| Owner's Operator and   Maintenance Manual   DEMAND OXYGEN DELIVERY   DEVICE AND SYSTEM   Venture 20EX, 20 PSIG Device   Venture 50EX, 50 PSIG Device   Venture Pack 211EX, 20 PSIG Shoulder Bag System   w/M6 Cylinder   Venture Pack 212EX, 20 PSIG Shoulder Bag System   w/M9 Cylinder   DEALER: THIS MANUAL MUST BE GIVEN TO THE USER OF THIS   DEMAND OXYGEN DELIVERY DEVICE.   USER: BEFORE USING THIS DEMAND OXYGEN DELIVERY   DEVICE, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.   Download from Www.Somanuals.com. All Manuals Search And Download.   TABLE OF CONTENTS   SAFETY SUMMARY ............................................. 4   UNPACKING/HANDLING ..................................... 8   INTRODUCTION ................................................ 10   Invacare Venture Demand Oxygen   Delivery Device ....................................................................................... 10   FEATURES ......................................................... 11   Venture Demand Oxygen Delivery Device   Components and Definition of Terms ................................................. 11   Venture Demand Oxygen Delivery Device System   Components and Definition of Terms ................................................. 13   SPECIFICATIONS ............................................... 14   OPERATION ...................................................... 16   The Use of Oxygen.................................................................................... 16   Set-up and Normal Pulse Operation .................................................... 16   MAINTENANCE................................................. 20   Prolonging the Life of Your Battery....................................................... 20   Care of Your Venture Demand Oxygen Delivery   Device and System.................................................................................... 21   Cleaning the Outer Case.......................................................................... 21   Carrying Case .............................................................................................. 21   ACCESSORIES AND SPARE PARTS ..................... 22   TROUBLESHOOTING.......................................... 23   OXYGEN CONSERVATION................................. 26   Battery Use Time........................................................................................ 26   Compatibility ............................................................................................... 26   LIMITED WARRANTY ........................................ 27   PART NO. 1093845   3 VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   SAFETY SUMMARY   WARNING   This device should ONLY be used under the supervision   of a physician. Verify your oxygen flowrate with your   physician when you begin using your Demand Oxygen   Delivery System.   If you have any questions about the oxygen delivery   system or its use, call your Home Health Care Dealer or   Service Representative.   If you suspect that an oxygen delivery system has been   damaged, contact your Service Representative   IMMEDIATELY. Your Service Representative MUST   handle all maintenance.   Keep all oxygen delivery systems out of the reach of   children.   Avoid creation of any spark near medical oxygen   equipment. This includes sparks from static electricity   created by any type of friction.   All electrical equipment near oxygen units MUST be   properly grounded.   This system only delivers medical grade (U.S.P.)   oxygen when used with oxygen cylinders marked as   such. When using this system with Invacare Venture   Homefill, oxygen delivery will be USP 93%+3%. Oxygen   supplied by the system is supplemental and is NOT to   be considered life-supporting.   Before using the Demand Oxygen Delivery Device, check   the contents gauge of the oxygen supply   equipment. The oxygen supply equipment MUST be filled   only by an authorized Service Representative, unless   being filled by the Invacare Venture Homefill system.   If at anytime the attendant or the end user concludes   that an insufficient amount of oxygen is being received,   VENTURE   4 PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   SAFETY SUMMARY   WARNING (CONTINUED)   the supplier and/or physician should be contacted   IMMEDIATELY. No adjustments should be made to   the flowrate unless prescribed by a physician.   DO NOT attempt to alter the switches in the battery   compartment. These switches are preset by your   Home Health Care Dealer and control the breath   delivery prescription.   RISK OF FIRE:   Oxygen does not burn. However, it will accelerate   burning of flammable materials.   TO PREVENT FIRE:   DO NOT SMOKE when operating the oxygen delivery   system. No Smoking signs should be prominently   displayed.   Keep oxygen delivery system at least 5-feet away from   ANY electrical appliances; such as radios, TVs, air   conditioners, fans, electric razors, etc.   Keep the oxygen delivery system away from open   flames or heating sources such as furnaces or stoves.   Keep flammable materials away from the oxygen   delivery system; such as oils, grease and aerosol sprays.   These will ignite and burn quickly in the presence of   concentrated oxygen.   TO PREVENT HIGH CONCENTRATIONS OF OXYGEN:   Keep the oxygen delivery system in a well ventilated   area at all times.   When the oxygen delivery system is not in use, press   the OFF button to place the Demand Oxygen Delivery   Device in the Continuous Mode, turn the Cylinder OFF,   bleed the remaining oxygen from the Regulator, turn   the Regulator OFF.   PART NO. 1093845   5 VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   SAFETY SUMMARY   WARNING (CONTINUED)   DO NOT wear the Demand Oxygen Delivery Device   under any clothing.   OTHER PRECAUTIONS:   Never attempt to lubricate the oxygen delivery system.   DO NOT insert any object (except suitable AC/DC   Adapter) in external power jack of Demand Oxygen   Delivery Device. This may cause an electrical shock   which could cause burns.   DO NOT subject the Demand Oxygen Delivery Device to   temperatures below -40°F (-40°C) or above 158°F (70°C).   To do so could damage the electronics and battery.   NEVER tamper with the oxygen delivery system.   Doing so could create a hazardous condition and   possibly cause the oxygen delivery system to malfunction.   If you have any questions, call your Home Health Care   Dealer or Service Representative.   ALWAYS connect the Demand Oxygen Delivery   Device to a pressure regulated oxygen source.   ONLY connect 20 psig Demand Oxygen Delivery   Device to a 20 psig oxygen source and a 50 psig   Demand Oxygen Delivery Device to a 50 psig oxygen   source.   DO NOT use any additional lengths of tubing between the   oxygen source and the Demand Oxygen Delivery Device.   ONLY use Invacare tubing supplied with the Demand   Oxygen Delivery Device, if any tubing is required.   DO NOT use a mask when using pulse mode delivery,   USE ONLY a standard nasal cannula.   When using the continuous flow delivery, a mask or   nasal cannula can be used.   VENTURE   6 PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   SAFETY SUMMARY   WARNING (CONTINUED)   DO NOT use a humidifier with the Demand Oxygen   Delivery Device.   DO NOT expose the Demand Oxygen Delivery Device   or system to water or any other liquids.   RADIO FREQUENCY INTERFERENCE   Most electronic equipment is influenced by Radio Frequency   Interference (RFI). CAUTION should be exercised with   regard to the use of portable communications equipment in   theareaaroundsuchequipment.   PART NO. 1093845   7 VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   UNPACKING / HANDLING   UNPACKING   1. Check for any obvious damage to the carton or its contents.   (If damage is evident, notify your local Invacare Home Health   Care Dealer/Carrier.)   2. Remove all the loose packing from the carton.   3. Carefully remove all the components from the carton.   The following is a listing of items included in the device and   system packages:   VENTURE50EX,   MODELIDD50EX   50 psig Demand Device with   Threaded Fitting Battery   Venture Operator's Manual.   VENTURE20EX,   MODELIDD20EX   20 psig Demand Device with Barb   Fitting and Connection Tubing   Battery   Venture Operator's Manual Optional   Threaded Fitting.   VENTURE PACK 211EX,   MODELIDD211EX   IDD20EX Demand Device With   Connecting Tube   AND VENTURE PACK 212EX,   MODELIDD212EX   20 psig Piston Type Flow Control   Regulator (0-6 L/min.)   Carrying Case: Dual Size Hunter   Green Shoulder Bag   Cylinder:   M6 Cylinder (IDD211EX)   M9 Cylinder (IDD212EX)   Nasal Cannula, Cylinder Wrench,   VENTURE   8 PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   HANDLING PROCEDURES   NOTE: Unless the Venture Demand Oxygen Delivery Device is to be   used immediately, retain boxes, containers and packing materials   for use in returning or storing the equipment until usage is required.   INSPECTION   1. Examine exterior of the equipment for nicks, dents,   scratches or other damages. Inspect all components.   STORAGE   1. Carefully repackage all components in the original packaging.   2. Store the repackaged equipment in a dry area.   3. DO NOT place other objects on top of the repackaged   equipment.   PART NO. 1093845   9 VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   INTRODUCTION   INVACARE VENTURE DEMAND OXYGEN   DELIVERY DEVICE   Ordinary oxygen supply units deliver oxygen continuously.   The normal breathing pattern is to inhale about one-third of the time,   and exhale two-thirds of the time. This means that continuous flow   units may be inefficient during user exhalation.   In contrast, the Venture Demand Oxygen Delivery Device (Venture   DODD) senses your inhalation and instantly releases a programmed   pulse of oxygen at a specified flowrate at the beginning of inhalation.   The specified flowrate of oxygen is delivered within the first second of   inhalation. Oxygen is not delivered during exhalation. This means that   your oxygen use time is extended by the ratio of breath duration over   the pulse duration. Venture DODD will extend user time an average of   four (4) times in the Variable duration breath mode.   Comfort is another advantage of the Venture DODD. Nasal drying caused   by continuous flow is eliminated, as oxygen is only delivered during   inhalation. Therefore, additional humidification is not necessary. Since the   oxygen is delivered at normal flowrates, there is not a high blast of oxygen   to the end user, which makes the Venture DODD more comfortable than   other pulse systems.   The Venture DODD can be used in the traditional continuous flow mode   simply by turning the Venture DODD device OFF. Effective flow deliveries of   0.5 to 6 liters per minute (L/min.) can be instantly selected, in pulse or   continuous mode, by each individual as prescribed by your physician.   During pulse mode, the Venture DODD runs on a disposable D-cell battery   which must be replaced after approximately 80 hours of normal use   (Refer to the OPERATION section in this manual for information on the   Low Battery Indicator LED). During continuous mode operation, the unit   does not need the battery to operate.   NOTE: Actual Battery use time depends on your breathing rate.   The Venture DODD can be used with compressed gas or liquid   oxygen systems. Both 20 and 50 psig units were designed for use on   20 and 50 psig oxygen delivery systems, respectively.   NOTE: The 50 psig system should ONLY be used with the threaded (DISS)   fitting, and NOT the 12-inch connecting tube, as the pressure may be too   great for the tube to stay connected.   VENTURE   10   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   FEATURES   VENTURE DEMAND OXYGEN DELIVERY   DEVICE COMPONENTS AND DEFINITION   OF TERMS (FIGURE 1)   ON/OFF Buttons - Determines flow selection method. The ON   button turns Power ON for normal pulse operation (   the OFF button disengages the electronics for continuous flow   operation ( ).   ) and   NOTE: The OFF button does NOT stop flow delivery and is NOT   recommended for unit shutdown. See SETUP AND NORMAL PULSE   OPERATION in the OPERATION Section of this manual.   Oxygen OUTLET Fitting - The delivery point of oxygen to the   end user from the Venture DODD. The fitting is barbed to   securely hold the cannula for oxygen delivery.   Oxygen INLET Fitting - The delivery point of oxygen from the   oxygen system to the Venture DODD.   External Power Input Jack - AC/DC Power Adapter connection port.   Status Indicators:   O2 - GREEN LED will illuminate with each breath indicating   that the Venture DODD is operating.   - Attention - YELLOW LED illumination indicates the   Venture DODD is NOT sensing end user's inhalation or   that there is an internal malfunction.   Refer to TROUBLESHOOTING Section.   - Low Battery - RED LED indicates two (2) low battery   levels.   O2 - Low/No Gas - GREEN LED illumination indicates   Venture DODD is NOT sensing O2 presence in pulse   delivery.   PART NO. 1093845   11   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   FEATURES   External Power   Input Jack   ç é é Oxygen OUTLET Fitting Oxygen INLET Fitting   Low O2 LED (GREEN)   Battery LED (RED)   Attention LED (YELLOW)   Pulse LED (GREEN)   êêêê   ON(Pulse ModeOperation)   è é OFF (Continuous Mode Operation)   NOTE: Battery Compartment located on   bottom back of Venture Demand Oxygen   Delivery Device.   è FIGURE 1 - VENTURE DODD FEATURES   VENTURE   12   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   FEATURES   VENTURE DODD SYSTEM COMPONENTS AND   DEFINITION OF TERMS (FIGURE 2)   Cylinder - Lightweight aluminum cylinder provided.   Regulator - Provides flowrate control of oxygen delivery with   compressed gas cylinders.   DISS Fitting - Connection point for direct connect to oxygen source.   Barb Fitting - Connection point for tubing used with compressed   gas cylinders.   Connection Tubing - Provides oxygen path from regulator to   the device inlet barb fitting for use with compressed gas cylinders.   Carrying Case with Padded Strap - Used for convenient   transportation and is adjustable to ensure comfort.   Optional AC/DC Adpater - Designed to provide AC current to device.   Barb Fitting   Regulator   DISS Fitting   M9 Cylinder   Tubing   M6 Cylinder   Case   Optional AC/DC   Adapter   (Cannula and Wrench not shown)   FIGURE 2 - VENTURE DODD SYSTEM FEATURES   PART NO. 1093845   13   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   SPECIFICATIONS   OPERATING VOLTAGES   EXTERNAL AC/DC POWER   ADAPTER:   Output: 12 + 3 VDC, 100 mA   Connector: 5.5mm OD x 2.1mm   ID x 12.0mm No. 2 Connector   ALARMS   Low Battery (Two [2] levels)   Cannula Disconnect (Breath not   sensed)   Device Malfunction   Low / No Gas Indicator   INDICATORS/SYMBOLS:   O2 LED:   Pulse Delivery   LED:   LED:   Cannula Disconnect/Missing   Breath Indicator, Audible Alarm   Battery Low Charge Indicator (1/2 to 2   hours of use time left), Audible Alarm   O2   LED:   Low or No Gas, Audible Alarm   Class BF   : BATTERY:   Approx. Battery Usage Time:   Alkaline: (1) "D" size cell   80 Hours @ 20 bpm   (Variable Breath Mode)   OXYGEN DOSE DURATION:   1-variable duration breathing:   Auto-adjusts between 0.5 and 1.0   seconds based on patient breathing rate.   3-fixed duration breath mode:   Fixed 1.0 seconds, 0.75 seconds,   0.50 seconds   FLOWRATERANGE(L/Min.):   0 to 6 L/Min.   OXYGENSOURCEPRESSURE:   Venture 20:   Venture 50:   20 + 3 psig (138 + 20 kPa)   50 + 5 psig (345 + 35 kPa)   REGULATOR   Flow Settings:   0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,   5.0, 6.0 L/Min.   Regulated Pressure:   20 psig   VENTURE   14   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   SPECIFICATIONS   ENVIRONMENTAL   Operating Temperature:   Storage Temperature:   40° to 104° F (5° to 40°C)   - 40° to 158°F (-40° to 70°C)   DEVICE DIMENSIONS   Length:   Thickness:   Width:   Weight:   5.4-inches (13.7cm)   1.6-inches (4.0cm)   3.6-inches (9.1cm)   16.0 ozs. (454gm)   PART NO. 1093845   15   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   OPERATION   THE USE OF OXYGEN   The type of oxygen administrating equipment to be used, as well as the   flowrate and time of usage, must be prescribed by a physician. Your   physician should verify your Venture DODD flowrate prescription.   WARNING   If at anytime the attendant or the end user concludes   that an insufficient amount of oxygen is being received,   the supplier and/or physician should be contacted im-   mediately. No adjustments should be made to the   flowrate unless prescribed by a physician.   NOTE: ONLY a NASAL CANNULA can be used in the PULSE MODE to   deliver oxygen. An oxygen mask or nasal cannula can be used with   Continuous Flow to deliver oxygen. Low flow cannula should not be used.   SET-UP AND NORMAL PULSE OPERATION   1. Install the battery into the Venture DODD or plug AC/DC   power adapter into external power jack.   A. To Install or Replace the Battery:   Open the battery cover on the back of the Venture   DODD by applying light inward pressure with your thumb   to the cover latch and then gently lift the cover open.   If replacing a battery: Remove the used battery from the   battery compartment by grasping the top and bottom ends   and pulling straight out.   To install the battery: Orient the new battery so that its   polarity matches the polarity marked in the battery compartment.   WARNING   DONOTattempttoaltertheswitchesinthebattery   compartment.Theseswitchesarepresetbyyourhome   healthcareproviderandcontrolthebreathdeliveryprescription.   Press the new battery in place and close the battery cover.   2. Check Oxygen Source to verify adequate oxygen supply.   VENTURE   16   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   OPERATION   3. If using an oxygen cylinder, check that the regulator is   properly attached before opening the cylinder valve.   4. Connect the Venture DODD to the regulator by performing   one (1) of the following:   A. If using an Oxygen Cylinder and Regulator:   On a 20 psig regulator, connect the tubing (supplied with   the Venture DODD) from the barb fitting of the regulator   to the barb inlet fitting on the Venture DODD.   On a 50 psig regulator, connect the threaded (DISS) fitting   (supplied with the Venture DODD) to the regulator DISS fitting.   NOTE: The 50 psig system should ONLY be used with the threaded   (DISS) fitting and not the 12-inch connecting tube, as the pressure may   be too great for the tube to stay connected.   B. If using a Liquid Oxygen Reservoir:   Directly connect the Venture DODD to the oxygen   source using a DISS fitting.   5. Connect the nasal cannula tubing to the barbed oxygen outlet   fitting on the Venture DODD and adjust cannula for comfort.   WARNING   When using nasal or masked type administering   equipment, oxygen or air tubing MUST be routed and   secured properly to ensure that the tubing does NOT   become entangled, kinked or severed.   NOTE: Use ONLY a NASAL CANNULA with PULSE MODE   delivery, NOT a mask. A mask may NOT fit your face well enough for the   unit to sense your inhalation.   NOTE: A special nasal cannula is NOT required. Any standard flow   crush-proof nasal cannula can be used.   NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.   NOTE: The cannula tubing can be up to 50 feet long to function properly   in the pulse mode. The continuous flow mode is NOT affected by the   length of the canula tubing.   PART NO. 1093845   17   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   OPERATION   6. Set the prescribed flow rate on the oxygen source by performing   one (1) of the following:   A. If using an Oxygen Cylinder and Regulator:   OPEN cylinder valve and adjust regulator to prescribed flow rate.   B. If using a Liquid Oxygen Reservoir:   Set flow control knob on reservoir to prescribed flowrate.   7. Press the ON button to engage pulse mode. The unit will   pulse each time the user inhales, the GREEN PULSE LED will   blink with each oxygen pulse delivery.   8. As battery power depletes, the unit will alarm to inform the end   user to replace the battery or switch to external power source.   A. INITIAL BATT. LED - The Red LED will FLASH and the   audible alarm will BEEP three (3) times. This indicates that   you have about two (2) hours of use time left on your battery.   B. SECOND BATT. LED - The Red LED will be SOLID and   the audible alarm will BEEP six (6) times. This indicates that you   have between 15 and 30 minutes of use time left on your battery.   NOTE: Remaining battery use times are dependent on battery type   being used. Alkaline battery type is recommended for longest life.   OFF   (Continuous Mode   Operation)   FIGURE 3 - SWITCHING TO CONTINUOUS FLOW   VENTURE   18   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   OPERATION   NOTE: Whenever the pulse mode is used, the battery is monitored.   If the SECOND BATT. LED illuminates, either replace the battery or   switch to continuous mode if oxygen supply is adequate. When in   continuous mode, the battery is NOT used.   NOTE: When in the continuous flow mode, the oxygen supply will be   used considerably faster then in the pulse mode. If using a portable   oxygen supply, make plans for an alternative source of oxygen. Replace   the battery as soon as possible.   9. If it is necessary to switch to continuous flow at any time,   press the OFF button (FIGURE 3).   10. If turning the system OFF for cylinder replacement or storage,   press the Venture DODD OFF button. Turn OFF the cylinder   post valve, bleed the oxygen from the regulator and turn OFF   the regulator prior to any disassembly.   PART NO. 1093845   19   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   MAINTENANCE   PROLONGING THE LIFE OF YOUR BATTERY   The Venture DODD battery is powered by a single D-cell battery.   The Alkaline battery type is recommended for longest life.   NOTE: The Alkaline battery should last for approximately 80 hours with   normal use, but will last even longer if you follow this procedure:   1. Use the battery to its fullest capacity prior to replacing. This is   accomplished by running the unit from the battery until it stops   working (approximately 15 to 30 minutes after the SECOND   BATT. indicator illuminates SOLID and the audible alarm   sounds six [6] times).   CAUTION   DO NOT store Venture DODD with the battery installed.   WARNING   DO NOT recharge, dispose of in fire, mix with different   battery type, connect improperly or disassemble   Alkaline batteries. Doing so may cause battery to   explode or leak resulting in injury.   2. Operate your unit from the AC/DC power adapter when at home.   3. Avoid exposure to extreme temperatures.   WARNING   DO NOT operate the unit in temperatures above   104°F (40°C) or below 40°F (4°C), or place near heat   sources such as radiators or fires.   Electrical shock hazard. DO NOT remove outer case   of this unit. All disassembly and maintenance of this   unit MUST BE done by an Invacare Dealer. If outer   case is removed by anyone EXCEPT an Invacare   Dealer the warranty will be voided.   VENTURE   20   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   MAINTENANCE   CARE OF YOUR VENTURE DEMAND   OXYGEN DELIVERY DEVICE AND SYSTEM   CLEANING THE OUTER CASE   WARNING   DO NOT submerge in water to clean.   DO NOT open outer case. If outer case is removed by   anyone EXCEPT an Invacare Dealer the warranty will   be voided.   CAUTION   Use no oil. Never apply waxes, polishes, or other similar   compounds to any surface of the Venture Demand   Oxygen Delivery Device, as these compounds may be   flammable.   Clean outer case with damp cloth. If necessary, a mild dish   detergent may be used. Dry outer case with lint free cloth.   CARRYING CASE   The carrying case can be hand washed in warm water. Air dry   ONLY. NO bleach. DO NOT dryclean.   PART NO. 1093845   21   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   ACCESSORIES AND SPARE PARTS   M6 Aluminum Cylinder, 164L Capacity   M9 Aluminum Cylinder, 248L Capacity   Piston Regulator, 0.5-6 L/Min., 20 psig, CGA 870   Shoulder Bag, Hunter Green   (Dual size for Cylinder diameters of 3 and 4-inches)   Venture Device Pouch   12-inch Connection Tubing   Barb Fitting, 1/4-28 UNF   7-Foot (2.1m) Nasal Cannula   AC/DC Adapter   Venture Operator's Manual   NOTE: Rechargeable batteries can be used to power Venture DODD,   however the batteries and charger are a user supply item not provided   by Invacare.   NOTE: If you have any questions regarding accessories, please call   Invacare or your Home Healthcare Dealer.   VENTURE   22   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   TROUBLESHOOTING   PROBLEM   POSSIBLE CAUSE   SOLUTIONS   Indicates Oxygen Pulse Flow   in normal pulse delivery.   O2 LED - (GREEN)   Blinking.   This is normal   operation.   Indicates Venture DODD   not sensing presence of   oxygen in pulse delivery.   O2 LED - (GREEN)   Turn unit OFF, check   cylinder pressure is above   refill limit (500 psig) and   post valve is open, check   regulator is set to correct   flowrate. If problem still   exists, call your Home   Health Care Dealer or   Service Representative   IMMEDIATELY and go to   continuous mode if   oxygen supply is adequate.   Indicates oxygen pulse   delivery, but Venture DODD   not sensing inhalation.   O2 and   (GREEN and YELLOW)   Blinking   LED -   Breathe normally. Check   cannula for kinking and   check tubing connection.   Indicates no pulse   delivery, Venture DODD   not sensing inhalation.   Breathe normally. Check   cannula for kinking and   check tubing connection.   LED - (YELLOW)   Blinking and   Intermittent Beep   Possible internal problem.   End user possibly not   receiving oxygen.   LED - (YELLOW)   Solid and Continuous   Beep   Turn unit off then on   again. If problem still   exists, call your Home   Health Care Dealer or   Service Representative   IMMEDIATELY and go to   continuous mode if   oxygen supply is adequate.   LED - (RED)   Battery low.   Replace battery as soon   as possible.   Blinking, Three (3) Beeps Approximately two (2) hours   at Beginning.   of battery use time left.   Approximately 30 minutes   of battery use time left.   LED - (RED) Solid,   Six (6) Beeps at Beginning.   Replace battery or turn   unit OFF to enable   continuous mode if   oxygen supply is adequate.   PART NO. 1093845   23   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   TROUBLESHOOTING   PROBLEM   POSSIBLE CAUSE   SOLUTIONS   Possible internal sensor   problem. End user   possibly not receiving   oxygen.   Turn unit off then on   again. If problem still   exists, call your Home   Health Care Dealer or   Service Representative   IMMEDIATELY and go to   continuous mode if   LED - (RED) Solid,   Continuous Beep.   oxygen supply is adequate.   No lights illuminate or   RED and YELLOW LEDs   on together.   Electronic failure.   Call your Home Health   Care Dealer or Service   Representative   IMMEDIATELY and go   to continuous mode if   oxygen supply is   adequate.   Check contents gauge on   oxygen source. Call your   Home Health Care Dealer   or Service Representative   IMMEDIATELY.   Oxygen supply exhausted.   Oxygen NOT being   delivered.   Cylinders - OPEN valve.   Liquid Oxygen Supply -   Make sure coupling is   FULLY engaged.   Oxygen supply NOT   turned ON.   Check all cannula   connections.   Cannula kinked or   crushed.   Turn regulator knob to   prescribed flowrate.   Regulator knob NOT   turned ON.   Call your Home Health   Care Dealer or Service   Representative   Internal Problem.   IMMEDIATELY.   External input will not   power device.   Push plug in all the way.   Connector jack not fully   inserted into unit.   Wall outlet bad or no   power.   Check that outlet power   switch is on. Check outlet   by plugging in another   appliance or lamp.   VENTURE   24   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   TROUBLESHOOTING   PROBLEM   POSSIBLE CAUSE   SOLUTIONS   External input will not   power device (Cont.).   Incorrect power adapter.   Check that power adapter   used within specification for   operating voltage of external   power supply. Call your   Home Health Care Dealer   or Service Representative   IMMEDIATELY.   Check battery   "Low" battery light   Battery life expired.   manufacturer's   comes on too soon after   battery replacement.   expiration date. See   Prolonging the Life of the   Battery in the MAINTE-   NANCE Section of this   manual. Call your Home   Health Care Dealer or   Service Representative   IMMEDIATELY.   External power adapter   plug does not go into   jack in unit.   Jack or plug damaged.   Call your Home Health   Care Dealer or Service   Representative   IMMEDIATELY.   Non-compatible adapter   plug.   Refer to SPECIFICATION   Section for external   power adapter connector   size.   Check if YELLOW and   GREEN O2 LEDS illumi-   nate. If so, see indicator   description in this section.   Call your Home Health   Care Dealer or Service   Representative   Unit pulses at times   other than inhalation.   Unit may be in backup   Pulse Mode.   IMMEDIATELY.   Verify compatibility of   model and pressure.   Call your Home Health   Care Dealer or Service   Representative   Operating pressure of   oxygen supply source too   high.   Continuous flow out of   cannula when in pulse   mode.   IMMEDIATELY.   NOTE: If any of the above problems occur and the solutions do not work,   please contact your Invacare Home Health Care Dealer for service.   PART NO. 1093845   25   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   OXYGEN CONSERVATION   OXYGEN CONSERVED AND BATTERY   USE TIME BASED ON BREATH RATE   BREATH RATE (BPM)   % OXYGENCONSERVED   Variable Breath   Fixed 1.0 sec. Breath Mode:   Fixed 0.75   Fixed 0.50 sec. Breath Mode:   10   20   30   40   Mode:   8277   70   63   75   83   Breath   91   66   Mode:   83   50   75   33   87   66   sec.   i.e. 77% Oxygen conservation over Continuous Flow achieved   with Venture DODD in Variable Breath Mode @ 20 BPM Rate.   Oxygen supply lasts 4.3-times longer. Venture DODDresponds to   individual breath rates. Therefore, use times stated are only   approximates. Actual savings depends on individual breathing   patterns.   COMPATIBILITY   Venture DODD requires using a flowrate controlled oxygen   source. Gaseous or liquid oxygen systems that meet the   following requirements are compatible to the Venture   DODD:   Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source.   Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source.   Flowrate Range: 0-6 L/Min.   VENTURE   26   PART NO. 1093845   Download from Www.Somanuals.com. All Manuals Search And Download.   LIMITED WARRANTY   PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED   TO COMPLY WITH FEDERAL LAW APPLICABLE TO   PRODUCTS MANUFACTURED AFTER JULY 4, 1975.   This warranty is extended only to the original purchaser/user of   our products.   This warranty gives you specific legal rights and you may also have   other legal rights which vary from state to state.   Invacare warrants the Demand Oxygen Delivery Device to be free   from defects in materials and workmanship for a period of one (1)   year from date of purchase. In addition, all internal electronic   components and switches carry a three (3) year warranty from   date of purchase. If within such warranty period any such product   shall be proven to Invacare satisfaction to be defective, product   shall be repaired or replaced, at Invacare option. This warranty   does not include normal wear and tear or shipping charges   incurred in replacement part installation or repair of such   product. Invacare sole obligation and your exclusive remedy   under this warranty shall be limited to such repair and/or   replacement.   For warranty service, please contact Invacare Service Department   at the Toll Free Number listed below during normal hours.   Upon receiving notice of an alleged defect in a product, Invacare   will issue a serialized return authorization. It shall be the   responsibility of the purchaser to return, at the purchasers cost,   to either Invacares plant or service center as specified by Invacare   in advance. Defective components part(s) must be returned for   warranty inspection using the serial number for identification   within 30 days of the return authorization date. DO NOT return   products to our factory without prior consent. C.O.D. shipments   will be refused - please prepay shipping charges.   LIMITATIONSANDEXCLUSIONS:THEFOREGOING   WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PROD-   UCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DE-   FACED, PRODUCTSSUBJECTEDTONEGLIGENCE,   ACCIDENT, IMPROPEROPERATION, MAINTENANCEORSTOR-   AGE, PRODUCTSMODIFIEDWITHOUTINVACARESEXPRESS   WRITTENCONSENT(INCLUDING, BUTNOT   LIMITEDTO, MODIFICATIONTHROUGHTHEUSEOF   UNAUTHORIZEDPARTSORATTACHMENTS;PRODUCTS   DAMAGED BY REASON OF REPAIRS MADE TO ANY   COMPONENTWITHOUTTHESPECIFICCONSENTOF   INVACARE, OR TO A PRODUCT DAMAGED BY   CIRCUMSTANCESBEYONDINVACARERESPIRATORY -   PART NO. 1093845   27   VENTURE   Download from Www.Somanuals.com. All Manuals Search And Download.   LIMITED WARRANTY   CORPORATIONSCONTROL, ANDSUCHEVALUATIONWILLBE   SOLELYDETERMINEDBYINVACARE. THEWARRANTYSHALL   NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR   FAILURETOADHERETOTHEFOLLOWINGINSTRUCTIONS.   THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN   LIEU OF ANY OTHER WARRANTIES WHATSOEVER,   WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED   WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A   PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR   VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE   LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE   PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.   THE APPLICATION OF ANY IMPLIED WARRANTY   WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION   OF THE EXPRESS WARRANTY PROVIDED HEREIN.   INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL   OR INCIDENTAL DAMAGES WHATSOEVER.   SOME STATES DO NOT ALLOW THE EXCLUSION OR   LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE,   OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY   LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY   NOT APPLY TO YOU.   THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH   STATE/PROVINCIAL LAWS AND REQUIREMENTS.   Invacare Corporation   _____________________________________________________________________________________________________________________   USA   One Invacare Way   Elyria, Ohio USA   44036-2125   Invacare is a registered trademark   of Invacare Corporation.   Yes, you can. is a trademark of   Invacare Corporation.   800-333-6900   ©2002 Invacare Corporation   Canada   Part No. 1093845   Rev C 10/02   570 Matheson Blvd E Unit 8   Mississauga Ontario   L4Z 4G4 Canada   800-668-5324   Download from Www.Somanuals.com. All Manuals Search And Download.   |