu, select [Import-Digitizer], and click [One Scan] or [Step Scan].  
The software opens the [File-Import-Digitizer-One Scan] or [File-Import-Digitizer-Step Scan] dialog.  
The dialog’s work window shows a field calibration image.  
4
Memo  
There is a possibility that the image of the Field Calibration Chart is displayed out of center in the Work window  
of the dialog box or the finder of the instrument due to individual differences of the VIVID 9i/VI-9i or Field Cal-  
ibration System. It does not have an effect on the result of Field Calibration.  
Note  
If the image of the Calibration Chart isn’t displayed in the angle of the VIVID 9i/VI-9i’s finder or Work window  
of the dialog box of the Polygon Editing Tool, setting of Field Calibration Chart or the instrument might be in-  
correct. Reinstall them correctly again.  
(The following example is from the [File-Import-Digitizer-One Scan] dialog.)  
[OK]  
button  
[Calibra-  
tion] button  
Work window  
Click the [Calibration] button.  
The system carries out field calibration.  
5
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Check that field calibration was successful.  
6
If you are using field calibration results to measure the dimensions of the field calibration system, the  
software shows the differences from the measurements made at time of shipping. Differences are shown  
for four places: X1, Y1, X2, and Y2.  
• If field calibration was not successful, one of the following messages appears. Proceed in accordance  
with the displayed message.  
Message  
Meaning  
What To Do  
Field Calibration:  
Scan Error  
Failed to recognize • Remove the laser barrier and the receiving lens cap.  
the field calibration • Check that you are using the correct system type for the lens  
system.  
used.  
• Check that foreign matter is not in the “Calibration Area”.  
• Check that there is no effect of light from outside.  
• Be sure that the Field Calibration Chart unit is correctly  
mounted to the Frame section.  
• Check that the image of the Field Calibration Chart is cor-  
rectly displayed in the view angle of the finger of the VIVID  
9i/VI-9i or the Work window of the dialog box of the Poly-  
gon Editing Tool.  
Field Calibration:  
Evaluate Error  
Failed to correctly • Check that you are using the correct system type for the lens  
measure the field  
used.  
calibration system.  
• Check that foreign matter is not in the “Calibration Area”.  
• Check that there is no effect of light from outside.  
• Be sure that the Field Calibration Chart unit is correctly  
mounted to the Frame section.  
• Check that the image of the Field Calibration Chart is cor-  
rectly displayed in the view angle of the finger of the VIVID  
9i/VI-9i or the Work window of the dialog box of the Poly-  
gon Editing Tool.  
Field Calibration:  
Calibrate Error  
Adequate field cali- • Remove the lens, and then set it again correctly.  
bration results can- • Check that foreign matter is not in the “Calibration Area”.  
not be obtained.  
• Check that there is no effect of light from outside.  
• Be sure that the Field Calibration Chart unit is correctly  
mounted to the Frame section.  
• Check that the image of the Field Calibration Chart is cor-  
rectly displayed in the view angle of the finger of the VIVID  
9i/VI-9i or the Work window of the dialog box of the Poly-  
gon Editing Tool.  
• When multiple units of the instrument are used, check to  
make sure that the Field Calibration System matches the  
serial number of the instrument.  
• If there is any dirt on the laser window or the receiving lens,  
clean these as explained on page 3.  
• Clean off any dust on the field calibration system.  
• If the field calibration system is scratched or dented, contact  
a KONICA MINOLTA SENSING authorized service facil-  
ity.  
• If the same error recurs when you retry the field calibration,  
contact a KONICA MINOLTA SENSING authorized ser-  
vice facility.  
Click the [Cancel] button.  
The dialog closes, and field calibration is completed.  
7
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Ending Field Calibration  
After the field calibration is finished, remove the unit from the Field Calibration System. Also close the Field  
Calibration System that was used and return it to its original condition.  
[Procedure]  
Remove the unit from the Attachment Section  
of the main unit.  
1
Close the Field Calibration System.  
1Close the Field Calibration System  
for the MIDDLE/TELE lens type.  
2
Note  
• Be careful not to pinch your fingers in the opening and  
closing section when you shut the Field Calibration Sys-  
tem.  
• The Field Calibration System may be degraded if left  
open. Always close and store the Field Calibration Sys-  
tem after use.  
• In the case where the MIDDLE/TELE lens type Field  
Calibration System has been opened after opening the  
WIDE lens type Field Calibration System, the Field Cal-  
2Close the Field Calibration System  
for the WIDE lens type.  
ibration System for the WIDE lens type and the Unit Sec-  
tion Cover will be separated so be sure to close the Unit  
Section Cover only after closing the WIDE lens type  
Field Calibration System.  
* In the case where the MIDDLE/TELE lens type  
Field Calibration System has been opened after  
opening the WIDE lens type Field Calibration Sys-  
tem, first close the WIDE lens type Field Calibra-  
tion System and then close the unit section cover.  
The Field Calibration System can be stored as is, without disassembling it into the Field Calibration System  
Unit Section and the Frame Section.  
In the case where the system has to be disassembled and stored for reasons related to installation space require-  
ments or to move or transport it, follow the procedure outlined below to disassemble it.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Field Calibration  
Disassembling the Field Calibration System  
In cases where the Field Calibration System needs to be disassembled to move or transport it or for reasons  
related to its installation space, separate the Field Calibration System into the Field Calibration System Unit  
and the Frame sections.  
Note  
Do not disassemble any sections that are not covered in this document. Doing so may lead to a reduction in precision.  
[Procedure]  
Remove the Unit Attachment Section from the  
Field Calibration System Unit Section.  
Use the included hex wrench to remove the 4 screws.  
1
Memo  
Carefully store the screws that are removed with the Field  
Calibration System Unit Section.  
Store the Field Calibration System Unit Section and the Frame Section in an appropriate  
place.  
2
Note  
• Do not store the Frame Section in an upright position.  
• Store the Field Calibration System Section in a closed condition.  
• Even when 2 or more persons are available to move or transport the Field Calibration System, always disassemble it  
before attempting to move it.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjusting the White Balance  
The color of an object varies slightly depending on the type of light source. To acquire accurate color image  
data, adjust the white balance as described below under the light source to be used.  
When scanning, we recommend that you use a stable, flicker-free light source (such as a high-frequency fluo-  
rescent lamp).  
Memo  
The white balance has been adjusted under fluorescent light before shipment.  
[Operating Procedure]  
Set the light source so that the lighting condition is the same as that under which scan  
is to be performed.  
1
Attach the white balance cap to the VIVID 9i as  
illustrated.  
2
Press  
.
STANDARD  
COLOR OFF  
3
HQ ON NF ON  
3
4
The MENU view will appear.  
W.B. DEFAULT  
SCSI 5 ENABLED  
Press  
“WHITE BALANCE”.  
to locate the arrow cursor to  
WHITE BALANCE  
INITIAL MODE  
INITIAL CONDITION  
SCSI SETTINGS  
INFO  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjusting the White Balance  
SELECT  
ENTER  
Press  
(or  
).  
5
6
WHITE BALANCE DEFAULT  
The WHITE BALANCE view will appear.  
Press  
as necessary to select one of the  
following settings.  
<Setting>  
• DEFAULT: Adjusts for fluorescent light.  
• USER’S: Uses the previously adjusted white bal-  
ance.  
• CALIBRATION: Adjusts white balance automati-  
cally.  
ENTER  
Press  
.
7
If [CALIBRATION] was selected, the digitizer dis-  
plays the screen shown at right and automatically ad-  
justs the white balance. When adjustment is complete,  
the display returns to the Menu screen. The adjusted  
white balance is saved into USER’S.  
CALIBRATING  
If DEFAULT or USER’S was selected, adjustment is  
completed and the display returns to the Menu page.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Displaying the Status Information  
The model name, version, lens, field calibration system information can be displayed.  
[Operating Procedure]  
Press  
.
STANDARD  
COLOR OFF  
3
HQ ON NF ON  
1
2
The MENU view will appear.  
W.B. DEFAULT  
SCSI 5 ENABLED  
Press  
FO”.  
to locate the arrow cursor to “IN-  
WHITE BALANCE  
INITIAL MODE  
INITIAL CONDITION  
SCSI SETTINGS  
INFO  
SELECT  
ENTER  
Press  
(or  
).  
3
INFO  
CAMERA  
The status information will be displayed as shown be-  
low.  
MODEL NAME  
9I  
1.00  
MIDDLE  
2001005  
VERSION  
LENS  
S/N  
Item  
Explanation  
Displays information  
FIELD CALIBRATION SYSTEM  
S/N 2001005  
CAMERA  
about the VIVID 9i digi-  
tizer.  
MODEL  
NAME  
Shows the model name  
(“VIVID 9i”).  
Displays the internal soft-  
ware’s version number.  
VERSION  
Identifies the lens that is  
currently mounted.  
TELE: Telephoto lens  
MIDDLE: Middle lens  
WIDE: Wide lens  
LENS  
S/N  
Shows the serial number.  
Shows information about  
the field calibration sys-  
tem.  
FIELD CALIBRATION  
SYSTEM  
Shows the serial number of  
the field calibration sys-  
tem.  
S/N  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjusting the Laser Power and CCD Gain  
Manually  
With certain objects, correct image data cannot be obtained.  
In this case, adjust the laser power and CCD gain as explained below.  
Adjustment must be made using the Polygon Editing Tool software. For information about how to proceed,  
refer to the software’s operation manual.  
Memo  
With the following objects, correct data may not be acquired, the laser power or CCD gain must be adjusted manually.  
• Glass etc. which transmit light  
• Mirrors etc. which cause mirror reflection  
• Electric bulbs etc. which themselves illuminate  
• Black or blue cloth which has a low reflectance for red and similar colors  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setting the Scan Mode  
The setting of scan mode can be set with Polygon Editing Tool software, and the setting is the top priority. It  
is not required to set by VIVID 9i.  
If you are not making scan settings from your PC’s software (for example, if using a customized program),  
please make the settings at the digitizer itself.  
Only settings are explained here.  
Note  
These settings become effective from the time of next start.  
STANDARD  
COLOR OFF  
3
HQ ON NF ON  
W.B. DEFAULT  
SCSI 5 ENABLED  
WHITE BALANCE  
INITIAL MODE  
INITIAL CONDITION  
SCSI SETTINGS  
INFO  
Item  
Explanation  
INITIAL MODE  
Selects one of the preset scan modes.  
Selects the scan mode.  
• STANDARD: Allowable distance is 600 to 1000 mm.  
The following initial conditions cannot be changed:  
NOISE FILTER: ON; HIGH QUALITY: ON  
• EXTENDED: Allowable distance is 500 to 2500 mm.  
The following initial conditions cannot be changed:  
NOISE FILTER: ON; HIGH QUALITY: OFF  
DIGITIZE  
• USER:  
Allowable distance is 500 to 2500 mm.  
Initial conditions can be changed as desired.  
Selects the color mode for image capture.  
COLOR  
• ON:  
Full color  
No image  
• OFF:  
Makes settings that are more detailed than the initial mode settings. Available set-  
tings vary according to the selected initial mode.  
Initial conditions that cannot be changed (as determined by the DIGITIZE scan  
INITIAL CONDITION  
mode) are not displayed.  
NUMBER OF  
SCANS  
Sets number of scans. Allows for increased dynamic range.  
NOISE FILTER  
If ON, noise filtering is applied to scan data.  
If ON, unreliable scan data is discarded.  
HIGH QUALITY  
If ON, dark correction is applied to color data. Use this feature when you want to  
remove the CCD offset components and get the signal components only. This set-  
ting is not shown if INITIAL MODE-COLOR is set to OFF.>  
DARK  
LOG  
If ON, log correction is applied to color data. Use this feature when you want to  
brighten intermediate colors. This setting is not shown if INITIAL MODE-COLOR  
is set to OFF.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4  
Appendix  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error Messages  
The following error messages appear if incorrect operation is performed or an abnormality occurs with the  
VIVID 9i. If an error message appears, take the appropriate corrective actions as shown in the table below. If  
the problem still does not disappear, contact a KONICA MINOLTA SENSING authorized service facility.  
Error Message  
(Alphabetically)  
Meaning  
Action  
AF ERROR  
LOW CONTRAST  
PRESS ANY KEY  
Auto-focus failed due to low contrast Set the manual focus.  
of the object. Press any key.  
BACKUP ERROR  
PRESS ANY KEY  
Failed to back up the data.  
Press any key.  
All the settings will be initialized, so set  
them again.  
BACKUP ERROR  
PRESS ANY KEY AND SCAN values.  
Failed to back up field calibration Field calibration values have been reini-  
tialized. Please try field calibration again.  
FIELD CALIBRATION SYS-  
TEM  
Press any key and scan field calibra-  
tion system.  
CAUTION  
The laser beam is not emitted prop- Highly dangerous! Turn off the power  
erly. Turn off the power immediately. immediately and contact a KONICA  
MINOLTA SENSING authorized service  
LASER SCANNER IS  
OUT OF ORDER  
PLEASE TURN OFF  
facility.  
CHANGE LENS  
S/N ********  
PRESS ANY KEY AND SCAN tion system.  
FIELD CALIBRATION SYS-  
TEM  
The lens is now being replaced.  
Press any key and scan field calibra-  
Attach the lens hood properly.  
Check the serial No. of the lens, and  
replace the lens properly.  
LIGHT IMPROPER  
SAME FILE NAME  
Inadequacy amount of light  
Illuminate the object brighter and scan it.  
A file of the same file name already Specify another file name.  
exists.  
SYSTEM ERROR  
PLEASE TURN OFF  
A system error has occurred.  
Turn OFF the power.  
Turn OFF the power, then turn it ON  
again. If the problem still does not disap-  
pear, contact a KONICA MINOLTA  
SENSING-authorized service facility.  
TERMINATE ERROR  
PRESS ANY KEY  
SET TERMINATE ENABLE  
SCSI terminator error has occurred.  
Press any key and set the terminator  
to “ENABLE”.  
Set the terminator to “ENABLE”.  
WB ERROR  
LIGHT IMPROPER  
PRESS ANY KEY  
Failed to adjust white balance.  
Press any key.  
Adjust white balance in a bright place.  
(White balance in effect before occur-  
rence of this error has not been updated.)  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Explanation of Measuring Principle  
Measuring Principle  
The VIVID 9i uses the light-stripe method to emit a horizontal stripe light through a cylindrical lens to the ob-  
ject. The reflected light from the object is received by the CCD, and then converted by triangulation into dis-  
tance information.  
This process is repeated by scanning the stripe light vertically on the object surface using a Galvano mirror, to  
obtain a 3D image data of the object.  
In addition, a color image of the object is also obtained by scanning the CCD through a RGB filter while the  
stripe light is not emitted. (A band pass filter is used when stripe light is emitted.)  
Filter set (R, G, B, BP)/rotary type  
CCD  
Receiving lens (Exchange type)  
Laser beam  
Emitting lens  
Object  
Galvano mirror  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Explanation of Measuring Principle  
High-Speed Image Processing Circuit  
The stripe light is scanned on the CCD image plane at one horizontal line per frame, and the CCD is driven so  
that the block readout start position is shifted one line per frame.  
Approximately 600 frames are acquired.  
· CCD drive rate: 12.5MHz  
· Block readout: 125 lines  
· Data acquisition speed: 2.5 sec.  
The output signal from the CCD is then converted into a digital signal, which is then subjected to digital signal  
processing. The signal processed data is finally saved in the frame memory.  
Frame  
Memory  
Frame interline  
Transfer CCD  
A/D  
FIFO  
DSP  
Clock Generator  
Host I/F  
Galvano  
Scanner  
Laser  
Diode  
CPU  
Driver  
CPU Data Bus  
Time center of gravity and Space center of gravity  
Time center of gravity is the center of gravity of the variation in the LD light passing CCD’s one picture ele-  
ment over a period of time.  
Space center of gravity is the center of gravity of the spatial variation in the LD light on the CCD at a certain  
time.  
With this instrument, 3D images are obtained by calculating the Time center of gravity of each picture element  
of CCD. Intersection of the viewing angle of each picture element of the CCD and passing angle (time) of the  
stripe light is calculated.  
Compared to the Space center of gravity, use of Time center of gravity can reduce influences of sensitivity vari-  
ation of CCD’s picture elements and variation of brightness of the object.  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dimension Diagram  
(Unit: mm)  
221  
282.6  
289.8  
300.3  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model Name  
Measurement Method  
AF  
Non-contact 3D digitizer VIVID 9i/VI-9i  
Triangulation, light block method  
Image surface AF (contrast method), Active AF  
Light-Receiving lens (Inter-  
changeable)  
TELE  
: Focal distance f=25 mm  
MIDDLE: Focal distance f=14 mm  
WIDE : Focal distance f=8 mm  
Object Distance Range  
Scan Range  
0.6 to 2.5 m  
0.6 to 1.0 m (In Standard mode), 0.5 to 2.5 m (In Extended mode)  
Galvano mirror  
Laser Scan Method  
Laser Class  
Class 2 (IEC60825-1), Class 1 (FDA), Max. 30 mW, 690 nm  
X Direction Input Range  
(In Extended mode)  
93 to 463 mm (TELE), 165 to 823 mm (MIDDLE),  
299 to 1495 mm (WIDE)  
Y Direction Input Range  
(In Extended mode)  
69 to 347 mm (TELE), 124 to 618 mm (MIDDLE),  
224 to 1121 mm (WIDE)  
Z Direction Input Range  
(In Extended mode)  
26 to 680 mm (TELE), 42 to 1100 mm (MIDDLE),  
66 to 1750 mm (WIDE mode)  
Accuracy  
TELE  
MIDDLE XYZ: ±0.10 mm/ ±0.20 mm  
WIDE XYZ: ±0.20 mm/ ±0.40 mm  
XYZ: ±0.05 mm/ ±0.10 mm  
(Conditions:Distance 0.6 m/ 1.0 m, KONICA MINOLTA SENSING’s standard 3D  
chart, KONICA MINOLTA SENSING’s processing software, temperature 20ºC, rela-  
tive humidity 65% or less)  
Precision (Z, σ)  
Input Time (per scan)  
TELE  
MIDDLE 0.016 mm/ ±0.048 mm  
WIDE 0.032 mm/ ±0.096 mm  
(Conditions: Distance 0.6 m/ 1.0 m, KONICA MINOLTA SENSING’s standard 2D  
chart, KONICA MINOLTA SENSING’s processing software, temperature 20°C, rel-  
ative humidity 65% or less)  
0.008 mm/ ±0.024 mm  
2.5 sec.  
Transfer Time to Host Com-  
puter  
Approx. 1.5 sec.  
Ambient Lighting Condition Office environment, 500 lx or less  
Imaging Element  
3D data  
: 1/3-inch frame transfer CCD (340,000 pixels)  
Color data : Common with 3D data (color separation by rotary filter)  
x
Number of Output Pixels  
Output Format  
3D data  
Color data : 640 × 480  
: 640 × 480 (640 460 in the HIGH QUALITY mode)  
3D data: Konica Minolta format, & (STL, DXF, OBJ, ASCII points, VRML)  
(Converted to 3D data by the Polygon Editing Tool Software / standard accessory)  
Color data : RGB 24-bit raster scan data  
Data File Size  
Viewfinder  
Total 3D and color data capacity: 3.6 MB per data  
5.7-inch LCD (320 × 240 picture elements)  
* There may be rare cases in which the point that is normally lit or unlit may break up  
(dot break up); however, this will not have any influence on the scanning data.  
Output Interface  
Power  
SCSI II (DMA synchronous transfer)  
Commercial AC power, 100-240 V  
221 (W) × 412 (H) × 282 (D) mm  
Approx. 15 kg (with lens attached)  
(50-60Hz), rated current 0.6 A (at 100 VAC)  
Dimensions  
Mass  
Temperature / Humidity  
Operating environment: 10 to 40°C (50°F to 104°F); 65% RH or less/no condensa-  
tion; Installation category: II; Pollution degree: 2; Maximum altitude: 2000 m  
Storage Temperature Range  
0 to 40°C, 85% or less (at 35°C)/no condensation  
The above specifications are subject to change without prior notice.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
© 2004 KONICA MINOLTA SENSING, INC.  
AFHCKR(4) Printed in Japan  
9222-1750-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Insignia Digital Photo Frame NS DPF0712G User Manual
Invacare Hot Tub 2001 User Manual
Jaton Computer Hardware VIDEO498PCITWIN User Manual
Jensen MP3 Player SMPV 4GBEB User Manual
Johnson Controls Refrigerator S1 TCZSD01A User Manual
JVC Car Stereo System GET0056 001A User Manual
JVC Projector Accessories PK CL200U User Manual
Kalorik Water Heater USK JK 32207 User Manual
Klipsch Home Theater System KPT 801 User Manual
Korg Computer Accessories EXB PCM09 User Manual